Upload
fabio-cardoso
View
215
Download
0
Embed Size (px)
Citation preview
8/3/2019 Shanaham Et Al - TENS x IFC
1/7
ARTICLE IN PRESSPHYST-222; No.of Pages 7
Physiotherapy xxx (2006) xxxxxx
Comparison of the analgesic efficacy of interferential therapy andtranscutaneous electrical nerve stimulation
Christine Shanahan a, Alex R. Ward b,, Val J. Robertson c,d
a School of Physiotherapy, La Trobe University, Vic. 3086, Australiab Department of Human Physiology and Anatomy, La Trobe University, Vic. 3086, Australia
c University of Newcastle, University Drive, Callaghan, NSW. 2308, Australiad North Sydney Central Coast Health Service, Gosford Hospital, Cnr Holden & Racecourse Rds, Gosford, NSW. 2250, Australia
Abstract
Objective To compare the analgesic efficacy of interferential therapy (IFT) and transcutaneous electrical nerve stimulation (TENS) using an
experimental cold pain model.
Design Randomised controlled trial with repeated measures design.
Setting University research laboratory.
Participants Twenty healthy subjects.
Interventions IFT and TENS applied to each subject on different days.
Main outcome measures Cold pain threshold (time), intensity and unpleasantness (visual analogue scales).
Results The mean cold pain threshold with a TENS intervention was higher than that with IFT. A training effect was evident as subjects
responses become more consistent with repeated exposure to stimulation and the testing procedure. Using data from the second testing
sessions, the differences in pain threshold between IFT and TENS for the two during-intervention (T3 and T4) measures were statistically
significant (T3 difference in the means 5.9 seconds, 99% confidence interval 3.1 to 8.7 seconds; T4 difference in the means 6.6 seconds, 99%
confidence interval 3.8 to 9.4 seconds). No significant differences were identified in pain intensity and unpleasantness ratings.
Conclusions TENS is more effective than IFT at increasing cold pain thresholds in healthy subjects, and this effect increases with repeated
exposures. Future trials should include a familiarisation session prior to testing to increase the consistency of subjects responses. The clinical
implications of these effects need investigation.
2006 Chartered Society of Physiotherapy. Published by Elsevier Ltd. All rights reserved.
Keywords: Transcutaneous nerve stimulation; Cold pain; Pain measures; Experimental pain; Electroanalgesia; Hypoalgesia
Introduction
Electrical stimulation is used regularly by physiother-
apists. Indications include pain management, muscle re-
education, strengthening and oedema reduction [1,2]. Ofthese, pain management is the most common use, and the
current types used are interferential therapy (IFT) and tran-
scutaneous electrical nerve stimulation (TENS) [3].
The physical properties of IFT and TENS are quite dif-
ferent. Conventional TENS is a type of low-frequency stim-
ulation, commonly applied at frequencies of 1200 Hz [1].
Corresponding author. Tel.: +61 3 9479 5814; fax: +61 3 9479 5784.
E-mail address: [email protected] (A.R. Ward).
The current is pulsed and the flow of charge is interrupted
rather than continuous. IFT is a type of medium-frequency
alternating current, with carrier frequencies of 110 kHz [4].
Unlike TENS, IFT current is sinusoidal and provides a con-
tinuous flow of charge. IFT is usually delivered in bursts witha frequency in the biological range (1150 Hz) [2]. Despite
the widespread use of IFT and TENS for pain management,
there is a lack of scientificevidence supporting aspects of their
claimed effectiveness. Consequently, in a clinical setting,
selection of the type and parameters of electrical stimulation
is largely based on theoretical or anecdotal evidence [2,4].
The aim of the present study was to compare the analgesic
effectiveness of IFT and TENS. A literature search was con-
ducted using the keywords interferential and/or TENS,
0031-9406/$ see front matter 2006 Chartered Society of Physiotherapy. Published by Elsevier Ltd. All rights reserved.
doi:10.1016/j.physio.2006.05.008
Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous
electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008
mailto:[email protected]://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008mailto:[email protected]8/3/2019 Shanaham Et Al - TENS x IFC
2/7
ARTICLE IN PRESSPHYST-222; No.of Pages 7
2 C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx
and pain or pain threshold (not dental, teeth, anaesthesia
or labour). The search was limited to studies of human sub-
jects, written in the English language, that reported the use
of ITF and/or TENS for treating musculoskeletal pain. The
databases searched were Embase, Medline, Cinahl, Sports
Discus, The Cochrane Library and PubMed (19662005).
Thirty-eight articles were identified.These studies used a range of pain-induction techniques:
cold [58], heat [9], mechanical [10] and ischaemic [11]. The
most consistent analgesic effects were observed with cold-
induced pain [6,7], with more varied results obtained using
other pain-induction techniques [1012].
Most studies examined the hypoalgesic effects of TENS
[7,1118], and a smaller number studied IFT [8,1922]. Only
four of the studies compared the effectiveness of TENS and
IFT directly [6,9,10,23]. Three of these reported that IFT
and TENS were significantly more effective at modulating
pain than sham interventions or control groups (the exception
being Alves-Guerreiro et al.[10]). However, no significant
differences in the analgesic effectiveness of IFT and TENSwere identified in any of the four studies. This was possibly
due to inadequate statistical power as the question addressed
was not do the interventions have an effect? but do the
effects differ? If both interventions had an effect, any differ-
ences could easily be obscured or more difficult to detect.
A common limitation across existing comparisons of
TENS and IFT is small sample sizes, ranging from 7 [6] to 18
[9] subjects per group. Unlessa within-subject design is used,
the risk of a type II error is high. For example, the Johnson
and Tabasam study [6] used seven subjects per group, and
each subject was tested with a single type of stimulation. The
observed difference (i.e. TENS more effective than IFT) wasstatistically insignificant (P = 0.09). However, a calculation
using 10 subjects and the same means and standard devia-
tions produces a P-value of less than 0.05, suggesting that
the group was too small given the effect size, resulting in a
type II statistical error.
The most effective and practical method to compare IFT
and TENS is to use a within-subject repeated measures
design, and to administer both to each subject. This design
eliminates between-subject variation and increases the statis-
tical power of the study. Existing comparative studies did not
use this approach.
The aim of the present study was to compare the analgesic
effectiveness of IFT and TENS using the same experimental
cold pain model as Johnson and Tabasam [6], but with larger
subject numbers and a within-subject design. A secondary
aim was to investigate whether familiarisation with electrical
stimulation and the testing procedure affected the consistency
of subjects responses.
The following hypotheses were tested:
TENS is more effective than IFT at increasing the cold
pain threshold of healthy subjects;
there is no difference in the effectiveness of IFTand TENS
at altering pain intensity and unpleasantness; and
familiarisation with stimulation and the testing procedure
affects the consistency of subjects responses (i.e. there is
a training effect).
Methods
Research design
A single blind randomised controlled trial with a repeated
measures design was used. The design was based on the
method used in previous research (e.g. [6,24]), but was modi-
fied to increase statistical power by using a larger sample size
and a training session identical to the testing session. Each
subject attended on two separate days and IFT and TENS
were given in a pre-determined order. Subjects were blinded
to the treatment modality being administered. The investi-
gator was not blinded as the investigators role was simply
to record the outcome measures as numerical values [mea-
sured time or reported visual analogue scale (VAS) rating],rather than make any subjective assessment or measurement;
thus, the risk of assessor bias being able to influence the
results was minimal. A sham group was not included as the
aim was specifically to compare the effectiveness of two
current types that had previously been shown to be more
effective than sham stimulation [57,9,24]. A single investi-
gator administered the electrical stimulation and recorded all
measurements.
The independent variable was the type of stimulation
administered (IFT or TENS). The dependent variables were
the time to pain threshold (seconds), and pain intensity and
unpleasantness as rated by subjects on two, 10-cm horizontalVASs.Painintensitywas operationally defined as howstrong
the pain in your hand feels, whilst pain unpleasantness was
how uncomfortable the pain in your hand feels.
Subjects
Twenty volunteers (10 males, 10 females) aged between
18 and 32 years (mean age 21.7 years, standard deviation 3.5
years) were recruited into the study. Potential participants
were excluded if their non-dominant arm was affected by
any orthopaedic, neurological or circulatory pathology, or
if the skin in the area under the electrodes was broken or
irritated.Participants were instructednot to takeany analgesic
or anti-inflammatory medication (or apply them as creams or
gels to their non-dominant forearm) in the 12 hours prior to
eachtesting session. No volunteers were excludedwhenthese
criteria were applied.
Apparatus
Electrical stimulation
A Chattanooga Intelect (Intelect Legend Stim, Chat-
tanooga Group Inc., Hixson, TN, USA) was used to deliver
the pre-modulated IFT current, and a Metron Multistimu-
Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous
electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008
http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.0088/3/2019 Shanaham Et Al - TENS x IFC
3/7
ARTICLE IN PRESSPHYST-222; No.of Pages 7
C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx 3
lator (Metron Medical Australia Pty. Ltd., Carrum Downs,
Victoria, Australia) was used for TENS. These stimulators,
similar in size and appearance, meant that testing conditions
were kept as uniform as possible for the two interventions.
The display screen and control panels were not visible to
subjects, and subjects were given no indication of which cur-
rent type was being delivered. Stimulation parameters werekept constant throughout the procedure, and only the current
intensitywasadjustedduring the session.Current was applied
via two Stimtrode 50 mm 90 mm rectangular self-adhesive
electrodes (Axelgaard Manufacturing Co. Ltd., Fallbrook,
CA, USA). TENS stimulation used 200-s biphasic pulses
(100-s phaseduration) anda pulse frequency of 100 Hz.Pre-
modulated IFT was applied with a carrier frequency of 5kHz
(100-s phase duration) and a beat frequency of 100 Hz.
Cold pain induction
Cold pain induction required two water baths of uniform
size andshape; onewas maintained at 37 C and the other was
maintained at 0 C with ice and water. A Heidolph heater stir-rer unit (AccurexEquipment,Brunswick, Victoria, Australia)
was used in each bath to ensure that the water temperatures
(both monitored with a thermometer) remained constant.
Procedure
Prior to the first session
Eligibility was confirmed and informed consent was
obtained. Each participant wasallocated at random to a group
by selecting a sealed envelope (marked for male or female)
containing a pre-arranged order of testing conditions. Within
the 10 envelopes designated to each sex, five specified thatIFT would be delivered in Session A and five specified that
TENS would be delivered in Session A. Subjects were not
informed of the group to which they were assigned. The sub-
ject was seated and their non-dominant forearm was cleaned
with an alcohol swab. The two electrodes were then posi-
tioned on the subjects non-dominant forearm, one anterior
and one posterior, with the centre point of the electrode lying
equidistant to the elbow and wrist. The electrode alignment
was standardised between subjects and between sessions.
Cold pain induction
The cold-induced pain cycle is based on a method used
in previous research [68,24]. Six cold pain cycles, each of
10-minutes duration, were conducted sequentially for a 1-
hour test session. The procedure is depicted in Fig. 1. Each
cycle commenced as the subject placed their hand (up to the
distal wrist crease) in a warm water bath. After 5 minutes,
the subject transferred their hand to a cold water bath. The
subject was instructed to focus on the sensations in their
immersed hand until they felt the onset of deep, dull, aching
pain and to then say pain. The time from immersion in cold
water to this point was recorded as the pain threshold (in sec-
onds). The subject was required to leave their hand in the
cold water for a further 30 seconds. Immediately upon with-
Fig. 1. Procedure for (a) one cold pain cycle and (b) one testing session
consisting of six cold pain cycles, based on Johnson and Tabasam [6].
drawal from the cold water, the subject completed a VAS
rating of pain intensity and pain unpleasantness (where zero
represented not intense/unpleasant at all and 10 represented
most intense/unpleasant pain imaginable). The subject then
rested until 10 minutes had elapsed since the beginning of the
cycle, at which time the next cycle was commenced with re-
immersion of the hand in warm water.
Electrical stimulation
At the start of Cycle 3 (i.e. 20 minutes into the session), the
allocated electrical stimulator was turned on. The investiga-tor increased the current intensity, as directed by the subject,
until they experienced a strong but comfortable level of
stimulation. The current intensity was adjusted upon request
to maintain this level, except when the subjects hand was
in the cold water bath. Stimulation continued uninterrupted
for 20 minutes, which constituted two complete pain cycles
(Cycles 3 and 4). Two further pain cycles (Cycles 5 and 6)
were completed once the stimulation was turned off. The pro-
cedure was identical for TENS and IFT delivery. Thus, two
pre-intervention measures, two during-intervention measures
and two post-intervention measures were obtained.
Session B
Session B was conducted no earlier than 12 hours after
Session A to minimise the risk of a carry-over effect. Subject
eligibility was reconfirmed, and the cold-pain-induction pro-
cedure was repeated using the alternative type of stimulation.
All 20 subjects participated in both sessions. After comple-
tion of Session B, subjects were asked in which, if either, of
the two sessions they felt stimulation was more effective and
also more comfortable. Subjects were also given an oppor-
tunity to make any other comments about the stimulation or
testing procedure.
Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous
electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008
http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.0088/3/2019 Shanaham Et Al - TENS x IFC
4/7
ARTICLE IN PRESSPHYST-222; No.of Pages 7
4 C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx
Data analysis
Data were analysed using SPSS Version 11. For data anal-
ysis purposes, the group who received IFT in Session A and
TENS in Session B were referred to as Group 1, and those
who received TENS in Session A and IFT in Session B were
referred to as Group 2.Results were analysed in two stages. First, descriptive
statistics were obtained for each group (1 or 2; TENS first
or IFT first) and each condition (TENS or IFT), and tests of
normality were made. Data from the first session for each
group were much more scattered than data from the second
session, with standard deviations being approximately dou-
ble in the first session. Consequently, the first session was
considered to be the training/familiarisation session and only
the data from the second session were used for subsequent
analysis.
Data from the second session were analysed using a
two-way repeated measures analysis of variance (ANOVA)
(cycle stimulus type). Where justified by the ANOVA, posthoc tests were performed.
Previous research has shown that both TENS and IFT
elevate the cold pain threshold [6,9,23]. The present sta-
tistical analysis was therefore restricted to testing whether
the two current types had significantly different effects.
Pre-planned comparisons were made to reduce the Bon-
ferroni correction needed when multiple comparisons are
performed and to reduce the risk of a type II error. For sin-
gle comparisons, a 95% confidence interval can be used,
but the confidence interval must be set higher for multiple
comparisons.
Pain threshold data
Pain thresholddata were analysed using changefrom base-
line (seconds)for each intervention. The two pre-intervention
measurements from Cycles 1 and 2 (T1 and T2) were aver-
aged to give a baseline for each session, which was then
subtracted from the pain threshold measurements taken dur-
ing Cycles 3 (T3), 4 (T4), 5 (T5) and 6 (T6). Resulting data
were subjected to preliminary analysis by calculating means
and standard deviations, and performing tests of normality.
An ANOVA was conducted using the data from the sec-
ond session, which clearly had greater consistency due to
subjects familiarisation and practice. Finally, comparisons
of the change in pain threshold at T3 and T4 were con-
ducted; the change from baseline data for IFT was compared
with that for TENS at these points. The rationale for this
was that Johnson and Tabasam [6] first noticed a differ-
ence between IFT and TENS at T3, which reached a peak
at T4 before returning to baseline. The choice of only two
statistical comparisons meant that with a Bonferroni cor-
rection, a 97.5% confidence interval ( = 0.025) could be
used to assess the significance of any TENS versus IFT
differences.
Pain intensity and unpleasantness data
Pain intensity and unpleasantness ratings were compared
in terms of change from baseline in the same manner as
the pain threshold data. Similarly, ANOVAs were conducted
using data from Session B alone. Post hoc testing was not
justified by the ANOVAs.
Subjects comments
The frequencies of subjects comments regarding the
effectiveness and comfort of the interventions were calcu-
lated and compared.
Results
One of the 20 subjects failed to reach a pain threshold
within the allocated 5-minute time period in 11 of the 12
measurement cycles. As such, this setof results was excluded,
leaving 19 data sets for analysis.
Pain thresholds
Fig. 2 shows the mean change in pain threshold with
electrical stimulation intervention (Cycles 3 and 4) and fol-
lowing intervention (Cycles 5 and 6) for the second set of
measurements (Session B, post-familiarisation). Thresholds
increasedwhen stimulation was applied, and the increase was
greater with TENS than with IFT. The difference between the
two current types was greatest at T4, near the end of stimu-
lation; the increase in pain threshold for TENS was nearlytwice that for IFT. A two-way ANOVA (cycle stimulus
type) using data from Session B found that between-cycle
differences were significant [F(3,1) = 3.88, P = 0.033] and
between-stimulus type effects were, on average, insignifi-
Fig. 2. Mean change in pain threshold (seconds) using data from the second
(post-familiarisation)testingsession.Error barsare standarddeviations. IFT,
interferential therapy; TENS, transcutaneous electrical nerve stimulation.
Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous
electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008
http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.0088/3/2019 Shanaham Et Al - TENS x IFC
5/7
ARTICLE IN PRESSPHYST-222; No.of Pages 7
C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx 5
cant [F(1,3) = 2.13, P = 0.164], but a significant interaction
effect (cycle stimulus type) was observed [F(1,3) = 4.36,
P = 0.023]. The significant interaction effect indicates the
possibility that there is a between-stimulus type effect, which
is only evident at particular cycles. Inspection of Fig. 2
suggests that there may be a between-stimulus type effect
at T3 and T4 but not at T5 and T6. At T3, the meanTENS threshold change was 6.25 seconds, compared with
0.35 seconds for IFT. At T4,when thedifference was greatest,
the mean TENS threshold change was 9.75 seconds, com-
pared with 3.15 seconds for IFT. The differences between
pain thresholds at T3 and T4 were significant at the
P = 0.01 level (T3 difference in the means = 5.9 seconds,
99% confidence interval 3.1 to 8.7 seconds; T4 difference
in the means= 6.6 seconds, 99% confidence interval 3.8
to 9.4 seconds). Ninety-nine percent confidence intervals
were used to meet the need for Bonferronis correction
and to further minimise the risk of a type I statistical
error.
Pain intensity
As with the threshold data, the pain intensity results
showed a much greater scatter in Session A than in Session B.
Standard deviations for Session A were typically 30% greater
than those for Session B. Fig. 3 compares Session B results
for pain intensity. Both IFT and TENS decreased the pain
intensity ratings but to a similar extent. There is the sugges-
tion that TENS is more effective but that the difference is not
significant.
A two-way ANOVA (cycle stimulus type) was con-
ducted using data from Session B. This indicated thatbetween-cycle differences were significant [F(1,3) = 3.64,
P = 0.04], but between-stimulus type effects were not
[F(1,3) = 0.195, P = 0.67] and there was no significant inter-
action effect [F(1,3) = 0.32, P = 0.81].
Fig.3. Mean change in painintensity [visual analoguescale (VAS)]showing
averageddata for the second (post-familiarisation)testing session.Error bars
are standard deviations. IFT, interferential therapy; TENS, transcutaneous
electrical nerve stimulation.
Fig. 4. Mean change in pain unpleasantness [visual analogue scale (VAS)]
showing averaged data for the second (post-familiarisation) testing session.
Error bars are standard deviations. IFT, interferential therapy; TENS, tran-
scutaneous electrical nerve stimulation.
Pain unpleasantness
Fig. 4 shows the results for pain unpleasantness, compar-
ing results from Session B. Both TENS and IFTdecreased the
pain unpleasantnessratings but to a similar extent. A two-way
ANOVA (cycle stimulus type) was conducted using results
from Session B. Between-cycle differences were insignifi-
cant [F(1,3) = 0.66, P = 0.59], as was the between-stimulus
type effect [F(1,3) = 0.014, P = 0.91]. There was no signifi-
cant interaction effect [F(1,3) = 0.39, P = 0.76].
Subjects comments
Comfort of stimulation
Twelve subjects (63%) reported that IFT was more com-
fortable than TENS, and four subjects (21%) reported that
TENS was more comfortable than IFT. Three subjects (16%)
reported no difference.
Effectiveness of stimulation
The effectiveness ratings were exactly the same as the
comfort ratings, i.e. 12 subjects (63%) reported that IFT was
more effective than TENS at altering their response to the
cold pain. Only four of the 19 subjects (21%) reported that
TENS was more effective than IFT. The remaining three sub-
jects reported no difference in perceived effectiveness. Thus,
a clear majority thought that IFT was more effective and also
that it was more comfortable than TENS.
Discussion
Training effect
Training had a major impact on the consistency of sub-
jects responses, as demonstrated by the decrease in standard
deviations across all outcome measures from Session A to
Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous
electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008
http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.0088/3/2019 Shanaham Et Al - TENS x IFC
6/7
ARTICLE IN PRESSPHYST-222; No.of Pages 7
6 C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx
Table 1
Pain threshold (time to pain onset) changes during (T3 and T4) and after (T5 and T6) interventions [interferential therapy (IFT) or transcutaneous electrical
nerve stimulation (TENS)]
IFT threshold changes (seconds)
T3 T4 T5 T6
Mean (S.D.), Group 1 (IFT first) 3.8 (9.6) 1.5 (4.7) 0.0 (6.8) 1.3 (6.1)
Mean (S.D.), Group 2 (IFT second) 0.4 (3.6) 3.2 (3.7) 3.6 (5.0) 4.1 (4.0)
TENS threshold changes (seconds)
T3 T4 T5 T6
Mean (S.D.), Group 2 (TENS first) 0.3 (9.4) 2.4 (9.9) 7.0 (8.6) 1.3 (12.2)
Mean (S.D.), Group 1 (TENS second) 6.3 (7.9) 9.8 (7.7) 2.6 (4.2) 3.8 (3.9)
Standard deviations (S.D.) were much reduced in the second experimental session.
Session B (Table 1). The effect of repeated exposures to the
test conditions was not reported in existing comparisons of
IFT and TENS, and may have contributed to the absence of
significant findings [5,6,9]. The present study highlights the
importance of familiarising subjects prior to testing so thatresponses are more consistent. Future evaluations should,
ideally, use three sessions per participant; one for training
or familiarisation with either type of stimulator and two for
testing.
The presence of a training effect also suggests that, in
a clinical context, the desired results of electrical stimula-
tion for pain management may not be achieved in an initial
application. The efficacy of stimulation may improve with
subsequent exposures, so treatment may be more effective
when used repeatedly. Clinically, this suggests that at least
two trials of electrical stimulation are needed to determine
the effectiveness of pain management reliably. This is con-sistent with a clinical finding that the effectiveness of IFT
for treatment of arthritic knee pain increased over an 8-week
treatment period [21].
Pain thresholds
The present study identified that TENS was significantly
more effective than IFT at increasing the experimental cold
pain threshold of healthy subjects, and that the difference
between the two current types was greatest during a 20-
minute period of stimulation (Fig. 2). AtT3 and T4, the mean
pain threshold increases were greater with TENS than IFT.
The difference was significant at the 99% confidence level.
Contrary to these findings, IFT is used clinically in prefer-
ence to TENS in the majority of physiotherapy departments
[3]. The present findings, however, suggest that IFT is less
effective butperhaps more comfortableand perceived as more
effective than TENS, as reported by Johnson and Tabasam
[23]. The choice of IFT or TENS in a clinical context depends
on the relative importance of effectiveness and patient com-
fort/acceptance. Unless patient acceptance or tolerance is a
major issue, the findings suggest that TENS should be the
clinical modality of choice and IFT should be used as a fall-
back alternative.
The properties of clinical pain, however, may differ from
those of experimental pain in healthy subjects. Healthy sub-
jects know that experimentally induced pain is strictly tempo-
rary, so the psychological impact may be different. Similarly,
the neurophysiological aspects of the pain, i.e. the originand pattern of neural activity, are also different. This means
that future trials using a clinical population are necessary to
progress this matter. If TENS is also shown to be more effec-
tive in clinical trials, it should be adopted in preference to
IFT unless comfort and acceptance are an issue.
Pain intensity and unpleasantness
The results for painintensityand unpleasantnesswere very
similar. Although both IFT and TENS apparently decreased
pain ratings to lower than baseline levels, the difference
between the effects of the two types of stimulation was notstatistically significant. These results are consistent with find-
ings from previous research usingthe samevariables [6,9,24].
This calls into question whether pain intensity and unpleas-
antness ratings are useful tools for measuring the effect of
electrical stimulation, as these measures do not show change
and they require an extension of the period of painful stimu-
lation after onset.
Subjects comments
The majority of subjects reported that IFT was their pre-
ferred type of stimulation, for both comfort and perceived
effectiveness, despite results indicating that TENS is actually
more effective at increasing the pain threshold. The subjects
preference for IFT warrants further investigation.
Conclusion
The effectiveness of TENS at increasing pain threshold
was greater than that for IFT. This difference in effectiveness
of the two current types was statistically significant when
the training effect was taken into consideration. The present
study established that repeated exposures to stimulation and
Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous
electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008
http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.0088/3/2019 Shanaham Et Al - TENS x IFC
7/7
ARTICLE IN PRESSPHYST-222; No.of Pages 7
C. Shanahan et al. / Physiotherapy xxx (2006) xxxxxx 7
cold-induced pain have a significant effect on the consistency
of subjects responses. Consequently, it is recommended that
future studies should include familiarisation sessions prior
to testing to allow for a training effect. The training effect
may also have clinical implications, and multiple uses may
be needed prior to evaluating effectiveness in a patient.
Neither TENS nor IFT was more effective at altering painratings of intensity and unpleasantness. This result is con-
sistent with the existing literature and raises questions about
whether the continued useof these outcome measures is justi-
fied in laboratory-based studies, particularly considering the
increased discomfort that they impose on participants.
The pain threshold results in the present study support the
use of TENS in preference to IFT for the treatment of pain.
This is contrary to current clinical practice,whichfavours IFT
despite TENS machines being considerably cheaper, more
portable, andmore practical for home use. Thepresent results
challenge existing beliefs and clinical practices, and indicate
the need for clinical trials.
Ethical approval: Faculty Human Ethics Committee, Faculty
of Health Sciences, La Trobe University, Victoria, Australia.
Funding: The Schools of Physiotherapy and Human Bio-
sciences, La Trobe University, Victoria, Australia.
Conflicts of interest: None.
References
[1] Selkowitz D. Electrical currents. In: Cameron M, editor. Physical
agents in rehabilitation. Philadelphia: WB Saunders Co.; 1999. p. 345
427.
[2] Robertson VJ, Ward AR, Low J, Reed A. Electrotherapy explained:
principles and practice. 4th ed. Oxford: Butterworth-Heinemann; 2006.
[3] Robertson VJ, Spurritt D. Electrophysical agents: implications of their
availability and use in undergraduate clinical placements. Physiother-
apy 1998;84:33544.
[4] Kitchen S. Electrotherapy evidence-based practice. 11th ed. London:
Churchill Livingstone; 2002.
[5] Johnson MI, Tabasam G. A single-blind investigation into the hypoal-
gesic effects of different swing patterns of interferential currents on
cold-induced pain in healthy volunteers. Arch Phys Med Rehabil
2003;84:3507.
[6] Johnson MI, Tabasam G. A double blind placebo controlled inves-
tigation into the analgesic effects of inferential currents (IFC) and
transcutaneous electrical nerve stimulation (TENS) on cold-induced
pain in healthy subjects. Physiother Theor Pract 1999;15:21733.[7] Johnson MI, Ashton CH, Bousfield DR, Thompson JW. Analgesic
effects of different frequencies of transcutaneous electrical nerve
stimulation on cold-induced pain in normal subjects. Pain 1989;39:
2316.
[8] Johnson MI, Wilson H. The analgesic effects of different swing pat-
terns of interferential currents on cold-induced pain. Physiotherapy
1997;83:4617.
[9] Cheing GLY, Hui-Chan CWY. Analgesic effects of transcutaneous
electrical nerve stimulation and interferential currents on heat pain in
healthy subjects. J Rehabil Med 2003;35:159.
[10] Alves-Guerreiro J, Noble JG, Lowe AS, Walsh DM. The effect of three
electrotherapeutic modalities upon peripheral nerve conduction and
mechanical pain threshold. Clin Physiol 2001;21:70411.
[11] WalshDM, Liggett C, Baxter D, AllenJM. A double-blindinvestigation
of the hypoalgesic effects of transcutaneous electricalnerve stimulation
upon experimentally induced ischaemic pain. Pain 1995;61:3945.
[12] Simmonds M, Wessel J, Scudds R. The effect of pain quality on the
efficacy of conventional TENS. Physiother Can 1992;44:3540.
[13] Barr JO, Nielsen DH, Soderberg GL. Transcutaneous electrical nerve
stimulation characteristics for altering pain perception. Phys Ther
1986;66:151521.
[14] Cheing GL, Hui-Chan CW. Transcutaneous electrical nerve stimula-
tion: nonparallel antinociceptive effects on chronic clinical pain and
acute experimental pain. Arch Phys Med Rehabil 1999;80:30512.
[15] Chesterton LS, Barlas P, Foster NE, Lundeberg T, Wright CC, Bax-
ter GD. Sensory stimulation (TENS): effects of parameter manipula-
tion on mechanical pain thresholds in healthy human subjects. Pain
2002;99:25362.
[16] Foster NE, Baxter F, Walsh DM, Baxter GD, Allen JM. Manipula-
tion of transcutaneous electrical nerve stimulation variables has no
effect on two models of experimental pain in humans. Clin J Pain
1996;12:30110.
[17] Roche PA, Gijsbers K, Belch JJ, Forbes CD. Modification of induced
ischaemic pain by transcutaneous electrical nerve stimulation. Pain
1984;20:4552.
[18] Walsh DM, Lowe AS, McCormack K, Willer J, Baxter GD, Allen JM.
Transcutaneous electrical nerve stimulation: effect on peripheral nerve
conduction, mechanical painthreshold, andtactile thresholdin humans.
Arch Phys Med Rehabil 1998;79:10518.
[19] Tabasam G, Johnson MI. Electrotherapy for pain relief: does it work?
A laboratory-based study to examine the analgesic effects of elec-
trotherapyon cold-inducedpain in healthy individuals.Clin Effect Nurs1999;3:1424.
[20] Johnson MI, Tabasam G. A single-blind placebo-controlled investiga-
tion into the analgesic effects of interferential currents on experimen-
tally induced ischaemic pain in healthy subjects. Clin Physiol Funct
Imaging 2002;22:18796.
[21] Adedoyin RA, Olaogun MOB, Fagbeja OO. Effect of interferential
current stimulation in management of osteo-arthritic knee pain. Phys-
iotherapy 2002;88:4939.
[22] Stephenson R, Walker EM. The analgesic effects of interferential (IF)
current on cold-pressor pain in healthy subjects: a single blind trial of
three IF currents against sham IF and control. Physiother Theor Pract
2003;19:99107.
[23] Johnson MI, Tabasam GI. An investigation into the analgesic effects
of interferential currents and transcutaneous electric nerve stimulation
on experimentally induced ischaemic pain in otherwise pain-free vol-unteers. Phys Ther 2003;83:20823.
[24] McManusFJ, WardAR, RobertsonVJ. Theanalgesic effects of interfer-
ential therapy on two experimental painmodels: coldand mechanically
induced pain. Physiotherapy 2006;92:95102.
Please cite this article as: Christine Shanahan et al., Comparison of the analgesic efficacy of interferential therapy and transcutaneous
electrical nerve stimulation, Physiotherapy (2006), doi:10.1016/j.physio.2006.05.008
http://dx.doi.org/10.1016/j.physio.2006.05.008http://dx.doi.org/10.1016/j.physio.2006.05.008