Upload
imad-aghila
View
219
Download
0
Embed Size (px)
Citation preview
7/28/2019 AAS 200504_030_031
1/2
author
articleHeike Gleisner, Alf Liebmann
Analytik Jena AGKonrad-Zuse-Strasse 1
07745 Jena, Germany
Tel.: +49/3641/7770
Fax: +49/3641/779279
Thomas Furche
H.C. Starck Hermsdorf GmbHRobert-Friese-Strasse 4
07629 Hermsdorf
Germany
Tel.: +49/36601/922101
Fax: +49/36601/922111
Measurement of Tungsten in Molybdenum using High-Resolution Continuum Source
Atomic Absorption Spectrometry (HR-CS AAS)
Introduction based on classical line source AAS (LS AAS) however has its Results and discussion
30Analytical Instrumentation
Looking back over the last few years at the development of AAS, there appear to have been no outstanding technical and analytical innovationsin this field. Changes have primarily involved accessories and software and serve to enhance automation and improve device handling. Firstly,
the optimisation of software improves user-friendliness, and secondly, the steady growth in quality control and data security standards are
addressed. Nevertheless, by virtue of its simple operability, rapid readiness for measurement, low operating costs and high degree of interference
immunity, AAS is still well established in inorganic analytics today. The commercial introduction of HR-CS AAS technology marks a developmental
leap opening up a whole new generation of atomic spectroscopy devices to the user.
7/28/2019 AAS 200504_030_031
2/2
31Analytical Instrumentation
Bibliography:
[1] B. Welz, H. Becker-Ross, S. Florek, U. Heitmann,
High-Resolution Continuum Source AAS The
better Way to do Atomic Absorption Spectrometry,
ISBN 3-527-30736-2, Wiley-VCH, Weinheim, 2005
[2] H. Gleisner, K. Eichardt, G. Schlemmer, U. Heitmann:
Die AAS wird neu definiert, LABO 4/ 2004, S.64-67
[3] U. Heitmann, H. Becker-Ro: Atomabsorptions-
Spektrometrie mit einem Kontinuumstrahler
(CS AAS), GIT 7/2001, S.728-731
[4] H. Gleisner: Applikationsberichte Analytik Jena,
CSAA_FL_01_04_d | 11 / 2004
Summary
The W content values measured for molybdenum samples lie in the anticipated range and show good correspondence with the
internal reference values. The performance of the method is simplified, because, as a result of the possibility of measuring the
spectral vicinity of the analysis line simultaneously, significantly more information is available on the sample under investigation.
Fluctuations in the radiant intensity of the lamp, in the detector sensitivity and in the permeability of the flame and therefore all
continuous background absorption are measured simultaneously and are automatically corrected at selected reference pixels.
Through the use of a high-resolution double monochromator, discontinuous spectral disturbances, e.g. through absorption lines
from the Mo matrix, are not identified by the pixels used for analysis and are thereby eliminated. There is therefore no necessity for
background correction as exists for LS AAS. The detection limit in HR-CS AAS is fundamentally improved, as no second lamp noise
source is present in the optical system. The use of an extremely low noise CCD array detector in the contrAA 300 is also superior
to the photomultipliers common in LS AAS and the use of a high-energy Xe short arc lamp with very high radiation intensity further
improves the signal-to-noise ratio significantly. The detection limit for tungsten was consequently improved by a factor of 5.
Finally, it can be seen that the measurement of tungsten in molybdenum using the contrAA 300 with HR-CS AAS can be
performed easily, correctly and without laborious sample preparation. The minimisation of investment and operating costs produces
in a significant increase in laboratory effectiveness and flexibility together with a noticeably higher sample throughput.
only the main resonance line is taken for analytical evaluation in
this example. The second absorption line can be used if the
concentration is exceeded in order to expand the dynamic working
range for calibration. As a comparison, an energy scan over 1.0 nm
around the analysis wavelength of the W-HCL used was performedin the classical LS AAS (Fig. 3). The two W lines can hardly be
distinguished given the significantly lower resolution of LS AAS. It
was also apparent that not every HCL can be deployed, as not all
lamp manufacturers use UV transparent quartz windows for their
W-HCLs., in which case one has to turn to the longer wavelength
and somewhat less sensitive W absorption line at 294.4 nm. In the
Mo sample spectrum (Fig. 4) there are three additional W
absorption lines in the spectral vicinity of the 255.135 nm analysis
line, which could be assigned to the molybdenum matrix. The
255.086 nm Mo line lies directly in the spectral transmission band
of the slit in the LS AAS instrument with deuterium background
correction and consequently further attenuates the D2 broadband
source. This attenuation distorts the background measurement onthe analysis line leading to over-correction of the background and
hence to a lower analysis result (Table 2). For this reason, the Mo
Fig. 4: Absorption spectrum of the W sample Mo-HZ, spectral
observation width 0.43 nm (W main resonance line 255.135 nm, W
secondary line 255,039 nm, Mo lines: 255.017 nm; 255.086 nm;
255.287 nm)
matrix must be separated using special sample preparation prior to
measurement of W with LS AAS. The high-resolution double
monochromator in the contrAA?300 separates the Mo absorption
lines significantly from the W absorption line. As HR-CS AAS
always identifies the background and the analysis line
simultaneously and measures selected correction pixels, the
measurement of tungsten using this method is not affected by the
molybdenum absorption line.
Table 2: Results of W measurement with HR-CS AAS and LS AAS