Shanaham Et Al - TENS x IFC

Embed Size (px)

Citation preview

  • 8/3/2019 Shanaham Et Al - TENS x IFC

    1/7

    ARTICLE IN PRESSPHYST-222; No.of Pages 7

    Physiotherapy xxx (2006) xxxxxx

    Comparison of the analgesic efficacy of interferential therapy andtranscutaneous electrical nerve stimulation

    Christine Shanahan a, Alex R. Ward b,, Val J. Robertson c,d

    a School of Physiotherapy, La Trobe University, Vic. 3086, Australiab Department of Human Physiology and Anatomy, La Trobe University, Vic. 3086, Australia

    c University of Newcastle, University Drive, Callaghan, NSW. 2308, Australiad North Sydney Central Coast Health Service, Gosford Hospital, Cnr Holden & Racecourse Rds, Gosford, NSW. 2250, Australia

    Abstract

    Objective To compare the analgesic efficacy of interferential therapy (IFT) and transcutaneous electrical nerve stimulation (TENS) using an

    experimental cold pain model.

    Design Randomised controlled trial with repeated measures design.

    Setting University research laboratory.

    Participants Twenty healthy subjects.

    Interventions IFT and TENS applied to each subject on different days.

    Main outcome measures Cold pain threshold (time), intensity and unpleasantness (visual analogue scales).

    Results The mean cold pain threshold with a TENS intervention was higher than that with IFT. A training effect was evident as subjects

    responses become more consistent with repeated exposure to stimulation and the testing procedure. Using data from the second testing

    sessions, the differences in pain threshold between IFT and TENS for the two during-intervention (T3 and T4) measures were statistically

    significant (T3 difference in the means 5.9 seconds, 99% confidence interval 3.1 to 8.7 seconds; T4 difference in the means 6.6 seconds, 99%

    confidence interval 3.8 to 9.4 seconds). No significant differences were identified in pain intensity and unpleasantness ratings.

    Conclusions TENS is more effective than IFT at increasing cold pain thresholds in healthy subjects, and this effect increases with repeated

    exposures. Future trials should include a familiarisation session prior to testing to increase the consistency of subjects responses. The clinical

    implications of these effects need investigation.

    2006 Chartered Society of Physiotherapy. Published by Elsevier Ltd. All rights reserved.

    Keywords: Transcutaneous nerve stimulation; Cold pain; Pain measures; Experimental pain; Electroanalgesia; Hypoalgesia

    Introduction

    Electrical stimulation is used regularly by physiother-

    apists. Indications include pain management, muscle re-

    education, strengthening and oedema reduction [1,2]. Ofthese, pain management is the most common use, and the

    current types used are interferential therapy (IFT) and tran-

    scutaneous electrical nerve stimulation (TENS) [3].

    The physical properties of IFT and TENS are quite dif-

    ferent. Conventional TENS is a type of low-frequency stim-

    ulation, commonly applied at frequencies of 1200 Hz [1].

    Corresponding author. Tel.: +61 3 9479 5814; fax: +61 3 9479 5784.

    E-mail address: [email protected] (A.R. Ward).

    The current is pulsed and the flow of charge is interrupted

    rather than continuous. IFT is a type of medium-frequency

    alternating current, with carrier frequencies of 110 kHz [4].

    Unlike TENS, IFT current is sinusoidal and provides a con-

    tinuous flow of charge. IFT is usually delivered in bursts witha frequency in the biological range (1150 Hz) [2]. Despite

    the widespread use of IFT and TENS for pain management,

    there is a lack of scientificevidence supporting aspects of their

    claimed effectiveness. Consequently, in a clinical setting,

    selection of the type and parameters of electrical stimulation

    is largely based on theoretical or anecdotal evidence [2,4].

    The aim of the present study was to compare the analgesic

    effectiveness of IFT and TENS. A literature search was con-

    ducted using the keywords interferential and/or TENS,

    0031-9406/$ see front matter 2006 Chartered Society of Physiotherapy. Published by Elsevier Ltd. All rights reserved.

    doi:10.1016/j.physio.2006.05.008

    Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous

    electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008

    mailto:[email protected]://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008mailto:[email protected]
  • 8/3/2019 Shanaham Et Al - TENS x IFC

    2/7

    ARTICLE IN PRESSPHYST-222; No.of Pages 7

    2 C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx

    and pain or pain threshold (not dental, teeth, anaesthesia

    or labour). The search was limited to studies of human sub-

    jects, written in the English language, that reported the use

    of ITF and/or TENS for treating musculoskeletal pain. The

    databases searched were Embase, Medline, Cinahl, Sports

    Discus, The Cochrane Library and PubMed (19662005).

    Thirty-eight articles were identified.These studies used a range of pain-induction techniques:

    cold [58], heat [9], mechanical [10] and ischaemic [11]. The

    most consistent analgesic effects were observed with cold-

    induced pain [6,7], with more varied results obtained using

    other pain-induction techniques [1012].

    Most studies examined the hypoalgesic effects of TENS

    [7,1118], and a smaller number studied IFT [8,1922]. Only

    four of the studies compared the effectiveness of TENS and

    IFT directly [6,9,10,23]. Three of these reported that IFT

    and TENS were significantly more effective at modulating

    pain than sham interventions or control groups (the exception

    being Alves-Guerreiro et al.[10]). However, no significant

    differences in the analgesic effectiveness of IFT and TENSwere identified in any of the four studies. This was possibly

    due to inadequate statistical power as the question addressed

    was not do the interventions have an effect? but do the

    effects differ? If both interventions had an effect, any differ-

    ences could easily be obscured or more difficult to detect.

    A common limitation across existing comparisons of

    TENS and IFT is small sample sizes, ranging from 7 [6] to 18

    [9] subjects per group. Unlessa within-subject design is used,

    the risk of a type II error is high. For example, the Johnson

    and Tabasam study [6] used seven subjects per group, and

    each subject was tested with a single type of stimulation. The

    observed difference (i.e. TENS more effective than IFT) wasstatistically insignificant (P = 0.09). However, a calculation

    using 10 subjects and the same means and standard devia-

    tions produces a P-value of less than 0.05, suggesting that

    the group was too small given the effect size, resulting in a

    type II statistical error.

    The most effective and practical method to compare IFT

    and TENS is to use a within-subject repeated measures

    design, and to administer both to each subject. This design

    eliminates between-subject variation and increases the statis-

    tical power of the study. Existing comparative studies did not

    use this approach.

    The aim of the present study was to compare the analgesic

    effectiveness of IFT and TENS using the same experimental

    cold pain model as Johnson and Tabasam [6], but with larger

    subject numbers and a within-subject design. A secondary

    aim was to investigate whether familiarisation with electrical

    stimulation and the testing procedure affected the consistency

    of subjects responses.

    The following hypotheses were tested:

    TENS is more effective than IFT at increasing the cold

    pain threshold of healthy subjects;

    there is no difference in the effectiveness of IFTand TENS

    at altering pain intensity and unpleasantness; and

    familiarisation with stimulation and the testing procedure

    affects the consistency of subjects responses (i.e. there is

    a training effect).

    Methods

    Research design

    A single blind randomised controlled trial with a repeated

    measures design was used. The design was based on the

    method used in previous research (e.g. [6,24]), but was modi-

    fied to increase statistical power by using a larger sample size

    and a training session identical to the testing session. Each

    subject attended on two separate days and IFT and TENS

    were given in a pre-determined order. Subjects were blinded

    to the treatment modality being administered. The investi-

    gator was not blinded as the investigators role was simply

    to record the outcome measures as numerical values [mea-

    sured time or reported visual analogue scale (VAS) rating],rather than make any subjective assessment or measurement;

    thus, the risk of assessor bias being able to influence the

    results was minimal. A sham group was not included as the

    aim was specifically to compare the effectiveness of two

    current types that had previously been shown to be more

    effective than sham stimulation [57,9,24]. A single investi-

    gator administered the electrical stimulation and recorded all

    measurements.

    The independent variable was the type of stimulation

    administered (IFT or TENS). The dependent variables were

    the time to pain threshold (seconds), and pain intensity and

    unpleasantness as rated by subjects on two, 10-cm horizontalVASs.Painintensitywas operationally defined as howstrong

    the pain in your hand feels, whilst pain unpleasantness was

    how uncomfortable the pain in your hand feels.

    Subjects

    Twenty volunteers (10 males, 10 females) aged between

    18 and 32 years (mean age 21.7 years, standard deviation 3.5

    years) were recruited into the study. Potential participants

    were excluded if their non-dominant arm was affected by

    any orthopaedic, neurological or circulatory pathology, or

    if the skin in the area under the electrodes was broken or

    irritated.Participants were instructednot to takeany analgesic

    or anti-inflammatory medication (or apply them as creams or

    gels to their non-dominant forearm) in the 12 hours prior to

    eachtesting session. No volunteers were excludedwhenthese

    criteria were applied.

    Apparatus

    Electrical stimulation

    A Chattanooga Intelect (Intelect Legend Stim, Chat-

    tanooga Group Inc., Hixson, TN, USA) was used to deliver

    the pre-modulated IFT current, and a Metron Multistimu-

    Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous

    electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008

    http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008
  • 8/3/2019 Shanaham Et Al - TENS x IFC

    3/7

    ARTICLE IN PRESSPHYST-222; No.of Pages 7

    C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx 3

    lator (Metron Medical Australia Pty. Ltd., Carrum Downs,

    Victoria, Australia) was used for TENS. These stimulators,

    similar in size and appearance, meant that testing conditions

    were kept as uniform as possible for the two interventions.

    The display screen and control panels were not visible to

    subjects, and subjects were given no indication of which cur-

    rent type was being delivered. Stimulation parameters werekept constant throughout the procedure, and only the current

    intensitywasadjustedduring the session.Current was applied

    via two Stimtrode 50 mm 90 mm rectangular self-adhesive

    electrodes (Axelgaard Manufacturing Co. Ltd., Fallbrook,

    CA, USA). TENS stimulation used 200-s biphasic pulses

    (100-s phaseduration) anda pulse frequency of 100 Hz.Pre-

    modulated IFT was applied with a carrier frequency of 5kHz

    (100-s phase duration) and a beat frequency of 100 Hz.

    Cold pain induction

    Cold pain induction required two water baths of uniform

    size andshape; onewas maintained at 37 C and the other was

    maintained at 0 C with ice and water. A Heidolph heater stir-rer unit (AccurexEquipment,Brunswick, Victoria, Australia)

    was used in each bath to ensure that the water temperatures

    (both monitored with a thermometer) remained constant.

    Procedure

    Prior to the first session

    Eligibility was confirmed and informed consent was

    obtained. Each participant wasallocated at random to a group

    by selecting a sealed envelope (marked for male or female)

    containing a pre-arranged order of testing conditions. Within

    the 10 envelopes designated to each sex, five specified thatIFT would be delivered in Session A and five specified that

    TENS would be delivered in Session A. Subjects were not

    informed of the group to which they were assigned. The sub-

    ject was seated and their non-dominant forearm was cleaned

    with an alcohol swab. The two electrodes were then posi-

    tioned on the subjects non-dominant forearm, one anterior

    and one posterior, with the centre point of the electrode lying

    equidistant to the elbow and wrist. The electrode alignment

    was standardised between subjects and between sessions.

    Cold pain induction

    The cold-induced pain cycle is based on a method used

    in previous research [68,24]. Six cold pain cycles, each of

    10-minutes duration, were conducted sequentially for a 1-

    hour test session. The procedure is depicted in Fig. 1. Each

    cycle commenced as the subject placed their hand (up to the

    distal wrist crease) in a warm water bath. After 5 minutes,

    the subject transferred their hand to a cold water bath. The

    subject was instructed to focus on the sensations in their

    immersed hand until they felt the onset of deep, dull, aching

    pain and to then say pain. The time from immersion in cold

    water to this point was recorded as the pain threshold (in sec-

    onds). The subject was required to leave their hand in the

    cold water for a further 30 seconds. Immediately upon with-

    Fig. 1. Procedure for (a) one cold pain cycle and (b) one testing session

    consisting of six cold pain cycles, based on Johnson and Tabasam [6].

    drawal from the cold water, the subject completed a VAS

    rating of pain intensity and pain unpleasantness (where zero

    represented not intense/unpleasant at all and 10 represented

    most intense/unpleasant pain imaginable). The subject then

    rested until 10 minutes had elapsed since the beginning of the

    cycle, at which time the next cycle was commenced with re-

    immersion of the hand in warm water.

    Electrical stimulation

    At the start of Cycle 3 (i.e. 20 minutes into the session), the

    allocated electrical stimulator was turned on. The investiga-tor increased the current intensity, as directed by the subject,

    until they experienced a strong but comfortable level of

    stimulation. The current intensity was adjusted upon request

    to maintain this level, except when the subjects hand was

    in the cold water bath. Stimulation continued uninterrupted

    for 20 minutes, which constituted two complete pain cycles

    (Cycles 3 and 4). Two further pain cycles (Cycles 5 and 6)

    were completed once the stimulation was turned off. The pro-

    cedure was identical for TENS and IFT delivery. Thus, two

    pre-intervention measures, two during-intervention measures

    and two post-intervention measures were obtained.

    Session B

    Session B was conducted no earlier than 12 hours after

    Session A to minimise the risk of a carry-over effect. Subject

    eligibility was reconfirmed, and the cold-pain-induction pro-

    cedure was repeated using the alternative type of stimulation.

    All 20 subjects participated in both sessions. After comple-

    tion of Session B, subjects were asked in which, if either, of

    the two sessions they felt stimulation was more effective and

    also more comfortable. Subjects were also given an oppor-

    tunity to make any other comments about the stimulation or

    testing procedure.

    Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous

    electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008

    http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008
  • 8/3/2019 Shanaham Et Al - TENS x IFC

    4/7

    ARTICLE IN PRESSPHYST-222; No.of Pages 7

    4 C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx

    Data analysis

    Data were analysed using SPSS Version 11. For data anal-

    ysis purposes, the group who received IFT in Session A and

    TENS in Session B were referred to as Group 1, and those

    who received TENS in Session A and IFT in Session B were

    referred to as Group 2.Results were analysed in two stages. First, descriptive

    statistics were obtained for each group (1 or 2; TENS first

    or IFT first) and each condition (TENS or IFT), and tests of

    normality were made. Data from the first session for each

    group were much more scattered than data from the second

    session, with standard deviations being approximately dou-

    ble in the first session. Consequently, the first session was

    considered to be the training/familiarisation session and only

    the data from the second session were used for subsequent

    analysis.

    Data from the second session were analysed using a

    two-way repeated measures analysis of variance (ANOVA)

    (cycle stimulus type). Where justified by the ANOVA, posthoc tests were performed.

    Previous research has shown that both TENS and IFT

    elevate the cold pain threshold [6,9,23]. The present sta-

    tistical analysis was therefore restricted to testing whether

    the two current types had significantly different effects.

    Pre-planned comparisons were made to reduce the Bon-

    ferroni correction needed when multiple comparisons are

    performed and to reduce the risk of a type II error. For sin-

    gle comparisons, a 95% confidence interval can be used,

    but the confidence interval must be set higher for multiple

    comparisons.

    Pain threshold data

    Pain thresholddata were analysed using changefrom base-

    line (seconds)for each intervention. The two pre-intervention

    measurements from Cycles 1 and 2 (T1 and T2) were aver-

    aged to give a baseline for each session, which was then

    subtracted from the pain threshold measurements taken dur-

    ing Cycles 3 (T3), 4 (T4), 5 (T5) and 6 (T6). Resulting data

    were subjected to preliminary analysis by calculating means

    and standard deviations, and performing tests of normality.

    An ANOVA was conducted using the data from the sec-

    ond session, which clearly had greater consistency due to

    subjects familiarisation and practice. Finally, comparisons

    of the change in pain threshold at T3 and T4 were con-

    ducted; the change from baseline data for IFT was compared

    with that for TENS at these points. The rationale for this

    was that Johnson and Tabasam [6] first noticed a differ-

    ence between IFT and TENS at T3, which reached a peak

    at T4 before returning to baseline. The choice of only two

    statistical comparisons meant that with a Bonferroni cor-

    rection, a 97.5% confidence interval ( = 0.025) could be

    used to assess the significance of any TENS versus IFT

    differences.

    Pain intensity and unpleasantness data

    Pain intensity and unpleasantness ratings were compared

    in terms of change from baseline in the same manner as

    the pain threshold data. Similarly, ANOVAs were conducted

    using data from Session B alone. Post hoc testing was not

    justified by the ANOVAs.

    Subjects comments

    The frequencies of subjects comments regarding the

    effectiveness and comfort of the interventions were calcu-

    lated and compared.

    Results

    One of the 20 subjects failed to reach a pain threshold

    within the allocated 5-minute time period in 11 of the 12

    measurement cycles. As such, this setof results was excluded,

    leaving 19 data sets for analysis.

    Pain thresholds

    Fig. 2 shows the mean change in pain threshold with

    electrical stimulation intervention (Cycles 3 and 4) and fol-

    lowing intervention (Cycles 5 and 6) for the second set of

    measurements (Session B, post-familiarisation). Thresholds

    increasedwhen stimulation was applied, and the increase was

    greater with TENS than with IFT. The difference between the

    two current types was greatest at T4, near the end of stimu-

    lation; the increase in pain threshold for TENS was nearlytwice that for IFT. A two-way ANOVA (cycle stimulus

    type) using data from Session B found that between-cycle

    differences were significant [F(3,1) = 3.88, P = 0.033] and

    between-stimulus type effects were, on average, insignifi-

    Fig. 2. Mean change in pain threshold (seconds) using data from the second

    (post-familiarisation)testingsession.Error barsare standarddeviations. IFT,

    interferential therapy; TENS, transcutaneous electrical nerve stimulation.

    Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous

    electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008

    http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008
  • 8/3/2019 Shanaham Et Al - TENS x IFC

    5/7

    ARTICLE IN PRESSPHYST-222; No.of Pages 7

    C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx 5

    cant [F(1,3) = 2.13, P = 0.164], but a significant interaction

    effect (cycle stimulus type) was observed [F(1,3) = 4.36,

    P = 0.023]. The significant interaction effect indicates the

    possibility that there is a between-stimulus type effect, which

    is only evident at particular cycles. Inspection of Fig. 2

    suggests that there may be a between-stimulus type effect

    at T3 and T4 but not at T5 and T6. At T3, the meanTENS threshold change was 6.25 seconds, compared with

    0.35 seconds for IFT. At T4,when thedifference was greatest,

    the mean TENS threshold change was 9.75 seconds, com-

    pared with 3.15 seconds for IFT. The differences between

    pain thresholds at T3 and T4 were significant at the

    P = 0.01 level (T3 difference in the means = 5.9 seconds,

    99% confidence interval 3.1 to 8.7 seconds; T4 difference

    in the means= 6.6 seconds, 99% confidence interval 3.8

    to 9.4 seconds). Ninety-nine percent confidence intervals

    were used to meet the need for Bonferronis correction

    and to further minimise the risk of a type I statistical

    error.

    Pain intensity

    As with the threshold data, the pain intensity results

    showed a much greater scatter in Session A than in Session B.

    Standard deviations for Session A were typically 30% greater

    than those for Session B. Fig. 3 compares Session B results

    for pain intensity. Both IFT and TENS decreased the pain

    intensity ratings but to a similar extent. There is the sugges-

    tion that TENS is more effective but that the difference is not

    significant.

    A two-way ANOVA (cycle stimulus type) was con-

    ducted using data from Session B. This indicated thatbetween-cycle differences were significant [F(1,3) = 3.64,

    P = 0.04], but between-stimulus type effects were not

    [F(1,3) = 0.195, P = 0.67] and there was no significant inter-

    action effect [F(1,3) = 0.32, P = 0.81].

    Fig.3. Mean change in painintensity [visual analoguescale (VAS)]showing

    averageddata for the second (post-familiarisation)testing session.Error bars

    are standard deviations. IFT, interferential therapy; TENS, transcutaneous

    electrical nerve stimulation.

    Fig. 4. Mean change in pain unpleasantness [visual analogue scale (VAS)]

    showing averaged data for the second (post-familiarisation) testing session.

    Error bars are standard deviations. IFT, interferential therapy; TENS, tran-

    scutaneous electrical nerve stimulation.

    Pain unpleasantness

    Fig. 4 shows the results for pain unpleasantness, compar-

    ing results from Session B. Both TENS and IFTdecreased the

    pain unpleasantnessratings but to a similar extent. A two-way

    ANOVA (cycle stimulus type) was conducted using results

    from Session B. Between-cycle differences were insignifi-

    cant [F(1,3) = 0.66, P = 0.59], as was the between-stimulus

    type effect [F(1,3) = 0.014, P = 0.91]. There was no signifi-

    cant interaction effect [F(1,3) = 0.39, P = 0.76].

    Subjects comments

    Comfort of stimulation

    Twelve subjects (63%) reported that IFT was more com-

    fortable than TENS, and four subjects (21%) reported that

    TENS was more comfortable than IFT. Three subjects (16%)

    reported no difference.

    Effectiveness of stimulation

    The effectiveness ratings were exactly the same as the

    comfort ratings, i.e. 12 subjects (63%) reported that IFT was

    more effective than TENS at altering their response to the

    cold pain. Only four of the 19 subjects (21%) reported that

    TENS was more effective than IFT. The remaining three sub-

    jects reported no difference in perceived effectiveness. Thus,

    a clear majority thought that IFT was more effective and also

    that it was more comfortable than TENS.

    Discussion

    Training effect

    Training had a major impact on the consistency of sub-

    jects responses, as demonstrated by the decrease in standard

    deviations across all outcome measures from Session A to

    Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous

    electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008

    http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008
  • 8/3/2019 Shanaham Et Al - TENS x IFC

    6/7

    ARTICLE IN PRESSPHYST-222; No.of Pages 7

    6 C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx

    Table 1

    Pain threshold (time to pain onset) changes during (T3 and T4) and after (T5 and T6) interventions [interferential therapy (IFT) or transcutaneous electrical

    nerve stimulation (TENS)]

    IFT threshold changes (seconds)

    T3 T4 T5 T6

    Mean (S.D.), Group 1 (IFT first) 3.8 (9.6) 1.5 (4.7) 0.0 (6.8) 1.3 (6.1)

    Mean (S.D.), Group 2 (IFT second) 0.4 (3.6) 3.2 (3.7) 3.6 (5.0) 4.1 (4.0)

    TENS threshold changes (seconds)

    T3 T4 T5 T6

    Mean (S.D.), Group 2 (TENS first) 0.3 (9.4) 2.4 (9.9) 7.0 (8.6) 1.3 (12.2)

    Mean (S.D.), Group 1 (TENS second) 6.3 (7.9) 9.8 (7.7) 2.6 (4.2) 3.8 (3.9)

    Standard deviations (S.D.) were much reduced in the second experimental session.

    Session B (Table 1). The effect of repeated exposures to the

    test conditions was not reported in existing comparisons of

    IFT and TENS, and may have contributed to the absence of

    significant findings [5,6,9]. The present study highlights the

    importance of familiarising subjects prior to testing so thatresponses are more consistent. Future evaluations should,

    ideally, use three sessions per participant; one for training

    or familiarisation with either type of stimulator and two for

    testing.

    The presence of a training effect also suggests that, in

    a clinical context, the desired results of electrical stimula-

    tion for pain management may not be achieved in an initial

    application. The efficacy of stimulation may improve with

    subsequent exposures, so treatment may be more effective

    when used repeatedly. Clinically, this suggests that at least

    two trials of electrical stimulation are needed to determine

    the effectiveness of pain management reliably. This is con-sistent with a clinical finding that the effectiveness of IFT

    for treatment of arthritic knee pain increased over an 8-week

    treatment period [21].

    Pain thresholds

    The present study identified that TENS was significantly

    more effective than IFT at increasing the experimental cold

    pain threshold of healthy subjects, and that the difference

    between the two current types was greatest during a 20-

    minute period of stimulation (Fig. 2). AtT3 and T4, the mean

    pain threshold increases were greater with TENS than IFT.

    The difference was significant at the 99% confidence level.

    Contrary to these findings, IFT is used clinically in prefer-

    ence to TENS in the majority of physiotherapy departments

    [3]. The present findings, however, suggest that IFT is less

    effective butperhaps more comfortableand perceived as more

    effective than TENS, as reported by Johnson and Tabasam

    [23]. The choice of IFT or TENS in a clinical context depends

    on the relative importance of effectiveness and patient com-

    fort/acceptance. Unless patient acceptance or tolerance is a

    major issue, the findings suggest that TENS should be the

    clinical modality of choice and IFT should be used as a fall-

    back alternative.

    The properties of clinical pain, however, may differ from

    those of experimental pain in healthy subjects. Healthy sub-

    jects know that experimentally induced pain is strictly tempo-

    rary, so the psychological impact may be different. Similarly,

    the neurophysiological aspects of the pain, i.e. the originand pattern of neural activity, are also different. This means

    that future trials using a clinical population are necessary to

    progress this matter. If TENS is also shown to be more effec-

    tive in clinical trials, it should be adopted in preference to

    IFT unless comfort and acceptance are an issue.

    Pain intensity and unpleasantness

    The results for painintensityand unpleasantnesswere very

    similar. Although both IFT and TENS apparently decreased

    pain ratings to lower than baseline levels, the difference

    between the effects of the two types of stimulation was notstatistically significant. These results are consistent with find-

    ings from previous research usingthe samevariables [6,9,24].

    This calls into question whether pain intensity and unpleas-

    antness ratings are useful tools for measuring the effect of

    electrical stimulation, as these measures do not show change

    and they require an extension of the period of painful stimu-

    lation after onset.

    Subjects comments

    The majority of subjects reported that IFT was their pre-

    ferred type of stimulation, for both comfort and perceived

    effectiveness, despite results indicating that TENS is actually

    more effective at increasing the pain threshold. The subjects

    preference for IFT warrants further investigation.

    Conclusion

    The effectiveness of TENS at increasing pain threshold

    was greater than that for IFT. This difference in effectiveness

    of the two current types was statistically significant when

    the training effect was taken into consideration. The present

    study established that repeated exposures to stimulation and

    Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous

    electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008

    http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008
  • 8/3/2019 Shanaham Et Al - TENS x IFC

    7/7

    ARTICLE IN PRESSPHYST-222; No.of Pages 7

    C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx 7

    cold-induced pain have a significant effect on the consistency

    of subjects responses. Consequently, it is recommended that

    future studies should include familiarisation sessions prior

    to testing to allow for a training effect. The training effect

    may also have clinical implications, and multiple uses may

    be needed prior to evaluating effectiveness in a patient.

    Neither TENS nor IFT was more effective at altering painratings of intensity and unpleasantness. This result is con-

    sistent with the existing literature and raises questions about

    whether the continued useof these outcome measures is justi-

    fied in laboratory-based studies, particularly considering the

    increased discomfort that they impose on participants.

    The pain threshold results in the present study support the

    use of TENS in preference to IFT for the treatment of pain.

    This is contrary to current clinical practice,whichfavours IFT

    despite TENS machines being considerably cheaper, more

    portable, andmore practical for home use. Thepresent results

    challenge existing beliefs and clinical practices, and indicate

    the need for clinical trials.

    Ethical approval: Faculty Human Ethics Committee, Faculty

    of Health Sciences, La Trobe University, Victoria, Australia.

    Funding: The Schools of Physiotherapy and Human Bio-

    sciences, La Trobe University, Victoria, Australia.

    Conflicts of interest: None.

    References

    [1] Selkowitz D. Electrical currents. In: Cameron M, editor. Physical

    agents in rehabilitation. Philadelphia: WB Saunders Co.; 1999. p. 345

    427.

    [2] Robertson VJ, Ward AR, Low J, Reed A. Electrotherapy explained:

    principles and practice. 4th ed. Oxford: Butterworth-Heinemann; 2006.

    [3] Robertson VJ, Spurritt D. Electrophysical agents: implications of their

    availability and use in undergraduate clinical placements. Physiother-

    apy 1998;84:33544.

    [4] Kitchen S. Electrotherapy evidence-based practice. 11th ed. London:

    Churchill Livingstone; 2002.

    [5] Johnson MI, Tabasam G. A single-blind investigation into the hypoal-

    gesic effects of different swing patterns of interferential currents on

    cold-induced pain in healthy volunteers. Arch Phys Med Rehabil

    2003;84:3507.

    [6] Johnson MI, Tabasam G. A double blind placebo controlled inves-

    tigation into the analgesic effects of inferential currents (IFC) and

    transcutaneous electrical nerve stimulation (TENS) on cold-induced

    pain in healthy subjects. Physiother Theor Pract 1999;15:21733.[7] Johnson MI, Ashton CH, Bousfield DR, Thompson JW. Analgesic

    effects of different frequencies of transcutaneous electrical nerve

    stimulation on cold-induced pain in normal subjects. Pain 1989;39:

    2316.

    [8] Johnson MI, Wilson H. The analgesic effects of different swing pat-

    terns of interferential currents on cold-induced pain. Physiotherapy

    1997;83:4617.

    [9] Cheing GLY, Hui-Chan CWY. Analgesic effects of transcutaneous

    electrical nerve stimulation and interferential currents on heat pain in

    healthy subjects. J Rehabil Med 2003;35:159.

    [10] Alves-Guerreiro J, Noble JG, Lowe AS, Walsh DM. The effect of three

    electrotherapeutic modalities upon peripheral nerve conduction and

    mechanical pain threshold. Clin Physiol 2001;21:70411.

    [11] WalshDM, Liggett C, Baxter D, AllenJM. A double-blindinvestigation

    of the hypoalgesic effects of transcutaneous electricalnerve stimulation

    upon experimentally induced ischaemic pain. Pain 1995;61:3945.

    [12] Simmonds M, Wessel J, Scudds R. The effect of pain quality on the

    efficacy of conventional TENS. Physiother Can 1992;44:3540.

    [13] Barr JO, Nielsen DH, Soderberg GL. Transcutaneous electrical nerve

    stimulation characteristics for altering pain perception. Phys Ther

    1986;66:151521.

    [14] Cheing GL, Hui-Chan CW. Transcutaneous electrical nerve stimula-

    tion: nonparallel antinociceptive effects on chronic clinical pain and

    acute experimental pain. Arch Phys Med Rehabil 1999;80:30512.

    [15] Chesterton LS, Barlas P, Foster NE, Lundeberg T, Wright CC, Bax-

    ter GD. Sensory stimulation (TENS): effects of parameter manipula-

    tion on mechanical pain thresholds in healthy human subjects. Pain

    2002;99:25362.

    [16] Foster NE, Baxter F, Walsh DM, Baxter GD, Allen JM. Manipula-

    tion of transcutaneous electrical nerve stimulation variables has no

    effect on two models of experimental pain in humans. Clin J Pain

    1996;12:30110.

    [17] Roche PA, Gijsbers K, Belch JJ, Forbes CD. Modification of induced

    ischaemic pain by transcutaneous electrical nerve stimulation. Pain

    1984;20:4552.

    [18] Walsh DM, Lowe AS, McCormack K, Willer J, Baxter GD, Allen JM.

    Transcutaneous electrical nerve stimulation: effect on peripheral nerve

    conduction, mechanical painthreshold, andtactile thresholdin humans.

    Arch Phys Med Rehabil 1998;79:10518.

    [19] Tabasam G, Johnson MI. Electrotherapy for pain relief: does it work?

    A laboratory-based study to examine the analgesic effects of elec-

    trotherapyon cold-inducedpain in healthy individuals.Clin Effect Nurs1999;3:1424.

    [20] Johnson MI, Tabasam G. A single-blind placebo-controlled investiga-

    tion into the analgesic effects of interferential currents on experimen-

    tally induced ischaemic pain in healthy subjects. Clin Physiol Funct

    Imaging 2002;22:18796.

    [21] Adedoyin RA, Olaogun MOB, Fagbeja OO. Effect of interferential

    current stimulation in management of osteo-arthritic knee pain. Phys-

    iotherapy 2002;88:4939.

    [22] Stephenson R, Walker EM. The analgesic effects of interferential (IF)

    current on cold-pressor pain in healthy subjects: a single blind trial of

    three IF currents against sham IF and control. Physiother Theor Pract

    2003;19:99107.

    [23] Johnson MI, Tabasam GI. An investigation into the analgesic effects

    of interferential currents and transcutaneous electric nerve stimulation

    on experimentally induced ischaemic pain in otherwise pain-free vol-unteers. Phys Ther 2003;83:20823.

    [24] McManusFJ, WardAR, RobertsonVJ. Theanalgesic effects of interfer-

    ential therapy on two experimental painmodels: coldand mechanically

    induced pain. Physiotherapy 2006;92:95102.

    Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous

    electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008

    http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008