Vela User�s Guide
Table of Contents
Vela Ventilator� Front Panel ................................................................................................................................1
Vela Ventilator� Rear Panel ................................................................................................................................3
Circuit Assembly....................................................................................................................................................5
Touch-Turn-Touch�/Touch-Turn-Accept�............................................................................................................9
Extended Function Testing ................................................................................................................................11
Fi02 Monitor Calibration ....................................................................................................................................13
New Patient Set-up ............................................................................................................................................15
Setting the Ventilation Breath Type and Mode ................................................................................................17
Monitored Parameters ......................................................................................................................................19
Graphics ..............................................................................................................................................................23
Synchronized Nebulizer......................................................................................................................................29
Oxygen Computer Chart ....................................................................................................................................31
Contact Information............................................................................................................................................33
Notes ..................................................................................................................................................................34
The Vela User�s Guide is not intended as a replacement for the Operator�s Manual. You must become completely familiar with the Vela Operator�s Manual before using the Vela ventilator.
This page intentionally left blank.
ACDC
DCSTATUS
CANCEL
ACCEPT
PANELLOCK
NEBULIZER 100%O2
MANUALBREATH
EXPHOLD
FREEZE INSPHOLD
ALARMSILENCE
ALARMRESET
P R E S S U R E A / C MAIN
ADVSETTINGS LIMITS SETUP
mLVti
mLVte
cmH2OPpeak
cmH2OPEEP
%FiO2
Man Brth Silence
5
PEEPcmH2O
5
Flow trigLPM
21
%O2%
22
Insp PresscmH2O
1.0
Insp TimeSec
20
RateBPM
39138028521
Lk Comp Vol Min
40
20
-208040
-40
0
0
-80
750500250
0-250
Paw (cmH O)2
V (lpm).
V (ml)
2 4 6 8 10 12
2 4 6 8 10 12
2 4 6 8 10 12
t
FROM PATIENT
TOPATIENT
PRESSURERELIEF
MONITORED PARAMETERS
MODEINDICATOR
SCREENINDICATOR
ALARM STATUSINDICATOR
DATA DIAL
PRIMARYCONTROLS
MESSAGEWINDOW
EFFORTINDICATOR
POWERINDICATORS
NEBULIZEROUTPUT
EXPIRATORYFLOW SENSOR
1 Vela User�s Guide L1536 Rev. C
Vela Ventilator Front Panel
Vela Front Panel
2 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
3 Vela User�s Guide L1536 Rev. C
Vela Ventilator Rear Panel
A - Power Switch
B - Fan and Fan Filter
C - Printer Port
D - VGA Port
E - Non-Operational Port
F - High Pressure Oxygen Fitting
G - Low Pressure Oxygen Fitting
H - Nurse Call System Connection
I - Alarm Speaker
J - Power Cord
K - Fuses
L - Ground
M - Alarm Volume Adjustment
N - Fiber Optic Output Port
A
B
C
D
E
F
G
H
I
J
K
L
MN
Vela Rear Panel
4 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
5 Vela User�s Guide L1536 Rev. C
Circuit Assembly
1. Install the diaphragm on the exhalation valve.Carefully seat the rim of the diaphragm on theexhalation valve and gently press around the rimto ensure it is seated evenly as shown.
2. Line up the fins on the exhalation valve bodywith the exhalation valve housing.
Circuit Assembly
6 Vela User�s Guide L1536 Rev. C
Circuit Assembly
3. Rotate clockwise until it clicks into place. 4. Attach the flow sensor with the flow tubes in anupright position.
7 Vela User�s Guide L1536 Rev. C
Circuit Assembly
5. To connect the smart connector for the variable orifice flow sensor, pull back the lockingsleeve. Push into the receptacle and slide thelocking sleeve forward to secure. To remove,repeat these steps in reverse order.
8 Vela User�s Guide L1536 Rev. C
FROM PATIENT
NEBULIZEROUTPUT
EXPIRATORY LIMB OF PATIENT CIRCUIT
INSPIRATORY LIMB OF PATIENT CIRCUIT
TOPATIENT PRESSURE
RELIEF
EXPIRATORYFLOW SENSOR
Circuit Assembly
Complete Patient Circuit Assembly
9 Vela User�s Guide L1536 Rev. C
Touch-Turn-Touch� / Touch-Turn-Accept�
Changes to controls and displays are accomplished with a three-step technique.
1. Touch the control or display to be changed. 2. Turn the data dial to reach the desired setting.
3. Touch the control again OR press "ACCEPT" toconfirm the change.
Changes not accepted by either method within 15 seconds revertto previous settings. Press "CANCEL" prior to accepting proposedchanges to return to previous settings.
BPMRate
15ml
Volume
500
Touch-Turn-Touch� /Touch-Turn-Accept�
10 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
11 Vela User�s Guide L1536 Rev. C
Extended Functions Testing
The extended functions menu allows access to stored data, performance calibrations and customization of the frontpanel. Touch the screen indicator to access the extended functions menu. When the screen select menu appears,press the Extended Functions button.
Extended Functions Menu
Extended Functions Testing
12 Vela User�s Guide L1536 Rev. C
Extended Functions Testing
Events Stores data events for Service evaluation & troubleshooting. *
Transducer Data Allows setting of transducer analog outputs for Service. *
Transducer Test Allows Service testing of transducer function. *
Version Info Displays software version information and turbine serial number.
Date/Time Displays total hours of ventilator operation and date/time configuration.
Fi02 Monitor Calibration Allows calibration of internal oxygen sensor.
(Comprehensive & Standard models only).
Vent Set-up: Allows setting of these functions:
Low Min Vol OFF Enable or disable an "OFF" setting
Panel Lock Enable or disable the front panel lock switch.
Fi02 Monitor Turns the Fi02 monitor on or off. (Comprehensive & Standard models only).
Altitude units of measure Toggles between feet and meters for the altitude setting.
Altitude setting Allows setting of altitude for accurate volume measurement.
Language buttons Select the desired language for the front panel.
Video Normal/Inverse Reverses the color configuration of the graphic interface.
For a complete explanation of each function, refer to the Vela Service Manual (P/N L1534).
* Denotes function for use by trained Service technician.
13 Vela User�s Guide L1536 Rev. C
Fi02 Monitor Calibration
When the Fi02 Mon Calibration button is touched, the calibration menu appears. You have a choice of doing an ambi-ent air calibration or a 100% oxygen calibration. Either can be done on or off a patient; however, be certain that thepatient can tolerate room air or 100% oxygen for the calibration procedure, which takes approximately 4 minutes tocomplete.
To exit the Fi02 Calibration screen, touch the EXIT button.
Fi02 Monitor Calibration
14 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
15 Vela User�s Guide L1536 Rev. C
New Patient Set-Up
1. Power on the ventilator and perform operational verification testing asdescribed in Chapter 2 of your Operator's Manual.
2. Select "NEW PATIENT". Touch "PATIENT ACCEPT"
3. Select Active or Passive Humidification.
4. Enter patient ID if desired.
5. Select �LEAK COMP� ON/OFF. Note: Default for NPPV A/C, NPPV SIMV; NPPV CPAP PSV; �LEAK COMPON�
6. Touch "SETUP ACCEPT"
The Set-up screen can be accessed at any time by pressing the "Setup" button on the bottom right corner of the touch screen.
PATIENT ACCEPT
NEWPATIENT
RESUMECURRENT
PATIENT SELECT
PATIENT
SETUP ACCEPT
IDENTIFICATION
ACTIVE
ONLEAK COMP
ON
HUMIDIFIER
Selecting "NEW PATIENT" clears all trends and saved loops, and sets all controls todefault values. Selecting "RESUME CURRENT" resumes ventilation at previous con-trol settings.
Touch the Patient ID box to open a virtual keyboard and use Touch-Turn-TouchTM/Touch-Turn-AcceptTM to enter alphanumeric characters.
New Patient Set-up
16 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
17 Vela User�s Guide L1536 Rev. C
MODE ACCEPT
CPAP PSV
APRV BiPhasic
APNEA SETTINGS
Pressure
PressureSIMV
VolumeSIMV
PressureA/C
PRVC SIMV
PRVCA/C
NPPV SIMV
NPPV CPAP PSV
NPPVA/C
VolumeA/C
MODE SELECT
APNEA MODE
Volume
Setting the Ventilation Breath Type and Mode1. Touch the Mode Indicator to open the mode window.
2. Select the desired mode and set the primary controls.
3. Touch "MODE ACCEPT" to accept the new mode and control settings.
4. Advanced settings and alarm limits may also be adjusted at this time. The buttons that open the advanced settingand alarm limits windows are located in the lower right hand corner of the touch screen.
Setting the VentilationBreath Type and Mode
18 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
19 Vela User�s Guide L1536 Rev. C
Monitored Parameters
Set Main Screen Monitored Parameters by using theTouch-Turn-TouchTM/Touch-Turn-AcceptTM technique.
Main Screen MonitorsFive monitored parameters are continuously
displayed to the left of the graphics display.Selections are made from a scrolling menu. Monitored parameters appearing on the
main screen may be different from those that appear on the optional loops screen.
Monitored Parameters
20 Vela User�s Guide L1536 Rev. C
Monitored ParametersTo configure an extended display for easy charting,
LOOP
MANEUVER
SCREEN SELECT
MONITOR
EXTENDEDFUNCTIONS
MAIN
TRENDS
1. Touch the Screen Indicator in the top center of the Main Screen display.
2. The Screen Select box will display.
3. Select MONITOR from the selection box that appears.
21 Vela User�s Guide L1536 Rev. C
Monitored Parameters
18.5mLVte
0.0L
Spon Ve
0cmH20
Ppeak
0mLVti
0.0bmpRate
0.00cmH20
Pmean
26.7mL
Spon Vt
0.0bmp
Spon Rate
0cmH20PEEP
0.0mL
Mand Vt
0secTi
8PSIG
02 Regulated
0.0L
Ve
0.00secTe
0%
Fi02
A 15-window display is configurable from a scrolling menu. Use the Touch-Turn-TouchTM/Touch-Turn-AcceptTM
technique to configure this display.
22 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
23 Vela User�s Guide L1536 Rev. C
GraphicsConfigure main screen graphics for scale and sweep speed.
1. Touch the vertical or horizontal axis to highlight. 2. Turn the data dial to adjust scale or sweep speed.
3. Touch the axis again or press accept to confirm the change.
Graphics
24 Vela User�s Guide L1536 Rev. C
GraphicsConfigure the optional Loops Screen
1. Touch the vertical or horizontal axis to highlight. 2. Turn the data dial to adjust the scale.
3. Touch the axis again or press accept to confirm
the change.
250 500 750
25 Vela User�s Guide L1536 Rev. C
Graphics
The FREEZE button freezes the current screenuntil pressed again. When the screen is frozenyou can scroll through displayed waveforms orloops using the Data Dial to move the cursor.
250 500 750
Cursor, currentlyoverlays the "X"axis at Zero
Flag showing Xand Y values atvarious pointsalong the looptracingDashed
Cursor Line
Flow/VolumeLoop Tracing
270
0.0
A flow/volume loop in "freeze" mode is depictedbelow. As the dotted line cursor traces the "frozen"loop curve, flags display the values along the curveof the loop.
26 Vela User�s Guide L1536 Rev. C
GraphicsSaving loops
1. Press the "Freeze" button to freeze the graphics display. 2. Press "Save Loop"
3. Saved loops will appear with a time reference. A total of 4 loops can be saved at any time. The oldest loop is discarded when the 5th loop is saved.
250
270
500 750
0.0
27 Vela User�s Guide L1536 Rev. C
GraphicsCreate Reference Loops
1. Press the "Freeze" button to freeze the graphics display.
2. Touch the saved loop you want as a reference.
4. Press the "Freeze" button to return to live loops display over the reference loop.
Discontinue Reference Loops
To discontinue the reference, press the "Freeze" button to freeze thegraphics display and press the "Ref Loop On" button to turn off the refer-ence loop. Press "Freeze" again to return to a live loop screen.
3. Press the "Ref Loop Off" button to toggle the reference loop on.
28 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
29 Vela User�s Guide L1536 Rev. C
Synchronized Nebulizer
A high pressure oxygen source gas connection is required to power the in-line nebulizer. Nebulizer flow delivery is synchronized with inspiration.Nebulization lasts for up to 60 minutes and is user adjustable in incrementsof one minute. Pressing the nebulizer button again ends nebulization.
Caution: Nebulizer output is not volume compensated during volume controlbreaths. Delivered volumes and peak inspiratory pressures will increase dur-ing nebulization. To ensure consistent tidal volume delivery, tidal volumeshould be reduced for the duration of the treatment. Nebulizer output isapproximately 50 ml per half second of inspiratory time. Tidal volume is unaf-fected during pressure control breaths.
Fi02 will increase slightly during nebulization. Delivered Fi02 during nebulization can be estimated using the following calculation;
New Fi02 = [(Vt x Fi02) + (0.1 x Ti)][Vt + (0.1 x Ti)]
NEBULIZER100%
O2
MANUALBREATH
EXPHOLD
FREEZE INSPHOLD
ALARMSILENCE
ALARMRESET
2
2
Synchronized Nebulizer
30 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
31 Vela User�s Guide L1536 Rev. C
Oxygen Computer Chart
To Find O2 Input Flow:
1. Select desired FiO2 on horizontal axis.
2. Project up to current minute volume (Ve from monitor display).
3. Project horizontally to left vertical axis andread oxygen flow.
To Find O2 Concentration:
1. Select current O2 input flow on vertical axis.
2. Project horizontally right to current minutevolume (Ve from monitor display).
3. Project vertically down to horizontal axisand read oxygen concentration.
30
25
20
15
10
5
0.21 .30 .40 .50 .60 .70 .80 .90 1.0
FiO2
1234
6
10
14
18Ve = 20
Oxy
gen
Inpu
t flo
w (
L/m
in)
12
8
16
Oxygen Computer Chart
32 Vela User�s Guide L1536 Rev. C
This page intentionally left blank.
33 Vela User�s Guide L1536 Rev. C
Contact Information
VIASYS Healthcare Customer Care-Technical ServiceHours: 7:30 AM to 4:30 PM (PST) M-FPhone: (760) 778-7200Fax: (760) 778-7377
VIASYS Healthcare Customer Care HelplineHours: 24 Hours, 7 days a weekPhone: (800) 934-2473Fax: (760) 778-7377
VIASYS Healthcare- Critical Care Division1100 Bird Center DrivePalm Springs, CA 92262-8099 U.S.A.Phone: (760) 778-7200, (800) 328-4139Fax: (760) 778-7274
www.ViasysCriticalCare.com
Contact Information
34 Vela User�s Guide L1536 Rev. C