41
European Linear Collider Workshop ECFA LC2013 BDS+MDI IPBSM Beam Size Measurement & Performance Evaluation May 29 , 2013 DESY ECFA LC 2013 Jacqueline Yan, S. Komamiya, M. Oroku, Y. Yamaguchi The University of Tokyo, Graduate School of Science T. Yamanaka, Y. Kamiya, T. Suehara The University of Tokyo, ICEPP T.Okugi, T.Terunuma, T.Tauchi, T.Naito, K.Kubo, S.Kuroda, S.Araki, J.Urakawa (KEK) 1 13/05/29

Jacqueline Yan , S. Komamiya, M. Oroku, Y. Yamaguchi

Embed Size (px)

DESCRIPTION

European Linear Collider Workshop ECFA LC2013 BDS+MDI IPBSM Beam Size Measurement & Performance Evaluation May 29 , 2013 DESY. Jacqueline Yan , S. Komamiya, M. Oroku, Y. Yamaguchi ( The University of Tokyo, Graduate School of Science ) - PowerPoint PPT Presentation

Citation preview

  • European Linear Collider Workshop ECFA LC2013BDS+MDI

    IPBSMBeam Size Measurement & Performance Evaluation

    May 29 , 2013DESYECFA LC 2013 Jacqueline Yan, S. Komamiya, M. Oroku, Y. Yamaguchi The University of Tokyo, Graduate School of ScienceT. Yamanaka, Y. Kamiya, T. Suehara The University of Tokyo, ICEPPT.Okugi, T.Terunuma, T.Tauchi, T.Naito, K.Kubo, S.Kuroda, S.Araki, J.Urakawa (KEK)*13/05/29

    ECFA LC 2013

  • Introduction

    Measurement SchemeExpected PerformanceRole in Beam Tuning13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • ATF2 Linear Collider FFS test facility@KEKRole of IPBSM (Shintake Monitor) at ATF2 IPBSM is crucial for achieving ATF2 s Goal 1 !!focus y to design 37 nm verify Local Chromaticity Correction FFSUltra-focused vertical beam size at IP !!Crucial for high luminosity

    IPBSM

    OutlineBeam Time Status Dec 2012 Spring 2013IPBSM PerformanceError studiesHardware UpgradesSummary & Goals and PlansIntroduction13/04/04ATFII Review *ATF 1.28 GeV LINAC , DR high quality e- beam with extremely small normalized vertical emittance yATF2 Goal 2: O(nm) beam trajectory stabilization

    ATFII Review

  • Compton scattered photons detected downstreamCollision of e- beamwith laser fringe upper, lower laser paths cross at IP form Interference fringesPiezouse laser interference fringes as target for e- beamOnly device able to measure y < 100 nm !!

    Crucial for ATF2 beam tuning and realization of ILCMeasurement SchemeECFA LC 2013 e- beam safely dumpedSplit into upper/lower paths phase scan by piezo stage13/05/29*

    ECFA LC 2013

  • Detector measures signal Modulation Depth M N + N -[rad][rad]ECFA LC 2013 measurable range determined by fringe pitch

    depend on crossing angle (and )N: no. of Compton photonsConvolution between e- beam profile and fringe intensityFocused Beam large MDilluted Beam small MSmall yLarge y13/05/29*

    ECFA LC 2013

  • Measures y* = 20 nm few m with < 10% resolution

    Expected Performanceselect appropriate mode according to beam focusingECFA LC 2013

    y and M for each mode

    13/05/29*

    Crossing angle1743082Fringe pitch266 nm1.03 m3.81 m15.2 mLower limit20 nm80 nm350 nm1.2 m

    Upper limit110 nm400 nm1.4 m6 m

    ECFA LC 2013

  • 174 deg. 30 deg. 2 - 8 deg Crossing angle continuously adjustable by prism 13/05/29ECFA LC 2013 *Vertical table 1.7 (H) x 1.6 (V) m

    InterferometerPhase control (piezo stage)

    path for each mode auto-stages + mirror actuators

    beam pipe Laser transported to IP optical delayhalf mirror

    ECFA LC 2013

  • transverse laser wire scanprecise position alignment by remote controlECFA LC 2013 Role of IPBSM in Beam Tuning13/05/29*beforehand . Construct & confirm laser paths, timing alignmentLongitudinalz scanAfter all preparations .

    continuously measure y using fringe scans Feed back to multi-knob tuning

    laser spot size t,laser = 15 20 m

    ECFA LC 2013

  • Beam Time Status

    13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • 12/20 : 1st success in M detection at 174 deg modeBeam time status in 2012 stable measurements of M 0.55 Feb 30 deg mode commissioned ( 1st M detection on 2/17)ECFA LC 2013 M = 0.52 0.02 (stat) y = 166.2 6.7 (stat) [nm]2 - 8 mode: clear contrast Mmeas 0.9)Prepared 174 deg mode commissioningSuppress systematic errorsHigher laser path stability / reliabilityHigh M measured at30 modeContribute with stable operation to ATF2 beam focusing / tuning study (10 x bx*, 3 x by* optics)Spring runMajor optics reform of 2012 summerWinter runLast 2 days in Dec runMeasured many times M = 0.15 0.25correspond to y 70 82 nm

    Large step towards achieving ATF2 s goal !!error studies ongoing aimed at deriving true beamsizepreliminary preliminary * IPBSM systematic errors uncorrected** under low e beam intensity ( 1E9 e / bunch)10 x x* , 1 x y* By IPBSM group@KEK13/05/29*

    ECFA LC 2013

  • measured M over continuous reiteration of linear /nonlinear@ tuning knobs @ 174 mode Beam time status in 2013 SpringECFA LC 2013 dedicated data for error studies under analysisexconsecutive 10 fringe scanspreliminary Time passed measure M vs time after all conditions optimized preliminary Stable IPBSM performance major role in beam tuning10 x bx*, 1 x by*13/05/29* 174 mode consistency scan moving towards goal of y = 37 nm :higher IPBSM precision and stability& looser current limits of normal / skew sextupoles current M 0.306 0.043 (RMS) correspond to y 65 nmBest recordfrom Okugi-sans Fri operation meeting slides

    ECFA LC 2013

  • Other studies using IPBSM13/05/29ECFA LC 2013 *Beam intensity scanothers: Test various linear / nonlinear tuning knobs IPBSM systematic error studies

    Reference Cavity scan in high region (ex: 30 deg mode) wakefield studies Check linearity of BG levels in IPBSM detector Observe steepness of intensity dependence compare with other periods to test effects of orbit tuning and / or hardware improvement for wake suppression(ex: 30 deg mode)beam intensity 5E9 / bunch

    BG level

    ECFA LC 2013

  • ex: spring 2012 : Adjust curvature of laser cavity mirrorsAim:Suppress systematic error sourcesHigher alignment precision & reproducibilityProved greatly effective in 2012 winter runECFA LC 2013 Optics reform of 2012 summerBy IPBSM group@KEKTuning of main laserAim for a more Gaussian profileby Spectra PhysicsReform laser profile and spatial coherence (adjust YAG rod & cavity mirrors)

    Exchange flash lampseeding laser tuning ( oscillation stability) 13/05/29*

    improvements details alignment precision match focal point to IPInjection position / angle into lensRe-optimize expander / reducer

    consistency , reproducibilitybefore / after mode switchingfocal point scan for all modes CW laser + reference lines on new base platesnew IP target (screen monitor)

    mode switching technique {small linear stage + mirror actuators } now: independent for each mode (before: shared rotating stages)balanced profiles suppress difference in path length & focal point

    ECFA LC 2013

  • ECFA LC 2013 Small linear stage+ mirror actuatorFirm lens holders just after injection onto vertical tableConfirm fine alignment using CW laser and transparent IP targetcheck positioning of lens, mirror, prism prismCW laser spotinside IP chamberlaser waist & crossing point13/05/29*

    ECFA LC 2013

  • Performance Evaluation #1: Stability

    Signal jitter sourcesphase drift / jitterLaser timing & power13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • Demonstration of stability in IPBSM operation : signal Jitterlong term stable performance is maintained under various scan conditions standard Long range scans dedicated to error studies : just as stable (jitter is not increased) compared to usual scans(beam & IPBSM conditions, analysis method kept consistent)

    *Usual scans immediately before & afterComp Sig. jitter is quite consistent at generally 20 25 % (@peak of fringe scans) Fine scanNav = 20 events at each phase stepLong range scans 60 rad(usually 20 rad)13/05/29ECFA LC 2013 *Long scans from other periods show similar stability

    datarangeComp sig jitter (@peak of fringe scans)130314_15575820 radNav = 1021.1 %130314_16573720 radNav = 1025.2 %130314_16342020 radNav = 2024.3%130314_16395260 radNav = 1025.4 %130314_16484060 radNav = 1026.3 %

    ECFA LC 2013

  • preliminary Signal jitter: 24.3 %(at peaks)

    1st of 2 consecutive long range scans Signal jitter: 25 %(at peaks)

    2nd of 2 consecutive long range scans 13/05/29*60 rad rangepreliminary S/N ~ 5.860 rad scans dedicated to error studyATFII Review Stability is maintained for long range scans (fluctuation / drift e.g. BG, phase, timing, power, ect)

    consecutive fringe scans : drift < 70 mrad / min ( negligible) Phase DriftECFA LC 2013 * final set of scans on 3/8 : very stable (initial phase) vs (time)(initial phase) vs (time)

    ECFA LC 2013

  • Comp Signal Jitter BGjitter

    signal jitter derived directly from actual fringe scans (peaks) 20 25% ECFA LC 2013 13/05/29*

    *scaled by S/NiCT monitor fluctuation

    Relative beam laser position* Intrinsic CsI detector energy resolution (GEANT4 sim.)detector energy resolutionSignal Jitter Sources < 10% under investigation < 1 % ~ 3 % 6 - 7 % (monitored by PIN-PD signal)< 5 % ICT monitor accuracy measured Comp sig energy normalized by beam intensityvaries with beam condition Spring, 2013: 174 deg modecontribution to Sig Jitter Esig / Esig, avgStudy of Signal Fluctuation ~ 1 % (from photo-diode)Prepared offline veto for large timing, power jittered events

    ECFA LC 2013

  • hard to separate from other fluctuation sources (laser pointing jitters, drifts, ect.) jitters can vary greatly over timePhase Jitter / Relative Position Jitter Cant push all fluctuation to phase jittersfitted energy jitters with contributions from statistics, timing, BG , and xpreliminary take high statistics scans (Nav ~ 100) under optimized conditions for dedicated analysisderive horizontal rel position jitter x using high statistic laserwire scanif y < 0.3 * y (ATF2 beamline design) Cy > 90 % for y* = 65 nmIssue 1: y M reduction Important to grasp residual M reduction factors in order to derive the true beamsizeIssue 2 : fluctuation source during fringe scanIf x 2.5 m cause 4 % signal jitters (assume Gaussian profile laser = 10 m)

    13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • possibly veto jittered points under clearly identified causesGoal: achieve precise Mmeas (y,meas )/ 2 plate setting 13/05/29ECFA LC 2013 *IP area:QD0, QF1MFB2FF : "vertical IP-phase BPM"ATF2 beamline & BPMsCheck for correlation of signal jitters with e beam orbit in BPMs e.g. MREF3FF (high location for ref cavity scan )synchronize fringe scan data with all ATF2 monitors e.g. BPMs, ICT monitors

    ex): check y position jitter@IP using MFB2FF : "vertical IP-phase BPM e beam orbitjitterRMS 1.3 nsRelative timing cut (beam laser)e.g. 1-sigma

    Observe Esig dependence on Esig : Investigate Signal FluctuationAnticipate O(nm) res. measurement of beam position jitter at IP by IPBPMs (under commissioning)

    [1] improve hardware [2] data selection

    ECFA LC 2013

  • Performance Evaluation #2:Modulation Reduction Factors13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • Direct Methodconsecutive mode switching , under same beam condition (e.g. : 2 7 30 ) use a y that yields very high M at low mode observe upper limit on Mmeas

    Note)apply to a particular dedicated data sample

    Indirect MethodEvaluate each individual factor offline and sum up

    Note) represents the typical conditions of a particular period however hard to derive overall M reduction (e.g. some factors lack quantitative evaluation, vary over time, only can get worst limit)

    Study of M reduction

    ModulationReduction FactorUnder-evaluate M, over-evaluate y How to evaluate M reduction?ECFA LC 2013 13/05/29*

    ECFA LC 2013

  • Plan for assessment of M reduction factorshow to find out bias due to uncertain individual factors: (e.g. relative position jitter, spatial coherence)

    At a low mode : measure a large M (near resolution limit) using a sufficiently small y compare results with higher modes

    example: if we measure M corresponding to y = 350 nm at 7 deg mode expect M = 0.98 at 2.75 deg mode (try to keep within 2-8 deg) what if we get only 0.95 ??? Ctotal 0.97 no individual bias factor worse than 0.97

    Note: conditions may vary over time confirm with repeated measurements need prove that these factors are really independent of priorities1st : suppress M reduction aim for Ctotal 12nd: precisely evaluate any residual errors derive the true beam size13/05/29ECFA LC 2013 *test using direct method

    ECFA LC 2013

  • Major bias if unattended torelative position jitter (phase jitter)Spatial coherenceLimited by alignment precisionCould be major bias Measured polarization and half mirror reflective propertiesResolution of mirror actuators aligning laser to beamECFA LC 2013 Spring 2013, 174 deg power measured directly for each path drift : < 70 mrad / min during consecutive fringe scans

    Still quantitatively uncertain under evaluation:Beamtime final optimization by tilt scanassume Gaussian laser profile (spot size)Individual M Reduction Factors Represent typical condition of a particular period13/05/29*

    Error source M reduction factor

    Fringe tilt (z, t) profile imbalance Cpro > 98.5% power imbalance Cpow > 99 %Laser polarization Optimized to S state using / 2 plate

    Phase drift not major issue

    Laser path alignment Ct,pos : ~ 99 %, Cz,pos : > 98 %

    ECFA LC 2013

  • laser polarization related measurementspolarization measured just after injection onto vertical tablevery close to linearly S polarization should be very little polarization related M reductionresults/ 2 plate setting 13/05/29ECFA LC 2013 90 deg cycleP contamination: Pp/Ps = (1.46 0.06) % Set-upIPBSM laser optics is designed for pure linear S polarization to precisely confirm there is no residual M reduction . next plan individual measurements for upper and lower paths near IP

    Hardware prepared carry out in June

    also measured reflective properties of half mirror

    Rs = 50.3 %, Rp = 20.1 % Match catalog specifications !!half mirror*power ratio

    ECFA LC 2013

  • ECFA LC 2013 *lower S peaks (maximum M) also yield best power balance Minimize M reduction*S peak P peak 45 deg between S and PDuring Beamtime /2 plate scan to maximize Mlaser polarization and power balanceRotate /2 plate angle lens upper power meter investigate power balance: U vs L path 90 deg180 degRotate /2 plate and measure high power Immediately in front of final focus lenses13/05/29M reduction factor due to power imbalance

    ECFA LC 2013

  • Mismatch in axis between fringe and beam transverse longitudinal

    laser path observed on lens: precision ~ 0.5 mm (few mrad) Fringe Tilt issues: Position drifted by the time we scan e beam may also be rotated in transverseCurrent method tilt scanfringe pitch / roll adjustment: observe M reduction Ctilt (70 - 80% if uncorrected) directly use e beam as reference for tilt adjustment

    *(study of fringe tilt by Okugi-san)important adjustment to eliminate M reduction 13/05/29ECFA LC 2013 ex fringe pitchM 0.07 0.32 Mirrors for adjusting tilt M174L Y (8.9 mm 9.01 mm )

    ECFA LC 2013

  • beamsize monitor using laser interferenceOnly existing device capable of measuring y < 100 nm Indispensible for achieving ATF2 goals and realizing ILC

    < Status > contribute with stable operation to continuous beam size tuningConsistent measurement of M 0.3 174 mode) at low beam intensity correspond to y ~ 65 nm (assuming no M reduction)Application of various linear / non-linear multi- knobsdedicated studies of e beam and IPBSM errors

    Performance significantly improved by laser optics reformssuppressed error sources, improved laser path reliability & reproducibilitySummaryECFA LC 2013

    Maintain / improve beamtime performance : e.g. stability, precisionAssess residual systematic errors derive the true beam size stable measurements of y < 50 nm within this runGoalsShintake Monitor (IPBSM)Towards confirming y = 37 nm13/05/29*

    ECFA LC 2013

  • ECFA LC 2013 Backup13/05/29*

    ECFA LC 2013

  • hard to separate from other fluctuation sources (laser pointing jitters, drifts, ect.) jitters can vary greatly over timePhase Jitter / Relative Position Jitter Cant push all fluctuation to phase jittersfitted energy jitters with contributions from statistics, timing, BG , and xpreliminary take high statistics scans (Nav ~ 100) under optimized conditions for dedicated analysisderive horizontal rel position jitter x using high statistic laserwire scanif y < 0.3 * y (ATF2 beamline design) Cy > 90 % for y* = 65 nmIssue 1: y M reduction Important to grasp residual M reduction factors in order to derive the true beamsizeIssue 2 : fluctuation source during fringe scanIf x 2.5 m cause 4 % signal jitters (assume Gaussian profile laser = 10 m)

    13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • Ex #2: check y position jitter@IP using MFB2FF : "vertical IP-phase BPMEX#1: MQD10BFF (high location near ref cavity MREF3FF)13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • simulationMeasures y* = 25 nm few m with < 10% resolution

    Expected Performancemust select appropriate mode according to beam focusingECFA LC 2013 Resolution for each mode13/05/29*

    ECFA LC 2013

  • Laser interference schemeTime averages magnetic field causes inverse Compton scatteringphase shift at IP wave number component along y-axis 2ky = 2k sin modulation depends on cosS-polarized laserWave number vector of two laser pathsFringe pitch13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • Calculation of beam sizeTotal signal energy measured by -detectorConvolution of Laser magnetic field Sine curveElectron beam profile GaussianM : Modulation depthLaser magnetic fieldElectron Beam profile with beam size y along y-directionS : Max / Min of Signal energy13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • Gamma detectorGammaBeam longitudinal direction: 33cm (17.7radiation length)

    Calorimeter like gamma detectorMulti layered CsI(Tl) scintillatorPMT R7400U (Hamamatsu Photonics)

    Width : 10 cmHeight : 5 cm13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • Phase control by optical delay lineOptical delay line (~10 cm) Controlled by piezo stageMovement by piezo stage : stagePhase shift13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • measurement schemeelectron beamTotal energy of gamma raywire positiongammawire scanner, laser wirePhase of laser fringemeasurablebeamsize ~ 1mmeasurablebeamsize < 100nmShintake monitorTotal energy of gamma rayCalculate beam size from Gaussian sigmaCalculate beam size from contrast of sine curve13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • 13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • 13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • laser path misalignmenttransverselongitudinalprecision of alignmnet by mirror actuator z,about 15-20% of z,laser (from zscan) tabout 5-10% of t, laser * (from laserwire scan)

    z,laser about half of t,laser

    longitudinal Cz- pos > 98.9 % transverse Ct-pos ~ 99.9 %*13/05/29ECFA LC 2013 *

    ECFA LC 2013

  • If y ~ 0.3 y C 88.4% for 70 nm @ 174 degC 96.2% for 150 nm@30 deg modeC97.7% for 500 nm@7 deg mode

    phase jitter observed from fringe scan: about 200 mrad ?? C 98 % (????) Phase (relative position) jitter13/05/29ECFA LC 2013 *

    ECFA LC 2013

    ** In order to achieve the high luminosity needed atILC necessary for the anticipated detection and detailed research of new physics the vertical electron beamsize must be focused down to 5.7 nm.

    At ATF2, the FFS test facility prototype for ILC, Goal 1 is to focus the vertical beam size down to 37 nmFor this purpose, the Shintake monitor, or called IPBSM, is installed at the virtual IP of ATF2 and plays an indispensible role in small beam size tuning there.

    Here is the outline of my talk:I will start with general introduction of IPBSMs measurement methods, expectations, and major rolesThen report on recent beam time results, and hardware upgrades , error studies, then a summary and some future goals----------------------------------------------------------------------------------------------------------------------*

    Here is the measurement method of IPBSMOwing to the scheme of using laser interference fringes as a target for the e beam, IPBSM is the only existing device capable of measuring beamsizes below 100 nm

    The laser transported to IP area is first split into upper and lower paths by a beam splitter we call the half mirror , than while controlling their relative phase with a delay line, they are crossed at IP to form laser interference fringes.

    The e beam, also focused to waist at IP, is collided perpendicularly against these fringes, and the resulting Comp scattered photons enter a downstream gamma detectorMeanwhile the beam, having finished its role , is safely disposed into a dump.

    *What is actually measured by the gamma detector is the modulation depth(in these slides written as M) in the Comp signal.

    Modulation is produced by scanning relative pos between laser phase and beam position

    If the beam is thin , or well focused , compared to the fringe pitches, M will be large, While M is small for a large beam size

    It can also be explained as that the no. of Comp scattered photons N is large when e beam interact with fringe mountain/valley, and small in between. This changes more significantly for a small beam .

    *What determines the measurable beam size is the fringe pitch (or d here), which is in turn depend on the crossing angle of the two paths.

    By switching between few crossing angle modes, IPBSM is by design capable of measuring with better than 10% resolution a wide range of beam sizes from 20 nm to as large as few micron.

    This plot shows the relationship between beam size and M. As you can see here, for larger beam sizes, we use small angle modes such as 2 - 8 deg, and for smaller beam sizes 30 and 174 deg modes.

    In order to always maintain better than 10 % resolution, we must select the appropriate crossing angle modes in accordance with beam focusing status. Simulation Condition90 bunches, statistical error 10% y = 0.3 yFringe phase jitter400 mradLaser pulse energy fluc. 6.8%

    *This is the laser path of each crossing angle mode.Mode switching is effectively carried out by individual sets of linear stages on remotely controlled movers, teamed up with mirror actuators.

    This is 174 deg, then 30 deg mode path,The 28 path is special. A prism is slid back and forth to enable continuous adjustment between 2-8 deg , to allow larger freedom in beam tuning for a larger beam size.*Now, here are some of the major steps in using IPBSM during beam tuning at AF2.

    Highly precise position alignment of laser to e beam in transverse plane and in longitudinal (or beam) direction.

    In the transv direction, we perform what is called a laserwire scan, in which one laser path at a time is scanned vs the beam in the transv direction , and laser pos is set to the peak of the resulting Gaussian Comp signal. The sigma here is supposed to correspond to laser spot size (in sigma) is about 15 20 micron.

    In the long direction, we perform the z scan, for which we actually conduct fringe scans along the way, and setl ong laser pos to where it yields the largest M.

    After all these preparations, IPBSM is ready to continuously and stably measure beam sizes using interference scans, and feeds back the result back to tune the e beam size down, for example in the multi-kobs tuning process shown here. ==================================================================**Here is the report on beamtime in 2012By Feb 2012, we have fully commissioned the 30 deg mode, and at the time we were stably measuring M of about 0.55, corresponding to about 150 nm.

    A major upgrade of the laser optics had been carried out during the summer shutdown period of 2012, which aimed at suppressing syst error sources and more reliable laser path construction

    During the following winter run, this was shown to have contributed significantly to performance improvement of IPBSM , and the remarkably stable IPBSM operation which supported various aspects of ATF2 beam tuning.

    High M was measured at 30 deg mode,And on the last 2 days of Dec run, we have finally achieved M detection at 174 deg mode, and measured many times M of 0.15 0.25, corresponding to beam sizes of 70 82 nm. (assuming no M degradation due to IPBSM related factors)

    In this way, the Dec run was a large step forward in achieving ATF2 goal 1.

    These data as well as data taken recently are ongoing undergoing detailed analysis *During beam time in 2013, IPBSM continued to contribute stably to beam tuning and studies, which includes measuring M at 174 deg modes over a few days at a time to enable application of various linear / nonlinear tuning knobs, aimed at further focuing down of beam size.

    This plot here shows one of the large M measured at 174 deg mode during this period

    Occasionally, after several rounds of tuning knobs, we take these consistency scans to test the effect of tuning knobs and of reliability & stability of IPBSM measurements.

    The best record for these consecutive scans is about M = 0.3, as in this example here , where 10 scans have been conducted within about 40 minutes.----------------------------------------------------------------------------------------In March operation:We could not apply Y24, Y46 ( normal sext. : current limit (>0 88A) . of SF5FF PS) We could not apply Y26 (skew sext current limit of SK2FF PS (use QK knobs instead).Plan to use four 20A PSsfor SK1FFSK4FF, and change SF5FF PS fromHAPS to 10A PS

    *Besides beam focusing, IPBSM was also used for dedicated study of the beam or errors in IPBSM itself.

    I will show some examples here.

    For example, for study of wakefield effects, we observe the dependency of M , or IP beam size on beam intensity. Another example is this scan of the pos of a ref cavity BPM in a high beta region is scanned and the response in M is observed.

    This plot here shows the dependence of BG levels with beam intensityHere we can quantatively observe steepness of the response to intensity by comparing the coefficinet of the linear fits between different periods of beam tuning. *The optimistic results reported so far owes largely to the major laser optics upgrade of 2012 summer, which aimed at suppressing syst error sources and more reliable and precise laser paths

    This table lists some main points.For ex, CW alignment laser was put to use in realigning a large number of optical components on the vertical table. Then the laser beam itself was made to pass with high precision along guide of alignment irises and strictly redefined reference lines.

    It is essential to maintain laser path precision when switching modes.A new IP target was installed which allowed us to check 30 deg and 174 deg mode path within the same screen at IPIn place of the rotating stages shared between different modes that used to be used for mode switching, we replaced them with more reliable small linear stages, an independent set for each mode, teamed up with actuator attached mirrors.

    Also, focal lenses for all modes were enabled with lens position scanning, to ensure precision in aligning laser focal point to IP

    Furthermore, tuning of the main laser itself is provides careful regular tuning and maintenance by Spectra Physics company for stabilizing oscillation, and improving laser profile and spatial coherence.

    **Next I will talk about stability of IPBSM operation based on fluctuation sources Such as Compton signal jitter, phase stability, and timing and power stability status*

    It can be demonstrated that during 174 deg mode operation this year, neither the measured M nor other fluctuation factors such as laser timing or power were significantly drifted when we prolonged the scan range in this way,

    In other words, long term stable performance is maintained under various scan conditions, whether it is many consecutive scans or single long scans, or fine step scans.

    This table compares comp signal jitter at peak of fringe scans for a certain set of 2 consecutive long 60 rad scans take on 3/14, compared to regular range scans immediately before an dafter to prove this point, as well as 2 otehr long range scans from other days in 2013.

    *In between small beamsize tuning, we gather data that enables a grasp of IPBSM operation stability, and how it affects fringe scan measurements.

    These are the 2 consecutive long range scans, during which conditions are proven to remain stable.

    These higher statistics data are used for analyzing signal jitters or phase drift and jitters.

    Phase stability, which can also be interpretted as relative position stability, is a factor that easily changes condition over time, and are analyzed using consistency scans, as shown here. The initial phase of consecutive fringe scans are plotted against times passed.The largest drift observed for this method is about 70 mrad / min. Since the typical scan for beam tuning is about 20 rad , taking about 80 sec, this is not a major issue.On the other hand there are stable periods with almost no drift al all.

    *This table shows the main statistical fluctuation sources , and how much each factor is estimated to contribute to overall signal jitters

    These consist of

    The effect from relative position jitter is still uncertain, and different methods are currently being tried out to investigate this.

    Meanwhile, signal jitter is analyzed separately from individual fringe scans themselves, for example at the peaks, mid points, or bottoms of fringe scans

    The sig jitter at fringe scan peaks are considerably consistent at about 20 25%. *A difficult error source to assess is relative pos jitter., or equivalently phase jitters. It is hard to separate rel pos jitter from other jitter sources, and it is not accurate to push all of the seemingly horizontal fluc on fringe scan plot to phase jitters.

    Vertical rel pos jitter (this y here) could potentially be a dominant M reduction factor, but we are not sure, so this is currently undergoing investigation.In general, by design, beam position jitter in ATF2 beamline is below 0.3 x beamsize, but since we are still awaiting full commissioning of IPBPMs, there hasnt been a way to confirm this confidently.

    On the other hand, horizontal pos jitter affect fringe scans by causing signal jitters. This has been investigated using high statistics laserwire scan (such as this one here, where 99 points are taken at each location) , and the standard deviation errors at each place is plotted as a function of the scanned positions and fitted with a function that takes into account the various dominant fluctuation sources, as well as relative position jitter.

    This rel pos jitter is a convolution between jitter of the e beam and of laser pointing stability. This of course also vary from diff data time to time, but assuming about 2.5 4.5 micron of H rel position jitter, it is estimated to contribute about 4-6% to signal jitter .

    Investigation of this issue is taken seriously and plans are to consider alternative methods, as well as acquire data of higher statistics to improve precision in analysis.

    *In order to resolve signal jitters, and achievemore stable scans for beam tuning and a more precisely fitted Modulation, we need to (1) imrpve IPBSM hardware. Another things is to try to apply offline data selection after a thorough investigation of the fluctuation sources.

    Some studies have been carried out where signal jitter dependence on sugnal energy is observed, such as in this model here, where there a statistics dominant terms proportional to sqrt of E, linear on E, which relates to laser conditions, and const terms.

    This plot shows laser timing gathered into a histogram for one single fringe scanWe can apply timing cut, such as vetoing data outside 1 sigma .

    Also, recently IPBSM related data have been synchronized with info from all ATF2 beampline BPMs all within the same fringe scan data, These can be checked for correlation of orbit fluctuations to signal jitters in the IPBSM. For example in a vertical IP phase BPM here, and some in high beta locations. **Next I would like to explain about a study of the various systematic errors affecting IPBSM measurements , which are interpretted using what is called M reduction factors.The c1, c2, so on. Here represent each individual type of bias factors, and these work to smear the fringe contrast and consequently lead to under-evaluation of M, or over-evaluation of beamsize.

    I have classified 2 ways to evaluate M reduction

    Method 1 I call direct method, where total M reduction is estimated using data where mode is switched under the same beam condition within a short time. M is measured at each mode, and consistency is compared to results from higher crossing angle modes. For example, a beamsize that is large enough to be able to measured, but small enough to give a very large M near the resolution limit at a low crossing angle , and we can observe the limit as to how high M can be measured.

    Another is what I call the indirect method, where for each individual bias factor, the corresponding contrast reduction factor is evaluated to best of what is known, .Although this is a more generalized than the 1st part, for some factors we do not yet have a clear way to assess quantatively, and from some we can only get the worst limit.So an overall M reduction may be difficult. *There are two main points to how to assess M reduction factors:The 1st priority is to suppress error sources through improvements in hardware and alignment methods.The 2nd priority is to evaluate any residual errors to the best precision possible, in order to be able to derive the so called true beam size from the measured value.

    So the next question here is how to assess the so called unknown or uncertain errors:Some of these can be proven to not depend on crossing angle mode, or not much affected by beam condition, in other words not much change over time.For uncertain individual factors: (e.g. Relative position jitter, spatial coherence, ect )Do we really have large bias ? how much ? how to find out ??

    For example, we are not sure yet of the quantative effect that spatial laser coherence has on fringe contrast.

    This goes back to my explanation of the direct method for evaluating total M reduction factor.For ex we can try to measure a beamsize that gives a very high theoretic value for M near resolution limit at a low crossing angle mode, like about 0.98 or so, Then lets say we get about 0.97 , not one time, but a average after a few almost stable scans, then that is the very worst contrast reduction factor overall we can get.In this case no individual factor can be worse than 0.97.In fact, as there are many small error sources, probably no big individual errors at all.

    But then again some other error sources are not so easy to generalize since they change over time, and depend on angle mode. *

    This table shows individual systematic errors affecting IPBSM measurements based on data taken at 174 deg mode in 2013

    These are derived using beam time data , or from measurements conducted during beam off time specifically for the purpose of error studies

    Some factors we particularly put effort into suppressing are those related to polarization and fringe tilt , which if left unattended to, would become a dominant M reduction factor. (this I will explain in detail in a moment)

    Others include

    Some factors , such as spatial coherence and phase jitter, are still difficult to evaluate quantatively

    Certain factors which have nearly no dependence on crossing angle mode, or does not vary much , while others need to be confirmed for individual scans.*The IPBSM laser optics is designed for linearly S polarized laser.A fine and precise measurement of the actual polarization of the IPBSM laser was carried out just before the continuous run of May 2013.

    Here is the setup: Just after injection onto the vertical table at IP, we use a half lambda plate to rotate the polarization state of the laser, in other words, its angle is scanned while a high intensity beamsplitter is set up to reflect S laser component upward into a power meter to be measured, while letting P component pass.The results show the IPBSM laser to be very close to purely S linearly polarized state, with P contamination of less than 1.5 %.

    There should be very little M reduction from polarization once we set the half lambda plate to the optimum setting.

    However to evaluate any residual error, if there is any, we need to investigate what happens after the laser pass through numerous optical components all the way to IP. The reflective properties of the esp important half mirror has been measured, and resulted to be well within catalog specifications. Furthermore we need to measure the polarization separately for U and L paths,, just immediately in front of the lenses, the setup for this has been prepared, and will be conducted in June. *In this way, we have determined the setting that yield this pure S polarized state

    We confirmed these particular angles, referred to as S peaks here, to be also the one that maximizes M, as confirmed using the beam during beamtime, where M is measured while scanning half lambda plate angle.

    If laser power is imbalanced between U and L paths, there will also be some M reduction . By measuring power as a function of half lambda plate angle immediately in front of the lenses near IP, it was found that the setting for S state also were the ones that realizes the best power balance. *Fringe tilt correspond to the mismatch in axis of laser interference fringes and beam axis.It takes the form of fringe pitch (in longitudinal direction) , or roll (in transv plane)

    This can be observed from the relative offset wrt lens center for U and L paths. However, the observation precision is limited to about 0.5 mm, and the fact that the e beam itself may be rotated in the transv plane, in which case it is difficult to say to what reference we are defining fringe tilt.Some other issues are that laser drift may well have occurred from time of pre-beamtime path making and when we actually conduct beamsize measurements with IPBSM.

    So, we put into practice what we call the tilt scan .Here, fringe is intentionally tilted by adjusting these mirrors in the 174 deg mode path and responding change in M is observed. Mirrors are set to where M is maximized.From this we can estimate tilt (either pitch or roll) with regard to e beam itself.

    *Shintake monitor (or IPBSM), a beamsize monitor using laser interference as target for the beam, is the only existing device capable of measuring y < 100 nm And plays an Indispensible role in achieving ATF2 goals and realizing ILC

    It has been contributing with remarkably stable operation to ATF2 continuous beam size tuning.Its performance have been improved by laser optics reforms and laser tuning. Consistent measurement of M > 0.3 174 mode) have been achievedcorresponding to y ~ 65 nm (assuming no M reduction)Other new records during recent beam run include application of various linear / non-linear multi- knobs.

    Dedicated studies have been done relating to e beam and IPBSM errorsOffline detailed analysis is still ongoing. As ATF2 is moving still closer to achieving the 37 nm goal, It becomes all the more imporatnt to maintain and improve performance in terms of stability and precision.Furthermore residual syst errors must be assessed to be able to derive the true beam size. **A difficult fluc source to assess is relative pos jitter., or equivalently phase jitters. It is hard to separate rel pos jitter from other jitter sources, and it is not accurate to push all of the seemingly horizontal fluc on fringe scan plot to phase jitters.

    We are not sure whether this is a dominant contribution to sugnal jitters, as well as a M reduction factor.

    It is generally said that by design, beam position jitter in ATF2 beamline is about 0.3 x beamsize, but since we are still awaiting full commissioning of IPBPMs, teher hasnt been a way to confirm this confidently.

    Recently a method is being tried out that compares the difference in sig jitter between peaks, mid points, and bottoms of a fringe scan.

    Investigation of this issue is taken seriously and plans are to consider alternative methods, as well as acquire data of higher statistics to improve precision in analysis.

    *What determines the measurable beam size is the fringe pitch (or d here), which is in turn depend on the crossing angle of the two paths.By switching between few sets of angle modes, IPBSM is by design capable of measuring with better than 10% resolution a wide range of beam sizes from 25 nm to as large as few micron.

    This plot shows the relationship between beam size and M. As you can see here, for larger beam sizes, we use small angle modes such as 2 - 8 deg, and for smaller beam sizes 30 and 174 deg modes.

    This right hand plot sjows beam size vs measurement resolution. In order to always maintain better than 10 % resolution, we must select the appropriate crossing angle modes in accordance with beam focusing status. Simulation Condition90 bunches, statistical error 10% y = 0.3 yFringe phase jitter400 mradLaser pulse energy fluc. 6.8%

    *

    Optical delay line

    8mm1buncht*

    Convolution

    Cos exp

    Y_yconstrast M

    *CsI(Tl)PMT,5cm 10 cm33 cm17.7 radiation length

    *Optical Delay line10cm

    *

    ()

    Id like to show you the difference between the conventional beam size monitor and shintake monitor.wire scanner or laser wire, normally used in accelerator study, scans the wire like this,and gamma peak and the size can be detected.Shintake monitor scans the laser interferrence as the ``slit'' of the strength of magnetic field.,This scheme enables to measure smaller beam than the wavelength of laser.

    Using laser interferrence as the ``slit'' of photons enables it to measure such a small beam size. The conventional beam size monitors, like laser wire or wire scanner cannot measure such small beam size.

    *misalignmentrlower

    misalignemnt

    laser **misalignmentrlower

    misalignemnt

    laser *300 nm

    Mard IPBPMcommissioning

    -------------------------------------------------------------------LorentzP S (e- rest frame

    *