Development of an analytical reversed phase column for characterization of intact antibodies using wide pore monolithic silica. Shigenori Ota, Yuko Yui, Shota Miyazaki, ChiakiAoyama, Ken Miyashita, Manami Takeda, Shunta Futagami, Tsutomu Sato GL Sciences Inc. Introduction The production and purification of monoclonal antibodies (mAb), comprehensive characterization can be performed at the intact protein on a mass spectrometry. Reversed phase columns with a wide pore (>30nm) particle are used for intact protein chromatography analysis. However, analytical column packed wide pore show low pressure resistance due to the physicochemical durability of support material. Monolithic silica have been widely studied due to their unique structures that have many advantages as separation media(Figure). Monolith structure and pore size were easily controlled by starting material. In this study, we have developed the monolithic silica column with wide pore size suitable for the intact protein analysis. The performance of the developed column was compared with columns packed with silica particles. Monolithic silica gel with an inner diameter of 2.1 mm was prepared via sol-gel processing of tetraethoxysilane. Then, monolithic silica was treated with alkali solution to prepare wide pore size (60nm). The resulting monolithic silica was cladded with a glass tube. After cutting it to an appropriate length, monolithic silica surface was modified with the phenyl group or the C4 group. Reversed phase modified monolithic silica column in this study was fabricated by sealing the cladded silica in a stainless steel tube. Intact protein (mAb) analysis were performed on a HPLC-UV and TripleTOF 6600 mass spectrometer (Sciex). Methods The recoveries of intact mAbs was compared on the two particle packed columns (silica gel type and polymer gel type) and developed columns, it appeared that adsorption was less pronounced on the silica gel particle packed, probably due to the size of the small pore size. Developed monolithic column was shown the excellent recoveries to intact mAb. The pressure drop of the developed column is typically low that of a column packed with polymer column. A mAb require buffering salt for stabilization in solution. However, to gain sensitivity in LC-MS analysis of intact mAbs, it is critically important to remove the salt before LC-MS analysis. Figures show the LC/MS data for the intact mAbs with developed column and silica gel columns, respectively. The developed column provided excellent total ion chromatogram peak shapes compared with silica gel column. The temperature effect for resolution was compared. The heat resistance of a monolithic column clad with resin was typically up to 50 °C, but clad the monolithic material with glass improved the columns heat resistance to 80. Recovery of mAb were significantly improved with elevated temperature, and adsorption was generally acceptable in the temperature range between 60 and 80. As temperature increases, peak widths of the protein peaks has decreased and absolute retention of each peak was reduced with increasing temperature. The Stability of the developed column at higher temperature and presence of trifluoroacetic acid has also improved by cladded with a glass tube. Results & Discussion 1 μm Through pore 60 nm Meso pore Φ2.1mm Monolithic silica Stainless steel tube Glass tube Assembly Schematic of the developed monolithic silica column Monolithic silica disk SPE Spin column Absorbent material Capillary HPLC column Monolithic silica application fields 0.0 1.0 2.0 3.0 4.0 5.0 6.0 Time (min) 0.00 0.10 Volts 0.0 1.0 2.0 3.0 4.0 5.0 6.0 Time (min) 0.00 0.10 Volts 0.0 1.0 2.0 3.0 4.0 5.0 6.0 Time (min) 0.00 0.10 Volts Developed column Silica gel packed column Polymer packed column 4.5MPa 10.5MPa Pressure drop 1.7MPa Peak height 135451 25136 113250 Conditions Eluent: A0.1 %TFA-H 2 O B0.1 %TFA-ACN Gradient: A/B A/B=90/10-2.5min-10/90 Flow Rate: 0.3 mL/min Oven Temp.: 80 Injection Vol.: 5 μL Retention time [min] Injection [times] Developed monolithic silica clumn Conditions Eluent: A) 0.075% HCOOH+0.025% TFA - H2O B) 0.075% HCOOH+0.025% TFA - ACN A/B=95/5-5min-10/90 Flow Rate: 0.3 mL/min Detection: UV 210 nm Oven.Temp: 80 Analyte: 1.4 mg/mL mAb Column performance of the developed column with competition. 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 4 0 500 1000 1500 2000 2500 3000 2.2 2.4 2.6 2.8 3 3.2 3.4 3.6 3.8 0 500 1000 1500 2000 2500 3000 Retention time [min] Injection [times] 0.0 1.0 2.0 3.0 4.0 5.0 6.0 Time (min) 0.0 0.2 0.4 Volts 50 60 80 0.0 1.0 2.0 3.0 4.0 5.0 6.0 Time (min) 0.0 0.2 0.4 Volts 0.0 1.0 2.0 3.0 4.0 5.0 6.0 Time (min) 0.0 0.2 0.4 Volts 0.0 1.0 2.0 3.0 4.0 5.0 6.0 Time (min) 0.0 0.2 0.4 0.6 Volts 40 Temperature effect Peak tailing Durability The graph below shows the durability test. In response to the increase of injection times (x-axis), the durability of each column is plotted at y-axis with the retention time of the mAb. The developed column modified phenyl groups shows great durability under low pH and high temperature conditions. The lifetime of phenyl groups modified column is much greater than that of the C4 modified columns. This assures stable analysis over a long period of time. LC/MS intact mAb analysis >10 μm <1 μm The pores can be sized according to application, resulting in a highly adaptive material for chromatography. Developed column Silica gel packed column 0.0 1.0 2.0 Time (min) 0 200 400 600 800 1000 mVolt 0.0 1.0 2.0 Time (min) 0 200 400 600 800 mVolt Conditions Eluent: A)0.08%HCOOH-0.02%TFA-H 2 O B)0.08%HCOOH-0.02%TFA-ACN A/B=80/20-2.5 min-10/90 Fab in CHO cell culture Purified sample Flow rate0.3 mL/min Col.Temp.60 Detection210 nm Injection Vol.1 μL M Electrophoresis ; Purified with MonoSpin ProL ; CHO cell culture M ; Marker Fab CHO cell culture direct analysis Purified by ProteinL immobilized Spin column Conclusions The developed monolith column has high porosity, it was possible to analyze the culture broth without clogging. C4 group Phenyl group Separation of Intact mAb using a developed monolithic column at different temperatures. Analyte: 1.4mg/mL mAb 5 uL injection Cladding Conditions Eluent: A0.1 %HCOOH-H 2 O B0.1 %HCOOH-ACN Gradient: A/B=90/10-10min-10/90 Flow Rate: 0.3 mL/min Detection: TripleTOF 6600 Oven Temp.: 80 Injection Vol.: 5μL Intact mAb mass analysis (total ion chromatogram) on (A) an developed column, (B) silica packed column. (A) (B) The developed column was shown to be very suitable for analysis of intact protein (mAb). The Novel glass cladded monolithic silica column was developed. Fab

GL Sciences Inc. 0 0.0 1.0 2.0 3.0 4.0 5.0 6.0 0 0.0 1.0 2 ... · GL Sciences Inc. Introduction The production and purification of monoclonal antibodies (mAb), comprehensive characterization

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Development of an analytical reversed phase column for characterization

of intact antibodies using wide pore monolithic silica.

Shigenori Ota, Yuko Yui, Shota Miyazaki, Chiaki Aoyama, Ken Miyashita, Manami Takeda, Shunta Futagami, Tsutomu Sato

GL Sciences Inc.

IntroductionThe production and purification of monoclonal antibodies (mAb), comprehensive characterization can be

performed at the intact protein on a mass spectrometry. Reversed phase columns with a wide pore

(>30nm) particle are used for intact protein chromatography analysis. However, analytical column packed

wide pore show low pressure resistance due to the physicochemical durability of support material.Monolithic silica have been widely studied due to their unique structures that have many advantages

as separation media(Figure). Monolith structure and pore size were easily controlled by starting material.

In this study, we have developed the monolithic silica column with wide pore size suitable for the intact

protein analysis. The performance of the developed column was compared with columns packed with

silica particles.

Monolithic silica gel with an inner diameter of 2.1 mm was prepared via sol-gel processing of

tetraethoxysilane. Then, monolithic silica was treated with alkali solution to prepare wide pore size

(60nm). The resulting monolithic silica was cladded with a glass tube. After cutting it to an appropriate

length, monolithic silica surface was modified with the phenyl group or the C4 group. Reversed phase

modified monolithic silica column in this study was fabricated by sealing the cladded silica in a

stainless steel tube.Intact protein (mAb) analysis were performed on a HPLC-UV and TripleTOF 6600 mass spectrometer

(Sciex).

Methods

The recoveries of intact mAbs was compared on the two particle packed columns (silica gel type and polymer

gel type) and developed columns, it appeared that adsorption was less pronounced on the silica gel particle

packed, probably due to the size of the small pore size. Developed monolithic column was shown the excellent

recoveries to intact mAb. The pressure drop of the developed column is typically low that of a column packed

with polymer column.

A mAb require buffering salt for stabilization in solution. However, to gain sensitivity in LC-MS analysis of

intact mAbs, it is critically important to remove the salt before LC-MS analysis. Figures show the LC/MS

data for the intact mAbs with developed column and silica gel columns, respectively. The developed

column provided excellent total ion chromatogram peak shapes compared with silica gel column.

The temperature effect for resolution was compared. The heat resistance of a monolithic column clad with

resin was typically up to 50 °C, but clad the monolithic material with glass improved the column‘s heat

resistance to 80℃. Recovery of mAb were significantly improved with elevated temperature, and adsorption

was generally acceptable in the temperature range between 60 and 80℃. As temperature increases, peak

widths of the protein peaks has decreased and absolute retention of each peak was reduced with increasing

temperature. The Stability of the developed column at higher temperature and presence of trifluoroacetic acid

has also improved by cladded with a glass tube.

Results & Discussion

1 µm Through pore

60 nm Meso pore

Φ2.1mm Monolithic silica

Stainless steel tube

Glass tube

Assembly

Schematic of the developed monolithic silica column

Monolithic silica disk

SPE Spin columnAbsorbent material Capillary HPLC column

Monolithic silica application fields

0.0 1.0 2.0 3.0 4.0 5.0 6.0Time (min)

0.0

00.1

0V

olts

0.0 1.0 2.0 3.0 4.0 5.0 6.0Time (min)

0.0

00

.10

Vo

lts

0.0 1.0 2.0 3.0 4.0 5.0 6.0Time (min)

0.0

00.1

0V

olts

Developed column Silica gel

packed column

Polymer packed

column

4.5MPa 10.5MPaPressure drop 1.7MPa

Peak height 135451 25136 113250

ConditionsEluent: A:0.1 %TFA-H2O

B:0.1 %TFA-ACN

Gradient: A/B A/B=90/10-2.5min-10/90

Flow Rate: 0.3 mL/min

Oven Temp.: 80 ℃Injection Vol.: 5 μL R

ete

ntio

n tim

e[m

in]

Injection [times]

Developed monolithic

silica clumn

Conditions

Eluent: A) 0.075% HCOOH+0.025% TFA - H2O

B) 0.075% HCOOH+0.025% TFA - ACNA/B=95/5-5min-10/90

Flow Rate: 0.3 mL/min

Detection: UV 210 nm

Oven.Temp: 80 ℃Analyte: 1.4 mg/mL mAb

Column performance of the developed column with competition.

2.4

2.6

2.8

3

3.2

3.4

3.6

3.8

4

0 500 1000 1500 2000 2500 30002.2

2.4

2.6

2.8

3

3.2

3.4

3.6

3.8

0 500 1000 1500 2000 2500 3000

Re

ten

tio

n tim

e[m

in]

Injection [times]

0.0 1.0 2.0 3.0 4.0 5.0 6.0

Time (min)

0.0

0.2

0.4

Vo

lts

50 ℃

60 ℃

80 ℃

0.0 1.0 2.0 3.0 4.0 5.0 6.0

Time (min)

0.0

0.2

0.4

Vo

lts

0.0 1.0 2.0 3.0 4.0 5.0 6.0

Time (min)

0.0

0.2

0.4

Vo

lts

0.0 1.0 2.0 3.0 4.0 5.0 6.0

Time (min)

0.0

0.2

0.4

0.6

Vo

lts

40 ℃

Temperature effect

Peak tailing

Durability

The graph below shows the durability test. In response to the increase of injection times (x-axis), the

durability of each column is plotted at y-axis with the retention time of the mAb. The developed column

modified phenyl groups shows great durability under low pH and high temperature conditions. The lifetime of

phenyl groups modified column is much greater than that of the C4 modified columns. This assures stable

analysis over a long period of time.

LC/MS intact mAb analysis

>10 μm <1 μm

The pores can be sized according to application, resulting in a highly adaptive material for chromatography.

Developed columnSilica gel

packed column

0.0 1.0 2.0

Time (min)

0200

400

600

800

1000

mV

olt

0.0 1.0 2.0

Time (min)

0200

400

600

800

mV

olt

① ②

Conditions

Eluent: A)0.08%HCOOH-0.02%TFA-H2O

B)0.08%HCOOH-0.02%TFA-ACN

A/B=80/20-2.5 min-10/90

Fab in CHO cell culture Purified sample

Flow rate: 0.3 mL/min

Col.Temp.: 60 ℃Detection: 210 nm

Injection Vol.: 1 μL

M

Electrophoresis

① ; Purified with MonoSpin ProL

② ; CHO cell culture

M ; Marker

Fab

CHO cell culture direct analysis

Purified by

ProteinL immobilized

Spin column

Conclusions

The developed monolith column has high porosity, it was possible to analyze the culture broth without clogging.

C4 group Phenyl group

Separation of Intact mAb using a developed monolithic column at different temperatures.

Analyte: 1.4mg/mL mAb

5 uL injection

CladdingConditionsEluent: A:0.1 %HCOOH-H2O

B:0.1 %HCOOH-ACN

Gradient: A/B=90/10-10min-10/90

Flow Rate: 0.3 mL/min

Detection: TripleTOF 6600

Oven Temp.: 80 ℃Injection Vol.: 5μL

Intact mAb mass analysis (total ion chromatogram) on (A) an developed column, (B) silica packed column.

(A) (B)

・ The developed column was shown to be very suitable for analysis of intact protein (mAb).

・ The Novel glass cladded monolithic silica column was developed.

Fab