The Optimum Operating Conditions for Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Ferrocene Pao-Tsai Kuo ( 郭郭郭 ), Chung-Min Lien ( 郭郭郭 ), Hau Lin ( 郭郭 ) Department of Chemical and Materials Engineering, Southern Taiwan University ABSTRACT A study was conducted to use the ferrocene to modify the carbon paste electrode. Because the ferrocene(Fe(C 5 H 5 ) 2 ) possesses the excellent catalytic ch aracteristic, it can be used with the graphite carbon powders which possesses the excellent conductivity to make the carbon paste electrode and to elev ate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution(PBS) and then the concentration of the hydrogen peroxide can be obtained and consequently, the concentration of the glucose can be determined. At optimum operating c onditions - 0.2V operating potential, 500 rpm stirring rate and in 0.05M PBS buffer solution ( pH = 7.4 ) , when the carbon paste electrode was modifi ed with the ferrocene [ferrocene : graphite carbon powders = 3 : 7 ( weight ratio )] , the detection limit was 0.02 mM H 2 O 2 , the linear range was 0.02 1.2 mM H 2 O 2 , R 2 = 0.9998 and the sensitivity was 161.43µA/cm 2 mM H 2 O 2 . INTRODUCTION Because hydrogen peroxide is widely used in the food industry for the purpose of preservation nowadays, a rapid and convenient sensor for detecting hydrogen peroxide is an important research subject. In recen t years, diabetes has become one of the top ten causes of death for th e people in Taiwan. Therefore developing a rapid and convenient glucos e biosensor also has become an important research subject. The glucose and oxygen can be catalyzed by the glucose oxidase and the glucose is o xidized to gluconic acid and the oxygen is reduced to hydrogen peroxide. The electrode releases the electrons and converts the mediator to the reductive state. Then the mediator at the reductive state releases the electrons to reduce the hydrogen peroxide to water and consequently the mediator at the reductive state is converted to the oxidizing state. Th e responding current for detecting the hydrogen peroxide is used to mea sure the amount of hydrogen peroxide and consequently determine the con centration of the glucose. Because the ferrocene(Fe(C 5 H 5 ) 2 ) possesses the excellent catalytic characteristic, it can be used with the graphite ca rbon powders which possesses the excellent conductivity to make the car bon paste electrode and to elevate the responding current of hydrogen p eroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution(PBS) and then the concentration of the hydrog en peroxide can be obtained and consequently, the concentration of the glucose can be determined. EXPERIMENTAL 1. Equipment Electrochemical Analyzer (CHI 401A, CH Instruments, Inc) was used to measure the activity of electrode by Cyclic Voltammetry ( CV ) and Time Base ( TB ) mode ; Electric Stirrer(Fargo) ; pH meter (Metrohm 731); Constant Temperature Thermal Bath (Wisdom BC-2DT 10L); Oven (DEN G YNG) ; Carbon Paste Electrode was used as the working electrodes, Co iled Platinum Wire was used as the counter electrode and Ag / AgCl was used as the reference electrode. 2. Chemicals and Reagents Ferrocene ; Hydrochloric Acid (HCl); Sodium Hydroxide (NaOH) ; H ydrogen Peroxide (H 2 O 2 ); Graphite Carbon Powder( C ); Carbon Paste; Cyc lohexanone(C 6 H 10 O); Potassium Dihydrogenphosphate (KH 2 PO 4 ); Potassium Ch loride (KCl). 3. Preparation of the Carbon Paste Electrode : Take one section of 7 cm electric wire with 0.05 cm inside diamete r. After depriving the coating 0.5 cm length from both ends, the nake- ended wire was washed, dried and ready for use. Then the ferrocene po wders, graphite carbon powders and carbon paste were mixed with the ap propriate ratio (ferrocene : graphite carbon powders : carbon paste = 0.3 : 0.7 : 1). After the mixing was complete, the mixture was evenly coated on the nake-ended electric wire and dried in the oven and then we obtained the carbon paste electrode. 7 cm 0.5 cm 0.05 cm Ferroc ene Graphite carbon powders Mixing with equal amount of carbon paste DEPARTMENT OF CHEMICAL AND MATERIALS ENGINEERING, SOUTHERN TAIWAN UNIVERSITY RESULTS CONCLUSIONS The results showed that the responding current for the carbon past e electrode modified with ferrocene was elevated significantly. The T B (Time Base ) graphs for different operating potentials, stirring rate s, and pH values were plotted to determine the optimum operating cond itions. The results showed that at the optimum operating conditions 0.2 V operating potential, 500 rpm stirring rate and in 0.05M PBS buffe r solution (pH=7.4), the detection limit was 0.02 mM H 2 O 2 , the linear ra nge was 0.02~1.2 mM H 2 O 2 , R 2 =0.9998 and the sensitivity was 161.43μA/cm 2 mM H 2 O 2 . This research can be further applied to the glucose biosensor i n the future. 1. L. Charpentier and N. El Murr, “Amperometric Determination of Chole sterol in Serum with Use of a Renewable Surface Peroxidase Electrode,” Analytica Chimica Acta, 318, 89 (1995). 2. R. Nagata, S. A. Clark, K. Yokoyama , E. Tamiya and I. Karube, “Amp erometric Glucose Biosensor Manufactured by a Printing Technique” Analytica Chimica Acta, 304, 157 (1995). 3. H. Liu, H. Li, T. Ying, K. Sun, Y. Qin, and D. Qi, “Amperometric Biosensor Snesitive to Glucose and Lactose Based on Co-Immobilization of Ferrocene, Glucose Oxidase, β-Galactosidase and Mutarotase in β-Cyc lodextrin Polymer,” , Analytica Chimica Acta, 358, 137 ( 1998). 4. Y.-M. Uang and T.-C. Chou , “Fabrication of Glucose Oxidase/Polypyr role Biosensor by Galvanostatic Method in Various pH Aqueous Solutions, Biosensors and Bioelectronics, 19, 141 (2003). REFERENCES (B ) (A) Fig 1. CV graphs for (A) carbon paste el ectrode modified with ferrocene ( the ra nge of scanning potential: -0.8 +0.8 V) (B) unmodified carbon paste electrode ( the range of scanning potential: -0.8 +0.8 V) Fig. 2 The TB graphs of carbon paste electrode for detection of H 2 O 2 at diff erent operating potentials (ferrocene : graphite carbon powders = 3 : 7); th e operating potentials are [ (A) 0V (B) –0.1V (C) –0.15V (D) –0.2V (E) –0. 25 V ] Fig. 3 The TB graphs of carbon paste electrode for detection of H 2 O 2 at diff erent stirring rates (ferrocene : gra phite carbon powders = 3 : 7); the st irring rates are [ (A) 100 rpm (B) 200 rpm (C) 300 rpm (D) 400 rpm (E) 500 rp m ] Fig. 4 The TB graphs of carbon paste electrode for detection of H 2 O 2 at diff erent pH values of PBS buffer soluti on (ferrocene : graphite carbon powder s = 3 : 7); the pH values are [ (A) pH = 4.0 (B) pH = 5.0 (C) pH = 6.0 (D) pH = 7.4 (E) pH = 8.0 ] Fig. 5 The TB graphs of carbon paste elect rode for determining the detection limit of H 2 O 2 (ferrocene: graphite carbon powders = 3 : 7); At 30 ℃; the operating potential = –0. 2 V; in 0.1 M KCl of 5 mL 0.05 M PBS buffer solution ( pH= 7.4 ) Fig. 6 The TB graphs of carbon paste el ectrode for determining the linear range of H 2 O 2 (ferrocene : graphite carbon pow ders = 3 : 7); At 30 ℃; the operating p otential = –0.2 V; in 0.1 M KCl of 5 mL 0.05 M PBS buffer solution ( pH= 7.4 ); stirring rate =500 rpm; 10μL of 100mM H 2 O 2 is injected per 100 seconds

The Optimum Operating Conditions for Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Ferrocene Pao-Tsai Kuo ( 郭寶財 ), Chung-Min

Embed Size (px)

Citation preview

Page 1: The Optimum Operating Conditions for Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Ferrocene Pao-Tsai Kuo ( 郭寶財 ), Chung-Min

The Optimum Operating Conditions for Detection of Hydrogen Peroxide for the Carbon Paste Electrode Modified with Ferrocene

Pao-Tsai Kuo ( 郭寶財 ), Chung-Min Lien ( 連崇閔 ), Hau Lin ( 林浩 )

Department of Chemical and Materials Engineering, Southern Taiwan University

ABSTRACT

A study was conducted to use the ferrocene to modify the carbon paste electrode. Because the ferrocene(Fe(C5H5)2) possesses the excellent catalytic characteristic, it can be used with the graphite carbon powders which possesses the excellent conductivity to make the carbon paste electrode and to elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution(PBS) and then the concentration of the hydrogen peroxide can be obtained and consequently, the concentration of the glucose can be determined. At optimum operating conditions - 0.2V operating potential, 500 rpm stirring rate and in 0.05M PBS buffer solution ( pH = 7.4 ) , when the carbon paste electrode was modified with the ferrocene [ferrocene : graphite carbon powders = 3 : 7 ( weight ratio )] , the detection limit was 0.02 mM H2O2 , the linear range was 0.02 ~ 1.2 mM H2O2 , R2 = 0.9998 and the sensitivity was 161.43µA/cm2 . mM H2O2.

INTRODUCTION

Because hydrogen peroxide is widely used in the food industry for the purpose of preservation nowadays, a rapid and convenient sensor for detecting hydrogen peroxide is an important research subject. In recent years, diabetes has become one of the top ten causes of death for the people in Taiwan. Therefore developing a rapid and convenient glucose biosensor also has become an important research subject. The glucose and oxygen can be catalyzed by the glucose oxidase and the glucose is oxidized to gluconic acid and the oxygen is reduced to hydrogen peroxide. The electrode releases the electrons and converts the mediator to the reductive state. Then the mediator at the reductive state releases the electrons to reduce the hydrogen peroxide to water and consequently the mediator at the reductive state is converted to the oxidizing state. The responding current for detecting the hydrogen peroxide is used to measure the amount of hydrogen peroxide and consequently determine the concentration of the glucose. Because the ferrocene(Fe(C5H5)2) possesses the excellent catalytic characteristic, it can be used with the graphite carbon powders which possesses the excellent conductivity to make the carbon paste electrode and to elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution(PBS) and then the concentration of the hydrogen peroxide can be obtained and consequently, the concentration of the glucose can be determined.

EXPERIMENTAL

1. Equipment Electrochemical Analyzer (CHI 401A, CH Instruments, Inc) was used to measure the activity of electrode by Cyclic Voltammetry ( CV ) and Time Base ( TB ) mode ; Electric Stirrer(Fargo) ; pH meter (Metrohm 731); Constant Temperature Thermal Bath (Wisdom BC-2DT 10L); Oven (DENG YNG) ; Carbon Paste Electrode was used as the working electrodes, Coiled Platinum Wire was used as the counter electrode and Ag / AgCl was used as the reference electrode. 2. Chemicals and Reagents Ferrocene ; Hydrochloric Acid (HCl); Sodium Hydroxide (NaOH) ; Hydrogen Peroxide (H2O2); Graphite Carbon Powder( C ); Carbon Paste; Cyclohexanone(C6H10O); Potassium Dihydrogenphosphate (KH2PO4); Potassium Chloride (KCl).

3. Preparation of the Carbon Paste Electrode : Take one section of 7 cm electric wire with 0.05 cm inside diameter. After depriving the coating 0.5 cm length from both ends, the nake-ended wire was washed, dried and ready for use. Then the ferrocene powders, graphite carbon powders and carbon paste were mixed with the appropriate ratio (ferrocene : graphite carbon powders : carbon paste = 0.3 : 0.7 : 1). After the mixing was complete, the mixture was evenly coated on the nake-ended electric wire and dried in the oven and then we obtained the carbon paste electrode.

7 cm

0.5 cm

0.05 cm

Ferrocene

Graphite carbon powders

Mixing with equal amount of carbon paste

DEPARTMENT OF CHEMICAL AND MATERIALS ENGINEERING, SOUTHERN TAIWAN UNIVERSITY

RESULTS

CONCLUSIONS

The results showed that the responding current for the carbon paste electrode modified with ferrocene was elevated significantly. The TB (Time Base ) graphs for different operating potentials, stirring rates, and pH values were plotted to determine the optimum operating conditions. The results showed that at the optimum operating conditions –0.2 V operating potential, 500 rpm stirring rate and in 0.05M PBS buffer solution (pH=7.4), the detection limit was 0.02 mM H2O2, the linear range was 0.02~1.2 mM H2O2, R2=0.9998 and the sensitivity was 161.43μA/cm2 ּmM H2O2. This research can be further applied to the glucose biosensor in the future.

1. L. Charpentier and N. El Murr, “Amperometric Determination of Cholesterol in Serum with Use of a Renewable Surface Peroxidase Electrode,” Analytica Chimica Acta, 318, 89 (1995).2. R. Nagata, S. A. Clark, K. Yokoyama , E. Tamiya and I. Karube, “AmperometricGlucose Biosensor Manufactured by a Printing Technique” Analytica Chimica Acta, 304, 157 (1995).3. H. Liu, H. Li, T. Ying, K. Sun, Y. Qin, and D. Qi, “Amperometric Biosensor Snesitive to Glucose and Lactose Based on Co-Immobilization of Ferrocene, Glucose Oxidase, β-Galactosidase and Mutarotase in β-Cyclodextrin Polymer,” , Analytica Chimica Acta, 358, 137 ( 1998).4. Y.-M. Uang and T.-C. Chou , “Fabrication of Glucose Oxidase/Polypyrrole Biosensor by Galvanostatic Method in Various pH Aqueous Solutions,” Biosensors and Bioelectronics, 19, 141 (2003).

REFERENCES

(B)

(A)

Fig 1. CV graphs for (A) carbon paste electrode modified with ferrocene ( the range of scanning potential: -0.8~ +0.8 V) (B) unmodified carbon paste electrode( the range of scanning potential: -0.8 ~ +0.8 V)

Fig. 2 The TB graphs of carbon paste electrode for detection of H2O2 at different operating potentials (ferrocene : graphite carbon powders = 3 : 7); the operating potentials are [ (A) 0V (B) –0.1V (C) –0.15V (D) –0.2V (E) –0.25 V ]

Fig. 3 The TB graphs of carbon paste electrode for detection of H2O2 at different stirring rates (ferrocene : graphite carbon powders = 3 : 7); the stirring rates are [ (A) 100 rpm (B) 200 rpm (C) 300 rpm (D) 400 rpm (E) 500 rpm ]

Fig. 4 The TB graphs of carbon paste electrode for detection of H2O2 at different pH values of PBS buffer solution (ferrocene : graphite carbon powders = 3 : 7); the pH values are [ (A) pH = 4.0 (B) pH = 5.0 (C) pH = 6.0 (D) pH = 7.4 (E) pH = 8.0 ]

Fig. 5 The TB graphs of carbon paste electrode for determining the detection limit of H2O2 (ferrocene: graphite carbon powders = 3 : 7); At 30 ; the operating potential = –0.2 V; in 0.℃1 M KCl of 5 mL 0.05 M PBS buffer solution ( pH= 7.4 )

Fig. 6 The TB graphs of carbon paste electrode for determining the linear range of H2O2 (ferrocene : graphite carbon powders = 3 : 7); At 30 ; the operating potentia℃l = –0.2 V; in 0.1 M KCl of 5 mL 0.05 M PBS buffer solution ( pH= 7.4 ); stirring rate =500 rpm; 10μL of 100mM H2O2 is injected per 100 seconds