Upload
lamnga
View
223
Download
0
Embed Size (px)
Citation preview
Center for proton therapy - PSI
Pedroni Eros
Paul Scherrer Institute - Villigen PSI, Switzerland
Center for proton therapy - PSI
10.06.2011PSI,
GANTRY DESIGN AND RECENT EXPERIENCE AT PSI
2nd Workshop on Hadron Beam Therapy of Cancer
ERICE
May 23th, 2011
on behalf of the Gantry 2 group
Layout of the presentation
–Gantry 1 experience
–Facility expansion at PSI - new SC cyclotron
–The new Gantry 2 of PSI
– Layout
– Beam optics and beam size
– Energy variations with the beam
– Sweeper magnets calibration
10.06.2011PSI, Page 2
– Sweeper magnets calibration
–Advanced beam scanning techniques
–Possible future clinical use of Gantry 2
Early 90's …
GANTRY 1
USED FOR PATIENT TREATMENTS SINCE 1996
10.06.2011PSI, Page 3
GOAL in 1989
BASIC FEASIBILITY OF PENCIL BEAM SCANNING
The long term experience of using GANTRY 1
• Designed in 1991 for protons
• On the basis of the scanning experience
with pion therapy 1981-1992
with inverse planning based on CT data
1981
1992
10.06.2011PSI, Page 4
Gantry 1
1989
• Upstream scanning applied only in the dispersion plane of the beam
• Magnetic scanning started before the last bending magnet
• Eccentric mounting of patient table on the gantry front wheel
(with counter-rotation)
• Gantry radius reduce to 2 m
• Still the smallest proton gantry in the world
System characteristics of Gantry 1
10.06.2011PSI, Page 5
α rotation
φ rotationβ rotation
If i could do it again …
• Eccentric mounting as with Gantry 1 (R = 2 m)
• But with rotation only on one side (0° 180°) as with the new Gantry 2
• Permanent floor underneath the patient table
• "Treatment cell" moving with the gantry
10.06.2011PSI, Page 6
• 2 m compact eccentric gantry
A new Gantry 3 ?
Pencil beam scanning
• Small pencil beam: 3 mm σ (7-8 mm FWHM)
• Cartesian scanning (infinite SSD)
• Discrete spot scanning
• “step and shoot” ,method - on a 5 mm grid
• From 1996 until May 2008
the only scanning gantry in the world
10.06.2011PSI, Page 7
X Sweeper magnet 5 ms / step fast
Y Range shifter 30 ms average
Z Patient table 1 cm/s slow
Dose Monitor+Kicker 100 us reaction time
Elements of scanning:
the only scanning gantry in the world
Patient handling with Gantry 1
• Use of a CT outside of the treatment area
• Position next patient while treating the previous one
• Transfer of positioned patients with automated transporter
• Very useful for treating children with anesthesia
• Patient positioning control with X-ray on Gantry 1 (optional)
10.06.2011PSI, Page 8
Clinical use of GANTRY 1
• In use since1996
• Full fractionation ~30 fractions
• Treatments 8:00 16:00
• Max 19 patients/day (2.5 per hour)
• 2.8 fields/fractions in average
• 1/3 of patients are children
• Under anesthesia
• 1/3 of treatments are IMPT
Courtesy of B. Timmermann
10.06.2011PSI, Page 9
• 1/3 of treatments are IMPT
• Weak points of Gantry 1
• Table motion is part of scanning
• Not possible to use collimators
• Not possible to apply repainting
• We treat only
• non moving targets!
Advantages of using scanning (vs. scattering)
• Automated treatments – all by software• Minimal simple equipment
• “The (pencil) beam "does it all"
• Same approach from small to large fields• No individualized hardware (no fabrication of collimators compensators )
• Apply dose fields in sequence without personnel entering the treatment room
• Simplify treatment - reduce treatment time and costs
• True 3-dimensional conformation
10.06.2011PSI, Page 10
• True 3-dimensional conformation
• Variable modulation of the range • Minimal neutron background (for the patient)
• Lower risk of second cancer than scattering (depends on sophistication)
• Important for pediatric treatments• Less activation of equipment in the nozzle (for the personnel)
• Less activation of the accelerator
• Better lateral fall-off as compared to scattering (?)
• No material in the beam path (except air and monitors) - Less MCS effects
Delivery of IMPT and IGPT
• Dose shaping within the target
• IMPT (intensity modulated PT)
• The term of comparison with IMRT (conventional therapy)
• Biological targeting (image guided proton therapy)
• Intentional non-homogeneous dose distributions
• Dose proportional to
the tumor activity
10.06.2011PSI, Page 11
the tumor activity
(biological signal)
• The topic for the
future…
Courtesy of A. Lomax PSI
Major disadvantage: the organ motion sensitivity
• Interference of organ motion with the scanning sequence
• Disturbance of the dose homogeneity within the target
• With Gantry 1 we treat only immobilized lesions attached to bony structures
• Tumors in the head, spinal chord and low pelvis
• We accept only movements < + -2 mm
• for treatments with full fractionation!
• Can we overcome this drawback?
The point of scanning to be improved
M.Phillips PSI 1992
10.06.2011PSI, Page 12
• Can we overcome this drawback?
• Possible remedies :
• Fast Scanning (Repainting)
• Gating
• Tracking?
• or a combination of those
• … the possible points to be developed
with Gantry 2…
2000
EXPANSION OF THE PROTON FACILITY AT PSI
SUPERCONDUCTING CYCLOTRON
FAST DEGRADER
LAMINATED BEAM LINES
10.06.2011PSI, Page 13
LAMINATED BEAM LINES
GOALS:
Stable DC beam
Modulation of the beam intensity at the ion source
Very fast energy changes
Layout of the newly expanded proton facility
• COMET - dedicated superconducting cyclotron [ACCEL - design H. Blosser]
• Beam for Gantry 1 all the year through
• patients treatments restarted in February 2007 no shut-downs since August 07
• Horizontal beam line for OPTIS 2 transfer from OPTIS 1 last year
• Next generation scanning gantry : Gantry 2 1. patient planned for 2012
Exp. Area (PIF)Medical cyclotron
10.06.2011PSI, Page 14
Gantry 2
Gantry 1
OPTIS 2
Disconnected from
PSI ring cyclotron in 2006
Facility specifications … mainly for the new Gantry 2
• Super-conducting cyclotron
• Very stable beam at the ion source
• Aiming at 2-3% at the 100-200 µs scale
• Deflector plate in the first orbit
• Dynamic control of the beam intensity
• 100-200 µs time scale
• Fast degrader (moving carbon wedges)
10.06.2011PSI, Page 15
• Fast degrader (moving carbon wedges)
• Continuous choice of the beam energy
• Beam line with laminated magnets
• Providing fast changes of beam energy
Excellent beam current stability … an example
• Change of paradigm of testing the beam monitors
• Use the inherent stability of the beam to check the fine
tuning of the electronics (capacitive matching)
• Result: improved linearity of the dose
• < 0.5% down to a 0.1 Gy dose
• Ideal beam for scanning 100% duty factor• 33
nF600 µs
10.06.2011PSI, Page 16
Dose measurements for diff. Plugin configurations
0.975
0.98
0.985
0.99
0.995
1
1.005
1.01
0 20 40 60 80 100 120
applied Dose [cGy]
mea
sure
d D
ose
/mea
sure
d D
ose
@ 1
Gy
ISO, 22pf
ISO, 47 pf
ISO, 33 pf
ISO, 39 pf
• 22
nF
nF600 µs
2005
A NEW PSI GANTRY - GANTRY 2
10.06.2011PSI, Page 17
MAIN GOAL
DEVELOP FURTHER SCANNING
AS A UNIVERSAL BEAM DELIVERY TECHNIQUE
GANTRY 2
TOPIC 1: INNOVATIVE GANTRY LAYOUT
10.06.2011PSI, Page 18
READINESS FOR
IMAGE GUIDED PROTON THERAPY
EFFICIENT PATIENT HANDLING
Beam Line
Support
Bearing axle
From -30°
to +180°
Design started from the patient table…
10.06.2011PSI, Page 19
Patient table
Room withfixed floor
0°to +180°would have been a better choice (cheaper) …
Layout of the Gantry 2 room: patient table, compact nozzle
Small compact nozzle
Same patient table as
at RPTC in Munich
(Schär Engineering)
10.06.2011PSI, Page 20
• Easy access to the patient table on fixed floor
• Fixed walls and ceiling for mounting commercial equipment (Vision RT?)
A system open on
both sides
- lateral and front -
Optimal for using an
in-room sliding CT
In-room sliding CT - within reach with the patient table
The beginning of
IGPT ?
10.06.2011PSI, Page 21
• Patient positioning (tumor in soft tissues region)
• Setup for treating moving targets - 4D CT (relation external gate - internal motion)
• BEV imaging - equivalent to portal imaging with photons• Very large field-of-view (26 cm x 16 cm)
• not masked by equipment or collimators in the beam path• QA control of gating and tracking
(scanning + pulsed X-rays)• Fluoroscopy mode
• Beam guidance?
BEV X-ray - synchronized with proton beam delivery
Scan and bend
IGPT
10.06.2011PSI, Page 22
a) compact gantry b) long throw gantry
Sweepers
X rays tubeProton beam
Bending
magnet
nozzle
Yoke hole
Patient
Imager
Sweeper
or
Scatterer
Collimator
UPSTREAM SCANNING
Bend and scan
BEV - imager
• Check patient
position at the
isocenter
10.06.2011PSI, Page 23
• Photograph of the retractable arm for holding the X-ray panel
• behind the patient on the side opposite to the nozzle (BEV X-ray).
GANTRY 2
TOPIC 2: PENCIL BEAM AND BEAM OPTICS
10.06.2011PSI, Page 24
GOALS:
SMALL PENCIL BEAM FOR PRECISION THERAPY
PARALLELISM OF SCANNING
GANTRY 2 beam optics
Q1
Q3
Q4Q5
Q6
Q7
A1
A2 A3
WU
WT
M1 M2
M3 P
P
S
S
3.2 m
Q2
QC
X
10.06.2011PSI, Page 25
• Rotational symmetric (large) phase space +- 3mm +- 10 mrad +- 0.5% dp/p
• Complete achromatism
• Scanning-invariant beam focus
• Orthogonal (x-,y-) focal planes in T and U
• Focus to focus with 1:1 imaging from the coupling point of the gantry to the iso-center
• 2D- parallelism of scanning (in T and U) (on Gantry 1 only U)
SPECIFICATIONS
A3A2A1QC
WTWU
H S Double parallel
scanning
Point to point
GANTRY 2 beam optics (TRANSPORT and RAYTRACE)
10.06.2011PSI, Page 26
TRANSPORT beam envelopes through the Gantry 2
Q2Q1 Q3 Q4 Q5Q6Q7
focus
Achromatic
Double parallel
scanning
Advantages of coupling the gantry at a small beam focus
• At a small beam focus the spot image is insensitive to scattering
• Decouple vacuum (gantry from facility)
• Collimate the beam at the coupling point
• Forced beam centering on axis
• Measure the beam characteristics on-line before the gantry
• On-line independently of Gantry rotation and scanning
• Manipulate the beam shape imaged on the scanning beam
10.06.2011PSI, Page 27
• Manipulate the beam shape imaged on the scanning beam
• Large Gaussian beam shape ?
• Virtual dynamic collimation at coupling point ?
• Disadvantage
• A focus coupling needs a longer beam line coupling with a parallel beam
Advantage of using parallel scanning
• Less skin dose
• Simplify treatment planning
• No inverse squared distance effects
• Simplify dosimetry and QA control
• Dose value and dose distribution in the patient and in the phantoms are “the same”
• Easier field patching (expansion of the scan range for treating long targets)
• Can be done without optimization within treatment planning
10.06.2011PSI, Page 28
• Just exchange magnetic scan position with patient table offset
• Compensators (simulated scattering)
• No dose errors due to inverse R-square effects
• Collimators
• No tapered faces necessary
A big beast - 34 tons
15 cm gap
To accommodate
a scan area of
20 x 12 cm
Lamination leaks …
The difficult piece: the last 90° bending magnet
10.06.2011PSI, Page 29
Photograph of the mounting of the 90° bending magnet.
Integrated vacuum chamber embracing the poles of the magnet
180 tons … mechanical isocenter within + - 0.4 mm
Lamination leaks …
but in the end
Vacuum problems
well solved !
• Vacuum “up to the patient”
• Sharp pencil beam - 3 mm sigma
• Two monitors and a strip monitor
• 2 mm strips (TERA collaboration)
• Removable pre-absorber
• IN and OUT of beam
• For ranges below 4 cm
Compact optimized nozzle
10.06.2011PSI, Page 30
• For ranges below 4 cm
• Telescopic motion of the nozzle
• To reduce air gap (keep patient at isocenter)
• Option to add collimator (compensator)
• To shield OAR on top of scanning
• To simulate passive scattering with a scanning beam
• Collision protection to treat patients remotely (multiple
fields in one go)
Results: pencil beam size (on axis)
• Minimize material in the nozzle for keeping To have a sharp lateral fall-off
the beam size between 3 and 4 mm sigma at all energies (100-200 MeV)
0.5
0.6
Adding piece by piecethe materials in the nozzle
Region 70 - 100 MeV
10.06.2011PSI, Page 31
0
0.1
0.2
0.3
0.4
0 50 100
GANTRY 2
TOPIC 3 :
ENERGY VARIATIONS WITH THE BEAM LINE
10.06.2011PSI, Page 32
COMPENSATED INTENSITY-ENERGY LOSSES
VERY FAST ENERGY CHANGES
SC cyclotron
Degrader
Beam analysis
Gantry 2
Compensation of beam losses in the range 100-200 MeV
200-400 nA Constant ratio
10.06.2011PSI, Page 33
Collimators forIntensity suppression
Coupling point
Rate Monitor 1
0
100
200
300
400
500
600
100 120 140 160 180 200
Energy [MeV]
kHz
0.2-0.4 nA
Goal : reserve the use of the deflector plate for intensity modulated dose painting
Optimization for maximum transmission at 100 MeV
10.06.2011PSI, Page 34
Beam envelopes at 100 MeV.
Horizontal (red) and vertical (blue) envelopes
Green line = transmitted beam intensity Dots beam sizes calculated with TURTLE
3 4
Intentional beam defocusing at higher energies
10.06.2011PSI, Page 35
1 2
Beam envelopes at 200 MeV.
yellow we four beam line sections.
Beam defocusing at collimators (black bars)
Same intensity losses as at 100 MeV
• Energy loop Red: up-down Blue: down-up
Taking into account hysteresis effects …(70-230 MeV)
Uncorrected
Corrected
1 mm
10.06.2011PSI, Page 36
Corrected
• But - we observe
• Slow change of lateral beam position after a
big energy change in the range 230-70 MeV
• 1-3 mm with exponential decay
with decay time of ~2 s
• < 0.3 mm position change after small energy
changes within the SOBP
With very fast energy changes …. 80 ms
10.06.2011PSI, Page 37
Scintillator block
the beam of Gantry 2 seen with a
TV camera
• We plan to correct these shift with sweeper
offsets as a function of the time after the last
energy change
• Strategy:
• Fixed targets - range precision of < 0.5 mm
while working with full ramp 70-230 MeV
• Moving targets - repainting - precision ~ 1mm
• repaint only the SOBP up and down
GANTRY 2
TOPIC 4 : SWEEPER MAGNETS
10.06.2011PSI, Page 38
COMMISSIONING ISSUES
NON LINEARITIES OF THE SWEEPERS
−8
−6
−4
−2
0
2
4
6
8210 MeV
T (
cm)
−6
−4
−2
0
2
4
6
8150 MeV
T (
cm)
210 MEV 150 MEV
Need of a very precise mapping of the sweeper's action
10.06.2011PSI, Page 39
−15 −10 −5 0 5 10 15−8
U (cm) −15 −10 −5 0 5 10 15−8
U (cm)
−15 −10 −5 0 5 10 15−8
−6
−4
−2
0
2
4
6
880 MeV
U (cm)
T (
cm)
Measured (red) and calculated (blue) spot maps of
the (linear input) action of the sweeper
magnets on the scanned beam position.
The non linearities are due to a curvature
of the effective boundary of the magnetic field
of the 90° bending magnet which is
changing with energy (changing sign)
80 MEV
100 MeV
U
T
QMF1: 0% QMF1: -10% QMF1: -20%
Invariant beam focusing … needs some "tricks"
First tests in 2008
10.06.2011PSI, Page 40
• Anticipated by Raytrace calculations
• Tapering of the poles surface of the U
sweeper
• Quadrupole corrector QMFC switched
in series with the U-Sweeper
• Results:
acceptable scanning-invariant focus
QMFC in series
70 MEV 120 MEV
After a proper mapping of the sweepers
10.06.2011PSI, Page 41
Beam spots (2 cm steps) at the isocenter covering the scan region of 12 cm x 20 cm
170 MEV 220 MEV
GANTRY 2
Topic 5 : the main goal of Gantry 2
NEW ADVANCED BEAM DELIVERY TECHNIQUES
10.06.2011PSI, Page 42
TO PROMOTE SCANNING AS A UNIVERSAL BEAM DELIVERY
TECHNIQUE
• For reducing organ motion errors
• Goal - Fast painting with volumetric repainting
• Painting lines
• < 10 ms per line (10cm + line change)
Aiming for highest scanning speed From spots
To lines
10.06.2011PSI, Page 43
• < 10 ms per line (10cm + line change)
• Painting energy layers
• 200 ms per plane (20 lines x 5 mm)
• Change of energy (100 ms - 5mm range)
• Painting of volume
• 6 s per liter (20 energies at 5mm steps)
• Volumetric repainting capability (aim)
• 10-20 repaintings / liter in 2 minutes
To contours
Conformation requires non-homogenous proton fluences
3d-Dosehomogeneous
U
T
depth
10.06.2011PSI, Page 44
Proton fluenceof energy layerisnon homogeneous
Repainted for coping with organ motion
Time driven beam delivery Beam path downloaded as tables of the sweepers U - T and beam intensity IDose control with feed-back loop:
Monitor 1 required dose -> vertical deflector plate
A) If sweeper speed is the limituse variable intensity (IM)
FPGA based delivery of tabulated energy layers
0.2 0.7 10
500
1000
1500
2000
Required Dose
Del
iver
ed M
Us
Dose linearity of simple T-lines
10.06.2011PSI, Page 45
B) If beam dose rate is the limituse variable sweeper speed (SSM)
C) or use both !
Full range of dynamic dose control
from zero IM
to any dose SSM
23 times
Max T speed
Variable intensity
Required Dose
5ms - 10 cm lines - painted with IM
GANTRY 2
POSSIBLE CLINICAL USE OF GANTRY 2
10.06.2011PSI, Page 46
POSSIBLE CLINICAL USE OF GANTRY 2
GOAL:
PROMOTE SCANNING AS A UNIVERSAL BEAM DELIVERY
TECHNIQUE
to completely replace scattering
• Often required
• sharp lateral fall-off at the boundary between tumor and sensitive structure
• Scanning alone better at high energy (edge enhancement with pencil beam)
• Scanning with added collimation better at low energy (beam size limitations)
• Scanning with varying beam energy superior to scattering ?… less material in the beam path
Idealized scattering
(zero phase space)
FWHMmm
Prostate
Brain stem
Improve lateral fall-off for treating static targets
10.06.2011PSI, Page 47
Factor 1.4 gain
Difference Gauss to
error-function (1.7)
Scanning with
collimation betterScanning alone better
(zero phase space)
Realistic scanning
Rangemm
Dorsal irradiation
Very big tumors
• Combine the use of fast scanning with patient table displacements
• Needs remote control of the patient table (collision detection)
• Take advantage of the parallelism of the beam
• trivial patching - shift table - and apply intensity filter to spot pattern
Medulloblastoma
10.06.2011PSI, Page 48
Example: a paediatric case
treated in 2004 with Gantry 1
COMPARE
PancreasRectumBreast
• Discrete spot scanning applied with iso-layered repainting (max dose per spot visit)
• Treatments in the trunk (pancreas, cervix, colon), breast, lymph nodes, etc.
• Advantage of having fast energy changes -> Volumetric repainting
• FAST SCANNING IS MORE THAN JUST REPEATING…
Moderately moving targets (0-5 mm)
10.06.2011PSI, Page 49
• From
• Simulation of repainting strategies …
• S. Zenklusen's thesis work
• PMB 55 (2010) 5103-5121
COMPARE
G1 spots, scaled
G2 spots, scaled
G2 spots, iso-layer
G2 lines, scaled
Largely moving targets
• Volume painting within a single breath hold” ? (repeated)
• Using Conformal line painting
• Speed of painting 0.2 Gy of a sphere of (17 layers) – ½ liter in 7 secs
• If not working, Gating
Lung liver
10.06.2011PSI, Page 50
7s
• Dose shaping with compensators
• Uniform scanning
• Energy layer with homogeneous fluence
• Magnetic line scanning at max. speed
• Very high repainting number
• Collimator optional
Simulated scattering
10.06.2011PSI, Page 51
• As a sub-mode of conformal scanning
• To simulate scattering on a scanning-gantry
• To replace completely scattering ??
Retinoblastoma ?
• 8 cm diameter sphere
Feasibility demonstration of "simulated scattering"
• Field shape = target projection
• minimal neutron background
• Variable modulation of the range
10.06.2011PSI, Page 52
• Distal layer: repainted 48 times in 30s
• Variable modulation of the range
• Layer shrinking
• Parallel beam
• no compensator dose errors
• Highly repainted
• From S. Zenklusen PhD - Medical Physics 2011
Status of the Gantry 2 project
• Significant project delays due to the
• Priority given to OPTIS 2 in 2008
• Difficulties with the vacuum of the 90° bending
magnet
• PSI internal development of the software for the
control of the patient table … insufficient resources
10.06.2011PSI, Page 53
THANK YOU