63

SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm
Page 2: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 1 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

All types of Caliper

Digital/

Dial/

Vernier-0.02 mm

GTS-WP-22 Based on

BS EN ISO 13385-1&

BS 887 (only for limits of

error reference)

0 - 600mm 20μm Laboratory/

Customer

Premises

(Std. room/

Metrology)

> 600mm to 1500mm 26μm

1500mm < to 2000mm 35μm

Vernier Caliper

(0.05 mm)

GTS-WP-22

Based on BS EN ISO

13385-1 & BS 887 (only

for limits of error

reference)

0 - 600mm (already

available)

± 36μm

600mm < to 1500mm ± 40μm

1500mm < to 2000mm ± 43μm

External Micrometer

(Digital)

LC:0.001 mm)

GTS-WP-23

Based on BS EN ISO 3611

&

BS 870 (only for limits of

error reference)

0-25mm

(already available)

1.7μm Laboratory/

Customer

Premises

(Std. room/

Metrology)

>25mm up to 100mm 2.2μm

>100mm up to 500mm 8.6μm

>500mm up to 925mm 15 μm

Page 3: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 2 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

External Micrometer

(Analogue

LC:0.01 mm)

GTS-WP-23

Based on BS EN ISO 3611

&

BS 870 (only for limits of

error reference)

0 to 25mm

(already available)

± 2.9μm Laboratory/

Customer

Premises

(Std. room/

Metrology)

>25mm up to 100mm ± 3.8μm

>100mm up to 500mm ± 8.7μm

>500mm up to 925mm 15μm

Micrometer Setting

Standard

GTS-WP-23

Based on BS EN ISO 3611

(using ULMS)

Up to 100 mm 1.1μm Laboratory

>100 Up to 600mm 3.1μm

GTS-WP-23

Based on BS EN ISO 3611

(using HMS)

Up to 100mm 1.8μm

>100mm Up to 600mm 5.2μm

>600mm Up to 900mm 9.2μm

Page 4: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 3 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Depth Micrometer

(Digital/analogue)

GTS-WP-28

Based on BS EN ISO 6468

Up to 25mm 2.0μm Laboratory >25mm up to 100mm 2.2μm

>100mm Up to 300mm 4.1μm

Up to 100mm /0.01mm 6.3μm

>100mm Up to 300mm

/0.01 mm

12μm

Tubular Micrometer

/Inside Micrometer

(Digital/ Analogue)

GTS-WP-24

Based on BS EN ISO 959

Up to 300mm 4.3μm Laboratory >300mm Up to 500mm 5.3μm

>500mm up to 1000mm 15μm

0.001mm to

300mm/0.01mm

7.2μm

300mm < to

2000mm/0.01mm

21μm

Inside Micrometer

(Caliper Type)

GTS-WP-24

Based on BS EN ISO 959

0mm to 50mm 2.4μm Laboratory

Page 5: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 4 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Dial/Digital

Indicators

GTS-WP-26

Based on BS EN ISO 463&

BS 907(only for limits of

error reference) (using

ULMS)

0.01mm to 100mm

/0.01mm

7.7μm Laboratory

0.001mm to 50mm

/0.001mm

2.3μm

Using Dial calibration

Tester

0.01mm to 25mm

/0.01mm

(Existing Scope )

25mm < to

100mm/0.01mm

7μm

Laboratory/

Customer

Premises

(Std. room)

Bore Gauge

(Ordinary/Digital)

GTS-WP-27

Based on JIS B 7515

(using Calibrations tester

& ULMS)

Up to 400mm/0.001mm 3.2μm Laboratory

Up to 400mm/0.01mm 6.5μm

Page 6: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 5 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

LVDT

(Ordinary/Digital)

GTS-WP-168

Based on ASTM F2537

(using ULMS/Slip Guges)

UP to 200mm 0.2+(5*L/1000)

μm Laboratory/

Customer

Premises

(Std. room)

Dial Test Indicator /

Lever Type Dial

Gauges

GTS-WP-172

Based on BS EN ISO 463,

BS 2795 & IS 11498

(using ULMS)

Up to 1mm/0.001mm 1.7μm Laboratory

Up to 1mm/0.01mm 5.8μm

Using Dial Calibration

Tester

Up to 1mm/0.01mm 6.9μm Laboratory/

Customer

Premises

Dial / Digital

Thickness Gauges

GTS-WP-36

Based on JIS B7503; JIS

B7524

(Using ULMS/slip gauge)

Up to 25mm /0.001mm 2.8μm Laboratory

Up to 25mm /0.01mm 6.1μm

Page 7: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 6 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Depth Gauge

(Dial/Digital/Vernier)

GTS-WP-29

Based on BS EN ISO

13385-2 (Using slip

gauges and HMS)

Up to 450mm/0.02mm ± 23μm Laboratory

Up to 450mm/0.01mm ±18μm

Height Gauge

(Digital/Dial/Analogu

e)

GTS-WP-25

Based on BS EN ISO

13225/

BS-1643 (Using slip

gauges and HMS)

Up to 600mm ± 14μm Laboratory/

Customer

Premises

(Std. room/

Metrology)

Up to 1000mm ± 17μm

Feeler Gauge GTS-WP-56

Based on BS 957

Up to 1mm 3.1μm

Radius Gauge GTS-WP-81

Based on IS 5273-1969

(using Profile Projector)

Up to 25mm 9.0μm Laboratory

Page 8: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 7 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Thread / Screw Pitch

Gauge (Metric/inch)

GTS-WP-126

Based on IS 4211 (using

Profile Projector)

Up to 7mm 5.2μm Laboratory

Up to 42 TPI 170μin

Thread Plug gauges

(Metric / Unified/BSP

(or) G threads )

For simple Pitch

Diameter

GTS-WP-70

Based on Euramet cg-10

(using ULMS & Profile

Projector) Metric Threads

1mm Up to 100mm 1.7μm

>100 Up to 200mm 2.4μm

Inch - Unified / BSP 1/16" to 4" 63μin

Above 4" and including 8" 91μin

Thread Ring gauges

(Metric / Unified/BSP

(or) G threads ) For simple Pitch

diameter

GTS-WP-137

Based on Euramet cg-10

(Using ULMS) Metric

Threads

3mm to 14mm 1.7μm Laboratory >14mm up to and

including 100mm

1.9μm

Inch - Unified / BSP 1/8" to 1/2" 67μin

1/2"< and including 4" 72μin

Thread Plug/Ring GTS-WP-173 1/8" to 1/2" 68μin Laboratory

Page 9: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 8 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

gauge – Taper

(NPT / BSPT)

For simple Pitch

Diameter

Based on JIS B 0262 &

Euramet cg-10 (using

ULMS)

1/2"< and including 4" 87μin

Setting / Plain Plug

gauge (Metric/Inch)

GTS-WP-80

Based on ASME B89.1.5

(using ULMS)

1mm to 50mm 0.85μm Laboratory >50mm Up to 100mm 1.1μm

>100mm Up to 400mm 4.2μm

Setting / Plain Ring

gauge (Metric/Inch)

GTS-WP-106

Based on BS EN ISO 4064

(using ULMS and Master

setting ring)

1mm to 14mm 0.68μm 14mm < to 400mm 2.1μm

Height Measuring

System (HMS)/

Digital Height Gauge

L.C: 0.001mm and

greater

GTS-WP-169

Based on BS EN ISO

13225/BS 1643

(Slip Gauges & Long slip

gauges)

Up to 1000mm 0.9+(6.5L/1000)

μm

Where “L” is in mm

Laboratory

Page 10: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 9 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Universal Length

Measuring System

(ULMS)

GTS-WP-139

(using slip gauges)

Up to 100mm (absolute)

Up to 600mm

(differential)

0.18 +

(2.5L/1000) μm

(Where L is in

mm)

Laboratory

Profile Projector

GTS-WP-158

Based on JIS 7540

(Using Glass scale and

Cross wire chart)

Up to 200mm 2.8 + (2.2

L/1000) μm

0.19°(11.4’)

(Where L is in

mm)

Laboratory

Thread Measuring

Cylinder

GTS-WP-170

(using ULMS) Based on

BS 3777 & BS 5590

up to 6.35mm 0.67μm Laboratory

Using High Precision

Micrometer

Page 11: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 10 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Cylindrical Standards

/ Measuring Pins

GTS-WP-170

(using ULMS)

Based on IS-11103

Up to 12mm 0.67μm Laboratory

Using High Precision

Micrometer

Steel Scale GTS-WP-171

Based onOIML R035-1-e&

BS 4372: 1968

(Using Profile Projector,

dig. Micrometer)

Up to 1000mm / 0.5mm 0.30mm Laboratory Up to 1000mm / 1mm 0.60mm

Caliper Checker GTS-WP-164

(using HMS and Ext. Slip

Gauge) Based on

Manufacturer Spec.

Up to 600mm 1+(6.7*L/1000)

μm

Where L is in

mm

Laboratory

Page 12: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 11 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Depth Micro Checker

GTS-WP-165 and 166

(using HMS and Ext. Slip

Gauge) Based on

Manufacturer Spec.

Up to 300mm 3.9μm Laboratory

Inside Micro checker GTS-WP-166

(using HMS and Ext. Slip

Gauge) Based on

Manufacturer Spec.

Up to 300mm

3.5μm Laboratory

>300mm Up to 600mm 5.4μm

Dial Calibration

Tester

GTS-WP-167

(using HMS and Slip

Gauge/ULMS)Based on

Manufactured Spec.

Up to 25mm 1.5μm Laboratory

Page 13: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Dimensional Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and Specification Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 12 of 12 DAC-70-02/F34-B, rev. b (24-11-2016)

Test Sieves GTS-WP-43

Based on ISO 3310-1

(Using Profile Projector)

50μm to 4.3mm

4.0μm Laboratory

GTS-WP-43

Based on ISO 3310-1

(using digital caliper)

5mm to 125mm 32μm Laboratory

Page 14: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Temperature Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 1 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)

Liquid-in-glass

thermometers

GTS-WP-13 based on BS

1041-2.1

-30 °C up to 150 °C 0.16 °C Laboratory

Direct reading

thermometers

GTS-WP-15 (in house method) -30 °C up to 160 °C 0.16 °C Laboratory

>160 °C up to 500 °C 0.4 °C

>500 °C up to 900 °C 1.3 °C

>900 °C up to 1200 °C 4 °C

Dial thermometers GTS-WP-14 based on EN

13190

-30 °C up to 160 °C 0.16 °C Laboratory

>160 °C up to 400 °C 2 °C

400 °C up to 800 °C 4 °C

Page 15: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Temperature Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 2 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)

Base metal

thermocouples

GTS WP-12 -40°C up to 250°C 0.3 °C Laboratory

250°C up to 600 °C 0.6 °C

600 °C up to 900°C 0.9 °C

900 °C up to 1200 °C 4 °C

Noble metal

thermocouples

GTS-WP-12 0 °C up to 600 °C 0.6 °C Laboratory

600 °C up to 900 °C 0.8 °C

900 °C up to 1200 °C 1.7 °C

Climatic Chamber GTS-WP-09

DKD-R5-7

(9 points)

30 °C up to 180 °C 1.1 °C Laboratory

/ Customer

Premises

Page 16: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Temperature Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 3 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)

Water bath, incubators GTS-WP-09

DKD-R5-7

(5 points)

5 °C up to 95 °C 1.1 °C Laboratory

/ Customer

Premises

Freezer / Chiller GTS –WP-154

DKD-R5-7

(9 points)

-30 °C up to 95 °C 1.1 °C Laboratory

/ Customer

Premises

Furnace, Oven GTS-WP 09

DKD-R5-7

(9 points, muffle furn.:

1 point)

30 °C up to 180 °C 1.1 °C Laboratory

/ Customer

Premises 180 °C up to 300 °C 1.3 °C

300 °C up to 800 °C 4.0 °C

800 °C up to 1200 °C 9.0 °C

Page 17: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Temperature Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 4 of 4 DAC-70-02/F34-B, rev. b (24-11-2016)

Auto Clave

(Temperature)

GTS-WP-155

DKD-R5-7

(5-9 points)

100 °C up to 140 °C 2 °C Laboratory

/ Customer

Premises

Page 18: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 1 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Sourcing

AC Voltage

Direct Method using Fluke

9100

U= Voltage

1 μV to 10 mV Laboratory

/Customer

Premises

10 Hz - 3 kHz 0.46 x 10-3 U + 0.44 mV

>3 kHz - 10 kHz 0.46 x 10-3 U + 0.59 mV

>10 kHz - 30 kHz 0.69 x 10-3 U + 1.1 mV

>30 kHz - 50 kHz 1.0 x 10-3 U + 2.2 mV

>50 kHz - 100 kHz 2.3 x 10-3 U + 5.9 mV

10 mV to 32 mV

10 Hz - 3 kHz 0.46 x 10-3 U + 0.11 mV

>3 kHz - 10 kHz 0.4 x 10-3 U + 0.15 mV

>10 kHz - 30 kHz 0.70 x 10-3 U + 0.28 mV

>30 kHz - 50 kHz 1.0 x 10-3 U + 0.56 mV

>50 kHz - 100 kHz 2.3 x 10-3 U + 1.5 mV

32 mV to 320 mV

10 Hz - 3 kHz 0.47 x 10-3 U + 22 μV

>3 kHz - 10 kHz 0.47 x 10-3 U + 29 μV

>10 kHz - 30 kHz 0.70 x 10-3 U + 56 μV

Page 19: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 2 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Voltage

Direct Method using Fluke

9100

U= Voltage

32 mV to 320 mV Laboratory

/Customer

Premises

>30 kHz - 50 kHz 1.1 x 10-3 U + 0.11 mV

>50 kHz - 100 kHz 2.3 x 10-3 U + 0.30 mV

0.32 V to 3.2 V

10 Hz - 3 kHz 0.48 x 10-3 U + 0.22 mV

>3 kHz - 10 kHz 0.47 x 10-3 U + 0.29 mV

>10 kHz - 30 kHz 0.70 x 10-3 U + 0.55 mV

>30 kHz - 50 kHz 1.1 x 10-3 U + 1.1 mV

>50 kHz - 100 kHz 2.3 x 10-3 U + 3.0 mV

3.2 V to 32 V

10 Hz - 3 kHz 0.48 x 10-3 U + 2.2 mV

>3 kHz - 10 kHz 0.71 x 10-3 U + 2.9 mV

>10 kHz - 30 kHz 0.93 x 10-3 U + 5.5 mV

>30 kHz - 50 kHz 1.7 x 10-3 U + 11 mV

>50 kHz - 100 kHz 4.1 x 10-3 U + 37 mV

Page 20: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 3 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Voltage

Direct Method using Fluke

9100

U= Voltage

32 V to 105 V Laboratory

/Customer

Premises

10 Hz - 3 kHz 0.47 x 10-3 U + 7.3 mV

>3 kHz - 10 kHz 0.70 x 10-3 U + 9.7 mV

>10 kHz - 30 kHz 0.93 x 10-3 U + 18 mV

>30 kHz - 50 kHz 1.7 x 10-3 U + 36 mV

>50 kHz - 100 kHz 4.1 x 10-3 U + 0.12 V

105 V to 320 V

40 Hz to 100 Hz 0.6 x 10-3 U + 22 mV

>100 Hz - 1 kHz 0.6 x 10-3 U + 22 mV

>1 kHz - 3 kHz 0.94 x 10-3 U + 22 mV

>3 kHz - 10 kHz 0.94 x 10-3 U + 37 mV

>20 kHz - 30 kHz 1.7 x 10-3 U + 74 mV

320 V to 800 V

40 Hz to 100 Hz 0.59 x 10-3 U + 73 mV

>100 Hz - 1 kHz 0.59 x 10-3 U + 73 mV

>1 kHz - 3 kHz 0.93 x 10-3 U + 73 mV

>3 kHz - 10 kHz 0.93 x 10-3 U + 0.12 V

Page 21: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 4 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Voltage

Direct Method using Fluke

9100

U= Voltage

800V to 1050 V Laboratory

/Customer

Premises

40 Hz to 100 Hz 0.59 x 10-3 U + 0.15 V

>100 Hz - 1 kHz 0.59 x 10-3 U + 0.15 V

>1 kHz - 3 kHz 0.93 x 10-3 U + 0.15 V

>3 kHz - 10 kHz 0.93 x 10-3 U + 0.24 V

AC Voltage

Direct Method using Fluke

5522A

U= Voltage

1.0 mV – 33 mV Laboratory

/Customer

Premises

>10 Hz - 45 Hz 0.93 x 10-3 U + 7.03 μV

>45 Hz -10 kHz 0.18 x 10-3 U + 7.03 μV

>10 kHz - 20 kHz 0.24 x 10-3 U + 7.03 μV

>20 kHz - 50 kHz 1.16 x 10-3 U + 7.03 μV

>50 kHz - 100 kHz 4.05 x 10-3 U + 13.9 μV

>100 kHz - 500 kHz 9.28 x 10-3 U + 57.8 μV

33 mV – 330 mV

10 Hz - 45 Hz 0.35 x 10-3 U + 9.20 μV

>45 Hz -10 kHz 0.18 x 10-3 U + 9.09 μV

>10 kHz - 20 kHz 0.20 x 10-3 U + 9.10 μV

Page 22: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 5 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Voltage

Direct Method using Fluke

5522A

U = Voltage

33 mV – 330 mV Laboratory

/Customer

Premises

>20 kHz - 50 kHz 0.41 x 10-3 U + 9.20 μV

>50 kHz - 100 kHz 0.93 x 10-3 U + 36.90 μV

>100 kHz - 500 kHz 2.38 x 10-3 U + 79.90 μV

0.33V - 3.3 V

10 Hz - 45 Hz 0.35 x 10-3 U + 57.60 μV

>45 Hz -10 kHz 0.18 x 10-3 U + 68.30 μV

>10 kHz - 20 kHz 0.23 x 10-3 U + 68.60 μV

>20 kHz - 50 kHz 0.35 x 10-3 U + 57.60 μV

>50 kHz - 100 kHz 0.81 x 10-3 U + 0.14 mV

>100 kHz - 500 kHz 2.80 x 10-3 U + 0.69 mV

3.3V – 33 V

10 Hz - 45 Hz 0.35 x 10-3 U + 0.75 mV

>45 Hz -10 kHz 0.18 x 10-3 U + 0.68 mV

>10 kHz - 20 kHz 0.28 x 10-3 U + 0.69 mV

>20 kHz - 50 kHz 0.41 x 10-3 U + 0.69 mV

>50 kHz - 100 kHz 1.05 x 10-3 U + 1.84 mV

Page 23: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 6 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Voltage

Direct Method using Fluke

5522A

U = Voltage

33 V-330 V Laboratory

/Customer

Premises

45 Hz - 1 kHz 0.23 x 10-3 U + 2.32 mV

>1 kHz - 10 kHz 0.24 x 10-3 U + 6.85 mV

>10 kHz - 20 kHz 0.30 x 10-3 U + 6.87 mV

>20 kHz - 50 kHz 0.36 x 10-3 U + 6.78 mV

>50 kHz - 100 kHz 2.23 x 10-3 U + 57.60 mV

330 V - 1020 V

45 Hz - 1 kHz 0.35 x 10-3 U + 12.10 mV

>1 kHz - 5 kHz 0.30 x 10-3 U + 12.10 mV

>5 kHz - 10 kHz 0.36 x 10-3 U + 11.70 mV

DC Voltage

Direct Method using Fluke

9100

U = Voltage

0.001 mV to 320.000

mV

87 x 10-6 U + 5 μV

Laboratory

/Customer

Premises 0.32001 V to 3.20000

V

89 x 10-6 U + 48 μV

3.2001 V to 32.0000 V 93 x 10-6 U + 0.46 mV

32.001 V to 320.000 V 93 x 10-6 U + 5 mV

320.01 V to 1050.00 V 88 x 10-6 U + 22 mV

Page 24: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 7 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

DC Voltage

Direct Method using Fluke

5522A

U =Voltage

0 - 329.9999 mV 58 x 10-6 U + 2 μV Laboratory

/Customer

Premises

0.33 V - 3.299999 V 59 x 10-6 U + 1 μV

3.3 V - 32.99999 V 60 x 10-6 U + 7 μV

33 V - 329.9999 V 62 x10-6 U + 0.069 mV

330 - 1020.000 V 61 x 10-6 U + 71 mV

DC Current

Direct Method using Fluke

9100

= Current

0.001 μA to 320.000

μA

0.17 x 10-3 + 0.013 μA Laboratory

/Customer

Premises 0.32001 mA to

3.20000 mA

0.18 x 10-3 + 0.094 μA

3.2001 mA to 32.0000

mA

0.18 x 10-3 + 1.0 μA

32.001 mA to 320.000

mA

0.20 x 10-3 + 11 μA

0.32001 A to 3.20000 A 0.70 x 10-3 + 0.14 mA

3.2001 A to 10.5000 A 0.64 x 10-3 + 1.1 mA

Page 25: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 8 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

DC Current

Direct Method using Fluke

9100

= Current

10-Turn Coil Laboratory

/Customer

Premises

3.2001 A to 32.0000

A

2.4 x 10-3 + 0.51 mA

32.001 A to 105.000

A

2.4 x 10-3 + 3.8 mA

50-Turn Coil

16.001 A to 160.000

A

2.4 x 10-3 + 2.5 mA

160.01 A to 525.00 A 2.4 x 10-3 + 19 mA

525.01 A to 1000.00

A

2.4 x 10-3 + 0.104 A

Direct Method using Fluke

5522A

=Current

0 - 329.999 μA 0.17 x 10-3 + 0.023 μA Laboratory

/Customer

Premises

0.33 mA - 3.29999

mA

0.12 x 10-3 + 0.059 μA

3.3 mA - 32.99999

mA

0.12 x 10-3 + 0.30 μA

Page 26: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 9 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

DC Current Direct Method using Fluke

5522A

=Current

33 mA - 329.999 mA 0.12 x 10-3 + 3.00 μA Laboratory

/Customer

Premises

0.33 A - 1.09999 A 0.23 x 10-3 + 0.047 mA

1.1 A - 2.99999 A 0.44 x 10-3 + 0.046 mA

3 A - 10.9999 A 0.59 x 10-3 + 0.57 mA

11 A - 20.5 A 1.21 x 10-3 + 0.84 mA

50- Turn Coil

0.2 A - 0.33 A 5.84 x 10-3 + 23 mA

0.33 A - 2.9999 A 5.83 x 10-3 + 0.16 A

3 A - 20.5 A 5.88 x 10-3 + 0.58 A

AC Current

Direct Method using Fluke

9100

= Current

1.0 μA to 320.0 μA Laboratory

/Customer

Premises

10 Hz - 3 kHz 0.83 x 10-3 + 0.35 μA

>3 kHz - 10 kHz 1.2 x 10-3 + 0.69 μA

>10 kHz - 20 kHz 2.3 x 10-3 + 2.3 μA

>20 kHz - 30 kHz 2.9 x 10-3 + 3.5 μA

Page 27: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 10 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Current

Direct Method using Fluke

9100

= Current

0.32 mA to 3.2 mA Laboratory

/Customer

Premises

10 Hz - 3 kHz 0.85 x 10-3 + 0.34 μA

>3 kHz - 10 kHz 1.2 x 10-3 + 0.68 μA

>10 kHz - 20 kHz 2.4 x 10-3 + 2.3 μA

>20 kHz - 30 kHz 2.9 x 10-3 + 3.5 μA

3.2 mA to 32 mA

10 Hz - 3 kHz 0.85 x 10-3 + 3.6 μA

>3 kHz - 10 kHz 1.2 x 10-3 + 7.3 μA

>10 kHz - 20 kHz 2.4 x 10-3 + 15 μA

>20 kHz - 30 kHz 2.9 x 10-3 + 26 μA

32 mA - 320 mA

10 Hz - 3 kHz 0.9 x 10-3 + 36 μA

>3 kHz - 10 kHz 1.2 x 10-3 + 54 μA

>10 kHz - 20 kHz 2.4 x 10-3 + 74 μA

>20 kHz - 30 kHz 2.9 x 10-3 + 0.11 mA

Page 28: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 11 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Current

Direct Method using Fluke

9100

= Current

0.32 A to 3.2 A Laboratory

/Customer

Premises

10 Hz - 3 kHz 1.2 x 10-3 + 0.55 μA

>3 kHz - 10 kHz 2.9 x 10-3 + 3 mA

3.2 A to 10.5 A

10 Hz - 3 kHz 2.3 x 10-3 + 3.5 mA

>3 kHz - 10 kHz 5.8 x 10-3 + 12 mA

10-Turn Coil

3.2001 A to 32.0000 A

10 Hz - 100 Hz 3.3 x 10-3 + 5.2 mA

>100 Hz - 440 Hz 9.3 x 10-3 + 31 mA

32 A to 200 A

10 Hz - 100 Hz 3.3 x 10-3 + 91 mA

>100 Hz - 440 Hz 8.1 x 10-3 + 0.28 A

Page 29: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 12 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Current

Direct Method using Fluke

9100

= Current

50-Turn Coil Laboratory

/Customer

Premises

16.001 A to 160.000

A/10 Hz - 100 Hz

3.3 x 10-3 + 27 mA

160.01 A to 1000.00

A/10 Hz - 100 Hz

3.3 x 10-3 + 0.45 A

Direct Method using Fluke

5522A

= Current

30 - 330 μA

10Hz-20Hz 2.31 x 10-3 + 0.12 μA

>20Hz - 45Hz 1.73 x 10-3 + 0.12 μA

>45Hz - 1kHz 1.45 x 10-3 + 0.12 μA

>1kHz - 5kHz 3.47 x 10-3 + 0.17 μA

>5kHz - 10kH 9.25 x 10-3 + 0.23 μA

>10kHz - 30kHz 18.5 x 10-3 + 0.46 μA

0.33 - 3.3 mA

10Hz - 20Hz 2.31 x 10-3 + 0.17 μA

>20Hz - 45Hz 1.45 x 10-3 + 0.17 μA

>45Hz - 1kHz 1.16 x 10-3 + 0.17 μA

Page 30: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 13 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Current

Direct Method using Fluke

5522A

= Current

0.33 - 3.3 mA Laboratory

/Customer

Premises

>1kHz - 5kHz 2.31 x 10-3 + 0.23 μA

>5kHz - 10kHz 5.78 x 10-3 + 0.35 μA

>10kHz - 30kHz 11.6 x 10-3 + 0.69 μA

3.3 - 33 mA

10Hz -20Hz 2.08 x 10-3 + 2.31 μA

>20Hz - 45Hz 1.05 x 10-3 + 2.31 μA

>45Hz - 1kHz 0.49 x 10-3 + 2.26 μA

>1kHz - 5kHz 0.97 x 10-3 + 2.25 μA

>5kHz - 10kHz 2.33 x 10-3 + 2.30 μA

>10kHz - 30kHz 4.64 x 10-3 + 2.31 μA

33 -330 mA

10Hz -20Hz 2.08 x 10-3 + 23.1 μA

>20Hz - 45Hz 1.04 x 10-3 + 23.1 μA

>45Hz - 1kHz 0.47 x 10-3 + 23 μA

Page 31: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 14 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Current

Direct Method using Fluke

5522A

= Current

33 -330 mA Laboratory

/Customer

Premises

>1kHz - 5kHz 1.16 x 10-3 + 57.7 μA

>5kHz - 10kHz 2.31 x 10-3 + 0.12 mA

>10kHz - 30kHz 4.63 x 10-3 + 0.23 mA

0.33 - 1.1 A

10Hz - 45Hz 2.08 x 10-3 + 12 mA

>45Hz - 1kHz 0.59 x 10-3 + 11 mA

>1kHz - 5kHz 6.96 x 10-3 + 1.15 mA

>5kHz - 10kHz 28.9 x 10-3 + 5.78 mA

1.11 - 3 A

10Hz - 45Hz 2.08 x 10-3 + 0.12 mA

>45Hz - 1kHz 0.70 x 10-3 + 0.12 mA

>1kHz - 5kHz 6.94 x 10-3 + 1.16 mA

>5kHz - 10kHz 28.9 x 10-3 + 5.78 mA

Page 32: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 15 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Current

Direct Method using Fluke

5522A

= Current

3 -11 A Laboratory

/Customer

Premises

45Hz - 100 Hz 0.71 x 10-3 + 2.30 mA

>100Hz - 1kHz 1.16 x 10-3 + 2.30 mA

>1kHz - 5kHz 34.7 x 10-3 + 2.31 mA

11 - 20.5 A

45Hz - 100 Hz 1.4 x 10-3 + 5.74 mA

>100Hz - 1kHz 1.75 x 10-3 + 5.74 mA

>1kHz - 5kHz 34.7 x 10-3 + 5.78 mA

Amp-50 turn

10 -16.5A @ 45Hz-

65Hz

4.08 x 10-3 + 2.83 mA

16.5 – 150 A @

45Hz-65Hz

4.04 x 10-3 + 25.00 mA

150 – 1025 A @

45Hz-65Hz

4.02 x 10-3 + 83.70 mA

10 -16.4999/ 65Hz-

440Hz

9.47 x 10-3 + 3.57 mA

Page 33: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 16 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Current

Direct Method using Fluke

5522A

= Current

Amp-50 turn Laboratory

/Customer

Premises

16.5 - 149.999/ 65Hz-

440Hz

9.45 x 10-3 + 30.40 mA

150 - 1025/ 65Hz-

440Hz

9.52 x 10-3 + 0.11 A

Resistance

Direct Method using Fluke

9100

R= Resistance

0.0001 Ω to 40.0000 Ω 0.33 x 10-3 R + 12 mΩ Laboratory

/Customer

Premises

40.001 Ω to 400.000 Ω 0.23 x 10-3 R + 23 mΩ

0.40001 kΩ to 4.00000 kΩ

0.17 x 10-3 R + 93 mΩ

4.0001 kΩ to 40.0000 kΩ

0.17 x 10-3 R + 0.93 Ω

40.001 kΩ to 400.000 kΩ

0.21 x 10-3 R + 9.3 Ω

0.40001 MΩ to 4.00000 MΩ

0.23 x 10-3 R + 0.12 kΩ

4.0001 MΩ to 40.0000 MΩ

0.59 x 10-3 R + 2.3 kΩ

Page 34: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 17 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Resistance

Direct Method using Fluke

9100

R= Resistance

40.001 MΩ to 400.000 MΩ

0.71 x 10-3 R + 46 kΩ Laboratory

/Customer

Premises

Direct Method using Fluke

5522A

R= Resistance

0 - 10.9999 Ω 46 x10-6 R + 12 mΩ

11 Ω - 32.9999 Ω 35 x10-6 R + 17 mΩ

33 Ω - 109.9999 Ω 33 x10-6 R + 17 mΩ

110 Ω - 329.9999 Ω 33 x10-6 R + 23 mΩ

330 Ω - 1.099999 kΩ 33 x10-6 R + 23 mΩ

1.1 kΩ - 3.299999 kΩ 33 x10-6 R + 0.23 Ω

3.3 kΩ -10.99999 kΩ 33 x10-6 R + 0.12 Ω

11 kΩ - 32.99999 kΩ 33 x10-6 R + 1.15 Ω

33 kΩ - 109.9999 kΩ 33 x10-6 R + 1.15 Ω

110 kΩ - 329.99999 kΩ 37 x10-6 R + 12 Ω

330 kΩ - 1.099999 MΩ 38 x10-6 R + 12 Ω

1.1 MΩ - 3.299999 MΩ 70 x10-6 R + 0.17 kΩ

3.3 MΩ - 10.99999 MΩ 0.15 x 10-3 R + 0.29 kΩ

11 MΩ - 32.99999 MΩ 0.30 x 10-3 R + 2.84 kΩ

Page 35: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 18 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Resistance

Direct Method using Fluke

5522A

R= Resistance

33 MΩ - 109.9999 MΩ 0.62 x 10-3 R + 3.30 kΩ Laboratory

/Customer

Premises

110 MΩ - 329.9999 MΩ 3.48 x 10-3 R + 0.12 MΩ

330 MΩ - 1100 MΩ 17.40 x 10-3 R + 0.58 MΩ

Capacitance

Direct Method using Fluke

9100

C= Capacitance

0.5000 nF to 4.0000 nF 25 pF Laboratory

/Customer

Premises

4.0001 nF to 40.000 nF 0.11 nF

40.001 nF to 400.00 nF 0.96 nF

400.01 nF to 4.0000 μF 4.7 x 10-3 C + 1.9 nF

4.0001 μF to 40.000 μF 5.8 x 10-3 C + 19 nF

40.001 μF to 400.00 μF 5.8 x 10-3 C + 0.19 μF

400.01 μF to 4.0000 mF 5.8 x 10-3 C + 1.8 μF

4.0001 mF to 40.000 mF 12 x 10-3 C + 69 μF

Direct Method using Fluke

5522A

C= Capacitance

220 pF - 399.9 pF 9.16 x 10-3 C+ 11 pF

0.4 nF - 1.0999 nF 6.21 x 10-3 C+ 12 pF

1.1 nF - 3.2999 nF 5.93 x 10-3 C + 12 pF

3.3 nF - 10.9999 nF 2.97 x 10-3 C + 12 pF

11 nF - 32.9999 nF 2.95 x 10-3 C + 0.12 nF

33 nF - 109.999 nF 2.96 x 10-3 C + 0.12 nF

Page 36: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 19 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Capacitance

Direct Method using Fluke

5522A

C= Capacitance

110 nF - 329.999 nF 2.96 x 10-3 C + 0.34 nF Laboratory

/Customer

Premises

0.33 μF - 1.09999 μF 2.97 x 10-3 C + 1.14 nF

1.1 μF- 3.29999 μF 2.96 x 10-3 C + 3.43 nF

3.3 μF - 10.9999 μF 2.96 x 10-3 C + 11.40 nF

11 μF - 32.9999 μF 4.68 x 10-3 C + 34.50 nF

33 μF - 109.999 μF 5.37 x 10-3 C + 0.12 μF

110 μF -329.999 μF 5.37 x 10-3 C + 0.34 μF

0.33 mF - 1.09999 mF 5.37 x 10-3 C + 1.14 μF

1.1 mF - 3.29999 mF 5.37 x 10-3 C + 3.41 μF

3.3 mF - 10.9999 mF 5.37 x 10-3 C + 11.4 μF

11 mF - 32.9999 mF 8.78 x 10-3 C + 34.4 μF

33 mF - 110 mF 12.80 x 10-3 C + 0.12 mF

Frequency

Direct Method using Fluke

9100

f= Frequency

0.5 Hz to 10.0 MHz 29 x 10-6 f Laboratory

/Customer

Premises

Page 37: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 20 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Frequency

Direct Method using Fluke

5522A

f= Frequency

0.01 Hz - 119.99 Hz 2.87 x 10-6 f + 13 μHz Laboratory

/Customer

Premises

120 Hz - 1199.9 Hz 2.91 x 10-6 f + 26 μHz

1.200 kHz - 11.999 kHz 2.91 x 10-6 f + 0.21 mHz

12.00 kHz - 119.99 kHz 2.91 x 10-6 f + 2.05 mHz

120.00 kHz - 1199.9 kHz 2.91 x 10-6 f + 20.40 mHz

1.200 MHz - 2.000 MHz 2.93 x 10-6 f + 0.29 Hz

Page 38: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 21 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Measuring

DC Voltage

Direct Method using Fluke

8846A

U= Voltage

100 mv 43 x 10-6 U + 4 μV Laboratory

/Customer

Premises

1 V 31 x 10-6 U + 8 μV

10 V 30 x 10-6 U + 57 μV

100 V 46 x 10-6 U + 0.69 mV

1000 V 49 x 10-6 U + 12 mV

AC Voltage

Direct Method using Fluke

8846A

U= Voltage

100 mV Laboratory

/Customer

Premises

5 Hz to 10 kHz 4.1 x 10-3 U + 46 μV

>10 Hz - 20 kHz 0.72 x 10-3 U + 46 μV

>20 kHz - 50 kHz 1.5 x 10-3 U + 58 μV

>50 kHz - 100 kHz 7 x 10-3 U + 93 μV

1 V

5 Hz to 10 kHz 4.1 x 10-3 U + 0.35 mV

>10 Hz - 20 kHz 0.70 x 10-3 U + 0.35 mV

>20 kHz - 50 kHz 1.4 x 10-3 U + 0.58 mV

>50 kHz - 100 kHz 6.9 x 10-3 U + 0.93 mV

Page 39: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 22 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Voltage

Direct Method using Fluke

8846A

U= Voltage

10 V Laboratory

/Customer

Premises 5 Hz to 10 kHz 4.1 x 10-3 U + 3.5 mV

>10 Hz - 20 kHz 0.7 x 10-3 U + 3.5 mV

>20 kHz - 50 kHz 1.4 x 10-3 U + 5.8 mV

>50 kHz - 100 kHz 7 x 10-3 U + 9.3 mV

100 V

5 Hz to 10 Hz 4.1 x 10-3 U + 35 mV

>10 Hz - 20 kHz 0.7 x 10-3 U + 35 mV

>20 kHz - 50 kHz 1.4 x 10-3 U + 58 mV

>50 kHz - 100 kHz 7 x 10-3 U + 93 mV

1000 V

5 Hz to 10 Hz 4.1 x 10-3 U + 0.35 V

>10 Hz - 20 kHz 0.72 x 10-3 U + 0.35 V

>20 kHz - 50 kHz 1.8 x 10-3 U + 0.55 V

>50 kHz - 100 kHz 7.03 x 10-3 U + 0.92 V

Page 40: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 23 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

DC Current

Direct Method using Fluke

8846A

= Current

100 μA 0.59 x 10-3 + 0.03 μA Laboratory

/Customer

Premises 1 mA 0.58 x 10-3 + 0.06 μA

10 mA 0.58 x 10-3 + 2.3 μA

100 mA 0.58 x 10-3 + 5.8 μA

1 A 0.59 x 10-3 + 0.23 mA

10 A 1.8 x 10-3 + 0.92 mA

AC Current

Direct Method using Fluke

8846A

= Current

100 μA/10 Hz to 1 kHz 1.2 x 10-3 + 0.05 μA Laboratory

/Customer

Premises 1 mA/ 10 Hz to 1 kHz 1.2 x 10-3 + 0.46 μA

10 mA /10 Hz to 1 kHz 1.2 x 10-3 + 4.6 μA

100 mA/10 Hz to 1 kHz 1.2 x 10-3 + 46 μA

1 A /10 Hz to 1 kHz 1.2 x 10-3 + 0.46 mA

10 A/ 10 Hz to 1 kHz 1.8 x 10-3 + 6.9 mA

Resistance

Direct Method using Fluke

8846A

R= Resistance

0.0001 Ω to 10.0000 Ω 0.12 x 10-3 R+ 3.5 mΩ Laboratory

/Customer

Premises 10.001 Ω to 100.000 Ω 0.12 x 10-3 R + 4.6 mΩ

0.1001 kΩ to 1.0 kΩ 0.12 x 10-3 R + 12 mΩ

1.0001 kΩ to 10.0000 kΩ

0.12 x 10-3 R + 0.12 Ω

Page 41: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 24 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Resistance

Direct Method using Fluke

8846A

R= Resistance

10.001 kΩ to 100.000 kΩ

0.12 x 10-3 R + 1.2 Ω Laboratory

/Customer

Premises 0.10001 MΩ to 1.00000 MΩ

0.12 x 10-3 R + 11.4 Ω

1.0001 MΩ to 10.0000 MΩ

0.47 x 10-3 R + 0.12 kΩ

10.001 MΩ to 100.000 MΩ

9.3 x 10-3 R + 12 kΩ

RTD - simulation Simulation Method using

Fluke 5522A

t= Temperature

-200 ⁰C to -80 ⁰C /

RTD-pt385,100 Ω

0.29 x 10-3 t + 0.36 m ⁰C Laboratory

/Customer

Premises -80 ⁰C to 0.003 ⁰C /

RTD-pt385,100 Ω

1.16 x 10-0 t + 3.47 m ⁰C

0.03 ⁰C to 100 ⁰C /

RTD-pt385,100 Ω

0.78 x 10-3 t+ 3.45 m ⁰C

100 ⁰C to 300 ⁰C / RTD-

pt385,100 Ω

0.35 x 10-3 t + 0.23 m ⁰C

Page 42: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 25 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

RTD - simulation Simulation Method using

Fluke 5522A

t= Temperature

300 ⁰C to 400 ⁰C / RTD-

pt385,100 Ω

0.29 x 10-3 t + 0.12 m ⁰C Laboratory

/Customer

Premises 400 ⁰C to 630 ⁰C / RTD-

pt385,100 Ω

0.22 x 10-3 t + 0.11 m ⁰C

630 ⁰C to 800 ⁰C / RTD-

pt385,100 Ω

0.33 x 10-3 t + 0.051 m ⁰C

-200 ⁰C to -80 ⁰C /

RTD-pt3926,100 Ω

0.29 x 10-3 t + 0.36 m ⁰C

-80 ⁰C to 0.003 ⁰C /

RTD-pt3926,100 Ω

1.16 x 10-0 t + 3.47 m ⁰C

0.03 ⁰C to 100 ⁰C /

RTD-pt3926,100 Ω

0.78 x 10-3 t + 3.45 m ⁰C

100 ⁰C to 300 ⁰C / RTD-

pt3926,100 Ω

0.35 x 10-3 t + 0.23 m ⁰C

300 ⁰C to 400 ⁰C / RTD-

pt3926,100 Ω

0.29 x 10-3 t + 0.12 m ⁰C

Page 43: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 26 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

RTD - simulation Simulation Method using

Fluke 5522A

t= Temperature

400 ⁰C to 630 ⁰C / RTD-

pt3926,100 Ω

0.22 x 10-3 t + 0.11 m ⁰C Laboratory

/Customer

Premises -200 ⁰C to -190 ⁰C /

RTD-pt3916,100 Ω

1.44 x 10-3 t+ 0.043 m ⁰C

-190 ⁰C to -80 ⁰C /

RTD-pt3916,100 Ω

0.24 x 10-3 t + 0.43 m ⁰C

-80 ⁰C to 0.003 ⁰C /

RTD-pt3916,100 Ω

1.16 x 10-0 t + 3.47 m ⁰C

0.03 ⁰C to 100 ⁰C /

RTD-pt3916,100 Ω

0.66 x 10-3 t + 3.45 m ⁰C

100 ⁰C to 260 ⁰C / RTD-

pt3916,100 Ω

0.31 x 10-3 t + 0.27 m ⁰C

260 ⁰C to 300 ⁰C / RTD-

pt3916,100 Ω

0.31 x 10-3 t + 0.14 m ⁰C

300 ⁰C to 400 ⁰C / RTD-

pt3916,100 Ω

0.26 x 10-3 t + 0.14 m ⁰C

Page 44: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 27 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

RTD - simulation Simulation Method using

Fluke 5522A

t= Temperature

400 ⁰C to 600 ⁰C / RTD-

pt3916,100 Ω

0.19 x 10-3 t + 0.13 m ⁰C Laboratory

/Customer

Premises 600 ⁰C to 630 ⁰C / RTD-

pt3916,100 Ω

0.42 x 10-3 t + 0.046 m ⁰C

-200 ⁰C to -80 ⁰C /

RTD-pt385,200Ω

0.23 x 10-3 t + 0.45 m ⁰C

-80 ⁰C to 0.003 ⁰C /

RTD-pt385,200 Ω

1.16 x 10-0 t + 3.47 m ⁰C

0.03 ⁰C to 100 ⁰C /

RTD-pt385,200 Ω

0.43 x 10-3 t + 3.46 m ⁰C

100 ⁰C to 260 ⁰C / RTD-

pt385,200 Ω

0.22 x 10-3 t + 0.37 m ⁰C

260 ⁰C to 300 ⁰C / RTD-

pt385,200 Ω

0.46 x 10-3 t + 0.093 m ⁰C

300 ⁰C to 400 ⁰C / RTD-

pt385,200 Ω

0.38 x 10-3 t + 0.093 m ⁰C

Page 45: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 28 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

RTD - simulation Simulation Method using

Fluke 5522A

t= Temperature

400 ⁰C to 600 ⁰C / RTD-

pt385,200 Ω

0.27 x 10-3 t + 0.093 m ⁰C Laboratory

/Customer

Premises 600 ⁰C to 630 ⁰C / RTD-

pt385,200 Ω

0.29 x 10-3 t + 0.067 m ⁰C

-200 ⁰C to -80 ⁰C /

RTD-pt385,500 Ω

0.23 x 10-3 t + 0.45 m ⁰C

-80 ⁰C to 0.003 ⁰C /

RTD-pt385,500 Ω

1.16 x 10-0 t + 3.47 m ⁰C

0.03 ⁰C to 100 ⁰C /

RTD-pt385,500 Ω

0.55 x 10-3 t + 0.3.45 m ⁰C

100 ⁰C to 260 ⁰C / RTD-

pt385,500 Ω

0.27 x 10-3 t + 0.31 m ⁰C

260 ⁰C to 300 ⁰C / RTD-

pt385,500 Ω

0.31 x 10-3 t + 0.14 m ⁰C

300 ⁰C to 400 ⁰C / RTD-

pt385,500 Ω

0.23 x 10-3 t + 0.15 m ⁰C

Page 46: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 29 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

RTD - simulation Simulation Method using

Fluke 5522A

t= Temperature

400 ⁰C to 600 ⁰C / RTD-

pt385,500 Ω

0.17 x 10-3 t + 0.14 m ⁰C Laboratory

/Customer

Premises 600 ⁰C to 630 ⁰C / RTD-

pt385,500 Ω

0.20 x 10-3 t + 0.097 m ⁰C

-200 ⁰C to -80 ⁰C /

RTD-pt385,1000 Ω

0.17 x 10-3 t + 0.60 m ⁰C

-80 ⁰C to 0.003 ⁰C /

RTD-pt385,1000 Ω

1.16 x 10-0 t + 3.47 m ⁰C

0.03 ⁰C to 100 ⁰C /

RTD-pt385,1000 Ω

0.43 x 10-3 t + 3.46 m ⁰C

100 ⁰C to 260 ⁰C / RTD-

pt385,1000 Ω

0.22 x 10-3 t + 0.37 m ⁰C

260 ⁰C to 300 ⁰C / RTD-

pt385,1000 Ω

0.23 x 10-3 t + 0.19 m ⁰C

300 ⁰C to 400 ⁰C / RTD-

pt385,1000 Ω

0.20 x 10-3 t + 0.17 m ⁰C

Page 47: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 30 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

RTD - simulation Simulation Method using

Fluke 5522A

t= Temperature

400 ⁰C to 600 ⁰C / RTD-

pt385,1000 Ω

0.14 x 10-3 t + 0.19 m ⁰C Laboratory

/Customer

Premises 600 ⁰C to 630 ⁰C / RTD-

pt385,1000 Ω

0.42 x 10-3 t + 0.046 m ⁰C

-80 ⁰C to 0.003 ⁰C /

RTD-ptNi385,120 Ω

1.16 x 10-0 t + 3.47 m ⁰C

0.003 ⁰C to 100 ⁰C /

RTDptNi385,120 Ω

0.89 x 10-3 t + 3.47 m ⁰C

100 ⁰C to 260 ⁰C / RTD-

ptNi385,120 Ω

0.62 x 10-3 t + 0.13 m ⁰C

-100 ⁰C to 260 ⁰C /

RTD-Cu427,10 Ω

0.60 x 10-3 t + 0.19 m ⁰C

Thermocouple Simulation Method using

Fluke 5522A

t= Temperature

600 C to 800 C / B-

T/C

0.64 x 10-3 t + 0.31 m C Laboratory

/Customer

Premises 800 C to 1000 C / B-

T/C

0.39 x 10-3 t + 0.38 m C

Page 48: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 31 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Thermocouple Simulation Method using

Fluke 5522A

t= Temperature

1000 C to 1550 C / B-

T/C

0.22 x 10-3 t + 0.49 m C Laboratory

/Customer

Premises 1550 C to 1820 C / B-

T/C

0.21 x 10-3 t + 0.38 m C

0.01 C to 150 C / C-

T/C

2.24 x 10-3 t + 12 m C

150 C to 650 C / C-

T/C

0.46 x 10-3 t + 1.18 m C

650 C to 1000 C / C-

T/C

0.36 x 10-3 t + 0.47 m C

1000 C to 1800 C / C-

T/C

0.32 x 10-3 t + 0.32 m C

1800 C to 2316 C / C-

T/C

0.42 x 10-3 t + 0.16 m C

-250 C to -100 C / E-

T/C

2.31 x 10-3 t + 0.40 m C

Page 49: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 32 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Thermocouple Simulation Method using

Fluke 5522A

t= Temperature

-100 C to -25 C / E-

T/C

1.84 x 10-3 t + 1.78 m C Laboratory

/Customer

Premises -25 C to 350 C / E-

T/C

0.56 x 10-3 t + 34.70 m C

350 C to 650 C / E-

T/C

0.28 x 10-3 t + 1.03 m C

650 C to 1000 C / E-

T/C

0.24 x 10-3 t + 0.70 m C

-210 C to -100 C / J-

T/C

1.48 x 10-3 t + 0.66 m C

-100 C to -30 C / J-

T/C

1.84 x 10-3 t + 1.55 m C

-30 C to 150 C / J-T/C 1.05 x 10-3 t + 79.1 m C

150 C to 760 C / J-

T/C

0.26 x 10-3 t + 2.01 m C

760 C to 1200 C / J-

T/C

0.22 x 10-3 t + 0.65 m C

Page 50: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 33 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Thermocouple Simulation Method using

Fluke 5522A

t= Temperature

-200 C to -100 C / K-

T/C

1.91 x 10-3 t + 0.53 m C Laboratory

/Customer

Premises -100 C to -25 C / K-

T/C

2.07 x 10-3 t + 1.59 m C

-25 C to 120 C / K-

T/C

1.29 x 10-3 t + 88 m C

120 C to 1000 C / K-

T/C

0.30 x 10-3 t + 2.02 m C

1000 C to 1372 C /

K-T/C

0.34 x 10-3 t + 0.34 m C

-200 C to -100 C / L-

T/C

2.14 x 10-3 t + 0.47 m C

-100 C to 0.01 C / L-

T/C

1.16 x 10-0 t + 11.6 m C

0.01 C to 800 C / L-

T/C

0.36 x 10-3 t + 11.6 m C

Page 51: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 34 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Thermocouple Simulation Method using

Fluke 5522A

t= Temperature

800 C to 900 C / L-

T/C

0.22 x 10-3 t + 0.72 m C Laboratory

/Customer

Premises -200 C to -100 C /

N-T/C

2.31 x 10-3 t + 0.43 m C

-100 C to -25 C / N-

T/C

2.53 x 10-3 t + 1.3 m C

-25 C to 120 C / N-

T/C

1.44 x 10-3 t + 96.8 m C

120 C to 410 C / N-

T/C

0.51 x 10-3 t +1.40 m C

410 C to 1300 C / N-

T/C

0.24 x 10-3 t + 0.89 m C

0.01 C to 250 C / R-

T/C

2.59 x 10-3 t +11.50 m C

250 C to 400 C / R-

T/C

1.01 x 10-3 t + 0.43 m C

Page 52: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 35 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Thermocouple Simulation Method using

Fluke 5522A

t= Temperature

400 C to 1000 C / R-

T/C

0.38 x 10-3 t + 0.61 m C Laboratory

/Customer

Premises 1000 C to 1767 C /

R-T/C

0.26 x 10-3 t + 0.40 m C

0.01 C to 250 C / S-

T/C

2.13 x 10-3 t + 11.5 m C

250 C to 1000 C / S-

T/C

0.42 x 10-3 t + 0.80 m C

1000 C to 1400 C /

S-T/C

0.31 x 10-3 t + 0.38 m C

1400 C to 1767 C /

S-T/C

0.30 x 10-3 t + 0.29 m C

-250 C to -150 C / T-

T/C

2.91 x 10-3 t + 0.25 m C

-150 C to 0.01 C / T-

T/C

1.16 x 10-0 t + 11.6 m C

Page 53: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 36 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Thermocouple Simulation Method using

Fluke 5522A

t= Temperature

0.01 C to 120 C / T-

T/C

1.45 x 10-3 t + 11.5 m C Laboratory

/Customer

Premises 120 C to 400 C / T-

T/C

0.40 x 10-3 t + 1.76 m C

-200 C to 0.01 C / U-

T/C

1.16 x 10-0 t + 11.6 m C

0.01 C to 600 C / U-

T/C

0.50 x 10-3 t + 11.6 m C

DC Power Direct Method using Fluke

5522A

with PQ Option

P= Power

33 mV / 0.33 mA 0.8 x 10-3 P Laboratory

/Customer

Premises

33 mV / 330 mA 0.7 x 10-3 P 1020 V / 0.33mA 0.6 x 10-3 P 1020 V / 330 mA 0.5 x 10-3 P 33 mV / 0.33 A 0.7 x 10-3 P

33 mV / 3 A 0.1 x 10-3 P 1020 V / 0.33 A 0.5 x 10-3 P

1020 V / 3 A 0.1 x 10-3 P

Page 54: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 37 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

DC Power Direct Method using Fluke

5522A

with PQ Option

P= Power

33 mV / 3 A 2.6 x 10-3 P Laboratory

/Customer

Premises

33 mV / 20.5 A 1.3 x 10-3 P 1020 V / 3 A 0.7 x 10-3 P

1020 V / 20.5 A 0.9 x 10-3 P

AC Power Direct Method using Fluke

5522A

with PQ Option

P= Power

45 Hz to 65 Hz @ PF=1 Laboratory

/Customer

Premises 33 mV / 3.3 mA 8.6 x 10-3 P

33 mV / 329.99 mA 0.02 x 10-3 P 329.999 mV / 3.3 mA 0.02 x 10-3 P 329.999 mV / 329.99

mA 0.01 x 10-3 P

330 mV / 3.3 mA 2.1 x 10-3 P 330 mV / 329.99 mA 0.01 x 10-3 P

1020 V / 3.3 mA 0.5 x 10-3 P 1020 V / 329.99 mA 0.01 x 10-3 P

33 mV / 0.33 A 0.6 x 10-3 P 33 mV / 20.5 A 0.2 x 10-3 P

329.999 mV / 0.33 A 0.01 x 10-3 P

Page 55: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 38 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

AC Power Direct Method using Fluke

5522A

with PQ Option

P= Power

329.999 mV / 20.5 A 0.05 x 10-3 P Laboratory

/Customer

Premises

330 mV / 0.33 A 0.4 x 10-3 P 330 mV / 20.5 A 0.08 x 10-3 P 1020 V / 0.33 A 0.21 x 10-3 P 1020 V/ 20.5 A 0.20 x 10-3 P

Resistance Meters

(micro, milli, ohm,

kohm, Meg ohm)

Direct Method using

Decade Resistance Boxes

R= Resistance

Up to 200 μΩ 1.6 x 10-3 R Laboratory

/Customer

Premises

>200 to 400 μΩ 1.5 x 10-3 R >400 μΩ to 600 mΩ 1.5 x 10-3 R

>600 mΩ to 600 Ω 1.9 x 10-3 R >600 Ω to 100 kΩ 2.1 x 10-3 R >100 kΩ to 300 kΩ 1.9 x 10-3 R >300 kΩ to 600 kΩ 1.9 x 10-3 R >600 kΩ to 3 MΩ 14 x 10-3 R

Insulation Resistance

Tester

Direct Method using

Decade Meg ohm Box

R= Insulation Resistance

Up to 9.99 MΩ 14 x 10-3 R Laboratory

/Customer

Premises

10 MΩ to 99.9 MΩ 12 x 10-3 R 100 MΩ to 1000 MxΩ 12 x 10-3 R

Page 56: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Electrical Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 39 of 39 DAC-70-02/F34-B, rev. b (24-11-2016)

Oscilloscope Direct Method using Fluke

5522A with SC1100

Option

Relative Deviation of the

vertical Axis

U= Voltage

6.6 V/ 50 Ω load 2.8 x 10-3 U Laboratory

/Customer

Premises

130 V/ 1 MΩ load 2.8 x 10-3 U

Oscilloscope Band Width

f= Frequency

Up to 1.1 GHz 14 x 10-3 f

Page 57: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Pressure Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 1 of 3 DAC-70-02/F34-B, rev. b (24-11-2016)

Pneumatic Pressure

Calibration of Digital &

Analogue Pressure

Gauges

GTS-WP-02

Based on the requirements of

BS EN 837-1 : 1998 Using

Druck DPI 610, DPI 620

up to 20 bar

0.02%

Laboratory

/ Customer

Premises

Pneumatic Pressure

Transducers,

Transmitters and

Switches

GTS-WP-02

Using Druck DPI 610, DPI 104

and Fluke 8846 multimeter

Up to 100 bar 0.11 %

Hydraulic Pressure

Transducers,

Transmitters and

Switches

GTS-WP-02

Using Budenberg DWT 580HX

Piston Cylinder 030L & Fluke

8846 multimeter

Up to 1200 bar 0.11 % Laboratory

Page 58: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Pressure Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 2 of 3 DAC-70-02/F34-B, rev. b (24-11-2016)

Hydraulic pressure

Digital & Analogue

Pressure Gauges &

pressure modules

GTS-WP-02

Using Budenberg DWT 580HX

Piston Cylinder 030L

Up to 1200 bar 0.02 % Laboratory

Vacuum gauge

calibration

GTS-WP-03

Based on the requirements of

BS EN 837-1 : 1998 and

ISO/TS 3567

Using Druck DPI 610, DPI 620

– 1 to 0 bar 1 mbar Laboratory

/ Customer

Premises

Calibration of hydraulic

pressure balance

GTS-WP-143 based on OIML

R110 and EURAMET cg-3

Version 1.0 (03/2011)

Up to 1400 bar 100 ppm

FSD

Laboratory

Page 59: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Pressure Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 3 of 3 DAC-70-02/F34-B, rev. b (24-11-2016)

Calibration of pneumatic

pressure balance

GTS-WP-143 based on

OIML R110 and EURAMET

cg-3 Version 1.0 (03/2011)

-1 to 25 bar 100 ppm

FSD Laboratory

Page 60: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Volume Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration. P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 1 of 1 DAC-70-02/F34-B, rev. b (24-11-2016)

Pipette

Fixed and Variable

Volume

Gravimetric Method Up to 100 μl 0.57μl Laboratory

100 to 2000 μl 0.57 μl to 1.30

μl

Page 61: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Force Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 1 of 1 DAC-70-02/F34-B, rev. b (24-11-2016)

Force

Verification /Calibration

of Compression testing

machines

GTS-WP-06

based on BS EN ISO 7500-1

3000 kN down to

300 kN

0.24%

of indicating

reading using

force transducer

class 1, ISO 376

Customer

Premises

Force

Verification /Calibration

of Compression testing

machines

GTS-WP-06

based on BS EN ISO 7500-1

3000 kN down to

200 kN

0.45%

of indicating

reading using

force transducer

class 2, ISO 376

Customer

Premises

Page 62: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Mass Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 1 of 2 DAC-70-02/F34-B, rev. b (24-11-2016)

concrete and asphalt

batching plants (Hopper

Scale)

Hopper Scale calibration of

concrete batching plants &

Euramet CG-18,

ASTM C94/C94M

& NIST Handbook 44

5000 kg 0.05% Customer

Premises

Class M masses Masses according to GTS-WP-

17

Up to 200 g ±0.50 mg Laboratory

1 kg ±2.3 mg

2kg ±9.7mg

5 kg ±12 mg

10 kg ±155 mg

20 kg ±170 mg

Page 63: SCOPE OF ACCREDITATION · 2020. 2. 28. · Based on JIS B 7515 (using Calibrations tester & ULMS) Up to 400mm/0.001mm 3.2μm Laboratory Up to 400mm/0.01mm 6.5μm

DDUUBBAAII AACCCCRREEDDIITTAATTIIOONN

DDEEPPAARRTTMMEENNTT

اعتماد تقييم المطابقة إدارة

SCOPE OF ACCREDITATION

Mass Calibration

General Const. Lab Calibration LLC

Industrial Area # 3, Sharjah- United Arab Emirates

Issue No: 06 Issue date: 18- 06- 2017 Accreditation Certificate No: LB-040-CAL

DETAILS OF THE APPLICABLE RANGE OF CALIBRATION AND MEASUREMENT CAPABILITY FOR

THE SCOPE OF ACCREDITATION

Calibration Field/

Measured Quantity

Calibration Method Range and

Specification

Calibration

Measurement

Capability

(CMC)*

Location

Note: For history details of accredited conformity assessment activities, please refer to Dubai Accreditation Department, Dubai Municipality.

Calibration and Measurement Capability (CMC) is the smallest uncertainty of measurement that a laboratory can achieve within its scope of

accreditation when performing more or less routine calibrations of nearly ideal measurement standards or nearly ideal measuring equipment.

Calibration and Measurement Capabilities represent expanded uncertainties expressed at approximately the 95 % level of confidence, usually using

a coverage factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory may be greater than the CMC

due to the behavior of the customer’s device and to influences from the circumstances of the specific calibration.

P.O Box: 67, DUBAI-UAE., TEL: 00971-4-3027445, FAX: 00971-4-3362381

Email: [email protected] web site: http://www.dac.gov.ae

Page 2 of 2 DAC-70-02/F34-B, rev. b (24-11-2016)

Top loading direct

reading weighing

balance

GTS-WP-01

Based on the requirements of

ASTM E 898 : 1988

(re approved 2005)

& Euramet CG-18

Calibrated weights

E1 (1 mg to 100g), E2(1 mg to

5 kg),

F1(1 mg to 5 kg),

F2 (10 kg & 20 kg) and

M1 (5 kg to 530 kg)

Up to 100 g 0.0001 g Laboratory/

Customer

Premises Up to 210 g 0.0002 g

Up to 0.5 kg 0.0006 g

Up to 1 kg 0.001 g

Up to 5 kg 0.009 g

Up to 10 kg 0.013 g

Up to 30 kg 0.23 g

Up to 100 kg 0.45 g

Up to 500 kg 2.27 g