8/13/2019 BA011CEN http://slidepdf.com/reader/full/ba011cen 1/10 BA 011C/07/en/11.99 No.50049295 DipFit W CPA 510 Immersion Assembly for pH/Redox Measurement Hauser + Endress The Power of Know How Operating Instructions

BA011CEN

Embed Size (px)

Citation preview

Page 1: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 1/10

BA 011C/07/en/11.99No.50049295 DipFit W

CPA 510Immersion Assembly forpH/Redox Measurement

Hauser+EndressThe Power of Know How

Operating Instructions

Page 2: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 2/10

Table of contents

1 General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1 Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

1.2 Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

2 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2.1 Scope of supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

2.2 Product structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

3 Mounting and installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

3.1 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

3.2 Electrode installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53.3 Mounting of the assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

3.4 Mounting with pendulum frame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

4.1 Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

4.2 Cleaning of the pH electrode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

5 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

5.1 Supplementary documentation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

TOC_BA.CHP

DipFit W CPA 510 Table of contents

1

Page 3: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 3/10

1 General

1.1 Application

DipFit W CPA 510 is an assembly intended to

hold a combined pH or redox electrode filledwith a gel electrolyte. It is universally suitable

for use in water and waste water treatment

systems as well as in process applications.

DipFit W CPA 510 can be used either as animmersion assembly in open containers and

channels or as a built-in assembly in tanksand closed containers.

1.2 Measuring system

A measuring system consists of: the DipFit W CPA 510 assembly

• a pH / redox electrode, 120 mm in length,suitable for the assembly and the medium

to be measured• a pH / redox measuring instrument

• a CPK 1, CPK 7 or CPK 9 measuring cable(terminated).

Option:

• Junction box VBA and measuring cable

(not terminated) for measuring cable

extension.

P R M

MEASE S CAL L

DI G

AssemblyDipFit W CPA 510

Cable gland Pg 16

Option:Junction box VBA

Electrode cable withelectrode plug (e.g. CPK 7)

Measuring instrument(e.g. Mycom CPM 152)

➀➁

Electrode plug

Electrode

Electrode holder

Protective cage     B     A     0     1     1     Y     0     1 .     C

     D     R

PAL conductor

Fig. 1.1 Completemeasuring system

➀   Power supply

(e.g. 230 V AC, 50 Hz)➁   Output (e.g. 0 / 4 ... 20 mA)

PAL plug

1 General DipFit W CPA 510

2

Page 4: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 4/10

Page 5: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 5/10

3 Mounting and installation

3.1 Dimensions

     1     4     0

     9

∅ 48

     I    m    m    e    r    s     i    o    n     d    e    p

     t     h

     1     6     0

     B     A

     0     1     1     Y     0     2 .     C

     D     R

Pg 16reduced

∅ 63

∅ 78.5

Fig. 3.1 DipFit W CPA 510

(➀) without mounting

     1     8

     B     A     0     1     1     Y     0     3 .     C

     D     R

     1     3     0

4 ×∅ 18

Pg 16reduced

∅ 90

∅ 125

∅ 165

Fig. 3.2 DipFit W CPA 510

(➁) with flange DN 50

     B     A

     0     1     1     Y     0     4 .     C

     D     R

120

     2     5     0    ±     5∅ 6

     1     1

     2     2

Fig. 3.3 DipFit W CPA 510(➂) with suspension bracket

     B     A     0     1     1     Y     0     5

 .     C     D     R

Pendulumframe

Knurledbanjo bolts

Fig. 3.4   DipFit W CPA 510

(➃) with pendulum frame

(to be ordered separately)

Fig. 3.6 DipFit W CPA 510(➅) with oval flange and

mounting plate

∅ 40

     B     A     0     1     1     Y     0     6 .     C

     D     R

     B     A     0     1     1     Y     0     9 .     C

     D     R

Fig. 3.5 DipFit W CPA 510 with

(➄) flange DN 50 with bayonet lock

150  ∅

     1     0     0 .     5

     7     5 .     5

     1     5     0

     1     8     0

150

180

180

Pg 16reduced

     2     0

     1     5     5

     B     A     0     1     1     Y     1     0 .     C

     D     R

Fig. 3.7 Dimensions of oval flange

(➆)

3 Mounting and installation DipFit W CPA 510

4

Page 6: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 6/10

3.2 Electrode installation

You may install electrodes with a threaded

Pg 13.5 plug-in head, a shaft length of

120 mm and a shaft diameter of 12 mm.

Fig. 3.8, ➀:

• Disengage the bayonet lock by turning theelectrode holder and pull the electrode

holder out of the assembly pipe.

Caution:

• Before installing a new electrode,

make sure that the electrode shaftis fitted with O-ring B and pressure

ring A and that the yellow

protection cap is removed.

• The electrode shaft must be

wetted before installation in the

assembly. Wetting with water issufficient.

Fig. 3.8, ➁:

• Screw in the electrode by hand and tighten it

with a size 17 socket wrench by a quarter turn.

Fig. 3.9, ➀:

• Screw the lid off the connection head and

push the electrode cable with the fittedelectrode plug through the assembly pipe.

• Screw the electrode plug onto the electrodescrew head and hand-tighten it.

• Install conductor for potential matching pin

on the existing potential matching connector(PMC).

Note:

Leave an extra measuring cablelength of approx. 10 cm in the

assembly for installation andremoval of the electrode holder.

Fig. 3.9, ➁:

• Lead the cable end through the Pg 16

clamp-type screw union of the connectionhead und fix it with the Pg union. Screw on

the lid of the connection head.

     B     A     0     1     1     Y     0     7 .     C

     D     R

➀ ➁

Fig. 3.8   ➀ Disassembly of

electrode holder

➁ Installation of

pH / redox electrode

     L    e    a    v    e    e    x     t    r    a    c    a     b     l    e     l    e    n    g     t     h     !

     B     A     0     1     1     Y     0     8 .     C

     D     R

➀ ➁

Fig. 3.9   ➀ Mounting of measuring

cable with plug

➁ Mounting of measuring

cable to electrode

A

B

PAL plug

PA510E01.CHP

DipFit W CPA 510 3 Mounting and installation

5

Page 7: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 7/10

3.3 Mounting of the assembly

The standard version of DipFit W 510 is

mounted using a DN 50 flange, a DN 50

flange and union nut, a DN 50 flange withbayonet lock or an oval flange and

mounting plate.

• To remove the assembly in case of flangemounting, loosen the flange screws.

• To remove the assembly in case of flangemounting with union nut, loosen the union nut

and then pull the assembly out of its holder.

• The bayonet version is mounted by a DN 50

flange and a bayonet lock on the assembly

head.This allows to remove the assembly with the

flange firmly mounted by a simple counter-

clockwise rotation.

• To remove the assembly in case of ovalflange mounting, loosen the tommy screws

and remove the assembly via the lateral

opening in the mounting plate.

3.4 Mounting with pendulum frame

The DipFit W assembly CPA 510 is designedfor pendulous suspension from the pendulum

frame (order no. 50080196):

• Screw stop nuts on both the knurled banjobolts (included in delivery of

pendulum frame)

• Remove suspension bracket of assembly.

• Screw knurled banjo bolts into the drill

holes M8 of the retaining ring and fasten

them according to the mounting instrucitonsof the pendulum frame.

Fig. 3.11:

• Hang assembly in pendulum frame.

     B     A     0     1     1     Y     1     1 .     C

     D     R

Fig. 3.10 Pendulum frame

(Order no. 50080196)

AssemblyCPA 510

pendulum frame

     B     A     0

     1     1     Y     0     5 .     C

     D     R

Fig. 3.11 Assembly CPA 510 mountedwith pendulum frame

3 Mounting and installation DipFit W CPA 510

6

Page 8: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 8/10

4 Maintenance

4.1 Assembly

• Remove the assembly at regular intervals

(depending on the medium)

• Clean the side in contact with the medium,

if required.

• Visual inspection of the O-rings:

Replace damaged O-rings!

Caution:

• Avoid any damaging of the new

O-rings and the O-ring seat.

• Pay special attention to an

appropriate material of the

O-rings in contact with themedium.

Order numbers for spare O-rings:

4.2 Cleaning of the pH electrode

• Clean the electrode at regular intervals(depending on the medium) to guarantee

reliable measurements.

• Poor response time, low sensitivity (slope)and unstable measured values indicate

soiling.

• Clean the electrode before each calibration.

Cleaning process:

• Remove light deposits with suitable

cleaning agents (see table).

• Use a soft brush and appropriate cleaningagents to remove adhering dirt.

• Remove stubborn dirt by soaking incleaning agent.

Caution:

• Do not use abrasive cleaning

agents! These may causeirreparable damage to the

electrode measuring surface.

• Vigorously flush the electrode withdistilled water after cleaning.

Cleaning agent residue may impairmeasurements.

• Recalibrate the measuring systemafter each cleaning.

Note:

Clean Redox electrodes only

mechanically! Chemical cleaning

cause measured errors because apotential is implicated. This abates

not before a few hours.

O-ring material in contact with the medium not in contact with the medium

EPDM (standard) 50044657 50044660

Viton 50044658 50044661

Soiling, deposit Cleaning agent

Grease and oil Tenside containing (alcaline) media or watersoluble organic solvents (e.g. alcohol)

Limestone deposits, metal hydroxides,

heavy biological deposits

Hydrochloric acid 3 %

Sulphide deposits Mixture of 3 % hydrochloric acid andthio-urea (usual commercial)

Protein deposits Mixture of 0.1 molar hydrochloric acid and

pepsin (usual commercial)

Fibres, suspended materials Pressurized water, possibly with wetting agents

Light biological deposits Pressurized water

PA510E01.CHP

DipFit W CPA 510 4 Maintenance

7

Page 9: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 9/10

5 Technical data

5.1 Supplementary documentation

❏   Technical InformationpH / redox combined electrodes

OrbiSint W CPS 11/12/13order no. 50054649

❏   Technical InformationpH / redox combined electrodes

OrbiTex W CPS 21order no. 50054650

General data   Manufacturer Endress+Hauser

Product designation DipFit W CPA 510

Attachment   CPA 510-xxx1 DN 50 flange

CPA 510-xxx2 DN 50 flange and union nut

CPA 510-xxx3 oval flange and mounting plate

CPA 510-xxx4 DN 50 flange and bayonet lock

CPA 510-xx15 suspension bracket

Materials in contact with the medium   Immersion tube PVC, PVDF

Flange DN 50 PVC, UP-GF

Oval flange with mounting plate PP-GF

Electrode holder PPS with 40 % GF

O-rings EPDM, Viton

Bayonet pin for electrode holder titanium

PAL plug Hastelloy C4

Operating pressure and temperature   PVC version

with flange DN 50

with oval flange and mounting plate

6 bar at 20 °C, 0 bar at 50 °C

0 bar at 50 °C

PVDF version

with flange DN 50

with oval flange and mounting plate

6 bar at 20 °C, 2 bar at 90 °C,

0 bar at 115 °C0 bar up to 80 °C

Immersion depth   Standard 1000 mm

Other lengths 500 to 3000 mm

Required mounting cross section DN 50

Weight 2.3 kg

Electrodes   Electrode type OrbiSint W CPS 11/12, OrbiTex W CPS

21

Installation via Pg 13.5 thread

Shaft length 120 mm

Shaft diameter 12 mm

5 Technical data DipFit W CPA 510

8

Page 10: BA011CEN

8/13/2019 BA011CEN

http://slidepdf.com/reader/full/ba011cen 10/10

Europe

Austria❑   Endress+Hauser Ges.m.b.H.WienTel.++43 (1) 8 8056-0, Fax (1) 880 56-35

BelarusBelorgsintezMinskTel.++3 75 (172) 263166, Fax (172) 2631 11

Belgium / Luxembourg❑   Endress+Hauser S.A./N.V.BrusselsTel.++32 (2) 24806 00, Fax (2) 248 0553

BulgariaINTERTECH-AUTOMATIONSofiaTel.++3 59 (2) 6648 69, Fax (2) 963 1389

Croatia❑   Endress+Hauser GmbH+Co.ZagrebTel.++3 85 (1) 663 7785, Fax (1) 6 6378 23

CyprusI+G Electrical Services Co. Ltd.NicosiaTel.++3 57 (2) 484788, Fax (2) 4846 90

Czech Republic❑   Endress+Hauser GmbH+Co.PrahaTel.++4 20 (26) 678 4200, Fax (26) 67841 79

Denmark❑  Endress+Hauser A/SSøborgTel.++45 (70) 1311 32, Fax (70) 1321 33

EstoniaElvi-AquaTartuTel.++3 72 (7) 422726, Fax (7) 4227 27

Finland❑  Endress+Hauser OyEspooTel.++3 58 (9) 859 6155, Fax (9) 8 5960 55

France❑  Endress+HauserHuningueTel.++33 (3) 896967 68, Fax (3) 8969 4802

Germany❑   Endress+Hauser Meßtechnik GmbH+Co.Weil am RheinTel.++49 (7621) 975 01, Fax (7621) 9 755 55

Great Britain❑  Endress+Hauser Ltd.ManchesterTel.++44 (161) 286 5000,

Fax (161) 99818 41GreeceI & G Building Services Automation S.A.AthensTel.++30 (1) 92415 00, Fax (1) 922 1714

HungaryMile Ipari-ElektroBudapestTel.++36 (1) 26155 35, Fax (1) 261 5535

IcelandVatnshreinsun HFReykjavikTel.++3 54 (5) 619616, Fax (5) 6196 17

IrelandFlomeaco Company Ltd.KildareTel.++3 53 (45) 8686 15, Fax (45) 868182

Italy❑   Endress+Hauser Italia S.p.A.Cernusco s/N Milano

Tel.++39 (02) 921064 21,Fax (02) 92107153

LatviaRaita Ltd.RigaTel.++3 71 (7) 312897, Fax (7) 3128 94

LithuaniaAgava Ltd.KaunasTel.++3 70 (7) 202410, Fax (7) 2074 14

Netherlands❑  Endress+Hauser B.V.NaardenTel.++31 (35) 69586 11, Fax (35) 695 8825

Norway❑  Endress+Hauser A/STranbyTel. ++47 (32) 8598 50, Fax (32) 8598 51

Poland❑   Endress+Hauser Polska Sp.z o.o.WarszawyTel. ++48 (22) 720 1090, Fax (22) 720 1085

PortugalTecnisis - Tecnica de Sistemas IndustriaisLinda-a-VelhaTel. ++351 (1) 4 1726 37, Fax (1) 418 5278

RomaniaRomconseng SRL

BucharestTel. ++40 (1) 410 1634, Fax (1) 410 1634

Russia❑   Endress+Hauser Moscow OfficeMoscowTel.++709 (5) 1 5875 71, Fax (5) 15898 64

Slovak RepublicTranscom Technik s.r.o.BratislavaTel. ++421 (74) 4 8886 84, Fax (74) 488 7112

Slovenia❑  Endress+Hauser D.O.O.LjubljanaTel. ++386 (61) 1 5922 17, Fax (61) 159 2298

Spain❑   Endress+Hauser S.A.BarcelonaTel. ++34 (93) 480 3366, Fax (93) 473 3839

Sweden❑  Endress+Hauser AB

SollentunaTel. ++46 (8) 5551 1600, Fax (8) 5551 1600

Switzerland❑   Endress+Hauser AGReinach/BL 1Tel. ++41 (61) 715 7575, Fax (61) 711 1650

TurkeyIntek Endüstriyel Ölcü ve Kontrol SistemleriIstanbulTel. ++90 (212) 275 1355,Fax (212) 26627 75

UkraineIndustria UkraïnaKievTel. ++380 (44) 268 81, Fax (44) 2 6908

YugoslaviaMeris d.o.o.BeogradTel. ++381 (11) 4 4461 64, Fax (11) 444 1966

Africa

EgyptAnasiaHeliopolis/CairoTel. ++20 (2) 41 7900, Fax (2) 41 7900

MoroccoOussama S.A.CasablancaTel. ++212 (2) 2413 38, Fax (2) 40 2657

NigeriaJ F Technical Invest.Nig. Ltd.LagosTel. ++234 (1) 6223 4546, Fax (1) 6223 4548

South Africa❑   Endress+Hauser Pty. Ltd.SandtonTel. ++27 (11) 444 1386, Fax (11) 444 1977

TunisiaControle, Maintenance et Regulation

TunisTel. ++216 (1) 7930 77, Fax (1) 78 8595

America

Argentina❑   Endress+Hauser Argentina S.A.Buenos AiresTel. ++54 (1) 145 2279 70,Fax (1) 1452279 09

BoliviaTritec S.R.L.CochabambaTel. ++591 (42) 569 93, Fax (42) 5 0981

Brazil❑   Samson Endress+Hauser Ltda.Sao PauloTel.++55 (11) 50 3134 55,Fax (11) 50 313067

Canada❑  Endress+Hauser Ltd.Burlington, OntarioTel.++1 (905) 68192 92,Fax (905) 68194 44

ChileDIN Instrumentos Ltda.SantiagoTel.++56 (2) 20501 00, Fax (2) 2 2581 39

ColombiaColsein Ltd.Bogota D.C.Tel.++57 (1) 23676 59, Fax (1) 6 1078 68

Costa RicaEURO-TEC S.A.San JoseTel.++506 (2) 961542, Fax (2 ) 961542

EcuadorInsetec Cia. Ltda.QuitoTel.++5 93 (2) 2691 48, Fax (02) 4618 33

GuatemalaACISA AutomatizacionY Control Industrial S.A.Ciudad de Guatemala, C.A.Tel.++5 02 (3) 3459 85, Fax (2) 32 7431

Mexico❑  Endress+Hauser I.I.Mexico CityTel.++52 (5) 5689 65, Fax (5) 5684 18

ParaguayIncoel S.R.L.AsuncionTel.++5 95 (21) 2139 89, Fax (21) 2265 83

UruguayCircular S.A.MontevideoTel.++5 98 (2) 9257 85, Fax (2) 92 9151

USA❑  Endress+Hauser Inc.Greenwood, IndianaTel.++1 (317) 53571 38,Fax (317) 53584 89

VenezuelaH. Z. Instrumentos C.A.CaracasTel.++58 (2) 94409 66, Fax (2) 9 4445 54

Asia

China❑  Endress+Hauser Shanghai

Instrumentation Co.Ltd.ShanghaiTel.++86 (21) 5490 2300,Fax (21) 54902303

❑   Endress+Hauser Beijing OfficeBeijingTel.++86 (10) 6834 4058,Fax (10) 68344068

❑   Endress+Hauser (H.K.) Ltd.Hong KongTel.++8 52 (2) 528 3120, Fax (2) 86541 71

India❑   Endress+Hauser India Branch OfficeMumbaiTel.++91 (22) 85214 58, Fax (22) 8 5219 27

IndonesiaPT Grama Bazita

JakartaTel.++62 (21) 79750 83, Fax (21) 7 9750 89

Japan❑   Sakura Endress Co., Ltd.TokyoTel.++81 (422) 5406 11, Fax (422) 5502 75

Malaysia❑  Endress+Hauser (M) Sdn. Bhd.Petaling Jaya, Selangor Darul EhsanTel.++60 (3) 73348 48, Fax (3) 7 3388 00

PakistanSpeedy AutomationKarachiTel.++92 (21) 77229 53, Fax (21) 7 7368 84

Papua New GuineaSBS Electrical Pty LimitedPort MoresbyTel. ++675 (3) 2511 88, Fax (3) 259556

PhilippinesBrenton Industries Inc.Makati Metro ManilaTel. ++63 (2) 638 8041, Fax (2) 638 8042

Singapore❑   Endress+Hauser (S.E.A.) Pte., Ltd.SingaporeTel.++65 (5) 6682 22, Fax (2) 6668 48

South Korea❑   Endress+Hauser (Korea) Co., Ltd.

SeoulTel. ++82 (2) 658 7200, Fax (2) 659 2838

TaiwanKingjarl CorporationTaipei R.O.C.Tel. ++886 (2) 2718 3938, Fax (2) 2713 4190

Thailand❑  Endress+Hauser Ltd.BangkokTel. ++66 (2) 99678 1120, Fax (2) 99678 10

VietnamTan Viet Bao Co. Ltd.Ho Chi Minh CityTel. ++84 (8) 833 5225, Fax (8) 833 5227

IranTelephone Technical Services Co. Ltd.TehranTel. ++98 (21) 874 6750, Fax (21) 873 7295

IsraelInstrumetrics Industrial Control Ltd.Tel-AvivTel. ++972 (3) 6 4802 05, Fax (3) 647 1992

JordanA.P. Parpas Engineering S.A.AmmanTel. ++962 (6) 4 6432 46, Fax (6) 464 5707

Kingdom of Saudi ArabiaAnasiaJeddahTel. ++966 (2) 6 7100 14, Fax (2) 672 5929

KuwaitKuwait Maritime & Mercantile Co. K.S.C.SafatTel.++9 65 (2) 4414 81, Fax (2) 4414 86

LebanonNabil IbrahimJbeil

Tel. ++961 (3) 2540 52, Fax (9) 548038Sultanate of OmanMustafa & Jawad Science & Industry Co.L.L.C.RuwiTel. ++968 (60) 2009, Fax (60) 7066

United Arab EmiratesDescon Trading EST.DubaiTel. ++971 (4) 6536 51, Fax (4) 653264

YemenYemen Company for Ghee an d Soap IndustryTaizTel. ++976 (4) 2306 64, Fax (4) 212338

Australia + New Zealand

AustraliaALSTOM Australia Ltd.

SydneyTel. ++61 (2) 9722 4777, Fax (2) 9722 4888

New ZealandEMC Industrial InstrumentationAucklandTel. ++64 (9) 415 5110, Fax (9) 415 5115

All other countries

❑   Endress+Hauser GmbH+Co.Instruments International

D-Weil am RheinGermanyTel. ++49 (7621) 975 02, Fax (76 21) 975345

BA 011C/07/en/11.99

Printed in Germany / CV5 / DT

❑   Members of the Endress+Hauser group

Endress HauserThe Power of Know How50049295