    DM-0101 DIGITAL MULTIMETER72x72 & 96x48mm SIZES

    INTRODUCTION

    These state-of-the-art panel meters are designed foraccurate measurement of the AC RMS voltage, ACRMS current, frequency, active power and powerfactor in single-phase systems. The currenttransformer ratio can be set between 5/5A and5000/5A via front panel buttons.

    Devices have separate inputs for supply andmeasurement.

    The unit allows both manual selection of the displaychannel and automatic scrolling.

    Changing Display Selection:

    By using the MENU button, voltage, current,active power, power factor and frequency can beselected and read from display. Every depression on

    the MENU button will display the nextmeasurement.

    60 seconds after the last keypress, the display willreturn automatically to the default display mode.However the defult display mode is selectable byprogramming.

    SETTING THE CURRENTTRANSFROMER RATIO

    AND THE DEFAULT DISPLAY

    Any measured parameter can be set as the defaultdisplay. Automatic parameter scrolling is also

    provided. Selections are stored into a non-volatilememory and is not affected from power failures.

    Press MENUbuttonfor 3 sec.

    Press Upor Downbuttons untilrequested CurrentTransformer Ratio isdisplayed.

    Press MENU buttonfor saving currenttransformer ratio anddisplaying default selection.

    Press Upor Downbuttons untilrequested display modeappears.0 Auto scroll values every 5sec.1 =voltage 4= power factor2= current 5= frequency

    3= active power

    Press MENU button forsaving default selection andexit for settings.

    PRECAUTIONS FOR INSTALLATIONAND SAFE USE

    Failure to follow below instructionswill result in death or serious injury

    Do not open the unit. There is no serviceable partsinside.

    Fuses must be connected to the power supply andphase voltage inputs, in close proximity of the unit.

    Fuses must be of fast type (FF) with a maximumrating of 6A.

    Disconnect all power before working on equipment.

    When the unit is connected to the network do nottouch terminals.

    Short circuit terminals of unused current

    transformers.Do not try to clean the device with solvent or the

    like. Only clean with a dry cloth.

    Verify correct terminal connections before applyingpower.

    Electrical equipment should be serviced only bycompetent service personnel. No responsibility isassured by the manufacturer or any of itssubsidiaries for any consequences arising out ofthe use of this material.

    Only for front panel flush mounting.

    Before installation

    Read the user manual carefully, determine thecorrect connection diagram.

    Remove all connectors and mounting bracketsfrom the unit, then pass the unit through themounting opening.

    Put mounting brackets and tighten. Do not tightentoo much, this can brake the enclosure.

    Make electrical connections with connectorsremoved from sockets, then place connectors totheir sockets.

    Note that the power supply terminal is separatedfrom measurement terminals, with common neutralpoint.