Author
joe-lozzoll
View
215
Download
0
Embed Size (px)
8/14/2019 Lecture 6 Dta & Dsc01
1/22
Lecture 6 DTA & DSC 01 1
DIFFERENTIAL THERMAL ANALYSIS (DTA)
AND
DIFFERENTIAL SCANNING CALORIMETRY (DSC)
The Differential Thermal Analysis (DTA) and Differential Scanning
Calorimetry (DSC) have always used to study the same phenomena.
Differential Thermal Analysis (DTA) is a thermal technique where the
sample temperature is compared with the temperature of a reference
(an inert material) and the temperature difference is recorded as a
function of the sample, reference or furnace temperature while the
sample is being heated or cooled at a constant rate.
The temperature change is detected by a differential method as shown
in Figure 6.1.
8/14/2019 Lecture 6 Dta & Dsc01
2/22
Lecture 6 DTA & DSC 01 2
Figure 6.1 The basic concept of DTA system
8/14/2019 Lecture 6 Dta & Dsc01
3/22
Lecture 6 DTA & DSC 01 3
The DTA method is very useful for the characterization of materials that willundergo physical or chemical changes by the heat effect without the mass
change.
The temperature change of the sample is due to either an endothermic orexothermic reaction or enthalpy change.
The related processes include: a phase change
fusion
melting crystal structural inversion
destruction of the crystal lattice
boiling
sublimation
evaporation
dehydration reaction
other reactions.
The phase change, dehydration, reduction and some of the decompositionprocesses produce endothermic effect, while crystallization, oxidation and
some of the decomposition reactions give exothermic effects.
8/14/2019 Lecture 6 Dta & Dsc01
4/22
Lecture 6 DTA & DSC 01 4
Comparison between the thermal method and the differential thermal
analysis are shown in Figure 6.2
The Normal Thermal Method:
In curves (a) and (b) the
temperature is measured as a
function of the furnace
temperature or time, while the
sample is continuously heated at a
constant rate of temperature
increase.
No enthalpy change is indicated by
the curve (a), but an endothermic
and exothermic enthalpy change is
indicated by the curves in (b).
Subsequently no deviation in curve
(a), but there are deviation in
curves (b): the change in Ti is dueto exothermic or endothermic
change.
These changes subside at Tf and
the sample temperature returns to
that of the furnace temperature.
The Differential Thermal Method:
In curve (c) the difference between the sample
temperature and that of the reference (TsTr) ismeasured as a function of the system
temperature, T.
At Ti, the curve deviates from horizontal and
forms an upward peak (exothermic) or
downward peak (endothermic).
The final reaction temperature Tf, does not occur
at the top of the peak, but lies on the higher
temperature side of the relevant peak. Its exactposition depends on the instrumental system
used.
Therefore, in a differential thermal method, a
very small temperature change can be detected
easily and the peak area is proportional to the
enthalpy change (H) and the sample mass.
http://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_2%20Compare%20Thermal%20DTA.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_2%20Compare%20Thermal%20DTA.ppt8/14/2019 Lecture 6 Dta & Dsc01
5/22
Lecture 6 DTA & DSC 01 5
DTA curve is a plot of the temperature difference (T= Ts - Tr) versus the temperature
(T). Endothermic process is shown as a downward peak, while exothermic process
is shown as an upward peak (Figure 6.3)
Figure 6.3
Exothermic
Endothermic
Four types of transitions
detected by DTA method:
I second order transition,
where a change in the
horizontal line is detected
(e.g. glass transition);
II a sharp endothermic curve
due to the fusion or melting
process;
III a broad endothermic curve
due to a decomposition or
dissociation process;
IV an exothermic curve due to
the crystalline phase
change.
8/14/2019 Lecture 6 Dta & Dsc01
6/22
Lecture 6 DTA & DSC 01 6
Two approaches in the temperature sensing system of DTA (Figure 6.4):
Temperature Sensor in DTA
1. Classic DTA: the thermocouple is immersed into the sample and the reference materials
2. Heat flux DTA: the thermocouple is placed outside the sample and the reference
materials
Figure 6.4
8/14/2019 Lecture 6 Dta & Dsc01
7/22
Lecture 6 DTA & DSC 01 7
Differential Scanning Calorimetry (DSC)
DSC is a technique in which the difference in the heat flow (power) to the sample (container) and the
reference (container) is detected versus time or temperature while the sample is continuously heated
in a certain atmosphere
i) Power Compensated DSC: the sample and the reference are heated separately by
using different heaters and the temperature difference between the sample and the
reference is maintained zero while the electrical power needed to maintain that both
temperatures are the same (P= d(Q/dt) is measured (Figure 6.5)
Two Types of DSC:
Figure 6.5
The heat absorbed by the sample per unit time is
compensated accurately by the difference in the
electrical power (P) supplied to the heaters
P = dq / dt = . (CSCR)
The peak area is, therefore, directlyproportional to the value ofH. In the power
compensated DSC the calibration constant
needed to convert the peak area into joule is
the electricalconversion factor.
8/14/2019 Lecture 6 Dta & Dsc01
8/22
Lecture 6 DTA & DSC 01 8
The Effect of High Temperature
Most DSC systems are designed to measure heat capacity and H attemperatures of less than 700 oC. At higher temperatures, radiation (which
increases with the temperature rise) becomes more significant.
Most DTA systems work on the principle of heat conductivity at lowertemperatures where the heat radiation is low. At higher temperatures the heat
transition from its source to the sample takes place at faster rate and if the
thermocouple is placed below the sample holder, the system becomes less
sensitive.
When a sample undergoes a reaction, there is a temperature gradient in the
sample. In order to obtain a meaningful calorimetric data, the sample size should
be minimum to reduce the heat gradient in the sample. However, if the sample
consists of different materials it is usually difficult to obtain a sufficiently small butrepresentative sample.
The Effect of the Sample Size
Power Compensated DSC
8/14/2019 Lecture 6 Dta & Dsc01
9/22
Lecture 6 DTA & DSC 01 9
ii) Heat Flux DSC: the sample and the reference are heated using the sameheat source and the temperature difference T is measured (Figure 6.6). The
signal is converted into the power difference Pand this value is plotted versus
temperature or time
Figure 6.6 Heat flux DSC (Source: Haines F 3.4)
The DSC curve is a plot of P versus
temperature (or time). Since an endothermic
process involves absorption of more heat by the
sample than the reference, the process is
indicated by the upward peak.
In a heat flux DSC system thethermocouple is placed under the
sample and the reference containers
(Figure 6.8 (a)-(e)). The base line
variation is due to a change in the heat
capacity of the sample and it is not
influenced by other properties of the
sample. The DSC system can be usedfor the determination of heat capacity
of the material. However, since the
base line variation is influenced by the
physical properties of the sample
container material, the instrument
needs to be calibrated using a
standard material.
8/14/2019 Lecture 6 Dta & Dsc01
10/22
Lecture 6 DTA & DSC 01 10
The calibration constant Kis a thermal factor (varies with temperature) that is
used to convert the peak area into joule.
The main components of the DTA/DSC system is shown in Figure 6.7.
The area under the peak is directly proportional to the heat of reaction:
H = KTdt = K. (peak area)
THE INSTRUMENTATION SYSTEM OF DSC/DTA
The main components ot the DTA/DSC system:
The DTA/DSC detector and amplifier
The furnace and temperature sensor
Computer and temperature programmer
Recorder, plotter or any data collection device
http://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_7%20DSC%20instrumentation.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_7%20DSC%20instrumentation.ppt8/14/2019 Lecture 6 Dta & Dsc01
11/22
Lecture 6 DTA & DSC 01 11
The DTA/DSC Detector
In the DCS/DTA method, a thermocouple is used to detect the temperature change while the sample
being heated under controlled conditions.
For a low temperature process, a copperconstantan or cromel-alumel thermocouple may be used,
while heating at high temperature or under corrosive atmosphere the Pt-Pt/13%Rh thermocouple isused.
The choice of thermocouple that can be used for a DSC/DTA system is shown in Figure 6.8.
A DTA/DSC experiment normally uses between 10 20 mg of sample.
A sample container may also be specially designed for use in high pressure thermal experiment. For
example in thepressure DSC(PDSC) technique, the sample container can should be able to stand
up to 70 atm pressure. Hence, reactions that produce gas at high temperature can be studied.
For example in the stability study of accelerated oxidative experiment by using a PDSC to a cooking
oil under high oxygen pressure or under catalytic reduction of organic compound by using hydrogen.
The Sample Container
Various materials may be used to make the sample plate and crucible, but most of the sample
container for use at lower temperatures (less than the melting point of Al: 660o
C), are made fromaluminum.
Platinum or ceramic may be used as the materials for the sample holder when the experiments
involve heating at higher temperatures. However, the thermal conductivity difference must be taken
into account in making a choice of the sample holder to be used.
http://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_8%20Thermocouple.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_8%20Thermocouple.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_8%20Thermocouple.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_8%20Thermocouple.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_8%20Thermocouple.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20Figure%206_8%20Thermocouple.ppt8/14/2019 Lecture 6 Dta & Dsc01
12/22
Lecture 6 DTA & DSC 01 12
Reference Materials
DTA and DSC are defined as the differential techniques because the behaviour
of the sample is compared with the behaviour of a reference.
All thermal properties that could be involved in the thermal analysis experiment
must be taken into account.
For example, emissivity of the sample might change when the sample changed
phase, or react or change in colour.
Among the commonly used materials for the reference are calcined alumina
Al2O3 or carborundum, SiC.
Sometimes the sample is diluted with the reference material.
In this case, the sample does not react with the reference material
Dilution of the sample with the reference material will normally improves the
base line and the the shape of the DTA peak.
8/14/2019 Lecture 6 Dta & Dsc01
13/22
8/14/2019 Lecture 6 Dta & Dsc01
14/22
Lecture 6 DTA & DSC 01 14
Figure 6.9: The normal DTA or DSC curve for a melting material
During the melting process,
completion of the reaction is
indicated by the tip of the
peak.
Curve A indicates the curve
that is obtained experimentallyreturns gradually to the
baseline (curve A)
Theoretically, the curve (peak)
should return immediately to
the baseline (curve B)
Active area
8/14/2019 Lecture 6 Dta & Dsc01
15/22
Lecture 6 DTA & DSC 01 15
Figure 6.10 A change in the temperature detected by the thermocouple
before and after the thermal reaction which is due to a change in the
thermal conductivity of the sample material (Source: Haines F 3.7)
8/14/2019 Lecture 6 Dta & Dsc01
16/22
8/14/2019 Lecture 6 Dta & Dsc01
17/22
Lecture 6 DTA & DSC 01 17
The Second Type Theory: The Relationsh ip BetweenTand The Sample Temperature
This theory takes into account various
relationship between the temperaturedifference T and the sample
temperature at certain times in order to
obtain the activation energy and the
order of the reaction.
The theory, however, does not take into
account the effect of heat transfer
There exists a relationship between the
activearea of the DTA peak thermogramand the quantity of heat involved.
The active area of the DTA/DSC curve
evolved from the thermal reaction process,
while the other areas are produced as the
T returns to the baseline after thereaction has completed (point E in Figure
6.9).
The shape of the curve (curve B) is
exponential due to the influence of natural
heating or cooling process and it depends
on the arrangement of the sample holder.
Therefore, the plot of ln(T) vs time
beyond the point E of the curve A will belinear.
The active area is the area that beginsfrom the point where the curve starts to
the point E of the curve.
The relationship between the active area
and the reaction heat is shown by the
following equation:
Heat = A(active area)
whereA = slope of the plot ln(T) vs time
The theory assumes that the physical
properties of the sample and the reference
do not change during the thermal
experiment
Therefore, the curve region after the point
E is influenced by the instrumental system
used.
8/14/2019 Lecture 6 Dta & Dsc01
18/22
Lecture 6 DTA & DSC 01 18
For melting, crystal change and zero order reactions:
The active region of the DTA curve (related to the reaction) ends near the tip of the
DTA peak
A portion of the DTA curve exponentially returns to the baseline which is influenced by
the instrumental parameters
Therefore, the end of the reaction and theA factor can be precisely determined.
For reactions other than melting, the Tposition before the curve returns to thebaseline when the reaction has been completed will depend on the type of the
reaction.
However, if the reaction is controlled by a different type mechanism, the reaction ends up
while the Tgets near to the baseline.
Under such conditions, determination of the constantA would be difficult.
The DTA curve that ends up at high temperature might contain a mixture of reaction
mechanisms and the final portion of the curve is exponential, hence, differ from one device
to another.
8/14/2019 Lecture 6 Dta & Dsc01
19/22
Lecture 6 DTA & DSC 01 19
Since the tip of the DTA peak might
not really represent the reaction
process difficulty could arise if two
close consecutive reactions
produce two overlapping peaks orcompletely combined peak.
i) due to the structural inversion of silica at
573 oC
ii) due to the crystal transition of potassium
sulphate at 583 oC.
The ability to separate such
overlapping peaks indicate the
resolution of the instrument.
For example: a mixture of silica and potassium
sulphate show two peaks:
Most DTA theory assume that:
a) during the thermal reaction, the physical properties of the material does not change
and it does not depend on temperature,
a) the baseline is the same before and after the reaction.
However, most DTA/DSC curves do not have the same baseline before and after
the reaction because physical changes always occur during the thermal reaction.
8/14/2019 Lecture 6 Dta & Dsc01
20/22
Lecture 6 DTA & DSC 01 20
The baseline change of the DTA/DSC
curve will affect:
calculation of the area under the DTA
peak
the magnitude of error in the areacalculation which depends on the peak
size: the smaller the peak the larger the
error
the actual determination of the baseline
If the baseline is changed, estimation of the
peak area may be carried out by first drawing a
straight line across the peak from where the
peak leaves the baseline to the point where the
peak meets the baseline again (Figure 6.11(a)) The result is that the measured peak area
tends to become larger
Two baselines approach may also be used as
shown in Figure 6.11(b).
Figure 6.11 Estimation of thearea under the DTA/DSC curve by
constructing a baseline before andafter the reaction as shown by the
dotted lines
(a)
(b)
8/14/2019 Lecture 6 Dta & Dsc01
21/22
Lecture 6 DTA & DSC 01 21
CALIBRATION
The instrument temperature must be properly calibrated in order to obtain accurate data.
For the DSC system, sensitivity of the calorimetry need to be calibrated.
Some of the reference materials recommended by the ICTAC (International Confederation for
Thermal Analysis and Calorimetry) for the calibration of DTA/DSC system is given in Table 6.1.
The DSC thermogram of the melting process of pure indium metal (99.999%) is
shown in Figure 6.12.
The melting point 156.6 oC is obtained by determining the intersection point Te.
The area marked under the peak, AS in the Figure is used for the determination of
calorimetry sensitivity constant, K:
K = HS . mS /AS
Where HS = the fusion enthalpy of indium (28.71 J/g)
mS = mass of the indium sample (g)
AS = peak area (cm2)
K = calorimetric sensitivity (J/cm2)
http://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20%20Table%206.1.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20%20Table%206.1.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20%20Table%206.1.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20%20Table%206.1.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20%20Table%206.1.ppthttp://localhost/var/www/apps/conversion/tmp/scratch_8/Lecture%206%20%20Table%206.1.ppt8/14/2019 Lecture 6 Dta & Dsc01
22/22
Lecture 6 DTA & DSC 01 22
Figure 6.12 DSC peak for indium melting process which shows the determination ofintersection point Te (6 mg, 10 K/min, nitrogen) (Source: Haines: F 3.11)
Calibration for a wide operational temperature
range using various materials allows
determination of relationship between K and
temperature.
Example: pure indium sample 6.68 mg gives a
peak with area of 21.94 cm2. Calculate the value
ofKat 156 oC.
Answer: K = (28.70 x 6.68 x 10-3) / 21.94
= 8.74 x 10-3 J / cm2
Note !
The DTA or DSC curve depends on the conditions of the sample
and the instrument:
Heating Rate Sample (chemical properties, purity, origin) Sample mass (volume, packing, distribution, dissociation) Crucible (material, geometry) Atmosphere (gas, static, flowing)