Electrochemical Analysis LaboratoryDepartment of Chemical Metrology and Biometry
Calibration and Maintenance of Conductivity Meter
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
2
Outline
What is Conductivity?
How is Conductivity Measured?
What is Conductive solution?
Conductivity Meter
Traceability
Standard method
Rules for reliable measurement
Calibration
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
3What is Conductivity?Conductivity is the ability of a solution to pass an
electric current. The current is carried by cations andanions.
RG
1
The electrical conductance(G, Siemens) is thereciprocal of the resistance(R, Ohm).
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
4What is Conductivity?
A
lK
K is the cell constant (m-1)
l
A
l
AG
1
A is the electrode area (m2)l is the electrode gap (m) is the specific conductivity (S/m) is the specific resistance (m)
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
5What is Conductivity?
Solution conducts electricity depends on
Concentration
Mobility of ions
Valence of ions
Temperature
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
6How is Conductivity Measured?
Conductivity may be measured by applying an
alternating electrical current (I) to two electrodes
immersed in a solution and measuring the resulting
voltage (V).
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
7What is Conductive solution?
Conductivity is measured in aqueous solutions of
electrolytes.
Electrolytes are substances containing ions or
compounds that ionise in solution.
Strong electrolytes: fully ionised in solution
Weak electrolytes: not fully ionised in solution
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
8What is Conductive solution?
nCC
is molar conductance (S cm2 mol-1) is conductivity (S/m)C is concentrationn is charge of ion
is equivalent conductivity of ion
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
9What is Conductive solution?Equivalent conductivity of various ions at 25 oC
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
10
Conductivity value of solution
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
11
What is Conductive solution?
At the low electrolytic conductivity, the
relationship between conductivity and concentration is
linear.
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
12
How is Conductivity Measured?
The cell constant (K) cannot be determined by
measuring the electrode gap and area.
The cell constant is determined using a calibration
solution with known conductivity (usually 0.01 mol/l
KCl solution)
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
13
Conductivity Meter
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
14
Conductivity Meter
The conductivity
meter uses the conductance
and cell constant to display
the conductivity
Conductivity = cell constant x conductance
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
15
Conductivity Cell
2 pole cell
3 pole cell
4 pole cell
Platinised cell
Flow through cell
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
16
Conductivity Cell
2 pole cell : universal used, interfere with
polarization
3 pole cell: better reproducibility
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
17
Conductivity Cell
4 pole cell: should measure with outer tube
to minimize volume effect
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
18
Conductivity Cell
Platinised cell: coated electrode
with platinum black to increase
surface area and decrease
polarization effect.
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
19
Conductivity Cell
Flow through cell: designed for flow
system or small volume sample,
suitable for measurement of
conductivity of pure water (avoid
carbon dioxide)
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
20
Storage
Before storing the cell, rinse with deionised water
Short-term : deionised water
Long-term : deionised water or store dry
After long-term storage, condition the cell for 8 hours in
deionised water before use.
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
21
Selection of Conductivity Meter
Factor affecting measurement accuracy
Formats and units of reports results
Suitability for the measuring environment
features assisting ease of use
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
22
Selection of Conductivity Cell
Advantages Disadvantages
- Easy to maintain - Field effect cell must be positioned in the middle of measuring vessel
- Economic - Polarization in high conductivity sample
- Recommended for viscous sample with suspension
2 pole cell
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
23
Selection of Conductivity Cell
Advantages Disadvantages
- Linear over a very large conductivity range
- Unsuitable for micro samples
- Calibration and measurement in different range
- Ideal for high conductivity measurement
4 pole cell
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
24
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
25
Conductivity Measurement
The accuracy of conductivity measurements can be influenced by
Polarization
Geometry
Temperature
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
26
Conductivity Measurement
Polarization
Polarization occurs on the electrode surface, which leads to errors on result.
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
27
Conductivity Measurement
Preventing polarization
increasing the electrode area: coating platinum black on platinum electrode
Selecting 4 pole conductivity cell
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
28
Conductivity Measurement
Geometry related errors field effect
Field linesThese field lines can effect the
measurement if something
interferes with the field lines
such as the sample vessel
wall.
This effect can be minimized
by using 4-pole conductivity
cell
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
29
Conductivity Measurement
Temperature
Conductivity measurements are the temperature
dependent, varying from 1.5 to 5.5 %/oC
Temperature increasing - conductivity increasing
Conductivity meter measures the actual
conductivity and temperature and then converts
it to reference temperature using temperature
correction function
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
30
Conductivity Measurement
Temperature Temperature correction function
Linear function
Non-linear function
User-Definable temperature compensation
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
31
Temperature correction function
Linear function
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
32
Determination of the temperature coefficient
The temperature coefficients of some electrolytes
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
33
Temperature correction function
Non-linear function
Non-linear
compensation data for
variety of solutions is
widely available such
as ISO 7888-1985
(natural waters)
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
34
Traceability = G K
Primary cell(defined dimension)
Primary reference solution
Secondary cell
Secondary reference solution
Electrolytic conductivity meter
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
35
Standard method for Conductivity Measurement of Water
ASTM D1195 95 (Reapproved 1999) Standard
Test Methods for Electrical Conductivity and
Resistivity of Water
ISO 7888 1985 Water quality Determination of
electrolytic conductivity
OIML R 68 Calibration method for conductivity
cells
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
36
Rules for reliable measurements
Calibrate frequently: cell constant is an important
factor for conductivity measurement. Calibration
interval for conductivity cell will depend on the
application, sample and operating condition.
Temperature and stirring conditions: For accurate
measurement, recommended to use thermostat bath
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
37
Rules for reliable measurements
Position of conductivity cell: All the pole of
conductivity cell are covered by the sample and stand
at the center of sample vessel.
Low conductivity measurement:
use a flow through cell, to avoid CO2 from air
use non-platinised and low cell constant (< 1 cm-1)
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
38
Rules for reliable measurements
High conductivity measurement:
use platinised cells to prevent polarization error
use high cell constant (> 1 cm-1)
do not dilute sample (conductivity is not
proportional to concentration at high levels)
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
39
General Condition
The conductivity of an electrolytic solution may be determined from the measured resistance of the cell, when it contains the solution, and the constant K of the cell
The value of the cell constant K depends onthe construction of the cellthe temperature and concentration of the
solutionthe measuring frequencyetc.
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
40
Preparation for Calibration
check the rigidity of the fixing of the electrode and
the state of their surface
the cell shall be degreased with the appropriate
solvent according to the materials of which the cell
and electrode are made and then wash with distilled
water
the standard solution shall be prepared in quantity
sufficient for calibration
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
41
Calibration
the cell shall be washed with standard solution at
least 5 times
the solution of the last washing shall be left in the
cell for 10 minutes
the cell containing the solution shall be brought in
thermostat bath at 25 0C and maintain at thermal
equilibrium
Create PDF files without this message by purchasing novaPDF printer (http://www.novapdf.com)
42
Calibration
the cell shall be washed with standard solution at
least 5 times
the solution of the last washing shall be left in
the cell for 10 minutes
the cell containing the solution shall be brought
in thermostat bath at 25 0C and maintain at thermal
equilibrium