DEVELOPMENT OF DETECTORS FOR PSEUDO- CHEMICAL WARFARE AGENT FORMAZAN COMPLEXES By Conor Rees 1

Dissertation Presentation

Embed Size (px)

Citation preview

Page 1: Dissertation Presentation

1

DEVELOPMENT OF DETECTORS FOR PSEUDO-CHEMICAL WARFARE AGENT FORMAZAN

COMPLEXES

By Conor Rees

Page 2: Dissertation Presentation

2

Chemical Warfare Agents • Most modern countries posses stockpiles of weaponized

agents • Increasing threats from extremist countries and groups on the

usage of agents e.g. Al-Qaeda, North Korea • Used as recently as 2013 in the Syrian capital • Counter measures need to be taken • Detectors need to be: Durable, selective, portable, cheap

Page 3: Dissertation Presentation

3

Existing Techniques: IMS / MS• Ion Mobility Spectroscopy• Mass spectroscopy • Disadvantages:

• Expensive • Training • Moderately large

Page 4: Dissertation Presentation

4

Existing Techniques: Chemical tests • Test papers

• Disadvantages• Easily contaminated• Subject to false positives• Identify by human eye

• Enzymes • Nerve agent immobilized-enzyme alarm detector (NAIAD) • Disadvantages

• Poor shelf life• Bulky • Storage problems

Page 5: Dissertation Presentation

5

HGV (Highly selective, Greatly enhanced responses & times, Very easy to use)• Developed by Dr. Ian Fallis Cardiff • Very fast responses • Small and easily portable• Swabs are already in use in airports etc. • Would be easy to integrate

Page 6: Dissertation Presentation

6

Need For Project • Colour change in swabs are very intense• However very hard for humans to see slight colour change over time • Swabs do not give an indication if antidote is needed• Detector needed to inform user on impending danger • Do not want to inject antidote if no presence of warfare agent, as can lead

to seizures and other side effects

N. B. Munro, A. P. Watson, K. R. Ambrose, and G.D. Griffin, Environmental Health Perspectives, 1990, 89, 205-215

(on left) Mark I NAAK kit

Page 7: Dissertation Presentation

7

Aims of Project • Begin to develop a sensor to detect significant colour

change over time • Detectors need to be:

• Cheap • Portable • Small • Simplistic • Sensitive

• 2 methods approached:• Image analysis program

• Potentially using a smart device (e.g. mobile phone) • Spectroscopic analysis

Page 8: Dissertation Presentation

8

Spectroscopic Analysis: Experimental • Construction of spectrometer

Page 9: Dissertation Presentation

9

Calibrating Micrometer to Wavelengths • First step to develop the spectroscopic analysis was to

calibrate the micrometer • Achieved through spectroscopic analysis of coloured

filters with specific wavelengths

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

Absorbance from Orange 550nm long pass filter

Micrometer position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

Absorbance from Green 495nm long pass filter

Micrometer Position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

Page 10: Dissertation Presentation

10

Calibrating Micrometer to Wavelengths • From the previous graphs, points of intense absorbance

were found and analyzed to determine the exact wavelength compared to the spectrometer

0.6 0.65 0.7 0.75 0.8 0.85 0.9 0.95 1 1.05

Orange 550nm long pass Filter

Micrometer position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

1.15 1.2 1.25 1.3 1.35 1.4 1.45

Absorbance from Green 495nm long pass filter

Micrometer Position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

Page 11: Dissertation Presentation

11

Analysis of Formazan Complexes • Calibrated spectrometer was now able to detect

absorbance based on micrometer positions • The next step was to analyze real formazan complexes • While also analyzing the sensitivity of the spectrometer

Page 12: Dissertation Presentation

12

Analysis of Formazan Complexes • Absorption becomes difficult to observe past a dilution

factor of 2 as the signal to noise ratio decreases as the dilution factor increases

• Therefore a more sensitive analysis was proposed

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 20

0.2

0.4

0.6

0.8

1

1.2

Absorbance from Non diluted formazan complex

Micrometer Position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 20

0.050.1

0.150.2

0.250.3

0.350.4

0.45

Absorbance from formazan complex at dilu-tion factor 2

Micrometer position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 20

0.020.040.060.080.1

0.120.140.160.180.2

Absorbance from formazan complex at dilu-tion factor 4

Micrometer position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

Page 13: Dissertation Presentation

13

“Boxed” Analysis • Moves a defined box along light spectrum seen through grating • Gives a higher wavelength resolution by separating wavelengths • This removes light contamination as the box concentrates on a

particular wavelengths

Page 14: Dissertation Presentation

14

“Boxed” Analysis• Improves sensitivity of the spectrometer • Very effective in measuring absorbance of complexes • Sensitivity of spectrometer stops at a dilution factor 4

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

Non diluted formazan complex

Micrometer Position / m-6

Abso

rban

ce /

[arb

irtar

y un

it]

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

non diluted formazan complex (box analysis)

Micrometer Position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

formazan complex at dilution factor 2 (boxed analysis)

Micrometer position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2

formazan complex at dilution factor 2

Micrometer position / m-6

Abso

rban

ce /

[arb

itrar

y un

it]

Page 15: Dissertation Presentation

15

Image Analysis Program • Program utilizes the CCID chip from a camera to analyze colours of

the solutions • Exact wavelengths of the solutions cannot be found from the pixels • Approximate wavelengths are used instead, e.g. red vs non red • An image is loaded to the program where it is converted to an RGB • It is then split into 3 separate image bands of red, green and blue

Page 16: Dissertation Presentation

16

Image Analysis Program • From the three split image bands, arrays of pixel values are

created and a histogram created depending on which colour is selected

• Program used “kmeans” clustering to determine the most dominant colours in an image

• A threshold was written in to the clustering to increase the sensitivity

• Program will then inform the user if there is a danger

Page 17: Dissertation Presentation

17

Moving Forward • Moving forward with the

spectroscopic analysis:• Using a SQUIGGLE motor • Small LED as light source • Replacing heavy steel components with

plastic • Encasing spectrometer components

with a casing to improve signal to noise ratio

Page 18: Dissertation Presentation

18

Moving Forward• Image analysis program:

• Adding a “centroiding” component to the program • Finding LD50 of warfare agent, swab testing this and analyzing • Developing to make more user friendly

• Using both methods in conjunction • Conducting spectroscopic analysis• Using program to identify clusters of light intensity

Page 19: Dissertation Presentation

19

Thanks for listening• Any Questions?