Upload
lyminh
View
214
Download
2
Embed Size (px)
Citation preview
2007/12/8 1
www.drummondsci.com/ products/oo_03a.html
Volume Microdispenser designed for DNA sequencing operations
www.elmat.lth.se/.../ Microdispenser.html
Microdispenser interfaces a capillary based separation system, dispensing picoliter fractions onto a chip based high density nano titre () plate.
7.5 Microdispensers
2007/12/8 2
(1) Closed-loop controlled dispensers(i) A fluidic actuator (ii) A flow sensor For continuous flow systems.
Classification scheme of microdispensers: Microdispenser
Close-loop controlled dispenser
Pump and flow sensorValve and flow sensor
Open-loop controlled dispenser
Droplet dispenser In-channel dispenser
ThermopneumaticElectrochemicalThermomechanicElectrostaticPiezoelectric
PneumaticThermopneumaticElectrokinetic
(2) Open-loop controlled dispensersTwo types:(i) Droplet dispensers: generating external droplet with constant volume.(ii) In-channel dispensers: preparing the fixed volume in the channel for further analysis on the same chip.
2007/12/8 3
7.5.1 Design consideration7.5.1.1 Droplet dispensers
Advantage: Small energy stored in the drop.
The total energy of a droplet:
d s kU U U= +Us: surface energy; Uk: kinetic energy
2s oU d=
do: droplet diameter; : surface tension 2 3 21 1
2 12k oU mv d v= =
: density of the liquid; v: droplet speed
Us/Uk increases with miniaturization. The surface energy dominates in the microscale.
2
12sk o
UU d v
=
2007/12/8 4
Nozzle: The most important part of a droplet microdispenser. To minimize the front surface effect (affecting the direction of dispensed droplets):
Making a ring-shaped nozzle with minimum front surface area.Coating the outer surface with a hydrophilic layer.
(a)(i) wet etching (ii) boron-doping (iii) blanket etching from the front(iv) selective wet etching from the front
2.
1.
3.
4.
(a)
Si Doped Si
(b)
SiO2Si
(b)(i) wet etching (ii) (ii) DRIE from the front(iii) selective oxidation to
form the nozzle shape(iv) selective dry etching
2007/12/8 5
Actuators:At ultrasonic driving frequencies:
13
2
80.34( )od f
=
f: frequency
At slow frequency or at a single droplet: o nozzled d=
Example 7-6: Thermal efficiency of a thermopneumaticdroplet dispenserGiven: A thermopneumatic droplet dispenser with:
(1)A heater resistance of 57 .(2)The ejection of a droplet following a voltage pulse
of 15 V and 0.8 s.(3)The droplet having a volume of 3.210-12 L and a
initial velocity of 9.6 m/s.Find: The thermal efficiency of this dispenser.Sol:(1) The total electrical energy of the heater:
2 26 615 0.8 10 3.16 10
57inVq t JR
= = =
2007/12/8 6
(2) The diameter of a spherical droplet:
2 3 21 12 12k o
U mv d v= = 316d o
V d= (volume of the droplet) 1 1153 36 6 3.2 10( ) ( ) 18.28 mdo
Vd
= = =
(3) The energy of droplet: 3 2
2 102.23 10 J12
od o
d vU d = + =
(4) The thermal efficiency:
0.007 %din
Uq
= =
2007/12/8 7
7.5.1.2 In-channel dispensersTwo steps in a measuring process:(1) Measuring the fluid amount needed.
The fluid amount is bordered by a stopper on one end and by outlet of delivering actuator on the other end.
(2) Pushing the measured amount to desired position using a common actuating principle.
Stopper structure: (A one-way valve)The surface tension can be adopted to realize such a passive valve. Pressure generated by surface tension:
2pr
= r: radius of curvature
and r can be used to manipulate the pressure of the stopper.
www.familyhealthmedia.com/VLU.htm
2007/12/8 8
(a) With hydrophoblic valve(b) With capillary valve(c) With electro-osmotic flow
Measuring Dispensing
Hydrophobic patch Dispensed liquid
Bubble delivered by an actuator
Capillary valve Dispensed liquid
Bubble delivered by an actuator
Electroosomotic flow
Electroosomotic flow
Dispensed liquid+
+
-
-
(a)
(b)
(c)
2007/12/8 9
Design Example: Mass Spectrometry Electro-Spray Tip
Mass Spectrometry
www.mhhe.com/.../carey/student/olc/ch13ms.html
2007/12/8 10
A mass spectrometer measures the mass of molecules and atoms. Mass spectrometers separate ions according to their mass-to-charge
ratio. Because most of the ions have a charge of +1, we can consider the mass spectrometer to separate ions with different masses.
Ions moving at high speeds through the mass spectrometer will travel along different trajectories from which their masses can be determined.
The detector of the mass spectrometer measures how many ions travel through the curved trajectory without colliding into the walls of the instrument. MagnetCurrent
regulator
Collector slit
Detector
Molecules in
Accelerator
2007/12/8 11
The mass spectrum for Boron
1. It indicates that out of 100 atoms of Boron there will be 81.7 atoms of Boron10 and 18.7 atoms of Boron11.
2.The average isotopic mass of Boron: (10 81.3/100) + (11 18.7/100) = 10.8
www.teachmetuition.co.uk/.../atomic1.htm
F (E v B)q= + F am=( / )a (E v B)m q = +
(Lorentz force law)
(Newton's second law of motion)
F is the force applied to the ion.m is the mass of the ion.a is the acceleration.q is the ionic charge.E is the electric field.v x B is the vector cross product of theion velocity and the magnetic field.
2007/12/8 12
www.sisweb.com/lc/ electrospray-tip.htm
www.nrc-cnrc.gc.ca/.../ nrc-imb_mass_spec-lr.jpg
electrospray needle
www.chemistry.nmsu.edu/.../ Image44.gif
touch.caltech.edu/.../ amish/trans97/nozzle.jpg
2007/12/8 13
Design Example: Mass Spectrometry Electrospray Tip
High performance High performance Taylor coneTaylor cone tiptip
Dead volume free tipDead volume free tip
2007/12/8 14
www.unige.ch/.../ PerretC/these_body.html www.soton.ac.uk/ ~lbrg/esims.htm
Gas and heat cause the droplets to evaporate prior to entering the instrument. As the solvent evaporates, the electric field density increases, hence the mutual repulsion between individual species becomes so great that it exceeds the surface tension and ions leave the droplet through what is known as a 'Taylor cone' .
Principle of the high performance Taylor cone tip
2007/12/8 15
1. Inside the micro channel (from the sample wall to the outlet of the spray-tip as indicated by d1)
2. From the micro channel outlet to the spray tip-head d23. From the spray tip-head to the mass spectrometry d34. Entering the the mass spectrometry
H
h'
MassspectrometerV=V1+V2+V3
S
d1 d2 d3
2007/12/8 16
Inside the micro channel d1
212
P Z constant + + =
1 12 1
2 2 2a aa aZ gZ gZ
= = =2 12b bgZ =
2p gH =(No electric field)
26p
ggH E Er
= + +
(Applied electric field)
Z a
Z b
h
H
2007/12/8 17
From the micro channel outlet to the spray tip-head d2
2 22
1 ( )2p p p T p
q V m gh m + =
sin cosea g >ea g
sin ( sin cos ) sin sin ( )sinp e p e p ea g t a t a t + + = +
cos ( cos sin ) cos cos ( )cosp e p e p ea g t a t a t + + + = +
Micro channel
sinea
cosea
ae
gsing cosg
2007/12/8 18
For
2007/12/8 19
Entering the the mass spectrometry ()
pMS
MS p
mm kq q
=
2 23
1 1 ( )2
MSMS p
MS
mV Kq k
= +T p
Due to colliding with the air molecular, the mass of the charged ions reduced.
2007/12/8 20
Mass Spectrometry Electrospray Tip Design
Design conceptSharp tip to form Taylor conePrevent the droplets from collecting at the tip
The spray-tip dimensions=20304560
7 cm 2 cm 0.024 cm
5cm
2007/12/8 21
Tip materialThe spray-tip is made of polymer PET with good hydrophobic characteristics.
Mass Spectrometry Electrospray Tip FabricationFabrication processes of the proposed spray-tip
PET film
bonding
Micromachining
cutting attaching
samplewall
Micromachining
PET film
PET film
(i) Prepare 3 pieces of 80 m thick PET film.
(ii) Clamp one of the three films using a specific fixture.
(iii) Micromachine the micro slot on the PET laminated film by a 100 m flat mill.
(iv) Bond the micro machined film with the two other pieces of film to make a micro channel
(v) Micro machine the incline with a larger flat mill.
(vi) Micro machine the tip into sharp point with the help of a CCD camera.
2007/12/8 22
Spray tip bonding
2007/12/8 23
Fixture
Micro CNC & Micro channel
2007/12/8 24
Temperature controllable roller 105~110 C
Micro machine the incline
2007/12/8 25
Micro machine the tip into sharp point with the help of a CCD camera
2007/12/8 26
High performance Taylor cone tipIincluded angle =20
Iincluded angle =30
2007/12/8 27
Iincluded angle =45
Iincluded angle =60
2007/12/8 28
Experimental Design
Sample preparation A 1 g/mL amphetamine and methamphetamine with 50% methanol containing 1% acetic acid () was selected as the sample solutions for these experiments. The molecular weights are 135 and 149, respectively.
Mass spectrometry LCQ ion-trap mass spectrometry (Finnigan MAT, San Jose, CA)
Experimental Verification
2007/12/8 29
Results
Charged liquid getting together at the tip
Taylor cone being formed
2007/12/8 30
The flat-head tip can only generate scattering Taylor cones around the outlet of the micro channel.
2007/12/8 31
Mass spectrograph of the proposed spray-tips with incline 60
2007/12/8 32
Mass spectrograph of the proposed spray-tips with incline 45
2007/12/8 33
Mass spectrograph of the proposed spray-tips with incline 30
2007/12/8 34
Mass spectrograph of the proposed spray-tips with incline 20
2007/12/8 35
Mass spectrograph of the conventional spray-tips with incline 90
2007/12/8 36
Dead Volume Free Tip
2007/12/8 37
Design of the dead volume free spray-chip
Design conceptTo embed a tiny needle penetrating the micro-channel of the spray-chip such that there is no connecting joint inside the micro channel; therefore, the dead volume problem can be avoided.
2007/12/8 38
3cm6cm75
m
0.1c
m
30o
Sample wall
Embedded needle
700 m
550 m
Micro needles with different inclined angles
10 cm 500
m
The spray-tip dimensions
2007/12/8 39
A stable and effective Taylor cone can be anticipated.
Tip materialOptical glass is selected to be the material of the spray-chip. To improve its hydrophoblic property, the micro channel surface is coated with hydrophoblic reagent. Material of the embedding needle is the high speed stainless steel (HSS).
2007/12/8 40
Samplewall
Dicing sawcutting
Opticalglass
UV-filmsticking
UVfilm
++
++
+
HSS needle LinearEDM
Fabrication of the dead volume free spray-chip
Fabricate the micro-channel on a 1mm thick optical glass by conducting dicing saw U-groove cutting.Cut out the side portions.Stick the UV-film to the front part of the already machined glass to seal the micro-channel.Sharp the high-speed stainless steel (HSS) tiny needle by linear EDM.Plug the sharpened needle into the micro-channel.Attach the sample wall to complete the dead volume free spray-chip
2007/12/8 41
Micro machined dead volume free chip
2007/12/8 42
Charged particles gather at the spray-tip and emit to the MS
Experimental Verification
2007/12/8 43
Mass spectrograph of the proposed spray-chip with needle angle =20
2007/12/8 44
Mass spectrograph of the proposed spray-chip with needle angle =30
2007/12/8 45
7.6 Microseparators
www.marstechusa.com/ tsk150.html
www.mpi-magdeburg.mpg.de/. ../1056/1071/microsep
2007/12/8 46
7.6.1 Design considerationTypes of microseparators:(1) Gas chromatography; (2) Liquid chromatography (); (3) Electrophoresis:
(i) Capillary electrophoresis (ii) Gel electrophoresis (iii) Free flow electrophoresis
Chromatography: Separation technique, which is based on the affinity differences between a stationary phase and the components of a mobile phase.
The stationary phase: Solid or liquid that interacts with the components in the mobile phase and causes their different flow velocities.
2007/12/8 47
The mobile phase: Consisting of the sample and a carrier medium.
(1)Gas chromatography: The carrier medium is a gas.
(2)Liquid chromatography: The carrier medium is a buffer solution.
Chromatogram: The time diagram of the concentration of different species in the sample. Each peak in the chromatography represents a component.
Analysis time: The time the slowest component needs to reach the sensor at the end of the separation channel.
slowest
Ltv
= L: channel length
2007/12/8 48
Separation speed:
The number of theoretical plates (N): 2
2
LN =
2: The variance contribution of longitudinal diffusion that represents the broadening of the peaks.
The smaller the signal peak compared to channel length, the higher is N and the better the separation.
The plate height (H): 2LH
N L
= =
The diffusion should be avoided to gain a better separation.
Recall 2xt
D=
The relation between the separation time and the dimension of the separation channels x.
2
1N Dt x
2007/12/8 49
Baseline resolution:The resolution between two close peak:
1 20.5( )xR
w w
=+
x: the distance between the peaksw1, w2: the widths of the peaksWhen R=1.5, the baseline resolution is reached.
2007/12/8 50
7.7 Cell Seeding of the Microvessel Scaffold
2007/12/8 51
Static Seeding
2007/12/8 52
(a) Initial stage (b) 24 hours in seeding
Cells died out due to the lack of metabolic nutrient and air.
(c) 36 hours in seeding (d) 48 hours in seeding
2007/12/8 53
Dynamic Seeding
2007/12/8 54
Major problems for long term cultivation:Aging and crack of the connection pipes and over evaporation of the cultivation medium.
(a) Initial stage (b) 12 hours in seeding
(c) 48 hours in seeding (d) 4 days in seeding
2007/12/8 55
Semi-dynamic Seeding
The cultivation medium is effected by periodically injecting fresh medium into, and sucking old medium out of, the scaffold.
Circulation frequency: 2-3 times daily.
The cultivation medium can be exchanged periodically and thus reduces the probability of virus infection.
Cultivation conditions: 37C, 5% CO2, relative humidity of 95%
2007/12/8 56
(a) 7 days in seeding
(b) 10 days in seeding
(c) 14 days in seeding
Cells can survive in the scaffold up to 4 weeks.
2007/12/8 57
PLGA Scaffold
The initial stage of seeding
24 hours of seedingThe PLGA scaffold becomes completely milky and opaque .
2007/12/8 58
Calcein-AM labeling
Fluorescence microscopy images after 7 days of seeding
Calcein-AM is the most suitable fluorescent probe for staining viable cells.
The cells are well adhered to and have grown in an orderly fashion along the circular microchannels.
2007/12/8 59
Quantum dots labeling
Fluorescence microscopy images after 96 hours of seeding
Scattered cells (dead) being leaked out during injection are observed.