8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 1/10
BA 011C/07/en/11.99No.50049295 DipFit W
CPA 510Immersion Assembly forpH/Redox Measurement
Hauser+EndressThe Power of Know How
Operating Instructions
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 2/10
Table of contents
1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.2 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
2 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.1 Scope of supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
2.2 Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
3 Mounting and installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.1 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
3.2 Electrode installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53.3 Mounting of the assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
3.4 Mounting with pendulum frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.1 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
4.2 Cleaning of the pH electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
5 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
5.1 Supplementary documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
TOC_BA.CHP
DipFit W CPA 510 Table of contents
1
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 3/10
1 General
1.1 Application
DipFit W CPA 510 is an assembly intended to
hold a combined pH or redox electrode filledwith a gel electrolyte. It is universally suitable
for use in water and waste water treatment
systems as well as in process applications.
DipFit W CPA 510 can be used either as animmersion assembly in open containers and
channels or as a built-in assembly in tanksand closed containers.
1.2 Measuring system
A measuring system consists of: the DipFit W CPA 510 assembly
• a pH / redox electrode, 120 mm in length,suitable for the assembly and the medium
to be measured• a pH / redox measuring instrument
• a CPK 1, CPK 7 or CPK 9 measuring cable(terminated).
Option:
• Junction box VBA and measuring cable
(not terminated) for measuring cable
extension.
P R M
MEASE S CAL L
DI G
AssemblyDipFit W CPA 510
Cable gland Pg 16
Option:Junction box VBA
Electrode cable withelectrode plug (e.g. CPK 7)
Measuring instrument(e.g. Mycom CPM 152)
➀➁
Electrode plug
Electrode
Electrode holder
Protective cage B A 0 1 1 Y 0 1 . C
D R
PAL conductor
Fig. 1.1 Completemeasuring system
➀ Power supply
(e.g. 230 V AC, 50 Hz)➁ Output (e.g. 0 / 4 ... 20 mA)
PAL plug
1 General DipFit W CPA 510
2
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 4/10
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 5/10
3 Mounting and installation
3.1 Dimensions
1 4 0
9
∅ 48
I m m e r s i o n d e p
t h
1 6 0
B A
0 1 1 Y 0 2 . C
D R
Pg 16reduced
➀
∅ 63
∅ 78.5
Fig. 3.1 DipFit W CPA 510
(➀) without mounting
1 8
B A 0 1 1 Y 0 3 . C
D R
1 3 0
4 ×∅ 18
Pg 16reduced
➁
∅ 90
∅ 125
∅ 165
Fig. 3.2 DipFit W CPA 510
(➁) with flange DN 50
B A
0 1 1 Y 0 4 . C
D R
➂
120
2 5 0 ± 5∅ 6
1 1
2 2
Fig. 3.3 DipFit W CPA 510(➂) with suspension bracket
B A 0 1 1 Y 0 5
. C D R
➃
Pendulumframe
Knurledbanjo bolts
Fig. 3.4 DipFit W CPA 510
(➃) with pendulum frame
(to be ordered separately)
Fig. 3.6 DipFit W CPA 510(➅) with oval flange and
mounting plate
∅ 40
B A 0 1 1 Y 0 6 . C
D R
B A 0 1 1 Y 0 9 . C
D R
➄
Fig. 3.5 DipFit W CPA 510 with
(➄) flange DN 50 with bayonet lock
➆
150 ∅
1 0 0 . 5
7 5 . 5
1 5 0
1 8 0
150
180
➅
180
Pg 16reduced
2 0
1 5 5
B A 0 1 1 Y 1 0 . C
D R
Fig. 3.7 Dimensions of oval flange
(➆)
3 Mounting and installation DipFit W CPA 510
4
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 6/10
3.2 Electrode installation
You may install electrodes with a threaded
Pg 13.5 plug-in head, a shaft length of
120 mm and a shaft diameter of 12 mm.
Fig. 3.8, ➀:
• Disengage the bayonet lock by turning theelectrode holder and pull the electrode
holder out of the assembly pipe.
Caution:
• Before installing a new electrode,
make sure that the electrode shaftis fitted with O-ring B and pressure
ring A and that the yellow
protection cap is removed.
• The electrode shaft must be
wetted before installation in the
assembly. Wetting with water issufficient.
Fig. 3.8, ➁:
• Screw in the electrode by hand and tighten it
with a size 17 socket wrench by a quarter turn.
Fig. 3.9, ➀:
• Screw the lid off the connection head and
push the electrode cable with the fittedelectrode plug through the assembly pipe.
• Screw the electrode plug onto the electrodescrew head and hand-tighten it.
• Install conductor for potential matching pin
on the existing potential matching connector(PMC).
Note:
Leave an extra measuring cablelength of approx. 10 cm in the
assembly for installation andremoval of the electrode holder.
Fig. 3.9, ➁:
• Lead the cable end through the Pg 16
clamp-type screw union of the connectionhead und fix it with the Pg union. Screw on
the lid of the connection head.
B A 0 1 1 Y 0 7 . C
D R
➀ ➁
Fig. 3.8 ➀ Disassembly of
electrode holder
➁ Installation of
pH / redox electrode
L e a v e e x t r a c a b l e l e n g t h !
B A 0 1 1 Y 0 8 . C
D R
➀ ➁
Fig. 3.9 ➀ Mounting of measuring
cable with plug
➁ Mounting of measuring
cable to electrode
A
B
PAL plug
PA510E01.CHP
DipFit W CPA 510 3 Mounting and installation
5
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 7/10
3.3 Mounting of the assembly
The standard version of DipFit W 510 is
mounted using a DN 50 flange, a DN 50
flange and union nut, a DN 50 flange withbayonet lock or an oval flange and
mounting plate.
• To remove the assembly in case of flangemounting, loosen the flange screws.
• To remove the assembly in case of flangemounting with union nut, loosen the union nut
and then pull the assembly out of its holder.
• The bayonet version is mounted by a DN 50
flange and a bayonet lock on the assembly
head.This allows to remove the assembly with the
flange firmly mounted by a simple counter-
clockwise rotation.
• To remove the assembly in case of ovalflange mounting, loosen the tommy screws
and remove the assembly via the lateral
opening in the mounting plate.
3.4 Mounting with pendulum frame
The DipFit W assembly CPA 510 is designedfor pendulous suspension from the pendulum
frame (order no. 50080196):
• Screw stop nuts on both the knurled banjobolts (included in delivery of
pendulum frame)
• Remove suspension bracket of assembly.
• Screw knurled banjo bolts into the drill
holes M8 of the retaining ring and fasten
them according to the mounting instrucitonsof the pendulum frame.
Fig. 3.11:
• Hang assembly in pendulum frame.
B A 0 1 1 Y 1 1 . C
D R
Fig. 3.10 Pendulum frame
(Order no. 50080196)
AssemblyCPA 510
pendulum frame
B A 0
1 1 Y 0 5 . C
D R
Fig. 3.11 Assembly CPA 510 mountedwith pendulum frame
3 Mounting and installation DipFit W CPA 510
6
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 8/10
4 Maintenance
4.1 Assembly
• Remove the assembly at regular intervals
(depending on the medium)
• Clean the side in contact with the medium,
if required.
• Visual inspection of the O-rings:
Replace damaged O-rings!
Caution:
• Avoid any damaging of the new
O-rings and the O-ring seat.
• Pay special attention to an
appropriate material of the
O-rings in contact with themedium.
Order numbers for spare O-rings:
4.2 Cleaning of the pH electrode
• Clean the electrode at regular intervals(depending on the medium) to guarantee
reliable measurements.
• Poor response time, low sensitivity (slope)and unstable measured values indicate
soiling.
• Clean the electrode before each calibration.
Cleaning process:
• Remove light deposits with suitable
cleaning agents (see table).
• Use a soft brush and appropriate cleaningagents to remove adhering dirt.
• Remove stubborn dirt by soaking incleaning agent.
Caution:
• Do not use abrasive cleaning
agents! These may causeirreparable damage to the
electrode measuring surface.
• Vigorously flush the electrode withdistilled water after cleaning.
Cleaning agent residue may impairmeasurements.
• Recalibrate the measuring systemafter each cleaning.
Note:
Clean Redox electrodes only
mechanically! Chemical cleaning
cause measured errors because apotential is implicated. This abates
not before a few hours.
O-ring material in contact with the medium not in contact with the medium
EPDM (standard) 50044657 50044660
Viton 50044658 50044661
Soiling, deposit Cleaning agent
Grease and oil Tenside containing (alcaline) media or watersoluble organic solvents (e.g. alcohol)
Limestone deposits, metal hydroxides,
heavy biological deposits
Hydrochloric acid 3 %
Sulphide deposits Mixture of 3 % hydrochloric acid andthio-urea (usual commercial)
Protein deposits Mixture of 0.1 molar hydrochloric acid and
pepsin (usual commercial)
Fibres, suspended materials Pressurized water, possibly with wetting agents
Light biological deposits Pressurized water
PA510E01.CHP
DipFit W CPA 510 4 Maintenance
7
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 9/10
5 Technical data
5.1 Supplementary documentation
❏ Technical InformationpH / redox combined electrodes
OrbiSint W CPS 11/12/13order no. 50054649
❏ Technical InformationpH / redox combined electrodes
OrbiTex W CPS 21order no. 50054650
General data Manufacturer Endress+Hauser
Product designation DipFit W CPA 510
Attachment CPA 510-xxx1 DN 50 flange
CPA 510-xxx2 DN 50 flange and union nut
CPA 510-xxx3 oval flange and mounting plate
CPA 510-xxx4 DN 50 flange and bayonet lock
CPA 510-xx15 suspension bracket
Materials in contact with the medium Immersion tube PVC, PVDF
Flange DN 50 PVC, UP-GF
Oval flange with mounting plate PP-GF
Electrode holder PPS with 40 % GF
O-rings EPDM, Viton
Bayonet pin for electrode holder titanium
PAL plug Hastelloy C4
Operating pressure and temperature PVC version
with flange DN 50
with oval flange and mounting plate
6 bar at 20 °C, 0 bar at 50 °C
0 bar at 50 °C
PVDF version
with flange DN 50
with oval flange and mounting plate
6 bar at 20 °C, 2 bar at 90 °C,
0 bar at 115 °C0 bar up to 80 °C
Immersion depth Standard 1000 mm
Other lengths 500 to 3000 mm
Required mounting cross section DN 50
Weight 2.3 kg
Electrodes Electrode type OrbiSint W CPS 11/12, OrbiTex W CPS
21
Installation via Pg 13.5 thread
Shaft length 120 mm
Shaft diameter 12 mm
5 Technical data DipFit W CPA 510
8
8/13/2019 BA011CEN
http://slidepdf.com/reader/full/ba011cen 10/10
Europe
Austria❑ Endress+Hauser Ges.m.b.H.WienTel.++43 (1) 8 8056-0, Fax (1) 880 56-35
BelarusBelorgsintezMinskTel.++3 75 (172) 263166, Fax (172) 2631 11
Belgium / Luxembourg❑ Endress+Hauser S.A./N.V.BrusselsTel.++32 (2) 24806 00, Fax (2) 248 0553
BulgariaINTERTECH-AUTOMATIONSofiaTel.++3 59 (2) 6648 69, Fax (2) 963 1389
Croatia❑ Endress+Hauser GmbH+Co.ZagrebTel.++3 85 (1) 663 7785, Fax (1) 6 6378 23
CyprusI+G Electrical Services Co. Ltd.NicosiaTel.++3 57 (2) 484788, Fax (2) 4846 90
Czech Republic❑ Endress+Hauser GmbH+Co.PrahaTel.++4 20 (26) 678 4200, Fax (26) 67841 79
Denmark❑ Endress+Hauser A/SSøborgTel.++45 (70) 1311 32, Fax (70) 1321 33
EstoniaElvi-AquaTartuTel.++3 72 (7) 422726, Fax (7) 4227 27
Finland❑ Endress+Hauser OyEspooTel.++3 58 (9) 859 6155, Fax (9) 8 5960 55
France❑ Endress+HauserHuningueTel.++33 (3) 896967 68, Fax (3) 8969 4802
Germany❑ Endress+Hauser Meßtechnik GmbH+Co.Weil am RheinTel.++49 (7621) 975 01, Fax (7621) 9 755 55
Great Britain❑ Endress+Hauser Ltd.ManchesterTel.++44 (161) 286 5000,
Fax (161) 99818 41GreeceI & G Building Services Automation S.A.AthensTel.++30 (1) 92415 00, Fax (1) 922 1714
HungaryMile Ipari-ElektroBudapestTel.++36 (1) 26155 35, Fax (1) 261 5535
IcelandVatnshreinsun HFReykjavikTel.++3 54 (5) 619616, Fax (5) 6196 17
IrelandFlomeaco Company Ltd.KildareTel.++3 53 (45) 8686 15, Fax (45) 868182
Italy❑ Endress+Hauser Italia S.p.A.Cernusco s/N Milano
Tel.++39 (02) 921064 21,Fax (02) 92107153
LatviaRaita Ltd.RigaTel.++3 71 (7) 312897, Fax (7) 3128 94
LithuaniaAgava Ltd.KaunasTel.++3 70 (7) 202410, Fax (7) 2074 14
Netherlands❑ Endress+Hauser B.V.NaardenTel.++31 (35) 69586 11, Fax (35) 695 8825
Norway❑ Endress+Hauser A/STranbyTel. ++47 (32) 8598 50, Fax (32) 8598 51
Poland❑ Endress+Hauser Polska Sp.z o.o.WarszawyTel. ++48 (22) 720 1090, Fax (22) 720 1085
PortugalTecnisis - Tecnica de Sistemas IndustriaisLinda-a-VelhaTel. ++351 (1) 4 1726 37, Fax (1) 418 5278
RomaniaRomconseng SRL
BucharestTel. ++40 (1) 410 1634, Fax (1) 410 1634
Russia❑ Endress+Hauser Moscow OfficeMoscowTel.++709 (5) 1 5875 71, Fax (5) 15898 64
Slovak RepublicTranscom Technik s.r.o.BratislavaTel. ++421 (74) 4 8886 84, Fax (74) 488 7112
Slovenia❑ Endress+Hauser D.O.O.LjubljanaTel. ++386 (61) 1 5922 17, Fax (61) 159 2298
Spain❑ Endress+Hauser S.A.BarcelonaTel. ++34 (93) 480 3366, Fax (93) 473 3839
Sweden❑ Endress+Hauser AB
SollentunaTel. ++46 (8) 5551 1600, Fax (8) 5551 1600
Switzerland❑ Endress+Hauser AGReinach/BL 1Tel. ++41 (61) 715 7575, Fax (61) 711 1650
TurkeyIntek Endüstriyel Ölcü ve Kontrol SistemleriIstanbulTel. ++90 (212) 275 1355,Fax (212) 26627 75
UkraineIndustria UkraïnaKievTel. ++380 (44) 268 81, Fax (44) 2 6908
YugoslaviaMeris d.o.o.BeogradTel. ++381 (11) 4 4461 64, Fax (11) 444 1966
Africa
EgyptAnasiaHeliopolis/CairoTel. ++20 (2) 41 7900, Fax (2) 41 7900
MoroccoOussama S.A.CasablancaTel. ++212 (2) 2413 38, Fax (2) 40 2657
NigeriaJ F Technical Invest.Nig. Ltd.LagosTel. ++234 (1) 6223 4546, Fax (1) 6223 4548
South Africa❑ Endress+Hauser Pty. Ltd.SandtonTel. ++27 (11) 444 1386, Fax (11) 444 1977
TunisiaControle, Maintenance et Regulation
TunisTel. ++216 (1) 7930 77, Fax (1) 78 8595
America
Argentina❑ Endress+Hauser Argentina S.A.Buenos AiresTel. ++54 (1) 145 2279 70,Fax (1) 1452279 09
BoliviaTritec S.R.L.CochabambaTel. ++591 (42) 569 93, Fax (42) 5 0981
Brazil❑ Samson Endress+Hauser Ltda.Sao PauloTel.++55 (11) 50 3134 55,Fax (11) 50 313067
Canada❑ Endress+Hauser Ltd.Burlington, OntarioTel.++1 (905) 68192 92,Fax (905) 68194 44
ChileDIN Instrumentos Ltda.SantiagoTel.++56 (2) 20501 00, Fax (2) 2 2581 39
ColombiaColsein Ltd.Bogota D.C.Tel.++57 (1) 23676 59, Fax (1) 6 1078 68
Costa RicaEURO-TEC S.A.San JoseTel.++506 (2) 961542, Fax (2 ) 961542
EcuadorInsetec Cia. Ltda.QuitoTel.++5 93 (2) 2691 48, Fax (02) 4618 33
GuatemalaACISA AutomatizacionY Control Industrial S.A.Ciudad de Guatemala, C.A.Tel.++5 02 (3) 3459 85, Fax (2) 32 7431
Mexico❑ Endress+Hauser I.I.Mexico CityTel.++52 (5) 5689 65, Fax (5) 5684 18
ParaguayIncoel S.R.L.AsuncionTel.++5 95 (21) 2139 89, Fax (21) 2265 83
UruguayCircular S.A.MontevideoTel.++5 98 (2) 9257 85, Fax (2) 92 9151
USA❑ Endress+Hauser Inc.Greenwood, IndianaTel.++1 (317) 53571 38,Fax (317) 53584 89
VenezuelaH. Z. Instrumentos C.A.CaracasTel.++58 (2) 94409 66, Fax (2) 9 4445 54
Asia
China❑ Endress+Hauser Shanghai
Instrumentation Co.Ltd.ShanghaiTel.++86 (21) 5490 2300,Fax (21) 54902303
❑ Endress+Hauser Beijing OfficeBeijingTel.++86 (10) 6834 4058,Fax (10) 68344068
❑ Endress+Hauser (H.K.) Ltd.Hong KongTel.++8 52 (2) 528 3120, Fax (2) 86541 71
India❑ Endress+Hauser India Branch OfficeMumbaiTel.++91 (22) 85214 58, Fax (22) 8 5219 27
IndonesiaPT Grama Bazita
JakartaTel.++62 (21) 79750 83, Fax (21) 7 9750 89
Japan❑ Sakura Endress Co., Ltd.TokyoTel.++81 (422) 5406 11, Fax (422) 5502 75
Malaysia❑ Endress+Hauser (M) Sdn. Bhd.Petaling Jaya, Selangor Darul EhsanTel.++60 (3) 73348 48, Fax (3) 7 3388 00
PakistanSpeedy AutomationKarachiTel.++92 (21) 77229 53, Fax (21) 7 7368 84
Papua New GuineaSBS Electrical Pty LimitedPort MoresbyTel. ++675 (3) 2511 88, Fax (3) 259556
PhilippinesBrenton Industries Inc.Makati Metro ManilaTel. ++63 (2) 638 8041, Fax (2) 638 8042
Singapore❑ Endress+Hauser (S.E.A.) Pte., Ltd.SingaporeTel.++65 (5) 6682 22, Fax (2) 6668 48
South Korea❑ Endress+Hauser (Korea) Co., Ltd.
SeoulTel. ++82 (2) 658 7200, Fax (2) 659 2838
TaiwanKingjarl CorporationTaipei R.O.C.Tel. ++886 (2) 2718 3938, Fax (2) 2713 4190
Thailand❑ Endress+Hauser Ltd.BangkokTel. ++66 (2) 99678 1120, Fax (2) 99678 10
VietnamTan Viet Bao Co. Ltd.Ho Chi Minh CityTel. ++84 (8) 833 5225, Fax (8) 833 5227
IranTelephone Technical Services Co. Ltd.TehranTel. ++98 (21) 874 6750, Fax (21) 873 7295
IsraelInstrumetrics Industrial Control Ltd.Tel-AvivTel. ++972 (3) 6 4802 05, Fax (3) 647 1992
JordanA.P. Parpas Engineering S.A.AmmanTel. ++962 (6) 4 6432 46, Fax (6) 464 5707
Kingdom of Saudi ArabiaAnasiaJeddahTel. ++966 (2) 6 7100 14, Fax (2) 672 5929
KuwaitKuwait Maritime & Mercantile Co. K.S.C.SafatTel.++9 65 (2) 4414 81, Fax (2) 4414 86
LebanonNabil IbrahimJbeil
Tel. ++961 (3) 2540 52, Fax (9) 548038Sultanate of OmanMustafa & Jawad Science & Industry Co.L.L.C.RuwiTel. ++968 (60) 2009, Fax (60) 7066
United Arab EmiratesDescon Trading EST.DubaiTel. ++971 (4) 6536 51, Fax (4) 653264
YemenYemen Company for Ghee an d Soap IndustryTaizTel. ++976 (4) 2306 64, Fax (4) 212338
Australia + New Zealand
AustraliaALSTOM Australia Ltd.
SydneyTel. ++61 (2) 9722 4777, Fax (2) 9722 4888
New ZealandEMC Industrial InstrumentationAucklandTel. ++64 (9) 415 5110, Fax (9) 415 5115
All other countries
❑ Endress+Hauser GmbH+Co.Instruments International
D-Weil am RheinGermanyTel. ++49 (7621) 975 02, Fax (76 21) 975345