104
221=192=010 POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K

POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

  • Upload
    ngotruc

  • View
    241

  • Download
    2

Embed Size (px)

Citation preview

Page 1: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

221=192=010

POLYMETRON Model 9210Silica Analyzer

USER MANUAL

May 2009, Revision K

Page 2: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs
Page 3: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

1

Table of Contents

Section 1 General Information ......................................................................................................... 51.1 Disclaimer .................................................................................................................................... 51.2 Safety information ........................................................................................................................ 5

1.2.1 Use of hazard information ................................................................................................... 51.2.2 Safety recommendations .................................................................................................... 51.2.3 Service and repairs ............................................................................................................. 61.2.4 Precautionary labels............................................................................................................ 6

1.3 Product recycling information....................................................................................................... 71.4 Product disposal .......................................................................................................................... 91.5 Restriction of hazardous substances (RoHS) ............................................................................ 10

Section 2 Specifications.................................................................................................................. 112.1 Technical specifications ............................................................................................................. 11

Section 3 Analyzer Overview ......................................................................................................... 133.1 Operation ................................................................................................................................... 133.2 Calibration.................................................................................................................................. 143.3 Analyzer outputs ........................................................................................................................ 143.4 Maintenance .............................................................................................................................. 143.5 Presentation............................................................................................................................... 15

3.5.1 Front panel ........................................................................................................................ 153.5.2 Rear panel......................................................................................................................... 16

Section 4 Installation........................................................................................................................ 174.1 Analyzer inspection and unpacking ........................................................................................... 174.2 Analyzer preparation.................................................................................................................. 174.3 Installation checklist ................................................................................................................... 174.4 Mounting the analyzer................................................................................................................ 18

4.4.1 Panel version .................................................................................................................... 184.4.2 Cabinet version ................................................................................................................. 19

4.5 Installing the canister holder ...................................................................................................... 204.6 Connecting the sample .............................................................................................................. 214.7 Connecting the drain tube.......................................................................................................... 224.8 Mains power connection ............................................................................................................ 224.9 External communications connection ........................................................................................ 264.10 Input/Output connections ......................................................................................................... 274.11 Reagent preparation ................................................................................................................ 29

4.11.1 Reagent 1M - Molybdate (2 liters)................................................................................... 294.11.2 Reagent 1A - Nitric acid (2 liters) .................................................................................... 294.11.3 Reagent 2 - Oxalic acid................................................................................................... 294.11.4 Reagent 3 - Reducing reagent ........................................................................................ 304.11.5 Calibration solution.......................................................................................................... 30

4.12 Connecting the canisters ......................................................................................................... 314.13 Analyzer startup ....................................................................................................................... 32

4.13.1 Reagents volume declaration.......................................................................................... 324.13.2 Flow rate adjustment....................................................................................................... 324.13.3 System and user setup procedures ................................................................................ 32

Section 5 Operating Instructions .................................................................................................. 335.1 Data Entry .................................................................................................................................. 33

5.1.1 Function Keys ................................................................................................................... 335.1.2 Modification of a value ...................................................................................................... 33

5.2 Measurement screens ............................................................................................................... 335.2.1 Main screen....................................................................................................................... 335.2.2 Display screen 2 - Measurement history........................................................................... 345.2.3 Display screen 3 - Alarms ................................................................................................. 345.2.4 Display screen 4 - Graph .................................................................................................. 34

5.3 Main menu ................................................................................................................................. 35

Page 4: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

2

Table of Contents

5.4 Analyzer menu overview ............................................................................................................36Section 6 System Setup...................................................................................................................37

6.1 Menu overview ...........................................................................................................................376.1.1 Time ..................................................................................................................................376.1.2 Display...............................................................................................................................386.1.3 Code..................................................................................................................................386.1.4 Soft issues.........................................................................................................................396.1.5 Default values....................................................................................................................396.1.6 mA adjustments.................................................................................................................396.1.7 Factory ..............................................................................................................................40

Section 7 User Setup ........................................................................................................................417.1 Measure menu overview ............................................................................................................41

7.1.1 Measuring mode................................................................................................................417.1.2 Grab sample......................................................................................................................427.1.3 Historic ..............................................................................................................................427.1.4 ADC values .......................................................................................................................43

7.2 Alarms menu overview ...............................................................................................................437.2.1 Alarms 1 to 6 .....................................................................................................................447.2.2 Warning alarm ...................................................................................................................457.2.3 System alarm ....................................................................................................................457.2.4 Alarm conditions................................................................................................................46

7.3 mA outputs menu overview........................................................................................................477.3.1 Affect .................................................................................................................................477.3.2 Special programming ........................................................................................................487.3.3 Test ...................................................................................................................................49

7.4 RS485 ........................................................................................................................................497.5 Sequence ...................................................................................................................................50

7.5.1 Channel activation.............................................................................................................507.5.2 Sequence ..........................................................................................................................50

Section 8 Calibration ........................................................................................................................518.1 Menu overview ...........................................................................................................................51

8.1.1 Programming (automatic calibration) ................................................................................528.1.2 Execution primary calibration ............................................................................................538.1.3 Execution manual calibration ............................................................................................538.1.4 Parameters........................................................................................................................548.1.5 Historic ..............................................................................................................................548.1.6 Calibration results..............................................................................................................55

Section 9 Maintenance.....................................................................................................................579.1 General.......................................................................................................................................579.2 Cleaning overview ......................................................................................................................57

9.2.1 General cleaning process..................................................................................................579.2.2 Chemical cleaning procedure............................................................................................58

9.3 Battery replacement ...................................................................................................................599.4 Fuse replacement.......................................................................................................................599.5 Yearly maintenance....................................................................................................................609.6 Maintenance menu overview......................................................................................................639.7 Startup........................................................................................................................................649.8 Reagents refill ............................................................................................................................649.9 Extended stop ............................................................................................................................659.10 Tube cleaning...........................................................................................................................659.11 Photometer test ........................................................................................................................669.12 Pumps - others .........................................................................................................................669.13 Pumps flowrate.........................................................................................................................66

Page 5: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

3

Table of Contents

9.14 Solenoid valves........................................................................................................................ 679.15 Relays ...................................................................................................................................... 679.16 Logical inputs ........................................................................................................................... 679.17 Preparation of the tubes for insertion in the fast connectors.................................................... 68

Section 10 Troubleshooting ........................................................................................................... 6910.1 Possible sources of the problem.............................................................................................. 6910.2 Possible solutions .................................................................................................................... 6910.3 Typical measurement values during calibration....................................................................... 70

Section 11 Spare Parts..................................................................................................................... 71Section 12 Default Configuration .................................................................................................. 73Section 13 Material Safety Data Sheets (MSDS)........................................................................ 77

13.1 Oxalic acid ............................................................................................................................... 7713.2 Sulfuric acid (98%)................................................................................................................... 7813.3 Nitric acid (50-70%) ................................................................................................................. 8013.4 Ammonium hydroxide (10-35% NH3) ...................................................................................... 8413.5 Sodium hexafluorosilicate ........................................................................................................ 8813.6 Ferrous ammonium sulfate hexahydrate ................................................................................. 8913.7 Sodium molybdate dihydrate ................................................................................................... 9113.8 Reagent R1A ........................................................................................................................... 9213.9 Reagent R1M........................................................................................................................... 9413.10 Reagent R2............................................................................................................................ 9613.11 Reagent R3............................................................................................................................ 9813.12 Lithium battery safety precautions ....................................................................................... 100

Page 6: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

4

Table of Contents

Page 7: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

5

Section 1 General Information

1.1 DisclaimerThe information in this manual has been carefully checked and is believed to be accurate. However, Hach Lange assumes no responsibility for any inaccuracies that may be contained in this manual. In no event will Hach Lange be liable for direct, indirect, special, incidental, or consequential damages resulting from any defect or omission in this manual, even if advised of the possibility of such damages. In the interest of continued product development, Hach Lange reserves the right to make improvements in this manual and the products it describes at any time, without notice or obligation.

Copyright © 2009 by Hach Lange. All rights reserved. No part of the contents of this manual may be reproduced or transmitted in any form or by any means without the written permission of Hach Lange.

1.2 Safety informationRead this entire manual before unpacking, setting up or operating this equipment. Pay attention to all danger and caution statements. Failure to do so could result in serious injury to the operator or damage to the equipment.

To make sure that the protection provided by this equipment is not impaired, do not use or install this equipment in any manner other than that specified in this manual.

Note: This equipment has been tested and found to comply with the limits for Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.

1.2.1 Use of hazard information

WARNINGA warning is used to indicate a condition which, if not met, could cause serious personal injury and/or death. Do not move beyond a warning until all conditions have been met

CAUTIONA caution is used to indicate a condition which, if not met, could cause minor or moderate personal injury and/or damage to the equipment. Do not move beyond a caution until all conditions have been met.

Note: A note is used to indicate important information or instructions that should be considered before operating the equipment.

1.2.2 Safety recommendationsFor safe operation, it is imperative that these service instructions be read before use and that the safety recommendations mentioned herein be scrupulously respected. If repairs or adjustments are necessary, the analyzer should be returned to an authorized Hach Lange service center.

If danger warnings are not heeded to, serious material or bodily injury could occur.

WARNINGIn accordance with safety standards, it must be possible to disconnect the power supply of the analyzer in its immediate vicinity.

WARNINGThe installation of the analyzer should be performed exclusively by personnel specialized and authorized to work on electrical installations, in accordance with relevant local regulations.

Page 8: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

6

General Information

1.2.3 Service and repairsNone of the analyzer’s components can be serviced by the user. Only personnel from Hach Lange or its approved representative(s) is (are) authorized to attempt repairs to the system and only components formally approved by the manufacturer should be used. Any attempt at repairing the analyzer in contravention of these principles could cause damage to the analyzer and corporal injury to the person carrying out the repair. It renders the warranty null and void and could compromise the correct working of the analyzer and the electrical integrity or the CE compliance of the analyzer.

If you have any problems with installation, starting, or using the analyzer please contact the company that sold it to you. If this is not possible, or if the results of this approach are not satisfactory, please contact the manufacturer’s Customer Service.

1.2.4 Precautionary labelsRead all labels and tags attached to the analyzer. Personal injury or damage to the analyzer could occur if not observed.

This symbol, when noted on a product enclosure or barrier, indicates that a risk of electrical shock and/or electrocution exists and indicates that only individuals qualified to work with hazardous voltages should open the enclosure or remove the barrier.

This symbol, when noted on the product, indicates that the marked item can be hot and should not be touched without care.

This symbol, when noted on the product, indicates the presence of devices sensitive to electrostatic discharge and indicates that care must be taken to prevent damage to them.

This symbol, when noted on the product, identifies a risk of chemical harm and indicates that only individuals qualified and trained to work with chemicals should handle chemicals or perform maintenance on chemical delivery systems associated with the equipment.

This symbol, if noted on the product, indicates the need for protective eye wear.

This symbol, when noted on the product, identifies the location of the connection for protective earth (ground).

Electrical equipment marked with this symbol may not be disposed of in European public disposal systems. In conformity with European local and national regulations, European electrical equipment users must now return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.

Products marked with this symbol indicates that the product contains toxic or hazardous substances or elements. The number inside the symbol indicates the environmental protection use period in years.

Page 9: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

7

General Information

1.3 Product recycling information

ENGLISH

Electrical equipment marked with this symbol may not be disposed of in European public disposal systems after 12 August 2005. In conformity with European local and national regulations (EU Directive 2002/96/EC), European electrical equipment users must now return old or end-of-life equipment to the manufacturer for disposal at no charge to the user.Note: For return for recycling, please contact the equipment manufacturer or supplier for instructions on how to return end-of-life equipment for proper disposal.

DEUTSCH

Elektrogeräte, die mit diesem Symbol gekennzeichnet sind, dürfen in Europa nach dem 12. August 2005 nicht mehr über die öffentliche Abfallentsorgung entsorgt werden. In Übereinstimmung mit lokalen und nationalen europäischen Bestimmungen (EU-Richtlinie 2002/96/EC), müssen Benutzer von Elektrogeräten in Europa ab diesem Zeitpunkt alte bzw. zu verschrottende Geräte zur Entsorgung kostenfrei an den Hersteller zurückgeben.Hinweis: Bitte wenden Sie sich an den Hersteller bzw. an den Händler, von dem Sie das Gerät bezogen haben, um Informationen zur Rückgabe des Altgeräts zur ordnungsgemäßen Entsorgung zu erhalten.

FRANCAIS

A partir du 12 août 2005, il est interdit de mettre au rebut le matériel électrique marqué de ce symbole par les voies habituelles de déchetterie publique. Conformément à la réglementation européenne (directive UE 2002/96/EC), les utilisateurs de matériel électrique en Europe doivent désormais retourner le matériel usé ou périmé au fabricant pour élimination, sans frais pour l'utilisateur.Remarque: Veuillez vous adresser au fabricant ou au fournisseur du matériel pour les instructions de retour du matériel usé ou périmé aux fins d'élimination conforme.

ITALIANO

Le apparecchiature elettriche con apposto questo simbolo non possono essere smaltite nelle discariche pubbliche europee successivamente al 12 agosto 2005. In conformità alle normative europee locali e nazionali (Direttiva UE 2002/96/EC), gli utilizzatori europei di apparecchiature elettriche devono restituire al produttore le apparecchiature vecchie o a fine vita per lo smaltimento senza alcun costo a carico dell’utilizzatore.Nota: Per conoscere le modalità di restituzione delle apparecchiature a fine vita da riciclare, contattare il produttore o il fornitore dell’apparecchiatura per un corretto smaltimento.

DANSK

Elektriske apparater, der er mærket med dette symbol, må ikke bortskaffes i europæiske offentlige affaldssystemer efter den 12. august 2005. I henhold til europæiske lokale og nationale regler (EU-direktiv 2002/96/EF) skal europæiske brugere af elektriske apparater nu returnere gamle eller udtjente apparater til producenten med henblik på bortskaffelse uden omkostninger for brugeren.Bemærk: I forbindelse med returnering til genbrug skal du kontakte producenten eller leverandøren af apparatet for at få instruktioner om, hvordan udtjente apparater bortskaffes korrekt.

Page 10: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

8

General Information

SVENSKA

Elektronikutrustning som är märkt med denna symbol kanske inte kan lämnas in på europeiska offentliga sopstationer efter 2005-08-12. Enligt europeiska lokala och nationella föreskrifter (EU-direktiv 2002/96/EC) måste användare av elektronikutrustning i Europa nu återlämna gammal eller utrangerad utrustning till tillverkaren för kassering utan kostnad för användaren. Obs! Om du ska återlämna utrustning för återvinning ska du kontakta tillverkaren av utrustningen eller återförsäljaren för att få anvisningar om hur du återlämnar kasserad utrustning för att den ska bortskaffas på rätt sätt.

ESPANOL

A partir del 12 de agosto de 2005, los equipos eléctricos que lleven este símbolo no deberán ser desechados en los puntos limpios europeos. De conformidad con las normativas europeas locales y nacionales (Directiva de la UE 2002/96/EC), a partir de esa fecha, los usuarios europeos de equipos eléctricos deberán devolver los equipos usados u obsoletos al fabricante de los mismos para su reciclado, sin coste alguno para el usuario. Nota: Sírvase ponerse en contacto con el fabricante o proveedor de los equipos para solicitar instrucciones sobre cómo devolver los equipos obsoletos para su correcto reciclado.

NEDERLANDS

Elektrische apparatuur die is voorzien van dit symbool mag na 12 augustus 2005 niet meer worden afgevoerd naar Europese openbare afvalsystemen. Conform Europese lokale en nationale wetgegeving (EU-richtlijn 2002/96/EC) dienen gebruikers van elektrische apparaten voortaan hun oude of afgedankte apparatuur kosteloos voor recycling of vernietiging naar de producent terug te brengen.Nota: Als u apparatuur voor recycling terugbrengt, moet u contact opnemen met de producent of leverancier voor instructies voor het terugbrengen van de afgedankte apparatuur voor een juiste verwerking.

POLSKI

Sprzęt elektryczny oznaczony takim symbolem nie może być likwidowany w europejskich systemach utylizacji po dniu 12 sierpnia 2005. Zgodnie z europejskimi, lokalnymi i państwowymi przepisami prawa (Dyrektywa Unii Europejskiej 2002/96/EC), użytkownicy sprzętu elektrycznego w Europie muszą obecie przekazywać Producentowi stary sprzęt lub sprzęt po okresie użytkowania do bezpłatnej utylizacji.Uwaga: Aby przekazać sprzęt do recyklingu, należy zwrócić się do producenta lub dostawcy sprzętu w celu uzyskania instrukcji dotyczących procedur przekazywania do utylizacji sprzętu po okresie użytkownia.

PORTUGUES

Qualquer equipamento eléctrico que ostente este símbolo não poderá ser eliminado através dos sistemas públicos europeus de tratamento de resíduos sólidos a partir de 12 de Agosto de 2005. De acordo com as normas locais e europeias (Directiva Europeia 2002/96/EC), os utilizadores europeus de equipamentos eléctricos deverão agora devolver os seus equipamentos velhos ou em fim de vida ao produtor para o respectivo tratamento sem quaisquer custos para o utilizador.Nota: No que toca à devolução para reciclagem, por favor, contacte o produtor ou fornecedor do equipamento para instruções de devolução de equipamento em fim de vida para a sua correcta eliminação.

Page 11: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

9

General Information

1.4 Product disposalNote: The following only applies to European customers.

Hach Lange is committed to ensuring that the risk of any environmental damage or pollution caused by any of its products is minimized as far as possible. The European Waste Electrical and Electronic Equipment (WEEE) Directive (2002/96/EC) that came into force on August 13 2005 aims to reduce the waste arising from electrical and electronic equipment; and improve the environmental performance of all those involved in the life cycle of electrical and electronic equipment.

In conformity with European local and national regulations (EU Directive 2002/96/EC stated above), electrical equipment marked with the above symbol may not be disposed of in European public disposal systems after 12 August 2005.

Hach Lange will offer to take back (free of charge to the customer) any old, unserviceable or redundant analyzers and systems which carry the above symbol, and which were originally supplied by Hach Lange. Hach Lange will then be responsible for the disposal of this equipment.

In addition, Hach Lange will offer to take back (at cost to the customer) any old, unserviceable or redundant analyzers and systems which do not carry the above symbol, but which were originally supplied by Hach Lange. Hach Lange will then be responsible for the disposal of this equipment.

Should you wish to arrange for the disposal of any piece of equipment originally supplied by Hach Lange, please contact your supplier or our After Sales Service department in Geneva for instructions on how to return this equipment for proper disposal.

Page 12: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

10

General Information

1.5 Restriction of hazardous substances (RoHS)The European Union RoHS Directive and subsequent regulations introduced in member states and other countries limits the use of six hazardous substances used in the manufacturing of electrical and electronic equipment.

Currently, monitoring and control instruments do not fall within the scope of the RoHS Directive, however Hach Lange has taken the decision to adopt the recommendations in the Directive as the target for all future product design and component purchasing.

Note: The following only applies to exports of this product into the People’s Republic of China.

Transmitter box X

CPU PCB (with battery) O O

Power PCB O O

RS 485 O

Profibus O

PCB CAN O

PCB Local controller O

PCB Measure O

PCB Pumps O

PCB Mixer O

PCB Channel O

O:

X:

Page 13: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

11

Section 2 Specifications

2.1 Technical specificationsSpecifications are subject to change without notice.

Sample

Number of channels 1 - 6

Measurement cycle < 10 min / channel

Sample pressure 0.2 to 6 bar (3 to 87 psi)

Temperature 5 - 50°C (41 - 122°F)

Sample flowMinimum 5L / hourMaximum 30L / hour

Connections

Sample inputSimple fittings for 6 mm O.D. tubing or ¼" O.D. in PE-low density. ¼" OD in PHED-PTFE-SS as option

Sample output Barbed stem for 12 mm (½" I.D.) hose

Conditions of use

Ambient temperature 5 to + 45°C (41 to 113°F)

Power supply voltage fluctuation ± 10%

Overvoltage category 2 (according to standard EN 61010-1)

Degree of pollution 2 (according to CEI 664)

Altitude ≤ 2000 m

Relative humidity 10 to 80%

Measurement category Cat II, Class 1 (overvoltage < 1500V)

Weight

Panel analyzer without canister 13 kg

Cabinet analyzer without canister 65 kg

Full canisters 8 kg

Standards

EMC EN 61326-1 (1997) & A1 (1998) & A2 (2001) & A3 (2004)

European safety standards IEC 61010-1 (2001) & EN 61010-1 (2001)

UL & CSA agreement Certified to UL & CSA safety standards, with cETLus mark (Control Number: 3164002)

UL & CSA standardsSafety of Electrical Equipment for Measurement, Control and Laboratory use; UL61010-1, Issued: 2004/07/12, Ed.2 Rev: 2005/07/22 and CAN/CSA C22.2 No. 61010-1, Issued: 2004/07/12

Analysis

Value measured Dissolved SiO2

Cycle time Approximately 10 minutes per channel

Measurement range (2 versions) 0 - 1000 ppb 0 - 5000 ppb

Repeatability ± 2% or ± 0.5 ppb ± 2% or ± 2 ppb

Detection limit 0.5 ppb 2 ppb

Page 14: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

12

Specifications

Inputs / Outputs

Alarms

Number 8

Functions 6 threshold alarms / lack of sample / active channel1 system alarm 1 warning alarms

Cut-off power 30 VDC, 0.5 A maximum

Logical inputsNumber 12

Remote alarm clearanceBy-pass measurement for a channel

Serial input interface RS 485, Modbus communication protocol

Outputs 4 / 20 mA

Number 8

4/20 mA or 0/20 mA programmableGalvanically insulatedMaximum charge of 800 ohms

Materials and protection

Transmitter Aluminium with polyester paint

Panels and other boxes Polystyrene

Protection transmitter box IP 65

Protection cabinet (optional) IP 54

Cabinet front door Plexiglas

Maintenance

Calibration Chemical zero, slope with calibration solution

MaintenanceNo particular maintenance is necessaryCleaning can be done with a soft non-aggressive cloth

Reagent consumption Approximately 1L per month and per reagent

Calibration solution consumption Approximately 200 ml / calibration

Time safeguard lithium battery CR 1220 (3 V)

Transmitter

Display

Programming by menuConcentration of 6 samplesAnalyzer statusAlarmsTrend curvesConcentration historicCalibration parameters historic (date, slope and offset)

Mains power supply

Mains100 - 240 VAC 50 - 60 HzAutomatic switchingMax. consumption: 80 VA

Page 15: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

13

Section 3 Analyzer Overview

3.1 Operation

The 9210 can analyze up to six different samples. The sample to be analyzed circulates in a fast loop permitting rapid renewal of the sample. Adjustment of the flow is carried out with the help of a needle valve (2). At the beginning of the analysis, the sample is introduced into the measuring cell (4) with the use of a solenoid valve (5).

The reagents R1M and R1A are first added using two of the reagent pumps (9). The silica contained in the sample then reacts with the molybdate and forms the silicomolybdic complex. The reaction will take up to 5 minutes.

Oxalic acid is then added using a reagent pump (9) to avoid phosphate interference and to intensify the color.

The silicomolybdic complex is reduced to a blue molybdenum complex by means of ferrous ions. A photometric measurement is carried out at the end of the reaction.

Figure 1 Working principal

1 - Overflow tank with level detector 6 - Mixing motor 11 - Calibration pump

2 - Flow adjustment valves (1 to 6) 7 - Magnetic bar 12 - Calibration solution canister

3 - Sample input (1 to 6) 8 - Drain 13 - Photometer

4 - Photometric measuring cell 9 - Reagent pumps (x4) 14 - Drain pump of measuring cell

5 - Measuring solenoid valve 10 - Reagent canisters (x4)

Page 16: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

14

Analyzer Overview

3.2 CalibrationIn order to ensure accurate measurements, the analyzer should be calibrated periodically. The zero is achieved chemically and internally in the analyzer. The system slope is controlled by comparison to a standard solution of known concentration. Further information regarding calibrations can be found in the section entitled Calibration on page 51.

3.3 Analyzer outputsThe instrument can measure silica concentration from three different sources:

• the process sample during a normal measuring cycle

• the calibration solution during an automatic calibration

• an external sample during the grab sample operation

Whatever the sample source, the measurement cycle structure is identical and all analyzer outputs (e.g. 4/20 mA outputs, screen display, RS485 data, alarm status, etc.) are refreshed during the last minute of the measurement.

3.4 MaintenanceTo ensure a high degree of accuracy is maintained by the analyzer, it is recommended to perform a specific maintenance procedure once a year. All replacement parts and fittings are available by purchasing the maintenance kit 09120=A=8000 which contains enough parts and fittings for two years.

Full details on this procedure and the contents of the maintenance kit can be found in Yearly maintenance on page 60.

Page 17: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

15

Analyzer Overview

3.5 Presentation

3.5.1 Front panel

Figure 2 Analyzer front panel

1 - Transmitter box 4 - Sample 1 to 6 flow adjustment valves

2 - Sample level detector 5 - Photometric measurement cell

3 - Overflow vessel

Page 18: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

16

Analyzer Overview

3.5.2 Rear panel

Figure 3 Analyzer rear panel

6 - Local controller box 9 - Measurement box

7 - Measurement solenoid valve 10 - Sample inlet valves

8 - Pumps box 11 - Drain

Page 19: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

17

Section 4 Installation

WARNINGThe analyzer should only be assembled by qualified staff. Mains power should only be connected once installation has been completed and checked.

4.1 Analyzer inspection and unpackingThe analyzer has been factory tested and checked prior to shipping. We nevertheless recommend that you perform a visual inspection in order to ensure that it has not been damaged. Any marked packaging is a potential sign of damage that may not be immediately visible. Keep all packaging in the event of claims.

Open the box. Take the packing list from the box and check the presence of all the items selected. If any parts or accessories are missing, refer to your distributor or to Hach Lange.

4.2 Analyzer preparationBefore installing the analyzer, think about the following:

• Place the analyzer close to the sample point. This will allow the response time to be reduced.

• The sample should be homogenous and representative.

• The temperature of the sample should be between 5 and 50°C.

• The pressure of the sample should be between 0.2 and 6 bar and remain relatively stable.

• The solution should be free of particles.

• Sample lines should be in PE/PTFE/FEP (4 x 6 mm).

• Avoid any location with a corrosive atmosphere or subject to liquid spills.

• Chose a dry and dust-free location.

• The ambient temperature of the analyzer should not exceed 45°C. If the temperature is below 5°C, the analyzer should be installed in a heated cabinet (not provided by Hach Lange).

4.3 Installation checklistFor a complete installation, perform the following actions in the order specified, following the instructions very carefully:

1. Mount the analyzer - section 4.4 on page 18

2. Install the canister holder - section 4.5 on page 20

3. Connect the sample lines - section 4.6 on page 21

4. Connect the drain tube - section 4.7 on page 22

5. Connect the mains power - section 4.8 on page 22

6. Connect the external communications - section 4.9 on page 26

7. Input/Output connections - section 4.10 on page 27

8. Reagent preparation - section 4.11 on page 29

9. Connect reagent canisters - section 4.12 on page 31

10. Analyzer stabilization and startup - section 4.13 on page 32

Page 20: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

18

Installation

4.4 Mounting the analyzer

CAUTIONWherever the analyzer is to be mounted, it is important to note that it must be placed in an upright position with the transmitter at the top. It is recommended to use a spirit level to ensure that the analyzer is correctly positioned and not leaning to one side or forward. This is essential to guarantee the accuracy of the analyzer.

4.4.1 Panel version For the panel version, the dimensions of the analyzer and fixation holes are shown in Figure 4.

A. Mains power supply cable gland, drilled ∅ 11 (5 to 10mm), 100…240V, 50-60Hz, 25VA

B. Field bus cable gland, drilled ∅ 13 (6 to 12mm)

C. 5 to 7 mm cable gland

• Alarm signals: 30V, 0.5A max.

• Logic inputs: Cutoff on channels 1 to 6, acknowledgement of alarm signals

• Analog outputs: 0-20 mA or 4-20 mA

D. Sample inlets, (channels 1 to 6), QR coupling for semi-rigid tubing ∅ 6mm, 5 to 50°C (40° to 120°F), P 0.2 to 6 bar (3 to 87 PSI)

E. Drain tube: silicon tubing ∅ 12 x 17mm, at atmospheric pressure

Figure 4 Panel version dimensions - mm [ins]

Page 21: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

19

Installation

4.4.2 Cabinet versionFor the cabinet version, the dimensions of the analyzer and fixation holes are shown in Figure 5.

A. Cable gland, 5 to 7mm

• Mains power supply cable gland, drilled ∅ 11 (5 to 10mm), 100-240V, 50-60Hz, 25VA

• Field bus cable gland, drilled ∅ 13 (6 to 12mm)

• Alarm signals: 30V, 0.5A max.

• Logic inputs: Cutoff on channels 1 to 6, acknowledgement of alarm signals

• Analog outputs: 0-20mA or 4-20mA

B. Sample inlets (channels 1 to 6), QR coupling for semi-rigid tubing ∅ 6mm, 5 to 50°C (40° to 120°F), P 0.2 to 6 bars (3 to 87 PSI)

C. Drain tube: silicon tubing ∅ 12 x 17mm, at atmospheric pressure.

Figure 5 Cabinet version dimensions - mm [ins]

Page 22: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

20

Installation

4.5 Installing the canister holderFor cabinet versions, a canister holder is pre-mounted in the rear of the cabinet. For the panel version, a canister holder is supplied separately and must be installed under the analyzer as illustrated in Figure 6 below.

Ensure the position of the canister holder does not exceed the maximum and minimum distances from the bottom of the analyzer that are given in Figure 6 above.

Figure 6 Canister holders

Page 23: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

21

Installation

4.6 Connecting the sampleUse new tubes for connections during installation

• Exterior ∅: 6 mm exactly (or ¼'')

• Material: polyethylene or PTFE or FEP

• Pressure: 0.2 to 6 bar

• Temperature: 5 to 50°C

At this stage of the installation, make sure that the flow valves (2) are all closed.

Connect the tubes by inserting into the quick release connections found under the sampling block as illustrated in Figure 8 below.

Figure 7 Sample connections schematic

1 - Solenoid valve 3 - Sample input 1 to 6

2 - Solenoid valve flow adjustment screw

Figure 8 Sample connections

Page 24: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

22

Installation

4.7 Connecting the drain tubeThe drain outlet is located on the bottom of the analyzer. A 12 x 17 mm pipe is delivered with the analyzer and should be connected to the drain outlet at one end and the other fed to a drain for sample evacuation.

Note: Evacuation should be free of any mechanical constraints, and any counter-pressure.

4.8 Mains power connection

WARNINGNo intervention should be carried out on the analyzer without first switching off the power.

The electrical installation should be carried out by duly qualified personnel. A supply voltage of 100-240 VAC is acceptable without changing the configuration. The power supply terminals can be removed from their housing to make connection easier.

For safety reasons, it is imperative to respect the following:

• Use a three-wire power lead (live + neutral + earth) for supplying the required power.

• The analyzer should be connected to the mains via a circuit-breaker or fuse whose value should be less or equal to 20 A. It should be located in the proximity and be identified. This connection should cut-off the live and neutral when electrical problems occur or when the user wishes to intervene inside the analyzer. However, the earth conductor should always be connected.

Figure 9 Drain connection

Page 25: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

23

Installation

The following procedure outlines the steps required to connect power to the analyzer. The main cabinet should be open with access to the interior:

1. Open the transmitter front door by unscrewing the four holding screws.

2. Swing open the door (it is hinged to the left) to reveal the inside of the transmitter.

3. Remove the metallic shielding plate protecting access to the main board.

4. Once the shielding plate is removed, the location of the power cable gland can be seen (indicated with the arrow).

Page 26: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

24

Installation

5. Next, run the power supply cable through one of the external cable glands located on the left outside of the cabinet (A), and into the analyzer.

6. Put your hand behind the local controller box and you will locate two cable glands to the left and two to the right on the bottom of the transmitter (B). The cable gland for the power cable is located left and nearest to you.

7. Unscrew the cable gland nut, pass the power cable through it, and then up through the cable gland and into the transmitter.

8. Screw back the cable gland nut to secure the power cable.

9. Remove the power supply connector and note where the earth, live and neutral must be connected.

Page 27: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

25

Installation

10. Connect the power supply cables to the connector.

11. Put the connector back in place.

12. Replace the metallic shielding plate, ensuring it is in front of the power cable just installed.

Page 28: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

26

Installation

4.9 External communications connection1. As with the power cable,

run the communications cable through an external cable gland on the left side of the cabinet, and into the analyzer.

2. Pass it through the cable gland located right and farthest from you on the base of the transmitter, so it appears inside the transmitter through the left front cable gland (as indicated).

3. Depending on the type of board installed in the analyzer, connect the communication cable as indicated.

Note: Connection is the same on the CPU board for both the JBUS/MODBUS and PROFIBUS options.

CPU board without RS485 option

CPU board with JBUS/MODBUS option - - - - - CPU board with PROFIBUS option

4. Close the transmitter door and screw securely back in place.

Page 29: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

27

Installation

4.10 Input/Output connectionsTo gain access to the I/O connections, open the back of the local controller box by unscrewing the 4 corner screws.

This will reveal the I/O board as illustrated in Figure 10 (which shows connections already in place).

Figure 10 Local controller connections

Page 30: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

28

Installation

Before using any of the cable glands, perforate first with a screwdriver. To ensure a good seal, the external diameter of the cables should be between 5 and 7 mm.

For information on the connections available refer to Table 1.

CAUTIONThe relay outputs can only supply power in very low safety voltage (30 VAC or 42.4 VDC maximum) and limited to 0.5 A.

The nomenclature given in the connections column of Table 1 below, refers to the same nomenclature that is printed on the I/O board against each available connection.

When all I/O connections have been made, put the cover back on the local controller box and secure in place with the 4 screws.

Connections Function Max. voltage Max. current

Mixer control 24 VAC 1 A

Heater control 24 VDC 1 A

Cell temperature measurement 5 VDC 1 mA

Re1 to Re6 User relay of thresholds 1 - 6

30 Veff or 42.4 VDC

0.5 A resistive charge

Re7 to Re10 Not used

Re11 Warning alarm

Re12 System alarm (NO and NC switches are available)

Note: Dry relays are used for all the following logical inputs. At best, provide 1 dry circuit (no voltage) which will be closed to launch an action.

In1 to In6By-pass sample measurement (channels 1 - 6)For input In1, use terminals " in " and " - "Terminals " + " and " out " are not used

5 VDC 5 mAIn7 to In10 Not used

In11 Remote acknowledgment of alarms

In12 Level sample detector

Iout1 to Iout6 Analog output 0-20 or 4-20 mA (channels 1 - 6)

24 VDC 23 mAIout7 Analyzer status

Iout8 Not used

Vout1 Sample electrovalve control

30 VDC 1 A

Vout2 to Vout7 Sample selection electrovalve controls (channels 1 - 6)

Vout8 Calibration pump

Vout9 to Vout11 Reagent pumps

Vout12 Drain pump

Table 1 Relay functions

Page 31: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

29

Installation

4.11 Reagent preparationThe four reagent solutions, and the calibration solution should be prepared as per the following instructions. Before refilling the five canisters with 2 liters of their respective solution, they should be thoroughly rinsed with ultra pure water.

Note: These solutions are best prepared with the aid of a magnetic stirrer.

WARNINGRead the Material Safety Data Sheets (MSDS) on page 77 and wear protective goggles and clothes before using chemical products.

4.11.1 Reagent 1M - Molybdate (2 liters)

• Rinse a 2 liter volumetric flask three times with ultra pure water

• Fill two-thirds of the flask with ultra pure water which has less than 5 ppb of SiO2

• Stirring continuously for about 5 minutes, gently dissolve 110g of the sodium dihydrate molybdate in the water. The turbidity of the solution should be very low

• Top up the solution to 2 liters with additional ultra pure water, and transfer the solution to the canister labelled R1M

4.11.2 Reagent 1A - Nitric acid (2 liters)

• Rinse a 2 liter volumetric flask three times with ultra pure water

• Fill two-thirds of the flask with ultra pure water which has less than 5 ppb of SiO2

• Stirring continuously for about 5 minutes, gently pour 300mL of nitric acid into the water. The solution should be clear

• Top up the solution to 2 liters with additional ultra pure water, and transfer the solution to the canister labelled R1A

4.11.3 Reagent 2 - Oxalic acid

• Rinse a 2 liter volumetric flask three times with ultra pure water

• Fill two-thirds of the flask with ultra pure water which has less than 5 ppb of SiO2

• Stirring continuously for about 5 minutes, gently dissolve 80g of oxalic acid in the water

• Top up the solution to 2 liters with additional ultra pure water, and transfer the solution to the canister labelled R2

Label Composition Concentration

R1M Sodium dihydrate molybdate Na2MoO4.2H2O >99%, neat (analytical quality; Fluka# 71756, Merck# 106521 or equivalent)

55 g/L

Label Composition Concentration

R1A Nitric Acid, HNO3 (65% Puriss analytical grade; Fluka# 84380 or equivalent)

150 mL/L, 15% V/V

Label Composition Concentration

R2 Oxalic Acid (Puriss analytical grade; Fluka# 75700 or equivalent) 40 g/L

Page 32: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

30

Installation

4.11.4 Reagent 3 - Reducing reagent

• Rinse a 2 liter volumetric flask three times with ultra pure water• Fill two-thirds of the flask with ultra pure water which has less than 5 ppb of SiO2

• Stirring continuously, gently pour 25mL of sulfuric acid into the water• Stirring continuously, gently add 40g of ammonium-iron(II) sulfate hexahydrate to the

solution• Top up the solution to 2 liters with additional ultra pure water, and transfer the solution to

the canister labelled R3

4.11.5 Calibration solution CONCENTRATED SOLUTION

Use of Titrisol® (Merck) cartridge

Dilute one cartridge of Titrisol® Merck 109947 solution with one liter of demineralized water to give a concentration of 1000 mg/liter of Si, i.e. 2139 mg/liter of SiO2. Dilute this solution 100 times, i.e. 10 mL per liter to give a concentration of 10 mg/liter of Si, i.e. 21.39 mg/liter SiO2.

CALIBRATION SOLUTION

By diluting the concentrated solution (at 21.39 mg/liters) with demineralized water, various calibration solutions with a known concentration of SiO2 may be prepared. The following table gives the nominal values when using demineralized water having a zero concentration of silica:

Given that the demineralized water may often have a non-zero silica content, use of one two methods for determining the nominal silica values of the final calibration solution. On completion, transfer 2 liters of the solution to the calibration canister labelled CAL.

Note: It is recommended to use demineralized water with a silica content of no more than 5 ppb when diluting the concentrated solution.

Method 1:

Determine the silica content of the demineralized water by using a laboratory measurement technique (e.g. Hach Lange laboratory photometer), or by using the grab sample mode of an accurately calibrated POLYMETRON 9210 if available on site. The concentration value can then be added to the values in the above table, giving the following revised table when (for example) using demineralized water with a silica content of 5 ppb (which equates to 2.33 ppb Si):

Label Composition Concentration

R3 Ammonium-iron(II) sulfate hexahydrate (NH4)2Fe(SO4)2, 6H2O (Puriss analytical grade; Fluka# 09720, Merck# 103792 or equivalent)

20 g/L

Sulfuric acid (H2SO4) (95-97%; Fluka# 84721 or equivalent) 12.5 mL/L

mL of concentrated solution 21.39 mg/l

Calibration solution ppb Si

Calibration solution ppb SiO2

Analyzer measuring range

10 50 107 0-200 ppb

20 100 213.9 0-200 ppb

25 125 267.4 0-500 ppb

50 250 534.8 0-1000 ppb

mL of concentrated solution 21.39 mg/l

Calibration solution ppb Si

Calibration solution ppb SiO2

Analyzer measuring range

10 52.3 112 0-200 ppb

20 102.3 218.9 0-200 ppb

25 127.3 272.4 0-500 ppb

50 252.3 539.8 0-1000 ppb

Page 33: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

31

Installation

Method 2:

Prepare the final calibration solution by diluting the concentrated solution (at 21.39 mg/liters) with demineralized water. Then determine the exact silica concentration value of this solution by using a laboratory measurement technique (e.g. Hach Lange laboratory photometer) or by using the grab sample mode of an accurately calibrated POLYMETRON 9210 if available on site.

4.12 Connecting the canistersEach reagent tubing is labelled individually and delivered already connected to the analyzer. They are fed through, and attached to, caps that attach to the reagent canisters. One end of the tube is fitted with a chemically inert PVC weight that is used to position the tube in the bottom of the canister.

Connect each cap to its canister:

• Tube R1M to reagent canister labelled R1M: Sodium molybdate

• Tube R1A on reagent canister labelled R1A: Nitric acid

• Tube R2 on reagent canister labelled R2: Oxalic acid

• Tube R3 on reagent canister labelled R3: Sulfuric acid and ferrous ammonium sulfate

Screw the cap on each canister so that the weight is resting at the bottom of the canister. Put the canister in the canister holder with the cap towards the exterior as shown in Figure 11 below.

The canister on the far right of the illustration is the calibration solution. This has a thicker tube compared to the reagents. Ensure that the cap has also been securely screwed on to this canister.

At this stage, the analyzer doors can now be closed and locked as access to the interior of the analyzer is no longer required.

Figure 11 Canisters connected

Page 34: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

32

Installation

4.13 Analyzer startupAt this stage the analyzer has been completely installed. However, you will be required to enter data using the display panel functions to make the analyzer operational. Therefore, in order to familiarize yourself with data entry procedures, it is advisable to read the section entitled Operating Instructions on page 33 before continuing.

Once you are familiar with data entry procedures, connect the analyzer to the mains power supply. Introductory screens will initially be displayed for about 60 seconds, before the main measurement screen is displayed.

4.13.1 Reagents volume declarationThe reagent and calibration solution canisters have been filled with 2 liters of solution. The analyzer must be set up to reflect this.

Select the Menu option from the display and select the MAINTENANCE option from the display, followed by the REAGENTS REFILL option. Set the BOTTLES FULL parameter to Yes. An example of this screen is shown in section 9.8 on page 64. You can also check that the values have been set to 2 liters by selecting the ADJUST VOLUMES option.

4.13.2 Flow rate adjustmentNext, the flow rates for each measurement channel must be set. When the sample tubes were connected, the flow valves were all closed (see section 4.6 on page 21). They must now be opened to allow sample to flow and be measured. To do this select the STARTUP option from the MAINTENANCE menu and follow the instructions described in section 9.7 on page 64.

4.13.3 System and user setup proceduresAt this point, the analyzer has been completely installed and can begin sample measurement. To do this, use the Esc function key to exit from the MAINTENANCE menu until you return to the measurement display screen. Press the Start function key to start the measurement process. The analyzer must be left to run for a period of two hours in order to stabilize.

During this stabilization period, you can setup the analyzer by first entering system data (see System Setup on page 37) and user data (see User Setup on page 41).

Once the system and user data has been input and the two hour period of stabilization has elapsed, the analyzer will need to be calibrated. Follow the procedures in Calibration on page 51.

The analyzer should now be ready for use. If any problem should arise, please refer initially to Troubleshooting on page 69. If the problem cannot be overcome, please contact your Hach Lange representative who will be happy to help.

Page 35: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

33

Section 5 Operating Instructions

5.1 Data Entry

5.1.1 Function KeysThe display panel of the 9210 has 5 function keys (illustrated below) to allow menu option selection, field selection, and data entry options.

• The Esc key cancels data input or goes back to the previous screen.

• The Enter key validates the input and goes on to the next step.

• The Up Arrow keys select the option displayed immediately above them on the screen.

5.1.2 Modification of a valueThe highlighted digit may be modified with the key. Each digit can then be modified by pressing the Enter function key. Repeat for each digit.

• On the first digit: possible to enter the " - " sign

• On the other digits: possible to enter a decimal point

Note: If you do not use the keyboard for at least 10 minutes, the analyzer returns to the measure mode except during the calibration and maintenance options. An access code may be required for the calibration, programming and service menus (see Code on page 38).

5.2 Measurement screens

5.2.1 Main screen

Figure 12 Function Keys

1. Name of last measured sample 7. Display of the last calibration data2. Current time 8. Function keys3. Concentration of last measured sample 9. Access to the main menu4. Concentration units 10. Go to display 25. Sample being analyzed (bar graph indicates progress) 11. Start or stop the measurement6. Measurement activity

Page 36: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

34

Operating Instructions

5.2.2 Display screen 2 - Measurement history

5.2.3 Display screen 3 - Alarms

5.2.4 Display screen 4 - Graph

Displays a list of the last measurements (sample name, concentration and time the measurement was taken).

Disp3 takes you to the next display screen.

Alarms S1-S6 relate to the six alarm outputs which are followed by the warning and system alarms.

The message against each alarm will indicate the status:

• OK to indicate no problems

• an appropriate error message

• or DISABLE to indicate the alarm has been disabled (this only applies to the warning and system alarms)

If a manual alarm has been set, press Enter to acknowledge and deactivate the alarm.

Refer to Alarm conditions on page 46 for a list of possible error messages.

Disp4 takes you to the next display screen.

This screen shows the measurement trend over a 24 hour period. The concentration scale is the same as defined by the analog output scale low and upper values (see Affect on page 47).

The sample name and associated analog output (in brackets) is displayed along the top.

To display the graph for the next sample, press Graph+ or press Main to return to the main measurement screen.

Page 37: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

35

Operating Instructions

5.3 Main menuFor a diagrammatic overview of the complete menu structure, refer to Analyzer menu overview on page 36.

The main menu is accessible from the main measurement screen (see illustration in Main screen on page 33). Press the function key under Menu to take you to the main menu.

Note: If a PROGRAMMING password has been set (see Code on page 38 for details on how to do this), you will be required to enter a password to gain access to the main menu.

In the illustration above, the symbol “.../...” is displayed in the bottom right. This indicates that there are more options available in this menu than those currently displayed. As you scroll down the other menu options will become visible in a rolling menu.

The first option in the menu will always be highlighted by default. To scroll to the required option, press the Up Arrow function key under the Select option.

Note: Scrolling past the last item in the list will take you back to the first item. It is not possible to scroll upwards through any list.

Detailed information on all the menu options is available throughout in this manual.

Page 38: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

36

Operating Instructions

5.4 Analyzer menu overview

Figure 13 Analyzer menu overview

Page 39: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

37

Section 6 System Setup

Before attempting to setup the analyzer, ensure you have read and understood how to enter and change data fields as described in the section entitled Data Entry on page 33.

6.1 Menu overviewSelect the SERVICE option from the main menu.

Note: If a PROGRAMMING password has been set (see Code on page 38 for details on how to do this), you will be required to enter a password to gain access to the main menu. In addition, if a SERVICE password has been set, you will also be required to enter this password to get access to the service menu.

6.1.1 Time

Set the date and the time for the analyzer.

Page 40: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

38

System Setup

6.1.2 Display

6.1.2.1 Customize

6.1.3 Code

Define the language, concentration units and customize the channel displays.

Languages: choice of the message language

• French: F

• English: GB

• German: D

• Spanish: Sp

• Italian: I

The unit choice is ppb/ppm, or µg/mg/l

Define each channel with a descriptive name.

You can use lower-case letters, upper-case letters, numbers, and punctuation signs.

Use this option to enter the codes (passwords) required to access the analyzer menus. These codes are independent of one another and can have the same value.

0000 indicates that no access code is required.

Page 41: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

39

System Setup

6.1.4 Soft issues

6.1.5 Default values

6.1.6 mA adjustments

This option displays the version number of the software installed on the analyzer.

This option allows you to reload all the factory set default values (see Default Configuration on page 73 for a complete listing).

Important Note: This operation will erase all the parameters (calibration, 4-20 mA scale, alarms, sequence, etc.), and the default values will be loaded. In addition all measurement values and alarm history will also be erased.

To exit the screen without loading default values, press Esc.

Each 4-20 mA output can be individually calibrated on a two point basis.

Connect a milliammeter to each of the 8 outputs in turn.

Use the " - " and " + " keys to make minor adjustments to the values on the 4 mA outputs.

Validate (press Enter) will move you to the 20 mA outputs screen.

Adjust the 20 mA value in the same way and then press Enter to complete the adjustment process and return to the screen to select the next output to adjust.

Page 42: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

40

System Setup

6.1.7 Factory

This option is reserved for qualified Hach Lange service personnel.

If you believe that changes to the factory defined settings may be required, please contact your local Hach Lange representative.

Page 43: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

41

Section 7 User Setup

Before attempting to setup the analyzer, ensure you have read and understood how to enter and change data fields as described in the section entitled Data Entry on page 33.

The user setup process consists of accessing five sub-menus from the main menu:

• MEASURE

• ALARMS

• mA OUTPUTS

• RS485

• SEQUENCE

7.1 Measure menu overviewSelect the MEASURE option from the main menu to define the measurement parameters.

Note: If a PROGRAMMING password has been set (see Code on page 38 for details on how to do this), you will be required to enter a password to gain access to the main menu.

7.1.1 Measuring mode

This option defines the measurement cycle parameters.

If the Interval cycle is selected, the measurements are made every 15 minutes by default. You can adjust the measuring cycle time by entering a different value e.g. 030 for a measurement every 30 minutes.

If the In loop cycle is selected, measurements are made on a continuous basis.

Page 44: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

42

User Setup

7.1.2 Grab sample

7.1.3 Historic

This option allows you to take measurements from an off-line sample. An initial “Please wait...” screen is displayed while the measuring cell is rinsed.

Prepare a beaker of 200mL of the sample to analyze.

When rinsing is complete, a screen will be displayed asking for the sample to analyze (illustrated left).

• First remove the sample level detector from the overflow vessel and fill the vessel with the sample (illustrated left).

• Replace the sample level detector.

• Press OK to start the measurement.

When the measurement completes, the results are displayed on screen and a message is displayed to inform you that the measurement has finished.

The Esc key can be used to abort the measurement at any time and return to the previous screen.

This option allows you to view the previously stored measurements and any alarms that were triggered.

The PROGRAMMING option sets the date (FROM) and the time (AT) from which the history will be displayed, and the measurement channels required.

• CH: all channels or a specific channel (1 to 6)

• ALARMS (yes or no): if set to yes, displays any system and warning alarms that were triggered

Page 45: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

43

User Setup

7.1.4 ADC values

7.2 Alarms menu overviewSelect the ALARMS option from the main menu to assign alarms and parameters to each relay.

Note: If a PROGRAMMING password has been set (see Code on page 38 for details on how to do this), you will be required to enter a password to gain access to the main menu.

The READ option lets you view the data as defined by the parameters set in the programming option.

Use the key to scroll through the list of stored data which displays the date and time of the measurement, the channel number, and the measured concentration value.

This option displays the parameters for the primary and last calibrations, and the last measurement.

The information given can be used for diagnostic purposes if the analyzer appears to be giving incorrect measurements.

Page 46: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

44

User Setup

7.2.1 Alarms 1 to 6

Value Description

MODE Limit Trigger the alarm when the measurement is above or below a pre-defined limit

Active channel Trigger the alarm when the channel is active i.e. during a measurement cycle

Level sample Trigger the alarm when there is no sample

AFFECT Channel 1 to 6 Define the channel on which the alarm is triggered

LIMIT XXXX Define the concentration limit when the alarm is triggered

DIRECTION Up Trigger the alarm when the measurement is above the limit

Down Trigger the alarm when the measurement is below the limit

HYSTERESIS nn% Hysteresis percentage

RELAY N.O. Normally open

N.C. Normally closed

If MODE = Limit

• Define the channel for which alarm 2 will be triggered

• Define the limit value.

• Define if the alarm is triggered when concentration goes above (Up) or below (Down) the limit.

• Set the hysteresis percentage.

• Define the relay type (normally open or normally closed)

If MODE = Active channel

The alarm is activated when the channel is active (i.e. when the sample is being measured).

If MODE = Level sample

The alarm is activated when the sample is missing.

Page 47: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

45

User Setup

7.2.2 Warning alarm

Refer to Alarm conditions on page 46 for a list of warning alarm conditions.

7.2.3 System alarm

Note: The system alarm can only be acknowledged manually, either by holding down the Enter key or, remotely, by using the digital input, In11 on the I/O card (see Figure 10 on page 27).

Refer to Alarm conditions on page 46 for a list of system alarm conditions.

Value Description

ALARM Yes or No Activate or deactivate the warning alarm

ACCEPT Manual When the alarm is triggered, it can only be turned off by pressing the Enter function key

Auto When the alarm is triggered, it will turn itself off when the reason for the alarm is no longer valid

RELAY N.O. Normally open

N.C. Normally closed

Define the warning alarm parameters.

Value Description

ALARM Yes or No Activate or deactivate the system alarm

RELAY N.O. Normally open

N.C. Normally closed

Define the system alarm parameters.

Page 48: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

46

User Setup

7.2.4 Alarm conditionsThe following table lists all the alarms conditions:

• Warning: the analyzer displays a warning on the screen and locks the warning relay.

• System: the analyzer locks the system alarm relay, displays a message and stops.

Message Description Type

Reset mod. Measure Reset the measurement module System

Reset module LC Reset the local controller module System

Error lc Local controller module error System

Err. measure mod. Measurement module error System

Comm error Communication error Warning

Table error Error during table transmission System

Task error Task execution error System

Measure error Measurement error Warning

Reference error Problem with current to the LED Warning

Extinction error No current to the LED Warning

Meas.time over Maximum cycle time for measurement has been exceeded Warning

Cal.time over Maximum cycle time for calibration has been exceeded Warning

No sample No sample available System

Sample x missing Sample x missing Warning

Reagent < 250 ml Low level of reagent left (2.5 days before no reagent alarm) Warning

Calib < 300 ml Low level of the calibration solution (max 1 calibration left) Warning

No reagent Empty reagent canisters <150 ml System

Slope drift Slight drift of slope calibration parameter (confirmation zone) Warning

Zero drift Slight drift of zero offset calibration parameter (confirmation zone) Warning

Zero error Zero calibration major error System

Slope error Slope calibration major error System

Sample x < C Sample concentration x is lower than C Warning

Sample x < 0 ppb Sample concentration x is negative Warning

Sample x > 1000 ppb or > 5000 ppb Sample concentration out of range Warning

Reference error The reference measurement lies outside its max/min values Warning

Table 2 System and warning error messages

Page 49: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

47

User Setup

7.3 mA outputs menu overviewSelect the mA OUTPUTS option from the main menu to set the analog output parameters.

Note: If a PROGRAMMING password has been set (see Code on page 38 for details on how to do this), you will be required to enter a password to gain access to the main menu.

7.3.1 AffectSelect the AFFECT option and then each output (1-6) in turn.

Description

AFFECT Define the measurement channel (1-6). The measurement on this channel will be carried over the selected analog output

TYPE Define the analog output scale (0-20 or 4-20 mA)

MODE Select between linear or dual (see graph in Figure 14)

LOW Value corresponding to the low end of the scale

MIDDLE Value corresponding to the mid-point of the scale (only available in dual mode)

UP Value corresponding to the upper end of the scale

Define the analog output parameters.

Page 50: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

48

User Setup

The following graph indicates the result of the 4-20 current in relation to the type, and the first, middle and end value of the scale.

7.3.2 Special programming

Note: In the case of a system alarm, all the analog outputs (1 to 7) are set to the preset value defined in the SYSTEM ALARM option.

Figure 14 Linear and dual slopes

This option lets you program the status of analog output 7 for the following special events:

• Maintenance

• Calibration

• Grab sample analysis

• System alarm

• Warning alarm

Use the Select function to choose the event for which analog output 7 is to be assigned.

MODE = Last: Analog output 7 is not changed.

MODE = Preset: Analog output 7 is set to the value defined in the VALUE parameter.

Page 51: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

49

User Setup

7.3.3 Test

7.4 RS485

Test all the analog outputs.

Adjust the current VALUE.

This value is then forced on all analog outputs and can be verified with the use of a multimeter connected to the analyzer.

This option allows you to adjust the communication parameters of the analyzer. The communication protocol is MODBUS-JBUS.

Description

NO Transmitter number (0 - 33)

BAUD Transmission speed in bauds (300, 600 1200, 2400, 4800, 9600)

PARITY Odd, even, no

STOP BIT 1 or 2

SWAP WORD Lets you reverse the send order of the data

Page 52: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

50

User Setup

7.5 Sequence

7.5.1 Channel activation

7.5.2 Sequence

The option allows you to activate or deactivate the measurement channels, and to define the order in which the samples are measured.

The CHANNEL ACTIVATION option defines which channels are active and which are not.

The SEQUENCE option displays the configured sequence and lets you change it.

The sequence is set up in 12 steps.

Use the key to change the number of the channel.

The Delete key replaces the number by the asterisk symbol. This symbol indicates no measurement for that step in the sequence.

Sequences examples:

1. 1 2 3 4 5 6 * * * * * *

4 5 6 3 2 1 * * * * * *

These two sequences analyze samples (1 to 6) in a different order.

2. 12 12 12 12 12 12

or 12 ** ** ** ** **

or 1* 2* ** ** ** **

These three sequences have the same effect and set the analyzer to measure channels 1 and 2 alternately.

Page 53: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

51

Section 8 Calibration

Before attempting to setup the analyzer, ensure you have read and understood how to enter and change data fields as described in the section entitled Data Entry on page 33.

8.1 Menu overviewSelect the CALIBRATION option from the main menu.

Note: If a PROGRAMMING password has been set (see Code on page 38 for details on how to do this), you will be required to enter a password to gain access to the main menu. In addition, if a CALIBRATION password has been set, you will also be required to enter this password to get access to the calibration menu.

For continued accurate measurements, the analyzer should be calibrated on a regular basis.

Calibrations allow the adjustment of:

• ZERO of the system

• SLOPE of the system

• Both ZERO and SLOPE of the system

The ZERO calibration is performed chemically by the analyzer. In order to avoid the use of water free of SiO2, the analyzer carries out a measurement without a colorimetric reaction. The slope of the system is then calibrated with a standard solution of known concentration of SiO2.

Three types of calibration are available:

• Primary

• Manual

• Automatic

In addition, the CALIBRATION menu also gives access to the current calibration parameters and a history of previous calibration results.

Page 54: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

52

Calibration

8.1.1 Programming (automatic calibration)

This option is used to define the parameters for an automatic calibration.

The programming screen lets you define the frequency of the zero calibration, the concentration of the calibration solution and to choose between an automatic or manual calibration.

The following table indicates the recommended frequency of the zero calibration in terms of the minimum concentration of the samples.

Sample minimum concentration Recommended frequency

< 5 ppb 24 hours

5 < x < 50 ppb 48 hours

> 50 ppb > 72 hours

For an automatic slope calibration, this screen lets you define the frequency.

• When the BASE option is set to Day, the WEEK option sets the calibration day(s).

Ex 1: SMTWTFS, for a daily calibration.Ex 2: SM*W*FS, for a calibration every day except Tuesday and Thursday.

• The HOUR option is used to set the time to start the calibration.

• When the BASE option is set to Hour, the INTERVAL option sets the frequency of the slope calibration. The recommended slope calibration frequency is 1 week (168 hours).

Page 55: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

53

Calibration

8.1.2 Execution primary calibration

8.1.3 Execution manual calibration

A primary calibration is carried out at the factory and is used as a reference for all following calibrations.

In general, this type of calibration does not need to be repeated unless important changes have been made to the analyzer.

Important Note: If you feel that a primary calibration is required, please discuss this with your Hach Lange representative for advice and assistance before performing this type of calibration on your analyzer.

The primary calibration sets the slope and zero references. All results of further manual or automatic calibrations will be compared to the zero and slope references set by this calibration. Large differences will signify an error in the calibration and will be highlighted with an error message.

The calibration of the slope is defined first, followed by the zero.

Press Esc to abort the calibration and return to the CALIBRATION menu.

The calibration is made in the following stages:

• Sample preparation

• Conditioning

• Measurement

• Rinse

A bar graph indicates the calibration progress.

At the end of the calibration, the results are displayed on screen.

Primary calibration results:

S = value of the slope (ppb / Abs)

O = value of the zero shift in milli-Abs

The default values show a general idea of the standard values according to the analyzer range (0-1000 or 0-5000 ppb).

Choose the type of calibration required:

• ZERO

• SLOPE

• ZERO + SLOPE

At the end of the calibration, the results will be displayed (see example in Historic on page 54).

Page 56: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

54

Calibration

8.1.4 Parameters

8.1.5 Historic

This screen displays the primary calibration parameters and the last calibration parameters.

The default values show a general idea of the standard values according to the analyzer range (0-1000 or 0-5000 ppb).

This screen displays the calibration history with the following information:

• Date and time.

• Type: manual or automatic.

• Slope (S) and offset (O) values.

Page 57: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

55

Calibration

8.1.6 Calibration resultsAt the end of a manual or automatic calibration, the analyzer compares the results with the primary reference calibration parameters. Depending on the results, you can accept, restart or cancel the calibration. There are 3 zones:

Zone ok

Normal variations compared to the reference values. The new parameters are acceptable.

In automatic mode, the analyzer returns to its measurement cycle. In manual mode, validate with OK and restart the analyzer (Start function key).

S: XXXXX DD:MM:YY - Slope

O: XXXXX DD:MM:YY - Offset

C: XXXXX DD:MM:YY - Concentration of the sample solution just measured, calculated using the calibration values from the Last calibration.

Confirmation zone

Abnormal variations compared to the reference values. The new parameters can be accepted but a warning is displayed and confirmation is required.

In automatic mode, the analyzer does not take the new parameters, but returns to measurement mode with the previous parameters. In manual mode, you must confirm (Conf) or cancel (Cancel) to accept or reject the new parameters.

Note: In automatic mode, where the calibration lies within the confirmation zone, the analyzer takes one measurement and then recalibrates. After 3 successive invalid calibrations, the analyzer waits for the next programmed date before attempting another calibration.

Rejection zone

Parameters are abnormal. There is a basic problem such as an invalid calibration solution preparation being used.

In automatic mode, the analyzer stops and triggers a system alarm. In manual mode, you can Cancel (return to measurement mode with the previous parameters) or Restart the calibration.

Page 58: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

56

Calibration

Page 59: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

57

Section 9 Maintenance

9.1 GeneralPeriodic maintenance will ensure accurate and consistent analysis results. Verify the level of the calibration and reagent solutions on a regular basis and refill if necessary.

9.2 Cleaning overview

9.2.1 General cleaning process

Figure 15 Cleaning overview

Page 60: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

58

Maintenance

9.2.2 Chemical cleaning procedure

Figure 16 Chemical cleaning procedure

Page 61: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

59

Maintenance

9.3 Battery replacementThe transmitter is delivered with a CR1220 (3V) lithium battery included to safeguard the time function in the analyzer when not powered up. Should this internal battery malfunction or require replacement it must be replaced with exactly the same battery type CR1220 (3V) UL Certified (compulsory VARTA, RENATA or MATSUSHITA).

WARNINGThere is a risk of explosion if the replacement battery does not correspond to the correct model as indicated above. Do not throw away used batteries, but send them to a local recycling center.

9.4 Fuse replacementThe following illustration shows the position of the fuse, which is located next to the mains connectors on the power supply board under the shielding plate.

Replacement fuses are supplied with the analyzer and these must be used to replace faulty or blown fuses. If for any reason these replacements are not available, only the following alternatives should be used:

• JDYX2, LITEFUSE, type 213, rated 2 AT, 250 VAC

• JDYX2, SIBA, Type 179120, rated 2 AT, 250 VAC

• JDYX2, SCHURTER, Type FST, rated 2 AT, 250 VAC

If there is any difficulty in obtaining these replacement fuses, please contact your local Hach Lange representative for advice and assistance.

Figure 17 Fuse location

Page 62: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

60

Maintenance

9.5 Yearly maintenanceOnce a year, maintenance should be carried out on the analyzer to ensure the high degree of accuracy of the analyzer is maintained. This maintenance requires the following maintenance kit which is available from your local Hach Lange representative:

The following step-by-step process should be followed:

Part No. Description

09210=A=8000 2 years spare part kit 921X (all ranges). Includes items and quantities below 8 x 589=010=015 Fitting for 1.5mm I.D tubing - 10/32 UNF thread8 x 590=050=060 Polyethylene tubing 1.6 x 3.2mm (per meter) 8 x 625=065=025 Printed stickers for tubing identification 1 x 09210=A=8050 Kit of (8) ferrules, (8) nuts8 x 09210=A=4362 Injectors 2 x 226=003=013 Teflon coated magnetic stirrer (for 921X analyzers)0.5 x 151400,22387 PE tubing 6 x 8mm (per meter)1.0 x 151575,00006 PE tubing 4 x 6mm (per meter)

1. Ensure that adequate warning has been given to all personnel that are involved in any way with the analyzer, that maintenance will be carried out on a pre-defined date/time.

2. Shut down the analyzer as follows:

a. Stop the analyzer measuring process by pressing the Stop function key

b. Press the Menu function key and select the MAINTENANCE optionc. Select the EXTENDED STOP optiond. Fill a beaker with 250 mL of demineralized watere. Open the analyzer to gain access to the reagent and calibration solution canistersf. Unscrew the caps on the 4 reagent canisters, remove the tube guides from the

canisters and place them in the beaker of demineralized waterg. Unscrew the cap on the calibration solution canister, remove the tube guide from

the canister and place it in a second (empty) beaker. Do not put it in the beaker of demineralized water containing the reagent tube guides

h. Press OK to start the rinsing processi. On completion press Esc to exit the process. The measuring cell and reagent

tubes will now be rinsed and emptyj. Disconnect the analyzer from the power source

3. With the analyzer still open, remove the 4 screws holding the pump box cover in place (item no. 8 in Figure 3 on page 16) and set to one side.

Note: The pump connections are described on the pump box cover. Each pump has an inlet tubing (from the canister to the pump) and an outlet tubing (from the pump to the measuring cell). The outlet tubing is also referred to as the injector.

4. Prepare 4 of the injectors (part no. 09210=A=4362 from the maintenance kit). The process is slightly different depending on the analyzer model in use (sulfuric acid or nitric acid chemistry).

9120 analyzer with sulfuric acid chemistry:

a. The injector is supplied with the ferrule and fitting nut attached. These are not required, so cut the tubing immediately after the ferule (90° angle) and discard the ferrule and nut. Do this on all 4 injectors

9120 analyzer with nitric acid chemistry:

a. No action is required and the tubing injectors should be used as supplied in the maintenance kit

Page 63: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

61

Maintenance

5. Prepare the inlet tubing (part no. 590=050=060 from the maintenance kit). Again, the process is slightly different depending on the analyzer model in use (sulfuric acid or nitric acid chemistry).

9120 analyzer with sulfuric acid chemistry:

a. Cut 4 lengths of tubing of 90 cm each and proceed to the step 6

9120 analyzer with nitric acid chemistry:

a. Using a scalpel, cut one end of the tubing to form a point approximately 30mm long, to enable the tube to be passed through the ferrule:

WARNINGAlways take care when using scalpels. Always make tube cuts away from the body and keep fingers away from the blade.

b. Slide the fitting nut onto the tube. Slide the ferrule over the pointed end and onto the tube as far as possible without forcing it, and ensuring the PTFE seal is facing towards the pointed end of the tube:

c. With the aid of long-nosed pliers or similar, pull the pointed tube end through the ferrule until the PTFE sealing surface has reached the uncut section of the tube. Keeping the ferrule as perpendicular as possible to the tube will ensure the best results. Rotate the ferrule around the tube 3 or 4 times to seat the ferrule on the tube correctly:

d. Using a scalpel, cut the pointed tube end as close to the ferrule as possible, ensuring the end of the ferrule is not cut:

Page 64: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

62

Maintenance

e. The tube assembly is now ready for use:

6. Renew all the reagent tubing.

Important Note: To avoid any contamination, ensure that each tube guide is always used for the same reagent as it was previously used for. Under no circumstances should they be interchanged.

Apply the following procedure for one reagent, then repeat for each reagent in turn. For easier access, start with the reagent pump on the far left (closest to the calibration pump).

a. Locate the inlet port (nearest to you) and outlet port (farthest from you) of the pump. Be aware that there is an additional port on each pump between the inlet and outlet ports but which is not to be used. Disconnect the inlet tubing and the injector tubing from the pump. Pull the injector tubing out of its position in the photometric cell (item no. 5 in Figure 2 on page 15) and discard it

b. Mount the newly prepared injector on the outlet port of the pump and screw in place. Slide the other end of the tubing through the panel cut-out at the front but do not yet connect to the photometric cell

c. Mount the newly prepared inlet tubing on the inlet port of the pump and screw in place

d. Retrieve the old inlet tubing and the tube guide from the beaker of watere. Unscrew the plastic fitting from the weight (position 1 in the illustration below) and

put the weight to one side on a clean tissue

f. Once the weight is removed, pull all the old reagent tubing out of the tube guide and discard

g. Slide the newly prepared reagent tubing through tube guide h. Place a new fitting (part no. 589=010=015 from the maintenance kit) on the end of

the tubing and screw into the weighti. Take the same sticker as the original one (part no. 625=065=025 from the

maintenance kit) and place it around the reagent tube approximately 10cm from the tube guide

j. Place the tube guide assembly back in its canister and screw down securelyk. Repeat these steps for the next reagent pump until all four have been completed

7. Change the magnetic stirrer in the photometer cell by first removing the 4 screws from the photometric cell cover. Put the screws to one side. Lift off the cap and then remove the old magnetic stirrer in the bottom of the cell.

Take a new stirrer (226=003=013 from the maintenance kit) and place in position at the bottom of the cell.

Replace the photometric cell cover and secure with the 4 screws.8. Place the ends of the injectors into the top of the photometric cell as far as they will go

without forcing them.

Page 65: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

63

Maintenance

9.6 Maintenance menu overviewThere are a number of maintenance and diagnostic procedures that can be performed using the analyzer software.

Select the MAINTENANCE option from the main menu.

Note: If a PROGRAMMING password has been set (see Code on page 38 for details on how to do this), you will be required to enter a password to gain access to the main menu.

9. Re-connect power to the analyzer. Once the measurement screen is displayed, select the Menu option and then MAINTENANCE. Then select the REAGENTS REFILL option and then the TUBES PRIMING option. This will purge and refill the new calibration and reagent tubing, and will take about 20 minutes to complete.

At the end of the process press Esc to return to the menu at which stage the new reagent tubes should be full of reagent.

10. Return to the measurement screen and press the Start function key to begin measurements. Leave the analyzer to run for a period of two hours after which the measurements will have stabilized.

11. Once stable, press the Stop function key to stop measurements. Then select the Menu option and then CALIBRATION. Perform a manual ZERO+SLOPE calibration.

After a successful calibration the analyzer is ready to resume measurements. Return to the measurement screen and press the Start function key.

Page 66: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

64

Maintenance

9.7 Startup

9.8 Reagents refill

The analyzer monitors the reagent and calibration solution consumption. Messages are displayed when the solution levels are low.

This option is used after the analyzer has not been used for an extended period of time, or when starting the analyzer for the very first time.

Using a screwdriver, set the flow rate for the first sample using the adjustment screws on the front panel (item 4 in Figure 2 on page 15). Regulate the flow (between 5 and 30 L/hour) to ensure a slight overflow into the overflow vessel.

Note: Do not completely unscrew the adjustment screw as it has no stop and may provoke a sample leak.

Select OK to continue to the next sample channel.

When all channels have been set, the pumps are primed. Check visually that the tubes are full and then press Enter to return to the menu.

Each time the reagent and calibration solutions are replenished, access this menu for:

ADJUST VOLUMES: the canister volumes are displayed and can be modified.

TUBES PRIMING: start the priming / renewal procedure of the reagents in the tubes.

BOTTLES FULL: choose this option each time the canisters have been refilled. The volumes are automatically reset to 2 liters.

If ADJUST VOLUMES has been selected in the menu above, the current volumes of the reagent canisters and the calibration solution are displayed.

These can be individually modified.

Message Reason Relays set

Reagent < 250ml Low reagent level (range < 10 days) Warning (Re 11)

Cal. Sol < 300ml Calibration low level (range < 2 calibration) Warning (Re 11)

Short reagent Reagent canister volume < 150ml System (Re 12)

Short cal. Calibration solution canister volume < 150ml

Page 67: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

65

Maintenance

9.9 Extended stop

9.10 Tube cleaning

In case of an extended stop (longer than 1 week) it is recommended to clean the tubes and the measuring cell.

Have a beaker of 250 ml of demineralized water available and then do the following:

• Stop the analyzer by pressing Stop

• Press the Menu key

• Select the MAINTENANCE menu

• Select the EXTENDED STOP option

• Submerge all the reagent tubes in the beaker of demineralized water

• Press OK to start each stage of the procedure

• Follow the instructions on screen and validate each stage with OK.

At the end of the process, the measuring cell is empty. Disconnect the power from the analyzer and replace the reagent tubes in their canisters.

Note: Do not interchange the tubes. They must be returned to the exact same canister they were taken from.

In case of an extended stop of over 6 weeks, you must discard the reagents and prepare another set for the restart.

If the tubes become clogged, use this procedure to clean them.

Have a beaker available containing 250ml of demineralized water and a second beaker containing 100ml of ammonia at 10% of normal commercial concentrations by volume i.e. approximately 3% V/V.

Immerse tubing R1A in the ammonia solution. Press OK to start the cleaning process.

Follow the instructions and press OK at each step.

The procedure is the following:

• Rinse the tube with demineralized water

• Clean with a 10% ammonia solution

• Rinse the tube again with demineralized water

On completion, press Esc to return to the previous screen.

Page 68: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

66

Maintenance

9.11 Photometer test

9.12 Pumps - others

9.13 Pumps flowrate

With this option, technicians can verify the correct functioning of the photometer. During the process, the mixer is operating and the transfer valve is open.

Displayed parameters:

• Dark current: Measurement value which corresponds to the LED emission.

• Reference: Reference signal after LED current adjustment.

• Led current: Current of the LED.

• Measure: Continuous photometric measurement signal.

• Absorbance: Absorbance in milli-Abs.

Function keys:

• bck: Begins the measurement which corresponds to switching off the LED.

• ref: Allows adjusting the LED current and to measure the reference voltage.

• meas: Starts the continuous photometric measurement.

This option verifies the correct functioning of the pumps, the mixer, the drain and the heater.

Use the Select key to verify an element and press the On key to operate it or Off key to stop it.

A label is affixed to each pump indicating the flow rate which has been factory defined.

Enter this value into the respective field:

• P1 - Reagents R1A and R1M

• P2 - Oxalic acid

• P3 - Reducer

• P4 - Calibration pump

Note: The information for the calibration pump is used to determine the consumption of calibration solution, so that an alarm can be triggered when the solution gets low.

Page 69: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

67

Maintenance

9.14 Solenoid valves

9.15 Relays

9.16 Logical inputs

This option verifies the correct functioning of the solenoid valves.

Use the Select key to verify a valve and press the On key to operate it or Off key to stop it.

Press Esc to return to the previous screen.

• Solenoid valves 01 to 06: Measuring channels 1 to 6.

• Solenoid valve 07: Sample transfer from overflow vessel to the measuring cell.

This option operates or stops the relays.

Use the Select key to select a relay and press the On key to operate it or Off key to stop it.

Note: For each relay function, refer to Table 1 on page 28.

This option displays the status of the logical inputs and verifies the functionality of the inputs In1 to In12.

Status: 1 = short circuit input

* = open input

Page 70: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

68

Maintenance

9.17 Preparation of the tubes for insertion in the fast connectors

Mark where the groove is to be located, using the following table:

1. Place the tube in the cutter with the cutting wheel on the mark.

2. Tighten the tool a quarter of a turn.

3. Rotate the tool around the tube and check a small groove has been cut all round the tube.

4. Remove the tube from the cutter.

Figure 18 Groove location

Tube length 6 mm 1/4'' 5/16''

L 7.3 ± 0.5 0.33 ± 0.02 0.36 ± 0.02

Figure 19 Make a groove in the tube

Page 71: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

69

Section 10 Troubleshooting

10.1 Possible sources of the problemThis list gives you an idea of where to look for problem solving if you are having problems with your analyzer.

10.2 Possible solutions

Problem Possible cause Problem N°

The system does not run Start / Stop 1

The measurement is not stable, too high or too low

SampleTubesPhotometerValvesPump

53842

Incorrect calibration

PumpTubesValvesCalibration solution Calibration programReagentsPhotometer

2346778

The photometer does not run or measure correctly Photometer 8

Problem Symptom Solution

1. START / STOP

Analyzer stoppedSystem alarm: Rectify the problem according to the message displayed (see Table 2 on page 46).

Start up the analyzer by pressing the Start key.

Mains power cut offConnect the analyzer to the mains power.

Check the fuse.

2. PUMP

Does not run

Test the pump. In the MAINTENANCE / PUMP - OTHERS menu the number of pulses is automatically set to 1,000. Start each pump one after the other and verify that the volume obtained per pump is between 2.5-3 mL (corresponding to 2.5-3 μL/pulse).

Change the pump.

IrregularThe analyzer may have been stopped without cleaning the tubes. Using a syringe, flush the pump with clean water.

Change the pump.

3. TUBESDefective tube

Change the tube.Clogged tube

4. SOLENOID VALVES

Solenoid valves leak Remove, test and clean.

The solenoid valves can't open Stuck membrane; cleaning required.

The solenoid valves can't close Blocked outlet; cleaning required.

5. SAMPLEDirty Disconnect the sample lines from the sample block.

Diminution of the flow Verify and adjust the flow (10 L/h), change the sample tube if it is clogged.

Page 72: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

70

Troubleshooting

10.3 Typical measurement values during calibrationThese values are indicative and depend on the ambient temperature. The real values may deviate by more than one volt without having any effect on the measurements.

6. CALIBRATION

Wrong configuration of the calibration solution concentration

Reprogram the calibration solution concentration (SiO2).

Wrong concentration of the standard solution Prepare a new standard solution.

Wrong laboratory analysis of the calibration solution concentration

Verify the SiO2 concentration of the laboratory calibration solution.

Wrong chemical products or old or contaminated Wrong purity? New supplier? Clean the reagent bottles.

7. REAGENTS

Reagent canisters connected in the wrong order.

Verify the flow of the circuit from the MAINTENANCE / PUMPS-OTHERS menu.

Reagent tubes connected in a wrong order. Verify the flow diagram.

Wrong concentration Prepare a new batch of reagents.

Wrong chemical products or old or contaminated Wrong purity? New supplier?

Wrong proportion of reagent Check the pumps.

Bacterial development in reagent R3 (greenish deposit) Add 40 ml of sodium benzoate at 1 g/L.

8. PHOTOMETER Wrong or irregular measurement

Verify the photometer in MAINTENANCE / PHOTOMETER TEST. In this menu measurement values of reference and measurement during the last calibration of the measurement cycle appear. The typical values of these are in the tables on page 70.

Is there an error message?

Version 0 - 1000 ppb

Sample concentration Reference (mV) Measure (mV)

107 ppb 3500 2515

214.9 ppb 3500 1754

267.4 ppb 3500 1472

534.8 ppb 3500 603

Version 0 - 5000 ppb

Sample concentration Reference (mV) Measure (mV)

1000 ppb 3500 1400

3000 ppb 3500 300

Page 73: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

71

Section 11 Spare Parts

Accessories

09125=A=1485 Kit for Profibus DP with board for 91xx / 92xx and Operator Manual09125=A=2485 Optional RS485 JBUS/MODBUS for transmitters model 91XX after February 2004 (CPU-ph2) or 92XX

after June 2006 (CPU-ph2): board + instruction manual (5 languages)09210=A=8000 2 years spare part kit 921X (all ranges). Includes items and quantities below

8 x 589=010=015 Fitting for 1.5mm I.D tubing - 10/32 UNF thread8 x 590=050=060 Polyethylene tubing 1.6 x 3.2mm (per meter) 8 x 625=065=025 Printed stickers for tubing identification 1 x 09210=A=8050 Kit of (8) ferrules, (8) nuts8 x 09210=A=4362 Injectors 2 x 226=003=013 Teflon coated magnetic stirrer (for 921X analyzers)0.5 x 151400,22387 PE tubing 6 x 8mm (per meter)1.0 x 151575,00006 PE tubing 4 x 6mm (per meter)

09210=A=8012 9210-ph2 Analyzer tech spare part kit. Includes items and quantities below:1 x 09210=A=0250 Sample level sensor for 921X1 x 09210=C=7010 9210 set of dry chemicals for 50 days operation1 x 09210=A=8072 Kit of 5 empty canisters w/caps & stickers for 9210-ph2 silica1 x 689=118=008 Sampling solenoid valve, NPS 0.8 (0.2 - 6 bars)1 x 689=118=024 Sampling electrovalve (overflow vessel)1 x 695=004=004 Calibration / flush pump1 x 09210=A=8040 micro-dosing diaphragm self-priming pump

09210=A=8072 Kit of 5 canisters w/caps & stickers for 9210-ph2 silica (for dry chemicals see 09210=C=7010) 09210=C=7010 9210 set of dry chemicals for 50 days operation (applicable for all 9210 generation)G-MC-9210 GOLD-Contract, Maintenance & Calibration, Analyzer Silica (SiO2), 9210P-MC-9210 PLATINUM-Contract, Maintenance & Calibration, Analyzer Silica (SiO2), 9210S-MC-9210 SILVER-Contract, Maintenance & Calibration, Analyzer Silica (SiO2), 9210MC-9210 Maintenance & Calibration, Silica (SiO2) analyzer, 9095/9097/9210

Spare parts in contact with sample

09210=A=0100 0.5 - 6 bars adaptation kit for 6 solenoid sample valves09210=A=0105 Sampling flow-rate adjustment screw (w/ its O-ring)09210=A=0110 Metric to imperial conversion assembly (for one channel) from connector 6 mm O.D. to 1/4'' O.D. acetal09210=A=0111 921X sample valve kit (uneven to even number)09210=A=0121 Complete 1 channel block for 921X09210=A=0122 Complete 2 channels block for 921X09210=A=0143 Complete 3 channels block for 921X09210=A=0144 Complete 4 channels block for 921X09210=A=0165 Complete 5 channels block for 921X09210=A=0166 Complete 6 channels block for 921X09210=A=0250 Sample level sensor for 921X356099,05040 Viton O-ring 4 x 1 mm of the sampling flow-rate adjustment screw.356099,35380 Viton O-ring 38 x 2 mm (around level detector)359110,52012 Tubing silicone 12 x 17 mm (per meter) - from 8810 vessel or 921x, 924x column to drain -689=118=008 Solenoid valve (24CC, 2/2way) as per 9210&9211 sample blocks [NPS 0.8 (0.2 - 6 bars)]. Delivered with

one connector P/N 350=005=377. For mounting, see special torx screw driver at P/N 820=000=006689=118=016 Solenoid valve (24CC, 2/2way, NPS 1.6 (0.2 - 2 bars)). Delivered with one connector P/N 350=005=377.

For mounting, see special torx screw driver at P/N 820=000=006689=118=024 Solenoid valve (24CC, 2/2way) as per 9210&9211 overflow vessel (AKA transfer valve). For connector,

see P/N 350=005=377. For mounting, see special torx screw driver at P/N 820=000=006

Spare parts in contact with cell or electrodes

09097=C=0560 Heating element09210=A=0300 Complete photometer cell for 921X09210=A=0400 Complete mixer for 921X analyzers09210=A=0500 Hollow fiber for 921X

Page 74: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

72

Spare Parts

09210=C=0310 PVC square cover for 921X photometric cell09210=C=0330 Spacer ring for 921X mirror09210=C=0340 Mirror for 921X analyzers 226=003=013 Teflon coated magnetic stirrer (for 921X analyzers)356099,05220 O-ring Viton 22 x 1 mm (around mirror)356099,15040 Viton O-ring for use on 8873 temperature sensor, 8362 pH-probe and 921X hollow fiber359000,40780 Pt100 temperature sensor431=100=210 Colorimetric cell nut PG21 - Brass coated nickel

Spare parts in contact with reagents or standard solution

08891=A=0040 Cap for 9210 5L canister with hole for reagent tubing09210=A=0600 Weight with filtration-grid for 921X reagent canisters09210=A=7100 Set of 4 stickers 4 for 9210 canisters "Declaration of content". Stickers only.09210=A=8040 Kit micro-dosing diaphragm pump, self priming, with tubing-connectors-screws ready for mounting on

9210-ph209210=A=8080 Tube guide including weight and filtration grid, for 2 liter canister of 9210, 9211, (for chemicals see

09210=C=7000, for canister-kit see 09210=A=8070, for reagent-tubing see 590=050=060)09210=C=0005 Canister holder for 921x (for canister and stickers, see 0921x=A=8070)09210=C=0630 Filtration-grid for weight of 8810 or 921X reagent canisters09210=C=7000 9210 set of dry chemicals for 45 days operation09210=C=7010 9210 set of dry chemicals for 50 days operation (applicable for all 9210 generation)148644,05125 Fitting - nipple type - 10/32 UNF for 1/8" or 4 mm tubing151065,08699 PTFE tubing Ø 0.8 x 1.6 mm (per meter)151575,00006 Tubing in PE (low density), Ø 4 x 6 mm (per meter)495=020=001 2 liter PE canister with plain cap584=024=024 Union fitting PP for 4 mm tubing589=010=015 Fitting for 1.7 mm I.D. tubing – 10/32’’ thread589=011=040 Fitting for 4 mm I.D. tubing – 10/32’’ UNF thread in PP590=050=060 Tubing in PE (low density), Ø 1.6 x 3.2 mm (per meter)695=004=004 Pump 24 VDC, membrane, 0.1 L/mn, for 921x calibration/flush pump

Electronics spare parts

09200=A=1001 CPU board with display for Monec E9200 09200=A=1010 CPU board with display for Monec E9200, loaded with 9210 latest software09200=A=2000 Power supply board for Monec E920009200=A=5000 User connections / Interface board for 9210 & 921109200=A=5500 921X Internal bus communication module09210=A=1234 Connection board for 921X mixer/heater09210=A=1500 Colorimetric measurement module for (0-1K_SiO2/ 0-5_PO4) applications.09210=A=1510 Colorimetric measurement module for (0-5K_SiO2) applications.09210=A=1520 Colorimetric measurement module for (0-50_PO4) application.425=210=022 Cable gland PG21 PP grey equipped with anti-bend protection for diameter sizes 13 to 18 mm

Additional hardware

820=000=006 Elbow torx screw driver T6

Documentation

221=092=010 Operator manual in French for POLYMETRON 9210 Silica, A4 format221=192=010 Operator manual in English for POLYMETRON 9210 Silica, A4 format221=292=010 Operator manual in German for POLYMETRON 9210 Silica, A4 format221=492=010 Operator manual in Italian for POLYMETRON 9210 Silica, A4 format221=592=010 Operator manual in Spanish for POLYMETRON 9210 Silica, A4 format

Page 75: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

73

Section 12 Default Configuration

User configuration

Parameter Default value Possible values Selected values Date

CALIBRATION / PROGRAMMING

Offset interval 24 hours 1 - 999 hours

Cal. sol 214 ppb ppb / ppm

Auto. cal No Yes / No

Base Day Day / Hour

Week SMTWTFS SMTWTFS

Hour 15:00:00 24 Hour clock

MAINTENANCE / REAGENTS REFILL / ADJUST VOLUMES

Reagent 1 2.00 L 0 - 5 L

Reagent 2 2.00 L 0 - 5 L

Reagent 3 2.00 L 0 - 5 L

Calibration solution 2.00 L 0 - 5 L

MEASURE / MEASURING MODE

Cycle In loop Interval / In loop

Interval 15 mins 1 - 999 mins

ALARMS / ALARM 1

Mode LimitLimitActive channelLevel sample

Affect Channel 1 Channel 1-6

Limit 1000 ppb

Direction Up Up / Down

Hysteresis 0% 0 - 10%

Relay N.O. N.O. / N.C.

ALARMS / ALARM 2

Mode LimitLimitActive channelLevel sample

Affect Channel 2 Channel 1-6

Limit 1000 ppb

Direction Up Up / Down

Hysteresis 0% 0 - 10%

Relay N.O. N.O. / N.C.

ALARMS / ALARM 3

Mode LimitLimitActive channelLevel sample

Affect Channel 3 Channel 1-6

Limit 1000 ppb

Direction Up Up / Down

Hysteresis 0% 0 - 10%

Relay N.O. N.O. / N.C.

Page 76: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

74

Default Configuration

ALARMS / ALARM 4

Mode LimitLimitActive channelLevel sample

Affect Channel 4 Channel 1-6

Limit 1000 ppb

Direction Up Up / Down

Hysteresis 0% 0 - 10%

Relay N.O. N.O. / N.C.

ALARMS / ALARM 5

Mode LimitLimitActive channelLevel sample

Affect Channel 5 Channel 1-6

Limit 1000 ppb

Direction Up Up / Down

Hysteresis 0% 0 - 10%

Relay N.O. N.O. / N.C.

ALARMS / ALARM 6

Mode LimitLimitActive channelLevel sample

Affect Channel 6 Channel 1-6

Limit 1000 ppb

Direction Up Up / Down

Hysteresis 0% 0 - 10%

Relay N.O. N.O. / N.C.

ALARMS / WARNING ALARM

Alarm Yes Yes / No

Accept Manual Manual / Auto.

Relay N.O. N.O. / N.C.

ALARMS / SYSTEM ALARM

Alarm Yes Yes / No

Relay N.O. N.O. / N.C.

mA OUTPUTS / AFFECT / OUTPUT 1

Affect Channel 1 Channel 1 - 6

Type 0-20 mA 0/20 mA / 4/20 mA

Mode Linear Linear / dual

Low 0 ppb

Middle 100 ppb

Up 1000 ppb

User configuration

Page 77: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

75

Default Configuration

mA OUTPUTS / AFFECT / OUTPUT 2

Affect Channel 2 Channel 1 - 6

Type 0-20 mA 0/20 mA / 4/20 mA

Mode Linear Linear / dual

Low 0 ppb

Middle 100 ppb

Up 1000 ppb

mA OUTPUTS / AFFECT / OUTPUT 3

Affect Channel 3 Channel 1 - 6

Type 0-20 mA 0/20 mA / 4/20 mA

Mode Linear Linear / dual

Low 0 ppb

Middle 100 ppb

Up 1000 ppb

mA OUTPUTS / AFFECT / OUTPUT 4

Affect Channel 4 Channel 1 - 6

Type 0-20 mA 0/20 mA / 4/20 mA

Mode Linear Linear / dual

Low 0 ppb

Middle 100 ppb

Up 1000 ppb

mA OUTPUTS / AFFECT / OUTPUT 5

Affect Channel 5 Channel 1 - 6

Type 0-20 mA 0/20 mA / 4/20 mA

Mode Linear Linear / dual

Low 0 ppb

Middle 100 ppb

Up 1000 ppb

mA OUTPUTS / AFFECT / OUTPUT 6

Affect Channel 6 Channel 1 - 6

Type 0-20 mA 0/20 mA / 4/20 mA

Mode Linear Linear / dual

Low 0 ppb

Middle 100 ppb

Up 1000 ppb

mA OUTPUTS / SPECIAL PROG / MAINTENANCE

Mode Last Last / Preset / Live

Preset 0 mA

mA OUTPUTS / SPECIAL PROG / CALIBRATION

Mode Last Last / Preset / Live

Preset 0 mA

mA OUTPUTS / SPECIAL PROG / GRAB SAMPLE

Mode Last Last / Preset / Live

Preset 0 mA

User configuration

Page 78: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

76

Default Configuration

mA OUTPUTS / SPECIAL PROG / SYSTEM ALARM

Mode Last Last / Preset / Live

Preset 0 mA

mA OUTPUTS / SPECIAL PROG / WARNING ALARM

Mode Last Last / Preset / Live

Preset 0 mA

RS485

N° 01 0 - 33

Baud 19200 600 / 1200 / 2400 / 4800 / 9600

Parity No No / Odd / Even

Stop bit 1 1 or 2

Swap word No Yes / No

SEQUENCE / CHANNEL ACTIVATION

Channel 1 On On / Off

Channel 2 On On / Off

Channel 3 On On / Off

Channel 4 On On / Off

Channel 5 On On / Off

Channel 6 On On / Off

SEQUENCE / SEQUENCE

Sequence 123456123456 123456*

SERVICE/TIME

Day Monday Monday - Friday

Date 01/01/01

Time 09:25:55

SERVICE / DISPLAY

Language GB F/GB/D/SP/I

Concentration ppb - ppm ppb - ppmμg - mg/l

SERVICE / CUSTOMIZE

Ch 1 sample 1 Free format

Ch 2 sample 2 Free format

Ch 3 sample 3 Free format

Ch 4 sample 4 Free format

Ch 5 sample 5 Free format

Ch 6 sample 6 Free format

User configuration

Page 79: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

77

Section 13 Material Safety Data Sheets (MSDS)

13.1 Oxalic acid

Name:OXALIC ACID 1. dihydrate Formula: 1. (COOH)2. 2H2O

2. anhydrous Formula: 2. (COOH)2.CAS N°144-62-7

PHYSICAL PROPERTIES

MELTING POINT 1. 101°C2. 190 °C

SOLUBILITY IN WATER About 14%FIRE AND EXPLOSION HAZARD Decomposition products include carbon monoxide (highly flammable and toxic) and

carbon dioxide.HEALTH HAZARD Classification of risk - Harmful

Harmful in contact with skinHarmful if swallowedHarmful if inhaledMinimum lethal dose (LD50) = 275 mg/Kg

SAFETY PRECAUTIONS

EYES Wash thoroughly with water; obtain medical attentionLUNGS Remove from exposure, rest and keep warm; obtain medical attentionINGESTION Wash out thoroughly with water; give plenty of water to drink; obtain medical

attention

CHEMICAL HAZARDS AND STABILITY

STABILITY StableSPECIAL HAZARDS Mixtures with sodium chlorite can explode with the addition of water.WASTE DISPOSAL Dissolve or mix the material with a combustible solvent and burn in a chemical

incinerator equipped with an after burner and scrubber. Dispose in a safe manner consistent with Local Authority chemical waste regulations. For guidelines on disposal consult the requirements of the Control of Pollution Act (1974) and the Control of Pollution (Special Wastes) Regulations (1980 S1 1709).

PROTECTIVE MEASURES

RESPIRATOR Dust mask or positive pressure hood - if handling large quantities.GLOVES Rubber or plasticEYE PROTECTION Goggles

SPILLAGE HANDLING

SMALL SPILLAGE Mop up with plenty of water and run to waste, diluting with running water.LARGE SPILLAGE Transfer to salvage canister and treat residue as for small spillage.The information on this sheet is believed to be correct at the time of printing. The details should be reviewed periodically and should be checked for special uses.

Page 80: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

78

Material Safety Data Sheets (MSDS)

13.2 Sulfuric acid (98%)

Name:SULPHURIC ACID Formula: H2 SO4

Hazard class: 8 UN N°: 1830CAS N° 7664-93-9

PHYSICAL PROPERTIES

MELTING POINT 10 °CBOILING POINT 330°CSPECIFIC GRAVITY 1,8MISCIBLE WITH WATERFIRE AND EXPLOSION HAZARD May evolve toxic fumes in a fire.HEALTH HAZARD Classification of risk - Corrosive.

Causes severe burns to eyes and skin.If ingested causes severe internal irritation and damage. Dilute acid irritates the eyes and skin and can cause burns and dermatitis.Minimum lethal dose, D50 = 2140 mg/Kg, via oral route.Occupational exposure standard, long term exposure limit (8 hours TWA) mg-3

Keep out of reach of children.

SAFETY PRECAUTIONS-FIRST AID

EYES Wash immediately and thoroughly with plenty of water for at least 10 minutes. Obtain medical attention.

LUNGS Remove from exposure, rest and keep warm, or if exposure has been great, obtain medical attention.

MOUTH Wash out mouth thoroughly with water; give plenty of water to drink; obtain medical attention.

SKIN Drench the skin thoroughly with water and give plenty of water to drink. Obtain medical attention.

CHEMICAL HAZARDS AND STABILITY

Reacts vigorously with evolution of heat. Never add water to this product.Will react violently with the following:• Acetonitrile.• Perchlorates.• Nitro-organic compounds.• Alkali metals.• Potassium permanganate • Metal halogenates.Avoid contact with bases.

PROTECTIVE MEASURES

Wear chemical-resistant gloves (nitrile), safety goggles or face shield and use fume cupboard. For large quantities use approved breathing apparatus, plastic apron and sleeves and wear boots.Ensure that safety shower and eye bath are available in the locality.

SPILLAGE HANDLING Wear appropriate protective clothing.Cover the spillage with soda ash. If local regulations permit, cautiously add plenty of water and mop up while diluting with more water and run to waste. Wash site of spillage thoroughly with water. Otherwise cover with dry lime, sand or soda ash and place in covered canisters. Arrange removal by disposal company. Ventilate area and wash spill site with water after material removal is complete.If material has entered drains it may be necessary to inform the Local Authorities.

Page 81: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

79

Material Safety Data Sheets (MSDS)

WASTE DISPOSAL

Dispose in safe manner consistent with Local Authority chemical waste regulation. For guidelines on disposal consult the requirements of the Control Pollution Act (1974) and the Control of Pollution (special wastes) Regulations (1980 SI 1709).

The information on this sheet is believed to be correct at the time of printing. The details should be reviewed periodically and should be checked for special uses.

Page 82: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

80

Material Safety Data Sheets (MSDS)

13.3 Nitric acid (50-70%)

PRODUCT AND COMPANY IDENTIFICATION

SYNONYMSAqua Fortis; Azotic Acid; Nitric Acid 50%; Nitric Acid 65%; Nitric Acid 69 - 70%

CAS NO. 7697-37-2 MOLECULAR WEIGHT 63.01 CHEMICAL FORMULA HNO3

COMPOSITION / INFORMATION ABOUT THE COMPONENTS

INGREDIENT CAS NO PERCENT HAZARDOUSNitric AcidWater

7697-37-27732-18-5

50 - 70%30 - 50%

YesNo

IDENTIFICATION OF HAZARDS

EMERGENCY OVERVIEW POISON! DANGER! STRONG OXIDIZER. CONTACT WITH OTHER MATERIAL MAY CAUSE FIRE. CORROSIVE. LIQUID AND MIST CAUSE SEVERE BURNS TO ALL BODY TISSUE. MAY BE FATAL IF SWALLOWED OR INHALED. INHALATION MAY CAUSE LUNG AND TOOTH DAMAGE.

SAF-T-DATA(TM) RATINGS Health RatingFlammability RatingReactivity RatingContact RatingStorage Color CodeLab Protective Equipment

4 - Extreme (Poison)0 - None3 - Severe (Oxidizer)4 - Extreme (Corrosive)White (Corrosive)Goggles & shield; lab coat & apron; vent hood; proper gloves

INHALATION Corrosive! Inhalation of vapors can cause breathing difficulties and lead to pneumonia and pulmonary edema, which may be fatal. Other symptoms may include coughing, choking, and irritation of the nose, throat, and respiratory tract.

INGESTION Corrosive! Swallowing nitric acid can cause immediate pain and burns of the mouth, throat, esophagus and gastrointestinal tract.

SKIN CONTACT Corrosive! Can cause redness, pain, and severe skin burns. Concentrated solutions cause deep ulcers and stain skin a yellow or yellow-brown color.

EYE CONTACT Corrosive! Vapors are irritating and may cause damage to the eyes. Contact may cause severe burns and permanent eye damage.

CHRONIC EXPOSURE Long-term exposure to concentrated vapors may cause erosion of teeth and lung damage. Long-term exposures seldom occur due to the corrosive properties of the acid.

AGGRAVATION OF PRE-EXISTING CONDITIONS

Persons with pre-existing skin disorders, eye disease, or cardiopulmonary diseases may be more susceptible to the effects of this substance.

FIRST AID

INHALATION Remove to fresh air. If not breathing, give artificial respiration. If breathing is difficult, give oxygen. Call a physician.

INGESTION DO NOT INDUCE VOMITING! Give large quantities of water or milk if available. Never give anything by mouth to an unconscious person. Get medical attention immediately.

SKIN CONTACT In case of contact, immediately flush skin with plenty of water for at least 15 minutes while removing contaminated clothing and shoes. Wash clothing before reuse. Thoroughly clean shoes before reuse. Get medical attention immediately.

EYE CONTACT Immediately flush eyes with plenty of water for at least 15 minutes, lifting lower and upper eyelids occasionally. Get medical attention immediately.

Page 83: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

81

Material Safety Data Sheets (MSDS)

FIRE FIGHTING MEASURES

FIRE Not combustible, but substance is a strong oxidizer and its heat of reaction with reducing agents or combustibles may cause ignition. Can react with metals to release flammable hydrogen gas.

EXPLOSION Reacts explosively with combustible organic or readily oxidizable materials such as: alcohols, turpentine, charcoal, organic refuse, metal powder, hydrogen sulfide, etc. Reacts with most metals to release hydrogen gas which can form explosive mixtures with air.

FIRE EXTINGUISHING MEDIA Water spray may be used to keep fire exposed containers cool. Do not get water inside container.

SPECIAL INFORMATION Increases the flammability of combustible, organic and readily oxidizable materials. In the event of a fire, wear full protective clothing and NIOSH-approved self-contained breathing apparatus with full face piece operated in the pressure demand or other positive pressure mode.

MEASURES IN CASE OF SPILLAGE

Ventilate area of leak or spill. Wear appropriate personal protective equipment as specified in Section 8. Isolate hazard area. Keep unnecessary and unprotected personnel from entering. Contain and recover liquid when possible. Neutralize with alkaline material (soda ash, lime), then absorb with an inert material (e.g. vermiculite, dry sand, earth), and place in a chemical waste container. Do not use combustible materials, such as saw dust. Do not flush to sewer! US Regulations (CERCLA) require reporting spills and releases to soil, water and air in excess of reportable quantities.

HANDLING AND STORAGE

Store in a cool, dry, ventilated storage area with acid resistant floors and good drainage. Protect from physical damage. Keep out of direct sunlight and away from heat, water, and incompatible materials. Do not wash out container and use it for other purposes. When diluting, the acid should always be added slowly to water and in small amounts. Never use hot water and never add water to the acid. Water added to acid can cause uncontrolled boiling and splashing. Containers of this material may be hazardous when empty since they retain product residues (vapors, liquid); observe all warnings and precautions listed for the product.

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION

AIRBORNE EXPOSURE LIMITS OSHA Permissible Exposure Limit (PEL): 2 ppm (TWA), 4 ppm (STEL)

ACGIH Threshold Limit Value (TLV): 2 ppm (TWA); 4 ppm (STEL)

VENTILATION SYSTEM A system of local and/or general exhaust is recommended to keep employee exposures below the Airborne Exposure Limits. Local exhaust ventilation is generally preferred because it can control the emissions of the contaminant at its source, preventing dispersion of it into the general work area. Please refer to the ACGIH document, Industrial Ventilation, A Manual of Recommended Practices, most recent edition, for details.

PERSONAL RESPIRATORS (NIOSH APPROVED) If the exposure limit is exceeded, wear a supplied air, full face piece respirator, air-lined hood, or full face piece self-contained breathing apparatus. Nitric acid is an oxidizer and should not come in contact with cartridges and canisters that contain oxidizable materials, such as activated charcoal. Canister-type respirators using sorbents are ineffective.

SKIN PROTECTION Wear impervious protective clothing, including boots, gloves, lab coat, apron or coveralls, as appropriate, to prevent skin contact.

EYE PROTECTION Use chemical safety goggles and/or a full face shield where splashing is possible. Maintain eye wash fountain and quick-drench facilities in work area.

Page 84: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

82

Material Safety Data Sheets (MSDS)

PHYSICAL AND CHEMICAL PROPERTIES

APPEARANCE Colorless to yellowish liquid.ODOR Suffocating, acrid. SOLUBILITY Infinitely soluble.SPECIFIC GRAVITY 1.41PH 1.0 (0.1M solution)% VOLATILES BY VOLUME @ 21C (70F) 100 (as water and acid)BOILING POINT 122C (252F)MELTING POINT -42C (-44F)VAPOR DENSITY (AIR=1) 2-3 VAPOR PRESSURE (MM HG) 48 @ 20C (68F) EVAPORATION RATE (BUAC=1) No information found.

STABILITY AND REACTIVITY

STABILITY Stable under ordinary conditions of use and storage. Containers may burst when heated.

HAZARDOUS DECOMPOSITION PRODUCTS When heated to decomposition, emits toxic nitrogen oxides fumes and hydrogen nitrate. Will react with water or steam to produce heat and toxic and corrosive fumes.

HAZARDOUS POLYMERIZATION Will not occur.INCOMPATIBILITIES A dangerously powerful oxidizing agent, concentrated nitric

acid is incompatible with most substances, especially strong bases, metallic powders, carbides, hydrogen sulfide, turpentine, and combustible organics.

CONDITIONS TO AVOID Light and heat.

TOXICOLOGICAL INFORMATION

NITRIC ACIDInhalation rat LC50: 244 ppm (NO2)/30M; Investigated as a mutagen, reproductive effector. Oral (human) LDLo: 430 mg/kg.

NTP CARCINOGENINGREDIENT KNOWN ANTICIPATED IARC CATEGORYNitric Acid (7697-37-2) No No NoneWater (7732-18-5) No No None

ECOLOGICAL INFORMATION

ENVIRONMENTAL FATE No information foundENVIRONMENTAL TOXICITY No information found

DISPOSAL CONSIDERATIONS

Whatever cannot be saved for recovery or recycling should be managed in an appropriate and approved waste facility. Although not a listed RCRA hazardous waste, this material may exhibit one or more characteristics of a hazardous waste and require appropriate analysis to determine specific disposal requirements. Processing, use or contamination of this product may change the waste management options. State and local disposal regulations may differ from federal disposal regulations. Dispose of container and unused contents in accordance with federal, state and local requirements.

TRANSPORT INFORMATION

PROPER SHIPPING NAME Nitric acid (with not more than 70% nitric acid)HAZARD CLASS 8UN/NA UN2031PACKING GROUP IIINFORMATION REPORTED FOR PRODUCT/SIZE 6.5GL

Page 85: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

83

Material Safety Data Sheets (MSDS)

REGULATORY INFORMATION

LABELLING TO EEC DIRECTIVES EU-no: 007.004.00.1 EINECS/ELINCS-no: 231-714-2

HAZARDOUS PRODUCT IDENTIFICATION C Corrosive" R " Phrases R35 Causes burns" S " Phrases S26 In case of contact with eyes, rinse immediately with plenty of

water and seek medical adviceS36 Wear suitable protective clothingS45 In case of accident or if you feel unwell, seek medical advice

(show the label where possible)X1 Reserved for industrial and professional use.X11 Keep in a cool place, protect against sunlight.

OTHER INFORMATION

NFPA RATINGS Health: 3 Flammability: 0 Reactivity: 0 Other: Oxidizer

LABEL HAZARD WARNING POISON! DANGER! STRONG OXIDIZER. CONTACT WITH OTHER MATERIAL MAY CAUSE FIRE. CORROSIVE. LIQUID AND MIST CAUSE SEVERE BURNS TO ALL BODY TISSUE. MAY BE FATAL IF SWALLOWED OR INHALED. INHALATION MAY CAUSE LUNG AND TOOTH DAMAGE.

LABEL PRECAUTIONS Do not get in eyes, on skin, or on clothing.Do not breathe vapor or mist.Use only with adequate ventilation.Wash thoroughly after handling.Keep from contact with clothing and other combustible materials.Do not store near combustible materials.Store in a tightly closed container.Remove and wash contaminated clothing promptly.

LABEL FIRST AID In case of contact, immediately flush eyes or skin with plenty of water for at least 15 minutes while removing contaminated clothing and shoes. Wash clothing before reuse. If swallowed, DO NOT INDUCE VOMITING. Give large quantities of water. Never give anything by mouth to an unconscious person. If inhaled, remove to fresh air. If not breathing, give artificial respiration. If breathing is difficult, give oxygen. In all cases get medical attention immediately.

PRODUCT USE Laboratory Reagent. The information in this sheet is believed to be correct at the time of printing. The details should be reviewed periodically and should be checked for special uses.

Page 86: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

84

Material Safety Data Sheets (MSDS)

13.4 Ammonium hydroxide (10-35% NH3)

PRODUCT AND COMPANY IDENTIFICATION

SYNONYMS Ammonium hydroxide solutions; ammonia aqueous; ammonia solutions

CAS NO. 1336-21-6MOLECULAR WEIGHT 35.05CHEMICAL FORMULA NH4OH in H2O

COMPOSITION / INFORMATION ABOUT THE COMPONENTS

INGREDIENT CAS NO PERCENT HAZARDOUSAmmonium HydroxideWater

1336-21-67732-18-5

21 - 72%28 - 79%

YesNo

Contains between 10 and 35% ammonia.

IDENTIFICATION OF HAZARDS

EMERGENCY OVERVIEW POISON! DANGER! CORROSIVE. MAY BE FATAL IF SWALLOWED OR INHALED. MIST AND VAPOR CAUSE BURNS TO EVERY AREA OF CONTACT.

SAF-T-DATA(TM) RATINGS

Health RatingFlammability RatingReactivity RatingContact RatingStorage Color CodeLab Protective Equipment

3 - Severe (Poison)0 - None1 - Slight4 - Extreme (Corrosive)White Stripe (Store Separately)Goggles & shield; lab coat & apron; vent hood; proper gloves

INHALATIONVapors and mists cause irritation to the respiratory tract. Higher concentrations can cause burns, pulmonary edema and death. Brief exposure to 5000 ppm can be fatal.

INGESTION

Toxic! May cause corrosion to the esophagus and stomach with perforation and peritonitis. Symptoms may include pain in the mouth, chest, and abdomen, with coughing, vomiting and collapse. Ingestion of as little as 3-4 mL may be fatal.

SKIN CONTACT Causes irritation and burns to the skin.

EYE CONTACT Vapors cause irritation. Splashes cause severe pain, eye damage, and permanent blindness.

CHRONIC EXPOSURE Repeated exposure may cause damage to the tissues of the mucous membranes, upper respiratory tract, eyes and skin.

AGGRAVATION OF PRE-EXISTING CONDITIONS

Persons with pre-existing eye disorders or impaired respiratory function may be more susceptible to the effects of this material.

FIRST AID

INHALATION Remove to fresh air. If not breathing, give artificial respiration. If breathing is difficult, give oxygen. Call a physician.

INGESTION If swallowed, DO NOT INDUCE VOMITING. Give large quantities of water. Never give anything by mouth to an unconscious person. Get medical attention immediately.

SKIN CONTACT Immediately flush skin with plenty of water for at least 15 minutes while removing contaminated clothing and shoes. Call a physician, immediately. Wash clothing before reuse.

EYE CONTACT Immediately flush eyes with gentle but large stream of water for at least 15 minutes, lifting lower and upper eyelids occasionally. Call a physician immediately. Immediate action is critical to minimize possibility of blindness.

Page 87: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

85

Material Safety Data Sheets (MSDS)

FIRE FIGHTING MEASURES

FIRE Auto-ignition temperatureFlammable limits in air % by volume

651C (1204F)lel: 16; uel: 25

EXPLOSION Flammable vapors may accumulate in confined spaces.FIRE EXTINGUISHING MEDIA Use any means suitable for extinguishing surrounding fire.

Use water spray to blanket fire, cool fire exposed containers, and to flush non-ignited spills or vapors away from fire.

SPECIAL INFORMATION In the event of a fire, wear full protective clothing and NIOSH-approved self-contained breathing apparatus with full face piece operated in the pressure demand or other positive pressure mode.

MEASURES IN CASE OF SPILLAGE

Ventilate area of leak or spill. Keep unnecessary and unprotected people away from area of spill. Wear appropriate personal protective equipment as specified in Section 8. Contain and recover liquid when possible. Do not flush caustic residues to the sewer. Residues from spills can be diluted with water, neutralized with dilute acid such as acetic, hydrochloric or sulfuric. Absorb neutralized caustic residue on clay, vermiculite or other inert substance and package in a suitable container for disposal.

HANDLING AND STORAGE

Keep in a tightly closed container, stored in a cool, dry, ventilated area. Protect against physical damage. Separate from incompatibilities. Store below 25C. Protect from direct sunlight. Containers of this material may be hazardous when empty since they retain product residues (vapors, liquid); observe all warnings and precautions listed for the product.

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION

AIRBORNE EXPOSURE LIMITS OSHA Permissible Exposure Limit (PEL): 50 ppm (NH3)

ACGIH Threshold Limit Value (TLV): 25 ppm (NH3) (TWA) 35 ppm (STEL)

VENTILATION SYSTEM A system of local and/or general exhaust is recommended to keep employee exposures below the Airborne Exposure Limits. Local exhaust ventilation is generally preferred because it can control the emissions of the contaminant at its source, preventing dispersion of it into the general work area. Please refer to the ACGIH document, Industrial Ventilation, A Manual of Recommended Practices, most recent edition, for details.

PERSONAL RESPIRATORS (NIOSH APPROVED) If the exposure limit is exceeded and engineering controls are not feasible, a full face piece respirator with an ammonia/methylamine cartridge may be worn up to 50 times the exposure limit or the maximum use concentration specified by the appropriate regulatory agency or respirator supplier, whichever is lowest. For emergencies or instances where the exposure levels are not known, use a full face piece positive-pressure, air-supplied respirator. WARNING: Air purifying respirators do not protect workers in oxygen-deficient atmospheres.

SKIN PROTECTION Wear impervious protective clothing, including boots, gloves, lab coat, apron or coveralls, as appropriate, to prevent skin contact. Neoprene and nitrile rubber are recommended materials. Polyvinyl alcohol is not recommended.

EYE PROTECTION Use chemical safety goggles and/or full face shield where dusting or splashing of solutions is possible. Maintain eye wash fountain and quick-drench facilities in work area.

Page 88: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

86

Material Safety Data Sheets (MSDS)

PHYSICAL AND CHEMICAL PROPERTIES

APPEARANCE Clear, colorless solution.ODOR Ammonia odor.SOLUBILITY Infinitely soluble.SPECIFIC GRAVITY 0.9 (28% NH4OH)PH 13.8 (29% solution).% VOLATILES BY VOLUME @ 21C (70F) No information found.BOILING POINT ca. 36C (ca. 97F)MELTING POINT -72C (-98F)VAPOR DENSITY (AIR=1) 0.60 NH3VAPOR PRESSURE (MM HG) 115 @ 20C for 10% solution; 580 @ 20C for 28% solution.EVAPORATION RATE (BUAC=1) No information found.

STABILITY AND REACTIVITY

STABILITY Stable under ordinary conditions of use and storage.HAZARDOUS DECOMPOSITION PRODUCTS Burning may produce ammonia, nitrogen oxides.HAZARDOUS POLYMERIZATION Will not occur.INCOMPATIBILITIES Acids, acrolein, dimethyl sulfate, halogens, silver nitrate,

propylene oxide, nitromethane, silver oxide, silver permanganate, oleum, beta-propiolactone. Most common metals.

CONDITIONS TO AVOID Heat, sunlight, incompatibles, sources of ignition.

TOXICOLOGICAL INFORMATION

AMMONIUM HYDROXIDE Oral rat LD50: 350 mg/kg; eye, rabbit, standard Draize, 250 ug; severe, investigated as a mutagen.

AMMONIA Inhalation rat LC50: 2000 ppm/4-hr; investigated as a tumorigen, mutagen.

NTP CARCINOGENINGREDIENT KNOWN ANTICIPATED IARC CATEGORYAmmonium Hydroxide (1336-21-6) No No None

Water (7732-18-5) No No None

ECOLOGICAL INFORMATION

ENVIRONMENTAL FATE This material is not expected to significantly bioaccumulate.ENVIRONMENTAL TOXICITY 24 Hr LC50 rainbow trout: 0.008 mg/L;

96 Hr LC50 fathead minnow: 8.2 mg/L;48 Hr LC50 bluegill: 0.024 mg/L;48 Hr EC50 water flea: 0.66 mg/L

DISPOSAL CONSIDERATIONS

Whatever cannot be saved for recovery or recycling should be managed in an appropriate and approved waste facility. Although not a listed RCRA hazardous waste, this material may exhibit one or more characteristics of a hazardous waste and require appropriate analysis to determine specific disposal requirements. Processing, use or contamination of this product may change the waste management options. State and local disposal regulations may differ from federal disposal regulations. Dispose of container and unused contents in accordance with federal, state and local requirements.

TRANSPORT INFORMATION

PROPER SHIPPING NAME Ammonia solutions (with 10-35% ammonia)HAZARD CLASS 8UN/NA UN2672PACKING GROUP IIIINFORMATION REPORTED FOR PRODUCT/SIZE 385LB

Page 89: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

87

Material Safety Data Sheets (MSDS)

REGULATORY INFORMATION

LABELLING TO EEC DIRECTIVES EU-no: 007.001.01.2 HAZARDOUS PRODUCT IDENTIFICATION C Corrosive" R " Phrases R34, R50 Causes burns. Very toxic to aquatic organisms" S " Phrases S26 In case of contact with eyes, rinse immediately with plenty of

water and seek medical adviceS36/37/39 Wear suitable protective clothing, gloves, and eye/face

protectionS45 In case of accident or if you feel unwell, seek medical advice

(show the label where possible)S61 Avoid release to the environment

OTHER INFORMATION

NFPA RATINGS Health: 3 Flammability: 1 Reactivity: 0

LABEL HAZARD WARNING POISON! DANGER! CORROSIVE. MAY BE FATAL IF SWALLOWED OR INHALED. MIST AND VAPOR CAUSE BURNS TO EVERY AREA OF CONTACT.

LABEL PRECAUTIONS Do not get in eyes, on skin, or on clothing.Do not breathe vapor or mist.Keep container closed.Use only with adequate ventilation.Wash thoroughly after handling.

LABEL FIRST AID If swallowed, DO NOT INDUCE VOMITING. Give large quantities of water. Never give anything by mouth to an unconscious person. If inhaled, remove to fresh air. If not breathing, give artificial respiration. If breathing is difficult, give oxygen. In case of contact, immediately flush eyes or skin with plenty of water for at least 15 minutes while removing contaminated clothing and shoes. Wash clothing before reuse. IMMEDIATE ACTION IS ESSENTIAL FOR EYE EXPOSURES. In all cases call a physician immediately.

PRODUCT USE Laboratory Reagent. The information in this sheet is believed to be correct at the time of printing. The details should be reviewed periodically and should be checked for special uses.

Page 90: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

88

Material Safety Data Sheets (MSDS)

13.5 Sodium hexafluorosilicate

Name:SODIUM HEXAFLUOROSILICATE (sodium fluorosilicate, disodium hexafluorosilicate disodique)

Formula: Na2SiF6Hazard class: 6.1 No. UN: 2674

CAS N° 16893-85-9

PHYSICAL PROPERTIES

White granular powderDensity (20 °C) approx. 2,7 g cm-3

Slightly soluble in water (6g 1-1 to 20°C)FIRE AND EXPLOSION HAZARD When heated to decomposition it may emit hydrogen fluoride.HEALTH HAZARD Classification of risk - toxic.

Avoid contact with skin.Avoid contact with eyes.Keep locked up.Keep out of reach of children.Toxic by ingestion and skin contactToxic by inhalationIrritating to skin and eyes, may cause damage if exposure is prolonged.Occupational exposure standard, long term exposure limit (8h TWA) 2,5 mg m-3

SAFETY PRECAUTIONS - FIRST AID If the casualty feels unwell, seek medical advice and show label where possible.EYES Rinse immediately and thoroughly with water for at least 10 minutes. Obtain

medical attention.LUNGS Remove from exposure, rest and keep warm. In severe cases, or if exposure has

been great obtain medical attention.MOUTH Wash out mouth thoroughly with water and give plenty of water to drink. Obtain

medical attention.SKIN Wash skin thoroughly with water and give plenty of water to drink. Obtain medical

attention.CHEMICAL HAZARD AND STABILITY Contact with acids liberates toxic gas.

PROTECTIVE MEASURES (as appropriate to quantity handled)

RESPIRATOR Dust respirator.VENTILATION Extraction hood.GLOVES Rubber or plastic.EYE PROTECTION Goggles or face shield.OTHERS MEASURES Plastic apron, sleeves, boots - if handling large quantities.

SPILLAGE HANDLING

PRECAUTIONS Wear appropriate protective clothing.Mix with sand, transfer carefully to canister and arrange removal by disposal company. Wash site of spillage thoroughly with water and detergent.For large spillages liquids should be contained with sand or earth and both liquids and solids transferred to salvage canisters. Any residues should be treated as for small spillages.If material has entered surface drains it may be necessary to inform Local Authorities, including fire services flammable.

WASTE DISPOSAL Dispose in safe manner consistent with Local Authorities or national chemical Waste Regulation. For guidelines for disposal in the UK consult the requirements of the Control of Pollution (special wastes) regulations (1980 SI 1709).

The information in this sheet is believed to be correct at the time of printing. The details should be reviewed periodically and should be checked for special uses.

Page 91: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

89

Material Safety Data Sheets (MSDS)

13.6 Ferrous ammonium sulfate hexahydrate

PRODUCT IDENTIFICATION

PRODUCT NAME FERROUS AMMONIUM SULPHATE HEXAHYDRATECAS No. (CE) 7783-85-9 233-151-8FORMULA (NH4)2Fe(SO4)2, 6H2OMOLECULAR WEIGHT 392,14 AMUALTERNATIVE NAMES Ferrous ammonium sulfate hexahydrate

Ammonium ferrous sulfate, hexahydrate Iron ammonium sulfate hydrate Sulfuric acid, ammonium iron (2+) salt, hexahydrate

PHYSICAL PROPERTIES

FORM SolidCOULOUR Greenish blueODOUR NilMELTING POINT 100°CDENSITY 1.86 g/cm3

FIRE AND EXPLOSION HAZARD Non-combustible.Likelihood of dangerous vapors being given off if close to a fire.In fire situations it may form nitrogen oxides and sulphur oxides.

HEALTH RISKS

SPECIFIC ENVIRONMENTAL AND HEALTH HAZARDS Irritant for the eyes, respiratory tract and skin.

SAFETY MEASURES

EYES - SKIN Flush copiously with water.Remove any contaminated clothing

RESPIRATORY TRACT Avoid continued exposure, breathe in fresh air; seek medical attention.IF SWALLOWED Drink plenty of water, provoke vomiting, and seek medical advice.

CHEMICAL RISKS AND STABILITY

STABILITY Conditions to avoid - high temperaturesCONDITIONS TO AVOID Sensitive to air, sensitive to light. PRODUCTS TO AVOID Powerful oxidizing agents and acids.DANGEROUS DECOMPOSITION PRODUCTS

Iron oxidesAmmoniaSulphur oxides

PREVENTIVE MEASURES

HANDLING Safe Handling Instructions: Do not breathe the dust. Avoid contact with the eyes, skin or clothing. Avoid prolonged or repeated exposures.

EXPOSURE LIMITS UK:Source - category - valueOEL - LTEL - 1 mg (Fe)/m3

OEL - STEL - 2 mg (Fe)/m3

BREATHING EQUIPMENT Use dust-masks or over-pressure breathing gear if dust formation is likely.GLOVES Essential.

Wear chemical resistant gloves (nitrile rubber, NBR).EYE PROTECTION Protective glasses/goggles are essential.OTHER Wear protective clothing and work in a well-ventilated area.

Remove any contaminated clothing and wash hands thoroughly after working with the product.

Page 92: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

90

Material Safety Data Sheets (MSDS)

DISPOSAL MEASURES

Contact a specialist waste disposal organization where disposal of this product is required. Dissolve, or mix, the product with a combustible solvent and burn in a chemical incinerator fitted with post-combustion and scrubber systems. Observe all applicable local, national and international environmental regulations.

FIRST AID MEASURES

FOLLOWING INHALATION In the event of inhalation, seek clean air. Where breathing has stopped, apply artificial respiration. Where breathing is difficult, give oxygen.

FOLLOWING CONTACT WITH SKIN In the event of skin contact, wash the skin immediately using soap and plenty of water.

FOLLOWING CONTACT WITH THE EYE

In the event of contact with the eyes, rinse out immediately with plenty of water for at least 15 minutes.

FOLLOWING INGESTION (IF SWALLOWED)

If swallowed, rinse out the mouth with water if the patient is still conscious. Call medical assistance.

FIRE FIGHTING - EXTINCTION METHODS

SUTABLE EXTINGUISHERS Water spray. Carbon dioxide, dry chemical powder or suitable foam.SPECIAL RISKS Specific Danger(s): Gives off toxic fumes if burnt.SPECIAL PROTECTIVE EQUIPMENT NEEDED FOR FIRE FIGHTING

Wear individual breathing apparatus and protective clothing to prevent any contact with the eyes or skin.

The information contained in this document is reviewed at regular intervals and checked for special cases.

Page 93: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

91

Material Safety Data Sheets (MSDS)

13.7 Sodium molybdate dihydrate

Name:SODIUM MOLYBDATE DIHYDRATE Formula: Na2MoO4.2H2OCAS N° 10102-40-6

PHYSICAL PROPERTIES

MELTING POINT 687°CSOLUBILITY IN WATER 840 g/LFIRE AND EXPLOSION RISK No information availableHEALTH RISKS Not hazardous according to Directive 67/548/EEC

SECURITY MEASURES

EYES Rinse immediately with plenty of water, keeping the eyelids open as much as possible

LUNGS In case of inhalation, remove from exposure and move to fresh air immediately.

INGESTION Drink plenty of water. If feeling unwell seek medical aid

CHEMICAL DANGER AND STABILITY

STABILITY StableSPECIFIC RISKS No information availableDISPOSAL Observe all federal, state, and local environmental regulations

for disposal of chemical products

PERSONAL PROTECTION

RESPIRITORY Wear appropriate breathing apparatus due to the formation of dust

GLOVES Not necessaryEYES Wear appropriate safety goggles

MEASURES TO TAKE IN CASE OF ACCIDENTAL SPILLAGE

PERSONAL Avoid inhalation of any dustENVIRONMENT Do not dispose of in any standard waste disposal systems

such as sinks or sewersCLEANING Vacuum or sweep up material and place into a suitable

disposal container. Avoid generating dusty conditions.The information listed above must be reviewed periodically and should not be considered a substitute for the MSDS for this product.

Page 94: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

92

Material Safety Data Sheets (MSDS)

13.8 Reagent R1A

PRODUCT AND COMPANY IDENTIFICATION

SUBSTANCE Product code :R1AName of product: 9210 R1A Reagent

COMPOSITION / INFORMATION ABOUT THE COMPONENTS

CHEMICAL PROPERTIES PreparationHAZARDOUS COMPONENTSCAS NUMBER SOLUTION QUANTITY CLASSIFICATION213-653-77697-37-2

WaterNitric acid

90%10% Yes

IDENTIFICATION OF HAZARDS

Causes severe irritation and burns. Harmful if swallowed. Avoid breathing vapor or dust. Use with adequate ventilation. Avoid contact with eyes, skin, and clothes. Wash thoroughly after handling. Keep container closed.

FIRST AID

CHRONIC EXPOSURE Long-term exposure to concentrated vapors may cause erosion of teeth and lung damage. Long-term exposures seldom occur due to the corrosive properties of the acid.

AGGRAVATION OF PRE-EXISTING CONDITIONS Persons with pre-existing skin disorders, eye disease, or cardiopulmonary diseases may be more susceptible to the effects of this substance.

FIRE FIGHTING MEASURES

FIRE Not combustible, but substance is a strong oxidizer and its heat of reaction with reducing agents or combustibles may cause ignition. Can react with metals to release flammable hydrogen gas.

EXPLOSION Reacts explosively with combustible organic or readily oxidizable materials such as: alcohols, turpentine, charcoal, organic refuse, metal powder, hydrogen sulfide, etc. Reacts with most metals to release hydrogen gas which can form explosive mixtures with air.

FIRE EXTINGUISHING MEDIA Water spray may be used to keep fire exposed containers cool. Do not get water inside container.

SPECIAL INFORMATION Increases the flammability of combustible, organic and readily oxidizable materials. In the event of a fire, wear full protective clothing and NIOSH-approved self-contained breathing apparatus with full facepiece operated in the pressure demand or other positive pressure mode.

MEASURES IN CASE OF SPILLAGE

Ventilate area of leak or spill. Wear appropriate personal protective equipment.Keep unnecessary and unprotected personnel from entering. Contain and recover liquid when possible. Neutralize with alkaline material (soda ash, lime), then absorb with an inert material (e.g. vermiculite, dry sand, earth), and place in a chemical waste container. Do not use combustible materials, such as sawdust.

HANDLING AND STORAGE

Store in a cool, dry, ventilated storage area with acid resistant floors and good drainage. Protect from physical damage. Keep out of direct sunlight and away from heat, water, and incompatible materials. Do not wash out container and use it for other purposes. When diluting, the acid should always be added slowly to water and in small amounts. Never use hot water and never add water to the acid. Water added to acid can cause uncontrolled boiling and splashing. Containers of this material may be hazardous when empty since they retain product residues (vapors, liquid); observe all warnings and precautions listed for the product.

Page 95: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

93

Material Safety Data Sheets (MSDS)

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION

RESPIRATORY PROTECTION A system of local and/or general exhaust is recommended to keep employee exposures below the Airborne Exposure Limits. Local exhaust ventilation is generally preferred because it can control the emissions of the contaminant at its source, preventing dispersion of it into the general work area.

SKIN PROTECTION Wear impervious protective clothing, including boots, gloves, lab coat, apron or coveralls, as appropriate, to prevent skin contact.

EYE PROTECTION Safety glasses recommended.

PHYSICAL AND CHEMICAL PROPERTIES

APPEARANCE Liquid.COLOR ColorlessODOR Suffocating, acrid. PH 1.0

STABILITY AND REACTIVITY

CONDITIONS TO AVOID Stable under ordinary conditions of use and storage. Containers may burst when heated.

TOXICOLOGICAL INFORMATION

NITRIC ACIDInhalation rat LC50: 244 ppm (NO2)/30M; Investigated as a mutagen, reproductive effector. Oral (human) LDLo: 430 mg/kg.

ECOLOGICAL INFORMATION

ENVIRONMENTAL TOXICITY No information found

DISPOSAL CONSIDERATIONS

PRODUCT

There are as yet no uniform regulations concerning the disposal of chemicals and their residues within the European Union.The chemicals and their residues must be considered as special waste; their disposal is regulated by each EEC member state. We advise you to contact the competent authorities (administration or specialist waste disposal companies) that will inform you of the disposal measures to take.

PACKAGING

Disposal in accordance with prevailing legislation. Contaminated containers must be processed in the same manner as the chemical itself. Unless specifically regulated, non-contaminated containers may be disposed of as household waste, or be recycled.

TRANSPORT INFORMATION

Not subject to transport provisions.

REGULATORY INFORMATION

Labelling to EEC directives

OTHER INFORMATION

Updated version : January 2009 Issue A.

Page 96: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

94

Material Safety Data Sheets (MSDS)

13.9 Reagent R1M

PRODUCT AND COMPANY IDENTIFICATION

SUBSTANCE Product code :R1MName of product: 9210 R1M Reagent

COMPOSITION / INFORMATION ABOUT THE COMPONENTS

CHEMICAL PROPERTIES PreparationHAZARDOUS COMPONENTSCAS NUMBER SOLUTION QUANTITY CLASSIFICATION213-653-77758-02-3

WaterSodium Molybdate Dihydrate

94%<5% Yes

IDENTIFICATION OF HAZARDS

INHALATION Vapors and mists cause irritation to the respiratory tract. INGESTION May cause gastrointestinal irritation with nausea, vomiting

and diarrhea.SKIN CONTACT Causes irritation and burns to the skin. EYE CONTACT Vapors cause irritation. Splashes cause severe pain, eye

damage, and permanent blindness.CHRONIC EXPOSURE Repeated exposure may cause damage to the tissues of the

mucous membranes, upper respiratory tract, eyes and skin. AGGRAVATION OF PRE-EXISTING CONDITIONS Persons with pre-existing eye disorders or impaired

respiratory function may be more susceptible to the effects of this material.

FIRST AID

EYE CONTACT Irrigate thoroughly with water. If discomfort persists obtain medical attention.

INHALATION Remove from exposure.SKIN CONTACT Wash off thoroughly with soap and water.INGESTION Wash out mouth thoroughly with water. In severe cases obtain

medical attention.

FIRE FIGHTING MEASURES

SPECIAL RISKS Not readily combustible.

MEASURES IN CASE OF SPILLAGE

Wear appropriate protective clothing.If local regulations permit, mop up with plenty of water and run to waste, diluting greatly with running water. Otherwise transferto container and arrange removal by disposal company. Wash site of spillage thoroughly with detergent and water.For large spillages liquids should be contained with sand or earth and both liquids and solids transferred to salvage containers.Any residues should be treated as for small spillages.

HANDLING AND STORAGE

HANDLING Wash thoroughly after handling. Avoid contact with eyes, skin, and clothing.. Avoid ingestion and inhalation. Wash clothing before reuse.

STORAGE Store in a tightly closed container.

Page 97: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

95

Material Safety Data Sheets (MSDS)

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION

RESPIRATORY PROTECTION If the exposure limit is exceeded and engineering controls are not feasible, a full facepiece respirator may be worn.

HAND PROTECTION Wear latex gloves.EYE PROTECTION Safety glasses recommended.HYGIENE MEASURES Wash hands after working with substance.

PHYSICAL AND CHEMICAL PROPERTIES

APPEARANCE Liquid.COLOR ColorlessODORPH ~7

STABILITY AND REACTIVITY

CONDITIONS TO AVOID Stable under ordinary conditions of use and storage

TOXICOLOGICAL INFORMATION

After eye contact May irritate.After skin contact May irritate.Hazardous properties cannot be excluded but unlikely when handled appropriately. The product should be handled with the care usual when dealing with chemicals.

ECOLOGICAL INFORMATION

ENVIRONMENTAL TOXICITY No information found

OTHER ECOLOGY-RELATED INFORMATION In the event of appropriate handling and use, there is no risk of ecological problems.

DISPOSAL CONSIDERATIONS

PRODUCT

There are as yet no uniform regulations concerning the disposal of chemicals and their residues within the European Union.The chemicals and their residues must be considered as special waste; their disposal is regulated by each EEC member state. We advise you to contact the competent authorities (administration or specialist waste disposal companies) that will inform you of the disposal measures to take.

PACKAGING

Disposal in accordance with prevailing legislation. Contaminated containers must be processed in the same manner as the chemical itself. Unless specifically regulated, non-contaminated containers may be disposed of as household waste, or be recycled.

TRANSPORT INFORMATION

Not subject to transport provisions.

REGULATORY INFORMATION

Labelling to EEC directives

OTHER INFORMATION

Updated version : January 2009 Issue A.

Page 98: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

96

Material Safety Data Sheets (MSDS)

13.10Reagent R2

PRODUCT AND COMPANY IDENTIFICATION

SUBSTANCE Product code: R2Name of product: 9210 R2 Reagent

SUPPLIER Manufacturer Name: HACH SASAddress: 29 rue de Bures, 14670 TroarmPhone : (33) 02.31.15.37.10Fax N° : (33) 02.31.15.37.30Emergency : (33) 02.31.15.37.35 (Unit of production): (33) 01.45.42.59.59 (N° ORFILA)

COMPOSITION / INFORMATION ABOUT THE COMPONENTS

CHEMICAL PROPERTIES PreparationHAZARDOUS COMPONENTSCAS NUMBER SOLUTION QUANTITY CLASSIFICATION213-653-7 6153-56-6

Water Oxalic acid

95%5% Xn ; R21/22

IDENTIFICATION OF HAZARDS

Injurious on contact with skin or by ingestion

FIRST AID

SKIN CONTACT Rinse affected area with plenty of water. Remove contaminated clothing.

EYE CONTACT Immediately flush eyes with plenty of water with eyelids openIF SWALLOWED IN LARGE QUANTITIES Give plenty of water to drink. In case of vomiting, take medical

advice.

FIRE FIGHTING MEASURES

EXTINGUISHING MEANS Adapt the extinguishing agent to the products stored in the immediate vicinity

SPECIFIC HAZARDS Non combustible. Risk of forming hydrogen if in contact with light metals

MEASURES IN CASE OF SPILLAGE

PRECAUTIONS FOR PEOPLE Avoid contact with skin and eyesENVIRONMENT PROTECTION Do not pour in sewers or rivers CLEANING Collect in specific containers. Clean soiled materials

HANDLING AND STORAGE

STORAGE RECOMMENDATIONS Store in hermetically sealed containers

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION

RESPIRATORY PROTECTION Required if steam / aerosolHAND PROTECTION Wear nitrile glovesEYE PROTECTION Safety glasses are requiredHYGIENE MEASURES Wash hands after contact

PHYSICAL AND CHEMICAL PROPERTIES

FORM LiquidCOLOR ColorlessODOR OdorlesspH < 1

Page 99: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

97

Material Safety Data Sheets (MSDS)

STABILITY AND REACTIVITY

CONDITIONS TO BE AVOIDED Strong heatingSUBSTANCES TO BE AVOIDED BasesDANGEROUS BY-PRODUCTS No available information

TOXICOLOGICAL INFORMATION

ACUTE TOXICITY DL50 (oral, rat): 7500 mg/KgIN THE EVENT OF SKIN CONTACT Slight irritationIN THE EVENT OF EYE CONTACT Slight irritationIF SWALLOWED IN LARGE QUANTITIES Injurious

ECOLOGICAL INFORMATION

ECOTOXIC Biologic: Fish CL50: 325 mg/l (anhydrous liquid)OTHER ECOLOGY-RELATED INFORMATION Damage:

DCO: 0.18 g/g (compared to the pure substance)DThO: 0.18 g/g (compared to the pure substance)DBO: 0.16 g/g (compared to the pure substance)Do not pour in sewer or in the rivers.

DISPOSAL CONSIDERATIONS

PRODUCT There are as yet no uniform regulations concerning the disposal of chemicals and their residues within the European Union.The chemicals and their residues must be considered as special waste; their disposal is regulated by each EEC member state.We advise you to contact the competent authorities (administration or specialist waste disposal companies) that will inform you of the disposal measures to take

PACKAGING Disposal in accordance with prevailing legislation. Contaminated containers must be processed in the same manner as the chemical itself. Unless specifically regulated, non-contaminated containers may be disposed of as household waste, or be recycled

TRANSPORT INFORMATION

Not subject to transport provisions

REGULAR INFORMATION

LABELLING TO EEC DIRECTIVESHAZARDOUS PRODUCT IDENTIFICATION

Xn Injurious" R " Phrases R21/22 Injurious in contact with skin

or if ingested" S " Phrases S36/37 Wear protective clothes and

appropriate gloves

OTHER INFORMATION

Updated version: November 2005 Issue A

Page 100: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

98

Material Safety Data Sheets (MSDS)

13.11Reagent R3

PRODUCT AND COMPANY IDENTIFICATION

SUBSTANCE Product code: R3Name of product: 9210 R3 Reagent

SUPPLIER Manufacturer Name: HACH SASAddress: 29 rue de Bures, 14670 Troarm

Phone : (33) 02.31.15.37.10Fax N° : (33) 02.31.15.37.30Emergency : (33) 02.31.15.37.35 (Unit of production): (33) 01.45.42.59.59 (N° ORFILA)

COMPOSITION / INFORMATION ABOUT THE COMPONENTS

CHEMICAL PROPERTIES PreparationHAZARDOUS COMPONENTSCAS NUMBER SOLUTION QUANTITY CLASIFICATION213-653-7 7783-85-97664-93-9

Water Ammonium iron III sulfatesSulfuric acid

96%2%2% C; R35; S26 -30 - 45

IDENTIFICATION OF HAZARDS

Product not considered as dangerous according to the 67/548/CEE directive

FIRST AID

SKIN CONTACT Rinse affected area with plenty of water. Remove contaminated clothing. Swab with polyethylene glycol 400

EYE CONTACT Immediately flush eyes with plenty of water with eyelids open for 10 minutes. Get medical aid if necessary

IF SWALLOWED IN LARGE QUANTITIES Give plenty of water to drink. In case of indisposition, take medical advice. Do not try to neutralize

FIRE FIGHTING MEASURES

EXTINGUISHING MEANS Adapt the extinguishing agent to the products stored in the immediate vicinity

SPECIFIC HAZARDS Non combustible. Risk of forming hydrogen if in contact with light metals

MEASURES IN CASE OF SPILLAGE

PRECAUTIONS FOR PEOPLE Avoid contact with skin and eyesENVIRONMENT PROTECTION Do not pour in sewers or rivers CLEANING Collect in specific containers. Clean soiled materials

HANDLING AND STORAGE

PRECAUTIONS TO TAKE Avoid contact with metalsUSEAGE ADVICE Respect conditions of useSTORAGE RECOMMENDATIONS Store in hermetically sealed containers

EXPOSURE CONTROLS / INDIVIDUAL PROTECTION

RESPIRATORY PROTECTION Required if steam / aerosolHAND PROTECTION Wear nitrile glovesEYE PROTECTION Safety glasses are requiredOTHER PROTECTION Acid-proof clothingHYGIENE MEASURES Wash hands after contact

Page 101: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

99

Material Safety Data Sheets (MSDS)

PHYSICAL AND CHEMICAL PROPERTIES

FORM LiquidCOLOR ColorlessODOR OdorlesspH < 1

STABILITY AND REACTIVITY

CONDITIONS TO BE AVOIDED Strong heatingSUBSTANCES TO BE AVOIDED Metals, base, acidDANGEROUS BY-PRODUCTS In case of fire, see FIRE FIGHTING MEASURES above

TOXICOLOGICAL INFORMATION

ACUTE TOXICITY DL50 (oral, rat): 7500 mg/KgIN THE EVENT OF SKIN CONTACT Causes burnsIN THE EVENT OF EYE CONTACT Causes burnsIF SWALLOWED IN LARGE QUANTITIES Strong pains with nausea, vomiting and diarrhea

ECOLOGICAL INFORMATION

ECOTOXIC No specific indicationsOTHER ECOLOGY-RELATED INFORMATION In the event of proper handling and use, there is no risk of

ecological problems.Injurious on aquatic life, by pH modification

DISPOSAL CONSIDERATIONS

PRODUCT There are as yet no uniform regulations concerning the disposal of chemicals and their residues within the European Union.The chemicals and their residues must be considered as special waste; their disposal is regulated by each EEC member state.We advise you to contact the competent authorities (administration or specialist waste disposal companies) that will inform you of the disposal measures to take

PACKAGING Disposal in accordance with prevailing legislation. Contaminated containers must be processed in the same manner as the chemical itself. Unless specifically regulated, non-contaminated containers may be disposed of as household waste, or be recycled

TRANSPORT INFORMATION

Not subject to transport provisions

REGULATORY INFORMATION

LABELLING TO EEC DIRECTIVES N° CE: 23-211-8HAZARDOUS PRODUCT IDENTIFICATION C Corrosive" R " Phrases R35 Causes burns" S " Phrases S26 - S30 - S36/37/39 - S45 If in contact with eyes, flush

eyes immediately with plenty of water.Wear protective clothes, gloves, goggles and a face protection device. In case of accident or feeling faint, contact a doctor

OTHER INFORMATION

Updated version: November 2005 Issue A

Page 102: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs

100

Material Safety Data Sheets (MSDS)

13.12Lithium battery safety precautionsThe lithium battery on the microprocessor PCB does not constitute a serious hazard providing the following instructions are observed:

NEVER:

• Attempt to recharge the battery.

• Crush, puncture, open or dismantle the battery.

• Store or operate equipment above the temperature specified.

• Short-circuit the battery.

• Dispose of the battery in a fire.

• Try to repair faults on the control PCB

In the event of a PCB fault or battery leakage:

ALWAYS:

• Wear suitable protective clothing.

• Remove the equipment to a well ventilated area.

• Contact your local agent.

• Refer to Health and Safety Notes on Lithium batteries.

Page 103: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs
Page 104: POLYMETRON Model 9210 Silica Analyzerœ¨线硅表... · POLYMETRON Model 9210 Silica Analyzer USER MANUAL May 2009, Revision K. 1 Table of Contents ... 3.3 Analyzer outputs