Installation user’s manual version 4.3 P-WS LOAD 24 Vdc 2004/108/EC (EMC) - 2006/95/EC (LVD) INSTRUMENT IDENTIFICATION ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH ENGLISH

PWS CARICO 24Vdc_en.pdf

Embed Size (px)

Citation preview

  • Installation users manual

    version 4.3

    P-WS LOAD 24 Vdc

    2004/108/EC (EMC) - 2006/95/EC (LVD)

    INSTRUMENT IDENTIFICATION

    ENGLISH ENGLISH ENGLISH ENGLISH

    ENGLISH ENGLISH ENGLISH ENGLISH

  • KEY

    The following symbols are used in the manual to draw the readers attention to important points:

    Caution! Risk of electric shock! Caution! This operation must be performed by specialist personnel.

    Pay particular attention to the following points.

    Further information.

    GUARANTEE 24 months from the delivery document date. The guarantee covers only defected parts and includes the replacement parts and labour. All shipping and packing costs are paid by the customer. It is possible to have the repair in guarantee on condition that the returned product has not been transformed, damaged or repaired without authorization. No guarantee is applicable on returned products without the original label and/or serial number. No guarantee against misuse. Batteries: Laumas provides 1 year guarantee from the date of delivery note, against material defects or battery manufacturing faults. Disposal of Waste Equipment by Users in Private Households in the European Union

    This symbol on the product or on its packaging indicates that this product must not be disposed of with your other household waste. Instead, it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for the recycling of waste electrical and electronic equipment. The separate collection and recycling of your waste equipment at the time of disposal will help preserve natural resources and protect human health and the environment. For more information about where you can drop off your waste equipment for recycling, please contact your local waste disposal Authority or the equipment retailer.

  • TABLE OF CONTENTS RECOMMENDATIONS FOR PROPER INSTALLATION............................................................................... 1

    TECHNICAL CHARACTERISTICS ................................................................................................................. 2

    ELECTRICAL CONNECTIONS ....................................................................................................................... 3

    OPERATION .................................................................................................................................................... 4

    STARTING UP THE INSTRUMENT ................................................................................................................ 4

    TARE ZERO-SETTING .................................................................................................................................... 5

    PROCEDURE FOR AN INSTRUMENT THAT HAS ALREADY BEEN CALIBRATED.................................. 5

    PROCEDURE FOR AN INSTRUMENT THAT HAS NOT YET BEEN CALIBRATED.................................... 6

    PROGRAMMING FORMULAS AND PARAMETERS.............................................................................. 7 - 10

    BATCHING.............................................................................................................................................. 11 - 12

    ALARMS......................................................................................................................................................... 13

    CONSUMPTION............................................................................................................................................. 14

    PRINTING....................................................................................................................................................... 15

    EXAMPLES OF PRINTOUTS ................................................................................................................. 16 - 17

    PROGRAMMING SECRET SYSTEM PARAMETERS....................................................................... 18 - 19

  • - 1 -

    RECOMMENDATIONS FOR CORRECT INSTALLATION OF WEIGHT INDICATORS

    - The entry into the cable board of cells must be independent ( on one side or the other of the

    board ) and directly connected to the terminal board of the device without breaking by bearing terminal boards or passing through troughs containing other cables.

    - Use the " RC " filters on the instrument-driven coils of the remote control switches. - Avoid inverter, if inevitable, use filters and separate with sheets. - The installer of the board is responsible for securing the electrical safety of the indicators. - It is a good norm to let the indicators always switch on to prevent the formation of condensation.

  • - 2 -

    TECHNICAL FEATURES Weight indicator with 4 output relays and 2 inputs, DIN container ( dimensions: 96 x 96 mm, depth 65 mm, hole 91 x 91 mm ).

    The instrument is able to read up to 19,999 divisions, but after exceeding the value 9,999 the weight reading restarts from zero and blinks to indicate that the value has been exceeded.

    POWER SUPPLY 24VDC POWER CONSUMPTION 5 VA FRONT PANEL PROTECTION IP64 DISPLAY semi-alphanumeric display 4 digits, 20 mm in seven segments LOAD CELL CONNECTIONS maximum 4 load cells 350 ohm LOAD CELL SUPPLY 5 VDC / 60 mA INTERNAL DIVISIONS 20000 DISPLAY RANGE -999 ; +19999 MEASURE RANGE - 4 mV + 16.5 mV READING RESOLUTION x 1 x 2 x 5 CONVERSION RATE 10 readings / sec. LOGICAL OUTPUTS voltage-free contacts max. 115Vca 2A LOGICAL INPUTS n. 2 HUMIDITY (condensate free) max 90 % STORAGE TEMPERATURE -20 + 70 C OPERATING TEMPERATURE -10 + 50 C ZERO-SETTING and CALIBRATION from keyboard START and STOP BATCHING from keyboard or external contact PRINT 24 columns serial TTL printer UNIT OF MEASURE kg or t FORMULAS max 12 formulas from keyboard with SET and PRESET OPTIONS on request: EC : 12 formulas selection from external commutator. E : 12 formulas selection from external contacts.

  • - 3 -

    ELECTRICAL CONNECTIONS

    CAUTION: The following procedures must be performed by specialist personnel. All connections must be made with the instrument turned off. Connect the load cells by means of a terminal board in a water-proof junction box, by connecting the leads having the same colour; perform the same operation for the shield keeping them isolated from the ground or metallic parts connected with the ground. By means of a shielded cable with a section not lower than 0,5 mmq connect the terminal board to the P-WS instrument. The routing shall be far away from the power cables and possibly protected by a metal pipe.

    11 1213 14 15 16 17 18

    Exc

    . +E

    xc. -

    S +

    S -

    1 1211

    10

    98

    765

    4

    3

    2

    OPT. ECS

    TOP

    / ZE

    RO

    STA

    RT

    LOAD CELLS

    N 3 4 5 6 7 8 9 10L

    SCRAP PRESET SET CYCLE END

    230 VAC

    Connect the shield to earth

    - + 24VDC

  • - 4 -

    OPERATION By closing the start contact or by pressing the DOSA key the instrument starts the batching and verifies that the weight is lower than the minimum weight programmed, it executes the auto-tare ( if it is enabled ) then closes the set and preset contacts. When the weight has reached the preset value the related contact will be opened, once it has reached the set value less the fall value the set contact will be opened and after the pause time is over (999,9 sec.) memorizes the consumption and closes the cycle end. When the weight has reached the minimum weight (unloading phase) and after the safe emptying time is over (999,9 sec.) the instrument will open the cycle end contact. If more than one cycle has been programmed, the instrument will automatically carry on. STARTING UP THE INSTRUMENT Switch on the instrument and wait about 5 minutes for all the components to reach a steady temperature. Check that the displayed weight is positive and that it increases when a weight force is applied to the container. If a negative value is displayed, check the load cell connections and ensure that the cells are correctly positioned (mounting direction). Also check that the container is correctly installed (tube position, sleeves, mounting accessories, etc.).

  • - 5 -

    TARE ZERO-SETTING: Make sure that the container is empty, then press and the display will show tArA. Hold down the key and at the same time press ENTER to set the tare to zero.

    INSTRUMENT THAT HAS ALREADY BEEN CALIBRATED

    Laboratory setting: CELL ( ); nU-U ( ); rISO ( ) In this case the instrument has already been calibrated, proceed to Instrument calibration check.

    INSTRUMENT CALIBRATION CHECK:

    Make sure that the container is empty and that the instrument displays zero. Place a significant quantity of product in the container (at least 50% of the maximum quantity of product that is to be weighed but in any case less than 9,999) and make sure that the instrument displays the correct value. - If the displayed value is different from the weight known to be in the container ( bigger than 1-

    2% ), make sure that this is not due to mechanical causes and again check the electrical connections and the direction in which the load cells are mounted.

    - If the difference in weight is negligible (about 1-2%), then proceed to Correcting displayed weight from keyboard.

    CORRECTING DISPLAYED WEIGHT FROM KEYBOARD:

    Procedure subject to entry of access password ( see item PASS on page 19).

    When the weight is displayed, press . The display will show CAL l. Hold down the key

    and press ENTER. The display will show the weight value blinking. Use the and keys to correct it and set the value of the real quantity loaded in the container. Press ENTER to confirm. The display will show CAL l for an instant, then the correct value will appear.

    If the display shows ErrO ( error ), this means that incorrect values have been set for the parameters CELL and nU-U ( calibration has been tampered with ). Repeat the calibration (see Theoretical calibration) restoring the laboratory setting.

  • - 6 -

    INSTRUMENT THAT HAS NOT YET BEEN CALIBRATED In this case the instrument has not yet been calibrated, proceed to Theoretical calibration, Tare zero-setting and Calibration check.

    THEORETICAL CALIBRATION :

    Procedure subject to entry of access password ( see item PASS on page 19). Switch off the instrument, then switch it back on again while holding down the DOSA key. The display will show C.O.S.c., press twice and the display will show:

    dECP , press ENTER and set the number of decimals (max. 3 decimals) using the and keys. Press ENTER to confirm and the display will show:

    nU-U , press ENTER and set the load cell sensitivity (expressed in mV/V) using the and keys. Press ENTER to confirm. The display will now show:

    unit , press ENTER and set the unit of measurement using the and keys: 0 = kg; 1 = t. Press ENTER to confirm. The display will now show PASS.

    Press DOSA to exit. Press MENU and then press several times until the display will show:

    CELL , press ENTER and set 10% of the systems full scale using the and keys (for example with 3 load cells of 100 kg ; full scale = 100 x 3 = 300.0 kg; 10% of 300.0 kg = 30.0 kg). Press ENTER to confirm. The display will show: rISO , press ENTER and set the system resolution: 1 = resolution 1 2 = resolution 2 (the last digit of the weight moves in steps of two: 0, 2, 4, 6 . . . . ). 3 = resolution 5 (the last digit of the weight moves in steps of five: 0, 5, 10, 15 . . . ).

    Press ENTER to confirm and return to the weight reading. After Theoretical calibration proceed as follows: - SET THE TARE TO ZERO (see Tare zero-setting). - CHECK THE INSTRUMENT CALIBRATION ( see Instrument calibration check ). - If necessary, CORRECT THE DISPLAYED WEIGHT (see Correcting displayed weight from keyboard).

    If the display shows ErrO ( error ), this means that incorrect values have been set for the parameters CELL and nU-U. Check them and if necessary, repeat the Calibration.

  • - 7 -

    PROGRAMMING FORMULAS AND PARAMETERS

    Press and to switch from one parameter to another without changing them. Carry out programming as follows: Press MENU while the weight is displayed. The corresponding LED will light up and the display will show: F l , programming formulas. Confirm by pressing ENTER, the LED R2 will light UP and the

    display will show P.SEt for 1 second followed by the PRESET value. Use the and keys to change the value, then press ENTER to confirm, the LED R3 will light up and the display will show

    SEt for 1 second followed by the SET value. Use the and keys to change the value, then

    press ENTER to confirm. The display will show F 2 (formula 2), by using the and the keys select the desired formula ( from F1 to F12 ) and press ENTER, you will now be able to program the SET and PRESET values. DELETING A FORMULA: By pressing the ENTER key for 3 seconds in any point of a formula, the CAnC. message will be displayed for 1 sec. and the formula will be cancelled.

    If in the "secret parameters" PASS has been set to zero (password enabled), access will be denied to parameters programming. By confirming the SET value of F12, the display will show F 1 again. Press MENU to exit. If in the "secret parameters" PASS has been set to 1 (password disabled) you will enter parameters programming and the display will show:

    P._ _ _ , minimum weight expressed in the unit of measure of the system. The container is considered empty if the weight is lower than this value. If the weight into the container is higher than this value the instrument doesn't allow the batching and displays the tArA message.

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: t.SIC , safe empty time ( max 999,9 seconds ); Time expressed in seconds which you have to give to the system during the unloading process after reaching the minimum weight and before opening the cycle-end contact (thus stopping the batching) to obtain a complete emptying of the container.

  • - 8 -

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: P. , maximum weight that the container can load. The operator can not program a formula whose addition of the products exceeds this value.

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: PAUS , pause time ( max 999,9 seconds ). This is the time elapsed between the conclusion of the batching of a product and the closing of CYCLE-END contact to allow for a steady weight. The instrument waits this time only if it must calculate the automatic fall or if the totals are enabled.

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: vo. , fall ( product in " fall " at the stop of the extractor ) expressed in the unit of measure of the system. When the weight reaches the value programmed in formula less the value programmed in this constant, the batching will be stopped. Warning: For each formula a different fall value will be programmed. If in the " secret " parameters the MANUAL FALL is enabled (o.VOL = 0 ):

    Press ENTER and the display will show F l for 1 second and then the fall value programmed for

    the formula1, change it with the and the keys and confirm by pressing ENTER. The display will show F 2 for 1 second and then the fall value programmed for the formula2, change

    it with the and the keys and confirm by pressing ENTER. So on for all the 12 formulas. If in the " secret " parameters the AUTOMATIC FALL is programmed (o.VOL = 1 - 99 ):

    The instrument will calculate on itself the fall value. If the power break down this value will be lost, when the power is fed again the instrument will calculate the fall value after the first batching ( until the weight batching will be included into the limit programmed at the voice divin the Secret Parameters ). AUTOMATIC FALL value visualization and memorization: Confirm by pressing ENTER , the display will show F l for 1 sec. and then the fall value memorized for the formula 1, press the DOSA key and the fall value for the formula1 calculated by the system will appear ( blinking ). Press ENTER if one wants to confirm the value, otherwise press DOSA to return to visualize the previous value. Confirm by pressing ENTER and the display will show F 2 for 1 sec. and so on for all the 12 formulas. Even if the instrument would be switched off, when the fall value calculated by the system has been memorized, the memorized dates will be maintained.

    Press ENTER on the fall value of F12, the following will appear:

  • - 9 -

    t.nC , time ( max 9,9 seconds ) that the system waits in batching after the PRESET value has been reached, before to compare the weight into the container with the SET value programmed ( usually program: 0,0 ).

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: LE.on , time expressed in seconds ( max 999,9 sec.). Time in which the SET contact remains closed after the PRESET contact has been opened ( fast batching ). Together with the LE.oF time it allow to activate an "tapping". Set 0 to inactivate this function.

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: LE.oF , time expressed in seconds ( max 999,9 sec.). Time in which the SET contact remains opened after the PRESET contact has been opened. Set 0 to inactivate this function.

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: A.tAr , auto-tare, one can enable the weight zero-setting at the beginning of the batching. The instrument set the weight to zero only if the weight is lower than the minimum weight programmed ( P._ _ _ ). Press ENTER and the following will appear: 0 = auto-tare disabled; 1 - 99 = auto-tare enabled every 1 - 99 batchings ( if one programs the value 1 the auto-tare will be executed every batching, if one programs 2 the auto-tare will be executed every two batchings, so on until max 99 batchings ).

    Warning: If the auto-tare is enabled, at the beginning of the batching the system waits for 2 seconds before executing the auto-tare.

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show: tOLL , tolerance ( expressed in the unit of measure of the system ). If at the end of the batching of the SET the difference between the weight reached and the SET value programmed exceeds the value programmed in this constant, the SCRAP contact will be closed for 3 seconds. Set 0 to inactivate this function.

    Press ENTER. Use the and keys to change the value, then press ENTER to confirm, the display will show:

  • - 10 -

    StA , print. Press the ENTER key and select by using the and the keys: 1 = print enabled ( at the end of every cycle the batching data with date and time will be printed ). 0 = print disabled. Press ENTER and the display will show: FILt , filter ( weight oscillation damper ). To reduce the weight oscillations increase the value.

    Press ENTER, set a value included between 0 and 9 seconds by using the and the keys. Confirm by pressing ENTER and the display will show: CELL , DO NOT CHANGE THIS VALUE ( if you do it the calibration of the system will be modified ) Press ENTER and the display will show: Program: ( ). Press ENTER to confirm and the display will show: rISO , DO NOT CHANGE THIS VALUE Program: ( ). Press ENTER, then press MENU to exit.

  • - 11 -

    BATCHING Press the DOSA key, the display will show F l, select the formula that one wants to execute

    ( from 1 to 12 ) by using the and the keys and confirm by pressing ENTER. The n.c. ( number of cycles ) message will appear, then the value; it is possible to program the number of the

    time one wants to re-execute the formula ( from 1 to 9999 ) by using the and the keys, then press ENTER. The instrument verifies the following conditions : - the formula must be programmed, otherwise the F._ _ _ blinking message will appear, press DOSA to exit ; - only if the auto-tare is enabled : the container must be " in tare " ( the weight present into the container must be lower than the minimum weight programmed, otherwise the tArA blinking message will appear. Press DOSA to exit ). Once all the conditions are met, the instrument starts the batching: if the auto-tare is enabled, at the beginning of the batching the system waits for about 2 seconds before executing the auto-tare , the SET and PRESET contacts will be closed contemporarly. Once the PRESET value has been reached the instrument opens the PRESET contact and if the LE.on and LE.oF values are programmed the "tapping" phase will start ( the SET contact will be opened and closed with the LE.on and LE.oF times ). Once the value programmed less the fall value has been reached, the instrument opens the SET contact and closes the cycle-end contact, it will be re-opened at the reaching of the minimum weight and after the safe emptying time is over. If the automatic fall or the consumption are enabled, after the SET opening, the system waits the pause time and then closes the cycle-end. Automatic fall: The system calculates the automatic fall every 1 - 99 batchings, until the weight batching will be included into the limit programmed at the voice div in the Secret Parameters. If a tolerance value different from zero has been programmed in the parameters and the difference between the SET value programmed and the reached weight is higher than the tolerance value, the SCRAP contact will be closed for 3 seconds. If the printer is enabled the batching data will be printed reporting date and time of the batching. Now if more than one batching cycle has been programmed, the instrument will carry on in automatic.

  • - 12 -

    KEYS FUNCTION DURING THE BATCHING ( batching interruption ): - By pressing the DOSA key, the HALt blinking message will appear and the batching will be temporarily interrupted. Press the ENTER key to start again the batching, by pressing the DOSA key again you can stop the batching definitely. - By pressing the MENU key the formula and cycles numbers will be displayed for 1 second. - By closing the STOP external contact the batching will be stopped. BATCHING FROM EXTERNAL CONTACT : Select the formula that one wants to batch from commutator ( option EC ) or from external contact ( option E ). Close the START contact for about 2 sec. ( the input 2 led will light up ), the number of the formula which will be executed will appear for 1 second. The instrument will verify the following conditions: formula must be programmed, container must be in tare , then the batching will start like show for the batching from the keyboard. RETURN FROM A BLACK-OUT: In case of a power supply black-out during the batching ( including the unloading phase, the CYCLE END contact closed ), the instrument displays the bLoC blinking message when the power supply will be restored. Press ENTER to start again the batching at the point in which it was interrupted, otherwise if you press the DOSA key for two times or if you close the STOP contact the batching will be interrupted and the weight will be displayed. In case of several batching cycles have been programmed from the keyboard, the power supply black-out will cancel the cycles which have still to be executed. Thus the operator will have to program the lost cycles

  • - 13 -

    ALARMS tArA ( blinking ), it indicates that the weight into the container is bigger than the minimum weight programmed in the parameters. Bringing the container back to the " in tare " status to restart the batch, otherwise press DOSA to exit. F. _ _ _ ( blinking ), it appears if at the start of a batching the formula recalled is not programmed. Press DOSA to exit. P. ( blinking ), it appears when the called formula has a weight total higher than the maximum weight programmed in the parameters. Press DOSA to exit. HALt ( blinking ), it means that during the batching the DOSA key has been pressed temporarily interrupting the cycle. Press ENTER to re-start it, otherwise press DOSA to interrupt it definitely. bLoC it indicates a power failure during the batching, press DOSA to cancel the batching, otherwise press ENTER to start the batching again from the point in which it was when the power failure has happened.

  • - 14 -

    CONSUMPTION

    PLEASE NOTE : IT IS POSSIBLE TO DISPLAY THE CONSUMPTION ONLY IF THEY ARE ENABLED IN THE " SECRET " PARAMETERS.

    Press the ENTER key for 3 seconds, the display will show tot for 2 seconds, then P l for 1 second and then the ' consumed ' quantity of product. The instrument can memorize until to 9999 divisions for every product. If this value is passed, the consumption will automatically be reset; when the product consumption will reach 9000 divisions, the related led will start blinking during the weight displaying phase. PRINTING:

    During the consumption display phase press the key to print the consumption with date and time, then the system will come back to the weight display. DELETING: During the consumption displaying phase press the ENTER key for 3 seconds, the CAnC message will appear for 3 second, confirm with ENTER again to cancel the consumption and come back to the weight display, by pressing the key the command will be deleted and you will automatically exit. If the printer is present the " AZZERAMENTO CONSUMI " ( CONSUMPTIONS ZERO-SETTING ) message will be printed with date and time. To exit from the consumption display press .

  • - 15 -

    PRINTING

    IN ORDER FOR THE PRINTER TO BE RECOGNISED AND ENABLED BY THE SYSTEM, IT MUST BE SWITCHED ON AT THE SAME TIME AS THE P-WS.

    If a printer is installed in the system and has been enabled under StA in the parameters, the printing sequence is the following:

    Press and hold down the MENU key. While holding down this key, press to print the formulas or the key to print the parameters.

    During the weight displaying phase press the key to print the consumption. At the end of every batching the products and the batched quantity will be printed. If the clock is available in the printer the date and time will be printed each time. Setting DATE and TIME on the printer:

    The time and date can be adjusted by using the PRINT and FEED keys on the front panel of the printer. Hold down the FEED key and simultaneously press PRINT. The printer will print the instructions, the time and the date with an arrow pointing to the digit that is being changed. Each time the PRINT key is pressed, the digit indicated by the arrow increases and the updated version is printed. Press FEED again to select the next digit to be changed. On completion of each operation, the printer will print the updated time and date. On reaching the last digit of the date on the right, press FEED again to exit date and time setting. The printer will now print the new set values.

  • - 16 -

    EXAMPLES OF PRINTOUTS PARAMETER PRINTOUT ( MENU + ): RISO CELL FILT STAMP TOLL A.TAR LE.OF LE.ON No.CO VO. 12 VO. 11 VO. 10 VO. 9 VO. 8 VO. 7 VO. 6 VO. 5 VO. 4 VO. 3 VO. 2 VO. 1 PAUSA P.MAX P.MIN t.EST divis o.VOL PASS unit mV/V dECP tot

    = 0001 = 1000 = 000.2 = 0001 = 0000 = 000.0 = 000.0 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 0000 = 001.0 = 9999 = 001.0 = 0010

    LAU2 costanti 11:34 22-04-99

    FORMULAS PRINTOUT ( MENU + ): SET = 0000 P.SET = 0000 formula 01

  • - 17 -

    LAU2 formule 11:07 22-04-99

    CONSUMPTION PRINTOUT ( while the consumption are displayed): COMP.01 = 0078 LAU 2 consumi 11:07 22-04-99 AZZERAMENTO CONSUMI ( Consumption zero-setting ) 10:25 23-04-99 BATCHING ( printout at the batching cycles' end ): A.TAR = 0 ( autotara = no ) ( disabled auto-tare ) LORDO = kg 0130 ( 0130 ) ( GROSS WEIGHT ) LAU2 F. 01 cic. 0001 / 0001 08:31 23-04-99 A.tAr = 1 ( autotara = si ) ( enabled auto-tare ) NETTO = kg 0130 ( 0130 ) ( NET WEIGHT ) LAU2 F. cic. 0001 / 0001 08:30 23-04-99

  • - 18 -

    PROGRAMMING SECRET SYSTEM PARAMETERS Press the DOSA key at any time to exit and return to the weight display with the new system parameters.

    Use the and keys to go from one parameter to another without making changes. Carry out programming as follows: Switch on the instrument while holding down the DOSA key. The display will show C.O.S.c., press

    and the display will show: tot ( consumption ) Confirm by pressing ENTER and the display will show the programmed value, now set: 0 = consumption disabled; 1 = kg x 10 ; 2 = kg x 100 ; 3 = kg x 1000, it means that if one programs 1 and the instrument batches 100 kg of a product, the instrument will memorize 10 in the consumption. The comma isn't considered, it means that if the instrument batches 10,0 kg the value 10 will be memorized in the consumption. Confirm by pressing ENTER , the following will appear : dECP ( system decimals )

    Press ENTER to confirm. The display will show the number of enabled decimals. Use the and keys to change the value (from 0 to 3 decimals max.).

    Press ENTER to confirm. The display will now show: nU-U ( load cell sensitivity expressed in mV/V )

    Press ENTER to confirm and display the value. Use the and keys to change it. Program: ( ) Press ENTER to confirm. The display will show: unit ( unit of measurement )

    Press ENTER to confirm. Use the and keys to select: 0 = kg ; 1 = t . Press ENTER to confirm. The display will now show:

  • - 19 -

    PASS (access protection password). When the password is enabled, access can be restricted/denied to parameters programming and calibration functions.

    Press ENTER to confirm and use the and keys to select: 0 = password enabled (access will be denied to parameters programming and calibration). 1 = password disabled. Press ENTER to confirm and the display will show: o.VOL manual fall or automatic fall selection,

    Press ENTER to confirm. Use the and keys to select: 0 = manual fall ; 1 - 99 = automatic fall every 1 - 99 batchings ( if one programs 2 the fall will be calculated every 2 batchings, if one programs 3 the fall will be calculated every 3 batchings, so on until max 99 batchings ). Press ENTER to confirm and the display will show: div. ( expressed in unit of measure of the system ), value, added or subtracted to the SET value, for which the system considers the value of the automatic fall valid. For example : div = 5 SET = 100, the system will continue to calculate the value of fall until the weight batched will be included between 95 and 105; then the instrument will calculate the value of fall every " x " batchings programmed at the voice " o.VOL ".

    Press ENTER and the display will show the value, use the and the keys to change it and confirm with ENTER, the display will show: t.ESt ZERO or STOP selection from external contact.

    Press ENTER and the display will show the value, use the and the keys to change it: 0 = ZERO ( zero-setting ) 1 = STOP. Warning: If you set 0 (ZERO), by closing the input the instrument will set the weight to zero only if the weight is inferior to the minimum weight value ( see Constants programming ). Press ENTER to confirm, then press DOSA to exit and return to the weight displaying phase.