63
M.Prasad Naidu MSc Medical Biochemistry, Ph.D.Research Scholar

HPLC instrument

Embed Size (px)

DESCRIPTION

instrument

Citation preview

Page 1: HPLC instrument

M.Prasad Naidu

MSc Medical Biochemistry,

Ph.D.Research Scholar

Page 2: HPLC instrument

Chromatography is a physical process whereby components ( solutes ) of a sample mixture are separated by their differential distribution between stationary & mobile phases .

Planar & column are two basic forms of chromatography .

High performance liquid chromatography is a form of column chromatography .

Page 3: HPLC instrument

During column chromatography process mobile phase carries the sample through the column containing stationary phase .

As the mobile phase flows through the stationary phase the solutes may

1) Reside only on stationary phase ( no migration ) ,

2) Reside only in the mobile phase ( migration with mobile phase ) ,

3) Distribute between two phases ( differential migration ) .

Page 4: HPLC instrument

The basis of all forms of chromatography is partition or distribution coefficient ( Kd ) .

Kd describes the way the solute distribute it self between two immiscible phases .

Distribution coefficient is a constant at a given temperature for two immiscible phases A & B .

concentration in phase A Kd = concentration in phase B

Page 5: HPLC instrument
Page 6: HPLC instrument

In column chromatography , the stationary phase may be pure silica or polymer , or it may be coated onto , or chemically bonded to, support particles .

The stationary phase may be coated into a tube , or it is coated on inner surface of the tube .

When the mobile phase is liquid it is called liquid chromatography ( LC ) .

When the stationary phase in LC consists of smaller diameter particles the technique is high performance liquid chromatography .

Page 7: HPLC instrument

In analytical liquid chromatography the mobile phase or eluent , exits from the column & passes through a detector or a series of detectors that produce a series of electronic signals that are plotted as a function of time distance or volume , the resulting graph is a chromatogram .

The retention time ( tR ) is the time taken for each analyte peak to emerge from the column .

Page 8: HPLC instrument

Under defined chromatographic conditions tR is a charcteristic of the analyte .

The volume of the mobile phase required to elute the analyte under defined chromatographic conditions is referred to as retention ( or ) elution volume ( VR ) .

VR = tR Fc

Page 9: HPLC instrument

Eluting solutes are displayed graphically as a series of peaks , they are frequently referred to as chromatographic peaks .

These peaks are described in terms of peak width , peak height & peak area .

The data represented by the chromatogram are used to help identify & quantify the solutes .

Page 10: HPLC instrument

Most important parameter in column chromatography is the partition ratio ( or ) capacity ratio K’ .

Capacity ratio has no units & it is a measure of the additional time the analyte takes to elute from the column relative to an unretained or excluded analyte that does not partition into stationary phase .

Page 11: HPLC instrument

K’ = tR – tM = VR – VM tM VM Capacity ratios characterize the column

performance . The success of any chromatographic procedure

is measured by it’s ability to separate completely ( resolve ) one analyte from a mixture of similar compounds .

Peak resolution ( Rs )is related the properties of the peaks .

Page 12: HPLC instrument
Page 13: HPLC instrument

Rs = 2 ( tRB – tRA ) WA + WB tRA & tRB are the retention times of compounds

A & B respectively , & WA & WB are base widths of peaks for A & B , respectively .

When Rs = 1.5 the separation of the two peaks is 99.7 % complete .

In most practical cases Rs value of 1.0 corresponds to 98 % of separation , are adequate for quantitative analysis .

Page 14: HPLC instrument

Peak asymmetry has many causes ,

1) Application of too much analyte to the column ,

2) Poor packing of the column , 3) Poor application of the sample to the column

or solute support interactions .

Page 15: HPLC instrument

Chromatography columns consists of number of adjacent zones each zone is called theoretical plate & its length in the column is called plate height .

The more efficient the column the greater the number of theoretical plates are involved .

N = 16 ( tR/W )2

Page 16: HPLC instrument

The plate number can be increased by increasing the column length, but there is a limit to this because the retention time & peak width increases proportionally L , where as the peak height decreases as the square root of N .

Page 17: HPLC instrument
Page 18: HPLC instrument

Good resolution is determined by the following 3 functions :

1) Selectivity , 2) Efficiency ,3) Capacity . Selectivity is a measure of inherent ability

of the system to discriminate between structurally related compounds .

Two structurally related compounds differ in Kd or K’ .

Ratio of partition coefficient of two compounds gives relative retention ratio ,α .

Page 19: HPLC instrument

Efficiency is the measure of diffusion effects that occur in the column to cause peak broadening & over lap .

Capacity is a measure of the amount of material that can be resolved without causing peaks to overlap irrespective of actions like gradient elution .

Page 20: HPLC instrument

The limit to the length of the column is due the problem of peak broadening .

The number of theoretical plates is related to the surface area of the stationary phase therefore smaller the particle size of the stationary phase , the better is the resolution.

The Smaller the paritcle size , the greater is the resistance to flow of the mobile phase .

Page 21: HPLC instrument

The resistance in flow causes back pressure in the column that is sufficient to damage the matrix structure of the stationary phase .

The new smaller particle size stationary phases that can withstand high pressures caused dramatic development in the column chromatography .

Page 22: HPLC instrument

The increased resolution achieved in HPLC compared to classical chromatography is primarily the result of adsorbents of very small particle size ( less then 20µm )& large surface areas .

The smallest gel beads used in gel exclusion chromatography are superfine grade with diameters of 20-50µm .

A combination of high pressure & adsorbents of smaller size leads to high resolution power & short analysis time in HPLC .

Page 23: HPLC instrument

(1) Solvent reservoirs, (2) Solvent degasser, (3) Gradient valve, (4) Mixing vessel for delivery of the mobile phase, (5) High-pressure pump, (6) Switching valve in "inject position", (6') Switching valve in "load position", (7) Sample injection loop, (8) Pre-column (guard column), (9) Analytical column, (10) Detector (i.e. IR, UV), (11) Data acquisition, (12) Waste or fraction collector.

Page 24: HPLC instrument

Solvent reservoir should have a capacity of at least 500 ml for analytical applications , but larger reservoirs are required for preparative work .

In order to avoid the bubbles in the column & detector the solvent must be degassed .

Several methods are there for degassing : 1) By warming the solvent ,2) By vigorous stirring with magnetic stirrer ,3) By ultrasonication ,4) By subjecting solvent to vacuum or by

bubbling helium gas through the solvent reservoir .

Page 25: HPLC instrument

Typical requirements for a pump are :1 ) it must be capable of pressure outputs of at

least 500 psi & preferably up to 5000 psi . The main feature of good pumping system is

that it can capable of outputs of at least 5x107 pascals ( 7200 psi ) .

2) Pump should have a controled , reproducible flow delivery of about 1ml/min for anlytical applications & up to 100ml/min for preparative applications .

3 ) it should yield pulse free solvent flow 4) It should have a small hold up volume .

Page 26: HPLC instrument

The correct application of the sample on to the HPLC column is particularly important factor in achieving successful separations .

Two injection methods are existing First method makes use of a microsyringe to

inject the sample either directly on to the column packing or onto a small plug of inert material immediately above the column packing .

The second method of sample injection retains the column pressure by use of a loop injector .

Page 27: HPLC instrument

Metal loop has as fixed small volume that can be filled with sample .

By means of an appropriate valve switching system , the eluent from the pump is channelled through the loop , the outlet of the loop leads directly onto the column .

Therefore sample is flushed on to the column by eluent without interruption of flow to the column .

Page 28: HPLC instrument
Page 29: HPLC instrument

Repeated application of highly impure samples such as sera , urine , plasma or whole blood are preferably deproteinated because they decrease the resolving power of the column .

To prevent the above problem a guard column is frequently installed between the injector & the analytical column .

Page 30: HPLC instrument

Guard column is a short column of the same internal diameter & packed with material similar to analytical column .

The packing in the guard column retains contaminating material & can be replaced at regular intervals .

Page 31: HPLC instrument

Sample preparation is essential preliminary action in HPLC , particularly for the test compounds in a complex matrix such as plasma , urine , cell homogenate .

For analysis of drugs in biological fluids sample preparation is relatively much simpler.

Sample preparation is done by clean up techniques they are :Solvent extraction ,Solid phase extraction ,Column switching & newer supercritical fluid extraction ( under research )Derivatization .

Page 32: HPLC instrument

For HPLC analysis many analytes are chemically derivatized before or after chromatographic separation to increase their ability to be detected .

Eluted amino acids are reacted with ninhydrin in post column reactor , the resulting chromogenic species are detected by photometer .

Page 33: HPLC instrument

Aliphatic amino acids , carbohydrates , lipids & other substances do not absorb UV can be detected by chemical derivatization with UV absorbing functional groups .

Precolumn derivitization for amino acids & peptides is by phenyl isothiocyanate , dansyl chloride for UV column detection .

Precolumn derivatization for fatty acids , phospholipids is by phenacyl bromide for UV column detection .

Post column derivatization for carbohydrates is by orsinol & sulphuric acid for UV column detection

Page 34: HPLC instrument

Column is made up of stainless steel . Column has to withstand pressures of up to 5.5

X 107 pascal . Straight columns of 15 – 50 cm length & 1 –

4mm diameter & has flow rate of 2 cm3/ min.

Preparative columns have an internal diameter of 25 mm & has flow rate of 100 cm3 / min.

Page 35: HPLC instrument

Three form of column packing matrices are available they are :

1) Microporous supports : ( micropores ramify through the particles which are generally 5 – 10 µm in diameter ),

2) Pellicular ( superficially porous ) supports : in which porous particles are coated on to an inert solid core such as a glass bead of 40 µm in diameter ,

3) Bonded phases : in which stationary phase is chemically bonded to an inert support such as silica .

Page 36: HPLC instrument

For adsorption chromatography , adsorbents such as silica & alumina are available as microporous or pellicular forms which are suitable for HPLC .

Pellicular forms have high efficiency but low sample capacity therefore microporous supports are preferred .

For partition chromatography bonded phases are used .

Page 37: HPLC instrument

In normal phase liquid chromatography the stationary phase is a polar compound such as alkylnitrile or alkylamine & the mobile phase is a nonpolar solvent such as hexane .

For reversed phase liquid chromatography stationary phase is a nonpolar compond such as octasilane (OS) or octadecylsilane (ODS), & the mobile phase is a polar solvent such a water / acetonitrile or water / methanol.

Page 38: HPLC instrument

Cross linked microporous polystyrene resins are widely used suitable ion exchange resins for HPLC .

Stationary phase for exclusion separations are porous silica , glass , polystyrene or polyvinylacetate beads & are available in a range of pore size .

Page 39: HPLC instrument

The support for affinity separation are similar to those for exclusion separations .

The spacer arm & ligand are attached to the supports by chemical bonding .

Chiral stationary phases contain proteins that are composed of amino acids each of which has a stereocenter ( except glycine ) commonly used are alfa 1 acid glycoproteins ( AGP ) ,human serum albumin ( HAS ) .

Semirigid as well as nonrigid gels have limitted role in HPLC stationary phase .

Page 40: HPLC instrument

The major priority in packing of a column is to obtain a uniform bed of material with no cracks or channels .

Rigid solids as well as hard gels should be packed as densely as possible but without fracturing the packing process .

Most widely used technique for column packing is the high pressure slurrying technique .

Page 41: HPLC instrument

The choice of mobile phase to be used in any separation will depend on the type of separation to be achieved .

Eluting power of the solvent is related to its polarity.

The components of the applied sample are separated by the continuous passage of the mobile phase through the column , this is known as elution development .

Page 42: HPLC instrument

Column development is of 2 types :1)Isocratic elution ,2)Gradient elution . Column development using a single liquid as the

mobile phase is known as an isocratic elution . In order to increase the resolving power of the

mobile phase , it is necessary continuously to change it’s pH , ionic concentration or polarity this is known as gradient elution .

Page 43: HPLC instrument

In order to produce a suitable gradient , two eluents have to be mixed in the correct proportions prior to their entering the column.

Gradient elution uses separate pumps to deliver two solvents in proportions predetermined by a gradient programmer .

All solvents for use in HPLC systems must be specially purified because traces of impurities can affect the column & interfere the detection system especially when measuring absorbance below 200nm .

Page 44: HPLC instrument

Purified solvents are available commercially , but even with these solvents 1 – 5 µm microfilter is generally introduced into the system prior to the pump .

All solvents are degassed before use .

Gassing can alter column resolution & interfere with continuous monitoring of the effluent .

Page 45: HPLC instrument
Page 46: HPLC instrument

The purpose of the pump is to provide a constant , reproducible flow of solvent through the column .

Two types of pumps are available :

1) Constant pressure pump ,2) Constant volume pump .

Page 47: HPLC instrument

Constant pressure pumps produce a pulseless flow through the column , but any decrease in the permeability of the column will result in lower flow rates for which the pumps will not compensate .

Constant pressure pumps are seldom used in contemporary liquid chromatography .

Constant displacement pumps maintain a constant flow rate through the column irrespective of changes within the column .

Page 48: HPLC instrument

Two types of constant displacement pumps are available :

1) Motor driven syringe type pump ,2) Reciprocating pump ( most commonly used

form of constant displacement pump ) . All constant displacement pumps have in built

safety cut off mechanisms , so that if the pressure within the chromatographic systems changes from preset limits the pump is inactivated automatically .

Page 49: HPLC instrument

The sensitivity of the detector system must be high & stable to respond to the low concentrations of each analyte in the effluent.

Most commonly the detector is a variable wave length detector based upon UV – visible spectrophotometry since few compounds are colored visible detectors are of limited value .

Detector is capable of measuring absorbance units down to 190 nm wave length & has sensitivities as low as 0.001 absorbance units for full – scale deflection ( AUFS ) .

Page 50: HPLC instrument

Variable wave length detector operates at a wave length selected from a given wave length range .

Thus the detector is tuned to operate at the absorbance maximum for a given analyte or set of analytes which enhances greatly the applicability & selectivity of the detector.

Acetonitrile & methanol two widely used solvents in reversed phase chromatography have minimum UV absorption at 200nm .

Page 51: HPLC instrument

Most biomolecules like proteins , nucleic acids, vitamins , steroids , pigments & aromatic amino acids absorb strongly in 220 – 365 nm range .

Aliphatic amino acids , carbohydrates , lipids & other substances do not absorb UV can be detected by chemical derivatization with UV absorbing functional groups .

Page 52: HPLC instrument

UV detectors have many positive characteristics : highly sensitive ,

small sample volumes , linearity over wide range concentrations , non destuctiveness to

sample & suitability for gradient elution.

Page 53: HPLC instrument

Fluorescence detectors are extremely valuable for HPLC because of their sensitivity but the technique is limited by the fact that relatively few compounds fluoresce .

Electrochemical detectors are extremely sensitive for electro active species .

The sensitivity of UV absorption , fluorescence & electrochemical detection can be increased significantly by the process of derivatisation , where by the analyte is converted pre or post column to a chemical derivative .

Page 54: HPLC instrument

Diode arrays are used as HPLC detectors because they rapidly yeild spectral data over the entire wave length range of 190 – 600 nm in about 10 milliseconds .

Incorporation of computer technology into HPLC has resulted in cost effective , easy to operate automated systems with improved analytical performance .

Page 55: HPLC instrument

The area or height of each chromatographic peak is determined from the stored data in computer & used to compute the analyte concentration represented by each peak .

Fast protein liquid chromatography :this provides a link between classical column chromatography ,& HPLC .

FPLC uses experimental conditions intermediate those of column chromatography & HPLC .

Page 56: HPLC instrument
Page 57: HPLC instrument
Page 58: HPLC instrument

Narrow-bore columns (1-2 mm) are used for in this application .

Liquid chromatography-mass spectrometry (LC-MS, or alternatively HPLC-MS) is an analytical chemistry technique that combines the physical separation capabilities of liquid chromatography (or HPLC) with the mass analysis capabilities of mass spectrometry.

Page 59: HPLC instrument

HPLC has had big impact on separation of oligopeptides & proteins .

FPLC a modified version useful in separation of proteins .

HPLC coupled with electrochemical detector is useful in assay of catecholamines ,vitamins (AD&E ,niacin , thiamine) & antioxidants .

HPLC has role in quantification of various hemoglobins in hemoglobinopathies .

HPLC coupled with MS is useful in measuring cortisol in blood & saliva .

Page 60: HPLC instrument

HPLC is useful in cytokine measurement . Useful in assay of HbA1c . Useful in assay of fructosamine . 5 – hydroxy idole acetic acid & serotonin can be

assayed. The pharmaceutical industry regularly employs

Reverse Phase HPLC to qualify drugs before their release.

Assay of plasma & urinary catecholamines , plasma & urinary metanephrines

Page 61: HPLC instrument

For diagnosis of different porphyrias .

Thyroxine , uric acid .

Nucleic acid analysis, oliginucleotides , steroids , amino acids , serotonin , measurement of isoenzymes .

Page 62: HPLC instrument

(1) Solvent reservoirs, (2) Solvent degasser, (3) Gradient valve, (4) Mixing vessel for delivery of the mobile phase, (5) High-pressure pump, (6) Switching valve in "inject position", (6') Switching valve in "load position", (7) Sample injection loop, (8) Pre-column (guard column), (9) Analytical column, (10) Detector (i.e. IR, UV), (11) Data acquisition, (12) Waste or fraction collector.

Page 63: HPLC instrument