Spiral CT Dr.shravani

Embed Size (px)

Citation preview

  • 7/30/2019 Spiral CT Dr.shravani

    1/82

    By

    Dr. Shravani

  • 7/30/2019 Spiral CT Dr.shravani

    2/82

    Basic principle

    The internal structure of an object can bereconstructed from multiple projections of the same.

    The ray projections are formed by thin cross sections

    of the scanning part with a narrow x-ray beam

    The transmitted radiation from this beam is

    measured by using a sensitive radiation detector.

    This measured radiation is fed into a computer and

    analysed by using mathematical algorithms andreconstructed as a tomographic image.

  • 7/30/2019 Spiral CT Dr.shravani

    3/82

    X-ray goes through

    collimator therefore

    penetrate only an

    axial

    layer of the object,

    called

    "slice"

    How does CT Work?

  • 7/30/2019 Spiral CT Dr.shravani

    4/82

    Patient is placed in the centerof the measurement field

    X-ray is passed through thepatients slice from many

    direction along a 360 path

    The transmitted beams arecaptured by the detectorswhich digitizes these signals

    These digitized signals calledraw data are sent to acomputer which create the CTimage

  • 7/30/2019 Spiral CT Dr.shravani

    5/82

    The object slice is divided

    into small volume

    elements called voxels.

    Each voxel is assigned a

    value which is dependent

    on the average amount of

    attenuation

    How is CT Image generated?

  • 7/30/2019 Spiral CT Dr.shravani

    6/82

    The attenuation values aretransferred to the computer where

    they are coded & used to create aslice image

  • 7/30/2019 Spiral CT Dr.shravani

    7/82

    CT Data Acquisition Components

  • 7/30/2019 Spiral CT Dr.shravani

    8/82

    DATA ACQUISITIONThe scanning process begins with

    data acquisition.

    Data Acquisition refers to amethod by which the patient issystematically scanned by the X

    ray tube and detectors to collectenough information for imagereconstruction.

    A basic data acquisition schemeconsists of

    X ray tube

    Filters

    Collimators

    Detectors

  • 7/30/2019 Spiral CT Dr.shravani

    9/82

    Image reconstruction

    Substance HU

    Air -1000

    Fat -120

    Water 0Muscle +40

    Contrast +130

    Bone +400 or more

    9

    The pixel itself is displayed according to the mean attenuation

    of the tissue(s) that it corresponds to on a scale from +3071

    (most attenuating) to -1024 (least attenuating) on the

    Hounsfield scale

    Voxel is a volumetric pixel

    The HU of common

    substances:

  • 7/30/2019 Spiral CT Dr.shravani

    10/82

    CT Gantry

  • 7/30/2019 Spiral CT Dr.shravani

    11/82

    CT gantry internal components

    1.X-ray tube &

    collimator

    2.Detector assembly

    3.Tube controller

    4.High freq. generator

    5.Onboard computer6.Stationary computer

  • 7/30/2019 Spiral CT Dr.shravani

    12/82

    X-RAY TUBE

    Rotating anode type

    More heat loading and

    heat dissipationcapabilities

    Small focal spot size

    (0.6mm) to improvespatial resolution

  • 7/30/2019 Spiral CT Dr.shravani

    13/82

    FILTERS

    Compensation filter is being used

    To absorb low energy x rays

    To reduce patient dose

    To provide a more uniform beam

  • 7/30/2019 Spiral CT Dr.shravani

    14/82

    COLLIMATORS

    To decrease scatter radiation

    To reduce patient dose

    To improve image quality Collimator width determines

    the slice thickness

  • 7/30/2019 Spiral CT Dr.shravani

    15/82

    DETECTORS The detectors gather information by measuring the x-ray

    transmission through the patient.

    Two types:

    Scintillation crystal detector

    (Cadmium tungstate+ Si Photodiode)

    Can be used in third and fourth generation scanners

    Xenon gas ionisation chamber

    Can be used in third generation scanners only

  • 7/30/2019 Spiral CT Dr.shravani

    16/82

    Scintillation crystal detector used in I & II gen. CT scanners

    When X ray Photons are absorbed by these materials the

    convert them into light photons and again converted into

    electric signals by photo multiplier diode.

  • 7/30/2019 Spiral CT Dr.shravani

    17/82

    Scintillation crystal detector used in III and IV gen. CT

    scanners

  • 7/30/2019 Spiral CT Dr.shravani

    18/82

    First Generation Scanners

    Translate - rotate one detector.

    Total Scan time : 25-30 minutes (5 minutes per section).

    Uses pencil beam to scan the body.Arc of rotation of the gantry by 1 degree.

    d

  • 7/30/2019 Spiral CT Dr.shravani

    19/82

    Second Generation Scanners

    Translate rotate multiple detector (30).

    Uses fan beam.

    Multiple detectors are used.Scan time for single section : 10 -90 seconds.

    hi d i

  • 7/30/2019 Spiral CT Dr.shravani

    20/82

    Third Generation Scanners

    Rotate rotate type.

    No linear motion of the Gantry.

    Multiple detectors are used.Single section Scan time : 4 9 seconds.

    F h G i S

  • 7/30/2019 Spiral CT Dr.shravani

    21/82

    Fourth Generation Scanners

    Rotate fixed type.

    Detectors are arranged in fixed circles around the

    patient and they do not move.Single section Scan time : less then 1 second.

  • 7/30/2019 Spiral CT Dr.shravani

    22/82

    Scan Modes

    22

    Conventional (incremental) CT(short pause , several minutes)

    Spiral (helical) CT

    (helical motion of X-ray tube,reducing time and radiation

    exposure)

  • 7/30/2019 Spiral CT Dr.shravani

    23/82

    What is Spiral Scan? -- just 4C

    Continuously rotating tube/detector system

    Continuously generating X-ray

    Continuously table feed

    Continuously data acquisition

  • 7/30/2019 Spiral CT Dr.shravani

    24/82

    Newer method

    X-ray tube moves continuously with slip-ring

    technique, the outer fixed detector ring and

    inner rotating x-ray tube ring having brushes thatremain in electrical contact with the inside of

    outer ring.

    X-ray tube moves continuously with patientmoving cephalad with constant rate producing a

    spiral or helix around the patient.

    Spiral CT

  • 7/30/2019 Spiral CT Dr.shravani

    25/82

    CT SCANNING IN SPIRAL-HELICAL

    GEOMETRY BASED ON SLIP RING

    TECHNOLOGY

    Sli i T h l

  • 7/30/2019 Spiral CT Dr.shravani

    26/82

    Slip-ring Technology

    Power is transmitted through parallel sets of

    conductive ringsinstead of electrical cables.

    Continuous Gantry RotationPrerequisite for Spiral CT

    Non Slip-ring Scanner Slip-ring Scanner

  • 7/30/2019 Spiral CT Dr.shravani

    27/82

    SPIRAL CT MAJOR STEPS

    DATA ACQUSITION-ENTIRE TISSUE IS BEING

    SCANNED DURING ONE BREATH HOLD

    IMAGERECONSTRUCTION-INTERPOLATION USED TO

    GENERATE SLICES. FILTERED BACK PROJECTION USED

    TO REDUCE BLURR.

  • 7/30/2019 Spiral CT Dr.shravani

    28/82

    DATA ACQUSITION

    Z-AXIS

  • 7/30/2019 Spiral CT Dr.shravani

    29/82

    Reconstruction of arbitrary

    slices (either contiguous or

    overlapping) within the

    scanned volume

    Distance between the

    slices is called Increment.

    A BVolume Data

    Continuous dataacquisition

    During the scanning procedure, the tube is

    rotating & emitting x-rays at all time.

    The patient is advanced at a constant speed

    through the rotating gantryThis produced a volume data set from which

    multiple slices can be reconstructed either in a

    contiguous or overlapping manner.

    The images can be produced at any position

    within the scanned volume A & B

  • 7/30/2019 Spiral CT Dr.shravani

    30/82

    Increment

    Slice Thickness

    Increment = Slice Thickness No Overlap

    No Gaps

    Contiguous Image

    Reconstruction

  • 7/30/2019 Spiral CT Dr.shravani

    31/82

    Increment

    Overlap

    Slice Thickness

    Overlapping Image

    Reconstruction

    Increment < Slice Thickness Overlap of slices

    Closer image interval

    More images created

  • 7/30/2019 Spiral CT Dr.shravani

    32/82

    Increment > Slice Thickness Gaps between slices

    Images are further apart

    Less images created

    Image Reconstruction with

    Gaps

    Increment

    Slice Thickness

  • 7/30/2019 Spiral CT Dr.shravani

    33/82

    Deep Inspiration Shallow Inspiration

    Misregistration due to differentrespiratory levels between slices

    Partial Volume EffectStandard CT / Slice Imaging

    Unable to reconstruct images atarbitrary position

    Slice imaging is slow

    Conventional CT scanning is slow because it

    requires multiple slice acquisition to cover the

    volume of interest

    Despite the fact that scanning are performed

    with consecutive slices, it is easy to miss small

    lesion when the patient varies his depth of

    inspiration from scan to scan.It is also difficult to reduce partial volume effect.

    In order to do so, the lesion must be captured

    fully, not partially within the slice. Several

    overlapping scans may be repeated, with the

    hope that the patient hold his breath at the

    same depth each time.

  • 7/30/2019 Spiral CT Dr.shravani

    34/82

    Why do we need spiral/helical CT scan

    It eliminates respiratory misregistration and motionartifacts by rapid acquisition in a single breath-hold.

    Produces overlapping images without additionalradiation. From single helical CT scan data, multiple,overlapping axial sections can be retrospectivelyreconstructed.

    Multidimensional imaging is possible 2D / 3Dreformations with maximum longitudinal resolutionoptimizing image quality.

  • 7/30/2019 Spiral CT Dr.shravani

    35/82

    SPIRAL CT MODES

    SINGLE SLICE

  • 7/30/2019 Spiral CT Dr.shravani

    36/82

    SPIRAL CT MODES

    MULTI - SLICE

  • 7/30/2019 Spiral CT Dr.shravani

    37/82

    Conventional CT Alternate translation & x-

    ray exposure is done.

    Each rotation of x-ray tubegenerates data of atransaxial image

    Prospective selection of

    slice positioning andspacing

    Spiral CT Both are done

    simultaneously

    Each rotation yields data

    Specific to an angledplane of section. Fromthis, by interpolation,transaxialimage is reconstructed.

    Done retrospectively .

  • 7/30/2019 Spiral CT Dr.shravani

    38/82

    Protocol factors of spiral CT scan

    Slice thickness / Collimation:

    5 to 8mm for abdomen scan.7 to 10 mm for chest scan.

    5 mm for neck scan.

    2 to 3 mm to scan small structures like lung nodules, smaller

    blood vessels.It affects effective slice thickness of reconstructed transaxial

    scan and its longitudinal resolution.

  • 7/30/2019 Spiral CT Dr.shravani

    39/82

  • 7/30/2019 Spiral CT Dr.shravani

    40/82

    Advantages of spiral overconventional CT

    Shorter scanning timeReconstruction along z axis with overlapping

    of successive images leading to detection of

    small lesionsFewer motion artifacts

    3D reconstruction of image

  • 7/30/2019 Spiral CT Dr.shravani

    41/82

    Applications:

    CT angiography.

    Cardiac CT with gating to freeze cardiac

    motion. Coronary artery imaging including calcium

    scoring.

    Tumor permeability studies.

    Functional and perfusion imaging. In Emergency acute abdomen cases.

  • 7/30/2019 Spiral CT Dr.shravani

    42/82

    Despite high heart rate of 90 bmp,

    motion free visualization of the

    coronary arteries.

    .

  • 7/30/2019 Spiral CT Dr.shravani

    43/82

    Hepatic Laceration & Adrenal hematoma due to abdominal trauma

    Fig: A Fig: B

  • 7/30/2019 Spiral CT Dr.shravani

    44/82

  • 7/30/2019 Spiral CT Dr.shravani

    45/82

  • 7/30/2019 Spiral CT Dr.shravani

    46/82

  • 7/30/2019 Spiral CT Dr.shravani

    47/82

  • 7/30/2019 Spiral CT Dr.shravani

    48/82

  • 7/30/2019 Spiral CT Dr.shravani

    49/82

    Hepatic focal fibrosis misdiagnosed as hepatocellular

  • 7/30/2019 Spiral CT Dr.shravani

    50/82

    carcinoma. Axial helical CT images obtained during

    hepatic arterial and portal venous phases reveal multiple

    round, hypo attenuating areas (arrows) in right and left

    liver lobes. This finding simulated hypo vascular massand was originally misdiagnosed as possible tumour.

    ARTIFACTS IN COMPUTED

  • 7/30/2019 Spiral CT Dr.shravani

    51/82

    ARTIFACTS IN COMPUTED

    TOMOGRAPHY

  • 7/30/2019 Spiral CT Dr.shravani

    52/82

    ARTIFACT

    Any discrepancy between the CT numbers

    represented in the image and the

    expected CT number based on the linear

    attenuation coefficient

  • 7/30/2019 Spiral CT Dr.shravani

    53/82

    PATIENT INDUCED ARTIFACTS

    Motion

    Beam Hardening Metal Artifact

    Out Of Field Artifact

  • 7/30/2019 Spiral CT Dr.shravani

    54/82

    MOTION ARTIFACT CAUSED BY:

    Voluntary motion.

    Involuntary motion.

    It produces ghosting effect object

    Ct image appears as if composed ofSuperimposed images

    MOTION

  • 7/30/2019 Spiral CT Dr.shravani

    55/82

    MOTION

    MOTION

    http://radiographics.rsnajnls.org/content/vol24/issue6/images/large/g04nv25g16x.jpeg
  • 7/30/2019 Spiral CT Dr.shravani

    56/82

    MOTION

    MOTION

  • 7/30/2019 Spiral CT Dr.shravani

    57/82

    MOTION

    MOTION

  • 7/30/2019 Spiral CT Dr.shravani

    58/82

    MOTION

  • 7/30/2019 Spiral CT Dr.shravani

    59/82

    REMEDY:

    VOLUNTARY MOTION: Explanation Of The

    Procedure And Good Communication With A

    Patient.

    INVOLUNTARY MOTION: Short Scan Time.

  • 7/30/2019 Spiral CT Dr.shravani

    60/82

    BEAM HARDENING

    It occurs when the average energy of an x-raybeam passing through the patient increases.

    Beam is hardened high energy photons are

    attenuated less by the tissue. As a result, they

    are pass through the patient and reach thedetectors.

    This artifact is also called cupping artifactbecause the hardening is most pronounced in

    the center and less at the periphery. It resembles a cup.

  • 7/30/2019 Spiral CT Dr.shravani

    61/82

    COMMON AREAS OF MANIFESTATION:

    Skull Petrous Pyramids

    Upper Chest And Shoulders Hips

    BEAM HARDENING

  • 7/30/2019 Spiral CT Dr.shravani

    62/82

    BEAM HARDENING

  • 7/30/2019 Spiral CT Dr.shravani

    63/82

    BEAM HARDENING - CONTRAST

    http://radiographics.rsnajnls.org/content/vol24/issue6/images/large/g04nv25g04x.jpeg
  • 7/30/2019 Spiral CT Dr.shravani

    64/82

    BEAM HARDENING

  • 7/30/2019 Spiral CT Dr.shravani

    65/82

    REMEDY

    INCREASE kvp

    Decrease slice thickness

    Increase filtration bowtie filter

  • 7/30/2019 Spiral CT Dr.shravani

    66/82

    METAL ARTIFACT

    Manifest itself as star

    streaking artifact. Its

    caused by presence of

    metallic objects inside oroutside the patient.

    Metallic object absorbs

    the photons causing anincomplete profile

    METAL ARTIFACT

  • 7/30/2019 Spiral CT Dr.shravani

    67/82

    METAL ARTIFACT

    REMEDY

    Removal of external metallic objects

  • 7/30/2019 Spiral CT Dr.shravani

    68/82

    OUT OF FIELD ARTIFACT

    Patient is not entirely enclosed in the scanningfield of view. Patients body can obstruct

    detectors. In addition, patient tissue outsidethe sfov will further harden the x-ray beam.Artifact appears as streaks and shading.

  • 7/30/2019 Spiral CT Dr.shravani

    69/82

    OUT OF FIELD ARTIFACT

  • 7/30/2019 Spiral CT Dr.shravani

    70/82

  • 7/30/2019 Spiral CT Dr.shravani

    71/82

    OUT OF FIELD ARTIFACT

  • 7/30/2019 Spiral CT Dr.shravani

    72/82

    REMEDY:

    Selection of larger sfov

    Taping patient tissue

    Raising patients arms above their headon the scan of chest and abdomen

    RING

  • 7/30/2019 Spiral CT Dr.shravani

    73/82

    RING

    RING

    http://radiographics.rsnajnls.org/content/vol24/issue6/images/large/g04nv25c19x.jpeg
  • 7/30/2019 Spiral CT Dr.shravani

    74/82

  • 7/30/2019 Spiral CT Dr.shravani

    75/82

    REMEDY

    Detector calibration

    Detector replacement

  • 7/30/2019 Spiral CT Dr.shravani

    76/82

    TUBE ARCING

    Tungsten vapor from anode and

    cathode intercepts the projectile

    electrons intended for collisions withthe target. Crackling sound!!!!

    TUBE ARCING

  • 7/30/2019 Spiral CT Dr.shravani

    77/82

    TUBE ARCING

  • 7/30/2019 Spiral CT Dr.shravani

    78/82

    REMEDY

    GAS BURNOFF

    TUBE REPLACEMENT

    LINE IN TOPOGRAM

  • 7/30/2019 Spiral CT Dr.shravani

    79/82

    LINE IN TOPOGRAM

    BAD DETECTOR CAUSESCONTINUOUS LINE ONTHE TOPOGRAM

    REMEDY:

    DETECTOR REPLACEMENT

    STAIRCASE

  • 7/30/2019 Spiral CT Dr.shravani

    80/82

    STAIRCASE

    Improper selection of slice thicknessand slice incrementation when

    generation mpr and 3-d images

  • 7/30/2019 Spiral CT Dr.shravani

    81/82

    REMEDY

    THIN SLICE USE

    50% OVERLAP ON RECON SLICE

    INCREMENTATION

  • 7/30/2019 Spiral CT Dr.shravani

    82/82

    Thank You