Upload
vanlien
View
226
Download
1
Embed Size (px)
Citation preview
Dr. Alexander MakarovThermo Fisher Scientific
September 1, 2009
18th IMSC Conference,
Bremen
Frontiers of Orbitrap Mass Spectrometry
Irkutsk
2
Zoology of ion traps
Ion trapsIon trapsThe art of confined motion
Down-side: imprisonment of innocent ionsUp-side: comparing to other MS, ions enter a world where they could live an eventful and meaningful lifeIon traps with
damping“Fishbowl”
Ion traps without damping
“Outer space”
(FAIMS traps)
• Paul traps• Linear radial• Linear axial
Magneto-static
RF Electrostatic
• FT ICR• Accelerators• Storage rings
• Multi-reflection• Multi-deflection• Orbital
• FT trap
Mass analysis:• excitation/ejection• image current• time-of-flight
Mass analysis:• excitation & ejejection
RF traps
Depends on free oscillations of narrow packets
3
Electrostatic traps: multi-reflection
Examples of Closed-type Examples of Open-type
Melzner F. US Patent 3,226,543; 1965.
Benner W.H. Anal.Chem.1997, 69, 4152-4168.
Behrisch R.; Blauth E.W.; Melzner F.; Meyer E.H. US Patent 3,174,034, 1965.
Wollnik H.; Przewloka A. Int. J. Mass Spectrom.1990, 96, 267-274.
Wollnik H., GB2,080,021, 1982
Verentchikov A.N.; Yavor M.I.; et al. Tech. Phys. 2005, 50, 73-81 and 82-86.
- image current detection- detection by excitation/ejection- time-of-flight detection
4
Electrostatic traps: multi-deflection
Examples of Closed-type Examples of Open-type
Toyoda M., Ishihara M. US7,148,473
Moller S.P. NIM A, 1997, 394, 281-286.
Matsuo T.; Toyoda M.; Sakurai T.; Ishihara M. I J. Mass Spectrom. 1997, 32, 1179-1185.
Satoh T. et al. J. Am. Soc. Mass Spectrom.2005, 16, 1969-1975.
Sakurai T.; Ito H.; Matsuo T. Anal. Chem.1994, 66, 2313-2317.
- image current detection- detection by excitation/ejection- time-of-flight detection
Bakker J.M.B. In: Advances in Mass Spectrometry. Vol. 5. 1971, 278-280.
5
Multi-reflection + Multi-deflection = Orbital
Examples of Closed-type Examples of Open-type
Makarov A. Anal. Chem. 2000, 72, 1156-1162.
Knight, R.D. Appl. Phys. Lett. 1981, 38,221-222.
Yang L.; Church D.A.; Tu S.; Jin J. Phys. Rev. A 1994, 50, 177-186.
Kingdon, K.H. A Phys. Rev. 1923, 21, 408-418.
Alimpiev S.S.; Makarov A.A. et al. SU 1,716,922, 1991.
- image current detection- detection by excitation/ejection- time-of-flight detection
6
Common features of electrostatic traps
Period of repetitive motion is T ~ (m/z)1/2
m/z-independent potential well In principle, unlimited mass rangeVery long mean free path must be provided:
• Loss of momentum widens the beam and deteriorates peak shape• Collisions take place at high energies (keVs) and could lead to prompt
or metastable ion fragmentation, thus reducing S/N Space-charge effects:
• Global mass shifts independent of m/z;• Diffusion of packet size • Synchronization of packets of different m/z (coalescence, self-
bunching)Short ion packets must be formed along the direction of m/zdispersionAccuracies of electrode shape, voltage and positioning are crucial for resolving power In closed traps, time-dependent fields are needed to introduce ions
++
++
Ø0.0001
7
What is so special about Orbitrap analyzer?
Image current detection naturally yields frequencies- an “easy” way towards high resolving power and mass accuracy along with all-mass detection and high sensitivityFor a given resolving power, Orbitrap field provides the smallest dimensions and the highest possible frequencies out of all types of electrostatic traps (harmonic oscillations)As a consequence of this, additional properties appear:
• high space charge capacity• robust and maintenance-free mass
analyzerFor other types of detection, other geometries might be more appropriate…
{ })/ln(2/2
),( 222mm RrRrzkzrU ⋅+−⋅=
z
f
r
8
Injection into the Orbitrap analyzer and Formation of Ion Rings
(r,φ) (r,z)
A short ion packet of one m/z enters the field Increasing voltage squeezes ions“Excitation by injection” is initiatedVoltage stabilises and ion trajectories are also stabilizedAngular spreading forms a ROTATING RING
9
Detection of Ions in the Orbitrap
Image current detection
I(t)
tI(t)
•All-mass detection (Felgett advantage)•Noise equivalent to <10 ē/√Hz
Frequency of axial oscillations of each ring induces an image current on split outer electrodes
10
zmk/
=ω
Detection of ion rings
Multiple ions in the Orbitrap analyzer generate a complex signal whose frequencies are determined using a fast Fourier Transformation
• Lighter ions enter Orbitrap analyzer earlier therefore they are squeezed closer to the central electrode than heavier ions
11
Experiments with individual ions (myoglobin +20)
Inte
nsity
848.3 848.4 848.5 848.6 848.7 848.8 848.9m/z
0
2000000
4000000
6000000
8000000
10000000
12000000
14000000848.70261.7 ppmR=50300
848.4498-4.5 ppm
R=38900
848.6003-2 ppm
R=46500848.7964-5.7 ppmR=22100
848.4 848.6 848.8 849.0 849.2m/z
R=46100
848.9516< 1 ppm
R=44800
0
2000000
4000000
6000000
8000000
848.70251 ppm
848.6515<1 ppm
R=45200
Inte
nsity
(1) (1) (1)
(f)
(1) (2)(1)
Velos
12
Overview of results for individual ions of +20 myoglobin
0
1
2
3
4
5
6
7
8
9
10
11
12
-5 0 5 10
Mass error, ppm
S/N
0
1
2
3
4
5
6
7
8
9
10
11
12
-5 0 5 10
Mass error, ppmS/
N
R=60,000 (0.76 sec) R=100,000 (1.52 sec)
Velos
Conclusion: Orbitrap could detect down to <10 elementary charges with mass accuracy still <3 ppm RMS
~5σ
2×
13
Importance of high transmission
0
1
2
3
4
5
1 10 100 1.000 10.000
RMS Mass accuracy, ppm
INJECTED ions/peak in 1 sec
High-resolution oaTOFOrbitrap
0
1
2
3
4
5
1 10 100 1.000 10.000
RMS Mass accuracy, ppm
DETECTED ions/peak in 1 sec
High-resolution oaTOF
Orbitrap
Assumptions:• oaTOF: resolving power 40,000, transmission 4% (grids x duty cycle x angular scattering on grids)• Orbitrap transmission 50%
14
0.8 sec
8 M record length10 Ms/s (borrowed LeCroy)0.8 s transientf =711 kHz,Δf=2.39 Hzf/Δf≈ 300000M/ΔM=½ f/Δf ≈ 150000
A.A. Makarov, Anal. Chem., v.72 (2000), No.6, p.1156-1162.
First steps: Orbitrap with Laser Ion Source
Field compensator Ion source
High voltageamplifier
15
1.E+00 1.E+01 1.E+02 1.E+03 1.E+04 1.E+05 1.E+06
Typical number of ions per injection 1.E-09 1.E-08 1.E-07 1.E-06 1.E-05 1.E-04 1.E-03 1.E-02
Typical width of ion pulse, s
For discontinuous analyzers, continuous ion beams require external storage and pulsed extraction
Orbitrap “fast inj.”Orbitrap“slow Inj.”TOFFT ICRRF ion trap
1.E-01 1.E+00 1.E+01 1.E+02
Typical entrance area, sq.mmTypical no. of ions per injection Typical width of ion pulse, sec Typical entrance area, mm2
16
Orbitrap mass spectrometry: state of the art
Sample in solution
Ion Source
Ion Source … C-trapC-trap Orbitrap
analyzerOrbitrapanalyzer
How does it work?
17
Curved Linear Trap (C-trap) as an Injection Device
CE
OE-1
OE-2
C-trap
Lenses
Deflector
Quasi-continuous ion beam enters a gas-filled RF-only quadrupole (C-trap) Ions collide with bath gas, loose energy and get stored in the C-trapRF is ramped down, radial DC is applied across rodsIons are ejected along lines converging on the orbitrap entrance. Ions enter Orbitrap analyzer as a small packet...
C-trap could be used also for: de-clustering/ removal of chemical noise passing ions through (T-piece)combining multiple populations, etc.
18
Biodiversity of Orbitrap mass spectrometry: status at 2009
+ HCD (XL)(QqQ-type fragmentation)
MALDI Source(high-power paradigm)
ETD(high-intensity beams)
+++ --
-
LTQ Orbitrap familyLTQ Orbitrap family
ExactiveExactive• HCD option
LC
Velos
FAIMS(new dimension of filtering)
19
FAIMS: separating isomers
Front-end Technology Extension: FAIMS
20
New Instrument Type: ESI-Orbitrap Exactive™
First serial bench-top FTMSHighest speed FTMS (10 spectra/ second) for fast LCNo precursor mass selection High resolution, mass accuracy and sensitivity allow not only fast screening and qualitative analysis, but also quantitative analysisIon population control using a pre-scanFast polarity switching without the loss of external calibrationAccurate mass is provided for all peaks in a single shot, from very low to very high S/N (>10,000)Additional dimension of analysis: all-ions fragmentation in an HCD cell
Exactive
21
hepc_96_std_2_5_ppb02 #345-350 RT: 4.29-4.34 AV: 3 SB: 48 3.61-3.96 , 4.72-5.51 NL: 4.90E4T: FTMS {0,0} + p ESI Full ms [600.00-1000.00]
697.5 698.0 698.5 699.0 699.5 700.0 700.5 701.0 701.5 702.0 702.5m/z
0
10
20
30
40
50
60
70
80
90
100
Rel
ativ
e A
bund
ance
700.
7747
700.
5249
700.
2743
701.
0246
701.
2750
701.
5245
701.
7757
698.
0179
700.
0232
698.
2677
697.
7677
698.
5178
702.
0258
701.
3460
699.
0185
698.
7676
702.
2986
699.
7824
699.
3641
698.
6295
699.
5554
Blow-up of the
internal standard
and
analyte ions (2.5ppb)
Example: Quantitation of Hepcidin using Exactive
2.5 ppb spiked and extracted plasma sample 100k resolution
H. Li1, M. J. Rose1, B. J. Hart2, S. Sharma2, C. A. James1
1Amgen Inc, Thousand Oaks , CA; 2Thermo Scientific, San Jose, CAExactive
22
Exactive HR-AM_MS QC Plate: Quantitation
RT: 0.00000 - 9.52012 SM: 7G
0 2 4 6 8Time (min)
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
RT: 4.34193
RT: 4.34193
RT: 6.32679
NL: 5.53E3m/z= 698.01436-698.02134+698.26425-698.27123 F: FTMS {0,0} + p ESI Full ms [600.00-1000.00] MS ICIS hepc_96_std_2_5_ppb02
NL: 1.01E5m/z= 700.52143-700.52843+700.77122-700.77822 F: FTMS {0,0} + p ESI Full ms [600.00-1000.00] MS ICIS hepc_96_std_2_5_ppb02
InternalStandard
Analyte2.5ppb
Exactive
23
Exactive HR-AM_ HCD MS QC Plate: Confirmation by HCD
RT: 0.00000 - 9.01554 SM: 7G
0 1 2 3 4 5 6 7 8 9Time (min)
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
0
20
40
60
80
100
Rel
ativ
e A
bund
ance
RT: 4.06004
RT: 4.06004
RT: 6.67110
NL: 6.08E2m/z= 501.19049-501.20051 F: FTMS {0,0} + p ESI Full ms2 [email protected] [330.00-1330.00] MS ICIS qc_plate_2_5_ppb51
NL: 4.12E4m/z= 511.22024-511.23046 F: FTMS {0,0} + p ESI Full ms2 [email protected] [330.00-1330.00] MS ICIS qc_plate_2_5_ppb51
InternalStandard
Analyte2.5ppb
Exactive
24
Quantitation using precursor scan: 2.5 ppb to 500ppb
Sample in spiked serum Calibration curves in triplicate 2.5 to 500 ng/mlQCs in triplicate at 4 levels (5, 15, 100 and 400 ng/ml)2/3’s of QCs must be within 25% of nominal at each level
Exactive
25
All-ion fragmentation for structure elucidation
Full scan spectra of Verapamil with and without HCD fragmentation.
Identification of precursors and fragments is facilitated by very high resolution and mass accuracy
Exactive
1.3 ppmO
O
N
O
O
1.4 ppm
1.2 ppm
N
N
O
O
2.1 ppmMH+
150 200 250 300 350 400 450m/z
0
20
40
60
80
1000
20
40
60
80
100 455.2890
167.0126
239.1614 311.2189
165.0912
303.2071
260.1642 455.2895
1.3 ppmO
O
N
O
O
1.4 ppm
1.2 ppm
N
N
O
O
2.1 ppmMH+
HCD off
HCD on
150 200 250 300 350 400 450m/z
0
20
40
60
80
1000
20
40
60
80
100 455.2890
167.0126
239.1614 311.2189
165.0912
303.2071
260.1642 455.2895
Quantitation
Confirmation
Replaces MRMs
26
New hybrid for proteomics: LTQ Orbitrap Velos
Main emphasis: • higher speed of MS/MS not on paper (e.g. for calibration mixtures)
but for real-life samples • higher quality and choice of MS/MS methods
Tuesday 3:00 – 3:20Eduard Denisov,, E. Damoc, J. Griep-Raming, O. Lange, A. Makarov, H. Kuipers, P. Remes, J. Schwartz, D. Taylor, T. Moehring and V. Zabrouskov:"A New Instrument for High-speed Proteomics: Orbitrap Mass Analyzer Interfaced to a Dual Linear Trap"
Velos
27
What to expect from Orbitrap mass spectrometry?
Faster and more powerful Orbitraps• High-field Orbitraps (dimensions, voltage)• Higher-performance (resolving power, mass accuracy) Orbitraps
Higher space charge capacity C-trap
“Intelligence and speed are better than haste”: multiple fills allow to improve quality of analysis
• Stepped collision energy in 1 scan• Multiple SIM windows in 1 scan • Multiple MS/MS in 1 scan• ....
28
Example 1: High-field Orbitrap
Smaller gap and higher voltage- 1.8x frequencySpace charge shifts: <1 ppm over the whole dynamic rangeMost “best” parameters could be demonstrated, but tuning ranges much narrower; more research is needed.
12 m
m
18 m
m
Research only!
30 m
m
526.26 526.28m/z
0
1
2
3
4
0
1
2
3
4
Rel
ativ
e A
bund
ance 526.2614
R=377733
526.2722R=375264
526.2756R=354997
526.2608R=378620
526.2717R=375454
526.2747R=359061
196.06 196.08 196.1002
4
6
8
10
12
02
4
68
10
12196.0913R=630118
196.0850R=598468
196.0910R=629410
196.0847R=629316R
elat
ive
Abu
ndan
ce
Experimental
Theoretical
RFWHM ≈630,000 RFWHM ≈370,000
A. Makarov et al. JASMS, 2009
A. Makarov, E. Denisov, O. Lange J. Am. Soc. Mass Spectrom. 2009, 20, 1391–1396.
29
Example 2: Overtone Orbitrap
Each of outer electrodes is split in two, all outer electrodes sustained at virtual ground
Research only!
30 m
m
30
SIM from overtone Orbitrap
100 200 300 400 500 600 700 800 900 1000 1100 1200 1300 1400
m/z
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
75
Rel
ativ
e In
tens
ity
157.9970
3rd harmonics: 41%R≈107,400
1422.9772
56.8789
355.4934
2nd harmonics: 1%R≈71,100
5th harmonics: 4%R≈177,300
1st harmonics: 54%R≈35,500
•Use of 3rd harmonics allows to drastically increase resolving power in short acquisitions-though at the expense of sensitivity (↓ x2.5)•Extensive filtering is required to remove undesired harmonics
31
Example 3: Compact Orbitrap
Smaller size- 1.5x frequency at the same voltage, 1.8x at higher voltage Higher tolerance requirementsNew lenses for focusing onto Orbitrap entranceLower capacitance
12 m
m
20 m
m
Research only!
30 m
m
2
3
4526,2600R=427119
526,2703R=429586
0
1
526.26 526.28m/z
0
1
2
3
4
R=378620
526.2717R=375454
526.2747R=359061
196.06 196.08 196.1002
4
6
8
10
12196.0910R=629410
196.0847R=629316R
elat
ive
Abu
ndan
ce Theoretical
Experimental
196,06 196,08 196,100
2
4
6
8
10
12
Rel
ativ
e A
bund
ance
196,0910R=667444
196,0848R=677597
32
0
1
2
3
4
5
6
7
8
9
10
‐6 ‐4 ‐2 0 2 4 6
S/N
Mass error, ppm
Lower capacitance improves sensitivity: detection of individual ions of +10 Ubiquittin
~5σ
R=120,000 (0.76 sec)
S/N≈6
S/N≈3
Conclusion: Combination of compact Orbitrap analyzer with a customized preamplifier allows increasing sensitivity by a factor of x1.7
33
Multiple fills: Stepped collision energy in 1 scan
500 1000 1500m/z
0
50
1000
50
1000
50
1000
50
1001421.9792
1189.9635977.9568653.9355247.9324
321.9290 977.9571 1421.9774765.9498
247.9324
321.9291 725.9576 937.9651 1318.06881421.9767
247.9322
1189.9630977.9565321.9289
a) CE=99 eV (NCE=35)
b) CE=121 eV (NCE=45)
c) CE=148 eV (NCE=55)
d) CE=99…148 eV (NCE=35…55)
Perfluoroalkylphosphazine, C28H19O6N3P3F44, m/z 1421.97842
Σ3 different MS/MS
spectra in 1 Orbitrap scan
34
MPX(432.9_3_530.8_2_379.2_2)_100k_01 NL: 3.30E5T: FTMS + p ESI Full ms [100.00-1980.00]
200 300 400 500 600 700 800 900 1000m/z0
10
20
30
40
50
60
70
80
90
100
Rel
ativ
e A
bund
ance
Multiple fills: multiple MS/MS
Bradikinin 1-7Angiotensin I (human)Bradikinin
y2
y2y6
y8y7b8
b5,b5
MH2+
b1,b1
y1
b62+,b62+
a8
a6a5
b6 b8
b6
a6
b4,b4b72+
Data acquired by Dr. M.Kellmann
35
Conclusion
Orbitrap mass analyzer offers a unique and very valuable combination of analytical parameters
New analytical methods are made possible by Orbitrap technology
Orbitraps are still evolving …..• Higher speed• Higher resolving power and mass
accuracy• Higher sensitivity• Mass production
Exciting new applications continue to emerge
36
Acknowledgements
My Family:Anna, Dasha, Nikita MakarovMy parents: Alla & Alexei Makarov
Manchester, UK:S. DavisA. HoffmannR. LawtherM. HardmanN. DemetriadesJ. HughesB. McKnightK. WorthingtonS. Smirnov
Bremen, DE:E. DenisovA. KholomeevW. BalschunO. LangeK. StrupatS. HorningR. PeschJ. SregaG. JungW. Huels F. CzemperO. HengelbrockA. Wieghaus
San Jose, CA:E. HemenwayM. Antonczak M. SenkoJ. SykaJ. SchwartzV. ZabrouskovT. SecondI. JardineT. ZibernaJ-J. DunyachE. WoutersM. SplendoreP. RemesD. TaylorB. TehlirianV. KovtounN. Izgarian
Development teams:LTQLTQ FTMALDIETDExactive
J. Griep-RamingE. SchroederU. FroehlichD. NoltingR. MalekT. MoehringH. MuensterM. KellmannM. ZellerS. StrubeS. MoehringR. SeedorfS. Kanngiesser
Colleagues:P. FongJ. GabelS. FenskeP. CardenasM. KonicekR. HermezianW. DeweyB. SiebertP. AthertonM. AhrensJ. SklenarS. ZanonW. WangJ. PhillipsJ. HornerJ. BlethrowH. Bui
37
Thank you !