Mass spectrometry for the chromatographer

Page 2: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

LC/MS in Drug Development

• Identify unknown peaks to support method development and validation

• Characterize degradation products in forced degradation studies and stability samples

• Elucidate drug degradation pathway to support excipient selection and formulation development

• Confirm peak identity for DS and related substances

• Support QA issues and NDA

• Support Manufacturing process and finished drug products

Page 3: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Features of LC/MS

Advantages:

• High sensitivity

• High selectivity

• Speed

• Combined power of separation and identification

• Molecular weight and structure information

Disadvantages:

• Some compounds are not ionizable

• Difficult to identify complete unknowns

• Limited structure information compared to NMR

Page 4: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Identification Strategy:When you start LC/MS analysis…..

• Is the method MS compatible?

• Conditions where degradation compounds form

• Synthetic scheme (Byproduct? Intermediate?)

• Structure of drug substanceactive sites and functional groups

Number of nitrogen atoms, Isotope ratio

• Full MS spectrum of unknown: Which is [M+H]?

• MS/MS spectra of DS and unknown

• UV spectra of DS and unknown

Page 5: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

LC/MS Analysis

Work Request

Confirm Chromatogram Modify HPLC conditions

MS/MS(n) StudyDetermine MW

Proposed Structure

Show Peak Equivalency

LC/MS compatible method LC/MS incompatible method

Isolation/Purification

NMR Confirmation

Select Ionization ModeESI vs APCI , (+) vs (-)

Confirmation ?No YesReport In-house Synthesis

Page 6: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

LC-MS

LC – Separation of the mixture of analytesInterface – Separation of the analyte from the solventMA (mass analyzer) – separation of the analyte molecular ion and fragments according to their mass to charge ratio

Interface MassAnalyzer DetectorLC

Extraction ofThe analyte fromThe solvent

Ion evaporationor ionization.Fragmentation

Page 7: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Infusion MS

LCpump

Sample solution

• Sample in continuous flow

• No sample injection

• No LC separation

• MS spectrum obtained over an given time

• Improved S/N due to the average of multiple spectra

• Select LC flow to assist ionization

1 μl/min

100 μl/min

MicroTee MS

Page 8: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

10-5 Torr

Richard B. Cole, Electrospray Ionization Mass Spectrometry, Fundamentals Instrumentation and Applications, Wiley Interscience, 1997

1 Torr10-2 –10-4 Torr

Formation of gas phase ions from solution phase

760 Torr

Ion optics

10-5 Torr

N

Basic Electrospray Schematic

Page 9: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

ELECTROSPRAY

Factors to consider

• Ionic strength • Surface tension of the solvent• Volatility of the solvent• Character of the analyte ions in solution: solvated, ion paired,

etc• Mobile phase composition and amount of water• pH of the mobile phase

Page 10: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Confirmation of molecular weightIn

tens

ity

0.0

2.0x106

4.0x106

6.0x106

8.0x106

1.0x107

1.2x107

1.4x107

1.6x107

1.8x107

m/z100.00 200.00 300.00 400.00 500.00 600.00 700.00 800.00

294.05

Inte

nsity

0

1x106

2x106

3x106

4x106

5x106

6x106

7x106

m/z100.00 200.00 300.00 400.00 500.00 600.00 700.00 800.00

292.04

ESI (+) ESI (-)

Page 11: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

1. Effect of instrumental parameters on ESI response

•Mobile phase flowrate*

•Capillary*/corona voltage

•Fragementor*/cone voltage

All play important roles in analyte ions formation and ionization efficiency

Page 12: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Effect of flow rate on ESI response

Increasing the flow rate increases droplet size which decreases the yield of gas-phase ions from the charged droplets.- Use shorter columns of same phase and proportionally decreaseflow rate.

NCH3

O

O

O

O

CH3

Dependence of mass analyzed ion intensity of the BH+ ion of protonated cocaine (10-5M) on flow rate.

P.Kebarle, et.al, Anal.Chem. 1993, 65, 972A

Page 13: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Formation of small micron sized droplets is not a problem if flow rate, surface tension are low. An increaseof these may make it difficult to for the electric field to produce the desired charged aerosol. The electric fieldstrength can be increased to try to overcome those effects. However if electric field is too high will giverise to electrical discharge and this is detrimental to ES signal. Nebulizing gas can be used to help focus the #of ions transported into vacuum envelope of the mass spectrometer.

Effect of capillary voltage

G.Valaskovic,J.Murphy, M.Lee,Milestone development

Page 14: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

FB 60CV+1

100 200 300 400 500 600 700 800 900 1000m/z0

100

%

TAE 226 FB-2 805 (8.205) 1: Scan ES+ 5.13e7438.16

403.22

372.20

193.08117.5399.94

359.54315.61

469.18

471.08

937.57491.09534.26635.18 720.40

913.51757.50 870.45 959.60

FB- 25CV +1

100 200 300 400 500 600 700 800 900 1000m/z0

100

%

TAE226 FB-1 807 (8.225) 1: Scan ES+ 4.86e7469.18

242.19214.09155.62100.19

255.11315.23

348.40 434.62

471.08

937.44

506.91 696.09577.06 652.27721.16

813.47 831.58891.09 941.49

976.06

25VSingle Quad

60VSingle Quad

DSDS

3:1

3:1

3:1

As potential difference (deltaV) between end of capillary (N) and first skimmer (S) is increased the ions are accelerated through the background gas, which leads to more collisions.

Effect of Cone/Fragmentor Voltage on fragmentation

Page 15: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

A. Kamel, P.R. Brown, B. Munson, Analytical Chemistry, vol.71, 968-977, 1999

Effect of experimental parameters on ESI response

A. Dependence of ion intensities on analyte concentration

Concentration effect of tetracycline on the signal intensity

Page 16: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

A B C

k1

ko

pH

k

pKa

B Cko

k1

pH

k

pKa

1 2 3 4 5 6 70

Basic compound: pKa=6 Acidic compound: pKa=3

pH = analyte pKa, Analyte is 50% ionized

B. Role of solution equilibria and solution-phase chemistry

Cs. Horvath, W. Melander, I. Molnar, Anal.Chem. 49 (1977) 142.

Page 17: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Effect of Solution pH (ESI)

•Variation of solution pH changes the acid-base equilibrium of an analyte species•Alters the degree of positive/negative charging via protonation/deprotonation•Should get increased intensity with decreasing pH for basic compounds in positive ion mode•Should get increased intensity with increasing pH for acidic compounds in negative ion mode•However ion-pairing, ion-association may lead to ion-suppression actually causing a decrease in ionization efficiency•There is a critical point at which increase in protonation ceases to increase ionization efficiency for protonated bases. •Effect of organic on the shift of ionization equilibria in the mobile phase.

Page 18: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Desired pH values were achieved using different combinationsof 0.1% formic acid and 0.1% ammonium hydroxide at constant organic composition. Flow injection analysis

0

100

200

300

400

500

600

2 4 6 8

pH

Ibup

rofe

n [M

-H]- s/

n

Effect of pH on [M-H]- negative-ion response of Ibuprofen.

Y.F. Cheng, Rapid Comm in Mass Spec., 2001; 15; 141-151

Effect of eluent pH

Page 19: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

0.00E+00

8.00E+05

1.60E+06

2.40E+06

3.20E+06

4.00E+06

0 20 40 60 80 100

Pe rce nt Ace tonitrile

ESI

Eff

icie

ncy

ImiDes iProClom

Full Scan ESI Efficiency of Antidepressants vs Concentration of Acetonitrile in the mobile phase

Mobile Phase: Aqueous//MeCNAqueous = 1 mMol Ammonium Formate, pH 4.32 with Formic Acid

Analyte Conc: 100 nM/mLInfusion Rate: 20 uL/min

Page 20: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Effect of type of acidic modifier on electrospray response

•Analyte ion intensity affected by the presence and concentration of other electrolytes

1) Depends on the propensity of ion-association which can lead to formation of neutral species

which will lead to ion-suppression causing a decrease in ionization efficiency:

CCl3COO->CF3COO->>CH3COO-~Cl- (Mirza and Chiat, Anal.Chem. 1994, 66, 2898-2904)

2) May depend on volatility of acid (bp of acid) Propionic > Acetic > Formic > TFA

Pindolol

0

5000000

10000000

15000000

20000000

25000000

0 20 40 60 80 100

% MeCN

MS

Inte

nsity

Pindolol pKa 8.8, TFA

Pindolol pKa 8.8/AA

Pindolol pKa 8.8/propionic

NH

ONH

OH

CH3

CH3

Page 21: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

The electrospray ion intensity can be enhanced by:

•Concentration of analyte and instrumental parameters

•Solution phase pH

•Type and concentration of acidic modifier or volatile buffer

•Type and concentration of organic modifier

•Gas phase Reactions are important, feasibility depends on

proton affinities of analyte versus modifier in eluent

•ESI is a complex process comprises of many different variables,

no set of parameters will guarantee large signal intensity

Conclusions

Page 22: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

MS Structure elucidation Basic principle

Nitrogen Rule

Isotope ratio (Cl, Br, etc.)

Adduct ions

LC/MS

Post column addition (acid or base)

MS/MS or CID : from fragments to substructures

Hydrogen-Deuterium exchange

Utilization of UV spectra to support MS identification

Other techniques for structure elucidation Exact MS : Elemental composition & chemical formulaGC/MS: EI mass spectrumNMR or LC/NMR : Detail structure

Page 23: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Calculating Molecular Weight

Three ways to calculate molecular weight

Average mass:

Average atomic weight for each element, (Merck index, DS profile)

12.01115*60+1.00797*122+14.0067*20+…………..=1443.8857

Monoisotopic (exact) mass:

Exact mass of the most abundant isotope of each element

12.0000*60+1.0078*122+14.0031*20+……………..=1442.8796

Nominal mass:

Integer nominal mass of the most abundant isotope of each element

12*60+1*122+14*20+………………………………….=1442

Mass defect 1444:1443:1442. This is as good as being a mile off!

ALWAYS use the monoisotopic mass when calculating the molecular weight in mass spectrometry!

Page 24: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Nitrogen Rule

• Compounds with an odd number of nitrogen atoms have an odd molecular weight (1,3,5,7…)

• Compounds with an even number of nitrogen atoms have an even molecular weight (0, 2,4,6…)

Page 25: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Positive and Negative Adduct Ions

Positive ion detection• [M+H]+ M+1• [M+NH4]+ M+18• [M+Na]+ M+23• [M+K]+ M+39• [2M+H]+ 2M+1• [2M+NH4]+ 2M+18• [2M+Na]+ 2M+23Negative ion detection• [M-H]- M-1• [M-H+HOAC]- M+59• [M-H+TFA]- M+113MeCN= +41

MeOH= +32

Page 26: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

NH

O

O

OO

O OO

O

O

OHO

O

C H 3

O H

O H

C H 3

C H 3

CH 3

C H 3

CH 3

H

ESI full scan mass spectrum of paclitaxel with mobile phase that contains 2mM ammonium acetate and acetonitrileKerns, et.al, 1994, American Chemical Society

Distinct molecular ionpattern can be usedto determine the molecularweight.

Page 27: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Relative Isotope Abundance of Common Elements

98.081Br10079BrBromine

32.537Cl10035ClChlorine

4.4034S0.7833S10032SSulfur

0.2018O0.0417O10016OOxygen

0.3815N10014NNitrogen

0.0162H1001HHydrogen

1.1113C10012CCarbon

RELATIVE ABUNDANCE

ISOTOPERELATIVE ABUNDANCE

ISOTOPERELATIVE ABUNDANCE

ISOTOPEELEMENTS

Page 28: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Isotope ratios of Chlorine and Bromine

Cl

100

32.6

0

20

40

60

80

100

120

M M+2

Cl2

100

65.3

10.6

0

20

40

60

80

100

120

M M+2 M+4

Br100 97.9

0

20

40

60

80

100

M M+2

Br2

100

195

95.5

0

50

100

150

200

250

M M+2 M+4

Page 29: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

P.R.Tiller, et.al, J.Chrom.A, 794 (1998) 15 - 25

Page 30: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

P.R.Tiller, et.al, J.Chrom.A, 794 (1998) 15 - 25

: Hydroxylation

Page 31: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

N

O O

CF3

CF3

FOHC

MW = 503.4

+ H O+

HH

N

O O

CF3

CF3

FC

MW = 521.1

H

OHHO

Acidic conditions

Using LCMS to determine artifact peaks formed on-column

Page 32: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

2 4 6 8 10 12 14 16 18

2000000

4000000

6000000

8000000

10000000

2 4 6 8 10 12 14 16 180

200000

400000

600000

800000

2 4 6 8 10 12 14 16 180

10000002000000300000040000005000000

m/z = 522, XIC

m/z = 504, XIC

TIC

Time (min.)

Inte

nsity

Inte

nsity

Inte

nsity

TIC vs. EIC

Page 33: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Exact Mass- Q-TOF

Trends in Analytical Chemistry, Vol. 22, No. 10, 2003, Imma Ferrer, et.al.

Page 34: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Post column addition

LC

1% Formic acid in IPA

• Sample injection

• LC separation

• No impact on LC chromatographic pattern

• Improve ionization with either acid or baseaddition after LC column

• Select appropriate solution to assist ionization

waste

MicroTee

1 ml/min

0.1 ml/min

MS

Page 35: Mass spectrometry for the chromatographerhplc.chem.shu.edu/NEW/Graduate/Spec.topics.2009/Lect 5 LC-MS-I... · Mass spectrometry for the chromatographer. ... is increased the ions

Nitrogen RuleThis rule is derived from the fact that, perhaps coincidentally, for the most common chemical elements in neutral organic compounds (hydrogen, carbon, nitrogen, oxygen, silicon, phosphorus, sulfur, and the halogens), elements with even numbered nominal masses form even numbers of covalent bonds, while elements with odd numbered nominal masses form odd numbers of covalent bonds, with the exception of nitrogen, which has a nominal (or integer) mass of 14, but has a valencyof 3.It should be noted that the nitrogen rule is only true for neutral structures in which all of the atoms in the molecule have a number of covalent bonds equal to their standard valency (counting each sigma bond and pi bond as a separate covalent bond for the purposes of the calculation). Therefore, the rule is typically only applied to the molecular ion signal in the mass spectrum.Mass spectrometry generally operates by measuring the mass of ions. If the measured ion is generated by creating or breaking a single covalent bond (such as protonating an amine to form an ammonium center or removing a hydride from a molecule to leave a positively charged ion) then the nitrogen rule becomes reversed (odd numbered masses indicate even numbers of nitrogens and vice versa). However, for each consecutive covalent bond that is broken or formed, the nitrogen rule again reverses.Therefore, a more rigorous definition of the nitrogen rule for organic compounds containing exclusively hydrogen, carbon, nitrogen, oxygen, silicon, phosphorus, sulfur, and the halogens would be as follows:An even nominal mass indicates that a net even number of covalent bonds have been broken or formed and an even number of nitrogen atoms are present, or that a net odd number of covalent bonds have been broken or formed and an odd number of nitrogen atoms are present. An odd nominal mass indicates that a net even number of covalent bonds have been broken or formed and an odd number of nitrogen atoms are present, or that a net odd number of covalent bonds have been broken or formed and an even number of nitrogen atoms are present.