Tonometry.4

  • Upload
    ghnah

  • View
    222

  • Download
    0

Embed Size (px)

Citation preview

  • 8/3/2019 Tonometry.4

    1/50

    Click to edit Master subtitle style

    4/28/12

    Tonometry

    Chapter 4Lecturer : Dr. Genalin Ang O.D.

  • 8/3/2019 Tonometry.4

    2/50

    4/28/12

    This is a clinical technique that provides ameasurement of the Internal pressure of theeye Called IOP ( Intra ocular pressure, ocular

    Tension )

    What is tonometer?

  • 8/3/2019 Tonometry.4

    3/50

    4/28/12

    High intraocular pressure causes damage tothe optic nerve, which can lead to glaucoma.But what is it? What makes it happen? Whatcan you do about it?

    First, let's dissect the term.Intra is the Latin word for within or inside.

    Ocular refers to the eye.Pressureis the result of applying a force

    onto a surface.

    What is IOP ( IntraocularPressure)

  • 8/3/2019 Tonometry.4

    4/50

    4/28/12

    Small reductions or increases in ballpressure would mean that the

    bounce potential would change andinfluence three key dimensions ofthe game: height of return after theball strikes the floor, bounce off the

    backboard, and bounce off the rim ofthe basket.

  • 8/3/2019 Tonometry.4

    5/50

    4/28/12

    HOW DOES THIS APPLY TO THEEYE?

    The aqueous humor is confined to asmall space in the front part of theeye. The remainder of the eye isfilled with vitreous humor.

  • 8/3/2019 Tonometry.4

    6/50

    4/28/12

    When we talk about intraocular pressure, weare referring to the pressure exerted bythese two fluids on the walls of the eye andon the structures inside.

    Similar to the basketball, the materials of theeye that contain the aqueous and vitreoushumors in a closed space have limitedflexibility and expansion capabilities. Thismeans that additional aqueous humorintroduced into the eye increases thepressure inside the eye.

  • 8/3/2019 Tonometry.4

    7/504/28/12

    There is three different principles :

    APPLANATIONIDENTATIONMANOMETRY

    Theoretical Principle ofTonometery

  • 8/3/2019 Tonometry.4

    8/504/28/12

    Is the most commonly used technique tomeasure IOP

    IOP= force/ areaMeaning in physics + force applied as well as

    the size of the area of the eye on which thisforce applied

    Applanation

  • 8/3/2019 Tonometry.4

    9/504/28/12

    GOLDMANN

    PERKINS

    NON-CONTACT ( NCT ) TONOMETERS

    APPLANATION

  • 8/3/2019 Tonometry.4

    10/50

    4/28/12

    Goldmann Tonometer

  • 8/3/2019 Tonometry.4

    11/50

    4/28/12

    Goldmann tonometerthe "gold standard" instrument attached to

    the slit lamp biomicroscope used in all eyedoctors' offices

    It requires a cobalt blue light source and asmall droplet of fluorescein on the ocularsurface.

    Applanation

  • 8/3/2019 Tonometry.4

    12/50

    4/28/12

    A tiny pressure sensor attached to a spring-loaded arm is gently placed against the tearfilm, and the doctor or technician reads thepressure through the microscope under the

    blue light.

    Applantion

  • 8/3/2019 Tonometry.4

    13/50

    4/28/12

    Maklokov TonometerIs a method of applanation that applies a

    constant force to the cornea

  • 8/3/2019 Tonometry.4

    14/50

    4/28/12

    Schiotz tonometer

    Indentation

  • 8/3/2019 Tonometry.4

    15/50

    4/28/12

    DisadvatagesMore invasiveIt is affected corneal rigidity

    Repeated measurement may be misleading due to aqueous humour being evacuated ateach reading

    Schiotz tonometer

  • 8/3/2019 Tonometry.4

    16/50

    4/28/12

    Is the most direct and accurate method tomeasure IOP

    Its not use due to invasive nature of theprocedure

    Manometry

  • 8/3/2019 Tonometry.4

    17/50

    4/28/12

    Goldmann Tonometer

    Common Techniques

  • 8/3/2019 Tonometry.4

    18/50

    4/28/12

    Goldmann applanation tonometry isconsidered the Gold Standard based onits accuracy and repeatability ofresults. A hand-held version called the

    Perkins tonometer

    Goldmann Tonometer

  • 8/3/2019 Tonometry.4

    19/50

    4/28/12

    Excessive pressure reading on the eye distortsthe mires Turning the reading drum

    Figure 4.4 a) Move Slit lamp up b) Move

    slit lamp rightc) Move slit lamp up and left

    I ffi i h

  • 8/3/2019 Tonometry.4

    20/50

    4/28/12

    Insufficient presurre on the eyeon indenting the probe

  • 8/3/2019 Tonometry.4

    21/50

    4/28/12

    Advantage you can use in different position

    Perkins Tonometer

  • 8/3/2019 Tonometry.4

    22/50

    4/28/12

    Using Perkins tonometer

  • 8/3/2019 Tonometry.4

    23/50

    4/28/12

    Uses both Applanation and Identationpriciple

    Portable, penshape hand held deviceshandheld device and calibrates digitally with

    the push of a button. It requires a disposablesterile cover for each patient. The steriledevice tip is gently placed against the tearfilm by the doctor or technician, and thepressure reading appears on

    Tonopen

  • 8/3/2019 Tonometry.4

    24/50

    4/28/12

    Tonopen

  • 8/3/2019 Tonometry.4

    25/50

    4/28/12

    Advisable in Children , un cooperativepatients or patient unable to be positionedbehind the lamp or long enough to performPerkins or NCT

    Contraindicated in patients with knownallergies to latex.

    Tips of Tonopen

  • 8/3/2019 Tonometry.4

    26/50

    4/28/12

    Which generally requires no anesthetic drop,is widely used in doctor's offices, clinics, andscreening facilities. It is very safe due to the"no touch" technology,

    Non Contact Tonometry

  • 8/3/2019 Tonometry.4

    27/50

    4/28/12

    Non Contact tonometer

    G ll NCT i

  • 8/3/2019 Tonometry.4

    28/50

    4/28/12

    A video monitoring system to view the imageof the eye

    - Some sort of visible mires in the viewer(reflections off the cornea)

    Mires must be focused and aligned- A button on the joystick is pressed (or an

    automatic mechanism is triggered) and theinstrument shoots its puff of

    air- The reading is visible in the viewer (can

    also be printed on some models)

    Table-

    Generally, most NCT instrumentsuse the following elements andsteps:

  • 8/3/2019 Tonometry.4

    29/50

    4/28/12

    Different types of NCT

  • 8/3/2019 Tonometry.4

    30/50

    4/28/12

    When applicable, the chin and headrest ofthe instrument must be disinfected

    Demonstration the small air puff onto thepatients fingers

    By convention, the right eye is usually testedfirst

    If applicable, a safety lock usually allowsthe instrument to stop at a safe distancefrom the eye

    Tips fro NCT

  • 8/3/2019 Tonometry.4

    31/50

    4/28/12

    3 readings or more are taken and averaged(to account for ocular pulse)

    NCT readings are often imprecise and arebest used as screening tools or if corneal

    contact is not possible.Goldmann or Perkins tonometry should be

    performed whenever possible, and each timean NCT measurement is

    abnormal or suspicious

  • 8/3/2019 Tonometry.4

    32/50

    4/28/12

    Schiotz tonometer

  • 8/3/2019 Tonometry.4

    33/50

    4/28/12

    The patient is placed in a supine position forthe procedure. A constant known weight isapplied to the cornea and the depth of thearea

    depressed is measured and converted to IOPby using a graph.

    Readings are placed on a calibration scalefrom which the IOP is determined

    The measurement is then corrected for errorinduced by corneal rigidity on a graph.

  • 8/3/2019 Tonometry.4

    34/50

    4/28/12

    For high IOP measurements, it is necessaryto use more weight to perform Schiotzbecause there is more resistance toindentation. The additional weight is needed

    to indent the cornea and obtain an accuratereading. The

    disadvantages of the Schiotz include: highrisk of corneal abrasion, supine position of

    patient, influence of technique onresults, assembly and disassembly of

    instrument, aqueous displacement affectingrepeat readings, and patient

    apprehension to procedure. Since there are

    Tips

  • 8/3/2019 Tonometry.4

    35/50

    4/28/12

    Finger Tension ( DigitalIOP )

  • 8/3/2019 Tonometry.4

    36/50

    4/28/12

    Finger tension IOP is a crude method togrossly assess IOP used in situations whereno other more precise method is

    available or possible (e.g. Non-cooperative

    patients). One can only approximate whetherthe eye is soft, normal or hard

    and compare both eyes for a notabledifference.

    Finger tension IOP

  • 8/3/2019 Tonometry.4

    37/50

    4/28/12

    The IOP value for each eyeType of anesthetic and instrument usedThe time that tonometry was performed. Repeated measurements and their time (if

    performed) Patient position (if pertinent)

    Recording

  • 8/3/2019 Tonometry.4

    38/50

    4/28/12

    Normals IOP

    The mean IOP in the normal population is16mmHg +/- 2.5mmHg

    Interpretation of Results

  • 8/3/2019 Tonometry.4

    39/50

    4/28/12

    Physical factorsPhysiological factors

    FACTORS INFLUENCING IOP

  • 8/3/2019 Tonometry.4

    40/50

    4/28/12

    Pressure on the globe

    Applying pressure on the globe tends toelevate the IOP. This is important becauseduring the procedure it may becomenecessary to control the patients lids

    Physical factors

  • 8/3/2019 Tonometry.4

    41/50

    4/28/12

    Patient blinking also results in an increase inIOP. Forceful blinking or blepharospasm cansignificantly elevate the

    IOP due to lid muscle compression on the

    globe. It is therefore necessary to instructpatients to blink gently to

    avoid an artificially elevated measurement.

    i fl i

  • 8/3/2019 Tonometry.4

    42/50

    4/28/12

    Trauma or inflammation of the eye can eitherdecrease or increase the IOP.

    If trauma affects the ciliary body, aqueousproduction may be reduced which woulddecrease the IOP while if the trauma affectsthe anterior chamber angle structures,aqueous outflow may be reduced which

    would increase the IOP.

    Trauma / inflammation

  • 8/3/2019 Tonometry.4

    43/50

    4/28/12

    If a patient has an intraocular inflammationsuch as uveitis, the ciliary body may becomeinflamed and decrease aqueous productiontherefore decreasing IOP.

  • 8/3/2019 Tonometry.4

    44/50

    4/28/12

    Certain medications may elevate or reduceIOP

    Long-term use of corticosteroids (especiallytopical and oral) has proven to induce

    increased IOP. Certain blood pressure medications such

    as Beta-blockers when taken orally canreduce IOP.

    The use of substances such as marijuanaand alcohol temporarily reduce the IOP.

    Medication

  • 8/3/2019 Tonometry.4

    45/50

    4/28/12

    Diurnal variation

    Vascular Integrity

    Physiological factors

  • 8/3/2019 Tonometry.4

    46/50

    4/28/12

    The IOP tends to vary throughout the daybeing generally highest during the morningand lowest in the evening.

    The average diurnal variation is 4mmHg.

    Persons with glaucoma tend to have agreater diurnal variation. Diurnal

    variation exceeding 8mmHg is a risk factorfor POAG. To measure diurnal variation in IOP

    several measurementsmay be taken at various times of the day in

    order to make a definitive assessment.

    Diurnal Variations

  • 8/3/2019 Tonometry.4

    47/50

    4/28/12

    Impaired venous drainage in the head-neckregion can result in a decrease of aqueousoutflow and subsequently

    an increase in IOP up to 4-5mmHg. This can

    be brought on by a tight collar or when thepatient holds their breath

    during tonometry.

    Vascular Integrity

  • 8/3/2019 Tonometry.4

    48/50

    4/28/12

    If there is poor circulation to the ciliary body,then aqueous production is reduced and IOPis decreased. This may occur in persons withcardiovascular conditions, including carotid

    occlusive disease. When there is carotidartery

  • 8/3/2019 Tonometry.4

    49/50

    4/28/12

    When the patient is in a supine position or ifthe head is below the heart, IOP is increased.

    The IOP is measured tobe 2-3mmHg higherlying down than when the patient is sitting

    upright. The increase in IOP can beattributed decrease the aqueous outflow.

    Patient Position Movement

  • 8/3/2019 Tonometry.4

    50/50

    Thank You

    The most pathetic person in the world issomeone who has sight, but has no vision. Helen Keller