Light Measuring
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
2/9
32
QUALITYPRECISION
GOSSEN Foto- und Lichtmesstechnik Your Guarantee or Precision and QualityGOSSEN Foto- und Lichtmesstechnik is specialized in the measurement o light, and has decades o experience in its chosen feld. Continuous innovation isthe answer to rapidly changing technologies, norms and markets. The outstanding quality o our products is assured by means o production in Germany and a
certifed quality management system in accordance with ISO 9001.
Illuminance (abbreviation:E, unit of measure:lux)indicates with how much intensity an area is illuminated. It amounts to one
lux when a luminous ux o one lumen illuminates a surace o one square
meter. This corresponds roughly to a normal candle ame at a distance o
one meter.
Luxmeters are used to measure illuminance at horizontal and vertical
suraces. However, illuminance does not indicate the brightness impression
o a room, because this depends to a great extent on the rooms reective
characteristics. A white room gives a much brighter impression than a dark
room.
With normal lighting, uniorm light distribution is not achieved as a rule,
or which reason specifcations in the standards usually make reerence to
average illuminance. This value is calculated as the weighted arithmetic
mean o all illuminance values in the room.
ApplicationsGOSSENs precision luxmeters are used or planning, installation, testing and
monitoring o lighting systems, as well as or adhering to lighting conditions
or hygienic, physiological, psychological and/or saety reasons. Scope o
validity, terminology, responsibilities, requirements and practical guidelines
are set orth to a great extent in the standards.
ApplicationsGOSSENs precision luminance meters are used or distance and contact
measurements or all types o light sources. For the purpose o adhering
to applicable norms, this allows or testing o minimum and maximum lu-
minosity, assurance o quality requirements specifed or products with
displays or lamps, ascertainment o maintenance requirements due to aging,
and optimization o illumination uniormity. Scope o validity, terminology,
responsibilities, requirements and practical guidelines are set orth to a great
extent in the standards.
Luminance(abbreviation:L, unit of measure:cd per m)indicates the brightness impression perceived by the eye when positioned in
ront o a back-lighted or illuminated surace. It describes the physiological
eect o light on the eye, and is used as a planning actor or outdoor lighting.
Reliable measurement results with defned error limits are guaran-
teed by luminance meters and luxmeters classifed in accordance with
DIN 5032-7 and DIN EN 13032-1. This ensures that quality inspections,
reerence measurements and assessments conducted with these meters
deliver accurate statements.
Calibration certifcates and recalibration at regular i ntervals are required or
use in quality-relevant applications and or assessments. The GOSSEN Light
Lab oers these services, even or products rom other manuacturers, andissues actory calibration certifcates. The optical bench used or this purpose
is subject to strict t est equipment monitoring, and is retraceable to the PTB
in Germany (Highest technical authority under the auspices o the Federal
Ministry o Economics and Technology). Aside rom the PTB, our lab is
the frst in Germany to be accredited or illuminance by DAkkS (German
accreditation authority), and is thus authorized to issue internationally
recognized DAkkS calibration certifcates. This assures that acquired
measured values comply with ofcial regulations and norms, which also
stand up to legal argumentation.
Longstanding customers rom industry, the public authorities and t he feldo medical engineering hold the products and services oered by GOSSEN
Foto- und Lichtmesstechnik in high esteem, are pleased to receive our expert
advice, and ask us to implement their special requirements.
Acceptance and constancy testing or image display devices in the feld
o medical engineering.
Contrast measurements at workstations (work saety directives)
Measurement o lighting or streets, tunnels, railways and airports, as well
as signal systems
Luminance measurements or CRTs, LCDs, LEDs and plasma displays.
Lighting in museums and public buildings
Lighting or sports acilities
Testing or consistent lighting o projection screens
Measurement o lighting installations, lightboxes and outdoor advertising
Monitoring o workstations and public buildings, as well as compliance
with regulations or working places, trade association directives and
ofcial norms
Measurement o emergency lighting
Repair and maintenance in production acilities, ofces and hospitals
Inspection o light sources or street lighting
Monitoring o sports acilities and parks
Quality assurance or manuacturers o illuminants and lamps
Planning o lighting eects by light designers and architects
Adherence to illuminance requirements in the felds o
arming and orestry.
Research and development at light engineering acilities
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
3/9
54
SpecifcationsMaximum precision Classifed measurement o illuminance in lx or c in
accordance with class B or class C per DIN 5032-7, appendix B o IEC
13032-1 and CIE 69.
Broad measuring range High initial sensitivity and a resolution o 0.01 lx
or 0.001 c or the MAVOLUX 5032 B USB, right on up to large illuminance
values o 199,900 lx or 19,990 c.
V() adaptation The spectral sensitivity o the silicon photodiode is
color corrected and corresponds to the spectral brightness sensitivity o the
human eye V(). The quality o this adaptation represents a signifcant
dierence between the class B and class C variants.
Cosine correction The luminosity o a at measuring surace is proportional
to the cosine o the incident angle o light. This relationship is taken into
consideration by the sensor during evaluation.
Simple expansion o unctions Unclassifed measurement o luminance
in cd/m2 or L is made possible by the optional luminance attachment.
An additional adapter disc prevents erroneous measurements due to
incidence o light rom the side.
Convenient everyday use Simple operation, easy to read display, compact
design, protection during transport in a high quality aluminum case.
Non-volatile memory 100 measured values can be saved, retrieved or
transmitted to a PC via the integrated USB port. Average illuminance values
can also be calculated as a result.
Computer-aided measurement The meters continuous operating mode is
assured thanks to power supply via the USB port. Meter control, as well as
reading, display and storage o measured values, is managed with GLUX 2
sotware, included in the delivery contents.
MAVOLUX 5032 B/C USB
Due to its outstanding accuracy in accordance with class B, the MAVOLUX
5032 B USB is used primarily or certifcation and inspection applications.
An additional measuring range with high initial sensitivity o 0.01 lx makes
it possible to measure extremely small light intensities. This even permits
reliable measurement o emergency lighting. Ater pressing the HOLD key,
the measured value is saved to memory and display illumination included
with this version is activated, making it possible to read the display in dark
environments. Adaptation to the spectral brightness sensitivity o the humaneye V() is highly precise with minimal deviation o just 1 < 3 %.
With accuracy according to class C, the industrial luxmeter Mavolux 5032 C
USB is primarily used or general applications.The smallest o our measuring
ranges begins with an initial sensitivity o 0.1 lx. V() adaptation deviation,
which amounts to 1 < 7.5 %, is considerably better than the admissible
error limit or class C.
Both variants can be used as unclassifed luminance meters with the help
o an optional luminance attachment with an acceptance angle o 15.
Luminance is measured in candelas per square meter (cd/m) or oot-
lamberts (L), and defnes the perceived brightness o a back-lighted or
reecting surace.
At GOSSEN, great emphasis is placed upon reliability by means ocalibration. For purposes o documented evidence, a actory calibration
certifcate or a DAkkS calibration certifcate can be ordered along with either
variant. Depending on how the meter is used, we recommend a calibration
interval o 12 to 24 months.
This high precision luxmeteris available in class B and class C versions in accordance with DIN 5032-7, appendix B o IEC 13032-1 and CIE 69. Both variants are equipped with V( )adaptation as well as cosine correction, and reliably measure the illuminance o daylight and artifcial sources o light. Even in the case o very bright sunlight or
light rom headlights, no accessories are required.
5
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
4/9
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
5/9
98
MAVO-SPOT 2 USBIs equipped with an excellent adaptation to the spectral brightness sensiti-
vity o the human eye V() and is highly precise with a minimal deviation o
just 1 < 3 %, which is signifcantly better than specifed in the standard.
By means o the mirror reex optics, the measuring subject may be targeted
precisely in a 15 feld o view and a sharply marked 1 measuring circle in
the center o the viewfnder. Focus can be set or distances ranging rom 1
meter to infnity. Shorter distances as o 34 cm are made possible by meanso optional close-up lenses. Alternatively, contact measurement can also be
perormed with an optional, top quality probe. The velvety coating on the
adapter disc prevents scratching o sel-luminous and back-lighted suraces.
Measurement unctions are selected and measurements are started with
convenient on-hand operation. The display on the viewfnder is activated
along with back-light by simply pressing a key. The comparative and ratio
measurements are especially advantageous, by means o which deviation
o measured value B rom reerence value A is evaluated and displayed.
The relationship A/B is used or contrast measurement at workstations.
Percentage deviation %A allows or an assessment o the consistency o
monitor screens and projection screen lighting, and dierence A-B is used to
detect deviation during the production process.Special tests or proving and observing norms and standards or quality,
general saety, work protection, as well as saety on monitors in the feld o
medical diagnostics and ofce technology may get its documented evidence
by means o our actory Calibration Certifcate. Depending on how the meter
is used, we recommend a calibration interval o 12 to 24 months.
This high precision luminance meteror distance measurement with an acceptance angle o 1 is assigned to class B in accordance with DIN 5032-7, appendix B o IEC 13032-1 and CIE 69.It measures the perceived brightness o back-lighted suraces in candelas per square meter (cd/m) or oot-lamberts (L) in consideration o ambient light.
MAVO-SPOT 2 USB
SpecifcationsMaximum precision Classifed spot metering o luminance with an
acceptance angle o 1 in cd/m or L in accordance with class B per DIN
5032-7, appendix B o IEC 13032-1 and CIE 69.
Precision ocusing The mirror reex with a 1 measuring circle and a 15
feld o view is laid out or distance measurements rom 1 meter to infnity,
with optional close-up lenses or ocusing as o 34 cm.
Contact measurement Contact measurement can be perormed with an
optional probe and adapter disc.
Broad measuring range High initial sensitivity and a resolution o 0.01 to
99,900 cd/m or 0.01 to 30,000 L.
Comparative and ratio measurement Contrast A/B, consistent lighting %A
and deviation A-B are calculated and directly displayed.
V() adaptation The spectral sensitivity o the silicon photodiode is
color corrected and corresponds to the spectral brightness sensitivity o the
human eye V().
Simple unctions expansion Unclassifed measurement o illuminance in lxor c i s possible with the optional reectance standard.
Convenient everyday use Simple operation with one hand, easy to readviewfnder display, compact design, tripod thread, protection duringtransport in high quality aluminum case.
Non-volatile memory 1000 measured values or 10 groups o 100measured values can be saved, retrieved or transmitted to a PC via theintegrated USB port.
Computer-aided measurement The meters continuous operating modeis assured thanks to power supply via the USB port. Meter control, as wellas acquisition, display and storage o measured values, is managed withincluded GLUX 2 sotware.
EIZO RadiCS The meter is integrated into the sotware and can be usedor comprehensive testing and automatic adjustment o RadiForce monitors,in order to assure constant and consistent image reproduction.
8
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
6/9
1110
MAVOMAX
The surveillance o the ambient light extends the standards-
specifed interval or constancy testing o image display devices or veiling
luminance and maximum contrast to 6 months. In the case o repeat
measurements within the ramework o acceptance or constancy testing,
veiling luminance need not be measured again, and the 60-minute waiting
time until the image display device has reached a stable state is eliminated.
Further applications would also be the surveillance o the conditions or
dimmable light as well as or external workstations or emergency diagnostics.
MAVOMAXcompletes the application o every diagnostic monitor or light box or medical applications according to norm DIN 6856-1. It guarantees consistent lighting andperorms light surveillance or monitors in accordance with norm DIN EN IEC 61223-2-5 (QS-RL dated 20 Nov. 2003) and DIN V 6868-57.
SpecifcationsConstant lighting conditions A green LED indicates adherence to theadmissible indoor light range or diagnostics, and a red LED indicates
violation o this range.
Reduced testing eort The Surveillance o ambient light extends the
standards-specifed interval or constancy testing o image display
devices or veiling luminance and maximum contrast to 6 months.
In the case o repeat measurements, there is no need to repeat the
veiling luminance measurement.
Selectable indoor light range Two variants are available: one or general
applications MAVOMAX 60 (20 to 60 lx), and one or use in the feld o
mammography MAVOMAX RK1 (10 to 50 lx) (corresponds to RK1 per drat
standard DIN 6868-157).
Flexible power supply Electrical power or the continuous operating mode
is supplied to the meter by connecting it to a ree USB port or the included
mains power pack.
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
7/9
1312
GLUX 2 SOFTWARE
SpecifcationsFree sotware GLUX supports the MAVOLUX 5032 B/C USB, the
MAVO-MONITOR USB and the MAVO-SPOT 2 USB, and is included in the
delivery contents.
International use German, English, French or Spanish can be selected.
Clear-cut display Measured value display as list or graphic.
Expressive profle Recording o selected measured quantities at an
adjustable interval.
Simple data export Storage as universal TXT fle or data transmission
Comortable continuous operation Power supply via USB.
Open device interace Detailed description included on the CD.
Helpul bonus material Various Excel spreadsheets or data transer with
sample applications are included on the CD. These serve as a basis or the
users own applications.
GLUX 2 SotwareIntuitive GLUX sotware is the link between the meter and customer-specifc urther processing at the PC. Momentary or stored measured values can betransmitted, saved as TXT fles and read in by word processing, spreadsheet and database applications.
This allows or inclusion in the users own reports, additional calculations
and extensive archiving. GLUX is also capable o recording measured values
at an adjustable interval or the creation o light profles and or long-term
monitoring. Ater connection to a USB port, the meter is supplied with
electrical power rom the PC. The complete interace description simplifes
integration into the users own sotware applications. Excel spreadsheets
or measured value transer, or reading out memory, or acceptance and
constancy testing o medical image display devices per DIN V 6868-57, and
or measurements at workstations are included on the CD as bonus material.
CERTIFICATION QM
Reliable Measured Values through Calibration at Regular IntervalsDIN EN ISO 9001-9004 demands test equipment monitoring, i this equipment is relevant or product quality or is used or the preparation o assessments.This test equipment must be calibrated at regular intervals and retraceable back to national standards. Calibration means to state and document the
deviation o the meter display to an illuminant which is normed by and retraceable to a National Standard. This must be done in a specifc procedure.
Depending upon how the respective device is used, a calibration interval o between 12 and 24 months is recommended. A combination including receiving
report, adjustment and fnal report is also possible or GOSSEN meters.
Strictest Standards or the GOSSEN Light LabThe GOSSEN Light Lab is equipped with a tested and monitored optical
bench, whose traceability to the national standard maintained by the PTB
(Highest technical authority under the auspices o the Federal Ministry o
Economics and Technology) is guaranteed by means o a Wi41G standard
lamp. The lab is subject to test equipment monitoring in accordance with
DIN EN ISO 9001-9004, and is additionally accredited or illuminance
by DAkks in accordance with DIN EN ISO/IEC 17025 under registrationnumber D-K-15080-01-00. Thereore the reliability o the product quality,
the competence o all personnel involved, continuous external auditing and
international recognition o the calibration service is a given act. GOSSEN
oers two dierent calibration certifcates.
Factory Calibration CertifcateThe actory calibration certifcate comprises testing and documentation o 14
dierent luminosities over the entire measuring range o 0.00 to 20,000 lx
with a measuring uncertainty o 3 %. Date o calibration, the devices serial
number, reerence conditions, identifcation o utilized test equipment and
admissible deviations are listed as well in accordance with the standards.
DAkkS Calibration CertifcateThe DAkkS calibration certifcate comprises testing and documentation
o luminosities including 10, 180 and 1800 lx within the DAkkS certifed
range o 1.75 to 2000 lx with a measuring uncertainty o 1.5 %. Type o
calibration, calibration device, calibration method, measurement conditions,
measurement results and measurement uncertainty are described in detail
as well. Calibration is limited to luxmeters which comply with at least class C
in accordance with DIN 5032 or DIN EN 13032.
Calibration o Devices romOther ManuacturersAter determining that devices rom other manuacturers are capable
o being calibrated, we are pleased to issue either a DAkkS or a actory
calibration certifcate. I the device does not comply with at l east class C, only
actory calibration can be oered.
Devices rom other manuacturers cannot be adjusted.
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
8/9
1614 15
MAVO-SPOT 2 USB
PrecisionLuminance Meter
Class B DIN5032-7/EN13032-1,appendixB
M508G
0.1 99,900lx
Withoptional reectance standard,classifed
0.01cd/m 99,990cd/m / 0.01L 30,000L
4
Automatic/ manual
Distancemeasurem.with mirrorreex optics,acceptance angle
o1, rom1 meterto infnity,contactmeasurement(with probe)
Siliconphotodiode withV () flter
Yes
1000measuredvalues or10groups o100measuredvalues
< 3%
8%
Multiunctional LCD
Yes
4keys,1 sliderswitch,1 DIPswitch
USB 2.0
GLUX 2
2ea.1.5 V mignon,type AA
Yes
30sec.
Approx.5000 measurements
USB
0C to 50C
1 9 0 m m x9 0 m m x5 7 m m
400g withoutbattery
Factorycertifcate H997B
Aluminum case,GLUX 2sotware,eyecup, lens cover,
USB cable,battery,operatinginstructions
MAVO-MONITOR USB
PrecisionLuminance Meter
Class B DIN5032-7/EN13032-1,appendixB
M504G
0.01cd/m 19,990cd/m / 0.001L 1,999L
4
Automatic/ manual
2persecond
Contactmeasurement
Siliconphotodiode withV () flter
Yes
1.5m, plug-in
100measured values
< 3%
8%
3 digitLCD
Yes
6keys
USB 1.1
GLUX 2
1,5V mignon,type AA
Yes
4min. /continuous operation
Approx.45 hours withalkaline manganese battery
USB
0C to 50C
65mm x120 mm x19mm (meter)
31mm x105 mm x30mm (probe)
265g withoutbattery
Factorycertifcate H997B
Aluminum case,adapterdisc,GLUX 2 sotware,
USB cable,battery,operatinginstructions
Model MAVOLUX 5032 B USB MAVOLUX 5032 C USB
Type P r ec i si onLuxm eter P r ec i si onLuxm eter
Classifcation Class BDIN5032-7/ EN13032-1,appendixB Class C DIN5032-7/EN13032-1,appendixB
Item number M 50 3N M 50 2N
Illuminance 0 .01 l x 1 9 9 ,9 0 0 l x/ 0 .0 01 c 1 9 ,9 9 0 c 0 .1 l x 1 9 9 ,9 0 0 l x/ 0 .0 1 c1 9 ,99 0 c
Luminance 0.1cd/m 1,999,000cd/m /0.01L 199,900L 1cd/m 1,999,000cd/m /0.1 L 199,900LWithoptional luminance attachment,notclassifed Withoptional luminance attachment,notclassifed
Measuringranges 5 4
Measuringrange selection Automatic/ manual Automatic/ manual
Measuringrate 2 pe rsecond 2 pe rsecond
Measuringmethod Distancemeasurement, Distancemeasurement,
Contactmeasurementwithoptional luminance attachment Contactmeasurementwithoptional luminance attachment
Measuringsensor Siliconphotodiode withV () flter Siliconphotodiode withV () flter
Probe withtripod thread Yes Yes
Measurementcable 1 .5 m , p lu g- in 1 .5 m , f rm ly c o nn ec te d
Measuredvalue memory 100measured values 100measured values
Errorl imit-V() adapted(1), typical < 3% < 7,5%
Errorlimit- overall error,typical 8% 15%
Display 3 1 /2 dig it LC D 3 1 /2 d ig it LC D
Back-Light Yes
Operatingelements 6keys 6keys
Interace U SB 1 .1 U SB 1 .1
Sotware G LU X 2 G LU X 2
Battery 1,5V mignon,type AA 1,5V mignon,type AA
Automaticbatterycontrol Yes Yes
Automaticshutdown 4 m i n . / cont i nuous oper ati on 4 m i n . / cont i nuous oper ati on
Batteryservice lie Approx.45 hours withalkaline manganese battery Approx.45 hours withalkaline manganese battery
Powersupply USB USB
Operatingtemperature 0C to 50C 0C to 50C
Dimensions 65mm x120 mm x19 mm (meter) 65mm x120 mm x19 mm (meter)
31mm x105 mm x30 mm (probe) 31mm x105 mm x30 mm (probe)
Weight 2 0 0 gw i thoutbat ter y 2 0 0 gw i thoutbat ter y
Certifcates Fact orycer ti fcat e - H9 9 7 B Fact orycer ti fcat e - H9 9 7 B
DAkkScertifcate -H997D DAkkScertifcate -H997D
Deliverycontents Aluminum case,GLUX 2sotware Aluminum case,GLUX 2sotware
USB cable,battery,operating instructions USB cable,battery,operating instructions
TECHNICAL DATA
Measuringfunctions
Operation
Powersupply
Miscellaneous
COMPONENT PART
MAVOLUX 5032 B USB MAVOLUX 5032 C USB MAVO-SPOT 2 USB
Luminance attachment M516G
Adapterdisc M499G
Measurementcables withspecial lengths:
3m -15146
5m -15147
10m -15148
Luminance attachment M516G
Adapterdisc M499G
Measurementcables withspecial lengths:
3m -15143
5m -15144
10m -15145
Probe orcontactmeasurement M511G
Close-uplens 1(51to 100cm) M496G
Close-uplens 2(34to 50cm) M497G
Stray-lightbae M513G
Carryingstrap M514G
MAVOMAX RK1 / 60
IndoorLight Surveillance Meter
M517G/M518G
1 0 l xt o 5 0 l x/ 2 0 l xt o 6 0 l x
1
Siliconphotodiode withV () flter
2LEDs
Bypermanently connectedUSB cable
0C to 50C
4 0 m m x3 3 m m x2 3 m m
1 5 0 g
Powerpack withUSB socket
90to 240V (50to 60Hz),operatinginstructions
Reectance standardor luxmeasurement M512G
7/31/2019 120403_GOSSEN_Lichtmesstechnikbroschre_E_FINAL05
9/91817