1
Testing Thermistors as Liquid Level Sensors
Kristy Lubinski
2
Circuit Diagram #1
V
5 V
PT 1000 PT 1000
750 Ω
3
Circuit Diagram #2
V
5 V
PT 1000 PT 1000
750 Ω
4
Set-Up
PT 1000’s 1 mm space
5
Boil Off Test
• Place PT 1000’s just into liquid• Collect data while the liquid boils off• Boil off rate measured separately– Liquid Nitrogen• 0.2 mm/min
– Liquid Argon• 0.2 mm/min
6
Circuit #1 – Liquid Nitrogen
7
Circuit #2 – Liquid Nitrogen
8
Circuit #2 – Liquid Argon Trial 1
Both Out
Both In
One In
9
Circuit #2 – Liquid Argon Trial 2
Both Out
Both In
One In
10
Time Resolution Test
• Quickly place both PT 1000’s in liquid at 1:30• Quickly pull both PT 1000’s out of liquid at
2:30 • Measure response time
11
Circuit #1 – Liquid Nitrogen
Both Out
Both In
Both Out
12
Circuit #2 – Liquid Nitrogen
Both Out
Both In
Both Out
13
Circuit #2 – Liquid Argon
Both Out
Both In
Both Out
14
Two Jump Test
• Lower PT 1000’s into liquid• Using only LabView Interface, stop apparatus
when 1 PT 1000 is in liquid• Repeat lifting out of liquid• Verify location of PT 1000 visually
15
Circuit #1 – Liquid Nitrogen
Both Out
Both Out
One In
One In
Both In
16
Circuit #2 – Liquid Nitrogen
Both Out
Both Out
One InOne In
Both In
17
Circuit #2 Liquid Argon
Both Out
Both Out
One InOne In
Both In
18
LN vs LArLiquid Boiling Point
(K)Density at B.P. (g/mL)
Surface Tension at B.P. (mN/m)
Nitrogen 77 0.807 8.94
Argon 87.3 1.40 14.19
19
Level Test
• Three sets of PT 1000’s on outside of plexiglass cylinder ring
• Level using three strings that suspend the ring above liquid (like a puppet)
PT 1000’s
Blue string used to hold ring up
20
21
Boil Off Test with Ring (not leveled)
22
Attempt to Level Resistors and then Boil off LN2
1. Trying to level ring by pulling on individual strings until one resistor was in the liquid and one was out of the liquid ~3.7 V
2. All three resistors around same level
3. Wiggled all three cords to see if they responded
4. Re-leveled all three and let sit to boil off liquid
5. Boil off points were different for all 3 sets, design set-up was not ideal, resistors were not all exactly 1 mm apart, not exactly on same location around the ring