EC-flouride

Embed Size (px)

Citation preview

  • 8/3/2019 EC-flouride

    1/5

    PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740

    ENVIROENERGY2009

    Abstract Electrocoagulation (EC) experiments were carried outin batch and continuous modes for removal of excess fluoride from

    ground water. During batch studies using 2, 4 ,6 and 8 electrodes, an

    optimal operating condition of 6V using 4 plates and 20 min

    electrolyze time (ET) was found appropriate removing >88% of

    fluoride from ground water. Repeated continuous EC runs at optimal

    conditions, showed stable performance of the electrochemical reactor

    (ECR) removing fluoride to the required drinking water quality

    standards between 1.0 and 1.5mgL-1. Experiments were also carried

    out by spiking ground water with excess fluoride from 10-20mgL -1

    which showed > 90% fluoride removal. The pH of the water before

    ECC was 7.2-7.4 and 8.3-8.6 depending on applied cell voltage.

    Fluoride removal during continuous studies for flow rates 1, 2 and 5Lph were 86, 82, and 78% respectively from its initial value of 5.4

    mgL-1.

    Scanning electron micrographs (SEM) showed changes in

    electrode structure before and after EC. Thermo gravimetric analysis

    (TGA) of sludge showed the completion of the oxidation process at

    5000C leaving behind an ash fraction of approximately 84%. Point of

    zero charge (pH PZC) of fluoride bearing sludge showed a value of

    8.2.

    KeywordsDefluoridation, Electrocoagulation, SEM, TGA,PZC.

    I. INTRODUCTION

    N India, endemic fluoride occurs in many parts of India.

    Fluoride concentration in ground water range from < 0.2

    18 mgL-1 in the States of Jammu and Kashmir, < 0.2 - 6.5

    mgL-1

    in Himachal Pradesh, > 1.5 mgL-1

    in Rajasthan, 0.2 - 0.6

    in Haryana, 0.35 - 15 mgL-1

    in Bihar, on an average 12 mgL-1

    in West Bengal, 15 - 20 mgL-1 in Chattisgarh, 8.2 to 13.2

    mgL-1

    in Orissa and 0.7 to 6.0 in Maharashtra. In Karnataka,

    fluoride hotspots are found in Raichur, Gulbarga, Bellary and

    parts of Chitradurga with values ranging from 2.0 7.5 mgL-1.

    Low fluoride concentration (< 1.0 mgL-1) could prevent dental

    problem, but higher fluoride concentration (> 1.5 mgL-1

    ) will

    cause dental and skeletal fluorosis [1]. Apart from these two

    major effects of fluorosis, other ailments include neurological,

    muscular and allergic manifestations. In addition to oxidative

    stress [2, 3], commonly observed effects of fluoride in animals

    include damaged sperms [4], and low birth rate [5].

    The prevailing most popular processes for drinking water

    Sanjeev. S is with Department of Environmental Engineering, NITK,

    Surathkal 575 025 , India (e-mail: [email protected])

    Rohini. J is with Department of Environmental Engineering, SJCE, Mysore

    570 006 , India (e-mail: [email protected])

    Naveen. S. M is with Department of Environmental Engineering, SJCE,

    Mysore 570 006 , India (e-mail: [email protected])

    Mahesh. S is working as Assistant Professor in the Department of

    Environmental Engineering, SJCE, Mysore 570 006, India. (e-mail:

    [email protected])

    defluoridation are the adsorption using activated alumina [6],

    bone char [7], activated carbon [8] and other adsorbents [9]

    and the coagulation using aluminum salts [10]. Other major

    processes for defluoridation include electro-dialysis [11]

    reverse osmosis [12] and nano-filtration [13]. Recently,

    Mameri, 2001, 1998 [14, 15] showed the effectiveness of ECT

    in drinking water defluoridation.

    The present research focuses on complete defluoridation of

    ground water using electrochemical coagulation technology.

    The main objective was to remove excess fluoride from

    drinking water( ground water) to the prescribed drinking

    water standards of 1 - 1.5 mgL-1.

    A. EC Mechanism with Aluminum Electrodes

    Defluoridation is achieved by forming Al (OH)3-XFX. Small

    hydrogen gas bubbles evolved at the cathode enhances the Al

    (OH)3-XFX flocs to form atop the ECR. Effective

    defluoridation is finally achieved when the flocs are removed

    by some mechanism.3+

    Al + 3e at anodeAl (1)

    ( )3 2 33H O Al OH 3Al H + +

    + + (2)

    ( ) ( )- -x3 3-xAl OH xF Al OH F + xOH+ (3)-

    2 22H O + 2e H + 2OH at cathode (4)

    Coexisting anions such as SO42- could affect waterdefluoridation in the EC process. Since some raw water,

    especially underground water may contain high concentrations

    of coexisting ions; it becomes necessary to quantify the effects

    of the coexisting ions on drinking water defluoridation in the

    EC process. If SO42-

    concentration is in excess of 250 mg L-1

    ,

    defluoridation tends to be difficult and cause localized

    corrosion of aluminum electrodes.

    The effect of Cl-on fluoride removal is small. Cl

    -can affect

    defluoridation in two different ways. On one hand, Cl-

    is

    known to rupture the passive films of aluminum electrodes,

    [16]. As Cl-

    concentration increases, the current efficiency of

    EC also increases, and therefore, more aluminum species get

    generated. This benefits defluoridation. On the other hand, ionexchange competition is likely to occur between Cl- and F-. As

    the Cl-

    concentration increases, defluoridation perhaps is

    slightly inhibited.

    II. EXPERIMENTAL

    The schematic arrangement of the experimental set up is

    shown in Fig. 1. Experiments were carried out in a cubical

    reactor (9.2cm x 9.2cm x 14.3cm) of capacity 1 L. Aluminum

    plate electrodes (of thickness 1 mm) having dimensions of

    67mm x 67mm were arranged in parallel in a monopolar

    Electrocoagulation Using DC Current for

    Removal of Fluoride from Ground Water

    Sanjeev. S, Rohini. J, Naveen. S. M, Mahesh. S

    I

    219

  • 8/3/2019 EC-flouride

    2/5

    PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740

    ENVIROENERGY2009

    configuration in the ECR. The completely submerged

    electrodes were connected to the positive and negative

    terminals of the DC power supply unit. Experiments were

    carried out by varying cell voltage. To attain uniform mixing,

    stirring was achieved using a magnetic bar placed between the

    bottom of the reactor. The gaps between the two neighboring

    electrode plates were varied between 5 and 15mm throughout

    the experimental runs.

    Fig. 1 Schematic diagram of the experimental setup.

    Batch EC experiments were carried out for a duration of 60

    min in each run. Samples were retrieved and filtered at regular

    time intervals, filtered and analyzed for F, pH, turbidity and

    conductivity. Fluoride determinations and other parameters

    were in accordance with standard methods [17]. All

    experiments were carried out at ambient temperature with

    different cell voltages (1, 2, 3, 4, 5, 6, 7, 8, 9,10V) as also by

    varying the number of electrodes (2, 4, 6, 8, and 10). Cell

    voltage was noted at regular intervals. After 60 min ET, the

    contents of the ECR were emptied in to a jar and a certain

    time was allowed for settling of the Al flocs in the sludge. The

    sludge contains a matrix of H2, O2, Al and other materials

    which settle down on little agitation in a Jar Test apparatus.The sludge was dried and subjected to further analysis. The

    sludge was dried in an oven and subjected to tests such as

    SEM, TGA and PZC. Filtered samples were analyzed using

    fluoride ion analyzer. Optimized process and operating

    parameters (4 electrodes, 6 volts for about 20 min analyses)

    were then utilized for continuous experiments. Defluoridated

    water was characterized for various other drinking water

    parameters after ECT.

    At low applied voltage, sedimentation dominates with a

    slow release of coagulant and gentle agitation provided by low

    bubble density and production [18, 19]. At higher current,

    faster removal is expected, because of the greater bubble

    density which result in a greater mass buoyed up to the surfacegiving higher rate constants [20] or greater mass flow-out

    causing prolonged EC times.

    III. RESULTS AND DISCUSSION

    Fluoride bearing ground water prior to its use in the EC

    experiments had the following characteristics: pH ~7.5-7.8,

    turbidity 1.0 1.5 NTU, fluoride 5.4 6.0 mgL-1

    , total

    hardness 300.0 305.0 mgL-1

    , chlorides 120.0 125.0 mgL-1

    ,

    nitrates 0.3 0.5 mgL-1

    , phosphates 1.25 1.50 mgL-1

    , total

    solids 8.0 10.0 mgL-1

    , and sulphates 5.0 6.0 mgL-1

    .

    A. Batch EC Experiments

    Fig. 2 shows fluoride degradation curves for cell voltages 1,

    2, 4 and 6V respectively for an initial fluoride concentration of

    5.4 mg L-1

    for a 4 electrode monopolar parallel configuration

    (surface area of electrodes to the volume of the bulk solution

    in the ECR - SA/V = 35.92m2

    m-3

    ). As seen, for an applied

    voltage of 1V, fluoride concentration decreases to 3.3 mgL-1

    from its initial value of 5.4 mgL-1

    . Also, for an applied voltage

    of 2V, fluoride concentration decreases from 5.4 to 2.3.Similarly, for an applied voltage of 4 V and 6V, fluoride

    concentration decreases to 1.8 and 0.8 mgL-1

    from its initial

    concentration of 5.4 mgL-1

    respectively. As seen, at higher cell

    voltages (> 6 V), marginal removal of fluoride is observed.

    80% of fluoride removal takes place within 20 minute of ET.

    The pH of the bulk solution in the ECR shows an increase

    from 7.76 to 8.75. As time progresses, anode dissolution take

    place, so also a small decrease in the applied current (I).

    0

    1

    2

    3

    4

    5

    6

    0 5 10 15 20 25 30

    ET, min

    Fluoridevalue,mgL-1

    1V

    2V

    4V

    6V

    Fig. 2 Percentage fluoride removal as function of ET. F0 = 5.4 mgL-1,

    Number of electrodes = 4; pH0 = 7.76; SA/V = 35.92 m2 m-3

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    0 5 10 15 20

    ET, min

    %F

    luorideremoval

    10 mgL-1

    15 mgL-1

    20 mgL-1

    Fig. 3 Fluoride removal as a function of ET. No. of electrodes= 4;

    pH0 = 7.76; SA/V = 35.92m2 m-3

    UPS

    + -

    DC rectifier

    Peristaliticpump

    Magnetic stirrer

    ECcellStirrer

    Feed tank

    collecting tank

    220

  • 8/3/2019 EC-flouride

    3/5

    PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740

    ENVIROENERGY2009

    Fig. 3 shows the fluoride removal as function of time. This

    experiment was carried out at higher fluoride concentrations

    of 10, 15, 20 mgL-1

    at the optimal operating conditions (4

    electrodes, SA/V = 35.92m2

    m-3

    )for 20 min ET. Sodium

    fluoride was used to spike up fluoride concentrations in the

    ground water samples. At a fluoride spike of 10 mgL-1

    ,

    fluoride removal is 93%. Similarly, at 15 mgL-1

    Fo, fluoride

    removal is 90% and lastly at 20 mgL-1, the removal was 87%.

    From this experiment, it was inferred that fluoride

    concentrations in excess of 10 mgL-1 could be effectively

    removed from water even at small cell voltages of 6V.

    At a cell voltage of 1 V, pH increases from 7.36 to 7.76 and

    for 3V, pH increases from 7.36 to 8.4, in case of third case i.e.

    for 4 V, pH increases from 7.36 to 8.7, and for 6V it increased

    from 7.36 to 8.9. It was observed that a higher voltage

    provides a higher end pH of the bulk solution. Overall, the pH

    of the cell shows increase from 7.36 to 8.9. Anode dissolution

    was found to be a function of cell voltage. As the cell voltage

    was increased, the anode dissolution also increased.

    B. Continuous EC Experiments

    Fig. 4 shows the percentage fluoride removal as a function

    of time in min at optimal operating conditions for differentflow rates i.e. 1, 2, 5 Lh

    -1. Samples were drawn at regular

    intervals up to 300 minutes (i.e. 5 hours treatment time).

    Samples so retrieved were filtered and fluoride concentration

    was analyzed using fluoride ion analyzer. At the flow rate of

    1 Lh-1

    , fluoride removal was ~82%; at 2 Lph flow rate

    fluoride removal was 82.4% and at 5 Lh-1

    ,82.7% as shown in

    Fig 5. Higher voltages were avoided as anode dissolution was

    high. A higher SA/V, i.e. over twice that of Mameri et al.,

    2001 was adopted in the present study, showed more

    consistent results on F removal.

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    0 50 100 150 200 250 300

    ET, min

    %

    Fluorideremoval

    1 Lph

    2 Lph

    5 Lph

    Fig. 4 Fluoride removal as function of ET at optimal conditions. F 0=

    5.4mgL-1; Number of electrodes=4; Applied voltage = 6V; pH0 =

    7.76.

    C. Sludge Analysis

    The mineral ions eliminated during EC from the bulk

    solution and aluminum from the electrodes were combined

    with the fluoro-complexes and insoluble compounds. These

    insoluble compounds get added with the aluminum hydroxide

    precipitate which produces the gel like insoluble precipitate

    which floats atop the ECR. Fig 5 shows sludge gel like

    emulsion after each EC experiment. Low voltages provide less

    sludge volume. The sludge obtained after settling in a beaker

    was oven dried and subjected to further analysis for TGA,

    SEM, and PZC. The pH PZC of fluoride bearing sludge showed

    a value of 8.20.

    Fig. 5 Gel like emulsion atop the ECR

    D. Thermal Analysis of Sludge

    TGA studies indicate the thermal stability of the sludge and

    provide information on the nature of the complex organic

    substances (Mahesh et al., 2006 Part I) in the sludge.

    Fig. 6 shows the thermogravimetric analysis (TGA),

    differential thermogravimetric analysis (DTGA) and

    derivative thermal analysis (DTA) curves for the precipitated

    sludge after ECT of fluoride contaminated drinking water. The

    thermal characteristics were observed both in the oxidizing

    (air) as well as inert (nitrogen) atmospheres at the heating rate

    of 10 K min-1 and air/nitrogen flushing rate of 0.4 dm3

    min-1

    .

    TempCel900800700600500400300200100

    DTA

    uV

    20.0

    0.0

    -20.0

    -40.0

    -60.0

    TG

    %

    120.00

    115.00

    110.00

    105.00

    100.00

    95.00

    90.00

    85.00

    DTGug/min

    100.0

    0.0

    -100.0

    -200.0

    -300.0

    -400.0

    63Cel-1.4uV

    60.9mJ/mg

    24Cel99.99%

    989Cel83.61

    99Cel94.87%

    200Cel90.35%

    301Cel87.44%

    401Cel85.15%500Cel

    84.66%600Cel84.30%

    699Cel84.11%

    799Cel83.95%

    899Cel83.78%

    150Cel92.55%

    250Cel88.74%

    376Cel85.51%

    66Cel97.11%

    55Cel

    86.2ug/min

    169Cel51.4ug/min 353Cel

    30.4ug/min

    Fig. 6 DTA-DTG -TG plots of EC sludge at various conditions in

    air atmosphere.

    221

  • 8/3/2019 EC-flouride

    4/5

    PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740

    ENVIROENERGY2009

    Figure shows the TG/DTG and DTA behaviour of the solid

    residue of the black liquor under oxidizing environment. The

    TG trace shows a gradual decrease in the residual sample mass

    upto a temperature of 401 oC shedding about 85% of the initial

    sample mass. The weight-loss rate is found to be extremely

    slow, and up to 500oC (over a temperature span of 99

    oC), the

    weight-loss is only 0.49%. This means that the sludge sample

    loses moisture at an almost steady rate along with

    volatalization of light volatiles upto 401oC and thereafter, the

    sample becomes dry and stable.

    The maximum weight-loss rate was 86.2 micrograms min-1

    at Tmax of 55oC (see DTGA trace). The peak temperature for

    the exothermic reaction as exemplified by the DTA curve was

    at Tp = 63oC with heat release of 60.9 mJ mg-1. Beyond

    500oC, the weight-loss is steady but very slow, giving off only

    ~0.88 % mass from 500-989oC (over a temperature increase

    of 489oC). It is found that the organics of the precipitate get

    oxidized leaving behind the ash fraction of 83.61%.

    E. SEM micrographs

    SEM micrographs of aluminium electrodes (anode), before

    and after ECT were obtained to compare their surface texture.

    The surface of the anode prior to its use in the EC experimentswas found to be uniform, bearing nano-sized crystals.

    Fig. 7a Surface of the electrode before ECT

    Fig. 7b Surface of the electrode after ECT

    Fig. 7a. shows the SEM micrograph of the same anode plate

    after several cycles of its use in EC experiments for a total

    duration of ET ~ 8 h. The anode plate surface was found to be

    rough, with a large number of small sized three dimensional

    dents of ~100- 200 micro meters in width and depth. These

    dents are formed around the nuclei of the active sites where

    the anode dissolution occurs producing aluminium

    hydroxides. The edges of the electrodes (anode) and the plate

    surfaces wear off releasing iron hydroxide for floc formation

    with an increase in the number of cycles of experiments. After

    repeated cycles of EC runs, these dents increase in size all

    over the active side of the plate leaving behind an eroded

    surface as shown in Fig. 7b.

    IV. CONCLUSION

    Batch and continuous EC experiments carried out have

    confirmed the efficacy of EC technique for the treatment of

    fluoride rich ground water. A 4 plate arrangement in parallel

    having an SA/V of 35.92 m2

    m-3

    operating at 6V cell voltage

    was found optimal in terms of effective defluoridation (>88%

    removal), minimum anode dissolution and manageable final

    pH of the treated water. Continuous EC experiments with

    spiking fluoride concentrations and at different flow rates 1

    Lh-1, 2 Lh-1 and 5 Lh-1 showed the fluoride removal of 86%,

    82%, 78% respectively. Other than the fluoride, other

    parameters like sulfates, chlorides, nitrates, phosphates,

    hardness were reduced to 30 -40%.

    Aluminium was not detected in the treated water pointing to

    fact that the electro- dissolved alumnium forms hydroxide

    during ECC and complex with organics to form insoluble

    precipitates the insoluble complex gets flocculated and floats a

    top the ECR. SEM micrographs showed changes in electrode

    structure before and after ECC. TGA of sludge showed the

    completion of the oxidation process 5000C behind an ash

    fraction of approximately 84%. pH PZC of fluoride bearing

    sludge showed a value of 8.2.

    ACKNOWLEDGMENT

    The financial support by MOEF/F. No. 19/23/2008 RE

    11.09.08. is gratefully acknowledged by the authors.

    REFERENCES

    [1] WHO Guidelines for Drinking Water Equality, World Health

    Organization. September, 2004.

    [2] J. Krechniak, I. Inkielewicz. Correlations between fluoride

    concentrations and free radical parameters in soft tissues of rats.

    Fluoride, 2005, vol. 38(4), pp. 293-296.

    [3] D. Chlubek. Fluoride and oxidative stress. Fluoride 2003, vol. 36(4),

    pp. 217-218.

    [4] H. Zakrzewska, J. Udaa, B. Baszczyk. In vitro influence of sodium

    fluoride on ram semen quality and enzyme activities. Fluoride. 2002,

    vol. 35(3), pp. 153-160.

    [5] S. Freni. Exposure to high fluoride concentration in drinking water is

    associated with decreased birth rates. J. Toxicol Environ Health 1994,

    vol. 42(1), pp. 109-112.

    [6] V. S. Chauhan, P. K. Dwivedi, L. Iyengar, Investigations on activated

    alumina based domestic de.uoridation units, J. Hazard. Mater. 2007,

    vol. 139, pp. 103-107.

    [7] H. Mjengera, G. Mkongo, Appropriate deflouridation technology for

    use in flourotic areas in Tanzania, Phys. Chem. Earth 2003, vol. 28,

    pp. 1097-1104.

    [8] S. Kumar, A. Gupta, J. P. Yadav, Fluoride removal by-mixtures of

    activated carbon prepared from Neem (Azadirachta indica) and Kikar

    (Acacia arabica) leaves, Ind. J. Chem. Techn. 2007, vol. 14, pp. 355-

    361.

    [9] K. Biswas, S. K. Saha, U. C. Ghosh, Adsorption of fluoride from

    aqueous solution by a synthetic Iron(III)-Aluminum(III) mixed oxide,

    Ind. Eng. Chem. Res. 2007, vol. 46, pp. 5346-5356.

    [10] M. Pinon-Miramontes, R. G. Bautista-Margulis, A. Perez-Hernandez,

    Removal of arsenic and fluoride from drinking water with cake alum

    222

  • 8/3/2019 EC-flouride

    5/5

    PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740

    ENVIROENERGY2009

    and a polymeric anionic flocculent, Fluoride 2003, vol. 36, pp. 122-

    128.

    [11] M. Tahaikt, I. Achary, M. A. Menkouchi Sahli, Z. Amor, M. Taky, A.

    Alami, A. Boughriba, M. Hafsi, A. Elmidaoui, Defluoridation of

    Moroccan groundwater by electrodialysis: continuous operation,

    Desalination, 2006, vol. 189, pp. 215-220.

    [12] M. Arora, R. C. Maheshwari, S. K. Jain, A. Gupta, Use of membrane

    technology for potable water production, Desalination, 2004, vol. 170,

    pp. 105-112.

    [13] K. Hu, J. M. Dickson, Nanofiltration membrane performance on

    fluoride removal from water, J. Mem. Sci, 2006, vol. 279, pp. 529-538.

    [14] N. Mameri, H. Lounici, D. Belhocine, H. Grib, D.L. Piron, Y. Yahiat,Defluoridation of Sahara water by small plant electrocoagulation using

    bipolar aluminium electrodes. Sep. Puri. Techno., 2001, vol. 24, pp.

    113-119.

    [15] N. Mameri, A. R. Yeddou, H. Lounici, D. Belhocine, H. Grib, B. Bariou,

    Deflouridation of Septentrional Sahara water of North Africa by

    electrocoagulation process using bipolar aluminium electrodes. Wat.

    Res. 1998, vol. 32(5), pp. 1604-1612.

    [16] C. Y. Hu, S. L. Lo, W. H. Kuan, Effects of co-existing anions on

    fluoride removal in electrocoagulation (EC) process using aluminum

    electrodes, Wat. Res. 2003, vol. 37 pp. 4513-4523.

    [17] APHA, Standard Methods for the Examination of Water and

    Wastewater, 21st ed., American Public Health Association,

    Washington, DC, 2005.

    [18] S. Mahesh, B. Prasad, I. D. Mall, I. M. Mishra, Electrochemical

    degradation of pulp and paper mill wastewater. Part 1. COD and color

    removal, Ind. Eng. Chem. Res, 2006, vol. 45(8), pp. 2830-2839.

    [19] S. Mahesh, B. Prasad, I. D. Mall, I. M. Mishra, Electrochemical

    Degradation of Pulp and Paper Mill Wastewater. Part 2. Characterization

    and Analysis of Sludge, Ind. Eng. Chem. Res, 2006, vol. 45(16), pp.5766-5774.

    [20] P. K. Holt, G. W. Barton, A.A. Mitchell, Mathematical analysis of a

    batch electrochemical reactor. Water supply. 2002, vol. 5-6, pp. 65-71.

    223