1 P.O. dei Pellegrini; A.S.L. Napoli 1Centro2 Dipartimento di Oftalmologia, Seconda Università degli Studi Napoli
Scheimpflug topography , Scanning-Slit topography and
Anterior Segment Oct in patients with keratoconus
Luigi Capasso, MD1; M Lanza, MD2 ; A Tortori, MD1; C Buono, MD1;L Gifuni, MD1;F Amoroso1; N Rosa, MD2
Authors have no financial interest.
BACKGROUND
Accurate measurement of corneal thickness
is crucial •in the monitoring ectatic dystrophies •in the preoperative assessment•in the planning of surgical procedures such as intrastromal corneal ring segment implantation, corneal collagen cross-linking with riboflavin (C3-R) and deep lamellar keratoplasty.
PURPOSE
To compare the measurements of the thinnest corneal
point (TCP) with three different devices
Orbscan II Computerised Topography, Visante Optical
Coherence Tomography (OCT) and Pentacam
in patients with keratoconus (KC)
prospective, clinical study, 71 eyes of 37 consecutive patients with keratoconus.
MATERIAL AND METHODS
25 males and 12 females
Keratoconus was diagnosed if there was scissoring reflex on retinoscopy and central or paracentral steepening of the cornea on computerized topography with at least one of the following slit lamp findings of keratoconus: central or paracentral thinning, anterior bulging or conicity, Fleisher’s ring. No patient presented Descemet’s breaks or apical scar and subepithelial fibrosis.We studied a spectrum of keratoconic eyes of different severity according to the Amsler scale automatically provided by the Pentacam as follows: early (Amsler 1) 23 eyes; moderate (Amsler 2) 26 eyes; and severe (Amsler 3-4) 22 eyes.
Age range: 12 – 29 years (mean= 22.01 ± 4.53) years
MATERIAL AND METHODS
Measurements were done sequentially by two different operators using Orbscan II computerised topography (Bausch and Lomb Inc), Visante anterior segment OCT (Carl Zeiss) and Pentacam (Oculus).
The software of the three devices automatically generates a pachimetry map.
All measurements were taken at the same day time, between 1 PM and 5 PM.
RESULTSThe TCP measured with the Orbscan II and Visante OCT was 424.74 ± 54.91 μm (mean ± SD) and 461 ± 38.48 μm respectively, with a statistically significant difference (p<0.001).
RESULTSThe TCP measured with the Pentacam and Visante OCT was 461.94 ± 38.94 μm (mean ± SD) and 461 ± 38.48 μm respectively, with no statistically significant difference (p>0.1).
RESULTSThe TCP measured with the Orbscan II and Pentacam was 424.74 ± 54.91 μm (mean ± SD) and 461.94 ± 38.94 μm respectively, with a statistically significant difference (p<0.001).
Discrepancies between the three devices were present, but Pentacam and Visante OCT, showing no significant differences, seem to be more reliable in measuring the corneal thinnest point in patients with keratoconus compared to Orbscan II.
CONCLUSIONS