Upload
roshan-ouseph
View
216
Download
0
Embed Size (px)
Citation preview
8/3/2019 EC-flouride
1/5
PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740
ENVIROENERGY2009
Abstract Electrocoagulation (EC) experiments were carried outin batch and continuous modes for removal of excess fluoride from
ground water. During batch studies using 2, 4 ,6 and 8 electrodes, an
optimal operating condition of 6V using 4 plates and 20 min
electrolyze time (ET) was found appropriate removing >88% of
fluoride from ground water. Repeated continuous EC runs at optimal
conditions, showed stable performance of the electrochemical reactor
(ECR) removing fluoride to the required drinking water quality
standards between 1.0 and 1.5mgL-1. Experiments were also carried
out by spiking ground water with excess fluoride from 10-20mgL -1
which showed > 90% fluoride removal. The pH of the water before
ECC was 7.2-7.4 and 8.3-8.6 depending on applied cell voltage.
Fluoride removal during continuous studies for flow rates 1, 2 and 5Lph were 86, 82, and 78% respectively from its initial value of 5.4
mgL-1.
Scanning electron micrographs (SEM) showed changes in
electrode structure before and after EC. Thermo gravimetric analysis
(TGA) of sludge showed the completion of the oxidation process at
5000C leaving behind an ash fraction of approximately 84%. Point of
zero charge (pH PZC) of fluoride bearing sludge showed a value of
8.2.
KeywordsDefluoridation, Electrocoagulation, SEM, TGA,PZC.
I. INTRODUCTION
N India, endemic fluoride occurs in many parts of India.
Fluoride concentration in ground water range from < 0.2
18 mgL-1 in the States of Jammu and Kashmir, < 0.2 - 6.5
mgL-1
in Himachal Pradesh, > 1.5 mgL-1
in Rajasthan, 0.2 - 0.6
in Haryana, 0.35 - 15 mgL-1
in Bihar, on an average 12 mgL-1
in West Bengal, 15 - 20 mgL-1 in Chattisgarh, 8.2 to 13.2
mgL-1
in Orissa and 0.7 to 6.0 in Maharashtra. In Karnataka,
fluoride hotspots are found in Raichur, Gulbarga, Bellary and
parts of Chitradurga with values ranging from 2.0 7.5 mgL-1.
Low fluoride concentration (< 1.0 mgL-1) could prevent dental
problem, but higher fluoride concentration (> 1.5 mgL-1
) will
cause dental and skeletal fluorosis [1]. Apart from these two
major effects of fluorosis, other ailments include neurological,
muscular and allergic manifestations. In addition to oxidative
stress [2, 3], commonly observed effects of fluoride in animals
include damaged sperms [4], and low birth rate [5].
The prevailing most popular processes for drinking water
Sanjeev. S is with Department of Environmental Engineering, NITK,
Surathkal 575 025 , India (e-mail: [email protected])
Rohini. J is with Department of Environmental Engineering, SJCE, Mysore
570 006 , India (e-mail: [email protected])
Naveen. S. M is with Department of Environmental Engineering, SJCE,
Mysore 570 006 , India (e-mail: [email protected])
Mahesh. S is working as Assistant Professor in the Department of
Environmental Engineering, SJCE, Mysore 570 006, India. (e-mail:
defluoridation are the adsorption using activated alumina [6],
bone char [7], activated carbon [8] and other adsorbents [9]
and the coagulation using aluminum salts [10]. Other major
processes for defluoridation include electro-dialysis [11]
reverse osmosis [12] and nano-filtration [13]. Recently,
Mameri, 2001, 1998 [14, 15] showed the effectiveness of ECT
in drinking water defluoridation.
The present research focuses on complete defluoridation of
ground water using electrochemical coagulation technology.
The main objective was to remove excess fluoride from
drinking water( ground water) to the prescribed drinking
water standards of 1 - 1.5 mgL-1.
A. EC Mechanism with Aluminum Electrodes
Defluoridation is achieved by forming Al (OH)3-XFX. Small
hydrogen gas bubbles evolved at the cathode enhances the Al
(OH)3-XFX flocs to form atop the ECR. Effective
defluoridation is finally achieved when the flocs are removed
by some mechanism.3+
Al + 3e at anodeAl (1)
( )3 2 33H O Al OH 3Al H + +
+ + (2)
( ) ( )- -x3 3-xAl OH xF Al OH F + xOH+ (3)-
2 22H O + 2e H + 2OH at cathode (4)
Coexisting anions such as SO42- could affect waterdefluoridation in the EC process. Since some raw water,
especially underground water may contain high concentrations
of coexisting ions; it becomes necessary to quantify the effects
of the coexisting ions on drinking water defluoridation in the
EC process. If SO42-
concentration is in excess of 250 mg L-1
,
defluoridation tends to be difficult and cause localized
corrosion of aluminum electrodes.
The effect of Cl-on fluoride removal is small. Cl
-can affect
defluoridation in two different ways. On one hand, Cl-
is
known to rupture the passive films of aluminum electrodes,
[16]. As Cl-
concentration increases, the current efficiency of
EC also increases, and therefore, more aluminum species get
generated. This benefits defluoridation. On the other hand, ionexchange competition is likely to occur between Cl- and F-. As
the Cl-
concentration increases, defluoridation perhaps is
slightly inhibited.
II. EXPERIMENTAL
The schematic arrangement of the experimental set up is
shown in Fig. 1. Experiments were carried out in a cubical
reactor (9.2cm x 9.2cm x 14.3cm) of capacity 1 L. Aluminum
plate electrodes (of thickness 1 mm) having dimensions of
67mm x 67mm were arranged in parallel in a monopolar
Electrocoagulation Using DC Current for
Removal of Fluoride from Ground Water
Sanjeev. S, Rohini. J, Naveen. S. M, Mahesh. S
I
219
8/3/2019 EC-flouride
2/5
PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740
ENVIROENERGY2009
configuration in the ECR. The completely submerged
electrodes were connected to the positive and negative
terminals of the DC power supply unit. Experiments were
carried out by varying cell voltage. To attain uniform mixing,
stirring was achieved using a magnetic bar placed between the
bottom of the reactor. The gaps between the two neighboring
electrode plates were varied between 5 and 15mm throughout
the experimental runs.
Fig. 1 Schematic diagram of the experimental setup.
Batch EC experiments were carried out for a duration of 60
min in each run. Samples were retrieved and filtered at regular
time intervals, filtered and analyzed for F, pH, turbidity and
conductivity. Fluoride determinations and other parameters
were in accordance with standard methods [17]. All
experiments were carried out at ambient temperature with
different cell voltages (1, 2, 3, 4, 5, 6, 7, 8, 9,10V) as also by
varying the number of electrodes (2, 4, 6, 8, and 10). Cell
voltage was noted at regular intervals. After 60 min ET, the
contents of the ECR were emptied in to a jar and a certain
time was allowed for settling of the Al flocs in the sludge. The
sludge contains a matrix of H2, O2, Al and other materials
which settle down on little agitation in a Jar Test apparatus.The sludge was dried and subjected to further analysis. The
sludge was dried in an oven and subjected to tests such as
SEM, TGA and PZC. Filtered samples were analyzed using
fluoride ion analyzer. Optimized process and operating
parameters (4 electrodes, 6 volts for about 20 min analyses)
were then utilized for continuous experiments. Defluoridated
water was characterized for various other drinking water
parameters after ECT.
At low applied voltage, sedimentation dominates with a
slow release of coagulant and gentle agitation provided by low
bubble density and production [18, 19]. At higher current,
faster removal is expected, because of the greater bubble
density which result in a greater mass buoyed up to the surfacegiving higher rate constants [20] or greater mass flow-out
causing prolonged EC times.
III. RESULTS AND DISCUSSION
Fluoride bearing ground water prior to its use in the EC
experiments had the following characteristics: pH ~7.5-7.8,
turbidity 1.0 1.5 NTU, fluoride 5.4 6.0 mgL-1
, total
hardness 300.0 305.0 mgL-1
, chlorides 120.0 125.0 mgL-1
,
nitrates 0.3 0.5 mgL-1
, phosphates 1.25 1.50 mgL-1
, total
solids 8.0 10.0 mgL-1
, and sulphates 5.0 6.0 mgL-1
.
A. Batch EC Experiments
Fig. 2 shows fluoride degradation curves for cell voltages 1,
2, 4 and 6V respectively for an initial fluoride concentration of
5.4 mg L-1
for a 4 electrode monopolar parallel configuration
(surface area of electrodes to the volume of the bulk solution
in the ECR - SA/V = 35.92m2
m-3
). As seen, for an applied
voltage of 1V, fluoride concentration decreases to 3.3 mgL-1
from its initial value of 5.4 mgL-1
. Also, for an applied voltage
of 2V, fluoride concentration decreases from 5.4 to 2.3.Similarly, for an applied voltage of 4 V and 6V, fluoride
concentration decreases to 1.8 and 0.8 mgL-1
from its initial
concentration of 5.4 mgL-1
respectively. As seen, at higher cell
voltages (> 6 V), marginal removal of fluoride is observed.
80% of fluoride removal takes place within 20 minute of ET.
The pH of the bulk solution in the ECR shows an increase
from 7.76 to 8.75. As time progresses, anode dissolution take
place, so also a small decrease in the applied current (I).
0
1
2
3
4
5
6
0 5 10 15 20 25 30
ET, min
Fluoridevalue,mgL-1
1V
2V
4V
6V
Fig. 2 Percentage fluoride removal as function of ET. F0 = 5.4 mgL-1,
Number of electrodes = 4; pH0 = 7.76; SA/V = 35.92 m2 m-3
0
10
20
30
40
50
60
70
80
90
100
0 5 10 15 20
ET, min
%F
luorideremoval
10 mgL-1
15 mgL-1
20 mgL-1
Fig. 3 Fluoride removal as a function of ET. No. of electrodes= 4;
pH0 = 7.76; SA/V = 35.92m2 m-3
UPS
+ -
DC rectifier
Peristaliticpump
Magnetic stirrer
ECcellStirrer
Feed tank
collecting tank
220
8/3/2019 EC-flouride
3/5
PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740
ENVIROENERGY2009
Fig. 3 shows the fluoride removal as function of time. This
experiment was carried out at higher fluoride concentrations
of 10, 15, 20 mgL-1
at the optimal operating conditions (4
electrodes, SA/V = 35.92m2
m-3
)for 20 min ET. Sodium
fluoride was used to spike up fluoride concentrations in the
ground water samples. At a fluoride spike of 10 mgL-1
,
fluoride removal is 93%. Similarly, at 15 mgL-1
Fo, fluoride
removal is 90% and lastly at 20 mgL-1, the removal was 87%.
From this experiment, it was inferred that fluoride
concentrations in excess of 10 mgL-1 could be effectively
removed from water even at small cell voltages of 6V.
At a cell voltage of 1 V, pH increases from 7.36 to 7.76 and
for 3V, pH increases from 7.36 to 8.4, in case of third case i.e.
for 4 V, pH increases from 7.36 to 8.7, and for 6V it increased
from 7.36 to 8.9. It was observed that a higher voltage
provides a higher end pH of the bulk solution. Overall, the pH
of the cell shows increase from 7.36 to 8.9. Anode dissolution
was found to be a function of cell voltage. As the cell voltage
was increased, the anode dissolution also increased.
B. Continuous EC Experiments
Fig. 4 shows the percentage fluoride removal as a function
of time in min at optimal operating conditions for differentflow rates i.e. 1, 2, 5 Lh
-1. Samples were drawn at regular
intervals up to 300 minutes (i.e. 5 hours treatment time).
Samples so retrieved were filtered and fluoride concentration
was analyzed using fluoride ion analyzer. At the flow rate of
1 Lh-1
, fluoride removal was ~82%; at 2 Lph flow rate
fluoride removal was 82.4% and at 5 Lh-1
,82.7% as shown in
Fig 5. Higher voltages were avoided as anode dissolution was
high. A higher SA/V, i.e. over twice that of Mameri et al.,
2001 was adopted in the present study, showed more
consistent results on F removal.
0
10
20
30
40
50
60
70
80
90
100
0 50 100 150 200 250 300
ET, min
%
Fluorideremoval
1 Lph
2 Lph
5 Lph
Fig. 4 Fluoride removal as function of ET at optimal conditions. F 0=
5.4mgL-1; Number of electrodes=4; Applied voltage = 6V; pH0 =
7.76.
C. Sludge Analysis
The mineral ions eliminated during EC from the bulk
solution and aluminum from the electrodes were combined
with the fluoro-complexes and insoluble compounds. These
insoluble compounds get added with the aluminum hydroxide
precipitate which produces the gel like insoluble precipitate
which floats atop the ECR. Fig 5 shows sludge gel like
emulsion after each EC experiment. Low voltages provide less
sludge volume. The sludge obtained after settling in a beaker
was oven dried and subjected to further analysis for TGA,
SEM, and PZC. The pH PZC of fluoride bearing sludge showed
a value of 8.20.
Fig. 5 Gel like emulsion atop the ECR
D. Thermal Analysis of Sludge
TGA studies indicate the thermal stability of the sludge and
provide information on the nature of the complex organic
substances (Mahesh et al., 2006 Part I) in the sludge.
Fig. 6 shows the thermogravimetric analysis (TGA),
differential thermogravimetric analysis (DTGA) and
derivative thermal analysis (DTA) curves for the precipitated
sludge after ECT of fluoride contaminated drinking water. The
thermal characteristics were observed both in the oxidizing
(air) as well as inert (nitrogen) atmospheres at the heating rate
of 10 K min-1 and air/nitrogen flushing rate of 0.4 dm3
min-1
.
TempCel900800700600500400300200100
DTA
uV
20.0
0.0
-20.0
-40.0
-60.0
TG
%
120.00
115.00
110.00
105.00
100.00
95.00
90.00
85.00
DTGug/min
100.0
0.0
-100.0
-200.0
-300.0
-400.0
63Cel-1.4uV
60.9mJ/mg
24Cel99.99%
989Cel83.61
99Cel94.87%
200Cel90.35%
301Cel87.44%
401Cel85.15%500Cel
84.66%600Cel84.30%
699Cel84.11%
799Cel83.95%
899Cel83.78%
150Cel92.55%
250Cel88.74%
376Cel85.51%
66Cel97.11%
55Cel
86.2ug/min
169Cel51.4ug/min 353Cel
30.4ug/min
Fig. 6 DTA-DTG -TG plots of EC sludge at various conditions in
air atmosphere.
221
8/3/2019 EC-flouride
4/5
PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740
ENVIROENERGY2009
Figure shows the TG/DTG and DTA behaviour of the solid
residue of the black liquor under oxidizing environment. The
TG trace shows a gradual decrease in the residual sample mass
upto a temperature of 401 oC shedding about 85% of the initial
sample mass. The weight-loss rate is found to be extremely
slow, and up to 500oC (over a temperature span of 99
oC), the
weight-loss is only 0.49%. This means that the sludge sample
loses moisture at an almost steady rate along with
volatalization of light volatiles upto 401oC and thereafter, the
sample becomes dry and stable.
The maximum weight-loss rate was 86.2 micrograms min-1
at Tmax of 55oC (see DTGA trace). The peak temperature for
the exothermic reaction as exemplified by the DTA curve was
at Tp = 63oC with heat release of 60.9 mJ mg-1. Beyond
500oC, the weight-loss is steady but very slow, giving off only
~0.88 % mass from 500-989oC (over a temperature increase
of 489oC). It is found that the organics of the precipitate get
oxidized leaving behind the ash fraction of 83.61%.
E. SEM micrographs
SEM micrographs of aluminium electrodes (anode), before
and after ECT were obtained to compare their surface texture.
The surface of the anode prior to its use in the EC experimentswas found to be uniform, bearing nano-sized crystals.
Fig. 7a Surface of the electrode before ECT
Fig. 7b Surface of the electrode after ECT
Fig. 7a. shows the SEM micrograph of the same anode plate
after several cycles of its use in EC experiments for a total
duration of ET ~ 8 h. The anode plate surface was found to be
rough, with a large number of small sized three dimensional
dents of ~100- 200 micro meters in width and depth. These
dents are formed around the nuclei of the active sites where
the anode dissolution occurs producing aluminium
hydroxides. The edges of the electrodes (anode) and the plate
surfaces wear off releasing iron hydroxide for floc formation
with an increase in the number of cycles of experiments. After
repeated cycles of EC runs, these dents increase in size all
over the active side of the plate leaving behind an eroded
surface as shown in Fig. 7b.
IV. CONCLUSION
Batch and continuous EC experiments carried out have
confirmed the efficacy of EC technique for the treatment of
fluoride rich ground water. A 4 plate arrangement in parallel
having an SA/V of 35.92 m2
m-3
operating at 6V cell voltage
was found optimal in terms of effective defluoridation (>88%
removal), minimum anode dissolution and manageable final
pH of the treated water. Continuous EC experiments with
spiking fluoride concentrations and at different flow rates 1
Lh-1, 2 Lh-1 and 5 Lh-1 showed the fluoride removal of 86%,
82%, 78% respectively. Other than the fluoride, other
parameters like sulfates, chlorides, nitrates, phosphates,
hardness were reduced to 30 -40%.
Aluminium was not detected in the treated water pointing to
fact that the electro- dissolved alumnium forms hydroxide
during ECC and complex with organics to form insoluble
precipitates the insoluble complex gets flocculated and floats a
top the ECR. SEM micrographs showed changes in electrode
structure before and after ECC. TGA of sludge showed the
completion of the oxidation process 5000C behind an ash
fraction of approximately 84%. pH PZC of fluoride bearing
sludge showed a value of 8.2.
ACKNOWLEDGMENT
The financial support by MOEF/F. No. 19/23/2008 RE
11.09.08. is gratefully acknowledged by the authors.
REFERENCES
[1] WHO Guidelines for Drinking Water Equality, World Health
Organization. September, 2004.
[2] J. Krechniak, I. Inkielewicz. Correlations between fluoride
concentrations and free radical parameters in soft tissues of rats.
Fluoride, 2005, vol. 38(4), pp. 293-296.
[3] D. Chlubek. Fluoride and oxidative stress. Fluoride 2003, vol. 36(4),
pp. 217-218.
[4] H. Zakrzewska, J. Udaa, B. Baszczyk. In vitro influence of sodium
fluoride on ram semen quality and enzyme activities. Fluoride. 2002,
vol. 35(3), pp. 153-160.
[5] S. Freni. Exposure to high fluoride concentration in drinking water is
associated with decreased birth rates. J. Toxicol Environ Health 1994,
vol. 42(1), pp. 109-112.
[6] V. S. Chauhan, P. K. Dwivedi, L. Iyengar, Investigations on activated
alumina based domestic de.uoridation units, J. Hazard. Mater. 2007,
vol. 139, pp. 103-107.
[7] H. Mjengera, G. Mkongo, Appropriate deflouridation technology for
use in flourotic areas in Tanzania, Phys. Chem. Earth 2003, vol. 28,
pp. 1097-1104.
[8] S. Kumar, A. Gupta, J. P. Yadav, Fluoride removal by-mixtures of
activated carbon prepared from Neem (Azadirachta indica) and Kikar
(Acacia arabica) leaves, Ind. J. Chem. Techn. 2007, vol. 14, pp. 355-
361.
[9] K. Biswas, S. K. Saha, U. C. Ghosh, Adsorption of fluoride from
aqueous solution by a synthetic Iron(III)-Aluminum(III) mixed oxide,
Ind. Eng. Chem. Res. 2007, vol. 46, pp. 5346-5356.
[10] M. Pinon-Miramontes, R. G. Bautista-Margulis, A. Perez-Hernandez,
Removal of arsenic and fluoride from drinking water with cake alum
222
8/3/2019 EC-flouride
5/5
PROCEEDINGS OF INTERNATIONAL CONFERENCE ON ENERGY AND ENVIRONMENT MARCH 19-21, 2009 ISSN: 2070-3740
ENVIROENERGY2009
and a polymeric anionic flocculent, Fluoride 2003, vol. 36, pp. 122-
128.
[11] M. Tahaikt, I. Achary, M. A. Menkouchi Sahli, Z. Amor, M. Taky, A.
Alami, A. Boughriba, M. Hafsi, A. Elmidaoui, Defluoridation of
Moroccan groundwater by electrodialysis: continuous operation,
Desalination, 2006, vol. 189, pp. 215-220.
[12] M. Arora, R. C. Maheshwari, S. K. Jain, A. Gupta, Use of membrane
technology for potable water production, Desalination, 2004, vol. 170,
pp. 105-112.
[13] K. Hu, J. M. Dickson, Nanofiltration membrane performance on
fluoride removal from water, J. Mem. Sci, 2006, vol. 279, pp. 529-538.
[14] N. Mameri, H. Lounici, D. Belhocine, H. Grib, D.L. Piron, Y. Yahiat,Defluoridation of Sahara water by small plant electrocoagulation using
bipolar aluminium electrodes. Sep. Puri. Techno., 2001, vol. 24, pp.
113-119.
[15] N. Mameri, A. R. Yeddou, H. Lounici, D. Belhocine, H. Grib, B. Bariou,
Deflouridation of Septentrional Sahara water of North Africa by
electrocoagulation process using bipolar aluminium electrodes. Wat.
Res. 1998, vol. 32(5), pp. 1604-1612.
[16] C. Y. Hu, S. L. Lo, W. H. Kuan, Effects of co-existing anions on
fluoride removal in electrocoagulation (EC) process using aluminum
electrodes, Wat. Res. 2003, vol. 37 pp. 4513-4523.
[17] APHA, Standard Methods for the Examination of Water and
Wastewater, 21st ed., American Public Health Association,
Washington, DC, 2005.
[18] S. Mahesh, B. Prasad, I. D. Mall, I. M. Mishra, Electrochemical
degradation of pulp and paper mill wastewater. Part 1. COD and color
removal, Ind. Eng. Chem. Res, 2006, vol. 45(8), pp. 2830-2839.
[19] S. Mahesh, B. Prasad, I. D. Mall, I. M. Mishra, Electrochemical
Degradation of Pulp and Paper Mill Wastewater. Part 2. Characterization
and Analysis of Sludge, Ind. Eng. Chem. Res, 2006, vol. 45(16), pp.5766-5774.
[20] P. K. Holt, G. W. Barton, A.A. Mitchell, Mathematical analysis of a
batch electrochemical reactor. Water supply. 2002, vol. 5-6, pp. 65-71.
223