LT1072X1 Cimarec 3

Embed Size (px)

Citation preview

  • 8/10/2019 LT1072X1 Cimarec 3

    1/16

    Cimarec Stirring Hot PlatesOPERATION MANUAL

    AND PARTS LIST

    Model # Top Plate Size Voltage SeriesSP46615 4 x 4 120 1066SP46610 4 x 4 240 1066

    SP46610-33 4 X 4 230 1066SP46614 4 x 4 100 1066

    SP46925 7 x 7 120 1069SP46920 7 x 7 240 1069

    SP46920-33 7 x 7 230 1069SP46924 7 x 7 100 1069SP47235 12 x 12 120 (20 amp cord set) 1072

    SP47235-60 12 x 12 120 (15 amp cord set) 1072SP47238 12 X 12 208 1072

    SP47230 12 X 12 240 1072SP47230-33 12 x 12 230 1072

    SP47234 12 X 12 100 1072

    1

    BARNSTEAD THERMOLYNE

    LT1072X1 5/27/99

  • 8/10/2019 LT1072X1 Cimarec 3

    2/16

    Safety Information ..............................................................................................................................................3Warnings ......................................................................................................................................................3

    Specifications ......................................................................................................................................................6Introduction..........................................................................................................................................................7

    General Usage..............................................................................................................................................7

    Operation ............................................................................................................................................................8To Select a Temperature ..............................................................................................................................8

    To Select a Speed ........................................................................................................................................8Maintenance........................................................................................................................................................9

    General Cleaning Instructions ......................................................................................................................9Replacement Parts............................................................................................................................................10

    Exploded View ..................................................................................................................................................11Ordering Procedures ........................................................................................................................................13One Year Limited Warranty ..............................................................................................................................14

    2

    Table of Contents

  • 8/10/2019 LT1072X1 Cimarec 3

    3/16

    Your Thermolyne Cimarec Stirring Hot Platehas been designed with function, reliability,and safety in mind. It is your responsibility toinstall it in conformance with local electricalcodes. For safe operation, please payattention to the alert signals throughout themanual.

    This manual contains important operating and

    safety information. The user must carefullyread and understand the contents of thismanual prior to the use of this equipment.

    WarningsTo avoid electrical shock, always:

    1. Use a properly grounded electricaloutlet of correct voltage and currenthandling capacity.

    2. Disconnect from the power supplyprior to maintenance and servicing.

    To avoid personal injury:1. Do not use in the presence of flame

    or combustible materials fire orexplosion may result. This device

    contains components which may ignitesuch materials.

    2. Keep top surface clean. Use a non-abrasive cleaner. Alkali spills, hydro-fluoric acid spills or phosphoric acidspills may damage top and lead tothermal failure. Unplug unit andremove spills promptly. Do notimmerse unit for cleaning.

    3. Replace the top immediately if

    3

    Safety Information

    Alert Signals

    WarningWarnings alert you to apossibility of personal injury.

    NoteNotes alert you to pertinent factsand conditions.

    Hot SurfaceHot surfaces alert you to apossibility of personal injury if youcome in contact with a surfaceduring use or for a period of timeafter use.

  • 8/10/2019 LT1072X1 Cimarec 3

    4/16

    damaged by etching, scratching or chip-ping. A damaged top can break in use.

    4. Do not use metal foil on hot plate. Donot use sand bath, metal containers orother insulating material on the stirringhot plate Top plate can be damagedand shock hazard can exist.

    5. Do not remove or modify grounded

    power plug. Use only properly groundedoutlets to avoid shock hazard. Not ratedfor use in hazardous atmospheres.

    6. Use caution when heating volatile mate-rials; top surface and element can reachthe Flash Point Temperature of manychemicals. These stirring hot plates arenot explosion proof. Fire or explosionmay result. Unit contains componentswhich may ignite such materials.

    7. Use appropriate hand and eye protec-tion when handling hazardous chemi-cals.

    8. Gross weight of items placed on top ofHot Plates should not exceed 25 lbs.(11.4 kg.) on the 7 x 7 and 12 x 12models, and 10 lbs. (4.5 kg.) on the 4 x4 model.

    9. Caution: Hot Surface. Avoid Contact.The top plate of the unit can remain hotfor some time after use with no visualindication.

    10. Refer servicing to qualified personnel.

    4

    SAFETY INFORMATION

  • 8/10/2019 LT1072X1 Cimarec 3

    5/16

    WarningTHIS PRODUCT CONTAINS REFRACTORY CERAMIC, REFRACTORY CERAMIC FIBEROR FIBERGLASS (GLASS WOOL) INSULATION WHICH CAN PRODUCE RESPIRABLEFIBERS AND DUST WHEN HANDLED. THESE FIBERS OR DUSTS CAN CAUSE IRRITA-TION AND CAN AGGRAVATE PREEXISTING RESPIRATORY DISEASE. REFRACTO-RY CERAMIC INSULATIONS MAY CONTAIN OR MAY FORM CRYSTALLINE SILICA(CRYSTOBALITE) WHICH MAY CAUSE LUNG DAMAGE (SILICOSIS).

    THE INTERNATIONAL AGENCY FOR RESEARCH ON CANCER (IARC) HAS CLASSI-FIED REFRACTORY CERAMIC FIBER AND FIBERGLASS AS (2B) POSSIBLY CARCINO-GENIC. IARC HAS CLASSIFIED CRYSTALLINE SILICA AS (2A) PROBABLYCARCINOGENIC.

    The insulating materials are located in the door, the hearth collar, in the chamber of theproduct or the top plate assembly. Tests performed by the manufacturer indicate that thereis no significant risk of exposure to dust or respirable fibers resulting from operation of thisequipment under normal conditions. However, there may be a risk of exposure to respirabledusts or fibers when repairing or maintaining the insulating materials, or when otherwisedisturbing the materials in a manner which causes release of dust or fibers therefrom.Through the use of proper handling procedures you can work safely with these insulatingmaterials and minimize any exposure. Accordingly, before you repair or replace any insulat-ing materials, or perform any other servicing on this product which could disturb or causeexposure to dust from insulating materials, you should consult the appropriate MaterialSafety Data Sheets (MSDSs) for such products with respect to proper handling and appro-priate protective equipment. For additional MSDSs, or additional information concerningthe handling of refractory ceramic products, please contact the Customer ServiceDepartment of Barnstead|Thermolyne Corporation at 1-800-446-6060 or 1-800-553-0039.

    5

    SAFETY INFORMATION

  • 8/10/2019 LT1072X1 Cimarec 3

    6/16

    Environmental ConditionsOperating: 17C to 27C; 20% to 80% relative

    humidity, non-condensing.Installation category II (overvolt-age) in accordance with IEC 664.Pollution degree 2 in accordancewith IEC 664.Altitude Limit: 2,000 meters.

    Storage: -25C to 65C10% to 85% relative humidity

    Declaration of Conformity (-33

    models only)Barnstead|Thermolyne hereby declares under itssole responsibility that this product conforms withthe technical requirements of the followingstandards:

    EMC: EN 50081-1EN 50082-1

    Safety: EN 61010-2-010

    per the provisions of the Low Voltage Directive73/23/EEC, as amended by 93/68/EEC.

    The authorized representative located within theEuropean Community is:

    Electrothermal Engineering Ltd.419 Sutton RoadSouthend On SeaEssex SS2 5PHUnited Kingdom

    Copies of the Declaration of Conformity areavailable upon request.

    6

    Specifications

  • 8/10/2019 LT1072X1 Cimarec 3

    7/16

    Your Cimarec stirring hot plate is a generalpurpose heating/stirring device intended forlaboratory procedures requiring temperaturesfrom 150C to 538C (SP47235, SP47235-60and SP47234 range from 120C to 370C)and stirring capability from 100 1000 rpm. Avessel with a flat bottom is recommended onthe stirring hot plate for ideal heat transfer.Avoid metal containers or vessels with a rim

    around the bottom as these could cause dam-age to the stirring hot plate.

    General UsageDo not use this product for anything otherthan its intended usage.

    7

    Introduction

    NoteThis unit contains materialswhich may emit a slight odor orsmoke during initial operation.This will dissipate within a fewminutes and will not effect theoperation or performance of thehot plate.

    These Cimarec stirring hotplates are not rated for use incorrosive environments. Use ina corrosive environment willdamage their components anddegrade their performance.

  • 8/10/2019 LT1072X1 Cimarec 3

    8/16

    Some shifting of motor bearings may haveoccurred during shipping and handling. Prior tofirst use, the stirrer should be run at maximumspeed for 10 minutes to realign the motorbearings.

    Both the temperature and speed controls arecombination ON/OFF switches and settingselectors.

    To Select a TemperatureTurn the control clockwise to the desired setting.The lighted heat indicator shows when the unitis turned on. The control maintains a set tem-perature of the stirring hot plate by turning thepower on and off over short time intervals. Thegreater the percentage of time the power is on,the higher the temperature. At the high position,power is on continuously and maximum temper-

    ature is obtained. From the OFF position, theknob may be turned to either the right or theleft. For fast heat up, turn the knob to the highposition. When the desired temperature isreached, reduce the setting to the point wherethe temperature will be maintained.

    To Select a SpeedTurn the speed control clockwise until the

    desired mixing action is achieved. The lightedstir indicator shows when the unit is stirring.Stirring capability from 100 to 1000 rpm. Unitcomes complete with a stir bar that is coatedwith Teflon Fluorocarbon Resin.*

    8

    Operation

    WarningGross weight of items placed ontop of Hot Plates should notexceed 25 lbs. (11.4 kg.) on the7 x 7 and 12 x 12 models,and 10 lbs. (4.5 kg.) on the4 x 4 model.

    Refer servicing to qualified per-sonnel.

    Disconnect from power supplybefore servicing.

    Do not use in the presence offlame or combustible materials -fire or explosion may result .This device contains compo-nents which may ignite suchmaterials.

    Do not use metal foil on hotplate. Do not use sand bath,metal containers or other insula-tion material on the stirring hotplate - Top plate can be dam-aged and shock hazard canexist.

    CautionWhen turning the control OFF,be sure the knob is in the OFFposition and the indicator light isout.

    Hot SurfaceCaution: Avoid contact.

    *Teflon Fluorocarbon Resin is a trademark of E.I. DuPontde Nemours and Company.

  • 8/10/2019 LT1072X1 Cimarec 3

    9/16

    General Cleaning InstructionsWipe exterior surfaces with lightly dampenedcloth containing mild soap solution.

    9

    Maintenance

  • 8/10/2019 LT1072X1 Cimarec 3

    10/16

    Model Number(s): SP46900, SP47200, SP46600Product Name: Cimarec Stirring Hot Plates

    4 x 4 7 x 7 12 x 12Series 622 Series 775 Series 817

    Key # Part # Part # Part # Description1 PT464X1 PT467X1 PT470X1 Top plate only2 EL691X1 EL694X1 EL472X1A Element, 120V2 EL691X2 EL694X2 EL472X2A Element, 240V

    2 ~ ~ EL472X3A Element, 208V2 EL1066X1 EL1069X1 EL472X4A Element, 100V3 CNX60 CNX60 CNX60 Control, 100/120V3 CNX61 CNX61 CNX61 Control, 208/240V4 MT1066X2A MTX104 MT1072X1 Motor, 100V4 MT1066X2A MTX98 MT1072X1 Motor, 120V4 MT1066X3A MTX103 MT1072X2 Motor, 208/240V5 RS1066X3A RS1069X3A RS1072X3A Rheostat, 100V5 RS1066X1A RS1069X1A RS1072X1A Rheostat, 120V5 RS1066X2A RS1069X2A RS1072X2A Rheostat, 208/240V6 KB1066X1A KB1072X2A KB1072X2A Knob, stir control7 KB1072X1A KB1072X1A KB1072X1A Knob, heat control8 CR586X1 CR586X1 CR612X4A Cord set, 240 volt8 CR584X1 CR177X1 CR612X5A Cord set, 120 volt8 CR584X1 CR177X1 CR177X1 Cord set, 100 volt8 ~ ~ CR612X7A Cord set, 208 volt8 ~ ~ CR612X8A Cord Set 120 volt (15 amp cord)9 PLX108 PLX108 PLX108 Pilot light, stir 100/120V9 PLX109 PLX109 PLX109 Pilot light, stir 208/240V10 PLX108 PLX108 PLX108 Pilot light, heat, 100/120V10 PLX109 PLX109 PLX109 Pilot light, heat, 208/240V11 5120-0025 FZX61 FZX28 Fuse, CE 230V

    10

    Replacement Parts

  • 8/10/2019 LT1072X1 Cimarec 3

    11/16

    Exploded View

    11

    6

    10

    8

    7

    9

    8

    9

    10

    7

    3

    OFF

    1

    0 2

    6

    4

    5

    OFF

    4

    10

    3

    2

    5

    11

    3

    See Top Plate

    Exploded View

    7

    10

    9

    6

    5

    4

    NoteSee Replacement Partslist for replacement fuseon 230V CE model.

  • 8/10/2019 LT1072X1 Cimarec 3

    12/16

    12

    EXPLODED VIEW

    Exploded View: Top Plate

  • 8/10/2019 LT1072X1 Cimarec 3

    13/16

    13

    Ordering Procedures

    Please refer to the Specification Plate for the complete model number, serial number, andseries number when requesting service, replacement parts or in any correspondence con-cerning this unit.

    All parts listed herein may be ordered from the Barnstead Thermolyne dealer from whomyou purchased this unit or can be obtained promptly from the factory. When service orreplacement parts are needed we ask that you check first with your dealer. If the dealercannot handle your request, then contact our Customer Service Department at319-556-2241 or 800-553-0039.

    Prior to returning any materials to Barnstead Thermolyne Corp., please contact ourCustomer Service Department for a Return Goods Authorization number (RGA). MaterialReturned without an RGA number will be returned.

  • 8/10/2019 LT1072X1 Cimarec 3

    14/16

    Barnstead|Thermolyne Corporation warrants that if a product manufactured byBarnstead|Thermolyne and sold by it within the continental United States or Canadaproves to be defective in material or construction, it will provide you, without charge, for aperiod of ninety (90) days, the labor, and a period of one (1) year, the parts, necessary toremedy any such defect. Outside the continental United States and Canada, the warrantyprovides, for one (1) year, the parts necessary to remedy any such defect. The warrantyperiod shall commence either six (6) months following the date the product is sold byBarnstead|Thermolyne or on the date it is purchased by the original retail consumer,whichever date occurs first.

    All warranty inspections and repairs must be performed by and parts obtained from anauthorized Barnstead|Thermolyne dealer or Barnstead|Thermolyne (at its own discre-tion). Heating elements, however, because of their susceptibility to overheating and contam-ination, must be returned to our factory, and if, upon inspection, it is concluded that failure isnot due to excessive high temperature or contamination, warranty replacement will be pro-vided by Barnstead|Thermolyne. The name of the authorized Barnstead|Thermolynedealer nearest you may be obtained by calling 1-800-446-6060 (319-556-2241) or writingto:

    Barnstead|ThermolyneP.O. Box 797

    2555 Kerper BoulevardDubuque, IA 52004-0797USA

    FAX: (319) 589-0516E-MAIL ADDRESS: [email protected]

    Barnstead|Thermolynes sole obligation with respect to its product shall be to repair or (atits own discretion) replace the product. Under no circumstances shall it be liable for inciden-tal or consequential damage.

    THE WARRANTY STATED HEREIN IS THE SOLE WARRANTY APPLICABLE TO

    Barnstead|Thermolyne PRODUCTS. Barnstead|Thermolyne EXPRESSLY DISCLAIMSANY AND ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDINGWARRANTIES OF MERCHANTABILITY OR FITNESS FOR USE.

    14

    One Year Limited Warranty

  • 8/10/2019 LT1072X1 Cimarec 3

    15/16

    15

  • 8/10/2019 LT1072X1 Cimarec 3

    16/16

    9001ISOREGISTERED

    Barnstead|Thermolyne2555 Kerper Blvd.

    P.O. Box 797Dubuque, IA 52004-2241

    Phone: 319-556-2241 800-553-0039

    Fax: 319-589-0516