Upload
frank-quitian
View
518
Download
93
Embed Size (px)
Citation preview
8/19/2019 GE Mac 400 Servicio
1/102
Service Manual
2036039-001 Revision D
GE Healthcare
MAC™ 400Resting ECG Analysis System
Version - Product Code SCT
8/19/2019 GE Mac 400 Servicio
2/102
T-2 MAC™400 2036039-001D
21 August 2008
NOTE
The information in this manual applies to MAC 400 resting ECG analysis systems with product code SCT.
Due to continuing product innovation, specifications in this manual are subject to change without notice.
MUSE, CASE, MAC, MARS, and 12SL are trademarks owned by GE Medical Systems Information Technologies,
a General Electric Company going to market as GE Healthcare. All other trademarks contained herein are the property
of their respective owners.
©2007, 2008. General Electric Company. All Rights reserved.
8/19/2019 GE Mac 400 Servicio
3/102
2036039-001D MAC™400 i
Contents
1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-1
Manual information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Revision history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Manual purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Intended audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
Contraindications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Definitions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3Safety messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Responsibility of the manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Equipment symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Service information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Equipment identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-5
Serial number format . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Serial number label (non-China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Serial number label (China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Product label (non-China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-7
Product label (China version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8CE marking information (non-Chinese only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-9
2 Equipment Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Description of the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Mechanical structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Description of the functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Power supply module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Main PCB assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
Internal interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Preparation for use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-16
The Setup menu overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Setting options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
Printing settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18
Restoring options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19
8/19/2019 GE Mac 400 Servicio
4/102
ii MAC
™
400 2036039-001D
Diagnostic tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Power on self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Device tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-20
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-1
Recommended maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Required tools and supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Inspections and cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Interior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Electrical safety tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Leakage current tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Earth leakage (AC line) current test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Enclosure (chassis) leakage current test . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6Patient leakage current tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Patient leakage current to ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Patient leakage current, mains on applied part . . . . . . . . . . . . . . . . . . . . . . . .3-8
Replacing components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
Replacing the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-10
Replacing the primary fuses in the power inlet module . . . . . . . . . . . . . . . . . . . .3-11
Opening the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Replacing the thermal printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-13
Replacing the display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-14
Replacing the PCB assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-15
Replacing the RTC battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-16Replacing the OTPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Replacing the 12V power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-18
Replacing the power inlet module along with the toroid assembly . . . . . . . . . . . .3-19
Adjustment instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-20
Functional checkout procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Tools needed . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-21
Visual inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-22
Functional checkout procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Operational checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-22
Diagnostic tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23
Electrical safety checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-23
4 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-1
Device connected to mains but does not turn on . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Green LED does not light up and device does not turn on . . . . . . . . . . . . . . . . . . .4-2
Green LED lights up and device does not turn on . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Battery only connection and device does not turn on . . . . . . . . . . . . . . . . . . . . . . .4-4
Blank display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Errors detected during self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
8/19/2019 GE Mac 400 Servicio
5/102
2036039-001D MAC™400 iii
Errors detected during printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Battery error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Printer door open . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Paper problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Paper is moving but not printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Data does not print on the upper or lower edges of paper . . . . . . . . . . . . . . . . . .4-10
Printing only baselines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-11
5 Parts List and Replacement . . . . . . . . . . . . . . . . . . . . . . . .5-1
Ordering parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Accessory kits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Patient Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Electrodes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Consumables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3Writer paper . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Service manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Operator’s guides . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Power cords . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-5
MAC 400 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Field replaceable units (FRUs) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-6
A Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . A-1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
B Electromagnetic compatibility (not applicable for China) . .
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Guidance and manufacturer’s declaration- electromagnetic emissions . . . . . . . B-2
Guidance and manufacturer’s declaration-electromagnetic immunity . . . . . . . . . B-3
Recommended separation distance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-5
8/19/2019 GE Mac 400 Servicio
6/102
iv MAC
™
400 2036039-001D
8/19/2019 GE Mac 400 Servicio
7/102
2036039-001D MAC™400 1-1
1 Introduction
8/19/2019 GE Mac 400 Servicio
8/102
1-2 MAC™400 2036039-001D
Introduction
Manual information
Revision history
The document part number and revision letter are on the bottom of each page. The
revision letter identifies the document’s update level. The revision history of this
document is summarized in the following table.
Manual purpose
This manual provides technical information for service representatives and technical
personnel so they can maintain the equipment to the assembly level. It serves as a
guide for maintenance and electrical repairs that are considered field-repairable.
Where necessary, the manual identifies additional sources of relevant information
and/or technical assistance. See the operator guides for the instructions necessary tooperate the equipment safely in accordance with its function and intended use.
Intended audience
This manual is intended for the person who maintains or troubleshoots this
equipment.
Intended use
The MAC 400 device is intended for use under the direct supervision of a licensed
healthcare practitioner. The system is intended to acquire, measure, and recordinformation from adult and pediatric populations. The basic system delivers 3-
channel ECG recordings in automatic and arrhythmia modes and 1- or 3-channel
ECG recordings in manual mode. The arrhythmia detection provides the
convenience of automatic documentation. It is not designed to provide alarms for
arrhythmia detection.
Table 1-1: Revision history, PN 2036039-001
Revision Date Comment
A 2 July 2007 Initial release of this document.
B 5 November 2007 Edited FRU part number.
C 2 January 2008 Added special content for China systems.
D 21 August 2008 Added new product label for non-China units; added IntendedUse and Contraindication; removed Equipment Symbols and
added a reference to the Operator's Guides; added a Functional
Checkout procedure.
8/19/2019 GE Mac 400 Servicio
9/102
8/19/2019 GE Mac 400 Servicio
10/102
1-4 MAC™400 2036039-001D
Introduction
Safety messages
Additional safety messages that provide appropriate operation information may be
found throughout this manual.
D NGER
EXPLOSION HAZARD — Use near flammables may cause
explosion.
Do not use in the presence of flammable anesthetics, vapors or
liquids.
W RNING
CONNECTION TO MAINS — The mains plug must be
connected to an appropriate power supply.
Operate the unit from its battery if the integrity of the protective
earth conductor is in doubt.
Responsibility of the manufacturer
GE Medical Systems Information Technologies (GE) is responsible for the effects of
safety, reliability and performance only if:
Assembly operations, extensions, readjustments, modifications or repairs are
carried out by persons authorized by GE.
The electrical installation of the relevant room complies with the requirements
of the appropriate regulations.
The equipment is used in accordance with the instructions for use.
General
Failure on the part of the responsible individual, hospital, or institution using this
equipment to implement a satisfactory maintenance schedule may cause undue
equipment failure and possible health hazards.
To ensure patient safety, use only parts and accessories manufactured or
recommended by GE.
Parts and accessories used must meet the requirements of the applicable IEC 60601
series safety standards and essential performance standards.
The use of ACCESSORY equipment not complying with the equivalent safety
requirements of this equipment may lead to a reduced level of safety. Consideration
relating to the choice should include:
Use of the accessory in the PATIENT VICINITY; and
Evidence that the safety certification of the ACCESSORY has been performed
in accordance to the appropriate IEC 60601-1 harmonized national standard.
8/19/2019 GE Mac 400 Servicio
11/102
Introduction
2036039-001D MAC™400 1-5
Equipment symbols
See the MAC ™ 400 Resting ECG Analysis System Operator’s Guides for
information about the symbols used on this product and its packaging.
Service informationRefer equipment servicing to GE authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids that warranty.
It is the user’s responsibility to report the need for service to GE or to one of their
authorized agents.
Equipment identification
025
Every GE device has a unique serial number for identification.
The serial number is formatted as shown in the following section.
NOTE
The following examples are representative only. Your product label may vary.
8/19/2019 GE Mac 400 Servicio
12/102
1-6 MAC™400 2036039-001D
Introduction
Serial number format
026
This manual applies to MAC 400 with product code SCT.
Serial number label (non-China version)
027
Table 1-4: Serial number description
Section Description
A Product Code for MAC 400 device
B Year of manufactureC Fiscal week of manufacture
D Production sequence number
E Manufacturing site code
F Miscellaneous characteristic
8/19/2019 GE Mac 400 Servicio
13/102
Introduction
2036039-001D MAC™400 1-7
Serial number label (China version)
079
Product label (non-China version)
The product label is located on the rear side of the device next to the power inlet
module.
Table 1-5: Label format (non-China version)
Section Description
A Date of manufacture in the format YYYY-MM
B Part number of product
C Product code description
D Serial number described above
E Barcode representation of serial number
Table 1-5: Label format (China version)
Section Description
A Date of manufacture in the format YYYY-MM
B Part number of product
C Production license
D Registration of product standards
E Product register number
F Serial number
G Barcode representation of serial number
AB
C
D
E
F
G
8/19/2019 GE Mac 400 Servicio
14/102
1-8 MAC™400 2036039-001D
Introduction
028
Product Label (Non-China Version)
Product label (China version)The product label is located on the rear side of the device next to the power inlet
module.
080
Product Label (China Version)
Table 1-6: Product label (Non-China Version)
Section Description
A Product name (MAC 400)
B Site of manufacture
C Rating of 12V power supply
D Dimension and rating of fuse
E See the MAC ™ 400 Resting ECG Analysis System Operator ’s
Guides for information about the symbols used on this
label.
A
B
C
D
F E
8/19/2019 GE Mac 400 Servicio
15/102
Introduction
2036039-001D MAC™400 1-9
CE marking information (non-Chinese only)
The MAC 400 device bears CE mark 0459 indicating its conformity with the provisions of the Council Directive 93/42/EEC concerning medical devices and
fulfills the essential requirements of Annex I of this directive. The product is in
radio-interference protection class A in accordance with EN 55011.The country of
manufacture can be found on the equipment labeling. The product complies with the
requirements of standard EN 60601-1-2 “Electromagnetic Compatibility - Medical
Electrical Equipment”.
EMC: immunity performance
Users should be aware of known RF sources, such as radio or TV stations and hand-
held or mobile two-way radios, and consider them when installing a medical device
or system. Adding accessories or components, or modifying the medical device orsystem may degrade the EMI performance. Consult with qualified personnel
regarding changes to the system configuration.
Configuration of instrument part number
There are four instrument part numbers associated with the MAC 400 device –
two for non-Chinese systems and two for Chinese systems. The language is
determined at the configuration centers and there are no language-specific variants.
Table 1-6: Product label (China Version)
Section Description
A Product name
B Company name
C Company address
D Rated current
E Fuse
F Type
Table 1-7: Instrument part numbers (non-China systems)
Instrument PartNumber
Description
2033547-001 MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement feature (non-Chinese only)
2033547-002 MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement and interpretation feature (non-Chinese only)
8/19/2019 GE Mac 400 Servicio
16/102
1-10 MAC™400 2036039-001D
Introduction
Table 1-7: Instrument part numbers (China systems)
Instrument Part
NumberDescription
2033547-003 MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement feature (China only )
2033547-004 MAC 400 resting ECG analysis system, 3-channel electrocardiograph for mains and battery
operation with measurement and interpretation feature (China only )
8/19/2019 GE Mac 400 Servicio
17/102
2036039-001D MAC™400 2-1
2 Equipment Overview
8/19/2019 GE Mac 400 Servicio
18/102
2-2 MAC™400 2036039-001D
Equipment Overview
General description
The MAC 400 resting ECG analysis system is a 12-lead, 3-channel system with a
2-line x 16 character display, passive patient cable and battery operation.
Top view
029
Table 2-1: Top view
Name Description
A Keyboard Press the keyboard keys to power on the system, change
operating parameters or control operations.
B Display View lead/lead group, heart rate, mode of operation, speed,
gain, and filter settings.
8/19/2019 GE Mac 400 Servicio
19/102
Equipment Overview
2036039-001D MAC™400 2-3
Side view
030
Table 2-2: Side view
Name Description
A ECG signal input connector Connect the patient cable.
8/19/2019 GE Mac 400 Servicio
20/102
2-4 MAC™400 2036039-001D
Equipment Overview
Rear view
031
Description of the device
The MAC 400 resting ECG analysis system is a portable electrocardiograph with an
integrated thermal printer and motor assembly. It is designed to acquire, process and
record ECG signals. This manual describes the version 1 of the software. A power
supply unit and battery are integrated into the device. It is designed primarily for
operating from the battery power source, and the print function only works on
battery power. The battery is a lithium-ion rechargeable battery, and charges when
the device is connected to mains.
The hardware consists of the following:
Power block
Battery detection circuit
Battery low indication circuit
Table 2-3: Rear view
Name Description
A Product label Displays product name, site of manufacture,
power supply specification, fuse specification
and relevant symbols.
B AC mains connector Connect the power cord to charge the battery.
8/19/2019 GE Mac 400 Servicio
21/102
Equipment Overview
2036039-001D MAC™400 2-5
Battery Charging Indication Circuit (Chinese systems only)
Keyboard interface
Thermal printer and motor assembly interface
Display interface
Clocks
PCB assembly CPU
Strobe compensation circuit
VCC monitor
Memory block
Over-temperature detection circuit
Thermal printer voltage VTPH control
Field-programmable gate array (FPGA)
Audible tone generator
The available functions and operation of the MAC 400 system are described in the
operator’s guides.
Mechanical structure
The major mechanical components of the MAC 400 device are the top and bottom
covers. The bottom cover is the basic element housing the following sub-assemblies:
Power inlet module with fuses
Aluminum plate with a power supply module and toroid assembly on the
bottom and the PCB assembly on the top. The display, the keyboard connector,
printer connector and patient cable connector are mounted to the PCB assembly
Battery pocket, battery cover and battery
Thermal printer and motor assembly
Paper pocket, paper lifting tape, and printer door
The top cover houses the display window, paper cutter and keyboard, which is
linked to the main PCB assembly via a flexible cable.
Description of the functionality
The following block diagram details the main PCB assembly containing all of the
circuitry for the MAC 400 resting ECG analysis system. The power line supply,
battery, thermal printer, keyboard and display are not on the main PCB assembly.
The PCB assembly is divided in to two sections—isolated and non-isolated.
The patient is directly connected to the isolated section of the device with the patient
cable. This section of the device is electrically isolated from the mains power supply
that is connected to the non-isolated section. The circuitry on the isolated side
acquires ECG data, amplifies it and converts it to digital format. Then, the line
frequency component and the muscle artifacts from the ECG data are filtered, and
opto-couplers transmit the digitized ECG data from the isolated side to the non-
isolated side. The transformer transfers power from the non-isolated side to the
isolated side.
8/19/2019 GE Mac 400 Servicio
22/102
2-6 MAC™400 2036039-001D
Equipment Overview
The circuitry on the non-isolated section processes, analyses and interprets digitized
ECG data. The non-isolated section interfaces with the keyboard, display and
thermal printer.
032
8/19/2019 GE Mac 400 Servicio
23/102
Equipment Overview
2036039-001D MAC™400 2-7
Power supply module
The power supply module is mounted on the aluminum plate on the bottom cover.
The power supply module is comprised of the power inlet and +12V power supply.
Power inletThe power inlet includes a three-pin IEC plug and fuse holder for two 2-pole
5x20mm fuse links. This system inlet has a snap fitting.
+12V power supply
The AC/DC power supply is a customized module designed for a universal range of
input voltages. The output supply is in the range of 12V +/- 2%. The specifications
of the power supply are outlined in “Table 2-4: Power supply specifications” on
page 2-7.
Battery
A customized, rechargeable Lithium-Ion battery with built-in 2S protection is
mounted to the battery pocket in the bottom cover. A dedicated battery charger is
used to charge the battery. The battery specifications are listed in “Table 2-5: Battery
specifications” on page 2-7.
The MAC 400 system can print only when the battery is connected to the device.
With a fully charged battery (8.3-8.4V), approximately 100 ECGs can be recorded in
auto mode.
NOTE
If the device cannot produce at least fifty ten-second ECG reports sequentially
on a fully charged battery, we recommend replacing the battery. For instructions
on replacing the battery, see “Replacing the battery” on page 3-10.
Table 2-4: Power supply specifications
Input voltage range 100-240V AC
Frequency 50-60Hz
Full load current 1.67A
Output voltage 12V
Line regulation 12V +/-0.5%
Load regulation 12V +/-2%
Table 2-5: Battery specifications
Rated voltage 7.2V
Battery voltage 5.4-8.4V
8/19/2019 GE Mac 400 Servicio
24/102
2-8 MAC™400 2036039-001D
Equipment Overview
Main PCB assembly
The MAC 400 system’s main PCB assembly contains all the circuitry for the device
except the power supply line, battery, thermal printer, keyboard and display. The
circuitry can be grouped by functionality.
Voltage supply and monitoring
The device requires several regulated voltages for operation of its various
components. Voltage supply and monitoring includes on/off electronics, 5V DC-DC
converter, +3.3V LDO and battery charger.
On/off electronics turn the device power on and off. The input voltage monitoring
circuit monitors the battery voltage.
A yellow LED on the keyboard indicates when it falls below 6.8V +/- 2%.
(non-Chinese systems)
A yellow LED will illuminate continuously, When battery is fully charged or
not charging. It will flash when battery voltage falls below 6.8V +/- 2%.(Chinese systems only)
If the voltage falls below 6.3V +/- 2%, the unit turns off automatically. (all
systems)
The output of the power supply is connected to an SMD2016 polyfuse resettable
1.5A fuse for over-current protection.
A Lithium-Ion charge management IC charges the battery when the device is
connected to mains. +12V is the input to the battery charger. LED DS1 and DS2
(shown in the component view in the following graphic) as outlined in “Table 2-6:
Charging conditions” on page 2-8 indicate the charging conditions.
070
Table 2-6: Charging conditions
Charger operation DS2 DS1
Charge in progress ON OFF
Charge complete OFF ON
Charge suspend, timer fault, over voltage, sleep mode, batter
absent
OFF OFF
8/19/2019 GE Mac 400 Servicio
25/102
Equipment Overview
2036039-001D MAC™400 2-9
CPU Motorola MC68832
The Motorola MC68832 (CPU) 32-bit micro controller is the core of the MAC 400
resting ECG analysis system. The CPU interacts with SRAM, PROM, EEPROM,
RTC, FPGA, thermal printer interface, keyboard interface, display interface, audible
tone and DSP (isolated side). A 32.768KHz crystal provides the reference clock
signal for the clock synthesizer of the CPU.
Memory
Memory on the MAC 400 system includes SRAM, PROM, and Serial EEPROM.
PROM stores the code for the CPU and serial EEPROM stores configuration data.
Real-time clock (RTC)
The RTC4574 has an internal 32.768KHz crystal unit. Serial communication
between CPU and RTC4574 through the I/O port of the CPU is used to exchange
time and date information between the CPU and RTC. RTC has an accuracy of +/-
one minute per month. When the device is turned off, the RTC battery powers the
RTC to keep track of the time and date. The battery is a 3V Lithium Ion coin battery
with 255mAH capacity, which can supply power to the clock for a minimum of five
years.
Audible tone (beep) generator
The audible tone generator is F/TCW 05 manufactured by DIGISOUND. The CPU
controls the beep by enabling it under the following circumstances:
The beep sounds once when the self-test is complete.
The beep sounds once every five seconds when the leads are not connected.
The beep sounds once every five seconds to indicate the printer door open,
battery and paper error conditions.
The beep sounds once for every QRS peak detected when the heart rate is
within the specified limits.
NOTE
Pressing the QRS beep key disables the audible tone. Turning off the QRS beep
in configuration settings also disables the audible tone for QRS complex.
Keyboard
The keyboard contains the following control keys: up cursor, down cursor, right
cursor, left cursor, enter, config, lead selection, copy, QRS beep, print start/stop.The on/off key is not part of the matrix.
There are two LEDs on the keyboard – one green and one yellow.
The green LED indicates that the system is connected to active mains.
(all systems)
On non-Chinese systems, the yellow LED indicates that the battery voltage is
below 6.8V +/- 2%.
8/19/2019 GE Mac 400 Servicio
26/102
2-10 MAC™400 2036039-001D
Equipment Overview
A yellow LED will illuminate continuously when battery is fully charged
or not charging. It will flash when battery voltage falls below 6.8V +/- 2%.
(Chinese systems only)
Display
The alphanumeric interface displays on a 16X2 character liquid-crystal display powered by +5V supply. The character format is 5X7 dots plus the cursor, and the
viewing angle is six o’clock. The display type is a reflective green STN.
FPGA
The CPU-thermal printer interface, CPU-display enable logic, CPU-motor driver IC
interface are all implemented in FPGA. The CLKOUT signal is the clock signal for
logic implemented and has an output frequency of 16.7MHz.
Thermal printer and motor assembly interface
The device is equipped with a Fujitsu thermal printer and motor assembly with 576
dots, thermistor for thermal printhead temperature, paper cue sensor and door platen
sensor. There is a serial data interface between the thermal printer and the FPGA.
The strobe compensation circuit modulates the strobe pulse width in accordance
with battery voltage, speed and temperature of the thermal printhead. The strobe
pulse width decides the time for which the dots are heated. The voltage supply to the
printhead and motor is cut off when the battery voltage falls below 6V and the
thermal printhead is overheated, or by the CPU control in case of error conditions.
The motor is a bipolar stepper powered by an IC driver that moves the paper at
speeds of 5, 25 or 50mm/s.
ECG recording and pre-processing
The ECG acquisition side is isolated from the non-isolated side by 8mm of
separation. There is no electrical conduction path between the isolated and non-
isolated sections, except a 68M-ohm resistor provided as a path for the leakage
current. Opto-couplers provide communication between the CPU and DSP. The
transformer provides magnetic coupling between non-isolated +5V supply and
+5V_ISO and +3.3V_ISO for power transfer. The main components of the isolated
side are:
ECG lead connector
Electrostatic discharge protection and conductive noise immunity
Probe ID circuit
Lead fail bias circuit
Right leg drive circuit
Instrumentation amplifiers
Multiplexer
ADC
Voltage reference
DSP
8/19/2019 GE Mac 400 Servicio
27/102
Equipment Overview
2036039-001D MAC™400 2-11
Serial EEPROM
Clocks
Opto-couplers
Transformer
+5V_ISO
+3.3V_ISO +1.8V
A step-up transformer is used to transfer power from the non-isolated side to the
isolated side. The voltages generated on the acquisition side are +5V_ISO,
+3.3V_ISO and +1.8V. Communication between DSP and CPU follows SPI
protocol by transmitting signals optically using Opto-couplers.
Connector J1 connects the ECG leads to the MAC 400 system. The system supports
10 leadwire ECG cables. Diodes connected in series offer electrostatic discharge
protection. A probe ID circuit detects whether the cable is connected. A lead-fail
bias circuit detects when a particular lead is not connected. The right leg drive
circuit generates the reference signal for all the leads. Instrumentation amplifiers
provide the required gain. An 8:1 multiplexer selects the eight channels one by one.
The ADC samples each sample and sends the digitized data to CPU. The signalchain in DSP includes notch filter, DC filter, low pass filter, down sampling and line
frequency filter. Serial EEPROM is used to store DSP code. A 20MHz external
crystal provides the reference clock signal to DSP.
Internal interfaces
This section describes the interface signals between the PCB assembly and external
modules.
Signal flow between 12V power supply and PCB assembly
J10 is a 4-pin horizontal snap-fit connector and connects a 12V power supply to the
PCB assembly. The mechanical structure of the connector does not allow reverse
connection of the power supply to the PCB. The connector signals are identified in
“Table 2-7: J10 connector signals” on page 2-11. Input output (I/O) is with respect
to the MAC 400 system PCB.
Table 2-7: J10 connector signals
Connector Signal Name I/O
J10-1 GND I
J10-2 GND I
J10-3 +12V I
J10-4 +12V I
8/19/2019 GE Mac 400 Servicio
28/102
2-12 MAC™400 2036039-001D
Equipment Overview
Signal flow between battery and PCB assembly
J11 is a compression connector that connects the battery to the PCB. See “Table 2-8:
J11 interface signals” on page 2-12 for details.
Signal flow between display and PCB assembly
J8 connects the display to the PCB. “Table 2-9: J8 connector signals” on page 2-12
identifies the signals involved in the interface between the display and the PCB.
Table 2-8: J11 interface signals
Connector Signal Name I/O
J11-1 VBAT + I
J11-2 VBAT + I
J11-3 VBAT - I
J11-4 VBAT - I
Table 2-9: J8 connector signals
Connector Pin Function Description Level/Polarity
J8-1 GND Ground pin of display 0V
J8-2 +5V VCC of display 5V
J8-3 CONTRAST CPU generates this signal as a PWMwave integrated by a low pass filter to DC
voltage in the range of 0-2V
0-2V
J8-4 ADR [0] This is the LSB of the address bus. It is
used to select between data and
instruction register
TTL/active high
J8-5 R/W* CPU generates R/W* signal. This signal
decides whether data is read from the
display or written into the display
TTL/active high
J8-6 DISPLAY The display enable signal generated by
CPU and further processed by FPGA
logic
3.3V/active high
J8-7 DATA [8] Part of the 16-bit bidirectional data bus
J8-8 DATA [9] Part of the 16-bit bidirectional data bus
J8-9 DATA [10] Part of the 16-bit bidirectional data bus
J8-10 DATA [11] Part of the 16-bit bidirectional data bus
8/19/2019 GE Mac 400 Servicio
29/102
Equipment Overview
2036039-001D MAC™400 2-13
Signal flow between thermal printer and PCB assembly
J7 is a positive lock 30-pin MOLEX connector, which connects the thermal printer
to the PCB. “Table 2-10: J7 connector symbols” on page 2-13 defines the J7connector symbols.
J8-11 DATA [12] Part of the 16-bit bidirectional data bus
J8-12 DATA [13] Part of the 16-bit bidirectional data bus
J8-13 DATA [14] Part of the 16-bit bidirectional data bus
J8-14 DATA [15] Part of the 16-bit bidirectional data bus
J8-15 Unused This pin is not used
J8-16 Unused This pin is not used
Table 2-9: J8 connector signals
Connector Pin Function Description Level/Polarity
Table 2-10: J7 connector symbols
Connector Pin Function I/O Description Level
J7-1 PHK I The cathode of LED in the cue mark
sensor
0-3.8V
J7-2 VSEN O The VCC of cue mark sensor and is
equal to 5V
5V
J7-3 VHE I The output of cue mark sensorconnected to the input of the
comparator, which is part of the cue
mark sensor circuit
800mV-5V
J7-4 SW I One terminal of the door sensor
input to door sensing circuit
TTL/active high
J7-5 SW I Other terminal of door sensor
connected to ground
0V
J7-6 VH O Print head voltage supply 6-8.4V
J7-7 VH O Print head voltage supply 6-8.4V
J7-8 DI O Data input to shift register TTL
J7-9 CLK O Clock input to shift register TTL, L->H
J7-10 GND Ground 0V
J7-11 GND Ground 0V
J7-12 STB 5 O Strobe signal TTL, active high
8/19/2019 GE Mac 400 Servicio
30/102
2-14 MAC™400 2036039-001D
Equipment Overview
Signal flow between PCB assembly and keyboard
J9 is an 18-pin molex vertical mount connector that connects the keyboard to the
main PCB. Reverse voltage protection is implemented mechanically. Interface
signals between the keyboard and the PCB are identified in “Table 2-11: J9
connector signals” on page 2-14.
J7-13 STB 4 O Strobe signal TTL, active high
J7-14 STB 3 O Strobe signal TTL, active high
J7-15 VDD O VCC for logic 5V
J7-16 TM I Thermistor
J7-17 STB 2 O Strobe signal TTL, active high
J7-18 STB 1 O Strobe signal TTL, active high
J7-19 AEO2 O Heater elements block select TTL, active low
J7-20 AEO1 O Heater elements block select TL, active low
J7-21 GND Ground 0V
J7-22 GND Ground 0V
J7-23 LAT O Latch signal TTL, active low
J7-24 DO I Data out signal TTL
J7-25 VH O Print head voltage supply 6-8.4V
J7-26 VH O Print head voltage supply 6-8.4V
J7-27 MT A O One end of phase A winding 6-8.4V
J7-28 MT A* O Other end of phase A winding 6-8.4V
J7-29 MT B O One end of phase B winding 6-8.4V
J7-30 MT B* O Other end of phase B winding 6-8.4V
Table 2-10: J7 connector symbols
Connector Pin Function I/O Description Level
Table 2-11: J9 connector signals
Connector Pin Function I/O Description Level
J9-1 GND Ground connection 0V
J9-2 Not connected
J9-3 Not connected
J9-4 GND Ground connection 0V
8/19/2019 GE Mac 400 Servicio
31/102
Equipment Overview
2036039-001D MAC™400 2-15
ECG lead connector
J1 is a 14-pin connector, which connects ECG leads to the MAC 400 system PCB.
The interface signals are identified in “Table 2-12: J1 interface signals” on page 2-
16.
J9-5 On/Off switch I On/Off switch independent of matrix
J9-6 ROW1 I Row 1 connection
J9-7 ROW2 I Row 2 connection
J9-8 ROW3 I Row 3 connection
J9-9 ROW5 I Row 5 connection
J9-10 COLUMN5 O Column 5 connection
J9-11 COLUMN1 O Column 1 connection
J9-12 COLUMN2 O Column 2 connection
J9-13 ROW4 I Row 4 connection
J9-14 COLUMN3 O Column 3 connection
J9-15 COLUMN4 O Column 4 connection
J9-16 GND Ground connection 0V
J9-17 Main indication LED O Connected to the green LED which
lights up when a valid mains is
connected to the device
J9-18 Battery indication
LED
O Non-Chinese systems:
Connected to the yellow LED which
lights up when the battery voltage
goes below 6.8V +/- 2%
Chinese systems:
Connected to yellow LED which will
illuminate continuously when battery
is fully charged or not charging or it
will flash when battery voltage falls
below 6.8V +/- 2%.
Table 2-11: J9 connector signals
Connector Pin Function I/O Description Level
8/19/2019 GE Mac 400 Servicio
32/102
2-16 MAC™400 2036039-001D
Equipment Overview
Preparation for use
The MAC 400 resting ECG analysis system does not require installation. However,
you may want to customize the system settings. Once the operating parameters are
set, the system saves the settings as the default parameters and activates them each
time the system is turned on.
The Setup menu overview
Use the setup menu to navigate between system settings and select the options you
want.
1. Press the configuration key to access the setup menu. The first menu
item is Language.
2. Use the up/down cursor to select a menu item.
3. Use the left/right cursor to select the desired setting.
Table 2-12: J1 interface signals
Connector pin Function
J1-1 V2
J1-2 V3
J1-3 V4
J1-4 V5
J1-5 V6
J1-6 Cable shield
J1-7 Cable shield
J1-8 GND
J1-9 RA
J1-10 LA
J1-11 LL
J1-12 V1
J1-13 Not connected
J1-14 RL
8/19/2019 GE Mac 400 Servicio
33/102
Equipment Overview
2036039-001D MAC™400 2-17
4. Press enter to confirm your selection.
5. Press the configuration key to save settings and exit the setup menu.
Setting optionsOperating parameters can be customized using the setup menu. To change or set
options, follow the steps for each menu item. Following are several of the most
common settings. For more information about setting operating parameters, see the
“MAC 400 Resting ECG Analysis System Operator’s Guide”.
To access the options menu and initiate any of the following setting selections, press
the configuration key.
Selecting a language
You can select the language for the display text and printed ECG reports.
1. Use the up/down cursor to select Language.
2. Use the right/left cursor to choose your preferred language.
3. Press enter to confirm your selection.
NOTE
When you choose Chinese or Russian, changes only affect printed reports. The
display remains in English.
Selecting the lead notations
There are two different lead notation options: AHA and IEC.
1. Use the up/down cursor to select Notation.
2. Use the right/left cursor to select AHA or IEC .
3. Press enter to confirm your selection.
Setting the date and time for printed ECG reports
To set the date:
1. Use the up/down cursor to select Date.
8/19/2019 GE Mac 400 Servicio
34/102
2-18 MAC™400 2036039-001D
Equipment Overview
2. Use the right/left cursor to set the date.
3. Press enter to confirm your selection.
To set the time:
1. Use the up/down cursor to select Time.
2. Use the right/left cursor to set the time.
3. Press enter to confirm your selection.
Selecting heart rate limit values
1. Use the up/down cursor to select HR control .
2. Use the right/left cursor to change the low limit value in increments of 5
BPM, between 30 and 120 BPM.
3. Press enter to confirm your selection.
4. Use the right/left cursor to change the high limit value (between 80 and
240 BPM).
5. Press enter to confirm your selection.
NOTE
After setting these options, you must press configuration to save the
options and exit the setup menu.
Printing settings
It is recommended that you print the current settings of the system before
performing any maintenance.
1. Press configuration to enter the setup menu.
2. Use the up/down cursor to navigate to Print .
3. Select Yes and press enter to confirm your selection and print the
settings.
8/19/2019 GE Mac 400 Servicio
35/102
Equipment Overview
2036039-001D MAC™400 2-19
Restoring options
When the main PCB assembly is replaced, you need to restore the original options.
Restoring the serial number
When the main PCB is replaced, the serial number needs to be restored.
1. Turn off the power.
2. Press and hold the enter and copy keys together and turn on the
device. Keep the key combination depressed until Enter serial number displays.
3. Use the left/right cursor keys to enter the alphanumeric serial number
and select the characters with the start/stop key. After you enter the
entire 13 digits, the display briefly goes blank and then displays Ser. No. Stored .
Restoring the interpretation option
When the main PCB is replaced, use the option activation code to restore the
original option. The option code label is located on the side of the device near the
patient cable connector.
1. Turn off the power.
2. Press and hold the lead selection key and left cursor together and
turn on the device. Keep the key combination depressed until Option code
displays.
3. Use the left/right cursor keys to enter the 12-digit numeric option code.
After you enter the entire code, the display briefly goes blank and then displays
Option Enabled .
To confirm that you have restored options, print the configuration list as described in
“Printing settings” on page 2-18.
Diagnostic tests
All messages displayed during self tests are in English. The system menu guides youthrough the functions and testing.
Power on self test
When the MAC 400 system is initially turned on, the Self-Test…Please wait screen
displays in English. The Self-test includes the following tests:
ROM test
8/19/2019 GE Mac 400 Servicio
36/102
2-20 MAC™400 2036039-001D
Equipment Overview
RAM test
DSP test
If any of these tests fail, the corresponding error code displays the message: Test
failed...4 ROM (2 RAM, 5 DSP) error .
Device testsThe device test function tests various modules of the system and is conducted
through the keyboard interface. To initiate device tests, press and hold the right
cursor and configuration key together until Device test displays. Press the right/
left cursor keys to display the previous/next test.
1. Key test (with beep test)
2. Display test
3. Motor test
4. Recording test
5. Results test
Key test
Device test KEY TEST displays. Press the enter key to initiate the key test. During
the test mode, when each key is pressed the corresponding name of that key
displays. The names are identified in the following table.
Table 2-13: Key names
Key Display message
Enter KEY_ENTER
Up cursor CUR_UP
Down cursor CUR_DOWN
Right cursor CUR_RIGHT
Left cursor CUR_LEFT
Lead selection KEY_LEAD
Copy KEY_COPY
QRS Beep KEY_BEEP (accompanied by a beep sound)Config KEY_CONFIG
Start/Stop Exits key test and displays Device test and moves to next
test
8/19/2019 GE Mac 400 Servicio
37/102
Equipment Overview
2036039-001D MAC™400 2-21
Display test
Device test DISPLAY TEST displays. To test the display:
1. Press the enter key to start the display test.
2. Press the right cursor key. Two rows of 16 black squares display.
3. Press the left cursor key to clear the display.
4. Press the enter key to exit and move to the next test.
Motor test
Device test MOTOR TEST displays. To test the speed of the motor:
1. Press enter to start the motor test.
2. Press the up/down cursor keys to select the speed: 5, 25, or 50 mm/s.
3. Press the start/stop key to begin printing short lines on the edges of the paper.
The distance between consecutive short lines indicates the selected speed.4. Press the start/stop key to stop the printing.
5. Press the enter key to move to the next test.
Recording test
Device test RECORDING TEST displays. To test the recording:
1. Press the enter key to start the recording test.
2. Press the up/down and right/left cursor keys to select the speed 5, 25, or 50
mm/s and the gain 5, 10, or 20mm/mV.
3. Press the enter key to print the two channel wave form.
4. Press the start/stop key to stop printing.
5. Press the enter key to move to the next test.
Results test
Device test RESULTS displays. Press the start/stop key to print one page results.
The printed output supplies data on the software contained in the unit (reference
number, version number, and date of preparation of the firmware), the results of the
tests carried out during power-up, and details on the connected patient lead.
To exit the device diagnostic tests, press the config key and the left cursor together.
8/19/2019 GE Mac 400 Servicio
38/102
2-22 MAC™400 2036039-001D
Equipment Overview
8/19/2019 GE Mac 400 Servicio
39/102
2036039-001D MAC™400 3-1
3 Maintenance
8/19/2019 GE Mac 400 Servicio
40/102
3-2 MAC™400 2036039-001D
Maintenance
Recommended maintenance
A regular equipment maintenance program helps prevent unnecessary equipment
and power failures and reduces possible health hazards. This chapter contains
instructions for the following recommended maintenance:
Inspecting and cleaning the device.
Electrical safety
Ground continuity test to verify all exposed metal is properly grounded.
Leakage tests to verify the equipment does not propose any health hazard.
W RNING
PROPER MAINTENANCE REQUIRED — Failure to
implement the recommended maintenance schedule may cause
equipment failure and possible health hazards.
All responsible individuals employing the use of this device
assume the responsibility for performing the recommended
maintenance unless an Equipment Maintenance Agreement exists.
Required tools and supplies
In addition to a standard set of hand tools, you need the following items:
Multimeter
Leakage current tester
Anti-static wrist strap
Inspections and cleaning
Check the device and accessories for mechanical defects, which can impair
their functioning.
Perform a functional check as detailed in “Device tests” on page 2-20.
Ensure that the labels and inscriptions on the device relating to safety are clearly
legible.
Visual inspection
Ensure that the fuses comply with the vendor’s specifications.
Check that the patient cable and lead wires are not defective or broken.
Check that the electrodes are not out of date.
Inspect cords and cables for fraying or other damage.
Check the case and display for cracks or other damage.
Inspect the keys and keyboard to make sure they have correct markings on them
and to ensure proper operation.
8/19/2019 GE Mac 400 Servicio
41/102
Maintenance
2036039-001D MAC™400 3-3
Confirm the Green LED is glowing when the power cord is connected to the AC
line and the AC switch is On.
Exterior cleaning
Clean the exterior surfaces once per month, and more frequently if needed.
Use a clean, soft cloth and water with a mild detergent. Wring excess water
from the cloth and wipe down the exterior surfaces of the device.
Dry the surfaces with a clean cloth or paper towel.
Interior cleaning
Check for dust on the interior components and clean as needed with commercially
available compressed air.
Clean the thermal printhead periodically to ensure good print quality and long
printer life. To clean the thermal printhead:
1. Use a disposable wrist strap to avoid static electricity.
2. Open the printer door and separate the platen from the printhead.
3. Wipe the glazed surface of the thermal printhead gently with cotton swabs and
ethyl alcohol.
4. When the surfaces are dry, fix the platen in place and close the printer door.
NOTE
Do not hit the thermal printhead with anything hard.
Do not use sand paper on the thermal printhead.
Do not add any unnecessary force on the thermal printhead.
Electrical safety tests
The suggested safety analysis tests comply to the international standard IEC 60601-
1. The tests should be performed with safety testing instruments with measuring
circuits calibrated according to IEC 60601-1. To ensure personal safety, refer to the
testing instrument’s user documentation.
Perform the tests under normal ambient conditions of temperature, humidity and
pressure, with line voltage.
The leakage currents correspond to 110% of rated voltage for the tested unit. If this
is not taken into account when performing leakage tests, the measured values must be calculated.
When performing the electrical tests, you can connect the ground probe of the safety
tester to the printer mounting base (A) or the aluminum insulation plate (B) as
indicated in the following graphic.
8/19/2019 GE Mac 400 Servicio
42/102
3-4 MAC™400 2036039-001D
Maintenance
071
Recommended test equipment
Safety tester for measurements according to 60601-1.
Testing connector as described for the individual tests.
GE recommended test schedule
After initial setup.
Once every 12 months thereafter as part of a regular maintenance plan.
Whenever power sources are affected by servicing.
Ground resistance test
This test determines whether the device has a power ground fault per IEC 60601-1
(Ed. 2.1), section 3, clause 18.
This test measures the ground resistance from the device power inlet ground
terminal to any exposed conductive part of the device.
Test circuit specifications: AC current source 50 Hz/60 Hz of 25A or 1.5 times
the rated current (whichever is greater) with a limited no-load output of 6V.
Per IEC 60601-1, test the device’s power inlet ground terminal. Any resistance
greater than 100 mOhm results in failure.
8/19/2019 GE Mac 400 Servicio
43/102
Maintenance
2036039-001D MAC™400 3-5
Leakage current tests
The following electrical diagram shows the Measuring Circuit [M] required for
leakage current. The reading in mV corresponds to mA (leakage current). The safety
tester instruments generally work with this Measuring Circuit [M] and the displayed
values are converted to leakage current.
033
Earth leakage (AC line) current test
078
This test measures leakage current of the device’s internal power supply. The device
has to be turned on and off, and connected to your safety-testing equipment.
Measurements should be taken under the following conditions (refer to the safety
tester diagram):
Device in “ON” state
Polarity switch NORM and RVS
S1 (neutral) closed and open
GND switch open
8/19/2019 GE Mac 400 Servicio
44/102
3-6 MAC™400 2036039-001D
Maintenance
The test has failed if the measured values are greater than:
Enclosure (chassis) leakage current test
034
This test measures leakage current from the chassis to ground during normal
condition (N.C.) and single fault conditions (S.F.C.). In all cases, the leakage current
is measured from any exposed conductive parts to ground; the device has to be
turned on and off, and connected to your safety-testing instrument.
During N.C., measurements should be taken under the following conditions (refer to
the safety tester diagram):
Device in “ON” state
Polarity switch NORM and RVS
GND switch GND closed
S1 (neutral) closed
During S.F.C., measurements should be taken under the following conditions (refer
to the following safety tester diagram):
Device in “ON” state
Polarity switch NORM and RVS
GND switch closed, S1 (neutral) open
GND switch open, S1 (neutral) closed
The test has failed if the measured values are greater than:
N.C. S.F.C.
Polarity: NORM & RVS
S1 (neutral): closed
GND: open
Polarity: NORM & RVS
S1 (neutral): open
GND: open
Total tests/combinations: 2 Total tests/combinations: 2
500 uA (IEC) 1,000 uA (IEC & UL)
300 uA (UL)
8/19/2019 GE Mac 400 Servicio
45/102
Maintenance
2036039-001D MAC™400 3-7
Patient leakage current tests
These tests measure leakage current of the device’s floating inputs
(patient connections) to ground.
Patient leakage current to ground
077
This test measures leakage current from the floating input (patient connection) to
ground during normal condition (N.C.) and single fault condition (S.F.C.). In all
cases, the leakage current is measured from the floating input to ground: The device
has to be turned on and off, and connected to your safety testing instrument.
During N.C., measurements should be taken under the following conditions (refer to
the safety tester diagram):
Device in “ON” state
Polarity switch NORM and RVS
GND switch GND closed
S1 (neutral) closed
During S.F.C., measurements should be taken under the following conditions (refer
to the safety tester diagram):
Device in “ON” state
N.C. S.F.C.
Polarity: NORM & RVS
S1 (neutral): closed
GND: closed
Polarity: NORM & RVS
S1 (neutral): closed I open
GND: open I closedTotal tests/combinations: 2 Total tests/combinations: 2
100 uA (IEC & UL) 500 uA (IEC)
300 uA (UL)
8/19/2019 GE Mac 400 Servicio
46/102
3-8 MAC™400 2036039-001D
Maintenance
Polarity switch NORM and RVS
GND switch closed, S1 (neutral) open
GND switch open, S1 (neutral) closed
The test has failed if the measured values are greater than:
Patient leakage current, mains on applied part
035
W RNING
PERSONAL SAFETY — MAINS VOLTAGE is applied to
floating input (patient connection) and patient cables/test plug
conductors and safety tester PATIN JACK.
For the protection of the person performing these tests, the
following values of resistor R may be used:
Type CF — 100 kOhm (220 to 240 V)
Disconnect safety tester from line voltage before connecting
patient cable/test plug to device.
After line voltage has been disconnected, connect the patient
cable/test plug to device.
This test performs leakage current tests during single fault conditions (S.F.C.) with
line voltage applied to the floating (patient connection) inputs.
N.C. S.F.C.
Polarity: NORM & RVS
S1 (neutral): closed
GND: closed
Polarity: NORM & RVS
S1 (neutral): closed I open
GND: open I closed
Total tests/combinations: 2 Total tests/combinations: 4
10 uA (IEC & UL) 50 uA (IEC & UL)
8/19/2019 GE Mac 400 Servicio
47/102
Maintenance
2036039-001D MAC™400 3-9
In all cases, the leakage current is measured from the floating (patient) inputs jack of
the device to ground. To setup this leakage current test:
1. Disconnect the safety tester from the line voltage
2. Ground all signal input/output connections (all SIPs/SOPs connected to ground)
3. Connect the patient cable/test plug to device
4. Connect the patient cable/test plug to the safety tester
5. Ensure that the person/tester in not touching the patient cable/test plug or
floating (patient) input or device conductive surfaces
6. Connect safety tester to line voltage —see warning on previous page.
During S.F.C, measurements should be taken under the following conditions (refer
to safety tester diagram):
Device in “ON” state
Polarity switch NORM and RVS
S1 (neutral) open
GND switch closedThe test has failed if the measured values are greater than:
After completing this test, disconnect the safety tester from line voltage before
disconnecting the patient cable/test plug and/or the device.
Replacing components
When it is determined that a component is faulty, refer to the following instructions
for replacement. Follow the safety instructions when replacing components.
W RNING
OPERATOR SAFETY — Repairing device when powered on
may cause injury.
Turn off the device before replacing components.
N.C. S.F.C.
N/A Polarity: NORM & RVS
S1 (neutral): open
GND: closed
Total tests/combinations: 0 Total tests/combinations: 2
N/A 50 uA (IEC & UL)
8/19/2019 GE Mac 400 Servicio
48/102
3-10 MAC™400 2036039-001D
Maintenance
Safety instructions
When repairing the MAC 400 device, the following considerations apply:
Only approved personnel may conduct repairs.
Before opening the device, turn off the power and disconnect the mains plug
from the power inlet module. Never connect the mains plug when the device isopen.
Before opening the device or completing any servicing, remove the battery.
Before replacing the primary fuses in the power inlet module, turn off the power
and disconnect the mains plug.
Only the original GE Healthcare components identified in the FRU list may be
used.
When replacing electronic components, implement ESD protection.
Defective or empty LI-Ion batteries should be disposed of in accordance with
the applicable legal stipulations or returned to the factory.
Replacing the battery
036
8/19/2019 GE Mac 400 Servicio
49/102
Maintenance
2036039-001D MAC™400 3-11
1. Remove the screw fastening the battery cover to the bottom cover.
2. Slide and raise the battery cover and remove the battery from its compartment.
3. Place the new battery in to the battery compartment with the battery label facing
downwards, and the battery terminals facing the battery contacts on the PCB.
Reverse voltage protection is implemented mechanically.
Replacing the primary fuses in the power inlet module
037
1. Depress the latch and slide the fuse holder from the power inlet module.
2. Remove the two fuses from the fuse holder.
3. Insert the two new fuses.
4. Secure the fuse holder back into its position.
UTION
EQUIPMENT PERFORMANCE — Use of unapproved
components may interfere with device performance.
Only use fuses indicated on the product label.
8/19/2019 GE Mac 400 Servicio
50/102
3-12 MAC™400 2036039-001D
Maintenance
Opening the device
038
1. Remove the battery as described in“Replacing the battery” on page 3-10.
2. Turn the device upside down and remove the four fastening screws on the
bottom.
3. With the keyboard facing up, open the printer door along with the platen.
4. Raise the top cover of the housing and disconnect the keyboard cable from the
J9 connector.
The display remains fixed on the PCB.
NOTE
If you need to replace the top cover assembly, dispose of the faulty (old)
one and reassemble with the new top cover assembly.
8/19/2019 GE Mac 400 Servicio
51/102
8/19/2019 GE Mac 400 Servicio
52/102
3-14 MAC™400 2036039-001D
Maintenance
Replacing the display
040
1. Open the device as described in “Opening the device” on page 3-12.
2. Disconnect the display cable connector from the J8 connector on the main PCB.
3. Remove the four screws fastening the display to the display mounting bracket.
4. Align the new display mounting holes to those on the display mounting bracket
and fasten the display with the screws.
5. Connect the display cable connector to the J8 connector on the main PCB.
8/19/2019 GE Mac 400 Servicio
53/102
Maintenance
2036039-001D MAC™400 3-15
Replacing the PCB assembly
041
NOTE
If possible, print out the current settings before replacing the PCB assembly. Forinformation about the configuration settings, see “Printing settings” on page 2-
18.
1. Open the device as described in “Opening the device” on page 3-12.
2. Disconnect the power supply connector from J10.
3. Disconnect the display cable from J8.
4. Remove the two screws fastening the display bracket assembly and remove it.
5. Disconnect the printer cable from J7.
6. Remove the four screws fastening the PCB assembly to the aluminum plate.
NOTE
Return the PCB assembly to the factory in ESD packaging.
7. Orient the new main PCB assembly, align the mounting holes with those in the
aluminum plate, and fasten the new PCB assembly to the plate with the screws.
8. Re-fasten the display and display bracket to the PCB and plate with the screws.
8/19/2019 GE Mac 400 Servicio
54/102
3-16 MAC™400 2036039-001D
Maintenance
9. Re-connect the printer cable to J7, the power supply cable to J10, the keyboard
cable to J9 and the display cable to J8.
10. Re-attach the upper section of the device to the lower section with the four
screws.
11. Replace the battery in the battery pocket, slide on the cover and re-fasten with
the screw.
12. Turn on the device and enter the settings as described in “Setting options” on
page 2-17.
13. Enter the serial number and option code (optional) as described in “Restoring
options” on page 2-19.
14. Adjust the display contrast as described in “Adjustment instructions” on page 3-
20.
15. Adjust the time and date.
16. If possible, set the configuration to the previous user settings. If not, select the
default settings.
Replacing the RTC battery
042
1. Remove the PCB assembly as instructed in steps 1-6 of “Replacing the PCB
assembly” on page 3-15.
2. With the display connector at the top and facing upwards, locate the RTC
battery BT1 on the left side of the PCB.
3. Turn the PCB assembly over and de-solder the battery contacts to remove the
battery.
NOTE
Dispose of battery according to local municipal waste guidelines.
8/19/2019 GE Mac 400 Servicio
55/102
Maintenance
2036039-001D MAC™400 3-17
4. Remove the two solder patches.
5. Solder the new battery into position on the PCB.
6. Replace the two solder patches.
7. Turn on the device and when you see the main menu, press the configuration
key.
8. Adjust the time and date.
Replacing the OTPROM
075
1. Remove the PCB assembly as instructed in steps 1-6 of “Replacing the PCB
assembly” on page 3-15.
2. With the display connector at the top and facing upwards, locate the OTPROM
in the upper center of the PCB.
3. Remove the OTPROM from the board.
4. Insert the new OTPROM and fit it into position.
8/19/2019 GE Mac 400 Servicio
56/102
3-18 MAC™400 2036039-001D
Maintenance
Replacing the 12V power supply
0432
1. Remove the main PCB as instructed in steps 1-6 of “Replacing the PCB
assembly” on page 3-15.
2. Remove the five screws fastening the aluminum plate to the bottom cover.
3. Remove the two screws on the power inlet module bracket.
4. Remove the screw fastening the ground wire to the printer bracket.
5. Turn over the aluminum plate and disconnect the yellow/green wire from the
yellow/green wire of the toroid.
6. Disconnect the white and black wires from the power inlet module.
7. Remove the four screws to unfasten the power supply from the mounting plate.
8. Align the new power supply mounting holes to the power supply mounting
plate and replace the four screws.
9. Connect the yellow/green wire of the power supply to the yellow/green wire of
the toroid.
10. Connect the black wire of the power supply to the terminal marked ‘N’ of the
power inlet module, and connect the white wire to the terminal marked ‘L’.
11. Complete steps 11-15 of “Replacing the power inlet module along with the
toroid assembly” on page 3-19.
8/19/2019 GE Mac 400 Servicio
57/102
Maintenance
2036039-001D MAC™400 3-19
Replacing the power inlet module along with the toroid assembly
044
1. Remove the main PCB assembly as instructed in steps 1-6 of “Replacing the
PCB assembly” on page 3-15.
2. Remove the power supply shield as described in steps 1-6 of “Replacing the
12V power supply” on page 3-18.
3. Disconnect the yellow/green connector from the power inlet module.
4. To remove the power inlet from the mounting bracket, depress the locking tabs
on the top and bottom and slide the inlet out.
5. Insert the new power inlet into the bracket and snap into place.
6. Replace the toroid:
a. Pry the adhesive-backed holder from the shield.
b. Place the new toroid into position on the shield, remove the plastic from the
adhesive-backed holder and place as shown in drawing.
NOTE
Ensure that the toroid is centered between the mounting holes.
8/19/2019 GE Mac 400 Servicio
58/102
3-20 MAC™400 2036039-001D
Maintenance
7. Connect the yellow/green wire of the toroid to the yellow/green wire of the
power supply.
8. Connect the yellow/green wire terminal to the power inlet terminal marked
9. Connect the white wire to the terminal marked “L” and the black wire to the
terminal marked “N” of the power supply to the power inlet module.
NOTE
Use care to ensure that the power inlet connections are properly assembled
as indicated in drawing.
10. Connect the earthing cable of the toroid supply to the thermal printer base
mounting hole.
11. Replace the shield and power supply assembly by turning it over and aligning
the screw holes.
NOTE
Before tightening the shield screws, confirm that the power supply cable
and grounding cable are properly positioned.
12. Replace the screws fastening the power inlet bracket to the lower housing.
13. Replace the screw holding the ground cable to the printer bracket.
14. Replace the screws on the shield.
15. Replace the main PCB assembly as described in “Replacing the PCB assembly”
on page 3-15.
Adjustment instructions
No adjustment of components is required for the MAC 400 device.
When the PCB assembly or display is exchanged, the contrast of the display must be
adjusted to optimum contrast. To adjust the contrast:
1. Press the configuration key.
2. Press the up/down cursor until contrast displays.
3. Use the left/right cursor to adjust the contrast.
4. When you set the desired contrast level, press the configuration key to exit the
setup menu.
Functional checkout procedureThese checkout procedures apply to all MAC 400 systems. See “Replacing
components” on page 3-9 for FRU replacement procedures.
8/19/2019 GE Mac 400 Servicio
59/102
Maintenance
2036039-001D MAC™400 3-21
Tools needed
1. Phillips screwdriver
2. Anti-static wrist strap
3. Service manual as needed for reference
FRU Replacement
FRU Description Tools Visual InspectionElectrical Safety
TestCheckout Procedure
Power Inlet module assembly 1 to 6 and 8 1,3,5 1, 2 1
Battery assembly 1, 3, 9 3 N/A 1, 5
Power supply module assembly 1 to 6 and 8 3 1, 2 1 and 8 to 12
Bracket kit assembly 1, 3 3 N/A 1
Main PCB assembly 1 to 6 and 8 3 1, 2 1 to 12
Software OTPROM assembly 1, 2, 3, 8 3 N/A 1, 11
LCD assembly with cable and
connector
1. 2. 3 3 N/A 1, 4, 9
Printer module assembly 1, 3, 8 3 N/A 1, 10
Printer door assembly 1, 3, 8 3 N/A 1, 10
Top cover assembly 1, 3, 8 3, 4, 5 N/A 1, 8, 11
Bottom cover assembly 1 to 6 and 8 3 1, 2 1 to 12
Hardware kit 1, 3 3 N/A 1
Fuse kit 3 1, 3 N/A 1
Clock battery assembly 1, 2, 3, 7, 8 3 N/A 1, 6
Axle for paper roll assembly 3 3 N/A N/A
Other Service Events
FRU Description Tools Visual InspectionElectrical Safety
TestCheckout Procedure
No parts replaced 3 3,5 N/A 1
Software update - Enable interpretation
option
3 N/A N/A 3
Post Installation Checkout 2, 3, 4, 5, 6, 8 2, 3 1, 2 1, 11
Annual Electrical Safety tests 2, 3, 4, 5, 6, 8 3 1, 2 1, 11
8/19/2019 GE Mac 400 Servicio
60/102
3-22 MAC™400 2036039-001D
Maintenance
4. Safety tester for measurements according to 60601-1
5. Testing connector as described in service manual for the individual electrical
safety tests
6. Multi-meter
7. Soldering iron
8. ECG simulator
9. MAC 400 Resting ECG Analysis System Operator’s Guides for reference
Visual inspection
Verify the following items have passed inspection.
Fuses. See “Fuse kit, PN 2034042-001” on page 5-13.
Cable/lead wires. See “Visual inspection” on page 3-2.
Electodes. See “Visual inspection” on page 3-2.
Part, cord, or cables. See “Visual inspection” on page 3-2. Case and display. See “Visual inspection” on page 3-2.
Keys and keyboard. See “Visual inspection” on page 3-2.
Green LED. See “Visual inspection” on page 3-2.
Functional checkout procedures
Perform the functional checkout procedures that are applicable to the procedure that
was performed.
Operational checks
1. Verify the power-up self-test passed. See “Power on self test” on page 2-19.
2. Verify the serial number restored successfully. See “Restoring the serial number”
on page 2-19.
3. For units with the interpretation option enabled, verify the interpretation option
restored successfully. See “Restoring the interpretation option” on page 2-19.
4. Verify the contrast adjustment was successful. See “Adjustment instructions” on
page 3-20.
5. Verify that the new battery was charged successfully. See the MAC 400 Resting
ECG Analysis System Operator’s Guides.
6. Verify the time and date were set successfully. See “Setting the date and time for
printed ECG reports” on page 2-17.
7. Verify that the last ECG was copied successfully. See the section Automatic Mode
in the MAC 400 Resting ECG Analysis System Operator’s Guides.
8/19/2019 GE Mac 400 Servicio
61/102
Maintenance
2036039-001D MAC™400 3-23
Diagnostic tests
Electrical safety checks
Verify the current leakage test results meet the requirements. Perform electrical
safety checks when indicated. All indicated electrical safety checks require a pass/
fail indication for the steps performed. Record the measurement values in your
debrief.
1. Verify the equipment does not propose any health hazard. See “Earth leakage
(AC line) current test” on page 3-5,“Patient leakage current tests” on page 3-7,
and “Enclosure (chassis) leakage current test” on page 3-6 for more
information.
2. Verify all exposed metal is properly grounded. See “Ground resistance test” on
page 3-4 for more information.
8. Verify the keyboard test was successful. See “Key test” on page 2-20 for more
information.
9. Verify the display test was successful. See “Display test” on page 2-21 for moreinformation.
10. Verify the printer motor test was successful. See “Motor test” on page 2-21 for
more information.
11. Verify the recording test was successful. See “Recording test” on page 2-21 for
more information.
12. Verify the results test was successful. See “Results test” on page 2-21 for more
information.
Electrical Safety Checks
Step Condition1 UUT - ON2 Result Leakage Current Limits
Earth Leakage Current
1 Forward Polarity NC ______ μΑ Pass/Fail 300 μΑ
2 Neutral Open, Forward Polarity SFC ______ μΑ Pass/Fail 1,000 μΑ
3 Neutral Open, Reverse Polarity SFC ______ μΑ Pass/Fail 1,000 μΑ
4 Reverse Polarity NC ______ μΑ Pass/Fail 300 μΑ
Enclosure Leakage Current
1 Forward Polarity NC ______ μΑ Pass/Fail 100 μΑ
2 Neutral Open, Forward Polarity SFC ______ μΑ Pass/Fail 300 μΑ
3 Ground Open, Forward Polarity SFC ______ μΑ Pass/Fail 300 μΑ
8/19/2019 GE Mac 400 Servicio
62/102
3-24 MAC™400 2036039-001D
Maintenance
4 Ground Open, Reverse Polarity SFC ______ μΑ Pass/Fail 300 μΑ
5 Neutral Open, Reverse Polarity SFC ______ μΑ Pass/Fail 300 μΑ
6 Reverse Polarity NC ______ μΑ Pass/Fail 100 μΑ
Patient Leakage Current To Ground
1 Forward Polarity NC ______ μΑ Pass/Fail 10 μΑ
2 Neutral Open, Forward Polarity SFC ______ μΑ Pass/Fail 50 μΑ
3 Ground Open, Forward Polarity SFC ______ μΑ Pass/Fail 50 μΑ
4 Ground Open, Reverse Polarity SFC ______ μΑ Pass/Fail 50 μΑ
5 Neutral Open, Reverse Polarity SFC ______ μΑ Pass/Fail 50 μΑ
6 Reverse Polarity NC ______ μΑ Pass/Fail 10 μΑ
Ground Resistance Resistance
1 AC mains power cord ground prong to
exposed metal surface (ground lug)
N/A ______ Ω Pass/Fail Less than 200m Ω
1 NC = Normal Condition; SFC = Single Fault Condition; N/A = Not Applicable2 UUT = Unit Under Test
Electrical Safety Checks
Step Condition1 UUT - ON2 Result Leakage Current Limits
8/19/2019 GE Mac 400 Servicio
63/102
2036039-001D MAC™400 4-1
4 Troubleshooting
8/19/2019 GE Mac 400 Servicio
64/102
4-2 MAC™400 2036039-001D
Troubleshooting
This section contains troubleshooting instructions for specific problems. When a
particular problem is observed, the service personnel should follow the sequence
described to pinpoint the area of failure.
Only approved personnel may conduct repairs. For safety information, refer to
“Safety instructions” on page 3-10.
Device connected to mains but does not turn on
There are two scenarios that might exist when the device does not power on. Refer
to the information for the one that you have observed, and follow the steps outlined
in the flow chart to pinpoint the faulty component.
Green LED does not light up and device does not turn on
When valid mains is connected to the MAC 400 device, the green LED on the
keyboard lights up. When the device is connected to mains and it does not turn on,
follow the steps in the following flowchart to pinpoint the defective component. Toreplace components, see “Opening the device” on page 3-12.
045
8/19/2019 GE Mac 400 Servicio
65/102
Troubleshooting
2036039-001D MAC™400 4-3
Green LED lights up and device does not turn on
When the device is connected to valid mains, the green LED on the keyboard lights
up. If the device is connected to mains, and the green LED lights up but the unit does
not turn on, follow the steps in the following flowchart to pinpoint the defective
component. Refer to“Opening the device” on page 3-12 when replacement is
required.
046
8/19/2019 GE Mac 400 Servicio
66/102
4-4 MAC™400 2036039-001D
Troubleshooting
Battery only connection and device does not turn on
The MAC 400 system only prints/records ECGs when a battery is installed. If the
device does not turn on, use the following flowchart to diagnose the problem.
047
8/19/2019 GE Mac 400 Servicio
67/102
Troubleshooting
2036039-001D MAC™400 4-5
Blank display
If the device is on and the audible tone (beep) sounds but the display is blank, follow
the steps in the following flowchart to pinpoint the faulty component.
048
Errors detected during self test
If any errors are detected during self-test, the error message Test failed… in addition
to the error code and brief description display.
Errors detected during printing
There are several errors that can occur when the device is printing. Refer to the error
on the display and use the troubleshooting flowcharts to determine the reason for the
error.
8/19/2019 GE Mac 400 Servicio
68/102
4-6 MAC™400 2036039-001D
Troubleshooting
Battery error
The MAC 400 device can only print when it is powered by the ba