Slide 1
Anritsu Master series OTDR
Presents:
The next generation inOTDRs
TheACCESS Master
MT9083A / B / C
TheNetwork Master
MT9090A
Slide 2
Anritsu Master series OTDR
OTDR 101
Before we start, lets review key
OTDR principles and terminology
8/11/2019 Anritsu OTDRs
2/45
Slide 3
Anritsu Master series OTDR
OTDR Operation
Bloc k Diagram
Slide 4
Anritsu Master series OTDR
d = distance
t = elapsed time
c = speed of light in a vacuum
n = Index of Refraction
Test pulse fired
Test pulse returned
d
d=t C
2 n If n is incorrect, thenthe distance measuredwill also be wrong!!
Must divide by
two to get one
way elapsed time.
DistanceMeasurements
8/11/2019 Anritsu OTDRs
3/45
Slide 5
Anritsu Master series OTDR
Fiber PhenomenaBackscattering
Transmitted Light
Random scattering
Backscattering
Fresnel Reflection
Slide 6
Anritsu Master series OTDR
Fiber Type
Multimode Fiber
Wavelength: 850nm and 1300nm
Short distance application such as in-building Ethernet
connection Graded-Index support higher data speed than Step-index
Singlemode Fiber
Wavelength: 1310 and 1625nm
High data speed over long distance application. Metro
Network, Long-haul network.
8/11/2019 Anritsu OTDRs
4/45
Slide 7
Anritsu Master series OTDR
Fiber Type
Slide 8
Anritsu Master series OTDR
Fiber Parameter
IOR (Index Of Refraction) obtain from fiber manufacturer, enter into OTDR
Affecting the measured cable distance
Backscatter coefficient obtain from fiber manufacturer, enter into OTDR
Affecting the measured reflectance.
8/11/2019 Anritsu OTDRs
5/45
Slide 9
Anritsu Master series OTDR
OTDR BasicsTypical OTDR trace with events
Slide 10
Anritsu Master series OTDR
OTDR Basics
Deadzones Why event deadzone matters
Verification of two close reflections
Provides ability to verify which
connector is bad (high reflectance)
With 80 cm deadzone, MT9083A can
detect two connectors at 1 m apart.
Why backscatter deadzone matters
Verification of a second event after areflective event
Provides ability to test vault splices
Average MT9083A spec
is 5 m (15 ft)
EDZ
ADZ
Mechanical connectors
Mechanical connectors
Fusion
splice
8/11/2019 Anritsu OTDRs
6/45
Slide 11
Anritsu Master series OTDR
OTDR BasicsEvent Deadzone
Connectors less than 80cm apart Connectors more than 1m apart
Slide 12
Anritsu Master series OTDR
OTDR BasicsBackscatter Deadzone
Connector Splice
5m backscatter deadzone
Undetectable splice
Connector Splice
5m backscatter deadzone
Detectable splice
8/11/2019 Anritsu OTDRs
7/45
Slide 13
Anritsu Master series OTDR
OTDR Basics
10m (33)Actual Splice Location
Measured Splice Location
Actual End Location
Measured End Location
Low Density (wider Data Point Sampling)
Resolution
Slide 14
Anritsu Master series OTDR
OTDR Basics
3m (10)Actual Splice Location
Measured Splice Location
Actual End Location
Measured End Location
High Density (narrow Data Point Sampling)
Resolution
8/11/2019 Anritsu OTDRs
8/45
Slide 15
Anritsu Master series OTDR
OTDR BasicsDynamic Range: How far can you measure.
Dynamic Range OK to measure
entire fiber.
Insufficient Dynamic Range
events near the end of the fiber
cannot be measured.
Slide 16
Anritsu Master series OTDR
OTDR BasicsHow far can our modules test?
Converting Dynamic Range into Distance
Can only be done on a straight piece of glass (no events)
Need to know the db/Km loss for each wavelength
Need to know the dynamic range at each wavelength.
Distance = Dynamic Range = 38dB/0.20dB/km = 190km
dB/km
Wavelength Typical Loss
850nm multimode 3.0dB/km
1300nm multimode 1.0 dB/km
1310nm single mode 0.35 dB/km
1550nm single mode 0.20 dB/km
1625nm single mode 0.25dB/km
8/11/2019 Anritsu OTDRs
9/45
Slide 17
Anritsu Master series OTDR
OTDR BasicsPulse Width
The single most important parameter Use the pulsewidth which still provides sufficient dynamic range
Too short
(short deadzonesbut insufficient DR)
Too long
(good DR
but large
deadzones)Just right
(good DR, short
deadzones)
Slide 18
Anritsu Master series OTDR
Pulse Width (PW), what does it effect, Smaller PW = Smaller Dead Zones,
Able to see close events,Not able to see longer distances. OTDRs specification deadzone is tested at the smallest PW.
Larger PW = Larger Dynamic Range,Able so large distances,Not able to see close events. OTDRs specification Dynamic Range is tested at the largest PW.
Its not possible to test close events with a large PW or largedistances with a small PW. Using the Autofunction on your OTDR sets the PWto the cable
length.Pulse width Dead Zone Dynamic Range
10ns 2 - 3m 6 - 9 dB
100ns 10 - 15m 11 - 14dB
500ns 50m 1821dB
1us 100m 2225dB
4us 400m 29 - 32 dB
OTDR BasicsPulse Width
8/11/2019 Anritsu OTDRs
10/45
Slide 19
Anritsu Master series OTDR
Presents:
The next generation inOTDRs
TheACCESS Master
MT9083A / B / C
Slide 20
Anritsu Master series OTDR
Historical Fiber Optic SystemBasic point to point deployment
OP-AMP/RegenerationStations
City A
City B
City A
City B
Splice points
Yesterdays
networks werevery simple and
most test
equipment was
adequate.
8/11/2019 Anritsu OTDRs
11/45
Slide 21
Anritsu Master series OTDR
But networks are evolvingCurrent Situat ion with Op tical Netwo rks
Increased complexity
Shift from long-haul to access installations
Increase in Metro and FTTx installations
Triple play (including sensitive video applications)
And
Budget reductions have forced service providersto get the job done with less people, equipment,
knowledge and time
Shouldnt your test equipment evolve too?
Slide 22
Anritsu Master series OTDR
Introduce the all- in-one concept
Power Meter
and Light Source
High Performance OTDR
MT9083 Series
ACCESS Master
Visible LD
IP Test function
Less equipment to carry and maintain!
8/11/2019 Anritsu OTDRs
12/45
Slide 23
Anritsu Master series OTDR
Introducing the MT9083 seriesdesignedwith the features that matter most.
Dual mode design (HR/ER mode) Compact
Lightweight
Fool-proof ease of use
All day battery operation
Quick boot-up
High resolution, but also goodrange
Quick testing time
Reliable
Rugged Complete testing tool
Inexpensive
Slide 24
Anritsu Master series OTDR
MT9083B/C: Dual Circuit Design Standard, high resolution (HR)
Current operation
All pulsewidths
Use for best deadzone
Enhanced Range (ER) Optimized for PON
Optimized for range
50ns-2us
User selectable
8/11/2019 Anritsu OTDRs
13/45
Slide 25
Anritsu Master series OTDR
MT9083B/C: FTTx/PON Measurement Using ER
In ER mode, MT9083B/C allows
user to measure entire span Up to a 64/128-branched splitter!
Section A Section CSection
B
MT9083B/C can detect
fault location beyond
splitter very accurately.
Slide 26
Anritsu Master series OTDR
Best Technical specs where it counts Small Dead Zone to allow
measurement of near endconnectors. Within CO In riser cables Premise
5cm Resolution Event plotted accurately Up to 150,001 sampling
points
dB Range also allows for over100km measurement range. Up to 45dB (1310/1550nm) Most Telco long haul links
are from 70km to 100km.
8/11/2019 Anritsu OTDRs
14/45
Slide 27
Anritsu Master series OTDR
Best Technical specs where it counts One touch Fault Locate mode,
Automatically sets all the Opticalparameters after a single touchthe user has the results.
Will offer a result with theminimum required averages.
When in Standard OTDR mode, By pushing the Start (Green)
button all optical parameters willbe configured.
Lost Test Mode, Allow for simple interface to
Power Meter and Light Sourcemeasurements.
High power meter option of+30dBm
Slide 28
Anritsu Master series OTDR
Other Technical specs Battery life
8hours of use battery life.
Boot-up time Ready to use in 15sec
Engineers dont waste time on thejob.
Easy file transfer Simple drag and drop file transfer via
USB.
Latest OTDR file format SR-4731 (issue 2).
PC Emulation software available, Multiple version of PC software
available included reporting softwaredownloadable from the web.
System upgrade software available fromthe web. Download the latest software from the
Anritsu web site.Your OTDR is never out of date.
8/11/2019 Anritsu OTDRs
15/45
Slide 29
Anritsu Master series OTDR
High Performance, Field Tested Hardware
Dedicated keys
SetupsSave/recall
Contrast
Number pad
Rotary encoder
Precise marker placement
Easy menu navigation
START button
True one-button fault
location
All parameters selected
Perfect for novice users
Dedicated softkeys
Quick, easy operation
ParametersWavelength
Display modes
Rugged Protector OptionSide bumpers
Rigid display cover
Shoulder strap
Easy to read, TFT color
displayStandard display or enhanced
indoor/outdoor display
LRUD keypad(left,
right, up, down) for
easy operation
User Friendly
Rugged Small
Lightweight
Easy to use
Impressive Battery
operation
8 hour typical
6 hour continuous
LiIon no memory
Slide 30
Anritsu Master series OTDR
Dedicated Testing Interfaces Clearly labeled Integrated dust caps A true all-in-one for optical fiber
construction and maintenance
OTDR ports
Low cost, interchangeable
adapters
Built-in light source (std)
Integrated power meter (std)
SM/MM
Optional power meters
Interchangeable adapters
For loop-back testing
+3 to -70 dBm (SM/MM)
+23 to -50 dBm (SM)
+30 to 43dBm (SM)
Visible laser diode
option
~0 dBm output
2.5mm universal
~5 km (3 mile) range
CW or modulated
Dual USB ports
Plug & play
mass storage
IP testing Option
RJ45 electrical
10/100MB
1000MB upgrade
Ping, download test,
throughput, frame loss
8/11/2019 Anritsu OTDRs
16/45
Slide 31
Anritsu Master series OTDR
MT9083B/C: Waveform Comparison Using ER Mode
MT9083B/C can capture
waveform accurately using
short pulse for entire fiberpath through PON splitter.
Electrical noise
Bad
Bad
Bad
Good!
Competitor
A
Competitor
B
Competitor
C
Slide 32
Anritsu Master series OTDR
MT9083B/C: Waveform Comparison Using ER Mode Competitor may be able to
see waveform beyond
splitter by using long
pulse. However
Zoom in
Very sharp waveform even
between close splitters
OK
?
Undefined waveformdue to poor dead zone
performance
Not
OK
OK! CompetitorA
Competitor
B
8/11/2019 Anritsu OTDRs
17/45
Slide 33
Anritsu Master series OTDR
MT9083B/C: Longest Distance Using ER Mode
MT9083B/C providessuperior range (100km) at100ns with excellent deadzone performance.
100km
100km
100km
100km
(with 100ns
Pulse)
Competitor
A
Competitor
B
Competitor
C
Slide 34
Anritsu Master series OTDR
MT9083B/C: Best Dead Zone Using HR Mode
There are faults our competitors can not see.
MT9083B quickly identifies location of faults at the installationand reduces risk of future trouble.
ANRITSU
15m
25m30m
MT9083B/C is the only
OTDR that can detect a
fault in this zone.
Competitor
B
Competitor
C
8/11/2019 Anritsu OTDRs
18/45
Slide 35
Anritsu Master series OTDR
Access Master MT9083[ ] Lineup
Opt Wavelength MT9083A MT9083B MT9083C
053 1310/1550 42/41 45/45
055 1310/1550/1650 37.5/36/33.5 42/41/35
056 1310/1490/1550 36/34.5/34.5 40/39/39
057 1310/1550/1625 36/34.5/31.5 40/39/38 45/45/43
058 1310/1490/1550/1625 38/37/37/36
059 1310/1383/1550/1625 38/37/37/36
063 M850/1300/S1310/1550 28/27/38/36.5 29/28/42/41
064 MMF850/1300 28/27
073 1310/1550 38/36.5
A = 35dB class
B = 40dB class
C = 45dB class
Slide 36
Anritsu Master series OTDR
Easy to use, task-oriented top menu
Dedicated testing
modes simplify
operation
1. Select the testing
to be performed
2. The GUI is then
optimized to
perform it
3. Saves time,
increases
productivity
Upon boot-up or whenthe Top Menu key ispushed, the Top Menuwill appear. A testingfunction to be performedcan be chosenimmediately.
8/11/2019 Anritsu OTDRs
19/45
Slide 37
Anritsu Master series OTDR
Fault Locate Mode for novice users
True one-button fault location
Quick test for cable damage
Perfect for CO fault locationReduces MTTR
Ideal for inexperienced users
Launch condition verified
Test parameters automatically selected
Quick data averageTypically less than 10 seconds
Easy to read results for even the most
inexperienced user
Includes break/end location, span loss Full trace view if desired
Slide 38
Anritsu Master series OTDR
Perform a Fault Locate test (1/2)1. Clean and connect fiber
2. Highlight fault locate using the
arrow keys & press enter
3. Connection check immediately
verifies connection
1. Press continue to begin test if
status bar is black
2. Clean & reconnect if red
3. User can not start test i f f iber is
conn ected incorrect ly nochance of incorrect resul ts
4. Automatically test sequence begins
1. longest wavelength is selected
2. testing parameters selected
3. Fiber test begins
1
SM OTDRMM OTDR
2
8/11/2019 Anritsu OTDRs
20/45
Slide 39
Anritsu Master series OTDR
Perform a Fault Locate test (2/2)
1. Results displayed in seconds1. End/fault location2. Event/loss3. Total span loss
4. Full event table with end fault
highlighted
2. Select view trace for more
information (if desired)1. Full event table2. Full trace
3. Or connect next fiber and
press
Slide 40
Anritsu Master series OTDR
Standard OTDR Modefor any task
Flexible operation for any skill level or task
Customize the unit Units of measure File save format File save location Reflectance calculation
Complete event table control Break/end threshold Loss specification Reflectance specification End to end loss
Full functionality OTDR testing Multiple trace comparison Automated or manual parameter
selectionyour choice Selectable display modes - from A, B,
Origin or Anywhere (zoom box)
8/11/2019 Anritsu OTDRs
21/45
Slide 41
Anritsu Master series OTDR
Run a quick test Press f1 and
All parameters are automatically selected
+
=
Complete analysislocation, type,
loss, pass/fail in seconds!
Slide 42
Anritsu Master series OTDR
Customize Test parameter(1/2) From the top menu, select
Press f4 Select 1km (3.3kft), low density, 3 ns
Press
Press Averaging Time, enter 10 sec, then
8/11/2019 Anritsu OTDRs
22/45
Slide 43
Anritsu Master series OTDR
Customize Test parameter (2/2) Press , then f1
Slide 44
Anritsu Master series OTDR
Real Time Test Select 1550 or 1625 nm
Select 2.5 km, low density, 10 ns & press
Contract as shown using the arrow keys Comment on the screen update rate (0.15 sec)
While still in real time, bend the fiber near the OTDR The trace will drop out
Release the bend
Trace will restore instantly
8/11/2019 Anritsu OTDRs
23/45
Slide 45
Anritsu Master series OTDR
All fiber information is on a single screennoneed to flip through tabs Battery status (%
and time remaining)Current
testing mode
Axis scaling
A and B
measurement
cursors
Softkeys
Loss mode
type and
results
Complete event
table:
Event #
Location
Type
Loss
Reflectance
Segment dB/km
Cumulative loss
Key testing parameters:
Wavelength
Distance range
PulsewidthIndex of refraction
Resolution
Number or averages
Slide 46
Anritsu Master series OTDR
Interpreting Event Table
1
2
3
8/11/2019 Anritsu OTDRs
24/45
Slide 47
Anritsu Master series OTDR
Construction OTDR Mode for cable instal lat ion
Automated testing of multiple fibers at multiple wavelength
Enter the information oncesaves times
Each fiber tested, named and saved automatically
Once the single page set-up is complete, the software becomes
the project manager
Identifies which fiber to connect
Verifies connection
Ensures consistency in filename and parameters
Eliminates user errors such as
missing files
Inconsistent naming Inconsistent parameter selection
Slide 48
Anritsu Master series OTDR
Perform a Construction Mode Test (1/2)
1. Select file storage location
2. Determine base file order
(example shown in filename
field)
3. Enter test site locations
4. Use other field for cable
ID, project, customer, etc.
5. Select wavelengths
6. Select testing options
7. Enter number of fibers to
test
1 2 3 4
5
67
8/11/2019 Anritsu OTDRs
25/45
Slide 49
Anritsu Master series OTDR
Perform a Construction Mode Test (2/2)
Once set-up is complete, connect thefiber as instructed and press
continue
Connection level is verified
Optimize and press continue if
bar is black
Clean and reconnect if red
Sit back and watch up to four
wavelengths tested, analyzed and
savedwithout pressing a button
Connect next fiber and presscontinue
Slide 50
Anritsu Master series OTDR
Software Overview Set-ups always available from front panel 5
Neatly arranged into categories
Notes explain each setting
Simply one-button
software upgrades via
USB!
8/11/2019 Anritsu OTDRs
26/45
Slide 51
Anritsu Master series OTDR
Loss Test Set Modefor opt ical bud get veri f icat ion
Stabilized light source standard More cost effective than buying stand-alone handheld
sources
CW, 1kHz or 2kHz output modes
Universal connector easily adapts to all connector
styles same port as the OTDR
High (-8 dBm) output provides exceptional dynamic
range for testing even the longest fibers (-8 dBm output) - (-65 dB sensitivity) = 57 dB of range
Integrated power meter standard
-5 to -50 dBm
High power CATV range (+20 to55 dBm) optional
Data table for easy storage of results
Slide 52
Anritsu Master series OTDR
Perform Optical Loss Test Set (OLTS) function
Select same wavelength for
Power Meter and Light source
Set Reference Connect patch cord between power
meter and light source through an
adapter.
Turn on the light source and record
the power in the Reference field
Connect patch cord to cableunder test
Measure loss
8/11/2019 Anritsu OTDRs
27/45
Slide 53
Anritsu Master series OTDR
Reference Procedure for Loss Measurement
Use two short patch cable
for referencing
Connect the short patch
cable to both end of the fiber
under test.
Light Source Access Master in
Power Meter Mode
Slide 54
Anritsu Master series OTDR
USB connection Can be used on PCs with USB
port
Fiberscope for x200/400
magnification
Adaptors for 1.25 & 2.5mm
patch cord.
Universe connectors including
APC
Image saving function
Video Inspection Probe (VIP)
8/11/2019 Anritsu OTDRs
28/45
Slide 55
Anritsu Master series OTDR
Visible Fault Locator Option
Integrated visible redlight fortroubleshooting within the OTDR dead
zone
Provides easy identification of a
particular fiber
Select output mode: CW or 2 Hz
modulated
Universal 2.5 mm connector fits all
standard size connectors
0dBm (1mW) max output
~5km (3 mile) range
Slide 56
Anritsu Master series OTDR
Dedicated field test solution for installation and
maintenance of optical links
OTDR Tester.
8/11/2019 Anritsu OTDRs
29/45
Slide 57
Anritsu Master series OTDR
Introducing the MT9090A with
MU909014x/15x OTDR Module
A truly revolutionary OTDR!
Pocket-sized
Easy Operation
Excellent OTDR Performance
Unique Battery Operation
Priced Right
Slide 58
Anritsu Master series OTDR
Overview. Product overview,
Operation, Performance and Options,
Specs,
Comparison to MT9083x,
8/11/2019 Anritsu OTDRs
30/45
Slide 59
Anritsu Master series OTDR
Value without compromise.
OCA
(Optical Channel
Analyzer for CWDM)
DCFL
(Drop Cable Fault Locator)
GiGE
(Gigabit Ethernet test)
Part of an ever-expanding platform!???
Modular platform ensures
maximum return on investment!
OTDR
Slide 60
Anritsu Master series OTDR
For Highly Portable Field Use. Easily held with one hand,
Palm-size (190mm x 96mm x 48mm), Approximately 700g (1.54 lbs).
Battery operation, Up to 8 hour operation
(with STD Battery pack), 4 hour recharge
(with STD Battery pack), NiMH battery pack or AA Alkaline Dry Battery
operation.
Rugged/sealed design, No vents or fans, Wrist strap to prevent accidental drops.
Color indoor/outdoor display standard, 4.3 high resolution, Landscape format naturally fits OTDR trace.
8/11/2019 Anritsu OTDRs
31/45
Slide 61
Anritsu Master series OTDR
The Foolproof testing device.
Easy to use,
True one-button fault location,
Automatic testing parameters selection in
Full-automatic mode (Full-manual
operation also available),
Quick fiber evaluation,
Event analysis table and full Trace view.
Quick boot-up,
Ready to test in
8/11/2019 Anritsu OTDRs
32/45
Slide 63
Anritsu Master series OTDR
Complete Optical Fiber Testing. Excellent OTDR Performance,
Capable of,Full parameter selection (5ns-20us),125,000 data points,Up to 37dB Dynamic range,High resolution accuracy,Dual wave 1310/1550nm config,Complete testing from Access to Core.
Complete PON testing tool,High resolution and extremely shortdead zones ensure entire PON (1X32, 1X64 branches)network evaluation,ITU-T L.41 compliant - 1625nm/1650nmwavelength for in-service maintenanceof PONs without external filters.
Slide 64
Anritsu Master series OTDR
1:32 Splitter
2km Fiber4km Fiber
2.2km Fiber
Complete PON Measurement for I&M.
8/11/2019 Anritsu OTDRs
33/45
Slide 65
Anritsu Master series OTDR
Long Battery Life,
Up to 8 hours / 4 hour recharge,
with Standard NiMH Battery pack (Supplied with the original
unit),
Alkaline battery (AA)operation also available.
Batteries.
Available 4 x AA Type
NiMH
Or
Alkaline Dry Batteries
Slide 66
Anritsu Master series OTDR
Additional Highlights.
Built-in 10m (30ft) launch fiber,
Measure front-end connection
without additional patchcords,
Can be enabled or disabled.
Complete data management,
Up to 1,000 traces internal/10,000+ with USB,
Easy drag and drop transfers to PC,
Reporting and printing with NETWORKS or
TraceView PC software.
8/11/2019 Anritsu OTDRs
34/45
Slide 67
Anritsu Master series OTDR
Optional Features. Visual Fault Locator Option,
Integrated 650nm laser,
Can also be used to identify a
particular fiber (in the splicing tray),
Can be used simultaneously with power meter or OTDR,
Up to 5km (3 mile) range.
Connector Inspection Option,
400x image displayed,
All software pre-loaded,
Eye-safe (Often an OH&S requirement),
Image save/recall,
Hardware includes probe/tips.
Slide 68
Anritsu Master series OTDR
Different Applications. FTTx,
Full network testing: ONT to OLT (incl. splitter(s)), In service maintenance (filters installed).
General fiberinstallation/certification, Premise/CO, Access, Metro, Core/Long Haul.
Backhaul, FTTc (fiber to the cell site), DAS, RRH.
Optical Fiber(when used Ethernet)
BaseTerminalStation :NodeB
Antenna
Optical
Fiber
BTS
RRH
RemoteRadioHead
Optical Drop CableFault Locator,OTDR
Optical Fiber Test
8/11/2019 Anritsu OTDRs
35/45
Slide 69
Anritsu Master series OTDR
MU909015B/B1-056/066 MU909014B/B1-056/066 MU909014A/A1-053/063 MU909014A/A1-054/064
1310/1550 20 nm * 1310/1550 20 nm * 1625 15 nm 1650 15 nm
PW=20sec 37/36dB 32.5/31dB 32.5dB*(1490/1550 cut filter included)
32.5dB*(1490/1550 cut filter included)
PW=500nsec 28/26dB 24.5/23dB 24.5dB*(1490/1550 cut filter included)
24dB*(1490/1550 cut filter included)
Modulation: CW, 1Hz
* Please see detailed on Product brochure
Fiber Type 10m/125m SMF (ITU-T G.652)
Loss Measurement mode 2 point loss, Splice loss, dB/km Loss LSA, ORL, Event
Environment Operating temperature and humidity: -5 to +40C*10,
8/11/2019 Anritsu OTDRs
36/45
Slide 71
Anritsu Master series OTDR
MU909011A Drop Cable Fault Locator Module
Slide 72
Anritsu Master series OTDR
PURPOSE-BUILT FOR TESTING SHORT FIBERS
Uniqueonly true short-fiber tester available
Perfect solution for FTTX drop cables, CO,
MDU risers, wireless towers & premise
Integrated 10m launch cord standard
Dynamic range of >7dB (10dB typ)
For testing up to 10km (~6 miles)
Unit will display transmission power levels,
then fault locate with the press of a single
button
Testing parameters are fixed
Operation simplified
Ensures proper settings and valid results
658 feet!
8/11/2019 Anritsu OTDRs
37/45
Slide 73
Anritsu Master series OTDR
Additional Features Integrated power meter option for FTTx
maintenance For measuring power to ONT Connect to MT9090A/MU909011A Verify power level and classify
PASSfiber OK, ONT faultFAILfiber damaged, launch fault locate
Power meter and fault locate from same port
Complete breakdown as to failure and responsibilityONT, drop cable or feed fiber
General purpose power meter -5 to -45dB range
Slide 74
Anritsu Master series OTDR
74
MT9090A SpecificationsMT9090A CONTROLLER SPECIFICATIONS
Display 4.3 inches TFT Color LCD (480 x 272 pixels, transmissive)
Interface USB (v1.1)
Battery type NiMH
Battery life time 3.5 hours (typ)
Battery charging time
8/11/2019 Anritsu OTDRs
38/45
Slide 75
Anritsu Master series OTDR
CMA5000aOTDRProduct Introduction
Slide 76
Anritsu Master series OTDR
CMA5000a The all-in-one, portable CMA5000a supports network
transmission and optical physical layer tests forinstallation and maintenance. Transmission
SDH/SONETEthernet
Optical physical layerFiber end testLoss test
OTDRChromatic dispersionPolarization mode dispersionSpectrum analysis
8/11/2019 Anritsu OTDRs
39/45
Slide 77
Anritsu Master series OTDR
Modular Structure One mainframe supports full range of versatile
measurements by changing modules
Transmission systems (modules)SDH/SONET (UTA, XTA)
Ethernet (UTA, GigE II)
Optical physical layer (modules)Fiber end face check (VIP)
Loss test (OTDR)
Optical Time Domain
Reflectometer (OTDR)
Chromatic dispersion
(CD-OTDR)Polarization mode dispersion (PMD)
Spectrum analyzer (OSA)
Slide 78
Anritsu Master series OTDR
Mainframe Types and Modules Two mainframes models allowing greater flexibility
SBA: Two bays up for up two modules
MBA: Four bays for up to four modules
Multiple measurement modules can be
set and operated simultaneously in
one mainframe.
1 bay used
GigE II
OTDR
CD-OTDR
SBA
2 bays used
UTA
XTA
Bays emptyUSB Connection
VIP
MBA
4 bays used
OSA
PMD
8/11/2019 Anritsu OTDRs
40/45
Slide 79
Anritsu Master series OTDR
CMA 5000a OTDR Avai lable Modu les
Model
Number
Wavelengths Dynamic Range
5261 QUAD SM/MM:
850/1300nm (62.5um)/1310/1550nm
24/26dB
35/35dB
5262 QUAD SM/MM:
850/1300nm (62.5um)/1310/1550nm
24/26dB
40/40dB
5266 850/1300nm62.5um 24/26dB
5269 850/1300nm50um 24/26dB
5283 1310/1383/1550/1625nm 35dB
5491 1310/1490/1550nm 40/40/40dB
5493 1310/1490/1550/1625nm 37dB
Model
Number
Wavelengths Dynamic Range
5225 1310/1550nm 37/36dB
5235 1310/1550nm 40/40dB
5236 1310/1550/1625nm 40/40/40dB
5245 1310/1550nm 43/45dB
5246 1310/1550/1625nm 43/45/43dB
5254 1550nm 50dB
5260 QUAD SM/MM:850/1300nm (50um)
/1310/1550nm
24/26dB35/35dB
Slide 80
Anritsu Master series OTDR
Key Features and B enefi ts
Unparalleled data sampling - Up to
256,000 data points
Twice as many as the nearest
competitor!
Unparalleled waveform creation
Ultra-fast real time screen update - 0.1
sec
Great for optimizing
splices/connectors
Gives perception of fast testing
Fully configurable to your customers
needs
Over 20 different optics choices
single mode or multimode, 25 to
50dB, dual, tri- or quad wavelengths
State of the art platform
Powerful PC, all common I/O
Active productcontinued
development
Designed for simplicity and ease of use
Dedicated testing modesno macros
One button testing (AUTO-TEST)
Minimal set-up time, maximum testing time
A strength for all skill levels
Superior trace linearity0.04 dB/dB
Provides greater measurement accuracy
and consistency
Overlay up to 8 traces simultaneous
Compare between wavelengths to seestresses effecting the cable
Compare between fibers in a cable to see
inconsistencies
8/11/2019 Anritsu OTDRs
41/45
Slide 81
Anritsu Master series OTDR
Test ing Modes
Fault Locate One Button
Automated Testing Standard
Traditional OTDRTesting
Construction Automated Multi-
Wavelength Multi-Fiber Testing
Loss Test Set Integrated Power
Meter and LightSource
Visual Fault Locator Visible Red Light
Video Inspection Probe Connector
inspectionmicroscope
Slide 82
Anritsu Master series OTDR
Fault Locate Mode One Button Au tomated Test ing
One-button break/end locator
Quick test for cable damage
Ideal testing mode forinexperienced users Launch condition verified Test parameters automatically
selected Data averaged just long enough
to produce valid results Easy to read results for even the
most inexperienced user Includes break/end location,
span loss and distance pastprior event
8/11/2019 Anritsu OTDRs
42/45
Slide 83
Anritsu Master series OTDR
Standard OTDR Mode Tradi t ional OTDR Testing
Full functionality OTDR testing Multiple trace comparison Automated or manual parameter
selectionyour choice Selectable display modes - from A, B,
Origin or Anywhere (zoom box)
Customize the unit Units of measure File save format File save location Reflectance calculation
Complete event table control Break/end threshold Loss specification Reflectance specification
End to end loss
Slide 84
Anritsu Master series OTDR
Construct OTDR ModeAu tom ated Mult i -Wavelength Testing
Automated testing of multiple fiber
cables
Test multiple fibers at multiple
wavelength quickly and easily
Enter the information once and
test/save each fiber with a single key
press
Ensures consistency in filename andparameters
Combines the best of CMA 4000
Construction and Multiple
Wavelength modes
8/11/2019 Anritsu OTDRs
43/45
Slide 85
Anritsu Master series OTDR
Loss Test Set ModeOptional Power Meter and Light Source
Adds a stabilized mode to the OTDRlaser for point to point testing More cost effective than buying
stand-alone handheld sources
CW, 1kHz or 2kHz output modes
Universal connector easily adapts toall connector styles Uses the same UC-130-XX
adapters as the OTDR
High (-8 dBm) output providesexceptional dynamic range fortesting even the longest fibers (8 dBm output) (65 dB
sensitivity) = 57 dB of range
High power CATV range +20 to 55 dBm Data table for easy storage of results
Slide 86
Anritsu Master series OTDR
Visual Fault Locator ModeOptional Red Light Test
Visible redlight fortroubleshooting withinthe OTDR dead zone
Provides easyidentification of aparticular fiber
Select output mode: CW
or 2 KHz modulated
Universal 2.5 mmconnector fits allstandard sizeconnectors
0dBm (1mW) max output ~5km (3 mile) range
8/11/2019 Anritsu OTDRs
44/45
Slide 87
Anritsu Master series OTDR
NetWorks Data Processing UtilityThe ult imate data proc essing to ol
Integrated software utility for viewing,printing and analyzing OTDR data
Analyze data on any PC
Complete report generating tool
Uni- and Bi-directional splice loss
Exception
Fiber Acceptance
Trace Summary
Batch processing for high count fiber
cables Full analysis and Bellcore/Telcordia file
format processing
Anritsu Master series OTDR
Slide 88
Anritsu Master series OTDR
The New AnritsuWhy Choose Us?
Now includes Nettest
The portable OTDR market share leader
Over 30 years in the optical industry
Over 25 years OTDR experience
Stable and debt-free: your partner for thelong haul
Focused on core competencynot ajack of all trades
Solutions and support that enable you tobe more competitive and profitable
Superior sales and technical support:anytime, anywhere
Anritsu Master series OTDR
8/11/2019 Anritsu OTDRs
45/45
Slide 89
Anritsu Master series OTDR
The right productplus the support yo u deserve
Easy and always FREE software upgrades Simply, one time registration
https://www1.anritsu.co.jp/Download/MService/Login.asp
Dedicated, knowledgeable technical support Via our worldwide network.Email : [email protected]
Local service centers Via our worldwide network.Email : [email protected]
Anritsu Master series OTDR
Anritsu Master series OTDRAnritsu Master series OTDR
https://www1.anritsu.co.jp/Download/MService/Login.aspmailto:[email protected]:[email protected]:[email protected]:[email protected]://www1.anritsu.co.jp/Download/MService/Login.asp