ABL800 FLEX Introduction. RTC, December 2004. Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, www.radiometer.com. Agenda. Front view Rear view Modular design Parameters Analyzer configurations Specifications Sampler types AutoCheck Screen design. - PowerPoint PPT Presentation

Citation preview

Page 1: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

1Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, www.radiometer.com

RTC, December 2004

ABL800 FLEXIntroduction

Page 2: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

2

Agenda

• Front view• Rear view• Modular design• Parameters• Analyzer

configurations• Specifications• Sampler types• AutoCheck• Screen design

Page 3: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

3

Front view

1

2

4

3

7

6

5

1. Aspiration inlet2. Barcode reader 3. Color touch screen4. Printer5. Solutions (cleaning

calibration, rinse and waste containers)

6. AutoCheck7. CD-RW drive8. Window to electrode

modules

5

8

Page 4: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

4

Rear view

1. Screen, rear2. Gas containers 3. Communication ports 4. Power switch5. Fan

1

3

2 2

54

Page 5: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

5

Communication ports

• The two USB communication ports provide connection to external devices, e.g. network, keyboard, USB storage device

• The two other communication ports provide connection to mouse and keyboard

• Furthermore there is a printer port, a network connector, a serial port and a port for an external CRT VG monitor

2 31

1 USB2 Mouse/keyboard3 Network connections

Page 6: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

6

Modular design

• Allows flexibility in parameter availability

1 Inlet2 pH/BG3 Met/Lyt4 Oximetry

2 31 4

Page 7: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

7

Parameters

• There are 10 electrode slots in the two electrode modules• One electrode slot for the reference electrode• Nine other electrodes available for the following

parameters:

pHpH, , ppCOCO22, , ppOO22, , ccKK++, , ccNaNa++, , ccCaCa2+2+, , ccClCl–– , , ccGlucoseGlucose, , ccLactateLactate

Page 8: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

8

Oximetry parameters

ctHb = concentration of total hemoglobin

sO2 = oxygen saturation %

FO2Hb = fraction of oxyhemoglobin

FCOHb = fraction of carboxyhemoglobin

FHHb = fraction of deoxyhemoglobin

FMetHb = fraction of methemoglobin

FHbF = fraction of fetal hemoglobin

ctBil = concentration of bilirubin

Page 9: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

9

Analyzer configuration

BG OXI ctHb/sO2 ctBIL/FHbF LYT MET

835 830 825 820 815 810 805 800

Models ABL835, ABL830 and ABL800 of the ABL800 FLEX analyzers are not available at the time of the release

Page 10: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

10

Specifications – analysis details

• Sample volumes Macromodes 195 L (all parameters) Micromodes 35-95 L (all parameters)

• Measuring time Macromodes 80 sec., cycle time 150 sec., approx. 24 samples/hr Micromodes 80-135 sec., cycle time 145-200 sec., approx. 18 samples/hr

• Derived parameters 47 different derived parameters to choose from. See the types and definition in

teh ABL800 FLEX Specification sheet

Page 11: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

11

Specifications - communications

• Computer specification Intel Celeron Processor Windows XP embedded

• True multi-tasking system Intel Celeron processor for DMS software Dedicated 80386 CPU for wet section

• Communication LIS/HIS communication based on ASTM and HL7 Network communication – Ethernet (TCP/IP protocol as communication protocol) True two-way communication

Page 12: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

12

Customizable

• Very comprehensive setup programs More functions than ever can be customized Examples: operators and passwords, corrective actions, reference and critical

ranges etc.

• Automatic restart after replacements• On-line help systems

Video tutorials Help troubleshooting

• Connection to RADIANCE Especially analyzer remote control module

Page 13: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

13

Sampler types

Syringe QC with adapterCapillary tube

Expired air Test tube

Page 14: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

14

AutoCheck

• Automated QC measurements • Eliminates user-to-user

variance - results only reflect analyzer performance

• Time-saving• Holds 20 ampoules

Page 15: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

15

Color touch screen

• Operation and management of the analyzer are performed via the color touch screen

• The screen is divided into three sections

1. Top section/status bar varies slightly according to the status of the analyzer and setup

2. Center section contains the main information and interaction area

3. Bottom section contains navigation buttons and varies slightly according to the screen in question

1

3

2

1 Top section2 Center section3 Bottom section

Page 16: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

16

Top section

• The status bar describes the current task

E.g. calibration, rinse, aspiration The time bar is present when the

analyzer is performing a task

• The parameter bar lists all the measured parameters available and activated

• The traffic light gives an overview of various analyzer activities and the traffic light color shows the overall status of the analyzer – see the “02 What is what” presentation for a further description

1

3

2

1 Status bar2 Parameter bar3 Traffic light

Page 17: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

17

Center section

• The center section is the main information and interaction area

• Each screen is identified by an icon and a name that matches the button that gives access to it

• The screen may contain text, check buttons, scroll bars, a screen keypad or screen keyboard

Page 18: Radiometer Medical ApS, Åkandevej 21, DK-2700 Brønshøj, Tel: +45 38 27 38 27, radiometer

18

Bottom section

• The bottom section changes with the different options, depending on the screen in question

1. In the Ready screen Access various information, e.g. the

current time, next QC, if connected to RADIANCE

The buttons can be selected to give access to e.g. Logon/Log Off, the Menu

2. General navigation buttons “Close” – close actual window “Back” – go one step backwards

3. Other navigation buttons Results, ID, messages, print, etc.

Quality Control Result screen:

1 Main menu2 General navigation buttons3 Other navigation buttons