Optical fiber
35
cm
Effective area: 128 x 128 mm2
35 cm
BL17
BL03
BL19
1-d large areascintillator detector
MLFビームラインにおける検出器の現状
He-3 PSD
2-d large area scintillator detectorusing WLSF
2-d compactscintillator detectorusing WLSF
BL18
BL01, BL02, BL08, BL09, BL11, BL12BL14, BL15, BL16, BL17, BL20, BL21
MWPC
He-3 detectors(commercially available)
Scintillator detectors
BL16, BL17RPMT
Multi-wire proportional counter
MLFにおけるシンチレーション検出器
2-D detector using WLSF
BL18(Senju)Single crystal diffractometer(2010 - 2013) BL03(iBIX)
Bio-single crystal diffractometer( 2007 – 2008, 2011 - )
BL19(Takumi) Residual stress diffractometer
(2007 – 2009)
Linear scintillation detector
2-D detector using WLSF
BL19における1次元シンチレーション検出器
Neutron
TAKUMI: Engineering Materials Diffractometer for Residual Stress Analysis
Detectors(each 5 installed at 2q=90 degree position)
Number of detectors: 10
Specifications of Detectors
The detector should have Sensitive area: 1000 mm x 200 mm Position resolution: 3mm 360 pixels / detector Detection efficiency: >50%@1Å Gamma sensitivity: <[email protected]
Sample
Developed under collaboration with ISIS, RAL
BL19における1次元シンチレーション検出器
Sensitive area:1m×20cm
Radius curvature:R=2m
Number of PMT:48/detector
Number of Pixels: 360
Resolution:3mm
Neutrons217mm
1370mm805mm
Detector size without PMT:805×1370×217mm
Detector size with PMT: 805×~2000×217mm
weight:~120kg(without PMT)weight:~180kg(with PMT)
Scintillation Detector using ZnS/LiF
Senstive Area(1m×20cm)
BL19検出器のヘッドの構造
Neutrons
Venetian Head Detection of scintillation light
from both sides of scintillator
Increase of detection efficiency
using scintillators tilted to neutron
direction(Effective thickness increased)
Scintillators
(ZnS/6LiF)
Reflectors
resolution: 3mm
Fibers
ReflectorsScintillators
(ZnS/6LiF)
Scintillators
6Li + n→3H + a+ 4.8 MeV
Neutron
B4C Block
Scintillator & grids
The scintillation light was collected with optical fibers and sent to PMT and Data acquisition system
PMT & DAQ
No.3
0
200
400
600
800
1000
1200
1400
0 50 100 150 200 250 300 350
BL19検出器の性能
No.4
0
200
400
600
800
1000
1200
1400
0 50 100 150 200 250 300 350
channel number
Counts
/100s
4.7%@std.devGamma sensitivity <10-6
with a 60Co source(at discri. level of 250mV)
10-8
10-7
10-6
10-5
10-4
100 150 200 250 300 350
Detector 1Detector 2
Ga
mm
a s
ensitiv
ity
Threshold level (-mV)
Gamma sensitivityUniformity measured with PE rod
5.5%@std.dev
BL19TAKUMI検出器
• 3mm position resolution
• Detection efficiency: >50%@1Å
• Gamma sensitivity: <[email protected]
• 3×196mm detector element
• 360ch/detector
• Total 10 detector
• Total 3600 elements
• 2.0m2 active area
• 75°<2q<105°
Total 10 detectors were
installed and have been in
service for several years
BL03における2次元シンチレーション検出器
• iBIX : Single crystal diffractometer for bio-molecular crystallography• A first WLS-fiber-based scintillator detector developed at the J-PARC.
Detection efficiency : > 50% (1.8Å)Spatial resolution : < 1 mmGamma sensitivity : < 10-6
Packing density : ~70%Compact
Specifications required to detector
High speed amplifierdiscriminator module
256 x 256 chimaging detectorwith WLS readout
Signal processingmodule with FPGAs
Prototype detector (in 2008)
iBIX検出器のヘッド構造
High detection efficiency
neutron
Neutron-sensitive scintillator
neutrons
XY
Conventional structure iBIX structure
Single Scintillator screen Double scintillator screens
WLS fibres
(2) Packed ½-mm diameter WLS fiber ribbons High spatial resolution
(1) Developed a ZnS/10B2O3 scintillator and double the number of screens
iBIX検出器の位置分解能
0
0.2
0.4
0.6
0.8
1
1.2
1.4
0 20 40 60 80 100
Rela
tive inte
nsi
ty
Position (x axis) [mm]
exprimental data
fitted cumulative distribution function
Spatial resolution (FWHM)= SQRT(2log2) σ~ 2.3548 σ= 1.19 [mm], including beam divergence
Edge Spread function:
A = 0.5 → 0.4483
B = 0.1 → 0.09343μ = 50 → 50.41
σ = 0.5 → 0.5082 with Gd mask
without Gd mask
Fitting parameters
0 2550
255
Y
X
Gd mask
BL18(Senju)の2次元シンチレータ検出器
* SENJU : A time-of-flight Laue single crystal diffractometer * A large area position-sensitive scintillator detector
using WLS fiber technology
Spatial resolution : ~4 mm
Detection area : ~30 x 30 cm2
Detection efficiency : >40%@1.8Å
Gamma-ray sensitivity : < 10-6
Specifications required to detector
Detector head
Signal processing
& Encoder electronics
DAQ
electronics
T. Nakamura, et al. NIMA686(2012)64.
BL18検出器のヘッド構造
In the design, to increase a detector coverage at an affordable costthe WLS fibers are placed at a regular pitch of 4 mm.
(1) WLS fiber placed at a 4-mm pitch Large area
(2) Two scintillator screens Detection efficiency
(3) Bend clear fiber backward Minimize dead area
Scintillator (front)
①WLS
fibers
③ Clear fibers
neutron
②ZnS/10B2O3
scintillators
Key Designs
0.5
0.6
0.7
0.8
0.9
1.0
1.1
1.2
0 100 200 300 400 500
rela
tive
co
un
ts
elapsed time, days
Det12-01(SCDD01_DS-01)
BL18検出器の性能
0
10
20
30
40
50
60
1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 39
dete
ctio
n ef
ficie
ncy,
%
detector number
production date: 2010.10 2013.1
Detector efficiency of 37+2 detectors
Count stability over time
Detectors were stableafter >400 days
0
0.2
0.4
0.6
0.8
1
1.2
20 25 30 35 40
position. mm
inte
nsity,
a.u
.
Image plate
SCD detector
Pencil beam ( ~3 mm in width, 4 Å)
CollimatorCollimator
Spatial resolution : 4 mm
Detection efficiency
~40%@1.8Å
Gamma-ray sensitivity
Mass production technique is acceptable.
Gamma sensitivity
vs Detection efficiency of 30 detectors
Earthquake
波長シフトファイバを用いたシンチレーション検出器
High spatial resolution detector
pixel size : 0.5 x 0.5 mm
detective area : 133 x 133 mm2
detector efficiency: ~50% for 1.8Å
gamma sensitivity: ~1 x 10-6
信号処理&データ収集系
BL18(SENJU )
Large area detector
pixel size : 4 x 4 mm
detective area : 256 x 256 mm2
detector efficiency: ~40% for 1.8 Å
gamma sensitivity: ~ 3 x 10-6
BL03(iBIX)
300 x 300 mm
Total 30 detectors were installed
and in service nowTotal 37 detectors were installed
and in service now
BL17におけるマルチワイヤ型検出器
Optical fiber
35
cm
Effective area: 128 x 128 mm2
35 cm
MWPC systemInstalled in MLF/BL17
Pressure vessel withstanding up to 8 atm
Multi-wire type detector elementWire pitch: 1 mmSensitive area: 128 x 128 mm2
ASD-ASIC for multi-channel signal processing
Individual line readout
Optical signal transmission
Sharaku: Polarized Neutron Reflectometer
開発したMWPCの性能評価
Irradiation experiments were performed using Cf-252at KUR/CN3at J-PARC/MLF BL17
MWPC MWPC
0
200
400
600
800
1000
10 15 20 25 30 35 40
Measu
red s
ignal
(counts
)
Position (mm)
X projection1.8 mm FWHM
0
2x104
4x104
6x104
8x104
1x105
1.2x105
0 20 40 60 80 100 120
Co
un
ts
Position (mm)
Standard deviation
1st. cathode: 3.8%
2nd. cathode: 4.6%
10-5
10-4
10-3
10-2
10-1
100
101
102
103
10-6
10-5
10-4
10-3
10-2
10-1
100
101
102
-500 0 500 1000
Measu
red c
ounts
(kcps) R
eacto
r pow
er (M
W)
Elapsed time (s)
● Measured by developed detector
■ Reactor monitor
10-1
100
101
102
103
104
105
10-5 10-4 10-3 10-2 10-1 100 101
Cou
nts
(cp
s)
Reactor Power (MW)
Position resolution Counting uniformity Output Linearity
beam size: 0.3 x 0.3 mm2
1D uniformityX: 3.5%Y: 4.6% > 5 decades!
With Cf-252 At KUR
Good linearity
Performances of MWPC