Development of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays

Cheng Su1, Lei Zhou1, Ning Weng1, Yu Wu1, Naren Chirmule2, Zhandong Don Zhong2

1: Research & Translational Sciences Biostatistics, Amgen Inc.

2: Clinical Immunology, Amgen Inc.

Page 2: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Outline

• Introduction of CAST

• Immunogenicity testing and Assay Cut Point

• The automated system

• Fit-for-purpose qualification test

• Conclusion

2

Page 3: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Customized Assay Statistical Tools (CAST)

3

Web base systems developed to automate specific types of repetitive experiments and intended to be used by scientific colleagues

• User friendly • Built-in statistical decision making• Outputs are report-ready

• Customization• data input format, analysis and output • user Interface

• Gain efficiency, ensure quality with reasonable system development cost

Page 4: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

4

CAST: Meeting Statistical Needs in Resource Limited Environment

3. CAST

Analysis and reports customized and automated for repetitive studies

Pros: quick, better quality and more flexible than 2; provides standards

Cons: only cost effective for repetitive studies, may fail on anomalies

2. Commercial Software

Scientific partners use software with/without training from statisticians

Pros: no statisticians FTE required

Cons: difficult for complex analysis, no quality control

1. Statistician hands-on analysis

Statisticians and scientific partners work closely together on individual studies

Design, analysis, report, interpretation

Pros: highest quality

Cons: FTE, turnaround time, subjectivity

To achieve significant efficiency gain with minimum quality loss

Page 5: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Immunogenicity Testing Strategy

5

Screening Ab Assay

Ab (-) Specificity Ab Assay

Screening NAb AssayAb (-)

NAb (-) Specificity NAb Assay

NAb (-) NAb (+)

≤DCP >DCP

≤ACP >ACP

≤ACP >ACP

≤ACP >ACP

Importance

• Immunogenicity may affect drug exposure, efficacy and safety

• Immunogenicity assessment affects many areas

• Preclinical Development• Clinical Development• Regulatory Affairs• Medical Affairs• Sales & Marketing

Multi-tier Strategy

ACP: assay cut pointDCP: depletion cut pointAb: antibodyNAb: neutralizing antibody

Page 6: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Statistical Applications for Immunogenicity Assay Validations

Assay cut point

Precision and accuracy

Non-linear curve fitting

Assay sensitivity

Assay acceptance criteria

Reagent equivalence test

Proficiency test

Assay transfer6

Page 7: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

CAST for Immunogenicity Assays Analysis

• Relieve resource constraint, reduce data turn-around time, and minimize subjectivity in data analysis• Data analysis not trivial• Repetitive, but has multiple steps• Versatile experimental designs• Current commercial software lack customization and require

statistical decision making

• Standardized and automated data analysis and reporting

• Achieve greater compliance with a qualified, fit-for-purpose system

7

Page 8: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

8

Statistical Background on ACP Estimation• ACP is a one-sided 95% prediction limit

• Base on assay validation experiment

• Ensure normality: outliers removal and data transformation

• Mixed effect modeling• Full model -> reduced model• Variance Component Analysis

• Calculation of prediction limit• Total variance• Degrees of freedom approximation

Page 9: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Standard Experimental Design and Model Library

9

• Donor samples randomized into six sets (S1, S2, …, S6)• Model:

Analyst + Date within Analyst + Plate within Analyst and Date + Donor

e.g., Bioassay Design 2

• Every design is matched with a (full) model• Consider all factors random effects

• Current system allows seven models• It is easy to add more designs

Drug Assay Run Analyst Date Plate(s)Random

SequencesAMGXXX 1 1 1 2

1 2 21 3 ≥2

2 2 4 22 5 22 6 ≥2

S1,S2, S3,S4, S5,S6, …...

S1,S2, S3,S4, S5,S6, …...

xxx

Page 10: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

10

Analysis Work Flow in CAST

Residual NormalityOutlier &

Transformation

Mixed Effect Modeling

Backward Selection,VCA

Reduced Model

Donor NormalityOutliers removal if

neededPredictive Interval

Estimation

p<0.05

p<0.05

only once

Data

Model terms

Design%outliers > 5

Page 11: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Challenges

11

• Prove that automation is feasible

• Maintain quality and ensured proper use

• Meet compliance requirements

• Changes in experimentation

Page 12: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

12

Page 13: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

13

Page 14: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Maintain Quality and Ensure Proper Use

• Extensive comparison of CAST-ACP and manual result

• System defers to Biostatistician in certain conditions

• Standardization to reduce potential error• Data templates• Standard designs

• Statistician review of critical results

• User manual, SOP, training, e-learning and certification

14

Page 15: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

. 15

Automation vs. Manual Analysis (Immunoassay)

Assays Statistical Methods TransformationPower

U95 bound #Outlier

Manual 1 1.10 0

CAST -1 1.11 0

Manual 1 1.11 2

CAST 1 1.20 3

Manual 1 1.10 4

CAST -2 1.13 7

Assays Statistical Methods TransformationPower

U95 bound #Outlier

Manual 0.5 58.80 0

CAST 0.5 58.80 0

Manual Poor data distribution

CAST

1

2

3

MSD

Could not find transformation, send data to biostat

Biacore

4

5

Page 16: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

16

Automation vs. Manual Analysis (Bioassay)

Assays Statistical Methods TransformationPower

U95 bound #Outlier

Manual 1 2.68 1

CAST -2 2.76 1

Manual 1 1.88 2

CAST 2 1.64 1

6

7

Bioassay

Page 17: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Fit-for-purpose Qualification and Compliance

17

• Good Software Engineering and Good Documentation Practices, Change Control, Qualification Test

• Trainings, certification and statistician signed off

Project Charter• Objectives• Scope• Assumptions• Risk

Change Record• Qualification

strategy• Test summary

Installation Verification - IV(TEST/PROD)

• Install and verify CAST per configurations

Qualification Test – QT

(TEST)• Requirement• Test against

requirements• Traceability to

requirements

Documents Created

Required Training and CertificationUser manual, instructor-led course, e-learning

Page 18: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

CAST ACP Web User Interfaces- InputInterface • 2 groups

• Immunoassay• Bioassay

• 3 types of cut points• Screening, Specificity,

Scr/Spec

• 2 input Data Format• Excel• Files from instrument

• 7 experimental designs

• Upper/lower bound

• 0 statistical decision

18

Page 19: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

CAST ACP Web User Interfaces - Result

19

• Analyzed by SAS on a server

• Click to get PDF outputs with unique id for traceability

• Outputs also sent via email to statistician for QC

Page 20: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

CAST ACP Summary Tables - Examples

20

<Study/analysis specific Titles >

Table 3 is the listing (not shown)

Page 21: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

CAST ACP Graphs - Examples

21

Log-Transf.

Page 22: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

22

CAST Development Process

SoftwareSAS, RG

Standardization• Exp designs• Data input/out put

Test plan• Acceptance Criteria• Requirements

Algorithms• Development• testing

Testing cases

TeamCI, MSB, BI

Release

Feasibility

Prototype Development

Fit-for-purposeQualification

Final Development

User Testing• bug report• feedbacks

Final Details• System• QC

TrainingMaintenance

Page 23: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Standard Process

23

User

data

result EDM

Signed offreport

result

BiostatisticianCAST-ACP <1 hour Vs.

6-8 hours

Benefit• Used widely within Amgen • Shorter turnaround and less time spent• Encourages broader use of proper statistical analysis• Improve compliance

Page 24: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

Summary

• CAST ACP standardizes data entry, study design, automates assay cut point estimation, produces standard reports and facilitates review process

• This approach increases efficiency, maintains quality and improves compliance

• It also encourages proper study design and broader use of statistical analysis on assay cut point estimation

• Similar approach is being applied to other applications

24

Page 25: Development of Fit-for-Purpose Customized Assay ... of Fit-for-Purpose Customized Assay Statistical Tools (CAST) for Immunogenicity Assays Cheng Su 1, Lei Zhou , Ning Weng , Yu Wu1,

25