4
EJX910 Multivariable Transmitter 13 EJX910 MULTIVARIABLE TRANSMITTER ITOU Akio *1 MIMURA Shin-ichi *1 KOYAMA Etsutarou *1 ODOHIRA Tetsu *1 NIKKUNI Masaaki *1 MIYAUCHI Tatsuhiko *2 We have developed the EJX910 multivariable transmitter, an all-in-one instrument that integrates the functions of a differential pressure transmitter, a pressure gauge, a thermometer and a flow computer, while featuring high performance and space-saving design. The transmitter employs a unique flow rate calculation method, achieving a mass flow calculation cycle of 100 milliseconds. By adopting a Reynolds number compensation algorithm, etc., all flow calculation parameters were optimized and a mass flow accuracy rate as high as 1% was achieved. Furthermore, EJX910 complies with a wide range of primary devices, including orifices, nozzles and venturi tubes, and various types of fluid, including general fluids, steam tables, and natural gas. Application information, such as the primary devices and fluid data required for mass flow calculation, is input using the EJXMVTool, a mass flow parameter configuration tool that runs on a PC and is downloaded to the transmitter by means of field communication. A field test performed on a British natural gas test line showed an excellent mass flow measurement accuracy of 1%. *1 Industrial Automation Products Business Division *2 Global Service Headquarters INTRODUCTION W hen evaluating mass flow rates using a primary device such as an orifice or nozzle in a differential pressure flowmeter in order to make fluid density compensations, the upstream pressure (static pressure) and fluid temperature are measured in addition to the output of a regular differential pressure transmitter. In the past in such cases, a differential pressure transmitter, a pressure transmitter, a temperature converter and a flow computer were all separately required. The DPharp EJX series of differential pressure transmitters developed by Yokogawa in 2004, can simultaneously measure both differential pressure and static pressure using an advanced form of silicon resonant sensors that comprise the multi-sensing function. Recently, we have developed the EJX910 multivariable transmitter as a new model that incorporates this series’ functions to full advantage. The EJX910 integrates the functions of a differential pressure gauge, a pressure gauge, a thermometer, and a flow computer into a single instrument, thereby achieving high space efficiency and multifunctionality. A Reynolds number compensation algorithm and other means have been adopted for the mass flow rate calculation of this transmitter to optimize all flow rate calculation parameters and achieve high-precision mass flow rate measurement. In this paper, we will focus on the functions related to mass flow rate calculation, one of the features a multivariable transmitter has to offer. Figure 1 shows an external view of the EJX910. Figure 1 External View of EJX910

Yokogawa Multivariable Transmitter EJX910

Embed Size (px)

Citation preview

Page 1: Yokogawa Multivariable Transmitter EJX910

EJX910 Multivariable Transmitter 13

EJX910 MULTIVARIABLETRANSMITTER

ITOU Akio *1 MIMURA Shin-ichi *1 KOYAMA Etsutarou *1 ODOHIRA Tetsu *1

NIKKUNI Masaaki *1 MIYAUCHI Tatsuhiko *2

We have developed the EJX910 multivariable transmitter, an all-in-oneinstrument that integrates the functions of a differential pressure transmitter, apressure gauge, a thermometer and a flow computer, while featuring highperformance and space-saving design. The transmitter employs a unique flow ratecalculation method, achieving a mass flow calculation cycle of 100 milliseconds. Byadopting a Reynolds number compensation algorithm, etc., all flow calculationparameters were optimized and a mass flow accuracy rate as high as 1% wasachieved. Furthermore, EJX910 complies with a wide range of primary devices,including orifices, nozzles and venturi tubes, and various types of fluid, includinggeneral fluids, steam tables, and natural gas. Application information, such as theprimary devices and fluid data required for mass flow calculation, is input using theEJXMVTool, a mass flow parameter configuration tool that runs on a PC and isdownloaded to the transmitter by means of field communication. A field testperformed on a British natural gas test line showed an excellent mass flowmeasurement accuracy of 1%.

*1 Industrial Automation Products Business Division*2 Global Service Headquarters

INTRODUCTION

When evaluating mass flow rates using a primary devicesuch as an orifice or nozzle in a differential pressure

flowmeter in order to make fluid density compensations, theupstream pressure (static pressure) and fluid temperature aremeasured in addition to the output of a regular differentialpressure transmitter. In the past in such cases, a differentialpressure transmitter, a pressure transmitter, a temperatureconverter and a flow computer were all separately required. TheDPharp EJX series of differential pressure transmitters developedby Yokogawa in 2004, can simultaneously measure bothdifferential pressure and static pressure using an advanced formof silicon resonant sensors that comprise the multi-sensingfunction. Recently, we have developed the EJX910 multivariabletransmitter as a new model that incorporates this series’ functionsto full advantage. The EJX910 integrates the functions of adifferential pressure gauge, a pressure gauge, a thermometer, and

a flow computer into a single instrument, thereby achieving highspace efficiency and multifunctionality.

A Reynolds number compensation algorithm and othermeans have been adopted for the mass flow rate calculation ofthis transmitter to optimize all flow rate calculation parametersand achieve high-precision mass flow rate measurement. In thispaper, we will focus on the functions related to mass flow ratecalculation, one of the features a multivariable transmitter has tooffer. Figure 1 shows an external view of the EJX910.

Figure 1 External View of EJX910

Page 2: Yokogawa Multivariable Transmitter EJX910

14 Yokogawa Technical Report English Edition, No. 42 (2006)

PRODUCT FEATURES

The EJX910 multivariable transmitter serves as a differentialpressure gauge, a pressure gauge, and a thermometer (with anexternal temperature sensor). In addition to thismultifunctionality, the fluid density compensation functionprovided by the transmitter itself and the PC-installedEJXMVTool’s mass flow parameter configuration tool enablehigh-speed, high-precision mass flow rate measurement. TheEJX910 supports a number of flow rate standards and a variety offluid types as target applications. Moreover, the EJX910 can beapplied to integrated flow rate measurement and variousdiagnoses that use many process variables (differential pressure,static pressure, temperature, etc.).

MASS FLOW RATE CALCULATION

Figure 2 shows an example of measuring mass flow rates fromorifices and temperature sensors installed in a process. The EJX910measures the difference between the upstream and downstreampressures of the orifice placed in the process, the upstream static

pressure and the fluid temperature. Then the transmitter calculatesthe mass flow rate from these measured values.

Figure 3 shows a configuration of a mass flow ratemeasurement system. Application information necessary for flowrate calculations (primary device and fluid information) is inputusing the EJXMVTool mass flow parameter configuration toolrunning on a PC. This information is then converted intoparameters that can be perceived by the transmitter anddownloaded to the transmitter by means of field communication.

Figure 4 shows a block diagram of a mass flow ratemeasurement system comprising the EJX910 and EJXMVTool.Differential pressure, static pressure and temperature measured bythe EJX910 can be directly output as process variables. The systemperforms fluid density compensation calculations according to thefollowing equation to determine the mass flow rate.

For this purpose the system employs a unique method ofcalculation that minimizes the transmitter’s calculation load, andachieves a flow rate calculation cycle of 100 milliseconds. In asimplified method of flow rate calculation, the parameters inEquation (1) are treated as fixed values, resulting in large massflow rate calculation errors as shown in Figure 5. The EJX910performs optimized calculations in real time using dynamicallychanging parameters, thereby realizing a high flow rate accuracyof 1%. More specifically, Reynolds number corrections are madeto the discharge coefficient (C) according to changing measuredvalues. The gas expansion correction factor (ε) is correctedagainst the effects of adiabatic expansion of gases. In addition,the fluid density (ρ1) is corrected for static pressure andtemperature variations.

Figure 2 Example of Mass Flow Rate Measurement Using Orifice

1

2

42

41ρεπ

βPdC

Qm ∆−

=

Mass flow rateDischarge coefficientBeta ratio (d/D)Bore of primary devicePipe inner diameterGas expansion factorDifferential pressureFluid density

::::::::

where

Dd

C

ε

β

Qm

˙˙˙˙˙˙ (1)

P∆1ρ

Figure 3 Mass Flow Rate Measurement System Configuration

EJX910 Multivariable Transmitter

Differential pressure, static pressure

Mass flow rate Temperature

Differential pressure, static pressure, temperature

Fluid

Temperature sensor

PC

EJXMVTool mass flow parameter configuration tool

Application information(primary device and fluid information)

Download

EJX910 multivariable transmitter

Temperature sensor

Mass flow rate

Figure 4 Mass Flow Rate Measurement Block Diagram

EJXMVTool EJX910

Physical property database(general fluids, steam tables, natural gases)

Primary device information

Flow rate calculation standards

Selection of fluid type and operating conditions

Flow rate calculation parameters

Sensor input

Out

put

Out

put

Opt

imiz

atio

n

Differential pressureStatic pressureTemperature

Flow rate calculation parameters

Flow rate calculation

Page 3: Yokogawa Multivariable Transmitter EJX910

EJX910 Multivariable Transmitter 15

SUPPORTED APPLICATIONS

The primary devices that the multivariable transmitter iscompatible with, i.e., orifices, nozzles and venturi tubes, complywith a number of flow rate standards. In addition, a fixed mode isavailable to set desired values to the discharge coefficient and thegas expansion correction factor. The transmitter can handle awide variety of fluid types as described below:(1) General fluids (gases and liquids)

The transmitter supports 12 types of general fluids andemploys the DIPPR® physical property database widely usedaround the world.

(2) Steam tablesThe transmitter supports IAPWS-IF97 Water and Steam(1997), the standard for steam tables used in each country(e.g., JSME and ASME).

(3) Natural gasesThe transmitter complies with the following typical naturalgas calculation standards:AGA8 Detail Characterization Method

Gross Characterization Method 1Gross Characterization Method 2

ISO 12213: 1997 First edition (Dec. 1, 1997)Part 2: molar-composition analysisPart 3: physical properties

(4) Custom fluidsUser-maintained physical property values can beincorporated in the transmitter.

STRUCTURE OF EJXMVTOOL MASS FLOWPARAMETER CONFIGURATION TOOL

Figure 6 shows the structure of the EJXMVTool mass flowparameter configuration tool used to set mass flow rateapplication information.(1) Flow rate parameter configuration function

This function comprises interactive graphic screens used toconfigure the flow rate calculation parameters shown in

Equation (1), including application information such asprimary devices and fluid types. Parameters can beconfigured easily as this function automatically converts theparameters to the transmitter’s format.

(2) Flow rate parameter management functionThis function saves flow rate parameters to a file, uploads anddownloads these files to and from the transmitter, obtainsresults of parameter calculation by the transmitter, andoutputs reports.

(3) General-purpose parameter management functionThis function sets general-purpose parameters such as rangesand units.

(4) Physical property databaseThe software contains a database of physical propertiesnecessary to calculate fluid density and viscosity.

(5) Field communications serverFor easy compliance with various methods of fieldcommunication, the software incorporates the same fieldcommunications server used in the PRM (Plant ResourceManager) field device management package.

FLOW OF APPLICATION INFORMATIONSETTING

As explained below, the operating screens of EJXMVToolhave been designed in association with specific applications.(1) Screen for setting primary device information

As shown in Figure 7, this screen is used to set the diameter ofan orifice or other primary device, the pipe diameter and thematerials.

(2) Screen for selecting fluid typeThis screen is used to select the fluid type to be measured.

(3) Screen for setting natural gas composition informationThis screen is used to set the composition information of anatural gas.

Figure 5 Comparison of Mass Flow Rate Errors betweenFixed-parameter Calculation and Optimized Calculation

Figure 6 Structure of EJXMVTool Software

Err

or [%

of R

eadi

ng]

Mass flow rate [kg/h]

–4

–6

–2

0

2

4

0 75,000 150,000

Fixed-parameter calculationOptimized calculation

Database

Field communications server

Interactive editor

EJXMVTool Mass Flow Parameter Configuration Tool

Flow rate parameter configuration Calculation of parameters for EJX910

Flow Rate Parameter Management

General-purpose Parameter

ManagementDownload to transmitter

PV monitorTags

RangesUnits

DampingLow-cut

Display settingMaintenanceAdjustment

Primary device information

Upload from transmitter

File management

Reporting

Flow rate calculation standards

Fluid types and operating conditions

Physical property databaseLow rate parameter

acquisition

Page 4: Yokogawa Multivariable Transmitter EJX910

16 Yokogawa Technical Report English Edition, No. 42 (2006)

(4) Screen for specifying fluid pressure and temperature rangeThis screen is used to specify fluid pressure and temperaturerange, which are necessary to calculate density or other dataitems.

(5) Screen for setting fluid density and viscosityAs shown in Figure 8, this screen is used to verify data to setfluid density and viscosity. This screen can be customized.

(6) Screen for downloading application informationThis screen is used to convert user-input primary deviceinformation and fluid information into a transmitter-specificformat and then download that information to the transmitter.

(7) Screen for simulated flow rate calculationsThis screen is used to verify the results of simulating massflow calculations by the transmitter with the sensor inputs inFigure 9 applied as simulated inputs, in order to predeterminethe results of applying the downloaded information.

(8) Screen for verifying flow rate calculations in an actual plantThis screen is used to verify the results of flow ratecalculation under actual plant operation with the respectivesensor inputs in Figure 9 actually applied.

FIELD TEST RESULTS

Figure 10 shows the results of an actual flow test performedon a British natural gas test line. The test results show a mass flowrate measurement accuracy level as excellent as 1%. Users have

also highly evaluated the EJX910 in other actual flow tests.

CONCLUSION

In future we intend to further accumulate experience in massflow rate measurement using multivariable transmitters, whilebroadening the range of applications to encompass an evengreater variety of primary devices and fluids. In addition to massflow rate measurement, we will further develop fluid levelmeasurement and multivariable measurement-based processdiagnosis.

REFERENCES

(1) ISHIKAWA Tamaki, et al., “New DPharp EJX Series ofPressure and Differential Pressure Transmitters,” YokogawaTechnical Report, No. 37, 2004, pp. 9-14

(2) KUROMORI Ken-ichi, “Industrial Flowmeters—RecentTopics—.” Measurement & Control, Vol. 42, No. 12, 2003,pp. 1015-1020 (in Japanese)

* ‘DPharp’, ‘EJXMVTool’ and ‘PRM’ are registered trademarksof Yokogawa Electric Corporation. All other product names,including software names, appearing in this document are thetrademarks or registered trademarks of their respectivecompanies or groups.

–2

–1

0

1

2

0 75,000 150,000

Mass flow rate [kg/h]

Err

or [%

of R

eadi

ng]

Figure 10 Results of an Actual Natural Gas Flow TestFigure 8 Settings of Fluid Density and Viscosity

Figure 7 Settings of Primary Device Information Figure 9 Verification of Results of Flow Rate Parameter Calculations