Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Embed Size (px)

Citation preview

Page 1: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Mass Spectrometry

시스템생물학과이선아

Page 2: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Mass Spectrometry • Widely used analytical technique• Within an accuracy of 0.01% of total weight of sample and within 5 ppm

for small organic molecules• Unequaled sensitivity –Nanomolar range routinely (1 x 10-9M) –Femtomolar range possible (1 x 10-15M) –Attomolar range claimed (1 x 10-18M)• Diversity of applications –Proteins –Oligonucleotides –Oligosaccharides –Lipids –Others• Proteomics –Identification of proteins expressed under specific conditions

Page 3: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

-3 fundamental parts: Ionization source, the ana-lyzer, the detector

-Ionization source

시료분자를 이온화시키고 더 작은 이온으로 쪼갠다 .

-Mass analyzer

이온들을 mass-to-charge(m/z) ratio에 따라 선택적으로 분리

-Ion detector

이온 흐름을 그 양에 비례하게 전기적인 흐름으로 전환 , 증폭시켜 signal을 생성

-Vacuum system

Page 4: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Basic components to MS•Ion source –Electrospray(ESI) –Atmospheric Pressure Ionization (APCI) –Chemical Ionization (CI) –Electronic Ionization (EI) –Matrix Assisted Laser DesorptionIonization (MALDI)

•Mass Selection –Quadrupole –Time of Flight (TOF) –Magnetic Sector –Ion Trap –Ion Cyclotron

•Detector –Phosphor / Photo Diode –Multi-channel Plate (MCP)

Page 5: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Ion Source: ESI

Electrospray ionization(ESI)- 용액 상태의 시료를 이온화 (LC-MS)- 기존의 방법으로는 얻기 힘들었던 intact

상태의 peptide 나 단백질을 이온화- 한 개 이상의 전하를 띤 이온을 생성

Page 6: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Ion Source: ESI

Page 7: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Ion Source: ESI

Page 8: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Ion Source: MALDI

Matrix Assisted Laser Desorption Ion-ization(MALDI)

Page 9: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

hn

Laser

+20 kV

Variable Ground Grid Grid

AH+

Sample plate 1. Sample (A) is mixed with excess matrix (M) and dried on a MALDI plate.

2. Laser flash ionizes matrix molecules.

3. Sample molecules are ionized by proton transfer from matrix:

MH+ + A M + AH+.

Page 10: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Why MALDI?

-Less sensitive to salts-Lower PRACTICAL detection limits-Easier to interpret spectra(less multi-

ple charges)-Quick and easy-Higher mass detection-Higher Throughput(1000>samples per

hour)

Page 11: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

MALDI/TOF Mass spec-trum

Re

lativ

e A

bu

nda

nce

m/z

0

10000

20000

30000

40000

50000 100000 150000 200000

(M+H)+

(M+2H)2+

(M+3H)3+

Page 12: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

The Mass Analyzer: TOFTime Of Flight(TOF)

Flight TubeIon Source

Principle: If ions are accelerated with the same potential at a fixed point and a fixed initial time and are allowed to drift, the ions will separate according to their mass to charge ratios.

20-25 kV

++

Page 13: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

The Mass Analyzer: TOF

Flight Tube

Detector

Ion Source

+

+ +

Page 14: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

The Mass Analyzer: TOF

Flight Tube

Detector

Ion Source

+

++

Page 15: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

The Mass Analyzer: Quadrupole

Quadrupole(Mass filter)-4 개의 molybdenum 막대로 이루어져 있으며 ,

한쌍 (1,2) 은 DC voltage, 다른 한쌍 (3,4) 은 Radio frequency voltage 가 가해진다 .

- 가해지는 전압의 진폭은 선택된 m/z 에 해당되는 ion 만 ion source 에서 detector 까지 통과하게 한다 .

- quadropole 의 전압을 바꾸면서 주어진 mass범위의 이온을 scanning 한다 .

Page 16: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

The Mass Analyzer: Quadrupole

Page 17: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Detectors

Page 18: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

+

-1000 V -100 V

Primary Ion from Flight Tube

L

D= 6-25 u

Ions are detected with a Microchannel Plate

Page 19: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

+

-1000 V -100 V

L

D= 6-25 u

Ions are detected with a Microchannel Plate

Page 20: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

+

-1000 V -100 V

e-

L

D= 6-25 u

Ions are detected with a Microchannel Plate

Multification by secondary emission

Secondary emissive materials: Beryllium oxide, magnesium oxide etc

Page 21: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

+

-1000 V -100 V

e-e- e-

e-

L

D= 6-25 u

Ions are detected with a Microchannel Plate

Page 22: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

+

-1000 V -100 V

e-e- e-

e-

L

D= 6-25 u

Ions are detected with a Microchannel Plate

Page 23: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

+

-1000 V -100 V

e-e- e-

e-

L

D= 6-25 u

e-

e-

e-

e-

e-

e-e-

e-

~103

Amplification

Ions are detected with a Microchannel Plate

Page 24: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Tandem MS(MS/MS)

Page 25: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Tandem MS(MS/MS)

Page 26: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Tandem MS(MS/MS)

Page 27: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Laser

Sampleplate

Analytemolecules in matrix

Accelerationgrids

Drift tube Ion detector

Mass spectrum

Vacuum system

Vacuumlock

MALDI TOF MS

Page 28: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

HiRes mass spectrum

Iondetector

MALDI ionsource

Ion reflector

The reflector focuses ion of same mass but different velocity on

detector; high resolution is obtained

Laser

High resolution TOF-MS with ion reflector

Page 29: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Daughter ion mass spectrum

Iondetector

Ion reflector

MALDI ionsource

CID

MS/MS spectrum of daughter ionsis measured in a single acquisition;

no pasting of segments;low sample consumption,

high speed, high sensitivity

LID

Laser

TOF/TOF-MS/MS with

Parent ionselector

Page 30: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Why interested in MALDI-TOF MS 분자량 측정

큰분자량 물질 분석

혼합물 분석 : 한 종류의 성분이 아닌 몇 종류가 혼재해 있어도 분석이 가능함

미량분석 : 매우 민감하여 미량의 시료도 분석 가능 함 : 펩타이드 경우 fmol 분석 가능

데이터 분석이 쉬움 : 분자 구조가 깨어 지지 않고 , 보통 다 전하를 (multiple charging) 띠지

않으므로 데이터가 다른 질량 분석기에서 보다 단순하여 해석이 용이함

염의 영향이 적음 : 생체단백질 분리에 이용되는 완충용액 , 염 등에 LC/MS 보다 영향을 덜 받음

분석이 신속함 : 시료와 Matrix 섞어 sample plate 에 떨어뜨려 용액을 말리는 시간 ( 약 5~10

분 ), MALDI-TOF 로 분석하는 시간 (1 분 이내 )

기기 사용 및 유지하기 위한 비용이 저렴 : LC/MS, GC/MS 처럼 질소 또는 아르곤 가스를 사용하지 않

고 , 미량의 Matrix 와 ACN, TFA 등을 이용함으로 시약 비용도 저렴함

Page 31: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Mass Spectrometry 분석-base peak-parent peak-radical cations-Isotopes

Page 32: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm
Page 33: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm
Page 34: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Biomolecule Analysis

* 과거에는 ?-Electrophoresis, chromatography, ul-

tracentrifugation-Not very precise*MS 이용하면 ?-Proteins, oligonucleotides, oligosac-

charides, lipids-Detect modifications and sequences

Page 35: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Post Translational Modifica-tions(PTM’s)

• PTM’s are very important in signaling as well as metabolic pathways (e.g. phosphorylation)

• Often we want to know not only which modi-fication a protein has undergone, but exactly where in the sequence the modification lies.

• Many of the search engines allow for “vari-able” modifications, but very few at one time (combinatorialy explosive)

• There is great opportunity here for robust searches that find PTM’s reliably!

Page 36: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Protein sequence Analy-sis

Page 37: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm

Peptide Sequencing

Page 38: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm
Page 39: Mass Spectrometry 시스템생물학과 이선아. Mass Spectrometry Widely used analytical technique Within an accuracy of 0.01% of total weight of sample and within 5 ppm