71
1 Department of Clinical Chemistry Faculty of Associated Medical Sciences Chiang Mai university < [email protected] > [email protected]

Iqc Eqa Pt มน

Embed Size (px)

Citation preview

Page 1: Iqc Eqa Pt มน

1

Department of Clinical Chemistry Faculty of Associated Medical Sciences

Chiang Mai university

<002aschi @chiangmai.ac.th

>

[email protected]

Page 2: Iqc Eqa Pt มน

2

Discuss and practicing about the following :

• Internal quality control (IQC)

• External quality assessment (EQA)

• Proficiency testing (PT)

Page 3: Iqc Eqa Pt มน

3

BMC Clin Pathol 2001; 1: 5

6 months monitoring, 935,896 - specimens 941902 analyses

- Pre analytical mistakes = 1048 (0.11%)

= Wrong identification :

Physician’s order mistakes, Inapp. Container used, Inapp. Quantity of specimen, Inapp. Q uality of specimens

Page 4: Iqc Eqa Pt มน

4

Lapworth, Ann Clin Biochem 1994, 31: 78-84 Clin Chem / 1 yr / ~9 9 7 0 0 0 tests / 249000 pts

No. of errors 1 2 0 => Frequency 0 .0 5 % of pts.- Pre analytical mistakes = 31.6% Analytical = 31.6%

Post analytical = 30.8% Multiple phases = 6%

Identification errors 31%

Page 5: Iqc Eqa Pt มน

5

Goldschmidt & Lent, Klin Biochem Metab 1995; 3: 131-40

Whole lab / 6 yr / No. of errors 133 - Pre analytical mistakes =

53% Analytical =

23% Post analytical = 24% Identification errors 58%

Impact on patient’s outcomes : None 43%

Mild 23% (delay) Moderate 26% (potential damage)

Severe 8% (errorneousmed .Intervention)

Very severe None

Page 6: Iqc Eqa Pt มน

6

Plebani & Carraro, Clin Chem 1997; 43: 1348-51

Stat lab / 3 mo / No. of total / errors 40490/189 Frequency = 0.47% of test results

- Pre analytical mistakes =68.2%

Analytical =13.3%

Post analytical =18.5% Identification errors 2.6%

Impact on patient’s outcomes : None 74%

Mild 19.6% (further investigation) Moderate 6.4% (inapprop. Therapy

modif.)

Page 7: Iqc Eqa Pt มน

7

Runner et al, Arch PatholLab Med 1993; 117: 573-7

Transfusion labs /4 wk / 712 hospitals No. of occasions 2 463 727 errors 67 289

Frequency = 2.7%

Types & Cause of errors Absent wristband 49.5%

Multiple wristband 18.3%

Incomplete data 17.5%

Partly errorneous data 0.6%

Illegible data 5.7%

Wrong wristband 0.5%

Page 8: Iqc Eqa Pt มน

8

McClelland & Phillips, BMJ 1994; 308: 1205-6

UK Transfusion /2 yr / 245hospitals/ No. of errors 111

Death 6 (1/550000 Tx)

Morbidity 12 (1/275000 Tx)

No adverse effect 93 (1/36000 Tx)

Types & Causes : Wrong blood tubes 23

Lab errors 6

Tx wrong blood 82

Results on Patient outcomes

Page 9: Iqc Eqa Pt มน

9

Hofgartner & Tait, Am J Clin Pathol 1999; 112: 14-21

2 Molecular (Survey)labs / 10 yr / No. of tests 4234 errors 16

Frequency = 0.38%

- Pre analytical mistakes = 44% Analytical = 31%

Post analytical = 12.5% Multiple phase = 12.5

Impact on patient’s outcomes :Mild 25%

Moderate 50 % Severe 25%

Page 10: Iqc Eqa Pt มน

10

Hofgartner & Tait, Am J Clin Pathol 1999; 112: 14-21

2 Molecular labs / 1 yr / (Questionaire) No. of tests 88394 errors 293

Frequency = 0.33 %

- Pre analytical mistakes = 60% Analytical = 19%

Post analytical = 15% Multiple phase = 6%

Impact on patient’s outcomes :Mild 20 %

Moderate 10.2 % Severe 6.4 %

Page 11: Iqc Eqa Pt มน

11

Pre-analytical errors/ yr 2000 / San Raffaele Hosp.

Types of errors Inpatients Outpatients Hemolysed serum 8494 256

Insufficient sample 3256 102

Incorrect sample 1824 289

Clotted sample 792 80 Incorrect ident 287 2

Lack of signature (Bl Gr) 266 Empty tubes 238 8

Sample not on ice 75 6 Lack/wrong compilation of accom. Tube 120

Tube broken in the centrifuge 57 36 Urine not acidified 24 Open container 20 13

Module withour signature 14 Urine volume not indicated 5

TOTAL 15 503 792

Page 12: Iqc Eqa Pt มน

12

Steps of IQC Process

@ การเตร�ยมผู้��ป่�วย @ เก�บร�กษาสิ่��งสิ่�งตรวจถู�กต�อง @ จ�ดระบบภายใน Lab. เหมาะสิ่ม @ ว�ธี�การว�เคราะห$เชื่&�อถู&อได� @ การค(านวณผู้ลถู�กต�อง

@ พิ�จารณาค�าที่��ได�ว�ายอมร�บได�หร&อต�องที่(าซ้ำ(.า@ การรายงานผู้ล ค�ดลอกผู้ลถู�กต�อง@ การต�ดต�อสิ่�มพิ�นธี$ก�บแพิที่ย$ พิยาบาล

Page 13: Iqc Eqa Pt มน

13

Physilogical factor

Pathological factors

Risk factors

Technical factors

Page 14: Iqc Eqa Pt มน

14

Page 15: Iqc Eqa Pt มน

15 Change (%) of the serum conc. of analytes after a standard meal

Page 16: Iqc Eqa Pt มน

16

Sample deteriorationReagentsPipetting systemTemperatureTransducerLight sourceOptical pathLight selectorWashing systemCalibration procedure

Analytical results

FACTORS

Page 17: Iqc Eqa Pt มน

17

การควบค0มค0ณภาพิภายในห�องป่ฏิ�บ�ต�การ

Internal Quality Control, IQCกระบวนการเพิ&�อควบค0มค0ณภาพิภายในห�องป่ฏิ�บ�ต�การให�ได�ผู้ลเป่2นที่��ยอมร�บ

ว�ตถู0ป่ระสิ่งค$ เพิ&�อควบค0มผู้ล Lab. ให�ถู�กต�อง รวดเร�ว ร�กษาสิ่ภาพิการที่(างานให�คงที่�� เป่2นที่��เชื่&�อถู&อ และ

ยอมร�บ

Page 18: Iqc Eqa Pt มน

18

ป่3จจ�ยที่��เก��ยวข้�องPreanalytical step - การเตร�ยมผู้��ป่�วย การเก�บสิ่��งสิ่�งตรวจ การ กระที่(าต�อสิ่��งสิ่�งตรวจ การเตร�ยมสิ่ารละลาย ฯลฯAnalytical step - การเตร�ยมข้�.นตอน สิ่ภาวะการที่ดสิ่อบ การ จ�ด/การเตร�ยม QC material การอ�านค�า ฯลฯPost-analytical step - การค(านวณ ค�ดลอกผู้ล รายงานผู้ล

และการแป่ลผู้ล ฯลฯ

องค$ป่ระกอบข้อง IQC - บ0คลากร - เที่คโนโลย� = ว�ธี�การ สิ่ารเคม� น(.า เคร&�อง ม&อ อ0ป่กรณ$ การจ�ดการ

Page 19: Iqc Eqa Pt มน

19

Steps of IQC Process

@ Optimum condition variance (OCV)

@ Routine condition variance - Known value

(RCV-K)

@ Routine condition variance - Unknown value (RCV-U)

@ Statistical use of patients’ results

@ Other applications : CUSUM plot, Youden

- plot, Westgard multirule, Mean range chart, etc.

Page 20: Iqc Eqa Pt มน

20

IQC for Glucose determination

Page 21: Iqc Eqa Pt มน

21

Routine Condition - Variance Unknown Value

(R -CV U) = ใชื่�ต�วอย�างที่��ก(าหนดค�าแน�นอน

= ม�ต�วอย�างที่��ม�ค�ามากกว�า 1 ระด�บ= ที่(าว�นละ 1 ต�วอย�าง= Plot ค�า %Variation แต�ละว�นโดย

%Variation =Observed value - Corrected value*

Corrected value

*Corrected value = Assigned value = Reference value

x100

Page 22: Iqc Eqa Pt มน

22

Page 23: Iqc Eqa Pt มน

23

Page 24: Iqc Eqa Pt มน

24

Page 25: Iqc Eqa Pt มน

25

949698

100

102104106

1 4 7 1013161922252831343740434649 วั�นที่��

mg / dl

ต�วอย�างกราฟควบค0มค0ณภาพิซ้ำ9�งแสิ่ดงว�าม�Accuracy เป่ล��ยนไป่ โดยที่�� Precision ไม�เป่ล��ยน ( คาดวั�า

เกิ�ดประมาณวั�นที่�� 27)

4

Page 26: Iqc Eqa Pt มน

26

90

95

100

105

1 4 7 1013161922252831343740434649วั�นที่��

mg / dl

ต�วอย�างกราฟควบค0มค0ณภาพิซ้ำ9�งแสิ่ดงว�าม�Precision เป่ล��ยนไป่ โดยที่�� Accuracyไม�เป่ล��ยน ( คาดวั�าเกิ�ด

ประมาณวั�นที่�� 26)

5

Page 27: Iqc Eqa Pt มน

27

More Considerations

@ Critical Value

@ Delta (Prior value) check

@ Cumulative Report

@ Clinical correlation

@ Lab. Data correlation

Page 28: Iqc Eqa Pt มน

28

Limitation in IQC

@ Control material ไม�เหม&อนต�วอย�างผู้��ป่�วย

@ Random error

@ การก(าหนดค�า True value ที่��ถู�กต�อง

@ Bias

@ ไม�สิ่ามารถูควบค0มข้�.นตอนก�อนการว�เคราะห$@ ว�ธี�การที่างสิ่ถู�ต� ถู&อเป่2น Normal distribution

@ สิ่�.นเป่ล&องที่�.งเวลา แรงงานและค�าใชื่�จ�าย

Page 29: Iqc Eqa Pt มน

29

Page 30: Iqc Eqa Pt มน

30

Page 31: Iqc Eqa Pt มน

31

Page 32: Iqc Eqa Pt มน

32

Page 33: Iqc Eqa Pt มน

33

Page 34: Iqc Eqa Pt มน

34

X

X +1SD

+1SD

-1SD

-1SD

-2SD

-2SD-3SD-3SD

+2SD

+3SD

+2SD +3SD

Acceptable range

Random error

Systemic

error

**

*

Page 35: Iqc Eqa Pt มน

35

Cumulative Data (Assigned value X = 10)

X X-X CUSUM 10 0 0 12 +2 +2 8 -2 0 13 +3 +3 11 +1 +4 12 +2 +6 11 +1 +7 9 -1 +6

Page 36: Iqc Eqa Pt มน

36

0 ••

••

• ••

5

10

-5

Page 37: Iqc Eqa Pt มน

37

MEAN - RANGE (X-R) GRAPH

DATE 2/5/95 2/5/95 3/5/95 4/5/95 5/5/95 5/5/95 6/5/95 7/5/95 7/5/95

TIME 0900 1300 0900 0900 0900 0300 0900 0900 1300 X 95, 96 95,97 97,95 95,93 96,96 92,99 97,99 97,97 99,94

96,95 94,96 99,94 95,96 95,94 93,98 96,98 96,92 93,96

Total 382 382 385 379 381 382 390 382 382

Mean 95.5 95.5 96.25 94.75 95.25 95.5 97.5 95.5 95.5

Range 1 3 5 3 2 7 2 5 6

Note

Page 38: Iqc Eqa Pt มน

38

Page 39: Iqc Eqa Pt มน

39

- Critical Values - Delta (Prior-value check) - Cumulative report- Lab. Data correlation- Clinical correlation- Peer Lab. Comparison

- External QA (Proficiency testing) program- etc.

Page 40: Iqc Eqa Pt มน

40

Date Protein Albumin Bilirubin ALP ALT1/10 6.7 3.1 32.8 127 956

3/10 5.4#R 20.6#R 0.6#R 55#R 26#R

4/10 7.2 3.3 39.5 147 1421

6/10 7.1 3.4 35.7 164 1500

8/10 6.8 3.4 31.8 163 1527

# = Delta check failureR = Repeat

Page 41: Iqc Eqa Pt มน

41

CBC : WBC 30,000 /cu.mm. Platelet 1,300,000 /cu.mm.

Serum PlasmaNa 138 139 mmol/L

K 6.4 4.6 mmol/LCl 95 98 mmol/LHCO3 27 25 mmol/LUrea 8.0 8.5 mmol/LCreatinine 0.09 0.09 mmol/L

Previous K 5 determinations were 5.2-6.4 mmol/L with unknown cause

Page 42: Iqc Eqa Pt มน

42

Lab. Results of a man with diabetic ketoacidosis

Na 132 mmol/L(132-144)

K 6.5 mmol/L(3.2-4.8)

Cl 89 mmol/L(98-108)

HCO3 6 mmol/L(22-32)

Urea 14.5 mmol/L(3.0-8.0)

Glucose 64 mmol/L(3.5-5.5)

Blood pH 7.02 (7.35-7.45)

AG 30.5 (7-17)

Page 43: Iqc Eqa Pt มน

43

Total anion = Total cation

Measured anion + Unmeasured anion = Measured cation + Unmeasured cation

AG = Unmeasured anion – Unmeasured cation

Unmeasured anion - Unmeasured cation = Measured cation - Measured anion

AG = Na+ + K+ - (Cl- + HCO-3)

Ref. Value = 7-17 mmol/L

OR AG = Na+ - (Cl- + HCO-3)

Ref. Value = 2-12 mmol/L

Page 44: Iqc Eqa Pt มน

44

1. Staffing pattern facilities and workload

2. Lack of personnel/ understanding/ Equipment

3. Reagent kits/Method used- vary & high cost

4. Too much tests to perform

5. Lack of support from executive personnel

6. Financial constraint, education, supply and service, transport and storage

Page 45: Iqc Eqa Pt มน

45

7. No regulations, reward/punishment Focus only on performance/ test system

8. Focus only on performance/tests system

not on total system

9. Data not shown in graph

10. EQA – prolonged response, no discussion or suggestion, low frequency, discontinuous of the program

11. Clinician understandings ?

12. Equivalent QC, Electronic QC, Eq (Big

error Less quality)

13. Etc.

Page 46: Iqc Eqa Pt มน

46

การตรวจค0ณภาพิสิ่�นค�า การตรวจสิ่อบผู้ล lab. ในห�องป่ฏิ�บ�ต�การ

ความอ�อนล�าข้องผู้��ตรวจสิ่อบค0ณภาพิสิ่�นค�า ความอ�อน ล�า เน&�องจากจ(านวนต�วอย�างมาก

สิ่ภาพิจ�ตใจและร�างกายข้องผู้��ตรวจสิ่อบฯ สิ่ภาพิจ�ตใจ และร�างกายข้องห�วหน�า lab.

ความเบ&�อหน�ายในการที่(างานซ้ำ(.าซ้ำาก ความเบ&�อหน�ายที่��ต�องตรวจงานซ้ำ(.าซ้ำาก

ความกดด�นที่��ต�องผู้ล�ตสิ่�นค�าให�ได�ป่ร�มาณ ม�ผู้��ป่ฏิ�บ�ต� งานน�อย ต�องร�บที่(าให�เสิ่ร�จตาม

และค0ณภาพิตามก(าหนด ก(าหนด ความถู�กต�องข้องเคร&�องม&อและเคร&�องว�ดที่�� ความถู�ก

ต�องข้องเคร&�องม&อที่��ใชื่�ในห�อง lab. เก��ยวข้�อง

ความล(าบากในการตรวจสิ่อบสิ่ภาพิพิร�อม ไม�สิ่ามารถู ตรวจสิ่อบความพิร�อมในการที่(า

ต�างๆ งาน หร&อความถู�กต�องได�ที่0กข้�.นตอน

ความชื่(านาญและป่ระสิ่บการณ$ข้องแต�ละ ข้าดป่ระสิ่บการณ$และความชื่(านาญ บ0คคล

Page 47: Iqc Eqa Pt มน

47

What is external qualityassessment?กิารที่��หน�วัยงานที่��เป�นศู�นย�กิลาง กิ�าหนดให�กิล �มห�อง ปฏิ�บั�ติ�กิาร วั�เคราะห�ติ�วัอย�างเด�ยวักิ�น ส่�งผลกิล�บัค'นมา

ศู�นย�กิลาง จะที่�ากิารวั�เคราะห�ผลที่��ได� เพื่'�อด�วั�า lab ใด ได�ผลกิารติรวัจวั�เคราะห�ใกิล�เค�ยงกิ�บัค�าที่��กิ�าหนด ที่��วั�เคราะห�โดยวั�ธี�อ�างอ�ง หร'อห�องปฏิ�บั�ติ�กิารอ�างอ�ง หร'อค�าที่��วั�เคราะห�ได�อย��ในกิล �ม หร'อแติกิติ�างไปจากิกิล �มมากิน�อยเพื่�ยงใด

Page 48: Iqc Eqa Pt มน

48

What is specimen for EQA program ?

1 .เคม�คล�น�กิ - serum – lyophilized, liquid

2. CBC - synthetic whole blood, picture

3. U/A - picture : crystals, cast, etc4. Microbiology : unknown culture,

picture

Page 49: Iqc Eqa Pt มน

49

Why do laboratories need external qualityassessment?

1. IQC เน�น precision ไม�ส่ามารถบัอกิ accuracy 2. EQA จะบัอกิ accuracy ที่างอ�อม เป�น relative เพื่ราะไม�

ได�ใช้�ส่ารมาติรฐานที่��ที่ราบั true value3 .เป�นกิารประเม�นค ณภาพื่ของห�อง lab ของตินเอง เม'�อเที่�ยบั

กิ�บั lab ในกิล �มเด�ยวักิ�น4. ใช้�อ�างอ�ง กิรณ�ที่��ผ��ป3วัยม�กิารส่�งติ�อไปย�งโรงพื่ยาบัาลในกิล �ม

เด�ยวักิ�น วั�าอาจจะใช้�ผล lab ด�วัยกิ�นได� โดยผ��ป3วัยไม�ติ�องติรวัจใหม�

5. ใช้�ในกิารประช้าส่�มพื่�นธี�ห�อง lab

Page 50: Iqc Eqa Pt มน

50

EQA แส่ดงโดย ด�ช้น�ควัามแปรปรวัน (Variance index, %V) ด�งน�4

%V = (X – mean)/mean * 100

เม&�อ X เป่2นค�าที่��ว�เคราะห$ได�โดย lab

Mean เป่2นค�าที่��ก(าหนด (Designed value)

Variance index (VI) = (V/CCV) * 100

เม&�อ V = % Variation

CCV = Chosen coefficient of variation

Page 51: Iqc Eqa Pt มน

51

Page 52: Iqc Eqa Pt มน

52

แปลผลเป�น

1. Variance Index Score (VIS)ถ�า V = CCV น��นค'อ VIS = 100 ----- ด� VIS < 100 ------ ย��งด�มากิข54น VIS > 100 ------ ควัรเร��มปร�บัปร ง

เม'�อ VI < 400 ------ VIS ม�ค�า = VI ที่��ได�

VI > 400 ------ VIS ม�ค�า = 400

2. Mean Variance Index Score (MVIS)

= ค�าเฉล��ยของ VIS ที่��ได�ส่�าหร�บัที่ กิ test ในคร�4งน�4น

3. Overall Mean Running VIS (OMRVIS)

= ค�าเฉล��ยของ VIS 3 0 test ล�าส่ ด

Page 53: Iqc Eqa Pt มน

53

กิารประมวัล ผลโดยใช้�กิราฟ ใช้�กิราฟ 45o

Page 54: Iqc Eqa Pt มน

54

ว�ธี�การค(านวณเพิ&�อลากเสิ่�น VIS = 100

ถ�าจะให�ได� VI ( ซึ่5�งกิ9ค'อ VIS ) = 100 จะติ�องให� % VARIATION ( V ) =

CCV พื่อด�

VI = x 100

ต�วอย�าง CCV ของ SODIUM = 1.6% กิ9ค'อค�า SODIUM 100 m mol / l ถ�าวั�เคราะห�แล�วัค�าที่��ได�เบั��ยงเบันไปจากิค�าที่��

กิ�าหนด + 1.6 m mol / l หา VIS แล�วัจะได� = 100X = 100 , X = 101.6 V = X - X x 100 = 101.6 - 100 x 100 = 1.6 X 100 VI = V x 100 = 1.6 x 100 = 100 VIS = 100 CCV 1.6

V CCV

30

Page 55: Iqc Eqa Pt มน

55

ด�งน�4นถ�าจะด�วั�าหากิค�า DESIGNATED VALUE = 150 mmol/LlVIS 100 จะม�ค�าเที่�าใด ? ค�านวัณได�

ด�งน�4 : 100 = + 1.6 150 = + 1.6 x 150 = + 2.4 mmol/L 100

เช้�นกิ�นให�ค�านวัณที่��ค�าติ��าๆ เช้�นที่�� 130mmol/L อ�กิช้ ดหน5�ง กิ9ส่ามารถลากิเส่�น VIS = 100 ด�านบัน & ด�านล�าง ของ เส่�น 45 องศูา ได�

น��นค'อ VIS 100 จะอย��ที่�� 150 + 2.4 = 152.4 mmol/L และ 150 2.4 = 147.6 mmol/Ll

31

Page 56: Iqc Eqa Pt มน

56120

130

140

150

160

170Your Result ( A ) SODIUM mmol / l

VIS 100VIS 100

VIS 0

120 130 140 150 160 Designated Value ( B ) mmol / l

127.9

132.1

147.6

152.4

VIS เก�น 100

VIS เก�น 100

( 1 )

( 1 )

( 2 )

( 2 )

( 2 )

( 2 )

( 2 )

( 3 ) ( 3 )

( 3 )

( 3 )

( 3 )

( 3 )

CCV = 1.6 %

32

Page 57: Iqc Eqa Pt มน

57120

130

140

150

160

170Your Result ( A ) SODIUM mmol / l

VIS 100VIS 100

VIS 0

120 130 140 150 160 Designated Value ( B ) mmol / l

* 3

* 1

* 2

* 6* 4

* 5

33

Page 58: Iqc Eqa Pt มน

58120

130

140

150

160

170Your Result ( A ) SODIUM mmol / l

VIS 100VIS 100

VIS 0

120 130 140 150 160 Designated Value ( B ) mmol / l

* 3* 1

* 5

* 4* 2

* 6

34

1

Page 59: Iqc Eqa Pt มน

59120

130

140

150

160

170Your Result ( A ) SODIUM mmol / l

VIS 100VIS 100

VIS 0

120 130 140 150 160 Designated Value ( B ) mmol / l

* 1

* 2

* 3

* 4

* 5* 6

35

Page 60: Iqc Eqa Pt มน

60

20

70

120

170

220

270

320

370Your Result CHOLESTEROL

( mg / dl )

VIS 100VIS 0VIS 100

20 70 120 170 220 270 320

Designated Value ( mg / dl )

1 *

2 *

5 * 4**6

3 *

36

Page 61: Iqc Eqa Pt มน

61

Page 62: Iqc Eqa Pt มน

62

Page 63: Iqc Eqa Pt มน

63

Page 64: Iqc Eqa Pt มน

64

External Quality Assurance (EQA) โดยกรมว�ที่ยาศาสิ่ตร$การแพิที่ย$ ที่0ก 3 เด&อน

25452547

25492551

25452547

25492551

25452547

25492551

25452547

25492551

EQC: VIS ต(�ากว�า 150 ถู&อว�า Acceptable VIS ต(�ากว�า 50 ถู&อว�า อย��ในเกณฑ์$ด�มาก

Page 65: Iqc Eqa Pt มน

65

กิารที่ดส่อบัควัามช้�านาญ เป�นเที่คน�คหน5�งของกิารประกิ�นค ณภาพื่ส่�าหร�บัห�องปฏิ�บั�ติ�กิาร

เป�นติ�วัติ�ดส่�นกิารกิระที่�าของแติ�ละห�องปฏิ�บั�ติ�กิาร ในกิารที่ดส่อบัที่��จ�าเพื่าะ หร'อกิารวั�ด และใช้�ติรวัจส่อบัห�องปฏิ�บั�ติ�กิารที่��ม�กิารปฏิ�บั�ติ�อย�างติ�อเน'�อง

หน�วัยงานที่��ร �บัผ�ดช้อบั (PT provider) ที่�ากิารแจกิจ�ายติ�วัอย�างให�กิ�บัห�องปฏิ�บั�ติ�กิาร เพื่'�อที่�ากิารติรวัจวั�เคราะห�

กิารประเม�นควัามส่ามารถของห�องปฏิ�บั�ติ�กิาร ส่ถ�ติ�ที่��ใช้�ค'อ Robust Z-Score ที่��ค�านวัณจากิค�าเฉล��ย Robust และค�าส่�วันเบั��ยงเบันมาติรฐาน Robust

Page 66: Iqc Eqa Pt มน

66

ค�านวัณหาอ�ติราส่�วันของกิารเบั��ยงเบันระหวั�างค�าอ�างอ�ง และค�าของห�องปฏิ�บั�ติ�กิารติ�อค�าเบั��ยงเบันมาติรฐานผลกิารเข�าร�วัมที่�4งหมด โดย

Z = (x – X)/s

โดยที่�� x = ผลของห�องปฏิ�บั�ติ�กิารที่��เข�าร�วัม

X = ค�าที่��กิ�าหนดหร'อค�าอ�างอ�ง

s = ค�าเบั��ยงเบันจากิกิารที่ดส่อบั (ได�แกิ�ค�าเบั��ยงเบัน มาติรฐานของผ��เข�าร�วัมโปรแกิรม เป�นติ�น)

Page 67: Iqc Eqa Pt มน

67

Z = + 2 แสิ่ดงว�าผู้ลการที่ดสิ่อบน�.นยอมร�บได� (Satatisfactory)

Z = < + 3 แสิ่ดงว�าผู้ลการที่ดลองน�.นน�าสิ่งสิ่�ย (Questionable) Z = > + 3 แสิ่ดงว�าผู้ลการที่ดลองน�.นยอมร�บไม�ได� (Unsatatisfactory)

Page 68: Iqc Eqa Pt มน

68

Proficiency test results are reported as standard deviation indexes (SDIs ). SDI is calculated by the following formula:

SDI = (Result – Mean)/ sd

Page 69: Iqc Eqa Pt มน

69

Page 70: Iqc Eqa Pt มน

Quality is a Journey, not a Destination

Q O

L O I O

GO

Page 71: Iqc Eqa Pt มน

Quality is a Journey, not a Destination