3
Tedea-Huntleigh www.vpgtransducers.com 1 Model 1510 Technical contact in Americas: [email protected]; Europe: [email protected]; Asia: [email protected] Document No.: 12024 Revision: 01-May-2013 FEATURES Capacity range: 100–500 kg Stainless steel construction Single-point 600 x 600 mm platform OIML R60 and NTEP approved IP68 protection Optional EEx ia IIC T6 hazardous area approval FM approval available Platform size 600 x 800 mm available APPLICATIONS Food industry platforms Marine and hybrid scales Process weighing hoppers Harsh environment DESCRIPTION Model 1510 is a high accuracy single-point load cell ideally suited to industrial applications which undergo regular washdown, typically platforms, wall scales and other process weighing applications in the food industry. Hermetically sealed against moisture, the all welded construction of the 1510 in combination with a polyurethane dual shielded cable, enables continuous operation in harsh environments while maintaining a high operating specification. The two additional sense wires feed back the voltage reaching the load cell. Complete compensation of changes in lead resistance due to temperature change and/or cable extension, is achieved by feeding this voltage into the appropriate electronics. OUTLINE DIMENSIONS in millimeters 38.1 150.0 86.5 38.1 25.4 25.4 73 38.1 25.4 12 8 Mounting Holes M8 x 1.25 Hermetically Sealed Single-Point Load Cell

1510

Embed Size (px)

DESCRIPTION

TEDEA-HUNTLEIGH HERMETICALLY SEALED SINGLE-POINT LOAD CELL

Citation preview

Tedea-Huntleigh

www.vpgtransducers.com1

Model 1510

Technical contact in Americas: [email protected]; Europe: [email protected]; Asia: [email protected]

Document No.: 12024Revision: 01-May-2013

Hermetically Sealed Single-Point Load CellFEATURES•Capacity range: 100–500 kg•Stainless steel construction•Single-point 600 x 600 mm platform•OIML R60 and NTEP approved• IP68 protection •Optional

❍❍ EEx ia IIC T6 hazardous area approval❍❍ FM approval available❍❍ Platform size 600 x 800 mm available

APPLICATIONS•Food industry platforms•Marine and hybrid scales•Process weighing hoppers•Harsh environment

DESCRIPTION Model 1510 is a high accuracy single-point load cell ideally suited to industrial applications which undergo regular washdown, typically platforms, wall scales and other process weighing applications in the food industry.

Hermetically sealed against moisture, the all welded construction of the 1510 in combination with a polyurethane dual shielded cable, enables continuous operation in harsh environments while maintaining a high operating specification.

The two additional sense wires feed back the voltage reaching the load cell.

Complete compensation of changes in lead resistance due to temperature change and/or cable extension, is achieved by feeding this voltage into the appropriate electronics.

OUTLINE DIMENSIONS in millimeters

38.1

150.0

86.5

38.1

25.4 25.4

7338

.1

25.4

12

8 Mounting Holes M8 x 1.25

Hermetically Sealed Single-Point Load Cell

Tedea-Huntleigh

www.vpgtransducers.com2

Model 1510

Technical contact in Americas: [email protected]; Europe: [email protected]; Asia: [email protected]

Document No.: 12024Revision: 01-May-2013

Hermetically Sealed Single-Point Load Cell

SPECIFICATIONSPARAMETER VALUE UNIT

Rated capacity—R.C. (Emax) 100, 250, 500 kg

NTEP/OIML accuracy class NTEP Non-approved C3* C4*

Maximum no. of intervals (n) 5000 single 1000 3000 4000

Y = Emax/Vmin 11425 1400 10000 12000 Maximum available 12500

Rated output—R.O. 2.0 mV/V

Rated output tolerance 0.2 ±% mV/V

Zero balance 0.2 ±% mV/V

Zero return, 30 min. 0.0170 0.0060 0.0170 0.0130 ±% of applied load

Total Error 0.0200 0.0300 0.0200 0.0150 ±% of rated output

Temperature effect on zero 0.0023 0.010 0.0014 0.0011 ±% of rated output/°C

Temperature effect on output 0.001 0.0040 0.0010 0.0008 ±% of applied load/°C

Eccentric loading error 0.0016 0.0035 0.0011 0.0008 ±% of rated load/cm

Temperature range, compensated –10 to +40 °C

Temperature range, safe –20 to +70 °C

Maximum safe central overload 150 % of R.C.

Ultimate central overload 300 % of R.C.

Excitation, recommended 10 VDC or VAC RMS

Excitation, maximum 15 VDC or VAC RMS

Input impedance 380±10 Ω

Output impedance 350±2 Ω

Insulation resistance >1000 MΩ

Cable length 3 m

Cable type 6-wire, braided, polyurethane, dual floating screen Standard

Construction Stainless steel

Environmental protection IP68Recommended torque 22.0 N*m

* 35% utilizationAll specifications subject to change without notice.

WIRING SCHEMATIC DIAGRAM

+VE INPUT (Green)

+VE OUTPUT (Red)

–VE INPUT (Black)–VE SENSE (Brown)–VE OUTPUT (White)

+VE SENSE (Blue)

Vishay Precision Group, Inc.

www.vpgsensors.com1

Legal Disclaimer Notice

Document No.: 63999Revision: 15-Jul-2014

DisclaimerALL PRODUCTS, PRODUCT SPECIFICATIONS AND DATA ARE SUBJECT TO CHANGE WITHOUT NOTICE.

Vishay Precision Group, Inc., its affiliates, agents, and employees, and all persons acting on its or their behalf (collectively, “VPG”), disclaim any and all liability for any errors, inaccuracies or incompleteness contained herein or in any other disclosure relating to any product.

The product specifications do not expand or otherwise modify VPG’s terms and conditions of purchase, including but not limited to, the warranty expressed therein.

VPG makes no warranty, representation or guarantee other than as set forth in the terms and conditions of purchase. To the maximum extent permitted by applicable law, VPG disclaims (i) any and all liability arising out of the application or use of any product, (ii) any and all liability, including without limitation special, consequential or incidental damages, and (iii) any and all implied warranties, including warranties of fitness for particular purpose, non-infringement and merchantability.

Information provided in datasheets and/or specifications may vary from actual results in different applications and performance may vary over time. Statements regarding the suitability of products for certain types of applications are based on VPG’s knowledge of typical requirements that are often placed on VPG products. It is the customer’s responsibility to validate that a particular product with the properties described in the product specification is suitable for use in a particular application. You should ensure you have the current version of the relevant information by contacting VPG prior to performing installation or use of the product, such as on our website at vpgsensors.com.

No license, express, implied, or otherwise, to any intellectual property rights is granted by this document, or by any conduct of VPG.

The products shown herein are not designed for use in life-saving or life-sustaining applications unless otherwise expressly indicated. Customers using or selling VPG products not expressly indicated for use in such applications do so entirely at their own risk and agree to fully indemnify VPG for any damages arising or resulting from such use or sale. Please contact authorized VPG personnel to obtain written terms and conditions regarding products designed for such applications.

Product names and markings noted herein may be trademarks of their respective owners.

Copyright Vishay Precision Group, Inc., 2014. All rights reserved.

Disclaimer

Legal Disclaimer Notice