Upload
ghnah
View
222
Download
0
Embed Size (px)
Citation preview
8/3/2019 Tonometry.4
1/50
Click to edit Master subtitle style
4/28/12
Tonometry
Chapter 4Lecturer : Dr. Genalin Ang O.D.
8/3/2019 Tonometry.4
2/50
4/28/12
This is a clinical technique that provides ameasurement of the Internal pressure of theeye Called IOP ( Intra ocular pressure, ocular
Tension )
What is tonometer?
8/3/2019 Tonometry.4
3/50
4/28/12
High intraocular pressure causes damage tothe optic nerve, which can lead to glaucoma.But what is it? What makes it happen? Whatcan you do about it?
First, let's dissect the term.Intra is the Latin word for within or inside.
Ocular refers to the eye.Pressureis the result of applying a force
onto a surface.
What is IOP ( IntraocularPressure)
8/3/2019 Tonometry.4
4/50
4/28/12
Small reductions or increases in ballpressure would mean that the
bounce potential would change andinfluence three key dimensions ofthe game: height of return after theball strikes the floor, bounce off the
backboard, and bounce off the rim ofthe basket.
8/3/2019 Tonometry.4
5/50
4/28/12
HOW DOES THIS APPLY TO THEEYE?
The aqueous humor is confined to asmall space in the front part of theeye. The remainder of the eye isfilled with vitreous humor.
8/3/2019 Tonometry.4
6/50
4/28/12
When we talk about intraocular pressure, weare referring to the pressure exerted bythese two fluids on the walls of the eye andon the structures inside.
Similar to the basketball, the materials of theeye that contain the aqueous and vitreoushumors in a closed space have limitedflexibility and expansion capabilities. Thismeans that additional aqueous humorintroduced into the eye increases thepressure inside the eye.
8/3/2019 Tonometry.4
7/504/28/12
There is three different principles :
APPLANATIONIDENTATIONMANOMETRY
Theoretical Principle ofTonometery
8/3/2019 Tonometry.4
8/504/28/12
Is the most commonly used technique tomeasure IOP
IOP= force/ areaMeaning in physics + force applied as well as
the size of the area of the eye on which thisforce applied
Applanation
8/3/2019 Tonometry.4
9/504/28/12
GOLDMANN
PERKINS
NON-CONTACT ( NCT ) TONOMETERS
APPLANATION
8/3/2019 Tonometry.4
10/50
4/28/12
Goldmann Tonometer
8/3/2019 Tonometry.4
11/50
4/28/12
Goldmann tonometerthe "gold standard" instrument attached to
the slit lamp biomicroscope used in all eyedoctors' offices
It requires a cobalt blue light source and asmall droplet of fluorescein on the ocularsurface.
Applanation
8/3/2019 Tonometry.4
12/50
4/28/12
A tiny pressure sensor attached to a spring-loaded arm is gently placed against the tearfilm, and the doctor or technician reads thepressure through the microscope under the
blue light.
Applantion
8/3/2019 Tonometry.4
13/50
4/28/12
Maklokov TonometerIs a method of applanation that applies a
constant force to the cornea
8/3/2019 Tonometry.4
14/50
4/28/12
Schiotz tonometer
Indentation
8/3/2019 Tonometry.4
15/50
4/28/12
DisadvatagesMore invasiveIt is affected corneal rigidity
Repeated measurement may be misleading due to aqueous humour being evacuated ateach reading
Schiotz tonometer
8/3/2019 Tonometry.4
16/50
4/28/12
Is the most direct and accurate method tomeasure IOP
Its not use due to invasive nature of theprocedure
Manometry
8/3/2019 Tonometry.4
17/50
4/28/12
Goldmann Tonometer
Common Techniques
8/3/2019 Tonometry.4
18/50
4/28/12
Goldmann applanation tonometry isconsidered the Gold Standard based onits accuracy and repeatability ofresults. A hand-held version called the
Perkins tonometer
Goldmann Tonometer
8/3/2019 Tonometry.4
19/50
4/28/12
Excessive pressure reading on the eye distortsthe mires Turning the reading drum
Figure 4.4 a) Move Slit lamp up b) Move
slit lamp rightc) Move slit lamp up and left
I ffi i h
8/3/2019 Tonometry.4
20/50
4/28/12
Insufficient presurre on the eyeon indenting the probe
8/3/2019 Tonometry.4
21/50
4/28/12
Advantage you can use in different position
Perkins Tonometer
8/3/2019 Tonometry.4
22/50
4/28/12
Using Perkins tonometer
8/3/2019 Tonometry.4
23/50
4/28/12
Uses both Applanation and Identationpriciple
Portable, penshape hand held deviceshandheld device and calibrates digitally with
the push of a button. It requires a disposablesterile cover for each patient. The steriledevice tip is gently placed against the tearfilm by the doctor or technician, and thepressure reading appears on
Tonopen
8/3/2019 Tonometry.4
24/50
4/28/12
Tonopen
8/3/2019 Tonometry.4
25/50
4/28/12
Advisable in Children , un cooperativepatients or patient unable to be positionedbehind the lamp or long enough to performPerkins or NCT
Contraindicated in patients with knownallergies to latex.
Tips of Tonopen
8/3/2019 Tonometry.4
26/50
4/28/12
Which generally requires no anesthetic drop,is widely used in doctor's offices, clinics, andscreening facilities. It is very safe due to the"no touch" technology,
Non Contact Tonometry
8/3/2019 Tonometry.4
27/50
4/28/12
Non Contact tonometer
G ll NCT i
8/3/2019 Tonometry.4
28/50
4/28/12
A video monitoring system to view the imageof the eye
- Some sort of visible mires in the viewer(reflections off the cornea)
Mires must be focused and aligned- A button on the joystick is pressed (or an
automatic mechanism is triggered) and theinstrument shoots its puff of
air- The reading is visible in the viewer (can
also be printed on some models)
Table-
Generally, most NCT instrumentsuse the following elements andsteps:
8/3/2019 Tonometry.4
29/50
4/28/12
Different types of NCT
8/3/2019 Tonometry.4
30/50
4/28/12
When applicable, the chin and headrest ofthe instrument must be disinfected
Demonstration the small air puff onto thepatients fingers
By convention, the right eye is usually testedfirst
If applicable, a safety lock usually allowsthe instrument to stop at a safe distancefrom the eye
Tips fro NCT
8/3/2019 Tonometry.4
31/50
4/28/12
3 readings or more are taken and averaged(to account for ocular pulse)
NCT readings are often imprecise and arebest used as screening tools or if corneal
contact is not possible.Goldmann or Perkins tonometry should be
performed whenever possible, and each timean NCT measurement is
abnormal or suspicious
8/3/2019 Tonometry.4
32/50
4/28/12
Schiotz tonometer
8/3/2019 Tonometry.4
33/50
4/28/12
The patient is placed in a supine position forthe procedure. A constant known weight isapplied to the cornea and the depth of thearea
depressed is measured and converted to IOPby using a graph.
Readings are placed on a calibration scalefrom which the IOP is determined
The measurement is then corrected for errorinduced by corneal rigidity on a graph.
8/3/2019 Tonometry.4
34/50
4/28/12
For high IOP measurements, it is necessaryto use more weight to perform Schiotzbecause there is more resistance toindentation. The additional weight is needed
to indent the cornea and obtain an accuratereading. The
disadvantages of the Schiotz include: highrisk of corneal abrasion, supine position of
patient, influence of technique onresults, assembly and disassembly of
instrument, aqueous displacement affectingrepeat readings, and patient
apprehension to procedure. Since there are
Tips
8/3/2019 Tonometry.4
35/50
4/28/12
Finger Tension ( DigitalIOP )
8/3/2019 Tonometry.4
36/50
4/28/12
Finger tension IOP is a crude method togrossly assess IOP used in situations whereno other more precise method is
available or possible (e.g. Non-cooperative
patients). One can only approximate whetherthe eye is soft, normal or hard
and compare both eyes for a notabledifference.
Finger tension IOP
8/3/2019 Tonometry.4
37/50
4/28/12
The IOP value for each eyeType of anesthetic and instrument usedThe time that tonometry was performed. Repeated measurements and their time (if
performed) Patient position (if pertinent)
Recording
8/3/2019 Tonometry.4
38/50
4/28/12
Normals IOP
The mean IOP in the normal population is16mmHg +/- 2.5mmHg
Interpretation of Results
8/3/2019 Tonometry.4
39/50
4/28/12
Physical factorsPhysiological factors
FACTORS INFLUENCING IOP
8/3/2019 Tonometry.4
40/50
4/28/12
Pressure on the globe
Applying pressure on the globe tends toelevate the IOP. This is important becauseduring the procedure it may becomenecessary to control the patients lids
Physical factors
8/3/2019 Tonometry.4
41/50
4/28/12
Patient blinking also results in an increase inIOP. Forceful blinking or blepharospasm cansignificantly elevate the
IOP due to lid muscle compression on the
globe. It is therefore necessary to instructpatients to blink gently to
avoid an artificially elevated measurement.
i fl i
8/3/2019 Tonometry.4
42/50
4/28/12
Trauma or inflammation of the eye can eitherdecrease or increase the IOP.
If trauma affects the ciliary body, aqueousproduction may be reduced which woulddecrease the IOP while if the trauma affectsthe anterior chamber angle structures,aqueous outflow may be reduced which
would increase the IOP.
Trauma / inflammation
8/3/2019 Tonometry.4
43/50
4/28/12
If a patient has an intraocular inflammationsuch as uveitis, the ciliary body may becomeinflamed and decrease aqueous productiontherefore decreasing IOP.
8/3/2019 Tonometry.4
44/50
4/28/12
Certain medications may elevate or reduceIOP
Long-term use of corticosteroids (especiallytopical and oral) has proven to induce
increased IOP. Certain blood pressure medications such
as Beta-blockers when taken orally canreduce IOP.
The use of substances such as marijuanaand alcohol temporarily reduce the IOP.
Medication
8/3/2019 Tonometry.4
45/50
4/28/12
Diurnal variation
Vascular Integrity
Physiological factors
8/3/2019 Tonometry.4
46/50
4/28/12
The IOP tends to vary throughout the daybeing generally highest during the morningand lowest in the evening.
The average diurnal variation is 4mmHg.
Persons with glaucoma tend to have agreater diurnal variation. Diurnal
variation exceeding 8mmHg is a risk factorfor POAG. To measure diurnal variation in IOP
several measurementsmay be taken at various times of the day in
order to make a definitive assessment.
Diurnal Variations
8/3/2019 Tonometry.4
47/50
4/28/12
Impaired venous drainage in the head-neckregion can result in a decrease of aqueousoutflow and subsequently
an increase in IOP up to 4-5mmHg. This can
be brought on by a tight collar or when thepatient holds their breath
during tonometry.
Vascular Integrity
8/3/2019 Tonometry.4
48/50
4/28/12
If there is poor circulation to the ciliary body,then aqueous production is reduced and IOPis decreased. This may occur in persons withcardiovascular conditions, including carotid
occlusive disease. When there is carotidartery
8/3/2019 Tonometry.4
49/50
4/28/12
When the patient is in a supine position or ifthe head is below the heart, IOP is increased.
The IOP is measured tobe 2-3mmHg higherlying down than when the patient is sitting
upright. The increase in IOP can beattributed decrease the aqueous outflow.
Patient Position Movement
8/3/2019 Tonometry.4
50/50
Thank You
The most pathetic person in the world issomeone who has sight, but has no vision. Helen Keller