2
Navigation and Search Key Specifications Comparison Serial Triggering and Decode Tektronix MSO/DPO3000 Series MSO/DPO3000 Series vs. Agilent MSO/DSO7000A Series Competitive Fact Sheet Serial Bus Support I 2 C, SPI, CAN, LIN, RS-232/422/485/ UART, and I 2 S/LJ/RJ/TDM available for all models . Bus Display – View up to 2 serial buses simultaneously. Bus Decoding – Decodes each packet on the bus and displays the value in hex, binary, decimal, or ASCII. Serial Triggering – Trigger on packet content. Event Table – View captured packets in a tabular listing view. Search – Wave Inspector® navigation automatically searches for and marks user-defined events on packet content. Serial Bus Support I 2 C, SPI, CAN, LIN, RS-232/UART and I 2 S (4-channel products only). Bus Display – Single serial bus display, anchored at the bottom of the screen. Bus Decoding – Hardware decode of each packet on the bus in hex, but not always time-aligned. Serial Triggering – Trigger on packet content. Limited serial triggering standard. Event Table – View captured packets in a tabular listing view. No Automated Search Tektronix MSO/DPO3000 Series Agilent MSO/DSO7000A Series Channels 2, 4 (+16 digita l MSO) 2, 4 (+16 digital MSO) Bandwidth 100, 300, 500 MHz 100, 350, 500 MHz, 1 GHz Max. Sample Rate (All channels on) 2.5 GS/s 4 GS/s (1 GHz models) 2 GS/s (all others models) Max. Record Length (All channels on) 5 M points 4 M points Input Impedance 1M, 75, 50 1M, 50 Navigation and Search Wave Inspector® controls Horizontal position, zoom Power Measurements (Optional) Built-in PC-Based Wave Inspector® controls. Pan/Zoom controls for easy scrolling through long records. Play/Pause button scrolls the waveform across the screen automatically. Search/Mark controls enable you to search for events of interest, mark them, then navigate through the record mark by mark. Only uses horizontal scale control to zoom. Manual scrolling of horizontal position to scroll through data. No automated search capability. Agilent MSO/DSO7000A Series Tektronix MSO/DPO3000 Agilent MSO/DSO7000A  Navigation by horizontal position  No Automated Search www.tektronix.com/mso3000 Pan right and some of the decoded text shifts; now it's not time correlated with data.

MSO3K Vs MSO7K

Embed Size (px)

Citation preview

Page 1: MSO3K Vs MSO7K

8/14/2019 MSO3K Vs MSO7K

http://slidepdf.com/reader/full/mso3k-vs-mso7k 1/2

Navigation and Search Key Specifications Comparison

Serial Triggering and DecodeTektronix MSO/DPO3000 Series

MSO/DPO3000 Series vs. Agilent MSO/DSO7000A SeriesCompetitive Fact Sheet

Serial Bus Support – I2C, SPI, CAN, LIN, RS-232/422/485/ UART, and I2S/LJ/RJ/TDM available for all models.

Bus Display – View up to 2 serial buses simultaneously.

Bus Decoding – Decodes each packet on the bus anddisplays the value in hex, binary, decimal, or ASCII.

Serial Triggering – Trigger on packet content.

Event Table – View captured packets in a tabular listingview.

Search – Wave Inspector® navigation automaticallysearches for and marks user-defined events on packetcontent.

Serial Bus Support – I2C, SPI, CAN, LIN,RS-232/UART and I2S (4-channel products only).

Bus Display – Single serial bus display, anchoredat the bottom of the screen.

Bus Decoding – Hardware decode of each packet

on the bus in hex, but not always time-aligned. Serial Triggering – Trigger on packet content.Limited serial triggering standard.

Event Table – View captured packets in a tabularlisting view.

No Automated Search

TektronixMSO/DPO3000 Series

AgilentMSO/DSO7000A Series

Channels 2, 4 (+16 digital MSO) 2, 4 (+16 digital MSO)

Bandwidth 100, 300, 500 MHz 100, 350, 500 MHz,1 GHz

Max. Sample Rate(All channels on) 2.5 GS/s 4 GS/s(1 GHz models)

2 GS/s (all others models)

Max. Record Length (Allchannels on) 5 M points 4 M points

Input Impedance 1M, 75, 50 1M, 50

Navigation and Search Wave Inspector® controls Horizontal position,

zoom

Power Measurements(Optional) Built-in PC-Based

Wave Inspector® controls.

Pan/Zoom controls foreasy scrolling throughlong records.

Play/Pause buttonscrolls the waveformacross the screenautomatically.

Search/Mark controlsenable you to search forevents of interest, markthem, then navigatethrough the record markby mark.

Only uses horizontal scalecontrol to zoom.

Manual scrolling ofhorizontal position to scrollthrough data.

No automated searchcapability.

Agilent MSO/DSO7000A Series

Tektronix MSO/DPO3000 Agilent MSO/DSO7000A

 

Navigation by

horizontal position

 

No Automated

Search

www.tektronix.com/mso3000

Pan right and some of thedecoded text shifts; now it'snot time correlated with data.

Page 2: MSO3K Vs MSO7K

8/14/2019 MSO3K Vs MSO7K

http://slidepdf.com/reader/full/mso3k-vs-mso7k 2/2

Measurement and Channel Math

Discovering an Intermittent Pulse

Agilent MSO/DSO7000A SeriesTektronix MSO/DPO3000 Series

Digital Debug with MSO

>50,000 wfms/s maximumwaveform capture rate

Channels represented withdifferent colors.

Intensity grading shows frequency

of occurrence.

Intensity grading is preservedwhen stopped.

Variable persistence alsoavailable.

 100,000 wfms/s maximum waveformcapture rate

No intensity grading in persistencemode – whether running or stopped.

Cannot distinguish channel history

when stopped – only last acquisition.

No variable persistence.

MSO/DPO3000 Series vs. Agilent MSO/DSO7000A SeriesCompetitive Fact Sheet

Can use all the record length – up to 5 M pointsfor measurements, averaging,and math, and up to 1 M for FFT.

Advanced math with arbitrary expression.

Gate measurements by screen or cursors.

Tektronix MSO/DPO3000 Agilent MSO/DSO7000A

Uses only the 1000 displayed pointsfor measurements, averaging, FFTand math.

Limited math choices.

No gating control, other than zoom.

10/09 GB/WOW 3GW-24009-1 © 2009 Tektronix

www.tektronix.com/mso3000

Tektronix MSO/DPO3000 Agilent MSO/DSO7000A

Sample rate not dependent on number ofchannels used.

Record length not dependent on numberof channels used.

Green trace for highs (1), blue trace forlows (0).

Bus waveforms can be moved within thedisplay.

Clocked or unclocked parallel bus decode.Event table for parallel bus decode.

Sample rate cut in half when using bothpods.

Record length compromised as digitalchannels are used.

There is no color difference between alow and a high.

Bus waveforms anchored to bottom ofthe display.

No clocked parallel bus decode.

No event table for parallel bus decode.

Intermediate states showbecause bus is unclocked.

Clocked bus shows no intermediate states.