Flanged Dilution Probe
New Product
Envent Engineering Ltd. 7060E Farrell Road. S.E., T2H 0T2 Calgary, Alberta, Canada. Telephone +1 403 253 4012 | Fax +1 403 253 4016 | Email [email protected]
Flanged Dilution Probe
The Envent Flanged Dilution Probe eliminatesproblems associated with sampling high H2Sconcentrations in wet, dirty gas at low pressure.
No sample pump requiredNo venting of high H2S sampleHighly resistant to contaminationSample is measured insitu at processtemperature and pressureNo H2S loss in the sample filterEnhanced safety, no exposure to process gasduring maintenanceNo power required (safe in hazardouslocations)Isolates the analyzer from the process gas
The ProblemSampling from low pressure process lines canpose problems when conventional samplingtechniques are used. Sample pumps requirefrequent maintenance, coalescing filters mayremove H2S as well as contaminants andrequired maintenance can create unsafe
working conditions when high H2Sconcentrations are analyzed. Process upsetscan contaminate the analyzer requiringmaintenance and contributing to analyzerdowntime.
The SolutionThe Envent Flanged dilutor allows percent levelH2S to be safely measured at low or nopressure. The probe completely isolates theanalyzer from process contaminants. Noregular maintenance is required.
The process H2S permeates across amembrane at a controlled rate. The permeablemembrane material is selected to be resistantto process contaminants found in natural gasand produced gas. The H2S is picked up by thecarrier gas and only the clean dry carrier istransported to the analyzer.
2" 150#316 SS RF Flange
2" 316 SS Full Port Ball Valve
BY CUSTOMERPROCESS PIPING
C ar r ie r ga s f ro m an al yz er C ar r ie r g as t o a na ly ze r
1/8" 316 SS tube 1/8" 316 SS tube
This port plugged
7/29/2019 flangeddilutor
2/2
SPECIFICATIONS
Utilities Instrument air at 100 cc/min
Ambient 0 to 100 C
Dilution Ranges 0 to 1% to 0 to 50%
Response Time 60 seconds to 90%
Process connection 2 NPT pipe or 2 150# RF flange (others on request)
1.5% of readingPrinciple of Operation Permeable membrane
Sample Pressure 0 to 10 psig
Repeatability
2" 150# 316 SS RF Flange
2" 316 SS Full Port Ball Valve
BY CUSTOMERPROCESS PIPING
Car rie r gas f rom analyz er Carr ier gas t o analyz er
1/8" 316 SS tube 1/8" 316 SS tube
This port plugged
Envent 331 H2S Analyzer (Division 2hazardous locations)
Envent 330 H2S Analyzer (Division1hazardous locations)
Calibration Integral calibration ports
Maintenance Isolation block valve