1
Bxxxx CYTO 2019 Pengfei Zheng 1 , Hongtai Gao 1 , Qiyao Wang 1 , Chuixin Liao 1 , Nan Li 2* , Xiaobo Wang 2 1. R&D, Agilent Biosciences (Hangzhou) Co. Ltd, Hangzhou, People's Republic of China 2. R&D, Agilent Technologies, San Diego, CA, United States * Contact: Nan Li, Ph.D [email protected] Continuous Sampling with High Absolute Counting Accuracy in Flow Cytometry Visit Agilent Technologies at Virtual Exhibit Hall of CYTO 2020 and visit our website at www.aceabio.com Introduction The primary function of the fluidics system in a flow cytometer is to transport the cells or microparticles from the sample tube to the flow chamber where they interrogate with the laser beam one by one for flow cytometry measurement. Conventionally, the sample and sheath fluidics are driven by air pressure supplied by a compressor. Thus, a tight seal on the sample tube is required. With this design, the sample volume cannot be accurately monitored. Therefore, the absolute counting can only be achieved by an indirect method of adding the known concentration reference beads into the sample to derive the true concentration of the cells in the sample. Recent new flow cytometers use the syringe pumps to drive the sample volume accurately, therefore direct volumetric absolute counting is achieved with high accuracy. The drawback of current syringe pump driven approach is that it takes a series of aspiration and injection actions to drive the sample from the sample tube to the flow chamber for analysis (such a cycle of action to aspirate and inject is called one stroke). This process is not continuous and takes longer to analyze samples of large volume when it is more than one stroke volume of the syringe pump. To conquer this issue, some systems use multiple syringe pumps (such as BD™ High Throughput Sampler (HTS)) to run alternatively to increase the sampling throughput. However, this increases the complexity of the system and the carryover is inevitably increased since it takes complex steps to clean the sampling line. In this study, we explore a new sampling line design with innovative fluidics control method to realize the continuous sampling using one syringe pump. With the new approach, the volume between the syringe pump and the three-way sampling valve (between the syringe pump and the flow chamber) is significantly reduced (delta V). The sample injection probe is immersed inside the sample tube all the time. The syringe pump repeatedly aspirates the volume of delta V from the sample and injects it into the flow chamber, until the final sample volume is injected and analyzed. During this procedure, minimal air gap is introduced into the sampling line and the sampling acts in a continuous mode, similar to the conventional compressed air driven system. By using the syringe pump, it can still achieve volumetric absolute count with guaranteed accuracy. Therefore, no costly reference counting beads are needed. The sampling line of a NovoCyte Quanteon flow cytometer is modified and tested in this study. The signal Mean Fluorescence Intensity (MFI) and Coefficient of Variation (CV) remain almost the same, and the accuracy of direct volumetric absolute count is maintained. Unlike the conventional aspiration injection method using the syringe pump sampling, the new approach can continuously drive the sample for analysis, therefore the assay throughput can be improved and flexible sample volume to be acquired can be used, without sacrificing the data quality of and accuracy of the direct absolute count. Results and Conclusions Principle and Method Procedure of Continuous Sampling Conclusions and Discussion 1. Power on solenoid valve. The common port (COM) is connected to normal-closed port (NC). 2. Sample Injection Probe 1 moves downwards to the sample tube. 3. During the course of Sample Injection Probe 1 moving downwards, the syringe pump aspirates 10 μL air from the tip of the Sample Injection Probe 1. Also, the shaker mixes the sample. 4. Syringe pump aspirates the sample into the Sample Injection Probe 1 and sample tubing until the front end of sample in the sample tubing reaches point . Define the sample volume of sample in tubing between point and COM port of the solenoid valve as ΔV. 5. Power off solenoid valve. The common port (COM) is connected to normal-open port (NO). 6. Syringe pump pushes the sample in the tubing between point and COM port of the solenoid valve into flow chamber with the sample flow rate that set by user. Start flow cytometer data acquisition. 7. If ΔV volume of the sample is used up before the Stop Condition is reached, the solenoid valve switches to the NC port, and instrument will repeat the step 4 to 6. The Sample Injection Probe 1 is immersed inside the sample tube all the time. 8. When the Stop Condition is reached or if user stops the test, syringe pump will push the sample remaining in the sample tubing and Sample Injection Probe 1 back to the sample tube. Sample Injection Probe 1 will then move upwards to the reset position. 9. Consequent sampling line washing procedure will follow. CYTO 2020 Poster: P182 Flow Chamber Sample Tube Sample Injection Probe 1 Sheath Fluid Waste Solenoid valve Syringe pump Sample Tubing COM NC NO Sample Injection Probe 2 Shaker The fluidics system of a NovoCyte® Quanteon flow cytometer is modified with the schematic illustration shown above. In order to reduce the time consumed by aspirating sample, the inner volume of sample injection probe 1 and sample tubing should be as small as possible. In order to save sample, the inner volume of sample injection probe 2 also should be as small as possible and the dead volume of the solenoid valve should be close to zero. Continuous Sampling Mode Normal Sampling Mode Comparison of Signal Mean Fluorescence Intensity (MFI) and Coefficient of Variation (CV) Sampling Mode E2 MFI FSC-H E2 MFI SSC-H E2 MFI FSC-A E2 MFI SSC-A E2 CV FSC-H E2 CV SSC-H E2 CV FSC-A E2 CV SSC-A Normal Sampling Mode 123,294 799,643 150,628 1,022,886 0.74% 0.78% 0.74% 0.90% Continuous Sampling Mode 123,314 801,696 150,732 1,026,012 0.74% 0.79% 0.74% 0.91% Deviation +0.02% +0.26% +0.07% +0.31% / / / / 1. In both mode, SPHERO™ Ultra Rainbow Fluorescent Particles are used for testing. Tthe Sample Flow Rate is 14 μL/min and Stop Condition is 12,000 Events in E1. 2. In Normal Sampling Mode, syringe pump will aspirate V volume of sample into instrument in one stroke depending on the capacity of the syringe pump (V=100 μL in this test). In the case of “# of Events” and “Time” as the Stop Condition, since the concentration of the sample is unknown, the full V volume will be aspirated. If the Stop Condition is reached before the V volume of sample is used up, this test is stopped. The rest of sample will be discarded as waste or be recovered (if the instrument has sample recovery function). If the Stop Condition is not reached by one stroke (V volume of sample is used up), the syringe pump will take another stroke and continue sample acquisition until the Stop Condition is reached. 3. In Continuous Sampling Mode, syringe pump will aspirate ΔV (= 20 μL) volume of sample into instrument in one stroke. If ΔV volume of sample is used up but the Stop Condition has not reached, the system will repeatedly aspirate ΔV sample until the Stop Condition is reached. 4. The data show that with Continuous Sampling Mode, the signal quality (i.e. Mean Fluorescence Intensity (MFI) and Coefficient of Variation (CV) ) remains unchanged from the Normal Sampling Mode. Normal Sampling Mode Continuous Sampling Mode Sample Flow Rate = 14 μL/min Stop Condition: 90,000 Events in R1 Sample: SPHERO TM AccuCount Particles SPHERO TM is a trademark of Spherotech, Inc. Comparison of Absolute Counting Mode Normal Sampling Mode (events/μL) Continuous Sampling Mode (events/μL) Test 1 959 968 Test 2 954 962 Test 3 951 968 Average 955 966 CV 0.42% 0.36% Target Value(events/uL) 967 Deviation from the Target Value -1.28% -0.10% A syringe pump-based continuous sampling method is achieved with this presented work. The method is validated to be effective and efficient based on the following results: 1. The signal quality, qualified by the Mean Fluorescence Intensity (MFI) and Coefficient of Variation (CV), remains the same compared to normal sampling mode using fluorescent calibration beads. 2. The Volumetric Absolute Counting results are the same compared to the normal sampling mode using the absolute counting beads. The advantages of this method include: 1. Reduces the sampling time, especially for high volume of sampling. 2. Reduce the sampling overhead volume usually associated with the syringe pump-based system. 3. Still maintains the advantages of the volumetric absolute counting enabled by the accurate volume dispensing capability using syringe pump. Signal Quality Volumetric Absolute Counting 1. The same sample is prepared and tested using two sampling modes for comparison. 2. The target absolute counting value is certified by the beads supplier. 3. The sample is run under each mode for three times and the Volumetric Absolute Counting results are obtained by the average of three tests. 4. The Volumetric Absolute Counting from the continuous sampling mode is no difference from the normal sampling mode. For Research Use Only. Not for use in diagnostic procedures. Fluidics System Schematics

Continuous sampling with syringe pump-Poster · 2020. 7. 15. · Bxxxx CYTO 2019 Pengfei Zheng1, Hongtai Gao1, QiyaoWang1, Chuixin Liao1, Nan Li2*, Xiaobo Wang2 1. R&D, Agilent Biosciences

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

  • Bxxxx

    CYTO 2019

    Pengfei Zheng1, Hongtai Gao1, Qiyao Wang1, Chuixin Liao1, Nan Li2*, Xiaobo Wang2

    1. R&D, Agilent Biosciences (Hangzhou) Co. Ltd, Hangzhou, People's Republic of China 2. R&D, Agilent Technologies, San Diego, CA, United States

    * Contact: Nan Li, Ph.D [email protected]

    Continuous Sampling with High Absolute Counting Accuracy in Flow Cytometry

    Visit Agilent Technologies at Virtual Exhibit Hall of CYTO 2020 and visit our website at www.aceabio.com

    Introduction

    The primary function of the fluidics system in a flow cytometer is to transport the cells or microparticles from the sample tube to the flow chamber where they

    interrogate with the laser beam one by one for flow cytometry measurement. Conventionally, the sample and sheath fluidics are driven by air pressure supplied by

    a compressor. Thus, a tight seal on the sample tube is required. With this design, the sample volume cannot be accurately monitored. Therefore, the absolute

    counting can only be achieved by an indirect method of adding the known concentration reference beads into the sample to derive the true concentration of the

    cells in the sample. Recent new flow cytometers use the syringe pumps to drive the sample volume accurately, therefore direct volumetric absolute counting is

    achieved with high accuracy. The drawback of current syringe pump driven approach is that it takes a series of aspiration and injection actions to drive the sample

    from the sample tube to the flow chamber for analysis (such a cycle of action to aspirate and inject is called one stroke). This process is not continuous and takes

    longer to analyze samples of large volume when it is more than one stroke volume of the syringe pump. To conquer this issue, some systems use multiple syringe

    pumps (such as BD™ High Throughput Sampler (HTS)) to run alternatively to increase the sampling throughput. However, this increases the complexity of the

    system and the carryover is inevitably increased since it takes complex steps to clean the sampling line.

    In this study, we explore a new sampling line design with innovative fluidics control method to realize the continuous sampling using one syringe pump. With

    the new approach, the volume between the syringe pump and the three-way sampling valve (between the syringe pump and the flow chamber) is significantly

    reduced (delta V). The sample injection probe is immersed inside the sample tube all the time. The syringe pump repeatedly aspirates the volume of delta V from

    the sample and injects it into the flow chamber, until the final sample volume is injected and analyzed. During this procedure, minimal air gap is introduced into the

    sampling line and the sampling acts in a continuous mode, similar to the conventional compressed air driven system. By using the syringe pump, it can still achieve

    volumetric absolute count with guaranteed accuracy. Therefore, no costly reference counting beads are needed. The sampling line of a NovoCyte Quanteon flow

    cytometer is modified and tested in this study. The signal Mean Fluorescence Intensity (MFI) and Coefficient of Variation (CV) remain almost the same, and the

    accuracy of direct volumetric absolute count is maintained. Unlike the conventional aspiration injection method using the syringe pump sampling, the new

    approach can continuously drive the sample for analysis, therefore the assay throughput can be improved and flexible sample volume to be acquired can be used,

    without sacrificing the data quality of and accuracy of the direct absolute count.

    Results and Conclusions

    Principle and Method

    Procedure of Continuous Sampling

    Conclusions and Discussion

    1. Power on solenoid valve. The common port (COM) is connected to

    normal-closed port (NC).

    2. Sample Injection Probe 1 moves downwards to the sample tube.

    3. During the course of Sample Injection Probe 1 moving downwards, the

    syringe pump aspirates 10 μL air from the tip of the Sample Injection

    Probe 1. Also, the shaker mixes the sample.

    4. Syringe pump aspirates the sample into the Sample Injection Probe 1 and

    sample tubing until the front end of sample in the sample tubing reaches

    point①①①①. Define the sample volume of sample in tubing between point①①①①

    and COM port of the solenoid valve as ΔV.

    5. Power off solenoid valve. The common port (COM) is connected to

    normal-open port (NO).

    6. Syringe pump pushes the sample in the tubing between point ①①①① and

    COM port of the solenoid valve into flow chamber with the sample flow

    rate that set by user. Start flow cytometer data acquisition.

    7. If ΔV volume of the sample is used up before the Stop Condition is

    reached, the solenoid valve switches to the NC port, and instrument will

    repeat the step 4 to 6. The Sample Injection Probe 1 is immersed inside

    the sample tube all the time.

    8. When the Stop Condition is reached or if user stops the test, syringe

    pump will push the sample remaining in the sample tubing and Sample

    Injection Probe 1 back to the sample tube. Sample Injection Probe 1 will

    then move upwards to the reset position.

    9. Consequent sampling line washing procedure will follow.

    CYTO 2020

    Poster: P182

    Flow Chamber

    Sample Tube

    Sample Injection Probe 1

    Sheath Fluid

    Waste

    Solenoid valve

    Syringe pump

    Sample Tubing

    COM

    NCNO

    ①①①①

    Sample Injection Probe 2

    Shaker

    The fluidics system of a NovoCyte® Quanteon flow cytometer is modified

    with the schematic illustration shown above. In order to reduce the time

    consumed by aspirating sample, the inner volume of sample injection

    probe 1 and sample tubing should be as small as possible. In order to save

    sample, the inner volume of sample injection probe 2 also should be as

    small as possible and the dead volume of the solenoid valve should be

    close to zero.

    Continuous Sampling Mode

    Normal Sampling Mode

    Comparison of Signal Mean Fluorescence Intensity (MFI)

    and Coefficient of Variation (CV)

    Sampling ModeE2 MFI

    FSC-H

    E2 MFI

    SSC-H

    E2 MFI

    FSC-A

    E2 MFI

    SSC-A

    E2 CV

    FSC-H

    E2 CV

    SSC-H

    E2 CV

    FSC-A

    E2 CV

    SSC-A

    Normal

    Sampling Mode123,294 799,643 150,628 1,022,886 0.74% 0.78% 0.74% 0.90%

    Continuous

    Sampling Mode123,314 801,696 150,732 1,026,012 0.74% 0.79% 0.74% 0.91%

    Deviation +0.02% +0.26% +0.07% +0.31% / / / /

    1. In both mode, SPHERO™ Ultra Rainbow Fluorescent Particles are used for testing. Tthe Sample Flow Rate is 14 μL/min and Stop Condition is 12,000 Events in E1.

    2. In Normal Sampling Mode, syringe pump will aspirate V volume of sample into instrument in one stroke depending on the capacity of the syringe pump (V=100 μL in this test). In the

    case of “# of Events” and “Time” as the Stop Condition, since the concentration of the sample is unknown, the full V volume will be aspirated. If the Stop Condition is reached before

    the V volume of sample is used up, this test is stopped. The rest of sample will be discarded as waste or be recovered (if the instrument has sample recovery function). If the Stop

    Condition is not reached by one stroke (V volume of sample is used up), the syringe pump will take another stroke and continue sample acquisition until the Stop Condition is reached.

    3. In Continuous Sampling Mode, syringe pump will aspirate ΔV (= 20 μL) volume of sample into instrument in one stroke. If ΔV volume of sample is used up but the Stop Condition has

    not reached, the system will repeatedly aspirate ΔV sample until the Stop Condition is reached.

    4. The data show that with Continuous Sampling Mode, the signal quality (i.e. Mean Fluorescence Intensity (MFI) and Coefficient of Variation (CV) ) remains unchanged from the Normal

    Sampling Mode.

    Normal Sampling Mode

    Continuous Sampling Mode

    Sample Flow Rate = 14 μL/minStop Condition: 90,000 Events in R1

    Sample: SPHEROTM AccuCountParticles

    SPHEROTM is a trademark of

    Spherotech, Inc.

    Comparison of Absolute Counting

    Mode Normal Sampling Mode (events/μL) Continuous Sampling Mode (events/μL)

    Test 1 959 968

    Test 2 954 962

    Test 3 951 968

    Average 955 966

    CV 0.42% 0.36%

    Target Value(events/uL) 967

    Deviation from the Target Value -1.28% -0.10%

    A syringe pump-based continuous sampling method is achieved with this presented work. The method is validated to be effective and efficient based on

    the following results:

    1. The signal quality, qualified by the Mean Fluorescence Intensity (MFI) and Coefficient of Variation (CV), remains the same compared to normal

    sampling mode using fluorescent calibration beads.

    2. The Volumetric Absolute Counting results are the same compared to the normal sampling mode using the absolute counting beads.

    The advantages of this method include:

    1. Reduces the sampling time, especially for high volume of sampling.

    2. Reduce the sampling overhead volume usually associated with the syringe pump-based system.

    3. Still maintains the advantages of the volumetric absolute counting enabled by the accurate volume dispensing capability using syringe pump.

    Signal Quality

    Volumetric Absolute Counting

    1. The same sample is prepared and tested using two sampling modes for

    comparison.

    2. The target absolute counting value is certified by the beads supplier.

    3. The sample is run under each mode for three times and the Volumetric

    Absolute Counting results are obtained by the average of three tests.

    4. The Volumetric Absolute Counting from the continuous sampling mode

    is no difference from the normal sampling mode.

    For Research Use Only. Not for use in diagnostic procedures.

    Fluidics System Schematics